You are on page 1of 16

Contents

Introduction...................................... 2
Membrane Type............................ 2
Pump Capacity............................. 2
Water Quality............................... 2
Installation....................................... 2
Pellicon® 3 88 cm2 and 0.11 m2
Cassette Holder Installation............ 2
Pellicon® 3 0.57 or 1.14 m2 Cassette

Installation and User Benchtop Holder Installation........... 3


Manual Process-Scale Holder

Guide
Installation.................................. 4
Hydraulic Process Scale Holder
Installation.................................. 5
Maintenance .................................... 7
Flushing and Cleaning........................ 7
Flushing...................................... 8
Cleaning...................................... 8
Sanitization...................................... 9
Pellicon 3 Cassettes
® Depyrogenation............................... 10
Integrity Testing.............................. 10
Measurement of Normalized Water
Permeability................................... 11
Storage.......................................... 12
Troubleshooting............................... 14
Warranty........................................ 15

The life science business of Merck KGaA, Darmstadt,


Germany operates as MilliporeSigma in the U.S. and Canada.
Pellicon® 3 Cassettes Installation and User Guide 2

Introduction Installation
This guide provides installation and Pellicon® 3 88 cm2 and 0.11 m2 Cassette Holder Installation
maintenance procedures for Pellicon® 3
NOTE
cassettes. It is not intended to provide
Pellicon® 3 88 cm2 and 0.11 m2 Cassettes must be installed in a Pellicon®
validation protocols or supporting data
Cassette Holder, Catalog Number XX42PMINI.
for validation purposes. Please refer
to the cassette Validation Guide for
Holder Installation
this information. Please refer to the
Certificate of Quality supplied with each 1. Loosen the nuts on the tie rods of the
cassette for specifications. holder and remove the nuts, washers,
tie rod spacers and end plate.
Pellicon® 3 Cassette
Membrane Type 2. Inspect the tie rods and nuts for signs
Install filters containing only one type of burrs or stripped threads. Replace
of membrane in the filter holder at one any damaged components. Nuts must
time. Do not mix filters with different turn freely on the tie rods to ensure
pore sizes or nominal molecular weight proper tightening of the holder.
cutoffs. The area of membrane used
depends on the filter surface area 3. Slide the Pellicon® 3 Cassette onto
required for your particular application. holder so that the tie rods pass
through the cutouts in the Pellicon®
Pump Capacity 3 Cassette. Repeat to install one or
When operating Pellicon® 3 cassettes, two additional Pellicon® 3 Cassettes
select a pump with adequate capacity. if required. Pellicon® Cassette Holder, XX42PMINI
Recommended feed crossflow rate for 4. Slide the end plate onto the tie rods
cassettes is 4 - 6 liter/min/m2. Optimal and press the end plate against the
cross flow will depend on the actual cassette.
solution being filtered.
5. Install the tie rod spacers, washers and nuts. Hand-tighten the nuts, alternating from one
Water Quality nut to the other nut.
Reverse osmosis (RO) water or water for NOTE
injection (WFI) is recommended Uneven tightening of the nuts can damage the Pellicon® Cassette. Nonparallel
plates or compression of the filters at one end can cause leakage.
6. Tighten the nuts using the torque wrench and 9/16 in. socket supplied with the
Pellicon® Cassette Holder. If your wrench has a ratchet head, move the lever to the
right lock position to tighten the nuts or to the left lock position to loosen the nuts.
Pellicon® 3 Cassettes Installation and User Guide 3

7. Set the torque wrench at 180 to 200 Pellicon® 3 0.57 or 1.14 m2 Cassette Benchtop Holder
in. lb. (20.3 to 22.6 Nm) by holding Installation
the spring loaded lock collar down
and turning the handle until the 180 XX42P0060 Pellicon® Cassette Acrylic Filter Holder
to 200 in. lb. (20.3 to 22.6 Nm) XX42PRV60 Pellicon® Cassette Acrylic Holder Low Retentate Volume
mark aligns with the 0 (zero) mark XX42P0080 Pellicon® Cassette Stainless Steel Holder
on the collar. XX42P0K80 Pellicon® Cassette Stainless Steel Holder Assembly

8. Rotate the handle slightly to spring 1. Loosen the nuts on the tie rods of the holder and remove the nuts, washers and end plate.
the lock collar back into place. 2. Inspect the tie rods and nuts for signs of burrs or stripped threads. Replace any
9. Attach the socket to the torque damaged components. Nuts must turn freely on the tie rods to ensure proper
wrench by firmly pressing the socket tightening of the holder.
onto the torque wrench drive. 3. Clean the manifold adapter plate (XXPEL3MAP) with an alcohol wipe.
10. Place the socket over one nut. With a 4. Slide the manifold adapter plate onto the holder against the holder’s manifold
continuous motion, turn each nut ¼ plate, with the solid light grey side labeled “Manifold Side” towards the manifold,
turn, alternating between the nuts, and the wings resting on the tie rods. Slide the manifold adapter plate directly
until the wrench clicks, which will up against the stainless steel manifold.
indicate it has reached the set force.
5. Slide the Pellicon® 3 cassette up against the manifold adapter plate. The dark
grey center with light grey perimeter of the manifold adapter plate should be up
against the Pellicon® 3 cassette.
6. Slide a Pellicon® 3 Cassette onto the holder.
7. Slide the end plate with the handle side up onto the tie rods.
8. Tie rod spacers (XX4200066) are available when operating the system with low
membrane area. Place enough spacers on each rod so that the length of exposed
threaded tie rod is less than the length of the hex socket. Replace the washers
and nuts. Hand tighten the nuts in a diagonal pattern.
NOTE
Uneven tightening of the nuts can damage the Pellicon® Cassette. Nonparallel
plates or compression of the filters at one end can cause leakage.
9. Tighten the nuts using the torque wrench and 15/16 in. deep hex socket supplied
with the Pellicon® Cassette Holder.
10. Set the torque wrench by turning the knurled locking collar clockwise or by pulling
the handle out to unlock. Turn the torque setting scale clockwise until the 350 in-lb
(40 Nm) mark on the torque scale is opposite the zero mark on the smooth collar.
Turn the knurled locking collar counterclockwise to lock.
Pellicon® 3 Cassettes Installation and User Guide 4

11. Using the torque wrench and 15/16 in.


deep hex socket, turn each nut 1/8 turn nut
in the sequence shown here. When the hex nut
washer
washer
wrench clicks, this indicates that the set
stainless steel
force has been reached. DO NOT OVER stainless steel
bushing top acrylic
bushing

TORQUE. Evenness of torque is critical pressure plate top

M
ILL
pressure plate

IPO
to system performance. top handle

RE
plate tie rod
manifold

Pellicon®
Filter

Pellicon
® manifold adapter
Gasket Plate
Cassette plate
threaded
tie rod

PE permeate
RM
EA manifolds
TE bottom (four holes)
manifold plate

manifoldGasket
adapter
Plate
plate
permeate ED
outlet port FE pressure
threaded
3/4 in. sanitary E gauge
holes for AT
insertion ® Tri-Clover ME
connection P ER
of tie rods bottom

TE
TA
pressure plate

N
TE
bottom retentate

E
R
outlet port
plate RE
manifold feed inlet
port
TE
NT
AT
Exploded View of Acrylic Pellicon®
E
Cassette Filter Holder.
1in. sanitary
Tri-Clover® FE
Pellicon ®
Holder Tightening Sequence connection ED

Exploded View of Stainless Steel Pellicon®


Cassette Filter Holder.
NOTE Pellicon10_2_06_1.eps
Depending on the results of the Manual Process-Scale Holder Installation
Pellicon10_2_06_2.eps
Integrity Test, the torque may need 1. Loosen the nuts on the tie rods of the holder and remove the nuts, washers and
to be increased at intervals of 50 end plate.
in-lb (5.6 Nm) up to a maximum of
550 in-lb (62 Nm). 2. Inspect the tie rods and nuts for signs of burrs or stripped threads. Replace any
damaged components. Nuts must turn freely on the tie rods to ensure proper
12. Recheck the torque prior to use. tightening of the holder.
NOTE 3. Clean the manifold adapter plate (XXPEL3MAP) with an alcohol wipe.
If the filter holder is tightened in a
cold area and moved to a warmer 4. Slide the manifold adapter plate onto the holder against the holder’s manifold plate,
area, loosen the nuts entirely, let with the solid light grey side labeled “Manifold Side” towards the manifold, and the
the filter holder warm up, and then wings resting on the tie rods. Slide the manifold adapter plate directly up against the
retighten to prevent warping of the stainless steel manifold.
end plates.
Pellicon® 3 Cassettes Installation and User Guide 5

5. Slide the Pellicon® 3 cassette up against the manifold adapter plate. The dark Hydraulic Process Scale
grey center with light grey perimeter of the manifold adapter plate should be up Holder Installation
against the Pellicon® 3 cassette.
The Hydraulic Closure with Manual
6. Replace the spacers, washers and nuts as shown in Figure 5. Hand-tighten the Controls is intended for use with the
nuts in a diagonal fashion (see Figure 4). Use a few drops of suitable lubricant on Millipore Process Scale Filter Holder. It
the tie-rods and between the nuts and the washers if required. will maintain appropriate pressure on
the holder during processing or during
Note Uneven tightening of the nuts can damage the Pellicon® 3 Cassette.
storage of installed filter cassettes.
Nonparallel plates or compression of the filters at one end can cause leakage.
RUN mode is required for process
7. Tighten the nuts using the torque wrench and 15/16 in. deep hex socket supplied
operations. The system hydraulic pressure
with the Pellicon® 3 Cassette Holder.
is maintained at 1800 psig ± 100 psig.
8. Set the torque wrench by turning the knurled locking collar clockwise or by pulling
In STORE mode, the system may be
the handle out to unlock. Turn the torque setting scale clockwise until the 350 in-lb
stored with the Pellicon® 3 cassettes
(40 Nm) mark on the torque scale is opposite the zero mark on the smooth collar.
installed. The system hydraulic pressure
Turn the knurled locking collar counterclockwise to lock.
is maintained at 1200 psig ± 100 psig.
9. Using the torque wrench and 15/16
in. deep hex socket, turn each nut
1/8 turn in the sequence shown in
Figure 4. When the wrench clicks,
this indicates that the set force
has been reached. DO NOT OVER
TORQUE. Evenness of torque is
critical to system performance.
Note Depending on the results of the
Integrity Test, the torque may need
to be increased at intervals of 50
in-lb (5.6 Nm) up to a maximum of Hydraulic Process Scale Holder
550 in-lb (62 Nm). Process Scale Holder

10. Recheck the torque prior to use.


1. Set the System Mode switch to DUMP.
Note
Check the Hydraulic Pressure Gauge
on the hydraulic power unit panel to be
Pellicon10_2_06_3
sure the pressure is relieved. Do not
proceed until the hydraulic pressure
reaches 0 psig as read on gauge PI-2.
Pellicon® 3 Cassettes Installation and User Guide 6

2. On the first holder level (if there Holder Pressurization and the System Pressure Manual Dump
are more than one), remove the Operation Valve, MV-1 on the left side of the
TC clamps and inserts holding the Hydraulic Power Unit.
1. The hydraulic fluid reservoir must be
hydraulic cylinders to the clamp rods.
full (indicated on the level indicator Caution
3. Loosen and remove the hand wheels inside the Hydraulic Power Unit). The hydraulic closure system must
on the clamp rods at the opposite remain connected to the holder
2. All hydraulic flex hose connections
end plate (of the same holder level). during processing and during
must be installed from the
storage of installed cassettes. The
4. Slide the end plate away from the Hydraulic Power Unit to the
hydraulic closure system must be
manifold segment, creating an hydraulic cylinders.
depressurized before connecting
opening for cassette insertion into
3. All hardware on the clamp rods or disconnecting fittings. If the
the holder.
must be in place. Tighten the hand position of the Operating Mode
5. Slide the manifold adapter plate wheels until they are hand-tight. Selector Switch (RUN or STORE)
or a stainless steel support plate is changed while the system is in
4. Turn on the air supply to the
onto the holder against the holder’s ENABLE mode, the hydraulic system
Hydraulic Power Unit. A minimum
manifold plate. pressure will drop to 1000 psig
of 80 psig air pressure should be
and then increase to the correct
6. Slide a Pellicon® 3 Cassette onto the available. Check the air pressure
pressure for the new selected mode.
holder. Up to ten 1.14 m2 cassettes on the pressure gauge PI-1 on
If the hydraulic system pressure
may be installed per side if required. Hydraulic Power Unit front panel.
drops below 1200 psig while in
7. Slide the end plate back against the 5. The Power Switch on the Hydraulic RUN mode or 800 psig while in
cassette assembly. Control Unit HS-1 must be in the STORE mode, the relevant Low
ON position, and the green Power Pressure Alarms (PAL-1 or PAL-2
8. Install the grooved clamp rods, TC Indicator must be illuminated. respectively) will be illuminated and
clamps and inserts and the hand the Alarm Horn will sound. Once
wheels. 6. Set the Operating Mode Selector the problem has been corrected and
Switch, HS-3 to RUN or STORE, as hydraulic pressure restored, the
NOTE The hand wheels require hand required.
tightening only. Ensure that each alarm condition may be cleared by
end plate is flush against the 7. Set the System Mode Selector pressing the Alarm Reset – Resume
cassette assembly and both hand Switch, HS-2 to ENABLE. The Operation Push-button.
wheels are snug against the end hydraulic closure system will close
plate. Not less than one inch of the holder assembly and maintain
excess thread on the clamp rod the appropriate pressure on the
exiting the hand wheel should be holder for either RUN or STORE
exposed. mode, as selected.

9. Repeat the procedure for any other 8. To relieve hydraulic system pressure
holder levels. at any time, set the System Mode
Selector Switch to DUMP or open
Pellicon® 3 Cassettes Installation and User Guide 7

Maintenance Flushing and


Pellicon® 3 Cassettes must be operated while under compression from the holder Cleaning
assembly to ensure that the cassettes will seal properly. The compression is supplied
by applying torque to the nuts on the threaded tie-rods of the holder assembly. If Flushing and cleaning are required for
proper holder maintenance is not performed, higher torque values will be required to following process steps:
achieve proper sealing.
Storage Solution Removal
The following steps are recommended to keep the holders in operational condition:
Pre-use flushing and cleaning steps are
1. Clean nuts, washers and tie-rods before every use to ensure that they are free of performed to remove shipping solution.
particles. A mild solvent such as IPA and water is generally effective. A minimum of 60 liters of solution per
m2 of membrane should be used in a
2. Use a few drops of suitable lubricant on the tie-rods and between the nuts and combination of flushes and recirculation
the washers, whenever permitted. This could be a food grade vegetable oil or cleanings.
glycerin.
3. Check with your Quality Assurance department before introducing a new Preprocess Equilibration
lubricant into your production facility. A flushing step is utilized to ensure that
the system and cassettes are properly
4. Protect the tie-rods from receiving any blows that could damage their threads,
prepared for processing. The flushing
which can lead to galling. Particular care should be taken when using a torque
solution may be something other than
wrench near the tie-rods.
water depending upon the initial solution
5. Replace any component that is visibly worn or does not spin freely. used for the process.
6. Rates of wear will vary from system to system, but replacement of nuts and Post Process Cleaning
washers should be considered after every ten installations, to maintain high
Cleaning and flushing steps are
sealing force in the holder.
performed after product recovery
7. When new nuts do not spin freely, refurbish the threaded rods with a 5/8 in–18 and is intended to clean and sanitize
die. the system and restore clean water
permeability and process flux to within
8. It is recommended that a set of spare parts be kept on hand at all times.
range. Some applications may require
a two-step cleaning procedure. In these
cases it is imperative that the primary
cleaning agent be flushed completely
from the system to avoid potentially
harmful chemical reactions between
cleaning agents. A full cleaning recipe
encompasses a sequence of flushes and
recirculations to achieve regeneration of
the membrane performance, sanitization,
Pellicon® 3 Cassettes Installation and User Guide 8

depyrogenation and residual flush out Flushing 6. Once the pump has slowly ramped
from the system. To select a cleaning up to the set point, partially close
The flushing step removes residuals
method, identify your application or the retentate valve (V4) to achieve
from the cassette and the system
suspected foulant from the Cleaning the retentate pressure listed above.
piping. Flushing is done primarily with
Agent Selection Chart. Select the
clean water but may be performed with 7. Flush the filter(s) until a total of 20
cleaning agent or agents compatible
cleaning or buffer solutions. The typical liters of water per m² of installed
with your membrane type and
volume of flushing solution for this step filter area has been pumped through
application requirements. Sanitization,
is 20 L/m2 of membrane area. the system to drain. Depending on
depyrogenation and storage agents are
individual removal criteria, the total
chosen similarly. In many instances,
flush volume can be adjusted.
sanitization and depyrogenation may be
accomplished in the same step. 8. At the end of the flush, turn off the
feed pump, fully open retentate and
With consumable agents such as chlorine,
filtrate lines and drain the retentate
monitor the concentration of the agent
and filtrate piping.
over the course of the cleaning cycle
and add additional cleaning agent as Cleaning
needed to maintain the recommended
concentration. The cleaning step applies to installations
®
where all parts of the system will be exposed
Suggested Retentate Pressures for a period of time to the process fluids.
for Flushing and Cleaning Membrane regeneration, system sanitization,
depyrogenation and system storage require a
Membrane Typical System Setup cleaning step. The typical volume of cleaning
Retentate Pressure
Cutoff solution for this step is approximately 10 to
2.8 - 3.4 bar 1. Close the tank drain valve (V1) and the 20 L/m2 of membrane area.
3, 5 kD tank valve (V2).
(40-50 psi)
1. Close the tank drain valves (V1 and V2).
0.5-1.0 bar 2. Fill the tank with the flushing
10, 30, 50 kD 2. Fill the tank with the cleaning
(5.0 - 15 psi) solution.
solution. (Cleaning solution
Suggested retentate pressures are 3. Fully open the retentate valve selection is noted in the Cleaning
recommended to achieve desired feed (V4) (and permeate valve (V5) Selection Chart.)
flow to permeate flow conversion rate. if present) and direct both the
Conversion rate of approximately 30% is retentate and filtrate to drain. 3. Fully open the retentate valve (V4)
desired for 10, 30 and 50 kD devices and (and permeate valve (V5) if present)
4. Open the tank drain valve (V1). and direct both the retentate and
conversion rate of approximately 10% is
desired for 3 and 5 kD devices. 5. Turn on the feed pump and pump filtrate to the tank.
water into the feed port of the 4. Open the tank drain valve (V1).
Pellicon® Holder. The recommended
feed flow rate is 4 to 6 Lpm/m². 5. Turn on the feed pump and pump
Pellicon® 3 Cassettes Installation and User Guide 9

water into the feed port of the Primary Cleaning Agents and Conditions for Organics, Biofilms,
Pellicon® holder. The recommended Biopolymer, Proteins and Polyphenolic
feed flow rate is 4 to 6 L/min/m².
Cleaning Temperature Time
Membrane Concentration pH
6. Once the pump has slowly ramped Agent (°C) (min)
up to the set point, partially close Ultracel® NaOH 0.1 N - 0.5 N 25 - 50 13 - 13.7 30 - 60
the retentate valve (V4) to achieve
NaOH 0.1 N - 0.5 N 25 - 50 13 - 13.7 30 - 60
the retentate pressure listed in the
table above. Biomax® 200 - 500
NaOCl 25 - 50 10 - 11 30 - 60
ppm
7. Recirculate the cleaning solution for the
prescribed time period as noted in the Stable NWP from run to run is achievable after initial exposure. There is an initial
Cleaning Conditions Chart. NWP decline in Pellicon® 3 Cassettes after initial exposure to 0.5N NaOH. Better
Note: If the system is complex and membrane life has been observed at lower concentrations. Total exposure time of
has other associated manifolds, Pellicon® 3 Cassettes with Ultracel® membrane must not exceed 100 hours. Total
ensure that all wetted surfaces in the exposure time of Pellicon® 3 Cassettes with Biomax® membrane must not exceed 200
manifolds are exposed to the cleaning hours. Contact Technical Service for additional Cleaning Agents and Conditions.
solution. All valves that have been
exposed to process fluids should also Sanitization
be exposed to the cleaning solution. It
is good practice to cycle (partially open Sanitization should be performed after the Pellicon® 3 Cassette installation has been
and the partially close) valves at least thoroughly cleaned and flushed to reduce bioburden. Sanitization pressures, flow rates
twice over the course of the cleaning and volumes are identical to those used during Cleaning. Repeat the flushing procedure
cycle to ensure that all wetted internal after sanitization to remove chemical residue from the cassette and system piping.
surfaces of the valve body are exposed NOTE
to the cleaning solution. Use sanitization agents and conditions that are suitable for the process
8. At the end of the cleaning requirements. Sanitization agents must comply with applicable local regulations.
recirculation, direct the retentate
and permeate to drain and continue
Recommended Sanitization Conditions and Agents
to run until the tank is empty Temp Time
Membrane Sanitizing Agent Concentration pH
(without running the pump dry), ( ° C) (min)
then turn off the feed pump, fully NaOCl 200 - 500 ppm 25 - 50 10 - 11 30
open the retentate and filtrate lines
Peracetic acid-containing
and drain the retentate, filtrate Biomax® 1% 10 to 40 3.5 30
Cold Sanitant*
piping and permeate piping.
NaOH 0.1 N - 0.5 N 25 - 50 13 - 13.7 30
9. Repeat the flushing procedure after
Peracetic acid-containing
cleaning to remove chemical residue 1% 10 - 40 3.5 30
Ultracel® Cold Sanitant*
from the cassette and system
piping. NaOH** 0.1N 25 - 50 13 30
Pellicon® 3 Cassettes Installation and User Guide 10

* Typical Peracetic acid-containing cold sanitant contain 20-24% Hydrogen Peroxide,


4.6% Peracetic acid, and 8-10% Acetic acid.
Integrity Testing
** Use milder conditions if effectiveness of agent is still observed. Better membrane Pellicon® 3 filter integrity should be
life is expected with lower concentration of caustic and/or at 20-25 °C for regenerated tested on a cleaned and thoroughly
cellulose membranes with a nominal molecular weight limit of 3 or 5 kDa. flushed system. The presence of residual
cleaning agents can significantly alter
Depyrogenation integrity test results. Integrity testing
should be performed on a water wet
If depyrogenation is required, it should only be performed after the system has been membrane only.
cleaned, sanitized, and flushed. Depyrogenation pressures, flow rates and volumes
are identical to those used during Cleaning. Water for injection should be used during
depyrogenation.

Depyrogenation Conditions and Agents


Depyrogenation Temp Time
Membrane Concentration pH
Agent (° C) (min)
Phosphoric Acid 0.1M 30 - 50 1 30
Biomax®
NaOH 0.1N 25 - 50 13 30
Ultracel® NaOH* 0.1N 25 - 50 13 30
* Use milder conditions if effectiveness of agent is still observed. Better membrane Integrity Test setup
life is expected with lower concentration of caustic and/or at 20-25 °C for regenerated
cellulose membranes with a nominal molecular weight limit of 3 or 5 kDa.

1. Drain the system of water. Drain


the retentate side of the system as
thoroughly as possible.
2. Attach a regulated and filtered air
supply to the feed or retentate
piping, preferably at the high point
of the system.
3. Isolate either the feed or the
retentate piping by closing a valve
or capping the piping if there is no
valve on it (to enable pressurization
of the filter feed channels).
4. Open the permeate line.
Pellicon® 3 Cassettes Installation and User Guide 11

5. Slowly raise the air pressure to the


recommended test value and wait
Measurement of screens is 4 to 6 L/min/m². (The
same conditions MUST be used each
5 minutes to purge residual water
in the permeate line. Permeate line
Normalized Water time NWP is measured in order to
ensure accuracy.)
must be completey emptied before Permeability 6. Recirculate the water for 10 minutes.
measuring air flow. Do not exceed
The normalized water permeability (NWP) Ensure that the pressure and the
the recommended air pressure
for Pellicon® 3 Cassettes should be temperature conditions are stable.
as this may displace water from
the membrane pores and result in established prior to the first product contact 7. Record the feed and permeate flow
excessively high air flow (a false of each filter. New membranes should be rate, feed, retentate, and permeate
failure). Rewet the membrane if this cleaned, flushed and integrity tested before pressures, and the temperature of
occurs. measuring NWP. The NWP measured at the water.
this point is used as a benchmark against
6. After the diffusion air flow pressure subsequent NWP measurements to 8. At the end of the recirculation, fully
stabilizes, measure and record the determine cleaning efficacy. open the retentate and filtrate lines,
air pressure, temperature and the turn off the feed pump, and drain
air flow rate exiting the permeate 1. Close the tank drain valve. the retentate and filtrate piping.
line. The air flow rate may be
2. Fill the tank with filtered, deionized 9. Calculate the NWP:
measured with an air flow meter or
water, water for injection, or
by measuring the air displaced into a
reverse osmosis permeate. The NWP= R•F/[A(Pin + Pout/2)-Pp]
submerged and inverted volumetric
flush water must be extremely pure
cylinder as shown in the figure above.
to avoid fouling the membranes These units yield
7. Compare the measured air flow or introducing other contaminants LMH/bar [liters/m2 •hours•bar]
rate to the specified flow value on into the system. The NWP may
the Certificate of Quality included be measured with solutions other Calculate:
with the cassette. If the measured than water (i.e. storage solution R = Permeate Flow Rate in L/hour
flow rate exceeds the specified flow or buffer) as long as the same P in = Feed Pressure in bar
value, refer to the Troubleshooting conditions and solution are used for P out = Retentate Pressure in bar
section of this guide. each measurement. T = Water Temperatures in °C
3. Fully open the retentate (and Pp = Permeate Pressure (if non-
permeate valve if present) and zero) in bar
direct both the retentate and A = Total Filter area in m2
permeate lines back to the tank. F = Temperature correction factor
from table below.
4. Open the tank drain valve
The acceptance criterion for cleaning
5. Turn on the feed pump and pump efficacy as measured by NWP is
water into the feed port of the membrane and application specific,
Pellicon® holder. The recommended and may vary between plants. Key
feed flow rate for Pellicon® A or C
Pellicon® 3 Cassettes Installation and User Guide 12

importance is stable process flux and no carry over.


Storage
If the NWP decreases significantly from run-to-run, cleaning procedures may be
inadequate. Alternative cleaning agents and procedures should be investigated. Storage solution must be introduced into
Contact a Technical Service Representative for assistance. the system through the cassettes and
the system piping to put the system in a
NWP Temperature Correction Factor (F)* bacteriostatic state in between process
runs. This procedure requires a minimum
T (°F) T (°C) F T (°F) T (°C) F T (°F) T (°C) F of 10 L/m2 of solution. Refer to Typical
125.6 52 0.595 96.8 36 0.793 68.0 20 1.125 System Setup diagram.
123.8 51 0.605 95.0 35 0.808 66.2 19 1.152 1. Close the tank drain valves (V1 and V2).
122.0 50 0.615 93.2 34 0.825 64.4 18 1.181
2. Fill the tank with the storage
120.2 49 0.625 91.4 33 0.842 62.6 17 1.212 solution.
118.4 48 0.636 89.6 32 0.859 60.8 16 1.243
3. Fully open the retentate valve
116.6 47 0.647 87.8 31 0.877 59.0 15 1.276 (V4) (and permeate valve (V5)
114.8 46 0.658 86.0 30 0.896 57.2 14 1.310 if present) and direct both the
113.0 45 0.670 84.2 29 0.915 55.4 13 1.346 retentate and filtrate to the tank.
111.2 44 0.682 82.4 28 0.935 53.6 12 1.383 4. Open the tank drain valve (V1).
109.4 43 0.694 80.6 27 0.956 51.8 11 1.422
5. Turn on the feed pump and pump
107.6 42 0.707 78.8 26 0.978 50.0 10 1.463 water into the feed port of the
105.8 41 0.720 77.0 25 1.000 48.2 9 1.506 Pellicon® holder. The recommended
104.0 40 0.734 75.2 24 1.023 46.4 8 1.551 feed flow rate for Pellicon® 3
cassettes is 4 to 6 L/min/m².
102.2 39 0.748 73.4 23 1.047 44.6 7 1.598
100.4 38 0.762 71.6 22 1.072 42.8 6 1.648 6. Once the pump has slowly ramped
up to the set point, partially close
98.6 37 0.777 69.8 21 1.098 41.0 5 1.699
the retentate valve (V4) to achieve
*Based on Water Fluidity Relative to 25 ºC (77 ºF) Fluidity Value F= (µT ºC/µ25 ºC) or the retentate pressure listed here.
(µT ºF/µ77 ºF)
7. Recirculate the storage solution for
5 to 10 minutes.
Pellicon® 3 Cassettes Installation and User Guide 13

NOTE to prevent possible microbial growth or


If the system is complex and has other associated manifolds, ensure that all contamination.
wetted surfaces in the manifolds are exposed to the solution. All valves that have
Ultracel® membranes can be stored in
been exposed to process fluids should also be exposed. It is good practice to
0.1 N NaOH only for the recommended
cycle (partially open and the partially close) valves at least twice over the course
time period. Longer exposure may result in
of the cycle to ensure that all wetted internal surfaces of the valve body are
membrane damage.
exposed to the solution.
8. Filters may be stored in holders with the feed, retentate and filtrate isolation valves
closed and enough compression on the cassettes to ensure that the fluid cannot
escape or evaporate.
9. The filters may also be removed from the system and placed in sealed bags
or containers. Submerge in storage solution in appropriate container. Filters
should be kept at 2-8 ºC. Process holders with hydraulics can properly hold and
maintain storage solution. Process holders are also designed with a store mode
compression setting that releases to about 60% of the process compression and
ensures that the units will not leak externally.
NOTE
The storage agent used must comply with applicable local regulations.

Recommended Storage Solutions for Pellicon® 3 Cassette Filters


Recommended
Membrane Storage Agent Concentration pH
Time Period
NaOH* 0.05 N 12.7 1 year
Ultracel®
NaOH 0.1 N 13 6 months
0.16M Acetic/0.12M
Acetic/Phosphoric Acid 2-3 1 year
Biomax® Phosphoric Acid
NaOH 0.1 N 13 1 year
*Use milder conditions if effectiveness of agent is still observed. Better membrane life
is expected with lower concentration of caustic and/or at 20-25 °C for regenerated
cellulose membranes with a nominal molecular weight limit of 3 or 5 kDa.
These recommended storage agents will ensure membranes remain wet and will
prevent microbiological growth without damage to the filter. Upon reinstallation, the
filters should be flushed, cleaned and sanitized prior to use.
Filters should be stored in liquid-tight bags or in containers at 2 - 8 °C and should
not be frozen. Fresh solutions should be prepared after the recommended time period
Pellicon® 3 Cassettes Installation and User Guide 14

Troubleshooting
Symptom Possible Cause Remedy Symptom Possible Cause Remedy
Integrity test Areas of membrane Repeat Flushing High NWP Damage to membrane Integrity test cassette
failure or low incompletely wetted Procedure. Flush with or cassette
retention water, retest System and cassette Reduce system
Manifold adapter plate Install manifold over pressurized pressure
not adapter plate Worn cassette Replace cassette
installed
Multiple Improper torque or Check torque
Inadequate pump Use larger pump when leaks in filter compressive force procedure, retorque.
capacity wetting filters stack Check for dirt or
Compression of filter Inspect threads, corrosion on tie rods
stack inadequate nuts and tie rods and or bolts
system compression. Tube fitting Tube/fitting cracked Inspect, replace part
Torque unit again; leaks NPT fitting not Remove old tape, re-
retest. Increase torque wrapped tape adequately with
if necessary. PTFE tape
Temperature change Torque unit again; End of tubing not cut Cut tube square and
since last time holder retest squarely reinstall
was torqued
Missing O-ring or Replace missing
Chemical compatibility Review chemical ferrule on fittings of component, gasket, or
problem compatibility of filter. defective sanitary entire fitting
Replace filter. gasket
Damage to membrane Visually inspect filter, Loose clamp Tighten clamp
or cassette replace as needed
Oversized clamp Replace clamp with
Improper membrane Contact Technical smaller diameter
selection Service. clamp
Low NWP System not vented Ensure system is Leak Dirt or debris in filter Disassemble stack,
Value properly vented so that air is between remove stack and
removed. holder, filters check for debris
System and cassette Flush system and Compression force too Check torque and
not completely flushed cassette. low compressive force
System and cassette Clean system and Immediate Tube or fitting leaks See remedy under
not completely cleaned cassette. complete Tube/Fitting Leaks
pressure loss above
Pellicon® 3 Cassettes Installation and User Guide 15

Symptom Possible Cause Remedy Warranty


Bubbling at Dirt, debris or loose See remedy under
external holder, or chemical Leak Between Filters The applicable warranty for the products
surface of incompatibility above. listed in this publication may be found
filter stack at www.millipore.com/terms (within the
“Terms and Conditions of Sale” applicable
Foaming in Fitting leak See remedy under to your purchase transaction).
system Tubing/Fitting Leaks
Loose holder Retighten holder
Improper flush out of Flush with additional
chemical water
Feed line sucking air Make sure feed line is
fixed below fluid/air
interface
Vortex in feed Add baffle to break
container vortex or reduce
mixing speed
Retentate splashing Fix retentate line
below fluid/in feed air
interface container
Pump cavitation Check feed tubing to
pump connection for
obstruction; remove
obstruction or replace
damaged/collapsed
tubing
Nuts do not Damaged thread on tie Check thread for nicks
move freely rod or nuts or dents; if found
replace tie rod or nuts.
We provide information and advice to our customers on application technologies and regulatory matters to
the best of our knowledge and ability, but without obligation or liability. Existing laws and regulations are
to be observed in all cases by our customers. This also applies in respect to any rights of third parties. Our
information and advice do not relieve our customers of their own responsibility for checking the suitability of
our products for the envisaged purpose.

For worlwide contact information and


technical assistance please visit:
www.sigma-aldrich.com

For additional information and


documentation please contact:
Merck KGaA, Darmstadt, Germany
Corporation with General Partners
Frankfurter Str. 250
64293 Darmstadt, Germany
Phone: + 49 6151-72 0

The vibrant M, Millipore, Pellicon, Ultracel and Biomax are trademarks of Merck KGaA, Darmstadt, Germany or its affiliates. All
other trademarks are the property of their respective owners. Detailed information on trademarks is available via publicly ac-
cessible resources.
MS_AN1065EN00 Rev 6, 02/2019. © 2006-2019 Merck KGaA, Darmstadt, Germany and/or its affiliates. All Rights Reserved.

You might also like