WinWdw Software
Operating manual
Time Group Inc.
Table of Contents
1. Overview…………………………………………………………………………1 1.1
Functions………………………………………………………………………1 1.2 Software
and hardware requirement………………………………………1 1.3
Application……………………………………………………………………1
2. Installation and unloading of software…………………………………………2 2.1
Installation and unloading of software……………………………………2 2.2 Usage of files
in software system……………………………………………2
3. Guideline of operation…………………………………………………………2 3.1 Main
window …………………………………………………………………2 3.2 Display panel of
load and deformation………………………………………6 3.3 Display panel of stroke and
test time…………………………………………9 3.4 Curve display
panel…………………………………………………………10 3.5 Control
panel…………………………………………………………………13 3.6 Specimen
info…………………………………………………………………17 3.7 Data
analysis…………………………………………………………………18 3.8 Test report
setting and print…………………………………………………22
4. Brief introduction for test procedure…………………………………………24
5. System calibration and parameters setting……………………………………25 5.1 System
calibration……………………………………………………………25 5.2 System parameter
setting …………………………………………………26 5.3
Adjusting……………………………………………………………………26
6. Instruction for report form……………………………………………………29
7. Maintenance and troubles……………………………………………………31 7.1
Maintenance…………………………………………………………………31 7.2 Troubles
and remedy…………………………………………………………32 Appendix
WinWdw Software Operating Manual
1.3 Application
● Suitable for microcomputer controlled electronic universal testing machine
● Others
2. Installation and unloading of software
2.1 Installation
The installation of software adopts the form of automatic guidance to instruct customer to
install step by step, so it is convenient. Herewith, no matter the software is stored by the form of
soft disc or CD, if the customer clicks the corresponding disc code, and then double click the file
of setup.exe in it to start to install the software. Also sometimes offer self compression-relieving
file. During installation, the customer can choose the route to install as self-request, otherwise as
the pre-setting route.
2.2 Usage of files in software system
The system grade files are stored under the present root directory, the main files as follows: 1.
WinWdw.exe : Executable files 2. wdw.sys, wdw.soft and wdw.dat : parameter-equipped files 3.
SoftwareDog_inst.exe : Software dog installing file The files under directory of software
installation cannot be deleted and destroyed. Due to some unidentified reasons to destroy the
software, it must use installing disc to restore. In addition, some files in system, such as
statement files, program control files, etc. put in individual folder, so don’t modify or delete
these folders and files in them, otherwise it may case serious problems. After the software
running, it will generate automatically two directories: Curve and Control used to store curve
files and control program files, it is prohibited to delete these directories, otherwise system will
not find necessary files causing troubles. 3.3 Uninstall of software
Delete all the files in the WinWdw folder (or other folder you installed them) can uninstall the
software quickly and safely.
3 . Guideline of operation
3.1 Main window Main window (as Diagram 1) is the control center of the programs, which
taking charge of managing each function window and the switch of system modes and displaying
information of test status.
-2-
Diagram 1. Main window
3.1.1 Main menu Operation: including test operation, test analysis, specimen information and
demonstrating status and exit. Parameters setting: including software parameter, control
parameter and data processing parameter setting. Adjustment: calibrating software and offering
gain of amplifier for adjusting person. Other: array sub-window and observe adjust window.
About: Version information Array window: used to array observe adjust window as the form of
Pre-setting. Some of above-mentioned menus cannot find under the authority of ordinary
operators.
3.1.2 Toolbar It is made up of several striking quick buttons, which are the corresponding items
of frequent used operating menu of the software, and the functions is same as corresponding
menu item. Using these buttons can expedite operating speed.
3.1.3 Operating menu Operating menu is as Diagram 2 below.
Diagram 2. Operating menu Diagram 3. Demo test menu
Specimen info, test operating, test analysis, enroll can enter corresponding windows. Demo
status: this is an automatic switch menu. Click, the tick signal will emerge in the front, so the
system enter into demo status, the menu of “Demo test” also will appear (as Diagram 3), it can
open history test data file without the testing machine, and test playback, at the time the testing
machine don’t make actual test, the collected signal adopts file date. Once again click, the tick
will disappear, the system restore the test status. Demo test: Open file: Open the demo history
test data file
-3-
WinWdw Software Operating Manual
ÏinÏdIrLE Electrunic Universal Testing Iachine Ieasureluntrul Sjrstel
Üperatiunlífnl) ParameterEetting Üther
Tegting Prucess Specimen Save Data Control Para
Specimen infurmatiun
Data Frutas:
gTesting É
Lug In
Üpen Demo Data File
Demo Status Dama Start
WinWdw Software Operating Manual
Demo start: After opening the file, click the menu to make the demo, also can click the “start”
button in the control panel. Test data and curve can playback. Demo end: Click to end the demo.
Also can click the “start” button in the control panel. Log In: The software has multi-grade
operating authority: ordinary operator, superior operator and executive. Choose the correct enroll
identity in the enroll window, and input corresponding password, and enter after confirmation.
3.1.4 Parameters setting menu 3.1.4.1 Control parameter setting Control parameter setting
window is as Diagram 4
Diagram 4. Control parameter setting window
Stop & protection setting: set overload protection parameters (generally setting better than
3%-5% of measuring range to protect) Specimen broken setting: after the test start, the software
will judge if the specimen is broken as this setting so that stop test. Other option: setting other
options. If the deformation is larger than some certain value, please dismantle the extensometer,
etc.
3.1.4.2 Data processing parameter setting Set data processing parameter. Please refer to
corresponding content in chapter of 3.7.2. 3.1.4.3—3.1.4.4 is only suitable for superior operator
or executive to operate.
3.1.4.3 Software parameter setting
-4-
Judgement uf E reak
WinWdw Software Operating Manual
Software parameter setting window is as Diagram 5. The user can set the relevant options of
clock cycle, displaying accuracy, change step and resolution.
Diagram 5. Software parameter setting window
Displaying accuracy:
Displaying ** bit behind decimal: when choose “NIL” in the frame “Auto” under this item, this
item will work to control the displaying bit number of load and stroke on the “display panel”.
Auto: The displaying accuracy has strict and specific stipulation in the relevant test standards, so
choose this item will identify the displaying bit number according to test load step automatically.
Clock cycle: The amplifier will connect the computer and testing machine together as the hinge.
The computer, as the data acquisition system, should record the test procedure real time and
accurately. The sampling cycle is the test process “step length” recorded by the computer, which
it should be better if shorter theoretically. The sampling cycle is shorter, the collected data will
more during the unit time, the curve will be more actual, but the limited store point number
(65000), so this will cause short store date time, please consider carefully and set this parameter
correctly. For static test, we suggest adopting 60ms, and the effect is excellent attested
practically.
Change step: Setting if change step and the parameter of change step delay time (Generally
500ms or above), due to specific meaning, so no redundant statement
-5-
WinWdw Software Operating Manual
herewith.
Resolution and others: Setting the resolution of load and deformation (the proportion of the
corresponding measuring range, generally 1/2000 which is 5/10000 of standard request). In this
item, also can set the parameter of “ stop machine if the load is better that some certain value
when ordinary operator adjusts the position”.
3.1.4.4 Storage and reuse of parameters
Reuse system parameter: When one set of computer control multi-set of testing machines or
under circumstances when the system setting should be changed (For example, equipped with
multi extensometers). After the parameters are set and calibrated ready under various kinds of
circumstances, store the system parameters. The user can reuse these parameters according to
applied testing machine and extensometer.
Store system parameter: Store the current equipped system parameters as the form of file. It had
better use the specific name when input. The measuring range can be as the name of the files if
quipped with several extensometers. After calibrating, generally store the calibrated parameters
for future reference and restore.
Reuse software parameter: Its function is similar with that of reuse system parameter, but the
reused is software parameters equipped.
Store software parameter: Its function is similar with that of store system parameter, but the
stored is software parameters equipped.
3.2 Display panel of load and deformation. The display window of load and deformation is as
Diagram 6.
Diagram 6. Display window of load and deformation
3.2.1 Load measuring real time display
The window displays load real time, according to chosen step to adjust automatically, also
corresponding with load in test curve. If no special statement, the unit is kN, and the unit is N
when measuring range is less than 2kN.
About zero: Before test, when no load imposed on the system, please make the load
-6-
WinWdw Software Operating Manual