Professional Documents
Culture Documents
Test preparation
Description Quantity
Deionized water
1. Connect the electrode 2. Turn on the meter. 3. Rinse the electrode in 4. Platinum-series
to the meter. Press PHMV to select mV. deionized water and electrodes only: Press
blot dry. the dispenser button on
top of the electrode until
the electrolyte get is visible
at the reference junction.
2. Substitute the electrode potential value (C) in the equation and solve for ESHE:
E SHE = E O + C
Where:
ESHE = oxidation reduction potential of the sample relative to the SHE, following the
international sign convention.
The Reference electrode potentials table shows the potentials, C, developed by the reference
electrode relative to the standard hydrogen electrode at various temperatures.
Interferences
Many factors limit the interpretation of ORP measurements in water. These factors include
irreversible reactions, electrode poisoning, the presence of multiple redox couples, very small
exchange currents and inert redox couples. ORP measurements in the field correlate poorly with
ORP values calculated from the redox couples present. Due to these factors, the interpretation of
ORP measurements will be specific to your particular application.
Accuracy check
Checking the electrode is necessary only when there is evidence of malfunction that cannot be
traced to other causes.
Procedure A
1. Open an ampule of Light's Solution or ORP verification solution*. Pour the contents of the
ampule into a beaker.
Procedure B
2. Prepare solution B (0.01 M potassium ferrocyanide, 0.05 M potassium ferricyanide and 0.36 M
potassium fluoride) as follows:
4. Rinse the electrode and repeat the measurement with solution B. The potential should be
about 66 mV greater in solution B than in solution A.
Summary of method
Redox measurements are made by determining the electron activity of a solution using an inert
indicator electrode and a reference electrode. The potential difference between the indicator
electrode and the reference electrode equals the redox potential of the system. The Gel-filled ORP
and Platinum Series ORP electrodes use a platinum indicator electrode and a silver/silver chloride
reference electrode.
Required apparatus
Description Quantity/Test Unit Catalog number
Select one meter and probe combination: — — —
sension™ 1 Portable pH/mV Meter each 5170010
sension™ 2 Portable pH/ISE Meter each 5172510
sension™ 3 Benchtop pH/mV Meter each 5175010
sension™ 4 Benchtop pH/ISE Meter each 5177510
Gel-filled Combination ORP Electrode, 5-pin 1 5193900
OR
Platinum Series Combination ORP Electrode, 5-pin 1 5193700
Optional reagents
Description Unit Catalog number
Light's Solution, ampules 20/pkg 26125-20
200 mV ORP Solution 500 mL 25M2A1001-115
600 mV ORP Solution 500 mL 25M2A1002-115
Hydrochloric Acid Standard Solution 0.1 N 1L 1481253
Sodium Hydrochloride Standard Solution, 0.1 N 1L 19153
Optional apparatus
Description Unit Catalog number
Beaker, 100-mL, polypropylene each .108042
Beaker, 250-mL, polypropylene each 108046
Beaker, 400-mL, polypropylene each 108048
Beaker, 600-mL, polypropylene each 108052
Beaker, 1000-mL, polypropylene each 108053
Digital Titrator each 1690001
Flask, Erlenmeyer, 125-mL each 50543
Electrode Holder, with electromagnetic stirrer each 4530001
Electrode Holder each 4530000
Stir Bar, Magnetic, 22.2 x 7.9 mm each 2095350
© Hach Company, 2007, 2010, 2012. All rights reserved. Printed in the U.S.A. Edition 7