INSTRUCTION MANUAL
FOREWORD VK-143P
RELATED MANUALS
VM Series Monitor Instruction Manual
! CAUTION
1. If insulation resistance (Megger) test is made on the signal cable between transducer and
monitor, disconnect the cable from transducer and monitor.
Be sure to discharge the charged electric load before connecting the cable to the transducer
and monitor.
If this caution is not adhered to, the transducer and monitor could be damaged.
2. The connector connecting the sensor cable and the extension cable shall be insulated with
the attached insulation sleeve (transparent shrink tube) or fluoro resin insulation tape.
The vinyl-insulating tape shall not be used, which may cause the wiring trouble in the case of
the temperature more than 80°C.
3. Do not measure insulation resistance and dielectric strength other than those at places
specified.
If measured, transducer damage may result.
4. Before collecting data to check output characteristics, set gap, etc., warm up the transducer
for more than 30 minutes to stabilize the output after the power is turned on.
5. Do not use radio transceiver or cellular phone near the equipment under the condition of
opened Driver Housing, or uninstalled Sensor or Extension Cable for maintenance or the like.
It may interfere in the output of the Driver.
6. Before suppling the power to the driver, make sure that the sensor, extension cable and
driver are connected as a transducer system.
Do not supply the power to the driver without connecting the sensor and extension cable.
7. Do not remodel this unit without permission. Otherwise the guarantee can not be made.
8. This unit is designed for use by specialists or persons thoroughly familiar with the field.
9. Make sure that the end user receives the Instruction Manual delivered with this unit.
10. Do not wipe off name plate with solvents, such as toluene and methanol. Characters may
disappear.
Before Use.....
When the unit is received, inspect it for damage suffered in transport and check whether it is the item you
ordered. In the unlikely event that it was damaged in transport or does not function according to specifications,
please contact the SHINKAWA Office or dealer nearest you.
Store the unit under the ambient conditions given in the specification.
Avoid places where it is exposed to high humidity or corrosive gases.
i
TABLE OF CONTENTS VK-143P
1. GENERAL INFORMATION
1.1 GENERAL -------------------------------------------------------------------------------------- 1
2. SPECIFICATIONS
2.1 STANDARD SPECIFICATIONS ---------------------------------------------------------- 2
2.2 OUTLINE DRAWING ------------------------------------------------------------------------ 5
2.3 NAME PLATE --------------------------------------------------------------------------------- 16
3. PRINCIPLE OF OPERATION
3.1 GENERAL -------------------------------------------------------------------------------------- 17
4. INSTALLATION
4.1 RECEIVING INSPECTION ---------------------------------------------------------------- 18
4.2 STORING --------------------------------------------------------------------------------------- 18
4.3 INSTALLATION ------------------------------------------------------------------------------- 18
4.4 INSTALLATION PROCEDURES --------------------------------------------------------- 24
4.5 VK DRIVER INSTALLATION AND REMOVAL --------------------------------------- 26
4.6 TROUBLE SHOOTING----------------------------------------------------------------------- 27
5. INTERCONNECTION
5.1 INTERCONNECTION OF SYSTEM COMPOSITION PARTS -------------------- 28
5.2 CONNECTOR INSULATION -------------------------------------------------------------- 29
5.3 CONNECTION OF SIGNAL TRANSMISSION CABLE ----------------------------- 30
5.4 FIELD WIRING DIAGRAM ------------------------------------------------------------------ 32
6. TECHNICAL DATA
6.1 STANDARD STATIC CHARACTERISTIC --------------------------------------------- 35
6.2 TEST DATA FOR DC POWER SUPPLY ---------------------------------------------- 36
6.3 TEST DATA FOR TARGET MATERIAL ------------------------------------------------ 37
6.4 TEST DATA FOR TARGET DIAMETER ----------------------------------------------- 38
6.5 TEST DATA FOR TARGET EDGE ------------------------------------------------------ 39
6.6 TEST DATA FOR SIDE BORE ----------------------------------------------------------- 40
6.7 TEST FOR SENSOR TEMPERATURE CHANGE ---------------------------------- 41
6.8 TEST FOR DRIVER TEMPERATURE CHANGE ------------------------------------- 41
ii
TABLE OF CONTENTS VK-143P
7. INSTALLATION
7.1 FM ------------------------------------------------------------------------------------------------ 42
7.2 CSA ---------------------------------------------------------------------------------------------- 43
7.3 ATEX -------------------------------------------------------------------------------------------- 45
iii
1. GENERAL INFORMATION VK-143P
1.1 GENERAL
Model VK-143P Transducer is the most suitable measuring system for differential expansion between rotor
and its case.
It detects displacement precisely without physical contact with a observed material by the eddy current
principle, and offers 13.5mm linear range for JIS SCM440 as a standard calibration material.
This transducer system consists of VL sensor, VW extension cable and VK driver. System cable length
between sensor and driver is selectable for 5m or 9m by combination with sensor cable and extension cable.
Casing
Gap VL sensor
Connector
Rotor
VW extension cable
(target)
Output
Power
supply
(-24VDC)
VK driver
-1-
2. SPECIFICATIONS VK-143P
No entry if additional
Model Code / Additional Spec. Code( spec. code is not specified. )
SPECIFICATIONS NOTICE
CALIBRATION JIS SCM440 flat surface 1. CALIBRATION MATERIAL
MATERIAL *3 MODEL VK-143P Transducers are calibrated for JIS SCM440 flat
LINEAR RANGE *3 Over 13.5mm(begins at approx.3mm from sensor tip) surface (more than 75mm dia.).
SENSITIVITY *3 0.8V/mm If the measured target is other than JIS SCM440 flat surface, it will
SENSITIVITY Within ± 4% present a different characteristics. In such a case, calibration by the
ERROR*3 connected equipment (e.g. monitor) side should be required for
LINEARITY *3 Within ± 200µm of 0.8V/mm straight line system operation.
(if calibrated as a system)
2. INSULATORS
Within ± 270µm of 0.8V/mm straight line
Prior to shipment, the insulators have been installed to the mounting
(including interchangeability errors)
holes of VK driver.
FREQUENCY DC to 3kHz (-3dB) at 12mm pk-pk Be sure to mount VK driver without removing insulators.
RESPONSE *3 Mounting without insulators could cause noise on driver output.
MAX. OUTPUT Approx.-22VDC Special caution to insulaters shall be paid for the intrinsic safety
VOLTAGE *3 specification so that a system shall be earth grounded only at the
OUTPUT 50Ω barrier strip.
IMPEDANCE *3
CURRENT Max. -20mA 3. SHIELD WIRE CONNECTION
CONSUMPTION Connect shield wire of signal cable (3-wire shielded cable between
(10kΩ load) driver and monitor) to COM terminal.
SENSOR TIP DIAMETER Approx. 25mm dia. If this is not adhered to, noise may be caused.
CABLE DIAMETER Approx. 6mm dia. (with armor : max. 10mm dia.) 4. CONNECTOR ISOLATION, etc.
CONNECTOR DIAMETER Approx. 9.5mm dia. The connector connecting the sensor cable and the extension cable
SYSTEM CABLE 5m ± 10% or 9m ± 10% shall be insulated with the attached insulation sleeve (transparent
LENGTH shrink tube) or fluoro resin insulation tape.
OPERATING Sensor : -40 to +125°C The vinyl-insulating tape shall not be used, which may cause the
TEMPERATURE Extension Cable : -40 to +125°C wiring trouble in the case of the temperature more than 80°C.
RANGE Driver : -38 to +80°C The connector shall not be located in the oil environment.
RANGE OF EX2,4 : -20 to 85°C (Sensor, Ext. Cable & Driver) The oil penetration to cable through the connector may cause the
TEMPERATURE AT EX5 : -38 to 85°C (Sensor, Ext. Cable & Driver) sensitivity change, due to the change of the cable capacitance.
EXPLOSION PROOF 5. MEGGER TEST OF SIGNAL CABLE
CONSTRUCTION If megger test is made on the signal cable (3-wire shielded cable), be
TEMPERATURE Sensor : Less than ± 3% of F.S. sure to discharge the charged electric load before connecting the
CHARACTERISTIC Extension Cable : Less than ± 2% of F.S. cable to driver.
Condition : Gap=11mm, Target : JIS SCM440 If this caution is not adhered the driver could be damaged.
0 to 80°C (at 20°C standard) 6. SENSOR INSTALLATION
Driver : Less than ± 4% of F.S. Not available for rain water at out door use.
Roop : Less than ± 4% of F.S. It may cause the sensitivity change and insulation down.
Condition : Gap=11mm, Target : JIS SCM440
0 to 60°C (at 20°C standard) 7. SAFETY BARRIER
OPERATING 30 to 95% RH (noncondensing, non-submerged) In case of the intrinsically safe specification, the approved following
HUMIDITY RANGE (sensor body : 100%) safety barrier is recommended.
POWER SUPPLY -24VDC ± 10% • MTL 796-
Espcially in the casse of FM approval、don’t , be used except MTL796-
DIELECTRIC STRENGTH Between each terminal and insulator:
OF DRIVER 1mA or less at 500VAC for one minute 8. CALIBRATED AS A SYSTEM
INSULATION Between each terminal and insulator: The sensor, extension cable and driver, which are calibrated as a
RESISTANCE OF 100MΩ or more at 500VDC system, shall be connected with each serial No. as specified in the
DRIVER inspection test report. If this is not adheared the output characteristics
SCREWS OF M4 may be out of specification.
TERMINAL BLOCK
APPLICABLE WIRE Other
0.75 to 2mm2
GAUGE
*3 The above specifications apply at 25°C with -24VDC power supply and
load resistance 10kΩ and JIS SCM440 target (thickness≥5mm).
CONFIGURATION
Sensor Connector Extension Cable
Driver
39341E3.7
-2-
2. SPECIFICATIONS VK-143P
No entry if additional
Model Code / Additional Spec. Code( spec. code is not specified. )
SPECIFICATIONS NOTICE
CALIBRATION JIS SCM440 flat surface 1. CALIBRATION MATERIAL
MATERIAL *3 MODEL VK-143P Transducers are calibrated for JIS SCM440 flat
LINEAR RANGE *3 Over 13.5mm(begins at approx.3mm from sensor tip) surface (more than 75mm dia.).
SENSITIVITY *3 0.8V/mm If the measured target is other than JIS SCM440 flat surface, it will
SENSITIVITY Within ±4% present a different characteristics. In such a case, calibration by the
ERROR*3 connected equipment (e.g. monitor) side should be required for system
LINEARITY *3 Within ±200µm of 0.8V/mm straight line operation.
(if calibrated as a system)
2. INSULATORS
Within ±270µm of 0.8V/mm straight line
Prior to shipment, the insulators have been installed to the mounting
(including interchangeability errors)
holes of VK driver.
FREQUENCY DC to 3kHz (-3dB) at 12mm pk-pk Be sure to mount VK driver without removing insulators.
RESPONSE *3 Mounting without insulators could cause noise on driver output.
MAX. OUTPUT Approx.-22VDC
VOLTAGE *3 3. SHIELD WIRE CONNECTION
OUTPUT 50Ω Connect shield wire of signal cable (3-wire shielded cable between
IMPEDANCE *3 driver and monitor) to COM terminal.
CURRENT Max. -20mA If this is not adhered to, noise may be caused.
CONSUMPTION 4. CONNECTOR ISOLATION, etc.
(10kΩ load) The connector connecting the sensor cable and the extension cable
SENSOR TIP Approx. 25mm dia. shall be insulated with the attached insulation sleeve (transparent
DIAMETER shrink tube) or fluoro resin insulation tape.
CABLE DIAMETER Approx. 6mm dia. (with armor : max. 10mm dia.) The vinyl-insulating tape shall not be used, which may cause the wiring
CONNECTOR Approx. 9.5mm dia. trouble in the case of the temperature more than 80°C.
DIAMETER The connector shall not be located in the oil environment.
SYSTEM CABLE 5m±10% or 9m±10% The oil penetration to cable through the connector may cause the
LENGTH sensitivity change, due to the change of the cable capacitance.
OPERATING Sensor : -40 to +125°C 5. MEGGER TEST OF SIGNAL CABLE
TEMPERATURE Extension Cable : -40 to +125°C If megger test is made on the signal cable (3-wire shielded cable), be
RANGE Driver : -38 to +80°C sure to discharge the charged electric load before connecting the
RANGE OF EX4 : -20 to +85°C (Sensor, Ext, Cable & Driver) cable to driver.
TEMPERATURE AT EX5 : -38 to +85°C (Sensor, Ext, Cable & Driver) If this caution is not adhered the driver could be damaged.
EXPLOSION PROOF 6. SENSOR INSTALLATION
CONSTRUCTION Not available for rain water at out door use.
TEMPERATURE Sensor : Less than ±3% of F.S. It may cause the sensitivity change and insulation down.
CHARACTERISTIC Extension Cable : Less than ±2% of F.S.
Condition : Gap=11mm, Target : JIS SCM440 7. SAFETY BARRIER
0 to 80°C (at 20°C standard) In case of the intrinsically safe specification, the approved following
Driver : Less than ±4% of F.S. safety barrier is recommended.
Roop : Less than ±4% of F.S. • MTL 796-
Condition : Gap=11mm, Target : JIS SCM440 8. CALIBRATED AS A SYSTEM
0 to 60°C (at 20°C standard) The sensor, extension cable and driver, which are calibrated as a
OPERATING 30 to 95% RH (noncondensing, non-submerged) system, shall be connected with each serial No. as specified in the
HUMIDITY RANGE (sensor body : 100%) inspection test report. If this is not adheared the output characteristics
POWER SUPPLY -24VDC ± 10% may be out of specification.
DIELECTRIC STRENGTH Between each terminal and insulator:
OF DRIVER 1mA or less at 500VAC for one minute
SCREWS OF M4
TERMINAL BLOCK
APPLICABLE WIRE
0.75 to 2mm2
GAUGE
*3 The above specifications apply at 25°C with -24VDC power supply and
load resistance 10kΩ and JIS SCM440 target (thickness≥5mm).
CONFIGURATION
30018E1.7
-3-
2. SPECIFICATIONS VK-143P
No entry if additional
Model Code / Additional Spec. Code( spec. code is not specified. )
SPECIFICATIONS NOTICE
CALIBRATION JIS SCM440 flat surface 1. CALIBRATION MATERIAL
MATERIAL *2 MODEL VK-143P Transducers are calibrated for JIS SCM440 flat
LINEAR RANGE *2 Over 13.5mm(begins at approx.3mm from sensor tip) surface (more than 75mm dia.).
SENSITIVITY *2 0.8V/mm If the measured target is other than JIS SCM440 flat surface, it will
SENSITIVITY Within ±4% present a different characteristics. In such a case, calibration by the
ERROR*2 connected equipment (e.g. monitor) side should be required for
LINEARITY *2 Within ±200µm of 0.8V/mm straight line system operation.
(if calibrated as a system)
2. INSULATORS
Within ±270µm of 0.8V/mm straight line
Prior to shipment, the insulators have been installed to the mounting
(including interchangeability errors)
holes of VK driver.
FREQUENCY DC to 3kHz (-3dB) at 12mm pk-pk Be sure to mount VK driver without removing insulators.
RESPONSE *2 Mounting without insulators could cause noise on driver output.
MAX. OUTPUT Approx.-22VDC
VOLTAGE *2 3. SHIELD WIRE CONNECTION
OUTPUT 50Ω Connect shield wire of signal cable (3-wire shielded cable between
IMPEDANCE *2 driver and monitor) to COM terminal.
CURRENT Max. -20mA If this is not adhered to, noise may be caused.
CONSUMPTION 4. CONNECTOR ISOLATION, etc.
(10kΩ load) The connector connecting the sensor cable and the extension cable
SENSOR TIP Approx. 25mm dia. shall be insulated with the attached insulation sleeve (transparent
DIAMETER shrink tube) or fluoro resin insulation tape.
CABLE DIAMETER Approx. 6mm dia. (with armor : max. 10mm dia.) The vinyl-insulating tape shall not be used, which may cause the
CONNECTOR Approx. 9.5mm dia. wiring trouble in the case of the temperature more than 80°C.
DIAMETER The connector shall not be located in the oil environment.
SYSTEM CABLE 5m±10% or 9m±10% The oil penetration to cable through the connector may cause the
LENGTH sensitivity change, due to the change of the cable capacitance.
OPERATING Sensor : -40 to +125°C 5. MEGGER TEST OF SIGNAL CABLE
TEMPERATURE Extension Cable : -40 to +125°C If megger test is made on the signal cable (3-wire shielded cable), be
RANGE Driver : -38 to +80°C sure to discharge the charged electric load before connecting the
TEMPERATURE Sensor : Less than ±3% of F.S. cable to driver.
CHARACTERISTIC Extension Cable : Less than ±2% of F.S. If this caution is not adhered the driver could be damaged.
Condition : Gap=11mm, Target : JIS SCM440 6. SENSOR INSTALLATION
0 to 80°C (at 20°C standard) Not available for rain water at out door use.
Driver : Less than ±4% of F.S. It may cause the sensitivity change and insulation down.
Roop : Less than ±4% of F.S.
Condition : Gap=11mm, Target : JIS SCM440 7. CALIBRATED AS A SYSTEM
0 to 60°C (at 20°C standard) The sensor, extension cable and driver, which are calibrated as a
OPERATING 30 to 95% RH (noncondensing, non-submerged) system, shall be connected with each serial No. as specified in the
HUMIDITY RANGE (sensor body : 100%) inspection test report. If this is not adheared the output characteristics
POWER SUPPLY -24VDC ± 10% may be out of specification.
DIELECTRIC Between each terminal and insulator:
STRENGTH OF 1mA or less at 500VAC for one minute
DRIVER
INSULATION Between each terminal and insulator:
RESISTANCE OF 100MΩ or more at 500VDC Other
DRIVER
SCREWS OF M4
TERMINAL BLOCK
APPLICABLE WIRE 0.75 to 2mm2
GAUGE
*2 The above specifications apply at 25°C with -24VDC power supply and
load resistance 10kΩ and JIS SCM440 target (thickness≥5mm).
CONFIGURATION
30305E1.2
-4-
2. SPECIFICATIONS VK-143P
2.2.1 VK DRIVER
Dimensions: mm
-5-
2. SPECIFICATIONS VK-143P
Dimensions: mm
-6-
2. SPECIFICATIONS VK-143P
Dimensions: mm
-7-
2. SPECIFICATIONS VK-143P
Dimensions: mm
-8-
2. SPECIFICATIONS VK-143P
Dimensions: mm
No. Name Material Quantity Remark
1 Coax. connector
2 Coax. cable Fluoro resin
3 Coax. connector
-9-
2. SPECIFICATIONS VK-143P
Dimensions: mm
-10-
2. SPECIFICATIONS VK-143P
Dimensions: mm
-11-
2. SPECIFICATIONS VK-143P
Dimensions: mm
-12-
2. SPECIFICATIONS VK-143P
Dimensions: mm
-13-
2. SPECIFICATIONS VK-143P
Dimensions: mm
-14-
2. SPECIFICATIONS VK-143P
Dimensions: mm
-15-
2. SPECIFICATIONS VK-143P
-16-
3. PRINCIPLE OF OPERATION VK-143P
3.1 GENERAL
Fig.3-1 below shows a block diagram of this differential expansion transducer. This is a non-contact type
differential expansion transducer applying eddy current, and a high frequency power of about 1MHz is
supplied to the sensor from an oscillator in the driver.
Thus, the sensor generates high frequency magnetic field so that eddy current flows in the target (observed
material).
The eddy current flow in the target causes magnetic field generation on the target side, resulting in sensor
impedance change.
Thus,if the relationship of the gap between the sensor and target vs. sensor impedance is known, the gap
between the sensor and target can be obtained through sensor impedance measurement.
The oscillator output is detected and voltage output linearized with respect to the gap is output from the
linearizer.
VK Driver
VL Sensor
Power Supply
FIG. 3-1
-17-
4. INSTALLATION VK-143P
! CAUTION
Keep defend tightening torque of under table without fail, in the case that the VL sensor
attaches to the installation brackets or sensor-sleeve. Failure to observe this precaution could
result in equipment damage.
-18-
4. INSTALLATION VK-143P
Connector
a)
Mounting Plate
Distance X ≥ 35mm
No effect.
TARGET
Best installation
X
-19-
4. INSTALLATION VK-143P
b)
Mounting Plate
Distance X ≥ 35mm
Y Y> 75mm
TARGET X No effect.
c)
Mounting Plate
d)
Mounting Plate
Distance X<35mm
TARGET
There is effect.
X
e)
Mounting Plate
f)
Mounting Plate
Y< 75mm
Y
TARGET There is effect.
g) Mounting Plate
X
Gap X ≤ 3mm
TARGET Maker dead zone therefore can’t measure.
Mounting Plate
h) X1
Gap X1 ≠ Gap X2
There is effect.
TARGET
X2
-20-
4. INSTALLATION VK-143P
a) to c)
h)
Output(V)
d) to f)
Gap (µm)
Dead zone
As the above results show, it is most preferable to install the sensor as shown in a) to c). However if the sensor
is required to be installed as shown in d) to f), calibration in installed conditions is necessary.
Also, the dead zone as shown in the above should be noted.
! CAUTION
The following conditions must be met to ensure accurate differential expansion detection.
1. The target surface must be more than 3 times the sensor tip diameter(Fig. 4.1(a)).
2. There must not be any conductive materials within a range of 3 times the sensor tip diameter
(Fig. 4.1(b)).
3. To avoid mutual interference between sensors, a minimum distance corresponding to 10
times the sensor tip diameter must be maintained between sensors (Fig. 4.1(c)).
4. Use a sufficiently rigid installation method to prevent vibration of the sensor itself (Fig. 4.1(d)).
3D 3D
(a)
(b)
10D
10D
(c) (d)
Fig. 4.1
-21-
4. INSTALLATION VK-143P
D : Sensor Tip
TARGET X
d>3D No Effect.
X>D No Effect.
R>10D No Effect.
CASING
X
TARGET
X>D No effect
-22-
4. INSTALLATION VK-143P
-23-
4. INSTALLATION VK-143P
! CAUTION
1. When adjusting the sensor set gap, refer to the TECHNICAL DATA, adjust and set sensor
set gap so that sensor does not contact target even at maximum proximity, and such that
gap is not beyond range of connected monitor.
2. Beware not to apply the impact force to the sensor, which may cause the damage.
Insert gauge for desired set gap between sensor and target.
Fix sensor at the position where the gauge can be slipped out
Installation accessory
smoothly.
! CAUTION
Do not scratch sensor tip or target surface.
Rotor
4.4.3
t Sensor
Gauge
-24-
4. INSTALLATION VK-143P
Connector
Digital Multimeter
VL Sensor
VK Driver
“OUT”
“COM”
! CAUTION
To prevent twisting, disconnect sensor cable from extension cable when mounting sensor.
First approach the VL sensor to the target and read the gap voltage indicated on a voltmeter etc. connected
across the VK Driver output terminals “COM-OUT”.
In accordance with this voltage and “6. TECHNICAL DATA” adjust VL sensor gap and fix sensor such that
voltage agrees with gap.
-25-
4. INSTALLATION VK-143P
! CAUTION
• The mounting holes are provided with insulators for insulation against earth. Make sure to
install the supplied 4 sets of mounting screws (M4) by inserting them through the insulators.
Installation without these insulators may cause interference noise in the output.
• Turn four attachment screws equally in the case of attachment and removal. When it is going to
loosen and remove one screw at a time especially in other than intrinsic safety, a screw may
separate from an insulator.
Tightening Torque 1.18N・m (12kgf・cm) or less
M4 Screw
Insulator (small)
Insulator (large)
Insulator
-26-
4. INSTALLATION VK-143P
-27-
5. INTERCONNECTION VK-143P
! CAUTION
1. The sensor, extension cable and driver, which are calibrated as a system at the manufacturer,
shall be connected with each serial No. as specified in the inspection test report.
If this is not adhered the output characteristics may be out of specification.
2. If the sensor and driver are connected without the extension cable, the output will differ from
the specification greatly.
VL sensor Connector
VW extension cable
Output
Power supply
(-24VDC)
Connector
VK driver
! CAUTION
1. Before tightening the connector, the connector internal should be checked to confirm that
there is no foreign particle, which may cause the bad characteristics due to the imperfect
contact.
2. As the connector needs to be firmly tightened to avoid looseness after being tightened, it is
recommended that the connector be tightened by the following torque.
Tightening torque : 0.8 to 1.1N•m (8 to 11kg•cm REF.)
Diagonal length of wrench used
Extension Cable•Sensor Side : 6.4mm (1/4 inch REF.)
Sensor•Connector : 8mm
It should be noted that the connector may be damaged if applied with an excessive torque.
3. Beware not to apply the excessive force on the thread part of the connector, which may
cause the damage.
4. After tightening connector, make sure that the cable torque does not cause the connector to
loosen again.
If the installation conditions cannot be changed to prevent this force acting on the connector,
the force should acts in the tightening direction of the connector.
If the installation direction of the extension cable is such that a force acts in loosening
direction of the connector, twist the extension cable lightly in the direction of that the resulting
repulsive force acts to tighten the connector. Then connect the connector and tighten.
(SHINKAWA Electric recommends to take up the surplus length of the extension cable by
winding it into a cable box. If the excessive extension cable needs to be accommodated in
the driver housing for some unavoidable reasons, do not push it forcibly in the housing, and
also do not cut it.)
-28-
5. INTERCONNECTION VK-143P
Connector
Sensor cable
Insulation sleeve
Extension cable
Collar
• If insulation sleeve tubing is not available, wrap the connector with insulating material such as
fluoro resin tape.
! CAUTION
The vinyl-insulating tape shall not be used, which may cause the wiring trouble in the case of
the temperature more than 80°C.
-29-
5. INTERCONNECTION VK-143P
• For interconnection between the VK driver and the monitor make sure to use 3-core shielded
cables or pair shielded standard 3-wire instrument cables, since the signal transmitted is weak.
! CAUTION
Do not use unshielded cables or non-pair shielded multicore cables as the measured values
could be affected by external noise.
Shield wire
! CAUTION
1. For cable wiring for pulse signals (phase marker, revolution) separate them from cables for
displacement and vibration signals.
2. Avoid running signal transmission cables together with control and power cables.
3. Install all cables in conduit pipes or on cable racks.
4. After the megger test of the signal cable, discharge the charged electric load before
connecting the cable.
If this caution is not adhered to, the transducer and monitor could be damaged.
-30-
5. INTERCONNECTION VK-143P
• Use cables with AWG No. 14 to 18 (0.75mm2 to 2mm2) gauge stranded conductors.
Cable rack
(conduit pipe)
! CAUTION
Earth grounding should be grounded at only one end.
-31-
5. INTERCONNECTION VK-143P
-32-
5. INTERCONNECTION VK-143P
-33-
5. INTERCONNECTION VK-143P
! WARNING
Beware of electric shock from high-voltage parts.
! CAUTION
Be sure to observe the following cautions at the time of interconnection.
Caution No.
Caution
in drawing
NOTE 1 Allowable bending radius
Without armor : 30 mm
With armor : 50 mm
Allowable tension : 98.1N (10 kgf REF.)
NOTE 2 To insulate the connector, cover it with the supplied shrink tube (insulating
sleeve) or insulating tape.
Take measures to prevent oil or water from entering the insulating sleeve or
tape. (refer to section “5.2”) in case of the intrinsically safe spesification.
NOTE 3 The terminal of the VK driver must be used with the terminal cover and the
connected cable must be fixed with the cable clamp at near the terminal
certainly. The VK driver must be installed in the housing.
NOTE 5 Use a 3-core shielded cable with AWG No. 14 to No. 18 gage stranded
conductors for wiring between the VK driver and the monitor.
Do not wire the cables together in high tension ducts. (refer to section “5.3”)
Refer to “7. INFORMATION ABOUT INTRINSICALLY SAFE
APPLICATION”in case of the intrinsically safe spesification.
-34-
6. TECHNICAL DATA VK-143P
-20
-18
-16
-14
-12
Output (V)
-10
-8
-6
-4
-2
0
0 5 10 15 20 25
GAP(mm)
-35-
6. TECHNICAL DATA VK-143P
6.2 EST DATA FOR DC POWER SUPPLY VARIATION (TARGET MATERIAL: SCM440)
-20
-18 -26.4V
-25.2V
-16
-24.0V
-14 -21.6V
-12 -22.8V
Output (V)
-10
-8
-6
-4
-2
0
0 5 10 15 20 25
GAP(mm)
-36-
6. TECHNICAL DATA VK-143P
Sensor
Target
Gap
-20
-18
SUS304
-16
-14 S45C
-12
Output (V)
-10
SCM440
-8
-6 SNCM625
-4
-2
0
0 5 10 15 20 25
GAP(mm)
-37-
6. TECHNICAL DATA VK-143P
Sensor
Target
-20
-18
D=30mm
-16
-14 D=50mm
D=75mm
-12
Output (V)
-10
D=∞
-8
-6
-4
-2
0
0 5 10 15 20 25
GAP(mm)
-38-
6. TECHNICAL DATA VK-143P
X X Sensor
Target
Gap
Target Sensor
Detailed drawing
-20
-18 X=20mm
-16
X=0mm
-14
X=10mm
-12
X=40mm
Output (V)
-10
X=30mm
-8
X=∞
-6
-4
-2
0
0 5 10 15 20 25
GAP(mm)
-39-
6. TECHNICAL DATA VK-143P
Side bore
X Sensor
Target
Gap
-20
-18
-16
X=∞
-14
X=40mm
X=30mm
-12
Output (V)
-10
X=20mm
-8
-6
X=15mm
-4
-2
0
0 5 10 15 20 25
GAP(mm)
-40-
6. TECHNICAL DATA VK-143P
6.7 TEST DATA FOR SENSOR TEMPERATURE CHANGE (TARGET MATERIAL: SCM440)
Target
Gap : 11mm
Driver
4
2
% of F.S.
0
-2
-4
0 20 40 60 80
Temperature(°C)
6.8 TEST DATA FOR DRIVER TEMPERATURE CHANGE (TARGET MATERIAL: SCM440)
Gap : 11mm
Driver
4
2
% of F.S.
0
-2
-4
0 20 40 60
Temperature(°C)
-41-
7. INFORMATION ABOUT INTRINSICALLY SAFE APPLICATION VK-143P
7.1 FM
-42-
7. INFORMATION ABOUT INTRINSICALLY SAFE APPLICATION VK-143P
7.2 CSA
-43-
7. INFORMATION ABOUT INTRINSICALLY SAFE APPLICATION VK-143P
-44-
7. INFORMATION ABOUT INTRINSICALLY SAFE APPLICATION VK-143P
7.3 ATEX
-45-
8. TABLE OF MODEL CODE VK-143P
●Driver
A B C D E
System Geothermal CE
System cable length Intrinsically safe
calibration spec. marking
1 5m 2 FM
2 9m 4 CSA
3 5m 5 ATEX
4 9m
5 5m
6 9m
● Extension Cable
VW-143P - / EX / GEO
A B C D
A B C D
Geothermal
Armor Cable length Intrinsic safety
Spec.
A With 1 2.0m 2 FM
L Without 2 3.0m 4 CSA
3 4.0m 5 ATEX
4 4.5m
6 6.0m
7 7.0m
8 8.0m
A 8.5m
● Sensor
-46-
8. CONTROL DRAWING VK-143P
A B C D E F
-47-
Warranty
The warranty period for the delivered product shall be 3 years from the date of delivery from
our factory. But the warranty period for the special order product and software shall be 1 year,
and for the repairing parts shall be 6 months. In the event that the delivered product develops
any defects within the warranty period for which Shinkawa Electric is liable, the defective part
will be replaced or repaired in our factory at our own expense.
In the event that an engineer needs to be dispatched for repair purposes during the warranty
period, travel expenses for the said engineer will be changed in accordance with the above
item. The cost of repair will be borne by Shinkawa Electric.
However, the following are excluded from our warranty: