Professional Documents
Culture Documents
Electrochemical Sensors
By
Dr.K.Venkateswara Rao
Associate professor of Nanotechnology
JNT University Hyderabad
Potentiometric Sensors
Principle:
The signal is measured as the potential difference (voltage) between the working
electrode and the reference electrode. The working electrode's potential must
depend on the concentration of the analyte in the gas or solution phase. The
reference electrode is needed to provide a defined reference potential
The experimental set-up
is simple. As reference
electrodes, those of the
second kind are useful.
They are created by a
combination of a
metal/metal ion electrode
and a stock of a sparingly
soluble salt of the
participating metal.
Among the practical
examples is the
silver/silver chloride
Experimental set-up for
electrode with a salt
potentiometry
bridge
So-called pH meters are well suited of since potassium
for this purpose they commonly
have input resistances in the gigaohm range.
chloride solution. The
instrument for potential
With the advent of the sensor age, efforts were
launched to miniaturize sensors and to mass
produce them. Simultaneously, a tendency
arose to apply the term sensor also for well-
known traditional potentiometric probes. It
seems better, however, to restrict the term
sensor exclusively to devices characterized by
miniature dimensions and by widespread
availability. Nevertheless, it is useful to start
with classical electrode forms when discussing
potentiometric sensors. Classical
potentiometric electrodes differ from
Nernst’s equation is the basis of all potentiometric measurements. The common form of
this equation is that given in below Equation for a working temperature of 25◦C.
•Type I sensors have an electrolyte containing mobile ions of the chemical species in the gas phase
that it is monitoring. The commercial product, YSZ oxygen sensor, is an example of type I.
•Type II sensors do not have mobile ions of the chemical species to be sensed, but an ion related to
the target gas can diffuse in the solid electrolyte to allow equilibration with the atmosphere. Therefore,
type I and type II sensors have the same design with gas electrodes combined with metal and an
electrolyte where oxidized or reduced ions can be electrochemically equilibrated through
the electrochemical cell. In the third type of electrochemical sensor, auxiliary phases are added to the
electrodes to enhance the selectivity and stability.
•Type III sensors make the electrode concept even more confusing. With respect to the design of a
solid state sensor, the auxiliary phase looks as part of the electrode. But it cannot be an electrode
because auxiliary phase materials are not generally good electrical conductor. In spite of this
confusion, type III design offers more feasibility in terms of designing various sensors with different
auxiliary materials and electrolytes.
Selectivity of Potentiometric Sensor
Selectivity is one of the most important advantages of potentiometric
sensors. The ideal case would be the specific sensor which responds
to only one single type of ion.
Galvanic cells can be set up with solid electrolytes rather than electrolytic
solutions. Such a cell is the basis for a well-known potentiometric gas
sensor, the lambda probe. The latter is designed to determine the oxygen
content of combustion gases, e.g. in motor vehicles. The lambda probe
can operate in two different modes, either potentiometrically or
amperometrically.
The Ion-Selective Field Effect Transistor (ISFET)
An ISFET is an ion-sensitive field effect transistor used for measuring ion
concentrations in solution; when the ion concentration (such as H+, see
pH scale) changes, the current through the transistor will change
accordingly. Here, the solution is used as the gate electrode. A voltage
between
For mass substrate
productionand oxide
of this surfaces
sensor arises
type, due
nearly allto anhighly
the ion sheath. .
developed
technological arsenal of microelectronics could be used. On the other
hand, all the achievements of potentiometry with ISEs also became
usable for ISFET sensors.
ISE with MOSFET and low
impedance connection to
measuring instrument (left).
Measuring set-up with ISFET
(right)
ISFET structure
(simplified)
Even the insulation layers of SiO2 and Si3N4 exhibit some pH sensitivity.
More common pH-sensitive layers are liquid membranes (polymers with
solvent) containing ion exchangers, e.g. amines attached with a
hydrophobic alkyl chain at the opposite end. pH-sensitive glass has also
been tested, in an adaptation of enamel electrodes. pH-sensitive ISFETs
Two problems have accompanied the development of the ISFET from the
very beginning.
The first one is the problem of direct contact for making contact with the
sensitive membrane, which is inherent in ISEs.
To ensure accurate measurement, instruments which work with ISEs must have a high
input impedance. As a rule of thumb, their input resistance should be ca. 100 times
higher than the internal resistance of the cell containing the ISE. Since several ISEs (e.g.
the glass electrode) have resistances in the range of megaohms, a good pH meter
should have at least an input resistance of gigaohms. Even low-cost pocket pH meters
fulfil this requirement today.
The lead from the working electrode to the meter must be shielded carefully to avoid
distortion by external electric fields of the surroundings.
With ISFETs, measurement is much easier than with ISEs. There are two ways to design
the measuring circuit.
In the circuit with a constant gate voltage, the drain current ID of the ISFET is a function
of the gate voltage UG according to ID = k · UG. The current follower shown in Fig. 7.16
sets the potential of the source terminal virtually to zero (mass potential) and outputs a
voltage proportional to the gate voltage. In the circuit with a constant drain current, the
voltage value between the reference electrode and the ion-selective membrane
appears unchanged at the output, but with low impedance. This circuit is most common
with ISFET applications. In this
circuit, all the details of a sensor signal are followed and may be manipulated by
microcomputers.
Connection of ISFETs with measuring instruments. Left: constant gate voltage, Right:
constant drain current
Experimental Conditions
Results of potentiometric measurements are always activity values. For chemical analysis,
however, concentration values are desirable. In order to obtain a clear relationship
between the concentration and the measured signal, the activity coefficient should be
kept constant. This requirement can be fulfilled by establishing a high value of ionic
strength I.
The actual value of ionic strength determines activity coefficients of all ion types present
in solution. The higher it is, the more stable it is against variation by addition of foreign
components. Independent of ionic strength, the pH should be kept constant during
measurement. Also, the temperature should not vary too much. The temperature
dependence of the Nernst factor will cause an error of 2% per Kelvin for single-charged
ions.
A common practice is to mix a metered volume of the sample solution with a known
amount of a buffer solution which also sets the ionic strength at a high value.
Such buffer solutions are commercially available under the name TISAB (total ionic
strength adjustment buffer). Commonly they also contain some antioxidant (e.g.
ascorbic acid) designed to exclude negative effects of atmospheric oxygen. Useful
compositions can be found in application notes of electrode suppliers.
All conditions should be as close as possible to the sample analysis. The graph E = f
(log c) is utilized as a reference to determine the concentration cx which belongs to
the measured value EM. If potentiometric calibration graphs are not linear, then a
non-Nernstian behaviour is perceptible, which means that the electrode may produce
inaccurate results.
This figure depicts the typical shape of calibration graphs for potentiometric sensors. At
the upper limit, i.e. at very high analyte concentrations, the electrode does not obey the
Nernst law.
The resulting curve is depicted schematically in below Fig At the intercept of both
extrapolated straight lines, the corresponding signal E(intercept) might be caused by
the interferent activity aj(intercept) as well as by the sample ion activity ai(intercept).
According to the Nikolskij equation,
the following condition is valid.
Gran plot for linearizing a series
of multiple standard additions
Determination of selectivity
coefficient by consecutive additions
of sample ion to solution of
Activity values are predetermined and well known, interferent ion
hence Kij can be calculated.
mperometric Sensors
With amperometric sensors, the electrode is polarized by a predetermined potential, and
the resulting electrolysis current is measured. Under certain conditions, saturation is
achieved, and the limiting current is proportional to the actual concentration.
Under these conditions, an electric current is an expression of the reaction rate. Now the
electrode is not necessarily in thermodynamic equilibrium with the analyte. It is not
necessary to wait until the equilibrium has been established.
Consequently, commonly the response time of amperometric sensors is magnitudes
shorter than that of potentiometric sensors (Morf and de Rooij 1995). The strict
concentration proportionality of the measured quantity (the limiting current) is
another appreciated property of this sensor group.
With catalysts, one of the possible electrode reactions is preferred to such an extent
that
the electrolysis current (as an expression of the reaction rate) depends nearly
exclusively on the analyte concentration. Oxides of copper, nickel, cobalt and
ruthenium have been used as catalytic additives for carbon paste electrodes (Chen et
al. 1993).
For carbohydrates, a good selectivity was achieved with Cu2O, for amines with NiO.
Some examples of selectivity enhancement of amperometric sensors are listed in below
table.
Table Selectivity enhancement of amperometric
sensors
Permselective membranes exclude specific
molecules from transport due to their size or
other properties. With dialysis membranes, the
molecular size is critical. An important
application is the exclusion of protein
molecules which may contaminate the
electrode surface.
Lipophilic membranes block the transport of
polar molecules, and in ion-exchanger
membranes, only ions
with a defined charge sign are mobile. Best
known are Nafion membranes (Du Pont), which
act as cation exchangers in wet state. Anions
cannot penetrate
Membranes
the membrane.with steric recognition of the
analyte have properties similar to that of
permselective membranes.
As mentioned in connection with potentiometric sensors, some molecules (ligands) can
coordinate with specific analyte molecules due to their size.
ctrode Design and Examples
Amperometric sensors commonly have the shape of microelectrodes. Such electrodes
are well suited for obtaining concentrations proportional signals in situ, e.g. when a
biological sample is punctured by the electrode. Most common shapes of
microelectrodes are microdisks, interdigitated lead structures, and narrow bands on a
support.
Also widely used are macroscopic shapes with a diffusion barrier or with a protecting
layer covering the active membrane which contains immobilized molecules.
Selectivity is increased since only gaseous components can reach the active area.
Membranes of macroporous polytetrafluorethylene (PTFE) are used preferably. The
material is strongly hydrophobic. Water in liquid state is unable to permeate the
membrane pores.
Unlike liquid water, gases easily diffuse through the pores. Membranes of silicon rubber
have also been used with the Clark sensor. They are permeable for oxygen since it is
dissolved homogeneously in the matrix.
An amperometric sensor is established by imposing a constant polarization potential
and measuring
the resulting limited current below figure. Of course, the imposed potential value
should be in a rather negative region where the current is diffusion limited.
Clark sensor: effect of diffusion barrier in front of active electrode area. Left: current-
potential diagram; right: concentration profile for working potential in limiting current
region.
Among cation-exchanger membranes, the cation exchager NAFION (Du Pont) plays an
outstanding role. Its basic matrix consists of PTFE, an extremely hydrophobic polymer
insoluble in nearly all known solvents. In NAFION membranes, at the surface of a PTFE
foil, fluorine containing side chains are attached which carry sulfonic groups.
This way, hydrophilic regions arise which in wet, swollen state can bind cations by
electrostatic forces. Cations are mobile in the molecular cavities of the polymer (below
figure). A common application of NAFION membranes is exclusion of anion
interferences. The surface of the working electrode is coated by the ion exchanger. For
that purpose, a NAFION solution is spread on the surface and allowed to dry.
Current load of the counter electrode is low and can be carried by usual reference
electrodes. Rather than using a three-electrode circuit, the variable voltage is simply
imposed between reference and working electrodes. Measurement of the current
response is a technical problem, however. Very low current is distorted by external
effects.
The current follower keeps the working electrode potential equal to zero (ground
potential), and it generates an output voltage proportional to the actual electrolysis
current. With low-current circuits, damping and filtering units are necessary and must
be adjusted carefully.
In commercially available potentiostats, specific additional electronic modules are
provided for picoampere measurements. For classic investigations, such modules
would be too sluggish. Also, additional preamplifier circuits based on special low-noise
amplifiers have been proposed which would be used prior to the input of a common
potentiostat (Fig. 7.26). In the input stage (OA1 in the figure), specific low-noise
operational amplifiers with high-input impedance are used.
Even the variation of the dielectric constant in a layer will affect the overall
impedance. If impedimetry is designed only to acquire data corresponding to ionic
properties or value of the dielectric constant, it is not really an electrochemical
method, in a strict sense.
Among the class of impedimetric sensors are real electrochemical sensors where a
chemical reaction is the source of information.
Electrochemical Biosensors
A biosensor is an analytical device which is used to determine the presence and
concentration of a specific substance in a biological analyte.
Output signals
Input Biosensor
signals
WHAT IS A NANOBIOSENSOR?
Potentiometric enzyme electrodes were the first biosensors. In such sensors,an IES
was coated by an enzyme layer acting as a biocatalyst for reaction of a specific
substance. The reaction product subsequently was detected by the ISE. This feature
had been transferred soon to ISFETs.
A special term, ENFET, has even been informally proposed for enzyme-modified
ISFETs. Attempts
were also made to utilize further biological interactions for recognition of analytes and
construction of potentiometric biosensors.
Immunologic sensors on the basis of antigen-antibody reaction may be called IMFETs
if they are
built on top of a MOSFET. Immunologic reactions, however, may be used much more
efficiently in combination with other transducers besides potentiometric ones. To this
day, potentiometric biosensors remain unique to a few enzymatic reactions.
Among the oldest biosensors is a urea sensor (Guilbault and Montalvo 1969). As
depicted in Fig.below, the enzyme urease is fixed in a hydrogel layer of polyacrylamide
which is coated on the surface of a glass electrode with nylon gauze as support. Urea
diffuses into the gel, where a catalysed hydrolysis takes place:
Immunosensors are bioaffinity sensors, i.e. antigen and antibody form a stable complex,
but there is no decomposition of reactants or formation of further reaction products.
For immobilizing the isolated and cleaned antibody, methods similar to those
described in can be used as. There is, however, a peculiarity. Since the most
common type of antibody molecules is Y-shaped, the molecules must assume a
specific orientation at the sensor surface. The antigen-binding sites should be
oriented towards the sample solution (below Fig.).
Living organisms have always been used to observe changes in the environment, i.e.
as biomonitors. Biosensors result from combining organisms with transducers in such
a way that measurable signals are obtained.
Due to these properties, they should be interesting receptors for chemical sensors.
On the other hand, it would be useful to have analytical sensors for nucleic acid
determination. The most important sensors in connection with nucleic acids are
hybridization sensors.
These allow one to search in solution for a specific DNA type and to identify in this way
an individual biological species, a group of individuals or even a single individual.
Hybridization sensors implement a specific form of the genetic fingerprint.
Immobilization of DNA. Nucleic acids can be immobilized by means of methods equal
to that used for protein molecules, i.e. adsorption, crosslinking, inclusion in gels or
polymers, covalent chemical bonding with formation of SAMs, and finally, also the
avidine-biotin complex formation. SAMs with nucleic acids have also been used on
silicon, where patterns have been generated by photolithography.
By
Dr.K.Venkateswara Rao
Associate professor of Nanotechnology
JNT University Hyderabad
THERMISTORS
(THERMAL RESISTORS)
• A thermistor is a type of resistor used to measure temperature changes,
relying on the change in its resistance with changing temperature. Thermistor
is a combination of the words thermal and resistor. The Thermistor was
invented by Samuel Ruben in 1930, and has U.S. Patent #2,021,491.
• Source: http://en.wikipedia.org/wiki/Thermistor
Thermistors:
• Source: http://en.wikipedia.org/wiki/Thermistor
Thermistors:
Source: http://www.science.uwaterloo.ca/~jddmarti/p352_w2007/Thermistor_50K.pdf
Thermistors:
Source: http://www.science.uwaterloo.ca/~jddmarti/p352_w2007/Thermistor_50K.pdf
Thermistors:
Example Applications:
1. Temperature measurement.
2. Time delay (self heating from large current ‘opens’ the thermistor so it can be
used as a slow switch). Heating = i2 R where R is the resistance and i is the
current.
3. Surge suppression when a circuit is first energized. Current needs to flow through
the thermistor for awhile to heat it so that it ‘opens’, and acts again as a switch.