Professional Documents
Culture Documents
3/28/18
Objective
The objective of this lab was to extract, isolate, and purify aldolase from rabbit
Methods
All of the procedures were performed at 4°C and sample aliquots from each fraction
Extraction
Ground rabbit muscle (80g) was hand stirred for 20 minutes with 150 ml of cold
0.03 M KOH and 0.001 M EDTA. The mixture was then centrifuged at 10,000 x g for
10 minutes and then the supernatant solution was decanted. The supernatant was
then filtered through glass wool to remove lipids and the pH of the filtrate was
adjusted to 7.5 with cold 0.1 N KOH. After the volume of the solution was recorded, a
1 ml aliquot of this first solution, Fraction I, was removed for later analysis.
Salt fractionation
An equal volume of 100% saturated ammonium sulfate was then slowly added to
Fraction I to yield a 50% saturated ammonium sulfate solution. After being slowly
stirred for an hour, Fraction I was centrifuged at 10,000 x g for 30 minutes. Its
supernatant was then filtered through glass wool and the volume of this solution,
Fraction II, was recorded. Again, a 1ml aliquot of Fraction II was removed for later
analysis. Ammonium sulfate (65g per liter of Fraction II) was gradually added with
slow magnetic stirring to bring the solution to 60% ammonium sulfate saturation.
The pH of this 60% ammonium sulfate saturated solution was then recorded and
Dialysis
Fraction II was centrifuged at 10,000 x g for 30 minutes and then the supernatant,
Fraction III, was removed to have its volume recorded as well as pH. A 1 ml aliquot
of Fraction III was saved for later analysis. The precipitate from the centrifugation of
Fraction II was dissolved in as little cold equilibrium buffer (10 mM Tris HCl, pH 7.5;
5mM EDTA) as possible. It was then centrifuged at 10,000 x g for 15 minutes. The
volume of the supernatant, Fraction IV, was recorded and a 0.1 ml aliquot diluted
with equilibration buffer to 1 ml was saved for later analysis. Dialysis tubing
(MWCO 14,000Da) that was already soaked in distilled water was filled with
The column (1.5 cm x 11 cm, 20 ml total volume) was set up by packing the
letting 100 ml of equilibration buffer and obtaining a 1 ml/min effluent flow rate,
Fraction V was added to the column and washed with equilibrium buffer ((2.5 mM
FBP [B-grade]: 50 mM Tris HCl, pH 7.5; 5 mM EDTA) until the effluent absorbance
at 280 nm fell below 0.1. The liquid following this change in absorbance was labeled
the “wash” which the volume was recorded and a 1 ml aliquot was saved for future
analysis. After the “wash” was eluted, 100ml of substrate elution buffer was used to
elute the aldolase. There were 50 1 ml fractions were collected and assayed by
reading A280. A 0.3 ml aliquot was removed from the fraction tube with the highest
absorbance reading, the “peak” Fraction. Another 0.1 ml aliquot was removed and
diluted to 1 ml with equilibration buffer. Both fractions were saved for future
analysis. Then all of the fractions that were part of the peak were pooled together
and called Fraction VIA. The same was done to the less concentrated fractions that
were part of the “tail” end of the peak and this was called Fraction VIB. The volumes
of both VIA and VIB were recorded and 0.1 ml aliquots from each were diluted to 1
Protein determination
A standard curve was created using 7 increments of bovine serum albumin from 1
to 25 μg. 0.15 M NaCl was added to adjust each final volume to 1 ml. The saved
fraction aliquots were each diluted with 0.15 M NaCl to a final volume of 1 ml and 1
ml of Coomasie Brilliant Blue was added. The absorbances were read at 595 nm
Aldolase assay
A reaction mixture with a final volume of 3 ml was used for every saved fraction
aliquot to determine absorbance rate when FBP is added. Each reaction mixture
A reaction mixture with a final volume of 3 ml was used for every saved fraction
aliquot to determine absorbance rate when Ga3P is added. Each reaction mixture
αGP dehydrogenase. Each sample also got 10 μl of each fraction and the absorbance
Results
Extraction
After being mixed with 0.03 M KOH and 0.001 M EDTA the resulting supernatant
contained the extracted proteins including aldolase. Fraction I was collected and the
final volume of Fraction I was 159 ml and had a concentration of 28.004 mg/ml.
Salt Fractionation
The following calculation was used to determine the amount of solid ammonium
The final volume of Fraction II was 311 ml and had a concentration of 1.60134
mg/ml.
Dialysis
Fractions III and IV were collected while preparing for dialysis. The final volume of
Fraction III was 320 ml and the pH was 7.28. The concentration of Fraction III was
6.441 mg/ml. The final volume of Fraction IV was 5.4 ml and the concentration was
72.106 mg/ml.
Fraction V, “wash”, “peak”, VIA, and VIA were all collected during the column
concentration of the saved aliquot was 0.1 M. The final volume of the “wash” was 14
m and the concentration was 0.26536 mg/ml. The final volume of the “peak” was 2
m and the concentration was 14.618 mg/ml. The final volume of both the VIA and
VIB fractions was 10 ml and the concentrations were 8.464 mg/ml and 1.1438
mg/ml respectively.
Protein determination
The Bradford Method was used to determine protein concentration. In order to use
this method a standard curve had to be created. Each fraction then compared its
standard curve. This means that absorbance and protein concentration have a
positive correlation.
Figure 1: Standard Curve for Bradford Method
0.5
0.4
0.3
0.2
0.1
0
0 5 10 15 20 25 30
-0.1
Concentration ( mol/L)
The protein content seemed to increase from Fractions I to IV, which made sense
since during the processes that they were collected, the solution was supposed to be
isolated and purified from any other components. The “wash” was found to have the
lowest protein content because it was supposed to be the portion before the
aldolase was eluted out of the column. The “peak” was found to have one of the
highest protein contents because it contained the fraction with the highest
Similar assays were set up, one with FBP mirroring the normal aldolase reaction,
and the other with TPI. Absorbance rates were recorded for each reaction in each
assay and then used to calculate enzyme activity, total protein, and specific activity.
Generally both enzyme activity and specific activity both increased with the
fractions.
some outliers including Fraction II. Reaction velocity did not seem to have a specific
10
6
Assay II
4 TPI
2
0
I II III IV V Wash Peak VIA VIB
Fraction
Figure 2. Using both the results from both the Bradford Method and the
Aldolase assays, total enzyme activity and total protein were able to be calculated.
These values were then used to calculate specific activity for each fraction.
Sample Calculations
Total Enzyme Activity (Units) = (ΔABS/min x 0.48 x dilution factor x total fraction
Discussion
First the aldolase had to be extracted from the rabbit muscle cells. To do this KOH
was used to lyse the cells to release the enzyme. Ammonium sulfate was then used
allowing the aldolase to precipitate out. To then get rid of any remaining salt, a
dialysis tubing of MWCO 14,000Da was used. This was because aldolase molecules
are bigger and therefore the tubing would allow any smaller molecules to be
separated out of the solution. During column chromatography, the aldolase was
bound to the phosphocellulose resin. The equilibrium buffer did not contain any
substrates of aldolase and therefore only washed out all other proteins. The elution
buffer was able to elute the aldolase because it contained FBP in which the aldolase
makes sense because the aldolase is supposed to be more purified in each fraction
meaning there should be less background proteins. Fraction III has a higher amount
of protein that doesn’t fit the trend. This fraction was the supernatant collected after
supernatant was supposed to contain all of the other proteins that did not have
water-protein interactions and therefore were not precipitated out. The total
amount of protein for this fraction could have been higher because of the large
amount of proteins that were separated from the aldolase as well as the solution not
being uniformly mixed and the saved aliquot being more concentrated than the
whole volume.
The affinity chromatography was the most efficient step in increasing specific
activity. In the aldolase assay the VIA fraction had one of the highest specific
activities. This fraction contained the chromatography fractions that surrounded the
peak fraction indicating that this specific activity was due to the affinity
chromatography. Fractions I and III have the lowest specific activity and the wash
should also be expected to have a low specific activity. Both the wash and Fraction
III are expected to contain proteins that were separated from the aldolase so it
makes sense that they would have a low activity in a reaction for aldolase.
The wash step/wash fraction should contain all background proteins that did not
bind to the phosphocellulose resin. The lack of aldolase would cause the specific
activity to be low in the aldolase assay while the presence of many other proteins
In the TPI assay Fraction III also had a relatively high specific activity. This could be
for the same reasoning as in why the wash fraction should have a higher specific
activity. The VIA fraction had one of the lowest specific activities, which makes
sense because it is supposed to be the most purified. The same goes for the peak
fraction.
When comparing this experiment to that of Penhoet and Rutter, they must have
used much larger amounts of aldolase since all of their total protein values are much
larger. Although those values are larger, it seemed as though the specific activities
were smaller. The increase from Fraction I to Fraction II was much larger in this
experiment that that of Penoet and Rutter. This also could have been because of a
calculation error seeing that the values from Fraction II are abnormally high. The
experiments were similar in finding that the phosphocellulose resin column created
Purification could have been increased by repeating any of the processes to ensure
that each step was completely successful. There was also an issue with not all of the
total fraction volumes being recorded, which then caused a problem with calculating
References
com.ezaccess.libraries.psu.edu/science/article/pii/0076687975421218
2.