Professional Documents
Culture Documents
1
ION SELECTIVE ELECTRODES
HISTORY
2
Until the 60’s the only ISE is in fact the pH electrode with a glass sensitive to
protons.
The fluoride electrode appeared in the middle 60’s. Since then a number of ISE
have been developed for cations and anions.
2
ION SELECTIVE ELECTRODES
WHERE?
3
Environment
Medecine, pharmaceutical ,
cosmetics
Power plants
Chloride and fluoride in mineral waters, potassium in fruit juices, chloride in beer,
Chloride traces (high sensitivity chloride electrode) for boiler water ( corrosion
protection)
3
ION SELECTIVE ELECTRODES
APPLICATIONS
4
4
ION SELECTIVE ELECTRODES
WHY?
5
Economical
Time saving
Instrumentation simple
5
ION SELECTIVE ELECTRODES
WHY?
6
According to the type of electrode and to sample matrix , the detection limit is at
10-6M for most models (high sensitivity chloride electrode) or even at 10-7 M like
the fluoride electrode.
6
ION SELECTIVE ELECTRODES
LIMITATIONS
7
When ions are bound, it is necessary to free them before hand by an adequate
handling such as for instance wet mineralisation
Other ions present in the sample may interfere with the measurement : either they
also respond to the same electrode like for instance iodide, bromide to the chloride
electrode or they will bind or precipitate the ion to be measured, like silver ions
with chloride ions. The selectivity degree of the electrode is expressed by the
selectivity constant K and the activity of the interfering ion.
7
ION SELECTIVE ELECTRODES
THEORY
8
E=E 0+
RT
*
S
nF 100
log (ax + Σkxia n ni )
The factor 0.198T/n will be ideally 59 mV for a monovalent ion, this represents
100% sensitivity. The Nernst equation can be modified by the sensitivity of the
electrode S/100%.
Second the selectivity of an electrode is never 100%, other ion may interfere and the
degree of selectivity to the ion to be measured x versus the interfering ion x1 is
expressed by the selectivity constant Kxi. The selectivity factor has to be included
for all interfering ions.
When Kxi = 2.10-2 for instance it means that the electrode is 50 times more sensitive
to x than to 1 (1/2 . 1/10-2 ).
When the value of Kxi is higher than one for a given ion it means that the electrode
is more sensitive to the interfering ion than to the ion to be measured.
8
ION SELECTIVE ELECTRODES
THEORY
9
E= electrode potential
E0 = standard electrode potential (temperature dependant)
T= absolute temperature
F= Faraday’s constant
R= gas constant
n= valence of ion x
ax= activity of ion x
S/100% = sensitivity of electrode
Kxi = selectivity constant of primary ion x to interfering ion i
9
ION SELECTIVE ELECTRODES
WHAT IS NEEDED?
10
Ionmeter
ISE (with filling solution)
Ionmeter
ISE with filling solution if it is a ion exchange type electrode
Reference electrode preferably with double bridge of type REF251 of with salt
bridge extension
Temperature sensor : like for pH the electrode behaviour is Nernstian and therefore
temperature has to be taken into account
Standard solution or addition solution (known concentration) to calibrate the
electrode
Ionic strength is an important parameter in ion mesurements,it expresses the
concentration of all ions present in the solution both as concerns molarity and
charge.
Ion Strength Adjuster (also called “electrolyte”) is needed in order to give the
sample and the standard solutions a high and constant ionic strength. It does not
interfere on the measurement but compensatse for the difference between activity
and concentration. A Total Ion Strength Adjuster Buffer (TISAB) like for the
fluoride electrode allows at the same time to adjust the pH range.
10
ION SELECTIVE ELECTRODES
TYPICAL PROPERTIES
11
Concentration range
pH range
Response time
Life time
The detection limit is different according to type of ion. This is determined by the
solubility constant of the crystal (or crystal powder ) used for the sensing
membrane. The less soluble the membrane, the lower the detection limit. Let us take
the example of chloride : at some point the Chlorides coming from the membrane
will be measured , this is the background noise or detection limit of the electrode.
The concentration range is an essential spec of the electrode
For some electrodes the pH range is essential as hydroxide ions or protons can
interfere : for instance OH- interfere on the fluoride crystal and H+ on the
measurement of sodium, as the sodium ion selective electrode is based on a
modified pH glass. The pH range is given in the specs of the electrode.
The response time of the electrode depends on its nature and design, and on the
concentration of the ion to be measured. At high concentration the response will be
faster than at low concentrations. An approximate response time is given in the
users manuals.
The life time depends on the working principle of the electrode and on its use: for
instance the Ag(CN)-2 building up when working with cyanide electrode is very
soluble and therefore the life time of the electrode is limited. Organic ion exchange
membrane electrodes also have a shorter life time than solid electrodes.
11
ION SELECTIVE ELECTRODES
TYPICAL PROPERTIES
12
The membrane resistance is normally high and therefore the electrode is connected
to the high impedance input of the meter.
Both the sensitivity and the check point can drift with time, therefore regular
calibration is necessary.
12
ION SELECTIVE ELECTRODES
CHECKING BULLETIN
13
sensitivity
Check point
All ISE of Radiometer Analytical are supplied with their checking bulletin showing
the calibration curve and the respective potential values for each of the standard
used.
The procedure used is a measurement in the electrolyte first, after stabilization a
100 µL addition of a concentrated standard solution is made in the electrolyte so as
to obtain the lowest concentration i.e. mostly 10 - 6 M, then successive additions ,
eventually of more concentrated solutions, are made in the same beaker so as to
cover the range of concentration of the electrode.
13
ION SELECTIVE ELECTRODES
VARIOUS TYPES
14
Glass
Solid state
Single or combined
Solid state: chloride, fluoride, silver/sulfide, lead, copper, bromide, iodide, cyanide
Combined : fluoride
14
ION SELECTIVE ELECTRODES
H+ and Na+
15
The hydrated layer is very narrow (order of 10-6 cm), it forms spontaneously when
electrode is left in a solution.
For H+, the sodium or lithium ions in the glass membrane are exchanged with the
protons.
The sodium electrode is just a modified pH glass. It is still very sensitive to H+ ions,
but this is overcome by adjusting the pH to alkaline values.
15
ION SELECTIVE ELECTRODES
H+ and Na+
16
Inner solution
The glass for the sodium electrode is modified compared to the glass used for pH.
16
ION SELECTIVE ELECTRODES
SOLID STATE
17
Cl-
AgCl
FF- -
Ag+/S--
LaF3
LaF3 Ag2S
Solid state
Electrodes
CN
CN,-,II-
- - Pb ++
Pb++
AgI
AgI PbS
PbS
Br-
Br- Cu ++
Cu++
AgBr
AgBr Cu
Cu1.8 Se
1.8Se
17
ION SELECTIVE ELECTRODES
SOLID STATE
18
18
ION SELECTIVE ELECTRODES
SOLID STATE
19
F- CN- Ag+
F-
TISAB pH adjuster and Ag(CN)2- soluble, limited life
decomplexer of AlF6 and MgF4 in high CN- concentrations
LaF3 ⇔ La 3+ + 3 F-
TISAB : total ion strength adjuster buffer. 3 Functions: adjusts the pH, decomplex
the bound F- ions and adjust the ion strength so as to make sure that measurements
in samples can be referred to calibration conditions.
For the cyanide electrode for instance the equilibrium is based on the CN-
exchanged with Ag+ (from the AgI polycrystal) following the reaction :
19
ION SELECTIVE ELECTRODES
SOLID CONTACT
20
Conductive pads
Sensing polycrystal
20
ION SELECTIVE ELECTRODES
LIQUID CONTACT
21
Filling solution
Insulating material
Sensing element
21
ION SELECTIVE ELECTRODES
ORGANIC ION EXCHANGER
22
Ammonium antibiotic
The potassium and ammonium ion exchanger have structure with a cavity where the
size of the respective ion fits perfectly.
22
ION SELECTIVE ELECTRODES
ORGANIC ION EXCHANGER
23
23
ION SELECTIVE ELECTRODES
COMBINED FLUORIDE ELECTRODE
24
The potassium and ammonium ion exchanger have structure with a cavity where the
size of the respective ion fits perfectly.
24
ION SELECTIVE ELECTRODES
REFERENCE ELECTRODE
25
Constant potential
Calomel electrode is very stable but goes only up to 70°C and shows hysteresis
when temperature changes, in turn the Ag/AgCl electrode goes up to 100°C fast
response to temperature variations, but slightly more instable and sensitive to light.
However in some cases, elements from the 2 electrodes may interfere on the
measurement : for instance for sodium measurement the potassium from the KCl
filling solution will act as an interfering ion, and when measuring chloride, the
chloride from the KCl will increase the result and bring an error.
Therefore reference electrode with double junction of type REF251 or the use of a
removable salt bridge is preferred. The advantage of using a salt bridge is that when
measuring different ions, it is easier to fill it with the relevant electrolyte. However
in order to avoid contaminations, it is recommended to use one dedicated salt bridge
or double junction reference for each different ion.
The solution filled in the second junction of in the salt bridge is the electrolyte (Ion
Strenght Adjuster) in order to minimize the junction potential.
25
ION SELECTIVE ELECTRODES
MEASURING METHODS
26
Standard addition
Standard subtraction
Analate addition
Analate subtraction
26
ION SELECTIVE ELECTRODES
SELECTION OF THE METHOD
27
Sample matrix
Concentration level
Accuracy requested
Time
Convenience
27
ION SELECTIVE ELECTRODES
PROCEDURE
28
Bracketing procedure
The sample and standard solutions must be similar in composition in order to avoid
matrix effects. This is the reason why ISA is added to both standard solutions and
sample.
The TISAB used for the fluoride measurements is alltogether an ISA ( NaCl), pH
adjuster between 5 and 6 and a decomplexing agent DCTA which builds complexes
with Aluminium, iron and magnesium to release fluoride ions.
28
ION SELECTIVE ELECTRODES
DIRECT POTENTIOMETRY
29
E = E 0 + S 25T / T 25 log(C + Bl )
E = measured potential
E0 = standard electrode potential (mV determined by calibration)
S25 = sensitivity at 25°C (mV/pC)
T = calibration temperature
T25 = 298.16 K
C = concentration of standard
Bl= blank detection limit of electrode
Csmp = C meas - Bl
The concentration of the sample is the total concentration measured less the value of
the Blank, this is specially significant for very low concentrations ( non linear part
of the electrode response range).
29
ION SELECTIVE ELECTRODES
CALIBRATION CURVE
30
E (mV)
E¡
90 Standard potential
60
C1
E1
dC/Std
ST
30 Deviation from
C2 -0
fitted curve
E2
for each point
-30
-90
-120
-150
Bl
Blank = experimental
-180 detection limit of the
electrode
log[C] -210
10-7 10-6 10-5 10-4 10-3 10-2 10-1 1
The blank is determined during the calibration. It represents the detection limit of
the electrode.
It is the intersection of the extrapolated linear response part and horizontal part of
the electrode. It allows to have results in the non linear response area of the
electrode which have the same accuracy as in the linear part.
30
ION SELECTIVE ELECTRODES
ACCURACY OF DIRECT POTENTIOMETRY
31
31
ION SELECTIVE ELECTRODES
STANDARD ADDITION PRINCIPLE
32
ISE = F-
Sample = F-
Standard addition = F-
In the standard addition method, the potential of the sample is measured first, then
several additions of identical volume of a concentrated standard solution are added
and each time the potential of sample + addition recorded.
The addition volume must be small with regard to the sample volume so that the
total ionic strength remains constant, the degree of complexation of the specific ion
remains constant, the liquid junction potential of the reference electrode remains
constant.
The electrode sensitivity is determined by 2or more additions. If only one addition
is made, the slope/sensitivity of the electrode must be known.
32
ION SELECTIVE ELECTRODES
STANDARD ADDITION WHEN?
33
33
ION SELECTIVE ELECTRODES
STANDARD SUBTRACTION PRINCIPLE
34
ISE = I-
Sample = I-
Standard addition = Ag+-
The potential measurement is taken first in the sample. A reagent which combines
with the specific ion is added ( here silver to iodide), i.e. the concentration of the
specification in the sample will decrease at each subtraction.
34
ION SELECTIVE ELECTRODES
STANDARD SUBTRACTION WHEN?
35
35
ION SELECTIVE ELECTRODES
STANDARD ADDITION/SUBTRACTION
36
± CstdVstd
Csmp =
(Vsmp + Vstd )10( E1− E ) / S − Vsmp
For one addition/subtraction
Csmp sample concentration
Cstd standard concentration
Vsmp sample volume
Vstd standard volume
E potential before 1st addition/subtraction
E1 potential after addition/subtraction
S sensitivity
36
ION SELECTIVE ELECTRODES
ANALATE ADDITION PRINCIPLE
37
ISE = F-
Standard = F-
Analate (sample) = F-
The sample concentration is calculated after the procedure has been completed.
37
ION SELECTIVE ELECTRODES
ANALATE ADDITION WHEN?
38
38
ION SELECTIVE ELECTRODES
ANALATE SUBTRACTION PRINCIPLE
39
ISE = Pb2+
Complexing Standard = Pb(NO3)2
Analate (sample) = SO42-
First a potential measurement is taken in the standard solution, which is in this case
a complexing reagent of known concentration like lead nitrate (in solution) , the
electrode is selective to lead.
Additions of the sample (here sulfate) are made. The reaction is the formation of
lead sulphate which precipitates, i.e. the lead detected by the electrode is decreasing
(difference between the solubility product of lead nitrate and lead sulphate). This is
a way to measure the sulphate in a sample by direct potentiometry with a lead
electrode, as no sulphate electrode is available.
39
ION SELECTIVE ELECTRODES
ANALATE SUBTRACTION WHEN?
40
40
ION SELECTIVE ELECTRODES
ANALATE ADDITION/SUBTRACTION
41
41
ION SELECTIVE ELECTRODES
BENEFITS OF ADDITIONS/ANALATES
42
42
ION SELECTIVE ELECTRODES
LIMITS OF ADDITIONS/ANALATES
43
43