Professional Documents
Culture Documents
7 Autoclave-Validation Gerard-Sheridan PDF
7 Autoclave-Validation Gerard-Sheridan PDF
Gerard Sheridan
• EU GMP
• HTM 2010
• EN 285
• ISO 17665
• EU GMP , Annex 1
• Currency?
• EN 285
“STERILIZATION - STEAM STERILIZERS –
LARGE STERILIZERS”
• Scope
Does not specify requirements for the validation and routine
control of sterilization by moist heat
• ISO 17665
“Sterilization of health care products - Moist heat - Part 1:
Requirements for the development, validation and routine control
of a sterilization process for medical devices”
• Scope
Medical devices
NOTE: “Although the scope of this part of ISO 17665 is
limited to medical devices, it specifies requirements and
provides guidance that may be applicable to other health
care products”
• Uniform understanding
• Misunderstandings = Deficiencies!
• Sterilization temperature
• Minimum acceptable temperature of the sterilization temperature band
Therefore : Minimum temperature on which the evaluation of the sterilisation
efficacy is based
• NOTE: The temperature setting on the automatic controller will not generally
be the sterilization temperature, but a higher temperature within the
sterilization temperature band.
• Chamber temperature
• The lowest temperature prevailing in the chamber.
• Equilibration time
• The period which elapses between the attainment of the sterilization temperature in
the chamber and the attainment of the sterilization temperature in all parts of the load
Note: EU GMP, Annex 1, Paragraph 92:
Sufficient time must be allowed for the whole of the load to reach the
required temperature before measurement of the sterilising time-period
is commenced. This time must be determined for each type of load to
be processed.
• Plateau period
• The equilibration time plus the holding time
• A common misunderstanding
• Thermometric test for a small load
during the plateau period the temperature measured above the test pack
does not exceed the temperature measured in the active chamber
discharge by more than 5ºC for the first 60 s and 2ºC for the remaining
period.
(Ref. HTM 2010 Part 3; 13.14 (b))
• Steam Quality
• Air Removal
• Load Configuration
• Steam Quality
• Key properties to be control
Dryness
>90% for normal porous loads
>95% for metal loads.
Superheat
<25oC of superheat
Non Condensable Gases (NCG)
<3.5% NCG
Condensate Quality
WFI quality when condensed
• Air Removal
Whilst a steam penetration test carried out on a periodic basis provides a very
useful equipment control function, provision should be made to ensure
adequate steam penetration occurs during every cycle.
• Air Detector
• Will detect:
Poor vacuum performance
High chamber leak rates
High levels of NCG
• Various types
Temperature type
Pressure type
Electronic Temperature/ Pressure correlation (Not appropriate)
• Load Configuration
• Consider:
Condensate collection
Occluded surfaces
• IQ, OQ
• ISO 17665
Generally references EN 285 in terms of tests to be performed.
• Commissioning
• HTM 2010
Tests defined within the document
Reference : HTM 2010 Part 3, Chapter 4 ‘Schedule of validation tests’
Table 2a ‘Validation tests for porous load sterilizers’ and
For each test specified the tables reference the relevant sections in
which they are detailed.
Microbiological tests
Microbiological tests are designed to show that sterilization
conditions are attained where thermometric methods
alone are not sufficient.
• HTM 2010
• In sterilizers with an active air removal system thermometric tests for a
‘small load’ are designed to demonstrate that the sterilizer is capable of
removing air from a small load in which air from a near-empty chamber has
been retained.
• ISO 17665 -2
• The validity of air dilution and steam penetration indirectly identified from
the performance tests specified for installation qualification and operational
qualification should be verified to be effective for a worst case sterilization
load, loading configuration……….
• EN 285
• The test loads described in this European Standard are selected to represent
the majority of loads…… for the evaluation of general purpose steam
sterilizer for medical devices. However, specific loads (e.g. heavy metal
objects or long and/or narrow lumen) will require the use of other test loads.
• Acceptance criteria
• During the plateau period the temperature measured above the load does not
exceed the temperature measured in the active chamber discharge by more than
5ºC for the first 60 s and 2ºC for the remaining period;
• The equilibration time determined from the measured temperatures does not
exceed 15 seconds for chambers up to 800 l and 30 seconds for larger chambers;
• The holding time determined from the measured temperatures is not less than
that specified for the sterilisation temperature;
• Acceptance criteria
• During the holding time the temperatures measured in the active chamber
discharge and in the load:
• are within the appropriate sterilization temperature band
• do not fluctuate by more than ±1ºC;
NOTE: Incorrectly indicated as ±0.5ºC in some printed versions
• do not differ from one another by more than 2ºC;
HTM 2010
• Microbiological testing may be necessary for any sterilizer where loading
conditions cannot be validated solely by thermometric methods.
ISO 17665 -2
• Data from which judgement is to be made should be established from
temperature measurements supplemented by chemical indicators and/or
biological indicators positioned in difficult-to-sterilize locations.
• HTM 2010
The F0 method may be used for assessment, or control, of processes where
difference in temperature is the only factor influencing the efficacy of the
cycle. For example, it may be applied to the steam sterilization of aqueous fluids
in sealed containers but it is not applicable to steam sterilization of porous
loads where air removal is also a key factor and failure to achieve direct contact
with Dry Saturated Steam can lead to failure, regardless of whether the required
temperature was achieved within the load.
• ISO 17665-2
If a medical device is wrapped and/or non-condensable gas can be trapped in
a part such as a lumen, tubing or crevice, F0 calculated from chamber
temperatures will not represent the lethality delivered and should not be
used to judge the results of a sterilization process……..In addition to the
measurement of process parameters, steam penetration should be assured for
each operating cycle, for example, by using an air detector fitted
to the sterilizer
• Periodic tests
• Appropriate schedule:
Reference HTM 2010 Part 3, Chapter 5,
Table 4a ‘Periodic tests for porous load sterilizers’
• Controls
• Change control
• Controls
• The materials in which parts are wrapped and the manner in which they
are wrapped