OBJECTIVE: To detect the presence of sugars. (Widely used method to monitor
column eluates for elution of sugar containing compounds such as LPS, 0-antigen, etc.)
MATERIALS:
Standards:glucose 1 mg/ml stock solution
use 10, 20, 40, 60, 80 µl and make up each sample with dH2O to a final volume of 100 µl. Blank 100 µl water Samples:50 µl (of 5 mg/ml LPS solution) and make up to 100 µl.
METHOD:
1. Add 50 ul of 80% Phenol solution (80% Phenol by weight)
2. Vortex. Add 2.0 ml concentrated Sulphuric Acid in a stream. 3. Stand 10 min. at room temperature. 4. Read absorbance an 490nm.