You are on page 1of 16

Contents

Introduction....................................... 2
Membrane Type............................ 2
Pump Capacity............................. 2
Water Quality............................... 2
Installation........................................ 2
Pellicon® 3 88 cm2 and 0.11 m2
Cassette Holder Installation............ 2
Pellicon® 3 0.57 or 1.14 m2 Cassette

Installation and User Benchtop Holder Installation........... 3


Manual Process-Scale Holder

Guide
Installation.................................. 4
Hydraulic Process Scale Holder
Installation.................................. 5
Maintenance ..................................... 7
Flushing and Cleaning......................... 7
Flushing...................................... 8
Cleaning...................................... 8
Sanitization..................................... 10
Depyrogenation................................ 10
Pellicon® 3 Cassettes Integrity Testing............................... 11
Measurement of Normalized Water
Permeability.................................... 11
Storage........................................... 13
Troubleshooting................................ 14
Warranty......................................... 15

The life science business of Merck KGaA, Darmstadt,


Germany operates as MilliporeSigma in the U.S. and Canada.
Pellicon® 3 Cassettes Installation and User Guide 2

Introduction Installation
This guide provides installation and Pellicon® 3 88 cm2 and 0.11 m2 Cassette Holder Installation
maintenance procedures for Pellicon® 3
cassettes. It is not intended to provide NOTE
validation protocols or supporting data Pellicon® 3 88 cm2 and 0.11 m2 Cassettes must be installed in a Pellicon® Cassette
for validation purposes. Please refer Holder, Catalog Number XX42PMINI.
to the cassette Validation Guide for
this information. Please refer to the Holder Installation
Certificate of Quality supplied with each
1. Loosen the nuts on the tie rods of
cassette for specifications.
the holder and remove the nuts,
Membrane Type washers, tie rod spacers and end
Pellicon® 3 Cassette
Install filters containing only one type plate.
of membrane in the filter holder at one
2. Inspect the tie rods and nuts for
time. Do not mix filters with different
pore sizes or nominal molecular weight signs of burrs or stripped threads.
cutoffs. The area of membrane used Replace any damaged components.
depends on the filter surface area Nuts must turn freely on the tie rods
required for your particular application. to ensure proper tightening of the
holder.
Pump Capacity
When operating Pellicon® 3 cassettes, 3. Slide the Pellicon® 3 Cassette onto
select a pump with adequate capacity. the holder so that the tie rods pass
Recommended feed flow rate for through the cutouts in the Pellicon®
cassettes is 4 - 6 liter/min/m2. Optimal 3 Cassette. Repeat to install one or Pellicon® Cassette Holder, XX42PMINI
cross flow will depend on the actual two additional Pellicon® 3 Cassettes
solution being filtered. if required.
NOTE 4. Slide the end plate onto the tie rods and press the end plate against the
Running at sub-optimal feed
fluxes can result in poor mass cassette.
transfer, and larger membrane area
requirements. Higher feed fluxes 5. Install the tie rod spacers, washers and nuts. Hand-tighten the nuts, alternating
can lead to device and/or system from one nut to the other nut.
pressure limitations, and require
larger pump and pipe diameters. NOTE
Recommendations are based Uneven tightening of the nuts can damage the Pellicon® 3 Cassette. Nonparallel
upon achieving optimal system plates or compression of the filters at one end can cause leakage.
performance.
Water Quality
Reverse osmosis (RO) water or water for
injection (WFI) is recommended.
Pellicon® 3 Cassettes Installation and User Guide 3

6. Tighten the nuts using the torque Pellicon® 3 0.57 or 1.14 m2 Cassette Benchtop Holder
wrench and 9/16 in. socket supplied Installation
with the Pellicon® Cassette Holder.
If your wrench has a ratchet style XX42P0060 Pellicon® Cassette Acrylic Filter Holder
head, move the lever to the right XX42PRV60 Pellicon® Cassette Acrylic Holder Low Retentate Volume
lock position to tighten the nuts or XX42P0080 Pellicon® Cassette Stainless Steel Holder
to the left lock position to loosen the XX42P0K80 Pellicon® Cassette Stainless Steel Holder Assembly
nuts. 1. Loosen the nuts on the tie rods of the holder and remove the nuts, washers and end plate.
7. Set the torque wrench at 180 to 200 2. Inspect the tie rods and nuts for signs of burrs or stripped threads. Replace any
in. lb. (20.3 to 22.6 Nm) by holding damaged components. Nuts must turn freely on the tie rods to ensure proper
the spring loaded lock collar down tightening of the holder.
and turning the handle until the 180
to 200 in. lb. (20.3 to 22.6 Nm) 3. Clean the manifold adapter plate (XXPEL3MAP) with an alcohol wipe.
mark aligns with the 0 (zero) mark 4. Slide the manifold adapter plate onto the holder against the holder’s manifold
on the collar. plate, with the solid light grey side labeled “Manifold Side” towards the manifold,
8. Rotate the handle slightly to spring and the wings resting on the tie rods. Slide the manifold adapter plate directly
the lock collar back into place. up against the stainless steel manifold.

9. Attach the socket to the torque 5. Slide the Pellicon® 3 cassette up against the manifold adapter plate. The dark
wrench by firmly pressing the socket grey center with light grey perimeter of the manifold adapter plate should be up
onto the torque wrench drive. against the Pellicon® 3 cassette.

10. Place the socket over one nut. With a 6. Slide additional Pellicon® 3 Cassettes onto the holder if more membrane area is required.
continuous motion, turn each nut ¼ 7. Slide the end plate with the handle side up onto the tie rods.
turn, alternating between the nuts,
until the wrench clicks, which will 8. Tie rod spacers (XX4200066) are available when operating the system with low
indicate it has reached the set force. membrane area. Place enough spacers on each rod so that the length of exposed
threaded tie rod is less than the length of the hex socket. Replace the washers
and nuts. Hand tighten the nuts in a diagonal pattern.
NOTE
Uneven tightening of the nuts can damage the Pellicon® Cassette. Nonparallel
plates or compression of the filters at one end can cause leakage.
9. Tighten the nuts using the torque wrench and 15/16 in. deep hex socket supplied
with the Pellicon® Cassette Holder.
10. Set the torque wrench by turning the knurled locking collar clockwise or by pulling
the handle out to unlock. Turn the torque setting scale clockwise until the 350 in-lb
(40 Nm) mark on the torque scale is opposite the zero mark on the smooth collar.
Turn the knurled locking collar counterclockwise to lock.
Pellicon® 3 Cassettes Installation and User Guide 4

11. Using the torque wrench and 15/16


in. deep hex socket, turn each nut
1/8 turn in the sequence shown nut

here. When the wrench clicks, this


hex nut
washer
washer
indicates that the set force has been
reached. DO NOT OVER TORQUE. stainless steel
stainless steel
bushing
Evenness of torque is critical to bushing top acrylic
pressure plate top
system performance.

M
ILL
pressure plate

IPO
top handle

RE
plate tie rod
manifold

Pellicon®
Filter

Pellicon
® manifold adapter
Gasket Plate
Cassette plate
threaded
tie rod

PE permeate
RM
EA manifolds
TE bottom (four holes)
manifold plate

manifoldGasket
adapter
Plate
plate
permeate ED
outlet port FE pressure
threaded
3/4 in. sanitary E gauge
holes for AT
insertion ® Tri-Clover ME
connection P ER
of tie rods bottom

TE
TA
pressure plate

N
TE
retentate
Pellicon® Holder Tightening Sequence bottom

E
R
outlet port
plate RE
manifold feed inlet
port
TE
NT
AT
Exploded View of Acrylic Pellicon®
NOTE E
Cassette Filter Holder.
Pellicon® 2 and Pellicon® 3 cassettes 1in. sanitary
Tri-Clover®
require 46 kN of axial compressive
FE
connection ED

force to function correctly. If Exploded View of Stainless Steel Pellicon®


installing in a holder from any
other supplier than MilliporeSigma Cassette Filter Holder.
please contact your local Technical Pellicon10_2_06_1.eps

Service team for guidance on the Manual Process-Scale Holder Installation


Pellicon10_2_06_2.eps
appropriate torque settings. 1. Loosen the nuts on the tie rods of the holder and remove the nuts, washers and
NOTE end plate.
Depending on the results of the
Integrity Test, the torque may need 2. Inspect the tie rods and nuts for signs of burrs or stripped threads. Replace any
to be increased at intervals of 50 damaged components. Nuts must turn freely on the tie rods to ensure proper
in-lb (5.6 Nm) up to a maximum of
550 in-lb (62 Nm). tightening of the holder.
12. Recheck the torque prior to use. 3. Clean the manifold adapter plate (XXPEL3MAP) with an alcohol wipe.
NOTE 4. Slide the manifold adapter plate onto the holder against the holder’s manifold
If the filter holder is tightened in a cold
area and moved to a warmer area, plate, with the solid light grey side labeled “Manifold Side” towards the manifold,
loosen the nuts entirely, let the filter and the wings resting on the tie rods. Slide the manifold adapter plate directly
holder warm up, and then retighten to up against the stainless steel manifold.
allow for thermal expansion of the end
plates and cassettes.
Pellicon® 3 Cassettes Installation and User Guide 5

5. Slide the Pellicon® 3 cassette up against the manifold adapter plate. The dark Hydraulic Process Scale
grey center with light grey perimeter of the manifold adapter plate should be up Holder Installation
against the Pellicon® 3 cassette. Slide additional cassettes into the holder against
eachother if more membrane area is required. The Hydraulic Closure with Manual
Controls is intended for use with the
6. Replace the spacers, washers and nuts as shown in Figure 5. Hand-tighten the Millipore Process Scale Filter Holder. It
nuts in a diagonal fashion (see Figure 4). Use a few drops of suitable lubricant on will maintain appropriate compression on
the tie-rods and between the nuts and the washers if required. the cassettes during processing or during
storage of installed filter cassettes.
NOTE
Uneven tightening of the nuts can damage the Pellicon® 3 Cassette. Nonparallel RUN mode is required for process
plates or compression of the filters at one end can cause leakage. operations, including flushing and
cleaning. The system hydraulic pressure
7. Tighten the nuts using the torque wrench and 15/16 in. deep hex socket supplied with
is maintained at 1950 psig ± 100 psig.
the Pellicon® 3 Cassette Holder.
In STORE mode, the system may be
8. Set the torque wrench by turning the knurled locking collar clockwise or by pulling
stored with the Pellicon® 3 cassettes
the handle out to unlock. Turn the torque setting scale clockwise until the 350
installed. The system hydraulic pressure
in-lb (40 Nm) mark on the torque
is maintained at 1200 psig ± 100 psig.
scale is opposite the zero mark on
the smooth collar. Turn the knurled
locking collar counterclockwise to
lock.
9. Using the torque wrench and 15/16
in. deep hex socket, turn each nut
1/8 turn in the sequence shown in
Figure 4. When the wrench clicks,
this indicates that the set force
has been reached. DO NOT OVER
TORQUE. Evenness of torque is
critical to system performance. Process Scale Holder Hydraulic Process Scale Holder
NOTE
Depending on the results of the Integrity Test, the torque may need to be 1. Set the System Mode switch to DUMP.
increased at intervals of 50 in-lb (5.6 Nm) up to a maximum of 550 in-lb (62
NOTE
Nm).
Check the Hydraulic Pressure Gauge
10. Recheck the torque prior to use. on the hydraulic power unit panel to be
Pellicon10_2_06_3
sure the pressure is relieved. Do not
proceed until the hydraulic pressure
reaches 0 psig as read on gauge PI-2.
Pellicon® 3 Cassettes Installation and User Guide 6

2. On the first holder level (if there Holder Pressurization and 8. To relieve hydraulic system pressure
are more than one), remove the Operation at any time, set the System Mode
TC clamps and inserts holding the Selector Switch to DUMP or open
1. The hydraulic fluid reservoir must be
hydraulic cylinders to the clamp rods. the System Pressure Manual Dump
full (indicated on the level indicator
Valve, MV-1 on the left side of the
3. Loosen and remove the hand wheels inside the Hydraulic Power Unit).
Hydraulic Power Unit.
on the clamp rods at the opposite
2. All hydraulic flex hose connections
end plate (of the same holder level). NOTE
must be installed from the Hydraulic
When the cassettes are compressed
4. Slide the end plate away from the Power Unit to the hydraulic
for the first time, reduction of
manifold segment, creating an cylinders.
excess space between the cassettes
opening for cassette insertion into
3. All hardware on the clamp rods can result in significant movement
the holder.
must be in place. Tighten the hand of the hydraulic piston. In some
5. Slide the manifold adapter plate wheels until they are hand-tight. cases, the entire functional length
or a stainless steel support plate of the piston (approximately 2
4. Turn on the air supply to the
onto the holder against the holder’s inches/5 cm) may be extended.
Hydraulic Power Unit. A minimum
manifold plate. If this happens, the piston will be
of 80 psig air pressure should be
exerting force against the hydraulic
6. Slide a Pellicon® 3 Cassette onto the available. Check the air pressure
cylinder instead of against the
holder. Up to ten 1.14 m2 cassettes on the pressure gauge PI-1 on
cassettes and the cassettes will not
may be installed per side if required. Hydraulic Power Unit front panel.
be properly compressed. To achieve
7. Slide the end plate back against the 5. The Power Switch on the Hydraulic correct cassette compression,
cassette assembly. Control Unit HS-1 must be in the relieve the hydraulic system
ON position, and the green Power pressure by setting the System
8. Install the grooved clamp rods, TC Indicator must be illuminated. Mode Selector Switch to DUMP,
clamps, inserts and hand wheels. manually tighten the hand wheels
6. Set the Operating Mode Selector again, and then re-set the Selector
NOTE Switch, HS-3 to RUN or STORE, as
The hand wheels require hand Switch to RUN or STORE. This
required. process may need to be repeated
tightening only. Ensure that each
end plate is flush against the 7. Set the System Mode Selector 1 or 2 times over the first 10-15
cassette assembly and both hand Switch, HS-2 to ENABLE. The minutes of compression to ensure
wheels are snug against the end hydraulic closure system will close the hydraulic piston remains within
plate. No less than one inch of the holder assembly and maintain its functional position during
excess thread on the clamp rod the appropriate pressure on the processing.
exiting the hand wheel should be holder for either RUN or STORE
exposed. mode, as selected.

9. Repeat the procedure for any other


holder levels.
Pellicon® 3 Cassettes Installation and User Guide 7

Caution
The hydraulic closure system must
Maintenance Flushing and
remain connected to the holder
during processing and during
Pellicon® 3 Cassettes must be operated
while under compression from the holder
Cleaning
storage of installed cassettes. The assembly to ensure that the cassettes Flushing and cleaning are required for
will seal properly. The compression is
hydraulic closure system must be supplied by applying torque to the nuts following process steps:
depressurized before connecting or on the threaded tie-rods of the holder
disconnecting fittings. If the position assembly. If proper holder maintenance is Storage Solution Removal
of the Operating Mode Selector not performed, higher torque values will
be required to achieve proper sealing. Pre-use flushing and cleaning steps are
Switch (RUN or STORE) is changed performed to remove shipping solution.
while the system is in ENABLE mode, The following steps are recommended to A minimum of 60 liters of solution per
the hydraulic system pressure will keep the holders in operational condition:
m2 of membrane should be used in a
drop to 1000 psig and then increase 1. Clean nuts, washers and tie-rods before combination of flushes and recirculation
to the correct pressure for the new every use to ensure that they are free cleanings.
selected mode. If the hydraulic of particles. A mild solvent such as IPA
and water is generally effective.
system pressure drops below 1200 Preprocess Equilibration
psig while in RUN mode or 800 psig 2. Use a few drops of suitable lubricant
on the tie-rods and between the A flushing step is utilized to ensure that
while in STORE mode, the relevant
nuts and the washers, whenever the system and cassettes are properly
Low Pressure Alarms (PAL-1 or PAL- permitted. This could be a food prepared for processing. The flushing
2 respectively) will be illuminated grade vegetable oil or glycerin. solution may be something other than
and the Alarm Horn will sound. Once
3. Check with your Quality Assurance water depending upon the initial solution
the problem has been corrected and department before introducing a new used for the process.
hydraulic pressure restored, the lubricant into your production facility.
alarm condition may be cleared by Post Process Cleaning
4. Protect the tie-rods from receiving any
pressing the Alarm Reset – Resume blows that could damage their threads,
Operation Push-button. Cleaning and flushing steps are
which can lead to galling. Particular
care should be taken when using a performed after product recovery and
torque wrench near the tie-rods. are intended to clean and sanitize
the system and restore clean water
5. Replace any component that is
visibly worn or does not spin freely. permeability and process flux to within
range. Some applications may require a
6. Rates of wear will vary from system to two-step cleaning procedure. In these
system, but replacement of nuts and
washers should be considered after cases it is imperative that the primary
every ten installations, to maintain cleaning agent be flushed completely
high sealing force in the holder. from the system to avoid potentially
7. When new nuts do not spin freely, harmful chemical reactions between
refurbish the threaded rods with a 5/8 cleaning agents. A full cleaning recipe
in–18 die. encompasses a sequence of flushes and
8. It is recommended that a set of spare recirculations to achieve regeneration of
parts be kept on hand at all times. the membrane performance, sanitization,
Pellicon® 3 Cassettes Installation and User Guide 8

depyrogenation and residual flush out Flushing 6. Once the pump has slowly ramped
from the system. To select a cleaning up to the set point, partially close
The flushing step removes residuals
method, identify your application or the retentate valve (V4) to achieve
from the cassette and the system
suspected foulant from the Cleaning the retentate pressure listed above.
piping. Flushing is done primarily with
Agent Selection Chart. Select the
clean water but may be performed with 7. Flush the filter(s) until a total of 20
cleaning agent or agents compatible
cleaning or buffer solutions. The typical liters of water per m² of installed
with your membrane type and
volume of flushing solution for this step filter area has been pumped through
application requirements. Sanitization,
is 20 L/m2 of membrane area. the system to drain. Depending on
depyrogenation and storage agents are
individual removal criteria, the total
chosen similarly. In many instances,
flush volume can be adjusted.
sanitization and depyrogenation may be
accomplished in the same step. 8. At the end of the flush, turn off the
feed pump, fully open retentate and
With consumable agents such as chlorine,
filtrate lines and drain the retentate
monitor the concentration of the agent
and filtrate piping.
over the course of the cleaning cycle
and add additional cleaning agent as Cleaning
needed to maintain the recommended
concentration.
®
The cleaning step applies to installations
where all parts of the system will be
Suggested Retentate Pressures exposed for a period of time to the
for Flushing and Cleaning process fluids. Membrane regeneration,
Typical System Setup system sanitization, depyrogenation and
Membrane Retentate Pressure system storage require a cleaning step.
Cutoff bar (psi) 1. Close the tank outlet valve (V1) and The typical volume of cleaning solution
2.8 - 3.4 the tank drain valve (V2). for this step is approximately 10 to 20 L/
3, 5 kD
(40 - 50) 2. Fill the tank with the flushing m2 of membrane area.
0.5 - 1.0 solution. 1. Close the tank outlet valve (V1) and
10, 30, 50 kD
(5.0 - 15) drain valve (V2).
3. Fully open the retentate valve
Suggested retentate pressures are (V4) (and permeate valve (V5) 2. Fill the tank with the cleaning
recommended to achieve desired feed if present) and direct both the solution. (Cleaning solution selection
flow to permeate flow conversion rate. retentate and permeate to drain. is noted in the Cleaning Selection
Conversion rate of approximately 30% is Chart.)
4. Open the tank outlet valve (V1).
desired for 10, 30 and 50 kD devices and
conversion rate of at least 10% and up 5. Turn on the feed pump and pump 3. Fully open the retentate valve
to 30% is desired for 3 and 5 kD devices. water into the feed port of the (V4) (and permeate valve (V5)
Pellicon® Holder. The recommended if present) and direct both the
feed flow rate is 4 to 6 Lpm/m². retentate and filtrate to the tank.
4. Open the tank outlet valve (V1).
Pellicon® 3 Cassettes Installation and User Guide 9

5. Turn on the feed pump and pump water into the feed port Primary Cleaning Agents and Conditions for
of the Pellicon® holder. The recommended feed flow rate is Organics, Biofilms, Biopolymer, Proteins and
4 to 6 L/min/m². Polyphenolic
6. Once the pump has slowly ramped up to the set point, Cleaning Temp Time
partially close the retentate valve (V4) to achieve the Membrane Concentration pH
Agent (°C) (min)
retentate pressure listed in the table above.
NaOH 0.1 N - 1.0 N 25 - 50 13 - 13.7 30 - 60
Biomax®
7. Recirculate the cleaning solution for the prescribed time NaOCl 200 - 500 ppm 25 - 50 10 - 11 30 - 60
period as noted in the Cleaning Conditions Chart. Ultracel® NaOH 0.1 N - 0.5 N 25 - 50 13 - 13.7 30 - 60
NOTE Stable NWP from run to run is achievable after initial exposure.
If the system is complex and has other associated There is an initial NWP decline in Pellicon® 3 Cassettes after
manifolds, ensure that all wetted surfaces in the manifolds initial exposure to 0.5 N NaOH for Ultracel® and 1.0 N NaOH
are exposed to the cleaning solution. All valves that have for Biomax®. Better membrane life has been observed at lower
been exposed to process fluids should also be exposed to concentrations. At the higher concentrations, total exposure
the cleaning solution. It is good practice to cycle (partially time of Pellicon® 3 Cassettes with Ultracel® membrane must
open and the partially close) valves at least twice over not exceed 100 hours and total exposure time of Pellicon® 3
the course of the cleaning cycle to ensure that all wetted Cassettes with Biomax® membrane must not exceed 200 hours.
internal surfaces of the valve body are exposed to the Contact Technical Service for additional Cleaning Agents and
cleaning solution. Conditions.
8. At the end of the cleaning recirculation, direct the
retentate and permeate to drain and continue to run
until the tank is empty (without running the pump dry),
then turn off the feed pump, fully open the retentate
and filtrate lines and drain the retentate and filtrate
(permeate) piping.
9. Repeat the flushing procedure after cleaning to remove
chemical residue from the cassette and system piping.
Pellicon® 3 Cassettes Installation and User Guide 10

Sanitization Depyrogenation
Sanitization to reduce bioburden should be performed after the If depyrogenation is required, it should only be performed
Pellicon® 3 Cassette installation has been thoroughly cleaned after the system has been cleaned, sanitized, and flushed.
and flushed. Sanitization pressures, flow rates and volumes are Depyrogenation pressures, flow rates and volumes are identical
identical to those used during Cleaning. Repeat the flushing to those used during Cleaning. Water for injection should be
procedure after sanitization to remove chemical residue from the used during depyrogenation.
cassette and system piping.
Depyrogenation Conditions and Agents
NOTE
Use sanitization agents and conditions that are suitable Depyrogenation Temp Time
Membrane Concentration pH
for the process requirements. Sanitization agents must Agent (°C) (min)
comply with applicable local regulations. Phosphoric
0.1 M 30 - 50 1 30
Biomax ® Acid
Recommended Sanitization Conditions and Agents NaOH 0.1 N 25 - 50 13 30
Sanitizing Temp Time Ultracel ®
NaOH* 0.1 N 25 - 50 13 30
Membrane Concentration pH
Agent (°C) (min)
*Use milder conditions if effectiveness of agent is still observed.
NaOCl 200 - 500 ppm 25 - 50 10 - 11 30
Better membrane life is expected with lower concentration of
Peracetic caustic and/or at 20-25°C for regenerated cellulose membranes
acid- with a nominal molecular weight limit of 3 or 5 kDa.
Biomax® containing 1% 10 - 40 3.5 30
Cold
Sanitant*
NaOH 0.1 N - 1.0 N 25 - 50 13 - 13.7 30
Peracetic
acid-
containing 1% 10 - 40 3.5 30
Ultracel® Cold
Sanitant*
NaOH** 0.1 N 25 - 50 13 30
* Typical Peracetic acid-containing cold sanitant contain 20-24%
Hydrogen Peroxide, 4.6% Peracetic acid, and 8-10% Acetic acid.
** Use milder conditions if effectiveness of agent is still
observed. Better membrane life is expected with lower
concentration of caustic and/or at 20-25°C for regenerated
cellulose membranes with a nominal molecular weight limit of 3
or 5 kDa.
Pellicon® 3 Cassettes Installation and User Guide 11

Integrity Testing 5. Slowly raise the air pressure to


the recommended test value and
Measurement of
Pellicon® 3 filter integrity should be tested wait at least 5 minutes to purge
residual water in the permeate line.
Normalized Water
on a cleaned and thoroughly flushed
system. The presence of residual cleaning Permeate line must be completey Permeability
agents can significantly alter integrity emptied before measuring air flow.
test results. Integrity testing should be Do not exceed the recommended air The normalized water permeability
performed on a water wet membrane pressure as this may displace water (NWP) for Pellicon® 3 Cassettes should
only. from the membrane pores and result be established prior to the first product
in excessively high air flow (a false contact of each filter. New membranes
failure). Rewet the membrane if this should be cleaned, flushed and integrity
occurs. tested before measuring NWP. The
NWP measured at this point is used as
6. After the diffusion air flow pressure a benchmark against subsequent NWP
stabilizes, measure and record the air measurements to determine cleaning
pressure, temperature and the air flow efficacy.
rate exiting the permeate line. The air
flow rate may be measured with an 1. Close the tank outlet valve.
air flow meter or by measuring the
2. Fill the tank with filtered, deionized
air displaced into a submerged and
water, water for injection, or
inverted volumetric cylinder as shown
reverse osmosis permeate. The
in the figure above.
flush water must be extremely pure
Integrity Test setup to avoid fouling the membranes
7. Compare the measured air flow rate
to the specified flow value on the or introducing other contaminants
Certificate of Quality included with into the system. The NWP may be
1. Drain the system of water. Drain the cassette. If the measured flow measured with solutions other than
the retentate side of the system as rate exceeds the specified flow value, water (i.e. storage solution or buffer)
thoroughly as possible. refer to the Troubleshooting section as long as the same conditions
2. Attach a regulated and filtered air of this guide. and solution are used for each
supply to the feed or retentate subsequent measurement.
piping, preferably at the high point of 3. Fully open the retentate (and
the system. permeate valve if present) and direct
3. Isolate either the feed or the both the retentate and permeate
retentate piping by closing a valve lines back to the tank.
or capping the piping if there is no 4. Open the tank outlet valve.
valve on it (to enable pressurization
of the filter feed channels).
4. Open the permeate line.
Pellicon® 3 Cassettes Installation and User Guide 12

5. Turn on the feed pump and pump The acceptance criterion for cleaning efficacy as measured by NWP is membrane and
water into the feed port of the application specific, and may vary between plants. Key importance is stable process
Pellicon® holder. The recommended flux and no carry over.
feed flow rate for Pellicon® A, C or
If the NWP decreases significantly from run-to-run, cleaning procedures may be
D screens is 4 to 6 L/min/m². (The
inadequate. Alternative cleaning agents and procedures should be investigated.
same conditions MUST be used each
Contact a Technical Service Representative for assistance.
time NWP is measured in order to
ensure accuracy.)
NWP Temperature Correction Factor (F)*
6. Recirculate the water for 10 minutes.
T (°F) T (°C) F T (°F) T (°C) F T (°F) T (°C) F
Ensure that the pressure and the
temperature conditions are stable. 125.6 52 0.595 96.8 36 0.793 68.0 20 1.125
123.8 51 0.605 95.0 35 0.808 66.2 19 1.152
7. Record the feed and permeate flow
rate, feed, retentate, and permeate 122.0 50 0.615 93.2 34 0.825 64.4 18 1.181
pressures, and the temperature of 120.2 49 0.625 91.4 33 0.842 62.6 17 1.212
the water. 118.4 48 0.636 89.6 32 0.859 60.8 16 1.243
8. At the end of the recirculation, fully 116.6 47 0.647 87.8 31 0.877 59.0 15 1.276
open the retentate and filtrate lines, 114.8 46 0.658 86.0 30 0.896 57.2 14 1.310
turn off the feed pump, and drain the 113.0 45 0.670 84.2 29 0.915 55.4 13 1.346
retentate and filtrate piping.
111.2 44 0.682 82.4 28 0.935 53.6 12 1.383
Calculate the NWP: 109.4 43 0.694 80.6 27 0.956 51.8 11 1.422
NWP = R * F / (A * [(Pin + Pout)/2 - Pp]) 107.6 42 0.707 78.8 26 0.978 50.0 10 1.463
105.8 41 0.720 77.0 25 1.000 48.2 9 1.506
These units yield 104.0 40 0.734 75.2 24 1.023 46.4 8 1.551
LMH/bar [liters/(m2 •hours•bar)] 102.2 39 0.748 73.4 23 1.047 44.6 7 1.598
100.4 38 0.762 71.6 22 1.072 42.8 6 1.648
Calculate:
R = Permeate Flow Rate in L/hour 98.6 37 0.777 69.8 21 1.098 41.0 5 1.699
Pin = Feed Pressure in bar *Based on Water Fluidity Relative to 25°C (77°F) Fluidity Value F= (µT °C/µ25 °C) or (µT
Pout = Retentate Pressure in bar °F/µ77 °F)
T = Water Temperatures in °C
Pp = Permeate Pressure (if non-zero)
in bar
A = Total Filter area in m2
F = Temperature correction factor
from table
Pellicon® 3 Cassettes Installation and User Guide 13

Storage solution in appropriate container. Filters should be kept


at 2-8°C when removed from the holder. Process holders
Storage solution must be introduced into the system through with hydraulics can properly hold and maintain storage
solution. Process holders are also designed with a store
the cassettes and the system piping to put the system in a
mode compression setting that releases to about 60% of
bacteriostatic state in between process runs. This procedure the process compression and ensures that the units will not
requires a minimum of 10 L/m2 of solution. Refer to Typical leak externally.
System Setup diagram.
NOTE
1. Close the tank outlet valve (V1) and drain valve (V2). The storage agent used must comply with applicable local
2. Fill the tank with the storage solution. regulations.

3. Fully open the retentate valve (V4) (and permeate valve Recommended Storage Solutions for Pellicon® 3
(V5) if present) and direct both the retentate and filtrate Cassette Filters
to the tank.
Storage Recommended
4. Open the tank outlet valve (V1). Membrane Concentration pH
Agent Time Period
5. Turn on the feed pump and pump water into the feed port NaOH* 0.05 N 12.7 1 year
of the Pellicon® holder. The recommended feed flow rate Ultracel®
NaOH 0.1 N 13 6 months
for Pellicon® 3 cassettes is 4 to 6 L/min/m².
Acetic/ 0.16 M
6. Once the pump has slowly ramped up to the set point, Phosphoric Acetic/0.12 M 2-3 1 year
partially close the retentate valve (V4) to achieve the Biomax® Acid Phosphoric Acid
retentate pressure listed here.
NaOH 0.1 N 13 1 year
7. Recirculate the storage solution for 5 to 10 minutes.
*Use milder conditions if effectiveness of agent is still
NOTE observed. Better membrane life is expected with lower
If the system is complex and has other associated concentration of caustic and/or at 20-25°C for regenerated
manifolds, ensure that all wetted surfaces in the manifolds
cellulose membranes with a nominal molecular weight limit of 3
are exposed to the solution. All valves that have been
exposed to process fluids should also be exposed. It is or 5 kDa.
good practice to cycle (partially open and the partially These recommended storage agents will ensure membranes
close) valves at least twice over the course of the cycle to remain wet and will prevent microbiological growth without
ensure that all wetted internal surfaces of the valve body damage to the filter. Upon reinstallation, the filters should be
are exposed to the solution. flushed, cleaned and sanitized prior to use.
8. Filters may be stored in holders with the feed, retentate and
Filters should not be frozen. Fresh solutions should be prepared
filtrate isolation valves closed and enough compression on
the cassettes to ensure that the storage fluid cannot escape after the recommended time period to prevent possible
or evaporate. microbial growth or contamination.

9. The filters may also be removed from the system and Ultracel® membranes can be stored in 0.1 N NaOH only for
placed in sealed bags or containers. Submerge in storage the recommended time period. Longer exposure may result in
membrane damage.
Pellicon® 3 Cassettes Installation and User Guide 14

Troubleshooting Symptom Possible Cause Remedy


Damage to membrane
Symptom Possible Cause Remedy
Integrity test cassette
or cassette
Repeat Flushing High NWP System and cassette Reduce system
Areas of membrane
Procedure. Flush with over pressurized pressure
incompletely wetted
water, retest Worn cassette Replace cassette
Manifold adapter plate Install manifold Check torque
not installed adapter plate Multiple procedure, retorque.
Improper torque or
Inadequate pump Use larger pump when leaks in filter Check for dirt or
compressive force
capacity wetting filters stack corrosion on tie rods
Inspect threads, or bolts
nuts and tie rods and Tube/fitting cracked Inspect, replace part
Compression of filter system compression. Remove old tape, re-
NPT fitting not
Integrity test stack inadequate Torque unit again; tape adequately with
wrapped
failure or low retest. Increase torque PTFE tape
retention if necessary. End of tubing not cut Cut tube square and
Temperature change squarely reinstall
Torque unit again;
since last time holder Tube fitting Missing O-ring or
retest Replace missing
was torqued leaks ferrule on fittings of
component, gasket, or
Review chemical defective sanitary
Chemical compatibility entire fitting
compatibility of filter. gasket
problem
Replace filter. Loose clamp Tighten clamp
Damage to membrane Visually inspect filter, Replace clamp with
or cassette replace as needed Oversized clamp smaller diameter
Improper membrane Contact Technical clamp
selection Service. Disassemble stack,
Ensure system is Leak Dirt or debris in filter remove stack and
System not vented
vented so that air is between check for debris
properly
removed. holder, filters Compression force too Check torque and
Low NWP
System and cassette Flush system and low compressive force
Value
not completely flushed cassette. Immediate See remedy under
System and cassette Clean system and complete Tube or fitting leaks Tube/Fitting Leaks
not completely cleaned cassette. pressure loss above
Pellicon® 3 Cassettes Installation and User Guide 15

Symptom Possible Cause Remedy Warranty


Bubbling
Dirt, debris or loose See remedy under The applicable warranty for the products listed in this
at external
holder, or chemical Leak Between Filters publication may be found at www.millipore.com/terms (within
surface of
incompatibility above. the “Terms and Conditions of Sale” applicable to your purchase
filter stack
transaction).
See remedy under
Fitting leak
Tubing/Fitting Leaks
Loose holder Retighten holder
Improper flush out of Flush with additional
chemical water
Make sure feed line is
Feed line sucking air fixed below fluid/air
interface
Add baffle to break
Foaming in Vortex in feed
vortex or reduce
system container
mixing speed
Fix retentate line
Retentate splashing below fluid/in feed air
interface container
Check feed tubing to
pump connection for
obstruction; remove
Pump cavitation
obstruction or replace
damaged/collapsed
tubing
Check thread for nicks
Nuts do not Damaged thread on tie
or dents; if found
move freely rod or nuts
replace tie rod or nuts.
We provide information and advice to our customers on application technologies and regulatory matters to
the best of our knowledge and ability, but without obligation or liability. Existing laws and regulations are
to be observed in all cases by our customers. This also applies in respect to any rights of third parties. Our
information and advice do not relieve our customers of their own responsibility for checking the suitability of
our products for the envisaged purpose.

For worlwide contact information and


technical assistance please visit:
www.sigma-aldrich.com

For additional information and


documentation please contact:
Merck KGaA, Darmstadt, Germany
Corporation with General Partners
Frankfurter Str. 250
64293 Darmstadt, Germany
Phone: + 49 6151-72 0

The vibrant M, Millipore, Pellicon, Ultracel and Biomax are trademarks of Merck KGaA, Darmstadt, Germany or its affiliates.
All other trademarks are the property of their respective owners. Detailed information on trademarks is available via publicly
accessible resources.

MS_AN1065EN00 Rev 7, 04/2021. © 2006-2021 Merck KGaA, Darmstadt, Germany and/or its affiliates. All Rights Reserved.

You might also like