Fetal Doppler BT-250 Quick Guide

Refer to the manual for safe and complete use.

1. Power button
(1) Connect power cord and adaptor.
(2) To turn on the device, push power button[ ].

Power Button

2. Probe connection
(1) Connect Doppler probe to the right panel.

3. Setup Mode
(1) Push the control knob in the main display,
then [Setup Mode] is displayed.
(2) In [Setup Mode], a user can set each item.
(3) Use the control knob to select or move.
[Select] [Move]

Control Knob Button

4. Measurement
(1) Place the transducer directly against the abdomen,
just above the point where the pelvic bones meet..
(2) Search for the fetal heart by slowly moving the
probe around until the fetal heart sounds are heard.
(3) Adjust the volume using the control knob.

5. Data saving and checking


(1) To save data, push [SAVE] button.
When you push this button( , ), [ ] icon
turns with light.
(2) You can check the saved data pushing
button.
(3) Select the ID then you can check the relevant record.

6. Display & Symbols


6-1. Display

P/N : 250-EN-QG-R00 Bistos Co., Ltd. 2014.03.14


[Graph mode] [Numeric mode]
6-2. Symbol
Symbol Name Description
Battery Status This symbol indicates the battery capacity.
Saving Status This symbol indicates data saving.
7.
Cleaning and disinfection
7-1. Main body
(1) Keep the external surface clean and free of dust, dirt, and residual liquids. Clean
with a damp cloth using mild soap and water or hospital approved nonabrasive
disinfectants.
7-2. Probe
(1) Wipe the device with a sterile wipe soaked in enzymatic detergent safe for use
with metal instruments. Wipe the exterior of the device three times. Prepare the
detergent according to the manufacturer’s probe recommendations.
(2) Scrub the transducer with enzymatic detergent using soft bristled brush for five (5)
minutes. DO NOT IMMERSE THE PROBE.
(3) Wipe the transducer three (3) times with sterile water to remove soap residue.
(4) Wipe the transducer with a sterile wipe soaked in Cidex™. Wipe all exterior
surfaces of the transducer three (3) times.
(5) Wipe the transducer three (3) times with sterile water to remove Cidex residue.
(6) Dry the device thoroughly with a sterile soft towel or gauze surgical sponge.
(7) Wrap the dry device in a fresh sterile soft towel or transparent sterile wrap for
storage until next use.

8. Troubleshooting and maintenance


If there are some problems in the device, please contact the service center.
(1) When the device power does not turn on.
① Check the connection status between power cord and adaptor.
② Change the adaptor.
(2) Change of internal Ni-MH battery.
① The internal battery must be handled by the company’s technician only. Do
not attempt to open the device.
② Avoid storing battery above 140°F. If clothing or skin comes in contact with
material from inside the battery, immediately wash with plenty of clean water.
th
7 Fl., A Bldg., Woolim Lions Valley 5-cha, 302, Galmachi-ro, Jungwon-gu,
Seongnam-si, Gyeonggi-do, Korea
[Global] TEL : 82-31-750-0340 www.bistos.co.kr bistos@bistos.co.kr
[US] TEL : 1-210-979-3883 / 1-210-439-6860 jhpark@bistosamerica.com

P/N : 250-EN-QG-R00 Bistos Co., Ltd. 2014.03.14

You might also like