Table of Contents

INSTRUCTION MANUAL I. Introduction…………………………................. 03

DUAL DISPLAY L C R METER II. Measuring Principle Introduction………… 04 ~ 07


III. Panel Illustration………................................ 08 ~ 10
Model : DE-5000 IV. LCD Display Illustration………………..……… 11 ~ 12
V. Operation Instructions………………..……….. 13 ~ 34
Thank you for purchasing our products. Please read this
1. Power ON/OFF …………………….…………. 13 ~ 14
instruction manual before using the meter and keep it
2. CAL ……………………………………………. 15 ~ 17
properly for contingent use. 3. LCR AUTO …………………………………….. 18 ~ 20
4. Inductance / Capacitance / Resistance
Measurement…………………………………. 20 ~ 21
5. FREQ ………………………………………….. 22 ~ 23
6. Backlight ………………………………………. 23
7. SORTING / SETUP / ENTER ……………..... 23 ~ 25
8. PC ……………………………………………… 26
9. D/Q/ESR/ș …………………………………….. 27 ~ 28
10. SER/PAL ………………………………………. 29
11. REL ……………………………………………. 30 ~ 31
12. HOLD …………………………………………... 32
VI. Replacing Batteries……………..……………… 33
VII. General Specification……………..…………… 34
VIII. Electrical Specification……………..…………. 35 ~36

-1-
Electrical Symbol I. Introduction
Risk of Danger. Important information.
This LCR Meter is a 19999/ 9999 counts dual display, high
See Manual. accuracy LCR meter, which could measure Inductance/
AC/DC power adaptor input polarity Capacitance/ Resistance with secondary parameters including
dissipation factor (D), quality factor (Q), phase angle (ș),
Current battery capacity indicator equivalent series/ parallel resistance (ESR or Rp).
This LCR Meter is fully auto ranging operation for AC
impedance & DC resistance measurement. The user could
measure the L/C/R components directly in “AUTO-LCR
Warning “ smart mode without selecting the function key.
Ɣ Discharge test device before measuring. Components could be measured in series or parallel mode
Ɣ Do not open the case of meter during measurement. according to the DUT (device under test) impedance
Ɣ After measurement, power off the meter. As there will be automatically.
slight power consumption under auto power off mode. User could select the desired test frequencies of
Ɣ Take the batteries out of the meter if it will be left idle for a 100Hz/120Hz/1kHz/10kHz /100kHz.
long time.
Ɣ Do not use organic solvent to clean the meter. Wipe it with The “SORTING ” mode could help the user to make a quick
a soft cloth, if necessary. sort for a bunch of components.
Ɣ means battery capacity is full. Once is
displayed which indicates power is low for operation and The LCD display with the backlight for easy reading in dark
please replace battery immediately to secure the accuracy places.
in specification.
With a 9V battery operation is standard for the meter, but a DC
9V adaptor can also be used as an optional power input.
The test data can be transferred to PC through an optional full
isolated optical USB-IR interface.

-2- -3-
II. Measuring Principle Introduction If ș > 0, the reactance is inductive. In other words, if ș < 0, the
reactance is capacitive.
There are two types for reactance. The one is the inductive
Impedance parameter introduction reactance XL and the other is the capacitive reactance XC.

䎷䏋䏈䎃 䏌䏐䏓䏈䏇䏄䏑䏆䏈䎃 䏆䏒䏑䏖䏌䏖䏗䏖䎃 䏒䏉䎃 䏕䏈䏖䏌䏖䏗䏄䏑䏆䏈䎃 䎋䏕䏈䏄䏏䎃 䏓䏄䏕䏗䎌䎃 䏄䏑䏇䎃 䏕䏈䏄䏆䏗䏄䏑䏆䏈䎃 䎷䏋䏈䏜䎃䏆䏒䏘䏏䏇䎃䏅䏈䎃䏇䏈䏉䏌䏑䏈䏇䎃䏄䏖䎝䎃 䎃
䎋䏌䏐䏄䏊䏌䏑䏄䏕䏜䎃 䏓䏄䏕䏗䎌䎑䎃 䎩䏒䏕䎃 䏈䏛䏄䏐䏓䏏䏈䎏䎃 䎽䏖䎃 䏕䏈䏓䏕䏈䏖䏈䏑䏗䏖䎃 䏗䏋䏈䎃 䏌䏐䏓䏈䏇䏄䏑䏆䏈䎃 䏌䏑䎃 䎋䏉䎃䎠䎃䏖䏌䏊䏑䏄䏏䎃䏉䏕䏈䏔䏘䏈䏑䏆䏜䎌䎃 䎃
䏖䏈䏕䏌䏈䏖䎃䏐䏒䏇䏈䎑䎃䎽䏖䎃䏆䏄䏑䎃䏅䏈䎃䏇䏈䏉䏌䏑䏈䏇䎃䏄䎃䏆䏒䏐䏅䏌䏑䏄䏗䏌䏒䏑䎃䏒䏉䎃䏕䏈䏖䏌䏖䏗䏄䏑䏆䏈䎃䎵䏖䎃䏄䏑䏇䎃 XL = 2 ʌf L䎃 䎃 䎃 䎃 䎃 䎋䎯䎃䎠䎃䎬䏑䏇䏘䏆䏗䏄䏑䏆䏈䎌䎃 䎃
䏕䏈䏄䏆䏗䏄䏑䏆䏈䎃䎻䏖䎑䎃䎬䏗䎃䏄䏏䏖䏒䎃䏆䏒䏘䏏䏇䎃䏅䏈䎃䏇䏈䏉䏌䏑䏈䏇䎃䏄䏖䎃䏄䎃䏟䎽䏟䎃䎋䏟䎽䏟䎃䎠䎃 䎵䏖䎕䎃䎎䎃䎻䏖䎕䎃 䎌䎃 XC =1/ (2 ʌf C)䎃 䎃 䎋䎦䎃䎠䎃䎦䏄䏓䏄䏆䏌䏗䏄䏑䏆䏈䎌䎃 䎃
䏒䏉䎃䏐䏄䏊䏑䏌䏗䏘䏇䏈䎃䏚䏌䏗䏋䎃䏄䎃䏓䏋䏄䏖䏈䎃䏄䏑䏊䏏䏈䎃䕠䎑䎃

Measurement mode
The impedance could be measured in series or parallel
mode. The impedance Z in parallel mode could be represented
as reciprocal of admittance Y. The admittance could be defined
as Y = G + jB. The G is the conductance and the B is the
susceptance.

Y = 1/Z = 1/Rp + 1/jXp = G+jB


Rs: Resistance in series mode Rp: Resistance in parallel mode
Zs = Rs + jXs or |Zs|Иș Xs: Reactance in series mode Xp: Reactance in parallel mode
Rs = |Zs| cosș Cs: Capacitance in series mode Cp: Capacitance in parallel mode
Xs = |Zs| sinș Ls: Inductance in series mode Lp: Inductance in parallel mode
Xs/Rs = tanș
ș = tan-1(Xs/Rs)

-4- -5-
There are two factors to provide the ratio of real part and ZM is defined as total impedance measured to DUT
imaginary part. Usually the quality factor Q is used for ( device under test ) by the special test fixture which has some
inductance measurement and the dissipation factor D is used parasitic impedance.
for capacitance measurement. D factor is defined as a ZM = (Rs + jȦLs) + ((Go+jȦCo)-1 || ZDUT)
reciprocal of Q factor.
ZDUT is the target impedance user wants to realize. It is
Q = 1 / D = tanș necessary to use the open/short calibration process to cancel
Q = Xs / Rs = 2ʌf Ls / Rs = 1 / 2ʌf Cs Rs the effect of Rs+jȦLs and Go+jȦCo.
Q = B / G = Rp / | Xp | = Rp / 2ʌf Lp = 2ʌf Cp Rp

Actually, Rs and Rp are existed in the equivalent circuit of


capacitor or inductor. If the capacitor is small, Rp is more Equivalent circuit
important than Rs. If capacitor is large, the Rs is more
important also. Therefore, use parallel mode to measure lower
value capacitor and use series mode to measure higher value
capacitor. For inductor, the impedance relationship is different
from capacitor. If the inductor is small, Rp is almost no effect. If
inductor is large, the Rs is no effect also. Therefore, use series
mode to measure lower value inductor and use parallel mode to
measure higher value inductor.

Open/short calibration
process to get the better accuracy for high/low impedance
measurement. The purpose of open/short calibration is to
reduce the parasitic effect of the test fixture.

-6- -7-
ijIJįġ ŕōĮijIJ Alligator
Panel Illustration Test Lead Case

ijijįġŕōĮijij SMD
tweezers (optional)

Rear

ijĴįġŕōĮijĴ Guard Line

Front panel Side


“GUARD” provides shield for DUT ( device under test ),
improvised test leads or equipments situated at places where
with high interference. IR to USB case
(optional)

-8- -9-
III. LCD Display Illustration
1. LCD display
2. POWER To power on/off the instrument
LCR auto mode, Inductance,
3. LCR AUTO Capacitance, Resistance and DC
resistance measurement selection key
4. FREQ Testing frequency selection key
5. Backlight display
6. SORTING Sorting mode control key
7. PC UART output control
8. CAL Open/Short calibration mode
9. D/Q/ ESR/ș D/Q/ ESR/ș parameters selection key
Setup menu control key LCD Display.
10. SETUP (in sorting mode )
11. SER/PAL Series and Parallel selection key 1. Sorting Sorting function is enabled
Setup menu control key Tolerance indication in sorting mode : ±0.25%, ±0.5%,
12. ENTER (in sorting mode ) 2. Tol
± 1%, ±2%, ±5%, ±10%, ±20%, & +80%-20%
13. REL% Relative mode Testing frequency indication –
3. kHz
14. HOLD Data hold 1kHz,10kHz,100kHz,100Hz & 120Hz
15. Input sockets and Terminals 4. PC Communication is active
16. AC/DC power adaptor 5. Battery capacity indication
17. Battery cover Range selection is enabled on setup menu of sorting
6. Range mode
18. Tilt-Stand
7. Auto Auto range for L, C or R measurement
19. IR to USB Slot
8. LCR Checking for L/C/R mode automatically
20. IR to USB case (optional)
9. ǻ Relative function is enabled
21. TL-21 Alligator Test Lead Case
22. TL-22 SMD tweezers (optional) 10. Ls/Lp Inductance in series or parallel mode is active
23. TL-23 Guard Line 11. Cs/Cp Capacitance in series or parallel mode is active
12. Rs/Rp AC Resistance in series or parallel mode is active
13. DCR DC resistance mode is selected

- 10 - - 11 -
Dissipation factor, Quality factor or Phase angle is
14. D/Q/ș active for L/C measurement mode IV. Operation Instructions
15. Rp AC Resistance in parallel mode is active To access optimum precision for all L, C and R measurements
especially at the highest and the lowest ranges, please use the
16. Cal Open/Short calibration mode
alligator test leads (TL-21) or improvised test leads applying to the
17. HOLD Data Hold measurement for OPEN and SHORT calibration (refer to page
18. APO Auto power off mode 15~17) to reduce the parasitic effect of the test fixture before going
19. ESR Series equivalent resistance mode forward the measurement.
20. Secondary Display To secure the accuracy in specification, please only connect DUT
21. ° Phase angle unit (device under test) to measuring socket or use TL-21 (standard
Unit for Resistance (ȍ, kȍ and Mȍ ) accessory) or TL-22 (optional accessory) for measurement.
22. Mkȍ – on secondary display If use improvised test lead for measurement please prevent using
Unit for Capacitance ( pF, nF, ȝF and mF ) long wire which may have incorrect result.
23. – on secondary display
Unit for Inductance ( ȝH, mH and H)
24. – on secondary display 1. Power ON/OFF
The percentage display in relative mode
25. % – on secondary display y When power on, all symbols will display for 2 seconds.
26. Primary Display
Unit for Resistance ( ȍ, kȍ and Mȍ )
27. Mkȍ – on primary display
Unit for Inductance ( ȝH,mH and H)
28. – on primary display
Unit for Capacitance ( pF,nF,ȝF and mF )
29. – on primary display
30. Bar-graph display y In Auto Power off mode, APO will be shown on the display. It
will power off automatically, if without any key operation or
y Special Indication Characters measurement for 5 min. Buzzer will beep three times to
remind auto power off is ready to activate and then “OFF”
ġ Indicates short calibration display to power off.

ġ Indicates open calibration

- 12 - - 13 -
2. CAL
y This function enables the LCR meter’s internal parameters
and external connector residues to be calibrated to have
better and precise measurements.
y To make better precision measurements on extremely high
or low impedance ranges of L,C,R, it is highly recommended
that do OPEN/SHORT calibration to reduce the parasitic
y The default setting is LCR auto mode at test frequency 1kHz. effect of the test fixture before measuring.
y Note: Remove any leads or DUT ( device under test ) from
y Battery condition is detected every second and displayed. the meter during this procedure. If leaving them connected,
suggests battery capacity is full suggests power is
low and needs to be replaced a new battery. will add impedance to the circuit causing the calibration to
fail indicated by FAIL showing on the display.
y If the key functions, the buzzer beeps once when key is
pressed. If the key has no function, the buzzer beeps twice Ex. Operation for open and Ex. Operation for open and short
when key is pressed. short calibration with TL-21 calibration with .TL-22

Using adaptor as power supply,


Auto Power off function disables
automatically and “APO”
disappears.

y Auto power off function is disabled automatically, when


using a 9V adaptor as the alternative power supply.
Open Cal. Short Cal.

Short Cal.
Using 9V- battery as power
Open Cal.
supply, “APO” displays.
y Connect TL-21 (or improved test leads applied to the
measurements) and press “CAL” key > 2 seconds to enter
the calibration mode ĺ OPEN ready ĺ OPEN calibration ĺ
y Auto power off function enables when using a 9V-battery as SHORT ready ĺ SHORT calibration. Follow below
power supply. procedure to complete the calibration.
y When start the open or short calibration, it will have 30
seconds countdown shown on the display.

- 14 - - 15 -
y If calibration is completed, PASS or FAIL appears on the
primary display.
y If both calibrations are passed, press ”CAL” key again to Enter short calibration mode.
store the calibration value to EEPROM.(Electrically Erasable * Short alligator test leads.
Programmable Read-Only Memory).
Press “ CAL ” key

Display the measurement reading


Start the short calibration with
30-seconds countdown

Press “ CAL ” key for 2sec.

SHORT calibration is completed


Enter to open calibration mode PASS or FAIL will appear on the
primary display
* If FAIL, run the procedure from
the first step again.
Press “ CAL ” key
Press “ CAL ” key

Start the open calibration with


30-seconds countdown Calibration is completed

Return to testing mode.

OPEN calibration is completed.


Pass or Fail will appear on the When open or short calibration
primary display fails. “FAIL” will appear on the
* If FAIL, run the procedure from primary display.
the first step again.
Press “ CAL ” key y If fails in calibration, please double check the DUT( device
under test ) is in the right calibration position and resume the
open and short calibration procedures.

- 16 - - 17 -
3. LCR AUTO
y The “LCR AUTO” key selects the primary parameter
measurement function. Each press of the key will select Enter Auto R mode
either Auto-LCR mode, Auto-L mode, Auto-C mode, Auto-R
mode or Auto-DCR mode *1.
y The default setting is LCR Auto *1 mode which can check the Press “LCR AUTO ” key
type of Impedance smartly and enter to the measurement
function automatically.
Enter DCR mode

In LCR AUTO test mode Press “LCR AUTO ” key

Press “LCR AUTO ” key Back to LCR AUTO test mode

Enter Auto L mode y When Auto-L or Auto-C mode is selected, the impedance
measurement is auto ranging.
y The primary display will show the parameter of inductance,
Press “LCR AUTO ” key capacitance or resistance of DUT ( device under test ).
y The secondary display will show the quality or dissipation
factor of DUT ( device under test ).
Enter Auto C mode y The secondary parameter will be based on the L/C/R
measurement. It means that (L + Q), (C + D) *2, (R + ș) *3 are
combined in one group respectively.
*1: If |Q| < 0.2, the Auto-R mode will be selected automatically. The
Press “LCR AUTO ” key parameter on sub-display is ș.
If Q Њ0.2, the Auto-L mode will be selected automatically. The
parameter on sub-display is Q.
If Q Њ-0.2, Auto-C mode will be selected automatically. The
parameter on sub-display is D.

- 18 - - 19 -
If C < 5pF. The parameter on sub-display is Rp.
*2: When LCR Auto mode is active, the secondary parameter will show c. To measure SMD component by SMD tweezers (TL-22,
the equivalent resistance in parallel mode (Rp) to replace the D option).
factor if the C value of DUT is less than 5pF.
*3: LCR AUTO mode only. In Auto-R mode or DCR mode, the
secondary parameter is not available.

4. Inductance / Capacitance / Resistance Measurement

y Power on the meter, the default testing mode is LCR Auto If required, Guard line (TL-23) can
mode that can enter to the L/C/R measurement provide shield for DUT (device
automatically. under test ) preventing from
interference when measuring high
y Each press of the key will select either Auto-LCR mode, impedance component
Auto-L mode, Auto-C mode, Auto-R mode or Auto-DCR
mode.
a. Insert DIP component leads to
the sockets directly. or
b. To measure DIP and SMD
component by Alligator Test
Lead Case (TL-21)

SMD Tweezers
(TL-22)

ġ
ɏʳ To make better precision measurements on extremely high or
If required, Guard line (TL-23) can Alligator Test Lead low impedance ranges of L,C,R, it is highly recommended
provide shield for DUT (device Case (TL-21) that do OPEN/SHORT calibration before testing. ġ
under test ) preventing from
interference when measuring high Warning : To avoid electric shock, please discharge the capacitor
impedance component before testing.

- 20 - - 21 -
5. FREQ
y The “FREQ” key selects the testing frequency. Each press of
the key will select either 1kHz, 10 kHz, 100kHz, 100Hz or The test frequency is 120Hz.
120Hz, total 5 frequency for selection.
y The default frequency setting is 1kHz.
Press “FREQ ” key

The testing frequency is 1kHz. Back to test frequency 1 kHz

Press “FREQ ” key


y The LCR impedance scale ranges are based on the test
frequency.
The testing frequency is 10kHz.
6. Backlight
y Press“ ”key to toggle backlight on/off.
Press “FREQ ” key
y The backlight will be disabled in 60 seconds automatically from
the last operation of pushing any key or measurement.
The test frequency is 100kHz.
7. SORTING / SETUP / ENTER
y These key functions are disabled in LCR AUTO mode.
Press “FREQ ” key
y Press “LCR AUTO ” key to select primary parameter
measurement function.
The test frequency is 100Hz. y In measurement (connect DUT), press “ SORTING ” key to
enter the sorting mode which will be set to 2000 digits display
automatically. If the reading is OL or less than 200 counts, the
sorting function is disabled.
Press “FREQ ” key

- 22 - - 23 -
y The primary display will show PASS or FAIL which based on
whether the impedance measured exceeds tolerance range. Last digit will flash
The secondary display will show the measurement reading. Press “Ż / Ź” key to select the
desired digit.
y When sorting mode is active, operate “ SETUP ” key along with Press “Ÿ / ź” key to adjust value.
“Ż / Ź” key, “Ÿ / ź” key and “Enter” key to set the reference
value, range and tolerance. Press “ ENTER ” key
y When the setup is finished, press “ ENTER ” key to confirm. “Tol ±” will flash.
Press“Ż / Ź”key to select Tol.
Figure : Setting up the reference value for sorting value : Each press of the key will
select either
±1%,±2%,±5%,±10%,
Display the measurement reading ±20%, -20%+80%, ±0.25% or
±0.5%.
Press “ ENTER ” key
Press “ SORTING ” key

Setup is completed.
The measurement reading is
shown on the secondary display.

Press “ SETUP ” key Ɣ If DUT ( device under test ) is


“Range” will appear and flash. in the specified Tol % , buzzer
Push “Ż / Ź” key to select the beeps once and the primary
position of decimal point and units display will show “PASS”.
based on the components you
measure
Press “ ENTER ” key y If DUT ( device under test ) is out
of the specified Tol %, the
primary display will show
“ FAIL ”.

- 24 - - 25 -
8. PC IR to USB case 9. D/Q/ESR/ș
(IR to USB case is optionel. Measuring
data can be transferred to PC with it. ) USB socket y In LCR Auto mode, this key has no function

y Snap on IR to USB case and connect y The “D/Q/ESR/ș ”key selects the secondary parameter
with USB cable to PC. measurement function. Each press of the key will select
either dissipation (D), quality (Q), ESR or phase angel (ș).
y Press “ PC ” key to start data
transmission, and the “PC” will appear on Example : In Cs testing mode
display. Press the “ PC ” key again to
cancel the transmission.
Secondary display parameter is
D.

Display the measurement reading


Press “ D/Q/ESR/ș ” key

Press “ PC ” key
Secondary display parameter is
Q.
PC communication is active.
“PC” will appear on the display
Press “ D/Q/ESR/ș ” key

Press “ PC ” key
Secondary display parameter is
ESR.
“PC” disappear on the display
when PC communication is
inactive
Press “ D/Q/ESR/ș ” key

- 26 - - 27 -
10. SER/PAL
y In LCR Auto mode, this key has no function.
Secondary display parameter is y Press the “SER/PAL” key to select parallel or series mode,.
ș. y When any one of primary parameter measurement function
is selected, parallel or series mode will be selected
automatically based on the total equivalent impedance
measured.
Press “ D/Q/ESR/ș ” key
y If the impedance is more than 100kȍ, will be in the parallel
mode automatically. Lp, Cp or Rp appears on the display.
y If the impedance is less than 100kȍ, will be in the series
Back to parameter D mode automatically. Ls, Cs or Rs appears on the display.
measurement Example : In Cs and Cp testing mode

When measure capacitance in


series mode.
Ɣ D/Q/ESR/ș setting is not applied to resistance
measurement- Auto-R mode and DCR mode. “Cs” will appear.

Press “SER/PAL ” key


In Auto-R mode, no parameter
When measure capacitance In
measurement on the secondary
Parallel mode.
display.
“Cp” is shown.

Press “SER/PAL ” key

In DCR mode, no parameter Back to Series mode.


measurement on the secondary “Cs” is shown.
display.

y The annunciators for Ls/Lp/Cs/Cp/Rs/Rp will be indicated based


on LCR measurement mode.

- 28 - - 29 -
11. REL Remove the current DUT
( device under test ) and insert
y In LCR Auto mode, this key has no function. another one
y Press the “REL” key to enter relative mode. The value on the The new reading is shown on the
display will be stored as reference value and “ǻ” annunciator primary display.
will appear. ***% is shown on the secondary
y The secondary display will show the percentage of reference display.
value for all subsequent measurements (All subsequent measurements
will be displayed as a percentage
value to the stored value )

Display the measurement Press “REL” key again


reading
Ex. 669.3 nF
“ǻ” flash
The reference value display on
Press “REL” key primary display.

“ǻ” appear
The reading on the display is
stored as reference value.
0.0% is shown on the secondary y REL% = (DCUR – DREF) / DREF * 100%.
display( since the measured
value and the reference are the *DCUR = Current DUT(device under test), DREF = Reference
same at this point ) DUT (device under test )
Remove the current DUT y The percentage range is from -99.9% to 99.9%. When the
( device under test ) and insert new measurement is double than reference value. OL% will
another one be shown on the secondary display.
y Press “ REL ” key for more than 2 sec. to exit Relative Mode.
The new reading is shown on the
primary display.
Minus ***% is shown on the
secondary display.

- 30 - - 31 -
12. HOLD
V. Replacing Batteries
y Press the “HOLD” key to freeze the reading on the primary
display and “HOLD” will appear. The meter is powered by a single 9V battery, with NEDA1604 ,
y In the Hold mode, only key and PC key have function. JIS006P and IEC6F22 carbon-zinc or alkaline battery. Alkaline
battery is recommended.
When replaces a new battery, the sign of ” “ displays. To
ensure the measuring data is within spec, it is suggested to replace
“HOLD” is shown. battery immediately once ” “ displays.

Warning
y Press the “HOLD” key again, to exit the Hold mode z Switch off the meter. Remove all test leads and external adaptor
before opening the case.
z Be sure to install the new batteries with correct polarities.

“Hold” disappears
1. Remove tilt-stand.
2. Loosen 4 screws on battery
cover with suitable
screwdriver.
3. Then take up the cover.
4. Replace the specified battery.
5. Reverse the procedure of
step2 and step1.

- 32 - - 33 -
VII. Electrical Specification
VI. General Specification Accuracy: ±(% of reading + number of least significant digits) at 23°C ± 5°C, <75% R.H.
Item Dual Display L C R Meter 1. Specifications are based on measurement performed at input sockets or terminals with Alligator
Test Lead Case (TL-21) after short & open calibration. (refer to page. 8~10 item 15)
Ls/ Lp/ Cs/ Cp/ Rs/ R p/ D/ Q/ Ĭ/ ESR
Parameters Measured 2. DUT ( device under test ) & Test leads should be properly shielded to GUARD if necessary.
L C R auto selection
3. To make better precision measurements on extremely high or low impedance ranges of L,C,R, it is
Measuring Circuit Mode Series / Parallel mode highly recommended that do OPEN/SHORT calibration before measuring for following ranges
Dual 19999/1999 display marked with *
Displays
Ranging Mode Auto Ϯʳ Resistance (Parallel / Series mode)
Range Resolution 100/120Hz 1kHz 10kHz 100kHz
Measuring Terminals 4 - wire Sockets ( Terminals ) & Guard 20.000ȍ 0.001ȍ ņ 1.0%+3* 1.0%+3* 2.0%+3*
200.00ȍ 0.01ȍ 1.0%+3 0.3%+2 0.3%+2 0.6%+3
L: 20.000ȝH ~ 2.000 KH 2.0000kȍ 0.0001kȍ 0.3%+2 0.3%+2 0.3%+2 0.6%+3
C: 200.00pF ~ 20.00mF 20.000kȍ 0.001kȍ 0.3%+2 0.3%+2 0.3%+2 0.6%+3
Auto LCR test range: 200.00kȍ 0.01kȍ 0.5%+2 0.5%+2 0.5%+2 1.0%+3
R: 20.000 ȍ ~ 200.0 Mȍ 2.0000Mȍ 0.0001Mȍ 1.0%+3 1.0%+3 1.0%+3 ņ
DCR: 200.00ȍ~200.0Mȍ (2.000Mȍ) 0.001Mȍ ņ ņ ņ 2.0%+3*
20.000Mȍ 0.001Mȍ 2.0%+3* 2.0%+3* ņ ņ
Test Frequency 100Hz/ 120Hz/ 1kHz/ 10kHz/ 100kHz (20.00Mȍ) 0.01Mȍ ņ ņ 2.0%+3* ņ
200.0Mȍ 0.1Mȍ 2.0%+3* 2.0%+3* ņ ņ
Backlight YES * Do open/short calibration before measuring for above ranges with * to have better precision
±0.25%, ±0.5%, ± 1%, ±2%, ±5%, ±10% measurements.
Tolerance mode
, ±20%, -20%+80% Ϯʳ DCR
Test Signal Level 0.5 Vrms Typ. Range Resolution Accuracy
200.00ȍ 0.01ȍ 1.0%+3*
Measuring Rate 1.2 /second, nominal 2.0000kȍ 0.0001kȍ 0.2%+2
20.000kȍ 0.001kȍ 0.2%+2
Response time Approx. 1 second/ DUT ( device under test ) 200.00kȍ 0.01kȍ 0.5%+2
Auto Power-Off 5 minutes approx. without key operation 2.0000Mȍ 0.0001Mȍ 1.0%+3
20.000Mȍ 0.001Mȍ 2.0%+3*
Temperature 200.0Mȍ 0.1Mȍ 2.0%+3*
0.15 x (spec.accur.) per °C (0-18°C, 28-50°C)
Coefficient * Do open/short calibration before measuring for above ranges with * to have better precision
measurements.
Operation Temperature 0°C to 50°C; 0-70% R.H.
Storage Temperature -20°C to +60°C; 0-80% R.H. Ϯʳ Capacitance (Parallel / Series mode)
Range Resolution 100/120Hz 1kHz 10kHz 100kHz
When 9V battery in good condition, 200.00pF 0.01pF ņ ņ 1.2%+5* 2.0%+5*
Battery voltage displays. 2000.0pF 0.1pF ņ 2.0%+3* 0.3%+2 0.6%+3
20.000nF 0.001nF 2.0%+3* 0.3%+2 0.3%+2 0.6%+3
Indication When low battery, displays. Please replace
200.00nF 0.01nF 0.3%+2 0.3%+2 0.3%+2 0.6%+3
new batteries immediately. 2000.0nF 0.1nF 0.3%+2 0.3%+2 0.6%+2 2.0%+5*
Alligator test lead case(TL-21)ΕAC/DC AdaptorΕ 20.000ȝF 0.001ȝF 0.3%+2 0.6%+2 1.2%+5* ņ
Standard Accessories (20.00ȝF) 0.01ȝF ņ ņ ņ 3.0%+5 (10ȝF max.)*
Guard Line (TL-23)ΕUser manualΕDC 9V Battery 0.01ȝF 0.6%+2 1.0%+3* ņ ņ
200.00ȝF
Option IR to USB caseΕSMD Tweezers(TL-22) (200.0ȝF) 0.1ȝF ņ ņ 3.0%+5 (100ȝF max.)* ņ
2000.0ȝF 0.1ȝF 1.0%+3* ņ ņ ņ
Dimensions (L/W/H) 188 / 95 / 52.5 mm (2000ȝF) 1ȝF ņ 1.2%+3* ņ ņ
Weight approx. 350g (LCR Meter only, excluding battery) 20.00mF 0.01mF 1.2%+3* ņ ņ ņ
ƔIf reading <2000, unit on display is pF
* Do open/short calibration before measuring for above ranges with * to have better precision
measurements.

- 34 - - 35 -
Ϯʳ Inductance (Parallel / Series mode).
Range Resolution 100/120Hz 1kHz 10kHz 100kHz Appearance and specification may be revised if needed
20.000ȝH 0.001ȝH ņ ņ ņ 2.5%+5*
200.00ȝH 0.01ȝH ņ ņ 1.2%+5* 0.6%+3 without prior notice.
2000.0ȝH 0.1ȝH ņ 2.0%+5* 0.6%+3 0.6%+3
20.000mH 0.001mH 1.2%+5* 1.0%+5 0.3%+2 0.6%+3
200.00mH 0.01mH 0.3%+2 0.6%+3 0.3%+2 1.2%+5*
2000.0mH 0.1mH 0.3%+2 0.3%+2 0.6%+3 ņ
20.000H 0.001H 0.3%+2 0.6%+3 1.2%+5* ņ
DER EE is assessed and certificated as meeting the
200.0H 0.1H 0.6%+3 1.2%+5* ņ ņ
2.000KH 0.001KH 1.2%+5* ņ ņ ņ
requirements of ISO 9001:2000. We produce meters in
ƔIf reading <2000, unit on display is ȝH
* Do open/short calibration before measuring for above ranges with * to have better precision various types and specification and all our product
measurements.
conforms to ISO standards. If any further inquiry needed,
Ϯʳ Accuracy v.s. Resistance (ZDUT)
DCR 100/120Hz 1kHz 10kHz 100kHz please visit our website at www.deree.com.tw
0.1~1ȍ 1.2%+5* 1.2%+5* 1.2%+5* 1.2%+5* 2.5%+5*
1~10ȍ 0.6%+3* 0.6%+3* 0.6%+3* 0.6%+3* 1.2%+5*
10~100kȍ 0.3%+2 0.3%+2 0.3%+2 0.3%+2 0.6%+3
100k~1Mȍ 0.6%+3 0.6%+3 0.6%+3 0.6%+3 2.5%+5*
1M~20Mȍ 1.2%+5* 1.2%+5* 1.2%+5* 2.5%+5* 100k~2Mȍ
>20Mȍ 2.5%+5* 2.5%+5* 2.5%+5* ņ ņ
* Do open/short calibration before measuring for above ranges with * to have better precision
measurements.
If D > 0.1, the accuracy should be multiplied by 1 D2
If D << 0.1 in capacitance mode ZC = 1/2ʌfC
If D << 0.1 in inductance mode ZL = 2ʌfL

Sub-display parameters accuracy


Ae = impedance (Z) accuracy
Definition: Q = 1/D & Rp = ESR * (1+1/D2)
D value accuracy De =ʳ ̈́ Ae *(1+D)
ESR accuracy Re= ̈́ ZM * Ae (ȍ)
ie., ZM = impedance calculated by 1/2ʌf C or 2ʌf L
Phase angle ș accuracy șe= ̈́ (180/ʌ) * Ae (deg)

HCA500000-00020
- 36 -