The tetraphenylborate-selective sensor is based on trinuclear chromium(III) pivalate. This
electrode consists of an electrode body containing an ion exchanger in a sensing module. This
sensing modul contains a liquid internal filling solution in contact with PVC membrane
containing a tetraphenylborate selective ion exchanger.
The electrode characteristics was fixed in order to elaborate a method for potassium
determinations in real objects. Plotting the electrode mV potential against the tetraphenylborate
concentration results in a straitght line with a slope of about 52-54 mV per decade. The time
needed to reach 95% of the stable electrode potential reading, the electrode response time, varies
from 20 s to 60s, when the concentration change is from low to high and vice versa, respectively.
The slight water solubility of the ion exchanger in the sensing module, which causes deviation
from theoretical response, determines the lower limit of detection (10 -5 M). The potentials are
constant at pH 3-11. Slight anion effects are noted with
hydrogenphosphate ions.
Following the completion of potentiometric titrations with made electrode were
determined optimal concentrations of tetraphenylborate and potassium solutions, participating in
the precipitation reactions (C 0.03 M).
Potentiometric method for quantitative analysis of potassium was developed by
potentiometric titration in direct titration method or by titration the rest, using an indicator made
electrode. By indirect potentiometric titration method can be determined a much lower
+
K
concentration of potassium ion ( ).
Cuprins
1. Introducere................................................................................................................................4-5
2. Investigaii bibliografice..............................................................................................................6
2.1 Metoda poteniometric de analiz.........................................................................................6-7
2.2 Electrozii ion-selectivi-structura............................................................................................7-9
2.3 Clasificarea electrozilor ion-selectivi...................................................................................9-10
2.4 Parametrii caracteristici pentru EIS....................................................................................10-16
2.5 Aplicarea EIS......................................................................................................................16-17
2.6 Parametrii cracteristici pentru o selecie de electrozi anion-selectivi................................17-20
2.7 EIS confecionai pentru determinarea ionilor de tetrafenilborat i potasiu.......................20-25
2.8 Metode cantitative de determinare a ionului de potasiu.....................................................26-27
3. Partea practic............................................................................................................................28
3.1 Materiale i metode............................................................................................................28-29
3.2 Rezultate i discuii.............................................................................................................29-36
4. Concluzii....................................................................................................................................37
5. Bibliografie...........................................................................................................................38-41
1. Introducere
Zi de zi tiina i industria se dezvolt cu pai rapizi, ns problema controlului analitic
rmne actual. Orice proces tehnologic sofisticat necesit un control analitic riguros, realizat
prin metode rapide, uor realizabile, puin costisitoare, sigure i nontoxice. Controlul analitic este
un sistem complicat care poate furniza informaii despre compoziia chimic a produselor,
permite s se foloseasc raional resursele naturale, s se evalueze starea mediului ambiant i s
se elaboreze msuri pentru protecia lui. Controlul analitic este o sfer n care se aplic teoria i
metodele chimiei analitice i urmrete trei scopuri: studierea, evaluarea i dirijarea compoziiei.
Pentru realizarea acestor scopuri n majoritatea cazurilor sunt necesare metode rapide, exacte,
selective, sensibile i la realizarea crora cheltuielile ar fi minime. Asemenea posibiliti ofer
metoda poteniometric cu folosirea electrozilor ion-selectivi (EIS) ce demonstreaz cererea
mereu crescnd pe piaa mondial.
Utilizarea EIS n multe domenii cum ar fi: chimia analitic, chimia fizic, biochimie, n
domeniul polurii apelor, n chimia agricol, criminologie, medicin, tehnologia produselor
alimentare, geologie, controlul calitii produilor etc. se datoreaz parametrilor lor: selectivitate
specific nalt; sensibilitate mare (limitele de detecie); durata de exploatare; domeniul pH n
care electrodul lucreaz, duratei analizei (ntruct exclud operaii ca: filtrarea, distilarea,
extracia); posibilitii efecturii analizelor n cantiti minime de probe; determinrii directe
(fr perturbarea mediului); uor pot fi recondiionai din punct de vedere al aparatajului; simpli,
uor de mnuit; ofer siguran n exploatare, economici.
Funcionarea normal i eficient a EIS este condiionat de anumii factori ca:
temperatura, natura materialului electroactiv, prezena ionilor interfereni, a agenilor tensioactivi
neionizai.
Membrana - componentul de baz a oricrui EIS la nivelul creia se realizeaz procesul
de recunoatere a speciei de determinat, are n componen SIA- variaia creia permite
cercetarea i descoperirea diferitor electrozi selectivi pentru anumite specii. SIA trebuie s
corespund anumitor cerine i anume s fie ct mai puin solubil n ap, ionizarea s se petreac
dup o singur specie, s fie stabil n soluiile folosite n confecionarea membranei; s asigure
un rspuns.
Variaia componenei sferei interne, diversificarea gamei de compui creeaz posibiliti
reale de a atinge valori ridicate a sensibilitii i selectivitii metodei, deci scopul analistului de
azi este de a gsi aa SIA ca s mbunteasc parametrii unui EIS i s corespund necesitilor
cotidiene n diferite domenii.
2. Investigaii bibliografice
2.1. Metoda poteniometric de analiz.
Metodele electrochimice de analiz snt metodele (poteniometria, amperometria,
culonometria, conductometria, voltamperometria, cronopoteniometria i altele) la care semnalul
analitic reprezint un parametru electric dependent de concentraia componentului analizat.
Un parametru esenial a metodelor electrochimice (mai sus menionate) servete diapazonul determinrii concentraiei i selectivitii (posibilitatea determinrii componentului
persistent n prezen de alii genani), n majoritatea cazurilor aceti parametri depind de
electrod . Metoda poteniometric de analiz se bazeaz pe utilizarea dependenei semnalului
analitic (potenialul) a senzorului specializat numit electrod de msurare , de componena soluiei
de analizat.
n caz ideal electrodul de msurare ales (selectiv) reacioneaz la determinarea unui ion
(sau grupe de ioni) i potenialul lui depinde de coninutul acestor ioni n soluie i se supune
legitii iui Nernst:
E=E o +S lga
(1)
unde: a-activitatea ionilor analizai din soluie; S - panta - sensibilitatea senzorului (poate
fi evaluat experimental) 2,3RT/nF (R - constanta gazelor; T - temperatura absolut; F - numrul
lui Faraday; n - numrul de electroni care particip n reacie).
n practic ns se observ unele abateri de la liniaritate datorit prezenei ionilor
interfereni (specifici fiecrui tip de electrod ).
Metoda poteniometric de analiz se divizeaz n mai multe tipuri:
a) Poteniometria direct.
Metod bazat pe fixarea dependenei potenialului electrodului de msurare de
concentraia soluiei (construirea graficului de calibrare ulterior utilizat pentru cercetare soluiilor
cu concentraii necunoscute ). Este o metod destul de simpl i expresiv, exactitatea creia
depinde n primul rnd de tipul electrodului de indicaie. Trebuie de menionat c odat cu
trecerea timpului, scade durata de exploatare a electrodului sensibilitatea scade i cresc erorile
rezultatelor.
n acelai timp avantajele i dezavantajele metodei constau n aceea c valoarea
potenialului msurat depinde de activitatea ionilor. Aceasta este unica metod ce permite
determinarea direct a activitii ionilor a soluiei, dar privind problema din alt punct de vedere
constatm c analitii deseori sunt interesai de concentraia i nu de activitatea ionilor.
Traducerea activitii ionilor n concentraie cu utilizarea anumitor coeficieni empirici cauzeaz
6
Membran
Soluie de analizat de
Electrod de
concentraie necunoscut
referin extern
de referin intern
ER,2
EM`
Faza
EM
Ej
Faza
ER,1
Faza
RT a A
EM=
ln
zF aA
aA
aA
(2)
R - constanta gazelor;
T - temperatura absolut;
E=const+ S lg a + k
k pot
A ,B
A B
pot
A ,B
zA
zB
B
(3)
z B - sarcina ionului B;
pot
k A ,B
a
b
c
I
II
1
membran impregnat
III
a
membranei;
b
IV
membran nglobat
Membrane solide
microcristaline
Membrane lichide
de sticl
policristaline
plastifiate
10
aion
unde p
aion
= -log
aion
II
p aion
).
Sensibilitatea EIS
Scopul cercetrilor n metoda poteniometric const n crearea i studierea noilor forme
S=2,303 RT / zF
(4)
Selectivitatea EIS
Selectivitatea este definit ca raportul dintre sensibilitatea parial referitoare la specia
z
Za
k pot
A , B aB
E = const 2,903RT/F lg a A +
A B
unde:
A-specia primar; B-specia inferioar;
pot
k A ,B - coeficient de selectivitate.
Determinarea selectivitii unui EIS se poate realiza prin :
- metoda soluiilor separate;
- metoda interferenei constante.
Metoda soluiilor separate -prezint dou variante
12
(5)
a) dac a AZa =a Zb
A
rezult c:
E
( HE B )/ 2,303 RT /F
log k pot
A , B=
(6)
unde EA i EB reprezint potenialul EIS n soluia care conine specia primar (A),
respectiv n soluia care conine numai interferentul (B). [5]
b) dac soluiile cu specia primar i interferen se prepar astfel nct potenialele
msurate s fie identice, atunci coeficientul de selectivitate se determin din relaia (3):
1
pot
k A ,B =
k pot
A ,B
a Za
A
1
a Zb
A
(7)
corespunztoare acestor activiti. Metoda soluiilor separate este foarte uor de aplicat, dar nu
este recomandat deoarece rezultatele astfel obinute nu sunt ntotdeauna relevante. Astfel
procesele determinante de potenial sunt adesea diferite n soluii pure ale ionilor A sau B fr de
cele existente n soluii mixte. Pe de alt parte, deoarece condiiile de determinare a E n soluiile
ionilor B difer de cele din proba de analizat, valorile obinute pentru acest coeficient sunt
deseori eronate. [5]
Metoda interferenei constante - const n msurarea potenialul EIS n soluii de
activitate a speciei primare (A) i de activitatea constant a speciei interferente (B):
Reprezentarea grafic E=f(-lg a) este o curb care prezint urmtoarele poriuni:
500
450
400
350
300
250
200
0
13
10
pot
k A ,B
se poate determina
n dou moduri:
Metoda 1 - const n localizarea grafic a punctului T, ca intersecie extrapolrii zonelor
liniare PQ i SR, unde electrodul rspunde n mod egal la ambii ioni
k pot
A ,B
se calculeaz ca
raportul dintre activitatea ionului A n punctul T, mai exact abscisa punctului T, ( A) i valoarea constant, conform ecuaiei:
pot
k A ,B
(aA) / (8)
Metoda 2 - pentru curbele de rspuns cu poriuni RS mai puin orizontale, cea mai simpl
metod este introducerea datelor experimentale n ecuaia (5) i determinarea prin calculul a
pot
constantei k A ,B .
Aceast metod este mai laborioas, dar mai riguroas, pricina erorii induse de prelucrri
grafice. Cunoscnd valoarea coeficientului de selectivitate a electrodului pentru anumii ioni - ne
permite s alegem condiiile optime pentru efectuarea msurrii.
irul selectivitii al EIS - anionici (teoretic) - este dircet proporional cu energia de
hidratare a acestor anioni:
ClO4- > SCN- > I- > BF4- > NO3- > Br- > l- > F
IV
Timpul de rspuns.
A fost propus estimarea timpului de rspuns prin intervalul de timp (t 95) necesar
senzorului s ating 95% din valoarea staionar a timpului de rspuns. n practic, ns,
determinarea t95 comport unele dificulti deoarece este necesar s se cunoasc valoarea
staionar a rspunsului; ceea ce nu este ntotdeauna convenabil sau existena unui model
matematic pentru funcionarea senzorului pe baza cruia s se evalueze t 95. [6] Varianta practic operaional consider timpul de rspuns ca definit prin intervalul de timp scurs de la contactul
14
senzorului cu sistemul investigat pin cnd variaia n timp a semnalului scade sub anumit
valoare, arbitrar stabilit. Practica a evideniat c timpul de rspuns al SP este puternic influenat
de concentraia speciei primare:
-
Pornind de la aceeai concentraie iniial spre concentraia final tot mai mari sau
tot mai mici, timpul de rspuns este tot mai mic, respectiv tot mai mare.
Dac raportul dintre concentraia final i cea iniial se pstreaz constant atunci
timpul de rspuns corespunde unei cretere de concentraie este mai mic dect cel pentru
scdere de concentraie;
n sistemele aflate n curgere, timpul de rspuns este ntotdeauna mai mic dect
cel determinat ntr-un sistem static. n prezena unui electrolit indiferent concentrat (C>0,1M),
agitarea nu influeneaz valoarea semnalului staionar;
Pentru EIS omogeni, compaci i fr scurgeri nu s-a constatat dependen de
geometria senzorului (dimensiuni i form).
Prezena ionilor interfereni agenilor tensioactivi neionizai i n deosebi formarea unui
strat difuzional - rezistiv pe suprafaa senzorului determin creterea timpului de rspuns.
Ultimul aspect menionat este des ntlnit n cazul EIS solizi de sticI sau lichizi cu membran
polimcric. ndeprtarea acestei bariere de difuzie duce inevitabil la micorarea drastic a
timpului de rspuns.
V
Timpul de folosire
Timpul de lucru a unui EIS este cuprins n intervalul de 4 - 10 sptmni. Dup ce trece
acest timp, panta senzorului va scdea, dar limita de detecie se va micora. Acest fapt este
cauzat de pierderea plastifiantului, purttorului sau plastinei ionice de pe filmul polimeric, ca
rezultat a introducerii n soluia de analizat. [7]
VI
unei multitudini de parametri fizici (solubilitate, presiune, viscozitate, mobilitate) sau chimici
(viteza de reacie, echilibru). Influena temperaturii asupra funcionrii SE este complex i greu
de evaluat:
-
analizat, prin deplasarea unor echilibre sau prin dilatarea volumului sistemului nvestigat.
-
VII
Reproductibilitatea
15
Condiia necesar pentru validitatea ecuaiei lui Nernst este echilibrul termodinamic ntre
soluia adiacent i faza organic. Aceast condiie este respectat atunci cnd faza de transfer
cinetic i formarea complexului este mai rapid dact procesul de difuzie. Din punct de vedere al
electrochimiei schimblu de current trebuie s fie mai mare n comparaie cu curentul care trece
prin memran. [7]
2.5 Aplicarea EIS [7]
Senzorii electrochimici (SE) servesc la efectuarea determinrilor analitice. Deoarece n
structura unui SE este prezent cel pufin interfa (senzor/prob de analizat), msurrile pot fi
nsoite uneori de efecte nedorite n timpul funcionrii SE. Dintre acestea pot meniona efectul
de memorie, de inhibare sau chiar de blocare a transferului de sarcin etc.
Deoarece asemenea efecte nu sunt prezente la senzorii poteniometrici, acetia sunt
competitori ai SE, fiind preferai acestora ntr-o serie de aplicaii. Dar, exist totui utilizri n
care SE nu au rivali. Este cazul, n primul rnd, al determinrilor de concentraii foarte sczute n
analiza de urme. n aceste situaii, hotrtoare pentru aplicare n favoarea SE au fost dou dintre
caracteristicele acesteia: pre foarte convenabil i respectiv un consum de putere foarte redus.
nsemntate practic foarte mare n metodele poteniometrice are determinarea pH-ului cu
electrod de sticl i ali electrozi, precum i la determinrile poteniometrice directe ale
concentraiilor (activitilor) altor ioni cu ajutorul electrozilor ion-selectivi (poteniometria).
Principalele domenii de aplicare:
Monitorizarea polurii: CN-, F-, S2-, l-, NO3- din afluente sau apele naturale;
Agricultur: CN-,l-, NO3-, I-, NH4-, Ca2+ din soluri, plante, fertilizani, adaosuri
alimentare;
16
Cele mai folosite modele de EIS plastifiate, cristaline sau gelatinoase, sunt
nepoluante i durabile n folosire, astfel fiind ideale pentru msurri n
condiii de laborator;
La conditii nefavorabile, cnd ionii determinai sunt n soluii relativ
diluate i ionii interfereni nu geneaz, msurarea poate fi efectuat simplu
i repede (ex: prin scufundarea n ru a electrozilor);
Cteodat sunt folosii pentru a afla dac un ion particular are
concentraie mai mic dect poate fi msurat de electrodul dat;
Sunt indispensabili pentru monitorizarea continu a schimbrii
concentraiilor (ex: titrare pbteniometric);
Sunt folosii particular n aplicaii biomedicinale, deoarece ei msoar
activitatea ionului direct, mai rar dect concentraia;
La folosire cu precauie i calibrare frecvent EIS au precizie de 2-3%;
EIS este unul din instrumente care poate msura activitatea ionilor att
negativi, ct i pozitivi;
EIS nu sunt afectai de speciile colorate sau tulburate;
Pot fi folosii n soluii apoase cu temperaturi mai mari de cea standard.
Cei cristalini la 0 - 80C, cei plastifiai 0 - 50C.
Anion
HCO3-
Panta,
mV/decad
-54
Diapazon liniar, M
10-2 - 10-5
1,010-6 - l,010-1 i 5,010-8 -
Ref.
[19]
F-
[20]
NO2
-58,4 - -60,8
NO3-
-59,1
110-6 - 2,2510-2
[21]
HPO42- - 1
-29,5
1,010-1 - 1,010-7
[22]
HPO42- - 2
-28,6
1,010-1 - 4,010-7
[23]
PO43-
-70,8
10-5-10-1
[24]
-2
5,010
17
CN-
-90
10-6-10-2
[25]
SCN- -1
-58,3
4,810-7-1,010-1
[26]
SCN- -2
-59,5
10-7-10-1
[27]
SCN -3
-57,8
1,010-7- 1,010-1
[28]
SO32-
-51,5
6,010-5-1,010-2
SO42- - 1
-29,2
10-2-10-6
[30]
S042- - 2
-29,2
10-2-10-6
[31]
S042-- 3
-29,6
10-2-10-6
[32]
F-
-62,7
1,010-6-1,010-1
[33]
Cl-
-54,5
1,010-6-1,010-1
[34]
ClO4- - 1
-60,6
1,010-6-1,0
[35]
ClO4- - 2
-59,3
10-1-5,010-7
[36]
Br- -l
-59,0
7,010-6-1,010-1
[37]
Br- -2
-59,2
2,210-6-1,010-1
[38]
I- - l
-58,9
5,010-7- 1,010-1
[39]
I- - 2
-60
8,010-7-1,010-1
[40]
I- - 3
-59,0
2,010-8-2,010-2
[41]
I3--1
-59,0
5,010-8-1,010-2
[42]
I3--2
-59,3
10-7-10-1
[43]
I3--3
-59
5,010-8-1,010-2
[44]
18
ClO4-
[29]
I3--4
-59,00,5
5,01081,0102
[45]
IO4-
-60,1
8,010-3-2,710-1
[46]
NO3-
HPO42-
SO42-
SCN-
F-
ClO4-
Fig. 4. Structurile unor ionofori folosii pentru confecionarea EIS pentru anioni [8]
19
Br-
I3-
I-
Fig. 5. Structurile unor ionofori folosii pentru confecionarea EIS pentru anioni [8]
Proprieti
Formula molecular
NaBC24H20
Masa molar
342,216 g/mol
Aspect
solid, alb
Punctul de topire
>300 C
Solubilitatea n ap
47 g/100 ml
Tetrafenilboratul de sodiu este un compus organic cu formula NaB(C 6H5)4. Este o sare, n
cadrul creia anionul este format din patru inele fenil unite de bor. Acest solid cristalin este
folosit pentru a pregti alte sruri de tetrafenilborat, care adesea sunt foarte solubile n solveni
organici. Compusul este folosit n chimie anorganic i organometalic ca un agent de
precipitare.
Sinteza
Tetrafenilboratul de sodiu este sintetizat conform reaciei dintre tetrafluoroboratul de
sodiu i reactivii aril Grignard:
NaBF4 + 4 ArMgBr 2 MgBr2 + 2 MgF2 + NaBAr4 (unde Ar = aril)
Utilizarea n sinteze chimice
Prepararea srii N-acilamoniu
Adaosul de tetrafenilborat de sodiu la o soluie de o amin teriar i clorur de un acid
acetonitril d sare acilamoniu prin precipitarea de NaCl din amestecul de reacie. Aceast metod
are un domeniu de aplicare larg:
RC(O)Cl + R'3N + NaB(C6H5)4 [RC(O)NR'3][B(C6H5)4] + NaCl
Tetrafenilboratul de sodiu este de asemenea utilizat ca un donator de fenil n catalizatorul
de paladiu n reacii cross-cuplare implicnd vinil i aril triflates pentru a oferi compuilor
arilalcani i biaril randamente bune i respectiv, n condiii uoare.
Utilizarea n chimia coordinativ
Tetrafenilboraii sunt deseori studiai de chimia organometalic datorit solubilitii lor
bune n solvenii nepolari i cristalinitii lor. Spre exemplu, complecii homoleptici trimetilfosfit
{M[P(OCH3)3]5}2+(Ni, Pd, i Pt) sunt preparai ca sruri de tetrafenilborat. n mod similar,
tetrafenilboratul de sodiu este folosit pentru a izola complecii care conin ca liganzi dinitrogen.
21
n reacia de mai jos, tetrafenilboratul de sodiu permite moleculei de azot s deplaseze ligandul
clorur, care este eliminat din soluie sub forma precipitatului de clorur de sodiu:
FeHCl(difosfin)2+ NaB(C6H5)4+ N2 [FeH(N2)(difosfin)2]B(C6H5)4+ NaCl
Utilizarea de tetrafenilborat este limitat la cationi non-acizi, deoarece acizii puternici
descompun acest anion la trifenilboran i benzen:
H++ B(C6H5)4- B(C6H5)3+ C6H6
Ionul de tetrafenilborat, de asemenea, formeaz precipitate cu ionii ce au dimensiuni i
sarcin similar: NH4+, Rb+, Cs+, Ag+ i Tl+. Tetrafenilboratul, din timp n timp, a fost propus ca
un extractant pentru msurarea cantitii i a fluxurilor de schimb i fixarea K + i NH4+ pentru
evaluarea fertilitii solului. De asemenea, a fost propus ca un instrument n curarea scurgerilor
radioactive, n care ceziul radioactiv, este imobilizat i recuperat. La fel, tetrafenilboratul a fost
propus ca un reactiv analitic pentru determinarea K+, fie gravimetric sau prin titrare, cu toate
acestea, tehnicile standarde de fotometrie n flacr, spectroscopia de absorbie atomic sunt, n
general, considerate ca fiind superioare la aceast utilizare. n lucrrile de laborator este fost
folosit ca un extractant pentru K+ din minerale.
Proprieti
Formula molecular
KBC24H20
Masa molar
358,3249 g/mol
Aspect
solid, alb
Produsul de solubilitate
diluat;
selectivitate la K+ n prezena Na+ (70 de ori mai mult); NH4+ (80 de ori
mai mult); Pb2+, Cu2+ (slab interfereni); Ca2+, Mg2+, Fe3+,Li+ nu interfer.
Confecionarea i caracterizarea electrodului potasiu - selectiv n baz de siliciu
poros [14]
Electrodul potasiu - selectiv n baz de siliciu poros a fost confecionat folosind procesor
microelectronic i tehnica de gravur prin anodizare electrochimic. Caracteristicile
electrochimice funcionale a electrodului sunt:
diluat;
coeficientul de selectivitate ki,j este: 1,8 pentru NH4+; 3,6 - Li+; 4,1 - Na+;
25
laboratoarele clinice pentru determinarea n lichidele biologice a metalelor alcaline (Na +i K+) i
metalelor alcalino-pmntoase (Ca2+i Mg2+), au fost: simplitatea metodei, rapiditatea metodei,
consumul minim de substane, limita de eroare (2-3%) apropiat de limita de eroare din
determinrile chimice.
Principiu:
Serul nehemolizat se dilueaz cu ap bidistilat, apoi se pulverizeaz ntr-un dispozitiv
special cu ajutorul unui curent de gaz purttor. Aerosolii formai n urma pulverizrii se amestec
cu unul din amestecurile de gaze: metan-aer, propan-aer sau acetilen-aer care arde ntr-un
arztor special (Brenner), dnd natere la spectre caracteristice. Liniile spectrale ale potasiului
pot fi selecionate cu ajutorul unui filtru. Lumina flcrii este proiectat cu ajutorul unei oglinzi
concave, prin filtrul de selecie, pe o celul fotoelectric, ceea ce produce un fotocurent, care se
poate msura cu ajutorul unui galvanometru. Intensitatea fluxului luminos este proporional cu
concentraia ionilor K+ din ser.
27
3. Partea practic
3.1 Materiale i metode
Tehnici experimentale folosite i modul de operare:
Substanele iniiale folosite la confecionarea membranei:
a)
[ Cr 3 O ( Piv )6 ( H 2 O )3 ][ B ( C 6 H 5 )4 ]
0,02g -
-SIA;
baz de PVC. Modul general de preparare a membranelor const n dizolvarea probei de PVC de
0,3 g prin agitare n THF. La omogenizarea acestora se adaug 0,3 g plastifiant (DOF), dup care
se adaug SIA i tetrafenilboratul de sodiu agitnd continuu pn cnd amestecul se
omogenizeaz total. Sistemul obinut se transfer ntr-o capsul Petri i se las timp de 12h
pentru ca s se evapore dizolvantul. [16]
b)
PVC. Din filmul se taie discuri, cu diametrul puin mai mare dect diametrul tubului din PVC,
care apoi se ncleie la acest tub folosind THF i pipet. Dup scurt timp se verific ermecitatea
electrodului prin adugarea n el a apei. Electrodul cptat se condiioneaz n soluie de 0,3 M
(n acest caz soluia srurilor solubile de BPh 4-), n interiorul electrodului analog se introduce
soluia de 0,3 M i dou picaturi de NaCl (pentru mrirea electroconductivitii). Timpul de
precondiionare a electrozilor a fost cel puin 24h. Datele experimentale au fost colectate
folosind ionomerul I-135MI cu electrozii:
Electrod de referin - Ag, AgClNaCl(saturat)
Electrodul de indicaie - electrodul de BPh4- - senzitiv
BPh4- (0,3M), NaCl(0,005M)AgCl, Ag;
Soluiile iniiate de 0,3 M au fost preparate din probe cntrite, dar soluiile cu
concentraii mai mici s-au preparat prin diluare.
c)
Potenialul celulei care cuprinde EIS i electrodul de referin este msurat cu dou
soluii separate, una coninnd ionul A cu activitatea aA, i alta coninnd ionul B cu
aceeai activitate aA=aB. Valorile msurate respectiv sunt EA i EB. Coeficientul de
selectivitate se calculeaz conform urmtoarei ecuaii:
pot
lg K A ,B =
( E A EB ) z A F
RTln 10
+ ( 1z A /z B ) lg a A
(9)
Au fost preparate soluii de 0,02 M a srurilor solubile de Cl-, CO32-, SO42-, NO3-, ClO4-,
HCO3-, HPO42-, H2PO4-, NO2-, Br-, I- a cror potenial a fost msurat folosind electrodul
confecionat.
timpul de rspuns 20 60 s;
29
La concetraii mici a soluiilor (~0,01 M) curba de titrare practic este liniar cu un salt de
titrare puin vizibil. Cu creterea concentraiei ambelor soluii (~0,03 M) curba are form clasic.
n aceste condiii volumul de titrant n punctul de echivalen poate fi determinat mai exact.
Pentru confirmare au fost construite curbe difereniale E/V n funcie de volumul titrantului.
30
45
40
35
30
25
~0,01
20
~0,02
~0,03
15
10
5
0
0
10
11
0,01013
0,01018
5,03
5,50
0,47
9,34
0,01999
0,02007
5,02
5,50
0,48
9,56
0,03020
0,02999
4,97
5,00
0,03
0,60
0,02022
0,03009
0,01003
1,67
1,70
Eroarea absolut V, ml
0,12
0,03
Eroarea relativ r, %
4,84
1,80
2,48
2,60
0,04011
1,25
1,30
32
0,02001
9,93
10,00
0,07
0,01008
0,03004
14,90
15,00
0,10
0,04001
19,85
19,75
0,10
0,71
0,67
0,50
HCO3-
NO3-
Cl-
H2PO4-
Br-
NO2-
ClO4-
HPO42-
SO42-
CO32-
0,20231
0,28660
0,32032
0,32032
0,3709
0,40461
0,45519
1,18438
1,38668
1,45412
33
Astfel a fost cptat irul de selectivitate HCO3-> NO3-> Cl-> H2PO4-> Br-> NO2-> ClO4->
HPO42-> SO42-> CO32-. Analiznd acest ir se poate de menionat c cel mai mare efect genant
manifest ionii HCO3-, NO3-i Cl-.
De toate acestea s-a inut cont la elaborarea metodei poteniometrice de analiz
cantitativ a potasiului prin titrare poteniometric, n varianta titrrii directe sau prin titrarea
restului, utiliznd electrodul confecionat.
n cazul titrrii directe se titreaz o soluie cu concentraia cunoscut a tetrafenilboratului
cu proba care conine ionii analizai. Concentraia uneia dintre soluii participante la reacia de
precipitare, care se produce n timpul titrrii poteniometrice, trebuie s fie 0,03 M. Pentru
verificarea metodei au fost titrate soluii de tetrafenilborat cu probe de sruri de potasiu cu
concentraii cunoscute. Au fost studiate diferite sruri pentru a pune n eviden efectul genant al
anionului asupra rspunsului electrodului sensibil la prezena tetrafenilborailor.
C(KNO3)=0,01002N
C(KCl)=0,01008N
C(K2SO4)=0,01009N
KNO3
0,01002
KCl
0,01008
0,03002
14,89
14,98
34
K2SO4
0,01009
14,88
Vexp., ml
Eroarea absolut V, ml
14,5
0,48
15,25
0,36
15,00
0,12
Eroarea relativ r, %
3,20
2,42
0,81
Deoarece ionii de sulfat interfereaz mai puin, curba de calibrare are un salt de titrare
mai mare. La fel exactitatea msurrilor crete odat cu scderea efectului genant al anionului.
n cazul titrrii indirecte la proba care conine cationii de potasiu se adaug un exces
cunoscut de soluie a tetrafenilboratului, iar excesul este titrat poteniometric cu soluie standard
de sulfat sau clorur de potasiu. n aceast variant pot fi analizate probe ce au un coninut mic
de K+, iar concentraiile soluiilor tetrafenilborat i clorur sau sulfat de potasiu s fie 0,03 M.
Pentru analiz s-au tratat cu exces de tetrafenilborat (5ml 0,03002 mol/l) diferite probe ce
conineau K+ cu concentraia cunoscut(~0,01N), iar excesul de anion a fost titrat cu soluie K +
(~0,04N).
C(KNO3)=0,04014N
C(KCl)=0,04011N
C(K2SO4)=0,04018N
Tab. 9 Volumul soluiilor KCl n punctul de echivalen obinut n urma titrrii poteniometrice
indirecte
X
C (X), mol-echiv./l
C(X), mol-echiv./l
C(NaBPh4), mol-echiv./l
0
KNO3
0,01002
0,04014
KCl
0,01005
0,040110
0,03002
35
K2SO4
0,01009
0,04018
Vteor., ml
Vexp., ml
Eroarea absolut V, ml
2,49
2,6
0,11
2,43
2,5
0,07
2,48
2,5
0,02
Eroarea relativ r, %
4,42
2,88
0,81
B (C6 H 5 )4
36
4. Concluzii
timpul de rspuns 20 60 s;
37
5. Bibliografie:
1.
. . . . 1, :
. 2004. 608 .;
2.
Oniciu L., Murean L. Electrochimie aplicat. ClujNapoca: Presa Universitar
Clujean. 1998. 68 p.;
3.
Lengyel T., Inczedy J., Ure A. M. Compendium of analytical nomenclature.
International Union of Pure and Applied Chemistry.2002;
4.
Recommendations for nomenclature of ion-selective electrodes. Pure and Applied
Chemistry. Vol. 48. Printed in Great Britain: Pergamon Press. 1976. p.127-132;
5.
Potentiometric selectivity coeficients of ion-selective electrodes. Pure and Applied
Chemistry. Vol. 72. Tokyo. 2000. p.1851-2082;
6.
. . . -
. . 2001. 400 .;
7.
Faribod F., Ganjali R., Dinardand R. The fabrication of the potentiometric
membrane sensors and their applications. In: African Journal of Biotechnology. Vol. 6.
2007. p.2960-2987;
8.
Faribod F., Ganjali R., Dinardand R. Developments in the Field of Conducting
and Non-conducting Polymer Based Potentiometric Membrane Sensors for Ions Over the
Past Decade. In: Sensors. Vol.8, 2008. p.2331-2412;
9.
http://en.wikipedia.org/wiki/Sodium_tetraphenylborate;
10.
http://en.wikipedia.org/wiki/Potassium_tetraphenylborate;
11.
Karrman K. J., Bladh E., Gedda P.-O. Potentiometric Micro-Titration of Potassium. In:
Pica E. M., Cosma V., Nacu H. Potassium ion-selective electrode based on decyl-18-
Zhang J., Zhu J., Lu W., Zi J. Fabrication and Characterization of Potassium Ion-
Selective Electrode Based on Porous Silicon. In: Sensors Journal, IEEE. Vol.7. 2007. p.38-42;
15.
http://www.scribd.com/doc/81172959/63/Determinarea-potasiului-seric;
38
16.
poteniometrici
anionici
pe
baza
pivalailor
trinucleari
ai
cromului(III).
Conferina
0 50 and its solubility in water 0 97,5 . Russian Chemical Bulletin, 1971, vol. 21 (1), p. 69;
18.
Harvey D. Modern analytical chemistry. McGraw-Hill Higher Education, 2000, 817 p.;
19.
Maj-Zurawska M., Ziemianek D., Mikolajczuk A., Mieczkowski J., Lewenstam A.,
Ganjali M. R.; Shirvani-Arani S.; Norouzi P.; Rezapour M.; Salavati-Niasari M. Novel
nitrite membrane sensor based on cobalt(II) salophen for selective monitoring of nitrite ions in
biological samples. Mikrochim. Acta 2004, 146, p.35-41;
21.
membrane ion-selective electrode (ISE) and its application for determination of nitrate in
tomatoes. J. Chil. Chem. Soc. 2004, 49, p.31-34;
22.
Ganjali M. R., Norouzi P., Ghomi M., Salavati-Niasari M. Highly selective and sensitive
Fabrication of a highly selective and sensitive HPO 4 2-PVC sensor based on a oxo-molybdenum
methyl-salen. J. Bra zil. Chem. Soc. 2006, 17, p.859-865;
24.
Goto Y., Miyazaki K., Hattori H., Yuchi A. Phosphate-selective electrode using
Marin M. A. B., da Silva R. C., Lehmkuhl A., da Silva, J. B. B., Ganzarolli E. M., de
Queiroz R. R. U. Automatic system for sequential determination of free and total cyanide using
homogeneous membrane tubular ion-selective electrode. Quim. Nova 2000, 23, p.23-29;
39
26.
Jeon S. W., Yeo H. Y., Jeong H. S., Oh J. M., Nam K. C. Polymeric ISE for hydrogen
amount of SO42- based on a zinc-Schiff's base. Anal. Lett. 2003, 36, p.881-894;
32.
ionophore for an anion selective electrode with a remarkable sulfate selectivity. Anal. Chim. Acta
1998, 358, p.35-44;
33.
Chandra S., Ruzicka A., Svec P., Lang H. Organotin compounds: An ionophore system
for fluoride ion recognition. Anal. Chim. Acta 2006, 577, p.91-97;
34.
Ganjali M. R., Pourjavid M. R., Rezapour M., Poursaberi T., Daftari A., Salavati-Niasari
membrane electrode based on a gold(I) complex: application to water and urine analysis. Anal.
Chim. Acta 2000, 415, p.159-164;
40
36.
Ganjali M. R., Yousefi M., Poursaberi T., Naji L., Salavati-Niasari M., Shamsipur M.
Highly selective and sensitive perchlorate sensors based on some recently synthesized Ni(II)hexaazacyclotetradecane complexes. Electroanalysis 2003, 15, p.1476-1480;
37.
Ganjali M.R., Norouzi P., Golmohammadi M., Rezapour M., Salavati-Niasari M. Novel
bromide PVC-based membrane sensor based on iron(III)-salen. Electroanalysis 2004, 16, p.910914;
38.
Singh A. K., Mehtab S., Saxena P. A bromide selective polymeric membrane electrode
electrode based on salophen complex of cobalt (III). J. Braz. Chem. Soc. 2005, 16, p.571-577;
40.
Poursaberi T., Hosseini M., Taghizadeh M., Pirelah H., Shamsipu M., Ganjali M.R. A
selective membrane electrode for iodide ion based on a thiopyrilium ion derivative as a new
ionophore. Microchem. J. 2002, 72, p.77-83;
41.
Chai Y. Q., Yuan R., Xu L., Xu W. J., Dai J. Y., Jiang, F. A highly sensitive PVC
membrane iodide electrode based on complexes of mercury(II) as neutral carrier. Anal. Bioanal.
Chem. 2004, 380, p.333-338;
42.
Ganjali M. R., Emami M., Javanbakht M., Salavati-Niasari M., Shamsipur M., Yousefi
between iodine and a new Schiff's base as anion-carrier in construction of a highly selective
triiodide PVC-based membrane electrode. Can. J. Anal. Sci. Spect. 2006, 51, p.108-116;
44.
Ganjali M. R., Moghaddamb M. R., Norouzi P., Shirvani-Arani S., Daneshgar P. Adib M.,
Sobhi H. R. Highly selective and sensitive triiodide PVC-based membrane electrode based on a
new charge transfer complex of 2-(((2-(((E)-1-(2-hydroxyphenyl) methylidine) amino) phenyl)
imino) methyl) phenol for nano-level monitoring of triiodide. Anal. Lett. 2006, 39, p.683-695;
41
45.
Ganjali M. R., Emami M., Javanbakht M., Salavati-Niasari M., Shamsipur M., Yousefi,
42