You are on page 1of 6

GE Medical Systems – Americas: 888-202-5528

P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.


GE Ultrasound - Europe: +49-212-28020
P.O. Box 1105 60 D-42665 Solingen
GE Ultrasound – Asia +86- 21-52080200
Shanghai Maxdo Center, No. 8 Xing Yi Road,
Shanghai, China 200336
http://www.gemedicalsystems.com

LOGIQ 9
20-November-2002
TruScan Imaging Technology

• Integrated locking mechanism that


Product Description provides rolling lock and caster
The LOGIQ 9 is our leadership swivel lock
• Integrated cable management
imaging system designed for • Front and rear handles
abdominal, vascular, obstetrics, • Easily removable air filter
gynecology, neonatal, urology,
transcranial, and small parts
applications. User Interface
System Architecture Operator Keyboard
• Floating keyboard adjustable in three
TruScan Architecture – our dimensions:
exclusive, software-intensive, • Height: adjustable 76.2 – 96.5 cm
ultrasound imaging platform gives you (30 - 38 in)
unsurpassed computational power, • Rotation: adjustable +/-75o from
image-manipulation capability, center
workflow flexibility and product • Extension: 31.8 cm (12.5 in) from
upgrade-ability. console
• Full-sized, backlit alphanumeric
TruAccess is the new, GE-exclusive, keyboard
raw-data processing technology that • Ergonomic hard key layout
will change the future of ultrasound • Interactive back-lighting
imaging. By accessing raw data, • Integrated recording keys for remote
TruAccess achieves excellent image control of up to 4 peripheral or
quality today and ensures unsurpassed DICOM devices
• Height - Min: 140.3 cm (55.25 in) - • Integrated gel warmer
image management tomorrow. Max: 160.7 cm (63.25 in)
CodeScan, Coded Technology, • Width: 64 cm (25.2 in) Touch Screen
improves image quality and penetration • Depth: 89.9 cm (35.4 in) • 10.4 in High Resolution color LCD
• Weight (no peripherals): 196.8 kg
while Multi-Level Codes provides (434 lb)
screen
greater bandwidth control. • Interactive dynamic software menu
Electrical Power • Brightness adjustment
SmartScan improves image • User-configurable layout
• Voltage:100-120 Vac or 220-240 Vac
acquisition and patient throughput • Frequency: 50/60 Hz
while increasing diagnostic confidence Monitor
• Power: Max. 1.2 KVA with on-board
and exam consistency. peripherals • 17′′ High-Resolution non-interlaced
• Maximum Thermal Output: 4095 scan flat CRT
ComfortScan, our most advanced BTU/hr. • Center of monitor height adjustable
ergonomic design ever, helps maximize (with keyboard): 125.7 – 146.1 cm
productivity and simplify every exam Console Design (49.5 – 57.5 in)
you perform. • 4 Active Probe Ports • Tilt/Rotate Adjustable Monitor
• Integrated HDD (40 GB) • Tilt Angle: Down 10°, Up 10°
• Integrated CD-RW drive • Rotate Angle: 90° right, 90° left
General Specifications •

Integrated MO drive • Wide Image area
On-board storage for peripherals • Integrated high-fidelity speakers
Dimensions and Weight • Wheels • Digital brightness & contrast
(Dimensions given with floating • Wheel diameter: adjustment.
keyboard stowed for transport) Front: 175 mm
Rear: 175 mm

LOGIQ 9 Product Data Sheet 2350082GSP Rev. 3 Page 1 of 6


System Options • Hospital Name: 23 Characters.
System Overview • B-Flow • Date: 2 Types selectable
Applications • LOGIQView MM/DD/YY,
• Abdominal • Coded Contrast Imaging DD/MM/YY
• Obstetrical • Advanced 3D • Time: 2 types selectable
• Gynecological • Tru3D – sensor-based and 24 hours, 12 hours
• Breast quantitative • Gestational Age from
• Small parts • DICOM 3.0 connectivity LMP/EDD/GA/BBT
• Vascular / Intraoperative/ Peripheral • Foot Switch, with programmable • Probe Name
• Transcranial functionality • Gray Map names
• Pediatric and Neonatal • Probe Orientation
• Musculoskeletal • Console Protective Cover
• Depth Scale Marker
• Urological Peripheral Options • Lateral Scale Marker
Operating Modes • Integrated mounting kits provided • Focal Zone Markers
• B-Mode for: • Image Depth
• M-Mode • Analog B&W thermal printer • Zoom Depth
• Color Flow Mode (CFM) • Analog A6 or A5 color thermal • B-Mode
• Power Doppler Imaging (PDI) with printer • Gain
topographic and directional maps • Digital B&W thermal printer • Dynamic Range
• PW Doppler with high PRF • Digital A6 color thermal printer • Imaging Frequency
• B-Flow Mode • Analog S-VHS VCR • Edge Enhance
• M-Color Flow Mode • External Color PC desktop printer & • Frame Averaging
connection kit • Gray Map
Scanning Methods • External video & audio connections • M-Mode
• Electronic Sector provided for other devices such as • Gain
• Electronic Convex Multi-format cameras, laser cameras, • Dynamic Range
• Electronic Linear and other peripherals • Time Scale
• Doppler Mode
Transducer Types Display Modes • Gain
• Sector Phased Array • Simultaneous Capability
• • Angle
Convex Array • B/PW • Sample Volume Depth and Width
• Microconvex Array • B/CFM or PDI
• Linear Array • Wall Filter
• B/M • Velocity and/or Frequency Scale
• Matrix Array
• Real-time Triplex Mode • Spectrum Inversion
System Standard Features (B + CFM or PDI/PW) • Time Scale
• State-of-the-art user interface with • B-Flow + PW • PRF
high resolution 10.4 inch LCD touch • Selectable alternating Modes • HPRF
panel • B/PW • Doppler Frequency
• Automatic Tissue Optimization • B + CFM (PDI)/PW • Color Flow Doppler Mode
• Fine Angle Steer • B-Flow/PW • Line Density
• Coded Excitation • Multi-image (split/quad screen) • Frame Averaging
• Coded Harmonic Imaging • Live and/or frozen • Packet Size
• Virtual Convex • B + B/CFM or PDI • Color Scale: 3 types
• Patient information database • PW/M Power, Directional PDI, and
• Image Archive on CD, MOD, and • Independent Cine playback Symmetrical Velocity Imaging
hard drive • Zoom Read/Write/Pan • Color Velocity Range and Baseline
• Integrated MO drive (supports 128, • Colorized Image • Color Threshold Marker
230, 540, 640 MB or 1.3 GB disks) • Colorized B • Color Gain
• Easy 3D (Freehand acquisition with • Colorized M • PDI
basic features) with 3D Movie • Colorized PW • Spectrum Inversion
• Raw Data Analysis • Time line display • Doppler Frequency
• Independent Dual B/PW Display • TGC Curve
• Real-time automatic Doppler calcs
• Display Formats • Acoustic Frame Rate
• OB Calcs
• Fetal Trending
Top/ Bottom selectable format (Size: • Cine Frame Number
1/2:1/2; 1/3:2/3; 2/3:1/3) • VCR Counter
• Multigestational Calcs
Side/Side selectable format • VCR Status
• Hip Dysplasia Calcs (1/2:1/2; 1/3:2/3; 0:1) • VCR Playback Counter
• Gynecological Calcs • Virtual Convex • Body Pattern: Up to 90 human types
• Vascular Calcs plus 13 animal types
• Urological Calcs Display Annotation • Application Name
• Renal Calcs • Patient Name: First, Last, & Middle • Measurement Results
• Insite capability name each store 27 characters. Up to • Operator Message
• iLinq capability 27 total characters displayed. • Displayed Acoustic Output
• On-board electronic documentation • Patient ID: 31 Characters. Up to 27 • TIS: Thermal Index Soft Tissue
total characters displayed. • TIC: Thermal Index Cranial
• Age, Sex and Birth Date (optional) (Bone)
• TIB: Thermal Index Bone

LOGIQ 9 Product Data Sheet 2350082GSP Rev. 3 Page 2 of 6


• MI: Mechanical Index Post-Processing • Scrolling timeline memory
• % of Power output • Max Read Zoom to 8x
• Biopsy Guide Line and Zone • Base Line Shift Image Storage
• Sweep Speed • On-board database of patient
• B/M-Mode information from past exams
• Storage Format: DICOM
System Parameters • Gray Map
• Image Storage Size:
• Colorized B and M
System Setup • Post Gain − Gray Image: ~300K to ~1.3 MB
• Pre-programmable Categories • Compression − Color Image: ~900K to 1.9 MB
• User Programmable Preset Capability • Anatomical M Mode • Multiframe
• Factory Default Preset Data • PW Mode • Display Format: Full size, 4x3, and
• Languages: English, French, German, • Gray Map "thumbnails"
Spanish, Italian, Portuguese, Russian, • Post Gain • Live image and stored image side-by-
Greek, Swedish, Danish, Dutch, • Compression side display
• Rejection • MO Disk Storage: 128, 230, 540, 640
Finnish, Norwegian
• Colorized D
• OB Report Format: 5 Types, Tokyo • Color Tag
MB or 1.3 GB
Univ., Osaka Univ., USA, Europe, • CD-RW storage
• Display Format
and ASUM • Conversion to JPEG and AVI file
• Color Flow
• EFBW: 8 types, Tokyo • Color Map
formats
University, Osaka University, USA • CFM Velocity Tag • Hard Drive Image Storage: 18.1 GB
and Europe (Shephard, Merz, • CFM Display Threshold
Hadlock/Shephard, Williams, Connectivity
• Angle Correct (PW mode) • Ethernet network connection
Brenner) • Quick Angle Correct (PW mode)
• • RS-232 serial data output
Up to 90 Programmable Annotations • Auto Angle Correct (PW mode)
• • DICOM support (option)
Body Patterns • Spectral Invert for Color and
• Customized Comment Home Position • Verify
Doppler
• Print
Pre-Processing • Store
• Write Zoom up to 4x Image Processing and • Modality Worklist
• B/M-Mode
• Gain
Presentation • Storage Commitment
• Modality Performed Procedure
• TGC • Digital Beamformer Step (MPPS)
• Dynamic Range • Media Exchange
• 1024 system processing channel
• Acoustic Output • Off network / mobile storage queue
technology
• Transmission Focus Position
• Displayed Imaging Depth: 0 – 36 cm Raw Data Processing
• Transmission Focus Number
• Minimum Depth of Field: 0 – 2 cm • Complete image flexibility allows
• Edge Enhancement
(Zoom) (probe dependent)
• Smoothing Control stored image optimization and
• Line Density Control • Maximum Depth of Field: 0 – 36 cm measurements. :
• Sweep Speed for M-Mode (probe dependent) • Overall B-Mode gain, dynamic
• PW-Mode • Transmission Focus range and gray scale maps
• Gain • 1- 8 Focus Points selectable (probe • Overall Doppler gain, baseline
• Dynamic Range and application dependent)
shift, sweep speed and inverted
• Acoustic Output • Focal Zone position, up to 12 steps
• Continuous Dynamic Receive Focus / spectral waveform.
• Transmission Frequency • Tru3D reconstruction from a stored
• PRF Continuous Dynamic Receive
Aperture CINE loop
• Wall Filter • Anatomical M-Mode
• Spectral Averaging • 256 shades of gray
• Up to 145 dB Dynamic • All Post-Processing capabilities
• Sample Volume Gate
Length Range/Displayed is 102 dB
Depth • Adjustable Field of View (FOV)
• Velocity Scale • Image Reverse: Right/ Left Scanning Parameters
• Color Flow Mode • Image Rotation: 4 steps,
• CFM Gain Rotation: 0°, 180° Digital B-Mode
• CFM Velocity Range • B Acoustic Power: 25 – 100 %
CINE Memory/Image Memory • B Gain: 120 dB range, 1 dB steps,
• Acoustic Output
• CINE Memory: 128 MB (includes slide pots)
• Wall Echo Filter
• Dual Image CINE Display • B Displayed Dynamic Range: 36 –
• Packet Size
• Frame Rate Control • CINE Gauge and CINE image 102 dB, 3-6 dB steps
• CFM Spatial Filter number display • B Frame Averaging: 7 steps
• CFM Frame Averaging • CINE Review Loop • B Gray Scale Map: 18 types
• CFM Line Resolution • CINE Review Speed: 9 types • B Edge Enhancement: 6 steps
• Frequency / Velocity Base Line (1/1, ½, 1/3, ¼, 1/5, 1/6, 1/7, 1/8, 1/9) • Frequency Selection
Shift • Selectable CINE Sequence for • Line Density: up to 5 steps
• CFM ACE-Filter CINE Review • Scanning Size (FOV or Angle
• Measurements/ Calculations & (Depending on the probe, see probe
Annotations on CINE Playback specifications)

LOGIQ 9 Product Data Sheet 2350082GSP Rev. 3 Page 3 of 6


• B Colorization: 4 color settings • CFM Window Size (same as B- Coded Excitation
• Image Softener: 4 steps Mode): • Available on the following probes:
• Reject: 6 steps • Convex 5° - 120° • M7C
• Suppression: 6 steps • Sector 10° - 90° • E8C
• Linear 5 mm - 46 mm • 10L
Digital M-Mode (Depending on the probe) • 10S
• M Acoustic Power: 25 – 100 % • Maximum Steerable Angle • 8C
• M Gain: 120 dB range, 1 dB steps, (Slant Scan): +/- 20 °
includes slide pots • CFM Spectrum Inversion Coded Harmonic Imaging
• M Displayed Dynamic Range: 36 – • Available on all probes
102 dB, 3-6 dB steps • CFM Baseline Shift: 11 steps
• M Gray Scale Map: 18 types • Pre-settable and independently
Coded Contrast Imaging (option
• M Edge Enhancement: 6 step- adjustable B-Mode Gain in B/CFM-
• M Sweep Speed: 8 types Mode available only outside the United
• M Colorization: 4 colors • CFM Frame Average: 7 steps States.)
• CFM/M-Mode capability • CFM Threshold: 0 - 100% • Available on 3.5C, 4S, M7C, 7L,
• Realtime Triplex Mode: 10L, & M12L probes
Anatomical M-Mode B + CFM/PW in any depth, at any • Contrast Timer
• M-mode cursor adjustable at any PRF • Timed Updates: 0.5 – 10 seconds
plane • 3 fundamental frequencies on 3.5C,
• Can be activated from a CINE loop • PRF setting for PW/CFM in Triplex
from a live or stored image Mode with user selectable ratio of 4S, M7C, 10L, and M12L
• M & A capability 1:1, 1:2, or 1:4 • 2 fundamental frequencies on 7L
• Not available for Linear transducers • Up to 4 B-mode harmonic settings:
Digital Power Doppler Imaging • Coded Harmonic Angio
Digital Spectral Doppler Mode • Power Doppler Maps: 10 maps • Coded Phase Inversion
• PW Acoustic Power: 0 – 100 % • 1 Directional map • Hi MI Harmonic
• PW Gain: 72 dB range, 1 dB step • Power Doppler Gain: 0-80 dB in • Coded Phase Inversion
• PW Displayed Dynamic Range: 24 – 1 dB steps • TruAgent Detection (3.5C, 4S)
40 dB, 4 dB step • PDI Frame Averaging: 7 steps • Multiple selections for CHA
• PW Gray Scale Map: 5 types • PDI Color Threshold: 0 - 100% tissue background
• PW Wall Filter: up to 10 steps, PRF • Wall Filter: 3 steps • Maximum Enhance Mode
(velocity scale) dependent • Simultaneous Real-time Triplex • The LOGIQ 9 is designed for
• PW Colorization: 4 steps Mode compatibility with commercially
• Color Tag B+PDI/PW in any depth, at any PRF available ultrasound contrast
• PRF: 1.0 – 30.0 kHz • PRF setting for PW/CFM in Triplex agents. Because the availability
• Velocity Scale Range: Mode with user selectable ratio of of these agents is subject to
• PD/HPRF +/- 5 m/sec (0°), 1:1, 1:2, or 1:4 government regulation and
+/- 10 m/sec (60°) • 8, 10, 12, 14 or 16 packets approval, product features
(Depending on the probe etc.) • Line Density: 3 steps intended for use with these agents
• Digital Wall Filter Frequency: 20 Hz
Auto Optimization may not be commercially
to 1800 Hz
• PW Sweep Speed: 8 types • Available in:
marketed nor made available
• Sample Volume Length: 1 – 16 mm • B-Mode before the contrast agent is
• Angle Correction: ± 0-90°, 1° step • B-Flow cleared for use. Contrast related
Available before Freeze and after • PW Doppler product features are enabled only
Freeze on systems for delivery to an
• Steered Linear: 0° - 20° B-Flow (option) authorized country or region of
(Depending on the probe) • Available on the following probes: use. GE Medical Systems makes
• Spectrum Inversion • 7L no claims concerning the safety
• Baseline Shift: 9 steps • 10L or effectiveness of contrast
• Doppler Auto Trace • i12L agents.
• Time Resolution: 2 to 25 msec, • M12L
typically 10 msec • M7C LOGIQView
• 3.5C • For use in B-Mode
Digital Color Flow Mode • E8C • LOGIQView Status
• Symmetrical Velocity Imaging • 8C • Auto detection of scan direction
for optimized 3D color images • Image Reverse: left/right • Pre or post-process zoom up to
• Color Maps: 8 velocity maps; • B-Flow sensitivity: 5 steps 10X
4 velocity-variance maps • Background ON/OFF • Rotation
• CFM Gain: 80 dB range, 1 dB steps • Displayed Dynamic Range: 36 – 102 • Auto best fit on monitor
• CFM Velocity Scale Range: dB • Measurements in B-Mode
- less than +/- 0.2 cm/s • Line Density: 3 steps • Up to 60 cm scan length
- max: +/- 3.0 m/s • Frame Averaging: 7 steps
• Wall Filter: 40 – 3500 Hz • Map: 18 selectable
3D (options)
• 8, 10, 12, 14 or 16 packets • Edge Enhance: 6 steps Advanced 3D
• Line Density: 3 steps • Flow Settings: Low / High • Acquisition of Color data
• CFM Spatial Filter: 6 selections • Automatic rendering

LOGIQ 9 Product Data Sheet 2350082GSP Rev. 3 Page 4 of 6


• 3D Landscape technology Real-time Doppler Auto Shephard, Hadlock, Hohler,
• 3D Movie Measurements / Calculations Campbell
• Main Mode • PS (Peak Systole) • Fetal Graphical Trending
• ED (End Diastole) • Growth Percentiles
Tru3D • MD (Minimum Diastole) • Multi-Gestational Calculations (4)
• Sensor-based acquisition • PI (Pulsatility Index) • Fetal Qualitative Description
• Display of data in: Main-, Parallel-, • RI (Resistivity Index) (Anatomical survey)
Angular-Mode • AT (Acceleration Time) • Fetal Environmental Description
• Render Modes: Gray Surface, • ACC (Acceleration) (Biophysical profile)
Texture, Min-, Max-, Average- • PS/ED (PS/ED Ratio) • Programmable OB Tables
Intensity • ED/PS (ED/PS Ratio) • Over 20 selectable OB Calcs
• Basic measurements: distance, angle, • HR (Heart Rate) • Expanded Worksheets
area • TAMAX (Time Averaged Maximum
• Volume measurements GYN Measurements/
Velocity)
• 3D Movie • PVAL (Peak Velocity Value) Calculations
• Volume Flow (TAMEAN and Vessel • Right Ovary Length, Width, Height
Area) • Left Ovary Length, Width, Height
Virtual Convex • Uterus Length, Width, Height
• Provides a convex field of view for OB Measurements / Calculations • Ovarian Volume
all linear and sector transducers • Gestational Age by: • ENDO (Endometrial thickness)
• M12L − GS (Gestational Sac) • Ovarian RI
• 10L • Uterine RI
− CRL (Crown Rump Length)
• 7L • Follicular measurements
• i12L − FL (Femur Length)
• Summary Reports
• 4S − BPD (Biparietal Diameter)
• 10S − AC (Abdominal Circumference) Vascular
− HC (Head Circumference) Measurements/Calculations
− APTD x TTD (Anterior/Posterior • SYS DCCA (Systolic Distal
Measurements / Trunk Diameter by Transverse Common Carotid Artery)
Trunk Diameter) • DIAS DCCA (Diastolic Distal
Calculations − LV (Length of Vertebra) Common Carotid Artery)
− FTA (Fetal Trunk Cross-sectional • SYS MCCA (Systolic Mid Common
General B-Mode Carotid Artery)
Area)
• Depth & Distance • DIAS MCCA (Diastolic Mid
• Circumference (Ellipse / Trace) − HL (Humerus Length)
− BD (Binocular Distance) Common Carotid Artery)
• Area (Ellipse / Trace) • SYS PCCA (Systolic Proximal
• Volume (Ellipsoid) − FT (Foot Length)
Common Carotid Artery)
• % Stenosis (Area or Diameter) − OFD (Occipital Frontal Diameter) • DIAS PCCA (Diastolic Proximal
• Angle between two lines − TAD (Transverse Abdominal Common Carotid Artery)
Diameter) • SYS DICA (Systolic Distal Internal
General M-Mode − TCD (Transverse Cerebellum Carotid Artery)
• M-Depth Diameter) • DIAS DICA (Systolic Distal Internal
• Distance
− THD (Thorax Transverse Carotid Artery)
• Time
Diameter) • SYS MICA (Systolic Mid Internal
• Slope
• Heart Rate − TIB (Tibia Length) Carotid Artery)
− ULNA (Ulna Length) • DIAS MICA (Diastolic Mid Internal
General Doppler • Estimated Fetal Weight (EFW) by: Carotid Artery)
Measurements/Calculations − AC, BPD • SYS PICA (Systolic Proximal
• Velocity − AC, BPD, FL Internal Carotid Artery)
• Time • DIAS PICA (Diastolic Proximal
− AC, BPD, FL, HC
• A/B Ratio (Velocities / Frequency Internal Carotid Artery)
− AC, FL • SYS DECA (Systolic Distal External
Ratio − AC, FL, HC
• PS (Peak Systole) Carotid Artery)
− AC, HC • DIAS DECA (Diastolic Distal
• ED (End Diastole)
• Calculations and Ratios External Carotid Artery)
• PS/ED (PS/ED Ratio)
• ED/PS (ED/PS Ratio) − FL/BPD • SYS PECA (Systolic Proximal
• AT (Acceleration Time) − FL/AC External Carotid Artery)
• ACCEL (Acceleration) − FL/HC • DIAS PECA (Diastolic Proximal
• TAMAX (Time Averaged Maximum − HC/AC External Carotid Artery)
Velocity − CI (Cephalic Index) • VERT (Systolic Vertebral Velocity)
• Volume Flow (TAMEAN and Vessel − AFI (Amniotic Fluid Index) • SUBCLAV (Systolic Subclavian
Area) • Measurements / Calculations by: Velocity)
• Heart Rate Jeanty, Merz, Tokyo University, • Summary Reports
• PI (Pulsatility Index) Mercer, Hansmann, Erickson, Hill,
• RI (Resistivity Index)

LOGIQ 9 Product Data Sheet 2350082GSP Rev. 3 Page 5 of 6


• Doppler Transmission Frequency: • M12L Matrix Array Wide Band
Probes 1.9 MHz Linear Probe
• 3.5C Wide Band Convex Probe • Harmonic Frequency: 4 MHz • Applications: Small Parts,
• Applications: Abdomen, OB/Gyn, • Biopsy Guide Available: Not Vascular, Neonatal, Pediatrics
Urology, Vascular Available • Maximum Band Width (-20dB):
• Maximum Band Width (-20dB): 5- 13 MHz
2 – 5 MHz • 10S Wide Band Sector Probe • Number of Elements: 960
• Number of Elements: 128 • Applications: Neonatal, Small • FOV: 39 mm
• Convex Radius: 40 mm Parts, Abdomen, Pediatrics • Foot Print: 45 x 10 mm
• FOV: 68° • Maximum Band Width (-20dB): • Doppler Transmission Frequency:
• Foot Print: 53 x 18 mm 4.0 – 11.0 MHz 5.6 MHz
• Doppler Transmission Frequency: • Number of Elements: 96 • Harmonic Frequency: 10.0 MHz
2.5 MHz • FOV: 90° • Biopsy Guide Available: Multi-
• Harmonic Frequency: 4 MHz • Foot Print: 14 x 11 mm Angle, Reusable
• Biopsy Guide Available: Single- • Doppler Transmission Frequency:
Angle, Reusable 5.0 MHz
• Harmonic Frequency: 10.0 MHz External Inputs and
• M7C Matrix Array Wide Band • Biopsy Guide Available: Not
Convex Probe
• Applications: Abdomen, OB/Gyn,
Available Outputs (not including on-board
peripherals)
Pediatrics • i12L Intraoperative Wide Band
• Maximum Band Width (-20dB): Linear Probe • Video Out
3 – 8 MHz • Applications: Intraoperative, Small • SVGA
• Number of Elements: 960 Parts, Vascular, Pediatrics • Composite Color
• Convex Radius: 50 mm • Maximum Band Width (-20dB): • Composite b/w
• FOV: 63° 5 – 12 MHz • S-Video
• Foot Print: 55 x 18 mm • Number of Elements: 96
• Doppler Transmission Frequency: • Audio Stereo Out
• FOV: 25 mm • Remote Expose (2)
3.1 MHz • Foot Print: 29 x 10 mm
• Harmonic Frequency: 7 MHz • 3 Position Footswitch
• Doppler Transmission Frequency: • RS-232 serial port for Line Printer
• Biopsy Guide Available: Multi- 5.0 MHz
angle, reusable • Ethernet
• Harmonic Frequency: 10.0 MHz • External microphone
• Biopsy Guide Available: Not • InsiteTM Modem Connection
• E8C Wide Band Microconvex Probe Available
• Applications: OB/Gyn • iLinq Modem Connection
• Maximum Band Width (-20dB): • 7L Wide Band Linear Probe
4 – 11 MHz • Applications: Vascular, Small
• Number of Elements: 128 Parts Safety Conformance
• Convex Radius: 11 mm • Maximum Band Width (-20dB):
• FOV: 133° 3 – 7 MHz The LOGIQ 9 is:
• Foot Print: 23 x 10x mm • Number of Elements: 192 • Listed to UL 2601-1 by a Nationally
• Doppler Transmission Frequency: • FOV: 46 mm Recognized Test Lab
4.4 MHz • Foot Print: 52 x 10 mm • Certified to CSA 22.2, 60601.1 by an
• Harmonic Frequency: 8 MHz • Doppler Transmission Frequency: SCC accredited Test Lab
• Biopsy Guide Available: Single 3.8 MHz • CE Marked to Council Directive
Angle, Disposable • Harmonic Frequency: 7.0 MHz 93/42/EEC on Medical Devices
• Steered Angle: Max. 20° • Conforms to the following standards
• 8C Wide Band Microconvex Probe • Biopsy Guide Available: Multi- for safety:
• Applications: Neonatal, Pediatrics Angle, Reusable • EN 60601-1 Electrical medical
Maximum Band Width (-20dB): equipment
4 – 11 MHz • 10L Wide Band Linear Probe • EN 60601-1-1 Electrical medical
• Number of Elements: 128 • Applications: Vascular, Small equipment
• Convex Radius: 11 mm Parts, Neonatal, Pediatrics • EN 60601-1-2 Electromagnetic
• FOV: 133° • Maximum Band Width (-20dB): compatibility
• Foot Print: 23 x 10x mm 4 – 10 MHz • EN 60601-1-4 Programmable
• Doppler Transmission Frequency: • Number of Elements: 192 medical systems
4.4 MHz • FOV: 39 mm • ISO 10993 Biological evaluation
• Harmonic Frequency: 8 MHz • Foot Print: 44 x 10 mm of medical devices
• Doppler Transmission Frequency: • NEMA UD3 Acoustic output
• 4S Wide Band Sector Probe 4.4 MHz display (MI, TIS, TIB, TIC)
• Applications: Abdomen, OB/Gyn, • Harmonic Frequency: 9.0 MHz
TCD • Steered Angle: Max. 20°
• Maximum Band Width(-20dB): • Biopsy Guide Available: Multi-
2 – 5 MHz Angle, Reusable
• Number of Elements: 128
• FOV: 90°
• Foot Print: 32 x 19 mm

LOGIQ 9 Product Data Sheet 2350082GSP Rev. 3 Page 6 of 6