Professional Documents
Culture Documents
Selection Guide-JTBaker Reagents For Mindray Analyzers-9052 PDF
Selection Guide-JTBaker Reagents For Mindray Analyzers-9052 PDF
PERFORMANCE MATERIALS
Mindray BC-5300
tWBC + 5 part differential analysis
Application Description Product Nr. Pack size
Diluent Diluid M5 3419.9020PC 20 L
Lysing reagent CyMet MD(I) 53 2984.1000PE 1L
CyMet MD(II) 3488.0500PE 500 ml
CyMet LH 53 2985.1000PE 1L
Cleaner Proclean MX5 3498.1000PE 5L
Hematology Control BC Diff 5 Control L/N/H 3613/3614/3615 3.0 ml
64
Medonic
TM
PERFORMANCE MATERIALS
melet schloesing
Mindray BC-2000, BC-2300, Semi-Automated
tWBC + 3 part differential analysis
Application Description Product Nr. Pack size
Diluent Diluid Mindray 3439.9020PC 20 L
Lysing reagent Lyzerglobin PCE 3771 10 x 10 ml
CyMet Mindray 3441.0500PE 500 ml for BC 2300
mindray
Cleaning ProClean Extra 3862.1000 1L
3862.5000 5L
Emergency Cleaning ProClean Plus 3901 100 ml
Hypochlorite 0.5% 3917 1L
3917.5000 5L
nihon kohden
Hematology Control 8 Parameter Control L/N/H 3427/3428/3429 2.5 ml
3463/3464/3465 2.5 ml, pierceable
3701/3702/3703 4.5 ml
Hematology Control 3 Diff Control L/N/H 3433/3434/3435 2.5 ml
3820/3821/3822 4.5 ml
Diluid, CyMet and ProClean are all trademarks of Avantor Performance Materials Inc.
orphee
Cross list: Mindray reagents vs. J.T.Baker
Mindray BC-3000 H Nr. J.T.Baker JTB Nr.
Lyse A12-000044 500 ml CyMet Mindray 3441.0500PE
seac
CyMet Mindray CN Free 3440.0500PE
Diluent A12-000042 5.5L Diluid Mindray 3439.9020PC
Diluent A12-000047 20L Diluid Mindray 3439.9020PC
Rinse A12-000043 5.5L Rinse Mindray 3442.5000PE
Rinse A12-000048 20L Rinse Mindray 3442.5000PE
swelab
E-Z Cleaner A12-000045100 ml ProClean Plus 3901
Probe cleanser A12-00004617 ml Hypochlorite 3917
sysmex
M-50 LHLysing reagent A12-000221 CyMet MH CN Free 3497.0500PE
M-50 LBA Baso reagent A12-000224 CyMet MBA 3489.1000PE
M-50 LEO (I) Diff reagent A12-000219 CyMet MD(I) 3418.1000PE
M-50 LEO (II) Diff reagent A12-000220 CyMet MD(II) 3488.0500PE
M-50 Cleanser A12-000225 ProClean MX5 3498.1000PE
urit
controls
cleaners
pbs
Stains
65
TM
PERFORMANCE MATERIALS
66
Medonic
TM
PERFORMANCE MATERIALS
melet schloesing
Mindray BC-5500
• T he separation between Monocytes and Neutrophils
1. Important NOTES (see figure 2) is improved when using the JT Baker
reagents in comparison to the Mindray reagents. The
Please read these notes prior to installation of the
mindray
Monocytes are positioned closer to the Lymphocyte
J.T.Baker reagent package on the Mindray BC5500 Auto area. Mindray reagents tend to give more often
Hematology analyzer. Monocytosis flagging, where JT baker reagents tend to
give more often Abn./Atypical Lym? The Lymphocyte
• T he recipe for the CyMet MD(I) has been improved on cluster is more stretched with JT Baker reagents, but the
nihon kohden
Eosinophilia detection. Therefore these reagent total amount counted is equal.
replacement instructions are intended for use with
CyMet MD(I) lots higher than lot 0928900030 and all lots
of CyMet MBA
orphee
software versions or any hardware modification of the
BC5500 model we cannot guarantee the performance of
the reagent.
J.T. BAKER MINDRAY
• Reagents for the Mindray BC5500 are bottled in similar Figure 2. Improved Monocytes / Neutrophils separation
square bottles as the original reagents. For the Diluid J.T. Baker vs. Mindray
cubitainer we will provide our customers a special
seac
adapter to overcome the difference in height due to the Therefore we advise to set the Sensitivity for this
fixed Diluent reagent aspiration tube. Abn./Atypical Lym flag to 4 instead of 5. Go to
SETUP, Advanced, Flag Sensitivity, Edit, Abn./
• T o guarantee optimal performance of the JT Baker Atypical Lym and change the value to 4.
reagents, the priming protocol mentioned in this
swelab
handout should be followed every time you switch from • J T Baker CyMet MBA reagent causes an effect on the
original Mindray reagents to JT Baker reagents or vice position of the mononuclear and polynuclear white
versa. Especially the priming protocol for the Diluent and blood cells and Basophil population in the BASO-
the CyMet MD(I) reagents should be performed correctly scattergraph (see figure 3). The Mindray Basophil reagent
to avoid cross contamination between the reagents. is much stronger than the JT Baker Basophil reagent
resulting in different gain settings for both reagent
sysmex
• Before changing any reagents on the Mindray BC5500 packages. Please follow the Optical System Setup
instrument, run at least 5-10 fresh human samples or procedure for Basophils mentioned in this handout.
BC-Diff 5 control blood with the original Mindray
reagents for reference and calibration purposes. Figure 3. Basophil subpopulation J.T. Baker vs. Mindray
urit
separation between the mononuclear (Lymphocytes
and Monocytes) and polynuclear (Neutrophils,
Eosinophils and Basophils) cells compared to Mindray
reagents. Figure 1 shows and example of the different
mononuclear and polynuclear white blood cells
controls
2. Priming protocol
Use next cross list for correct reagent installation and usage.
cleaners
67
TM
PERFORMANCE MATERIALS
For the complete Deprime of the instrument we advise to MAS = High angle scatter (horizontal X-axis), related to
use the Pack-up function from the Overall maintenance complexity
screen. This is the best method to flush out the original LAS = Low angle scatter (vertical Y-axis), related to size
reagents, flush the system with water and prime the system
with JT Baker reagents. 7µm standard particle test (DIFF & BASO channel):
:
Pack-up protocol (see page 10-33 operator manual)
Click the “service” icon in the “Main” screen and click the 1) In the count screen select “Mode” and change to the
“Overall” button to enter the maintenance screen. Proceed as “OV-PD” mode and click “Diluent” to dispense diluent
follows: into an empty tube 3 times.
1. Click the “Pack-up” button at the “Overall” screen.
2. Click “Yes” to proceed with the pack-up. 2) Dispense 1 drop of 7µm standard particle into the
3. Remove all reagent pickup tube assemblies from their centrifugal tube and mix them.
bottles and containers when prompted on the screen.
4. Click “OK” to start the fluidic system emptying. 3) Change the mode to “OV-Whole blood-CBC+5DIFF” and
5. A progress bar pops up and shows the remaining time. test the dilution as whole blood
6. When prompted place all reagent pickup assemblies
into a container with distilled water. 4) Then go to the “Review > OpAdjust” screen and select
7. Click “OK” to start to flush the analyzer with distilled the “DIFF” channel (see figure 5).
water.
8. After this flushing process a dialog box will ask to
remove all reagent pickup assemblies from the distilled
water.
9. Remove them and place them in a clean beaker.
10. Click “OK” to start to empty the fluidic system.
11. The instrument will power off automatically.
12. Power on the instrument and wait for initialization to
finish.
13. Several alarms will sound.
14. Click the “Count” button and click on the upper alarm
box.
15. Place all reagent pickup assemblies in their correspon-
ding J.T. baker reagents.
16. Click on the “Remove error” button and reagent priming
will start for all the reagents.
17. When ready, perform at least 5 background counts. Figure 4. Optical Adjust screen (OpAdjust)5)
18. Perform the Optical System setup protocol.
19. After that perform the calibration of the instrument. 5) Click “Calculate” to check the four parameters:
Mindray reagents: AS-peak, 88 ± 4; LAS-peak, 45 ± 4
JT Baker reagents: MAS-peak, 88 ± 4; LAS-peak, 45 ± 4
3. Optical System Setup
MAS 0.1max width, =14
Next methods should be used setting up the new types of LAS 0.1max width, =14
Optical systems of the BC5500 instruments with JT Baker
reagents correctly: 6) Select the “BASO” channel and click “Calculate” to check
1. Using standard latex particles of 7µm. the two parameters:
2. Using fresh human samples. Mindray reagents: MAS-peak, 147 ± 4; LAS-peak, 58 ± 4
Both methods are perfectly described in Mindray instruction JT Baker reagents: MAS-peak, 89 ± 4; LAS-peak, 45 ± 4
materials.
7) If the MAS/LAS-peak value is out of normal range, input
3.1. I nstruction on using the standard particles of 7µm . the target value and press “>>>” to calculate the new
Gain. Please write down the new Gain; then access
For the new optical system, 7µm standard particle is used for “Setup > Gain” screen to modify the gains accordingly.
both channels (DIFF & BASO).
68
Medonic
TM
PERFORMANCE MATERIALS
melet schloesing
3.2. Instructions using fresh samples
IIt is advised to setup the instrument using standard There are several important checks to see if the MAS-peak and
latex particles. are not available, gains can be adjusted for LAS-peak gains for the BASO are set up correctly:
the DIFF and BASO channels according to experience. • There should be a clear separation between the Basophils
Experienced customers can used next quick setup method and the rest of the white blood cells.
to adjust the optical gains for DIFF and BASO channels. • If the bottom of the total WBC population is cut by a
mindray
straight line, the total WBC count is not ok and the LAS
Quick Optical Gain Setup method: PEAK gain has to be increased.
• See figure 7 for an incorrect separation between Basophils
WBC LAS DIFF J.T. Baker = WBC LAS DIFF Mindray and other White Blood Cell populations. MAS PEAK gain
WBC MAS DIFF J.T. Baker = WBC MAS DIFF Mindray has to be lowered.
nihon kohden
WBC LAS BASO J.T. Baker = WBC LAS BASO Mindray -71
WBC MAS BASO J.T. Baker = WBC MAS BASO Mindray -66
Example:
Gain Mindray J.T. Baker
WBC LAS DIFF 97 97
orphee
WBC MAS DIFF 70 70
WBC LAS BASO 161 90 (161 - 70)
WBC MAS BASO 191 125 (191 - 66)
After correct Optical Gain Setup a normal scatter graph for Figure 7. Incorrect separation Basophil population
the 4-part Differential with JT Baker reagents is shown in
seac
figure 5.
swelab
sysmex
Figure 5. WBC 4-part Differential
urit
• There should be a clear separation between
Neutrophils+Baso and Eosinophils populations.
• The DIFF scattergraph should be 3/4 or 4/5 of the DIFF
screen (LAS-peak).
• The centre of the “Neu+Baso” should be in the right middle
controls
A normal scatter graph for the Basophils with J.T. Baker reagents
is shown in figure 6.
cleaners
pbs
69
TM
PERFORMANCE MATERIALS
4. Calibration
70
About Avantor™ Performance Materials
Avantor Performance Materials manufactures and markets high-performance chemistries and materials
around the world under several respected brand names, including the J.T.Baker®, Macron Fine Chemicals™,
Rankem™, Diagnova™, BeneSphera™, and POCH™ brands.
Avantor products are used in a wide range of industries. Our biomedical and life science solutions are
used in academic, industry and quality control laboratories for research, pharmaceutical production and
medical lab testing, while our electronics solutions are used in the manufacturing of semiconductors and
flat panel displays. Based in Center Valley, Pennsylvania (USA), Avantor is owned by an affiliate of New
Mountain Capital, LLC.