Order Number 1143-9 Environment Condition of Calibration
Received Date October 29,20L6 Temperature : (26+2)oC
Customer Politeknik Negeri Banjarmasin Humidity : (60 r 5) o/o

Address Jl.Brigjen H.Basry Reference : SMM-WI-F-02


Banjarmasin, Kalimantan Selatan Calibration Date : October 29,2016
Calibration Location : Politeknik Negeri Banjarmasin

Equipment Name Proving Ring


Manufacturer c.M.H. Ltd Dial Information
Type Analog Manufacturer : Borletti
Serial Number 3000-4 Capacity :10 mm
Capacity 3000 tbf Resolution : 0.01 mm
Serial Number
Calibration Reoort
Nominal
Uul lndication Calibration Factor Repeatability Unceftainty
Standard
tbf 1 Div-x 0.01 mm lbf/Div o/o o/o

0 0 0.00 0.00
300 12 25.00 0.00
500 24 25.00 0.00
900 35 25.47 0.58 t 1,0
1200 47 25.35 0.58
1500 59 25.57 1.15
1800 7t 25.47 1.15
2100 B3 25.40 1.15
7400 95 25,35 1.15

The Uncertainty is taken at a Convidence Level 95 o/o for Coverage Factor ( k ) = 2


Mean Factor Calibration ProJjtg BjnS [25-JJ-l )blf I dli;u

1,20

E loo
?so
o
s60
.9

:40
;20
0
0 1200 1800

x= Standard lndc ( lbf )

Calibrator used
Callbrator Name Manufacturer Type Serial Number Traceable to SI through
1. Load Cell KELI PST -A 508590 KIM-LIPI

Date of Issued
Calibrated by ; LHS
Typist : LHS

Director

-End of Ceftificate-