Professional Documents
Culture Documents
Technical Manual
Release 8 March 2006 (Rev. 7)
TECHNICAL MANUAL
Revision history
Revision history
2 04.09.01 i, ii, iii, 20, 23, 47, 73, I.I. 9”-12”-16” wiring and components plate
93, 95, 97, 99, 101, unification.
102, 103, 104, 105, Toshiba I.I. 16” new model.
106, 107, 108, 109, (Ref. RDM 5101, RDM 5218)
110, 111, 112
3 08.10.03 All New procedure for adjusting the HV power
supply.
Change in adjustment/checks procedures.
Spare Parts list update.
(Ref. RDM 5283, RDM 5432)
4 09.12.03 i, ii, iii, 10, 12, 13, New Philips I.I. 12" STD.
15, 16, 25, 28, 33, 34, New CCD camera N23.
35, 47, 48, 51, 52, (Ref. RDM 5686, RDM 5715)
54, 56, 57,
from 59 to 63,
from 65 to 124
5 15.03.04 All New Power Supply board A1.
New CCD camera N23 DRF (Dynamic Recursive
Filter).
(Ref. RDM 5763, RDM 5780)
(Rev. 7) IB SYSTEM
TECHNICAL MANUAL
Revision history
IB SYSTEM (Rev. 7)
TECHNICAL MANUAL
Contents
Contents
1. INTRODUCTION 1
1.1 Icons appearing in the manual ............................................................... 2
2. SAFETY ASPECTS 3
2.1 Risks associated with the installation, maintenance and
use of the equipment .............................................................................. 4
2.2 Electromagnetic compatibility................................................................. 6
2.3 Electromagnetic environment ................................................................. 7
2.4 Environmental hazards and Disposal ..................................................... 8
2.5 Symbols used ......................................................................................... 9
4. TECHNICAL CHARACTERISTICS 15
4.1 Standard versions................................................................................. 15
4.1.1 General characteristics ........................................................................ 15
4.1.2 Electrical characteristics...................................................................... 16
4.1.3 Image intensifier.................................................................................. 16
4.1.4 N23 TV chain ...................................................................................... 17
4.1.5 N33 TV chain ...................................................................................... 18
4.1.6 N23 TV chain Monitor.......................................................................... 19
4.1.7 N33 TV chain Monitor.......................................................................... 20
4.2 Digital Acquisition System versions ...................................................... 21
4.2.1 General characteristics ........................................................................ 21
4.2.2 Electrical characteristics...................................................................... 21
4.2.3 Image intensifier.................................................................................. 22
4.3 Environmental conditions (valid for all models)..................................... 23
4.4 Applicable standards ............................................................................ 24
5. ASSEMBLY 25
5.1 Installing the component plate.............................................................. 25
5.1.1 Remote-controlled tables ..................................................................... 25
5.1.2 Tilting tables ....................................................................................... 25
5.2 Assembling the I.I. on the radiological table.......................................... 26
5.3 Assembling the monitor ........................................................................ 27
(Rev. 7) i IB SYSTEM
TECHNICAL MANUAL
Contents
7. CONNECTION OF ACCESSORIES 38
7.1 Connection of generators ...................................................................... 38
7.1.1 Connection to generators..................................................................... 38
7.1.2 Connection of VARIAC Module with I.I. system and with conventional
generators........................................................................................... 40
7.2 Connecting the tables ........................................................................... 42
7.2.1 Connection with APOLLO remote controlled table................................. 42
7.2.2 Connection with MERCURY 332 remote controlled table ...................... 43
7.2.3 Connection with SYMPHONY remote controlled table ........................... 44
7.2.4 Connection with VISION tilting table.................................................... 45
7.2.5 Assembling and wiring the keypad for over-table serial changers
(SAG 4 - SAG 6) on VIROMATIC ........................................................... 46
7.3 Connection with digital acquisition system (DIVA-D)............................. 48
8. FUNCTIONAL CHECKS 49
8.1 Description of the controls .................................................................... 49
8.2 Functional checks................................................................................. 51
8.3 Checking the image quality ................................................................... 52
8.3.1 Jumpers X10 and X11 setting on Power Supply board A1 .................... 52
8.3.2 Spatial resolution ................................................................................ 53
8.3.3 Grey scale (Leeds Test Object GS2) ...................................................... 54
8.3.4 Loss of contrast and noise (Leeds Test Object N3)................................. 55
8.3.5 Detail/contrast ratio (Leeds Test Object TO.10) .................................... 56
9. ADJUSTMENT 58
Camera adjustment ...................................................................................... 58
9.1.1 Focal distance ..................................................................................... 60
9.1.2 Orientation and centering.................................................................... 61
9.1.3 Diaphragm and automatic kV adjustment............................................ 62
9.1.4 Adjusting the focus ............................................................................. 64
9.1.5 Cleanliness of the optics and output window and blemishes................. 65
IB SYSTEM ii (Rev. 7)
TECHNICAL MANUAL
Contents
IB SYSTEM (Rev. 7)
TECHNICAL MANUAL
Introduction
1. INTRODUCTION
G
NOTE:
The present manual is updated for the product it is sold with in order to
grant an adequate reference in performing diagnostics and repair
operation normally carried out by the service engineer.
The manual may not reflect changes to the product not impacting service
operations.
G
NOTE:
The operating instructions are contained in the manuals of the
equipment associated with the IB SYSTEM, because the IB SYSTEM itself
does not incorporate any commands that can be used by the operator.
A brief overview of the available commands is given in paragraph 8.1.
(Rev. 7) 1 IB SYSTEM
TECHNICAL MANUAL
Introduction
The equipment must be used according to the procedures set out in this
manual, and must never be used for any purposes other than those
specified in the manual.
The equipment must be operated by qualified personnel with the
necessary knowledge of radiation safety procedures.
The manufacturer shall not be held responsible for any breakage, injuries
or malfunctions resulting from improper use of the equipment, or from
failure to perform the prescribed maintenance.
The IB SYSTEM is a medical electrical device, and as such may only be
used under the direct supervision of a doctor or highly qualified
technician, with the necessary knowledge of radiation safety procedures.
This manual is intended to provide authorised technicians with the
instructions for correctly installing and servicing the equipment.
G
THIS ICON INDICATE A NOTE; please read thoroughly the items
marked by this picture.
THIS ICON INDICATE A WARNING; the items marked by this icon refer
to the safety aspects of the patient and/or of the operator.
IB SYSTEM 2 (Rev. 7)
TECHNICAL MANUAL
Safety aspects
2. SAFETY ASPECTS
The I.I. (Image Intensifier) essentially consists of a glass vacuum tube
(similar to the tube of a television set) protected against mechanical
damage by a metal enclosure designated the envelope.
On the side of the I.I. there is a square enclosure which houses the
HV (high voltage) power supply. This power supply receives a 24Vdc
input and generates the working voltages for the I.I. which are in the
order of 30kV.
G
NOTE:
Installation of the IB SYSTEM an a VSM radiological remote or tilting
table has been designed in such a way to grant maximum protection
against X-ray radiation. Manuals of the radiological tables indicate areas
of exposure and typical doses delivered in those areas during operation.
(Rev. 7) 3 IB SYSTEM
TECHNICAL MANUAL
Safety aspects
WARNING:
When handling the I.I., always be aware of the following risks:
• Implosion risks: if dropped or subjected to excessive stress, the I.I.
tube may break, ejecting potentially toxic gases, dust and pieces of
glass. The entrance window and the output window are particularly
vulnerable areas which must therefore be handled with great care.
• High voltages: the HV power supply board includes voltages as high
as 30kV (30,000 Volts). Do not touch any part of the power supply
with bare hands or with metal objects to avoid the risk of electrical
shock. Note that these voltages persist for a few dozen seconds after
the power supply has been disconnected. There are no power on
indicators on the I.I. or on the HV power supply board. Therefore,
before attempting any operations on the I.I., always make sure that
the power cable has been physically disconnected.
The above considerations also apply to the monitor. In this case, the
delicate part is the tube. Some parts with dangerous voltages are
accessible through the rear cover plate, which must be removed for
connecting the cables.
• X Rays: certain maintenance operations require the performance of an
X-ray exposure. Although the I.I. envelope is sufficiently shielded with
lead, the service technician must in any case take every precaution to
minimise his exposure to radiation.
IB SYSTEM 4 (Rev. 7)
TECHNICAL MANUAL
Safety aspects
(Rev. 7) 5 IB SYSTEM
TECHNICAL MANUAL
Safety aspects
The IB SYSTEM must be kept turned off when high frequency surgical
devices or similar devices are in use.
IB SYSTEM 6 (Rev. 7)
TECHNICAL MANUAL
Safety aspects
Electromagnetic
Compliance EMC environment requirements
emissions
IB SYSTEM can be connected to a line
Class A
other than for domestic appliance.
Radiated and conducted IB SYSTEM generates radio frequency
emissions energy only internally. R.F. energy
CISPR 11 Group I generated does not cause interference
with electrical appliances used in its
vicinity
Harmonic emission IB SYSTEM can be connected to a line
Compliant
EN 61000-3-2 other than for domestic appliance.
Voltage spikes/flicker
IB SYSTEM can be connected to a line
emission Compliant
other than for domestic appliance.
EN 61000-3-3
Susceptibility
EN 60601-1-2 IB SYSTEM can be connected to a line
Compliant
(EN 61000-4-2 /3 /4 /5 other than for domestic appliance.
/6 /8 /11)
10mW 0.3
100mW 1
1W 3
10W 8
100W 30
(Rev. 7) 7 IB SYSTEM
TECHNICAL MANUAL
Safety aspects
The equipment contains materials which, at the end of its life cycle, must
be disposed of at the collection centres designated by the Local Health
Units.
In particular, the equipment contains the following materials and/or
components:
• control and power supply unit: iron, copper, fibreglass, electrolyte,
non biodegradable plastic material, rubber
• monitor: iron, copper, fibreglass, glass, non biodegradable plastic
material
• image intensifier and camera: iron, lead, aluminium, copper, non
biodegradable plastic material, fibreglass, glass, phosphorus
compounds
IB SYSTEM 8 (Rev. 7)
TECHNICAL MANUAL
Safety aspects
Symbol Description
∼ Alternating current
Protective ground
Functional ground
(Rev. 7) 9 IB SYSTEM
TECHNICAL MANUAL
Identification of the equipment
IB SYSTEM 10 (Rev. 7)
TECHNICAL MANUAL
Identification of the equipment
(Rev. 7) 11 IB SYSTEM
TECHNICAL MANUAL
Identification of the equipment
Both the manufacturer label and the Villa Sistemi Medicali label are
present. The latter identifies the monitor as a part of the IB SYSTEM. The
code and serial numbers on this label therefore coincide with those on
the main label.
IB SYSTEM 12 (Rev. 7)
TECHNICAL MANUAL
Identification of the equipment
Both the manufacturer label, and the Villa Sistemi Medicali labels are
present. The latter identifies the I.I. tube as part of the IB SYSTEM. The
code and serial number on this label therefore match those on the main
label.
(Rev. 7) 13 IB SYSTEM
TECHNICAL MANUAL
Identification of the equipment
Made in Italy
Code: N23XXXXX Type: N23 CCD
Line : 24V DC 20W MANUFACTURER: NICAL SPA
Serial Number: XX.XXX Date: MM/YY
IB SYSTEM 14 (Rev. 7)
TECHNICAL MANUAL
Technical characteristics
4. TECHNICAL CHARACTERISTICS
Description Data
Manufacturer Villa Sistemi Medicali S.p.a.
Via della Azalee, 3
20090 – Buccinasco – MI – Italia
Installation under-table (remote-controlled tables)
over-table (tilting tables)
Table compatibility Apollo, Mercury, Symphony, Vision, Viromatic
Generator typical configuration High frequency generators: Genius HF-A /
(*) G 100 RF/A
Genius conventional generators
Available commands • select manual/automatic kV adjustment
• up/down image inversion
• left/right image inversion
• select I.I. field (zoom)
Position of controls • on remote console (Apollo, Mercury,
Symphony)
• on serial changer keyboard (Vision)
• on dedicated keypad on board the serial
changer (Viromatic)
Monitor controls • brightness
• contrast
(*) Other generator models may be compatible, but it is left to the installer to assess
compatibility on the bases of the information contained in this manual and in the radiological
table manual.
(Rev. 7) 15 IB SYSTEM
TECHNICAL MANUAL
Technical characteristics
Description Data
Power supply 220/230/240Vac ±10% mono-phase
Frequency 50/60Hz
Current 0.5A
Power 120VA
Isolation class Class I with type B applied parts (IEC 601-1)
Type of installation Permanent installation
IB SYSTEM 16 (Rev. 7)
TECHNICAL MANUAL
Technical characteristics
Description Data
Model Nical N23 50 NH 0A (625 lines 50Hz)
Nical N23 50 NH 3A (625 lines 100Hz)
Type of camera Solid state CCD ½” low persistence
Acquisition matrix 582x752 pixel, 10 bit
Scanning system CCIR 625 lines 50Hz
Aspect ratio 4:3
Bandwidth 20MHz ± 3dB
Electronic circle Internal generator
Antivignetting circuit I.I. brightness distortion compensation: up/down,
left/right, vertical/horizontal parabola
Video gain Automatic control
Automatic kV/mA adjustment Standard
Last image hold Standard
Recursive digital filter for image Variable weighting k=2, 4, 8, 16
noise reduction
DRF (Dynamic Recursive Filter) Dynamic filter for noise reduction which
function optimises image quality both statically and for
moving images.
Activated by default.
Image inversion Vertical and horizontal
Automatic kV/mA adjustment Support for automatic generator (high frequency
or conventional) brightness adjustment (ABC).
The function can be disabled by the operator.
(Rev. 7) 17 IB SYSTEM
TECHNICAL MANUAL
Technical characteristics
Description Data
Model Nical N23
Type of camera Solid state CCD ½” high resolution
Acquisition matrix 1024x1024 pixel, 12 bit
Scanning system Progressive scanning
Aspect ratio 1:1
Bandwidth 40MHz ± 3dB
Electronic circle Internal generator
Antivignetting circuit I.I. brightness distortion compensation: up/down,
left/right, vertical/horizontal parabola
Video gain Automatic control
Automatic kV/mA adjustment Standard
Last image hold Standard
Recursive digital filter for image Variable weighting k=2, 4, 8, 16
noise reduction
DRF (Dynamic Recursive Filter) Dynamic filter for noise reduction which
function optimises image quality both statically and for
moving images.
Activated by default.
Image inversion Vertical and horizontal
Automatic kV/mA adjustment Support for automatic generator (high frequency
or conventional) brightness adjustment (ABC).
The function can be disabled by the operator.
IB SYSTEM 18 (Rev. 7)
TECHNICAL MANUAL
Technical characteristics
Description Data
Model Nical NF17 F6 AAB (625 lines 50Hz)
Nical NF17 F2 AAB (625 lines 100Hz)
Nical NF21 F6 AAB (625 lines 50Hz)
Nical NF21 F2 AAB (625 lines 100Hz)
Dimensions 17” (414mm)
21” (511mm)
Scanning 625 lines, 50Hz standard
625 lines, 100Hz optional
Bandwidth 40MHz for 50Hz version
40MHz for 100Hz version
Vertical/horizontal linearity < 1.5%
error
Brightness and contrast Remote controlled table:
adjustment - with keys on the monitor front panel
- controlled through keys on the table console
- automatic adjustment of the brightness,
depending on the ambient light.
Tilting table:
- with keys on the monitor front panel
- automatic adjustment of the brightness,
depending on the ambient light.
Dimensions 17”: 405x359x350mm (LxHxP)
21”: 489x418x380mm (LxHxP)
Weight 17”: 17,5kg (excluding trolley)
21”: 21,8kg (excluding trolley)
Monitor trolley height: adjustable 900-1300 mm
adjustable monitor angle
transport wheels with parking brake
(Rev. 7) 19 IB SYSTEM
TECHNICAL MANUAL
Technical characteristics
Description Data
Model Nical NF17 F2 BAA
Nical NF21 F2 BAA
Dimensions 17” (414mm)
21” (511mm)
Scanning Progressive
Bandwidth 40MHz
Vertical/horizontal linearity < 1.5%
error
Brightness and contrast Remote controlled table:
adjustment - with keys on the monitor front panel
- controlled through keys on the table console
- automatic adjustment of the brightness,
depending on the ambient light.
Tilting table:
- with keys on the monitor front panel
- automatic adjustment of the brightness,
depending on the ambient light.
Dimensions 17”: 405x359x350mm (LxHxP)
21”: 489x418x380mm (LxHxP)
Weight 17”: 17,5kg (excluding trolley)
21”: 21,8kg (excluding trolley)
Monitor trolley height: adjustable 900-1300 mm
adjustable monitor angle
transport wheels with parking brake
IB SYSTEM 20 (Rev. 7)
TECHNICAL MANUAL
Technical characteristics
Description Data
Manufacturer Villa Sistemi Medicali S.p.a.
Via della Azalee, 3
20090 – Buccinasco – MI – Italia
Installation under-table (remote controlled tables)
table-top (tilting tables)
Table compatibility Apollo, Mercury, Symphony, Vision
Generator compatibility Genius HF-A and G100RF-A high rrequency
generators
Image acquisition system Support for interfacing with digital radiographic
compatibility image acquisition system
Camera compatibility Supports interfacing with CCD camera for Digital
Acquisition System
Description Data
Power supply 220/230/240Vac ±10% mono-phase
Frequency 50/60Hz
Current 0.15A
Power 30VA
Isolation class Class I with type B applied parts (IEC 601-1)
Type of installation Permanent installation
(Rev. 7) 21 IB SYSTEM
TECHNICAL MANUAL
Technical characteristics
IB SYSTEM 22 (Rev. 7)
TECHNICAL MANUAL
Technical characteristics
Description Data
Operating temperature From +10°C a +35°C
Storage temperature From –15°C a +45°C
Relative humidity for operation From 10% to 80% RH non-condensing
and storage
Atmospheric pressure for From 630hPa a 1060hPa
operation and storage
Protection class IEC 529 IP X0
(Rev. 7) 23 IB SYSTEM
TECHNICAL MANUAL
Technical characteristics
IB SYSTEM 24 (Rev. 7)
TECHNICAL MANUAL
Assembly
5. ASSEMBLY
Fix the special cover on the component plate using the four M4 screws.
The plate can be positioned on top of the generator cabinet or on the floor
next to it. Alternatively, it can be wall mounted using the two eyebolts.
When positioning the plate, make sure that the ventilation holes are not
obstructed.
In case the unit will be installed with a conventional generator, the plate
will be 2: one for the management of the TV chain and the Monitor, the
second for the command of the kV and automatic mA (Variac module).
(Rev. 7) 25 IB SYSTEM
TECHNICAL MANUAL
Assembly
The I.I. is supplied with the camera already assembled and with the
cables already connected. The orientation of the I.I. is different for
assembly on remote-controlled (under-table assembly) or tilting (table-top
assembly) tables.
IB SYSTEM 26 (Rev. 7)
TECHNICAL MANUAL
Assembly
G
NOTE:
In the case the I.I. is used on tilting tables Viromatic or Vision, the
remote control cable has not to be connected inside the monitor.
12. Position the connecting cables, running the corrugated tube up to the
component board
13. Connect the BNC cable to connector X5 on the TV CCU controller A3
14. Connect the live and neutral conductors of the power supply cable to
terminal block X1 of power supply board A1. Connect the ground
conductor to the main ground bolt
15. Connect the conductors of the remote control cable to terminal block
X4 of table-generator interface board A2
G
NOTE:
When N33 TV chain is present the brightness and contrast remote
control for in room monitor connection must be realised directly to table
general terminal board (X0).
(Rev. 7) 27 IB SYSTEM
TECHNICAL MANUAL
Assembly
G
NOTE:
The jumpers X12, X13, X14, X15, X16, X17, X18 are set in position
1-2 (TH) when a Thomson or Toshiba I.I. is installed in the system.
The jumpers X12, X13, X14, X15, X16, X17, X18 are set in position
2-3 (PH) when a Philips I.I. is installed in the system.
The jumpers X12, X13, X14, X15, X16, X17, X18 are factory adjusted.
The power supply for the IB SYSTEM is generally taken from the
radiological table and must be brought to terminal block X0 as follows:
Terminal Connection
X0.1 Live
X0.2 Neutral
X0.3 Functional ground
The protective ground must be connected from the ground bolt on the
component plate to the main ground node on the remote controlled or
tilting table.
IB SYSTEM 28 (Rev. 7)
TECHNICAL MANUAL
Assembly
Connect the bolt to the ground bar of the radiological table using a
yellow/green conductor with a minimum cross section of 4 mm2.
WARNING:
The equipment is factory set to operate with a mains voltage of 230Vac.
Before suppling the system check the mains voltage. If it is different from
the voltage set, it can be adapted by changing the connection of
transformer T1 on the component plate:
Mains voltage T1 primary T1 secondary
220 Vac 220 26
230 Vac 230 26
240 Vac 230 24
To check the power supply voltage to the I.I. and the CCU, disconnect the
terminals X3 and X6 and measure the voltage without load.
On both X3, pin 4 (positive) and pin 5 (negative), and X6, pin 1 (positive)
and pin 2 (negative), the voltage read must be 24Vdc ± 0.2V for N23 TV
chain and 26Vdc± 0.2V for N33 TV chain.
(Rev. 7) 29 IB SYSTEM
TECHNICAL MANUAL
Assembly
The connections to the A1 board are carried out through the following
terminal blocks:
X1.1 Monitor power supply output: Live Monitor 230Vac 0.5A max
X1.2 Monitor power supply output: Monitor 230Vac 0.5A max
Neutral
X2.1 Auxiliary power supply output: Live Not used 230Vac 0.5A max
Auxiliary power supply output:
X2.2 Not used 230Vac 0.5A max
Neutral
X4.1 Select Zoom 1 from table (see Note) Radiological table Active at 24Vdc or 0Vdc
X4.2 Select Zoom 2 from table (see Note) Radiological table Active at 24Vdc or 0Vdc
X4.3 Select Zoom 3 from table (see Note) Radiological table Active at 24Vdc or 0Vdc
X4.4 Common zoom I.I. (+24Vdc or Radiological table 24Vdc or 0Vdc
0Vdc) (see Note)
X5.1 I.I. field analogic signal for Digital Digital Acquisition Voltages on 1 kOhm
Acquisition System System load:
no zoom: 0Vdc
Zoom 1: 3.6Vdc
Zoom 2: 6.5Vdc
Zoom 3: 11.4Vdc
X5.2 I.I. field analogic signal for Digital Digital Acquisition Common 0Vdc
Acquisition System (0Vdc) System
X6.1 CCU power supply (+24Vdc) Board A2, X1.1 24Vdc max 0.8A
N23
IB SYSTEM 30 (Rev. 7)
TECHNICAL MANUAL
Assembly
G
NOTE:
The functioning principle of the I.I. zoom signals is the following:
If none of ZOOM 1, ZOOM 2, ZOOM 3 signal is active (at +24Vdc or 0Vdc
in according with the Image Intensifier in use), the bigger field is set in
the I.I.. If ZOOM 1 is active, the first zoom is active, and so on (e.g. for a
12" I.I., signal ZOOM 1 selects the 9" field and signal ZOOM 2 selects the
6" field). Signal ZOOM 3 is used only for the 4 field I.I.'s.
The following table shown the possible setting of the outputs on the
connector X4.1, X4.2, X4.3 in according with the settings of the X12,
X13, X14, X15, X16, X17, X18 jumpers on the power supply board A1.
X12 -- ON
X13 -- ON X4.1 = Active at 0Vdc
X14 -- ON
X4.2 = Active at 0Vdc
X15 -- ON
X4.3 = Active at 0Vdc
X16 -- ON
X17 -- ON X4.4 = 0Vdc
X18 -- ON
(Rev. 7) 31 IB SYSTEM
TECHNICAL MANUAL
Assembly
Default
Jumper Pos. Description Note
setting
X7 X16 Recursive filter with OFF
weighting 16
X7 x8 Recursive filter with OFF If none of these
weighting 8 jumpers is closed,
X7 x4 Recursive filter with OFF the filter is enabled
weighting 4 with weighting 2
X7 x0 Recursive filter with OFF
weighting 0
X7 DRF Dynamic Recursive OFF OFF: Dynamic
Filter Recursive filter
ON: Standard
Recursive filter
X7 Agc_off Disable the automatic OFF
DO NOT CLOSE
gain control
X8 All Spare functions OFF NOT ACTIVE
IB SYSTEM 32 (Rev. 7)
TECHNICAL MANUAL
Assembly
The connections of the A2 board are carried out through the following
terminal blocks:
See electrical
X5.1…6 Variac board interface signals Variac board
diagram
See electrical
X6.1…25 CCU interface signals CCU
diagram
G
NOTE:
Connections on connectors X2 and X4 must be made using shielded
cable.
(Rev. 7) 33 IB SYSTEM
TECHNICAL MANUAL
Assembly
Default
Jumper Pos. Description Note
setting
X5 A Analogic V ref. for fluoro kV/mA to ON For Genius and G100
generator with positive signal generators set “Reverse
analog input”
X5 B Analogic V ref. for fluoro kV/mA to OFF
generator with negative signal
IB SYSTEM 34 (Rev. 7)
TECHNICAL MANUAL
Assembly
Default
Dip-sw Pos. Description Note
setting
S1 1 “Neutral Density” filter enable ON Standard configuration:
(Optional) “N.D.” not present or
disable
S1 2 Automatic white level ON Standard configuration:
Auto white level enable
S1 3 Video signal area control for fluoro OFF Standard configuration:
kV auto adjustment (R.O.I.) large area
S1 4 Recursive filter x0 OFF Standard configuration:
recursive filter x2
S1 5 Recursive filter x4 OFF Standard configuration:
recursive filter x2
S1 6 Recursive filter x8 OFF Standard configuration:
recursive filter x2
Standard configuration:
S1 7 Recursive filter x16 OFF
recursive filter x2
Standard configuration:
S1 8 Dynamic recursive filter (D.R.F.) OFF dynamic recursive filter
enable
(Rev. 7) 35 IB SYSTEM
TECHNICAL MANUAL
Assembly
The connections of the A2 board are carried out through the following
terminal blocks:
X2.1 OUT V ref. for fluoro auto kV Generator (with 0-12Vdc. Set-point at
with positive or negative signal shielded cable) 6.5Vdc
depending by jumper X5
position (see paragraph 6.3)
X2.2 Common 0Vdc Generator (with
shielded cable)
X2.3 Fluoroscopy call OUT Generator Active with continuous
and pulsed fluoro
X2.4 X-ray ON IN Generator Active with pulsed
fluoro
X2.5 Fluoroscopy request IN Radiological table Active with continuous
and pulsed fluoro
X2.6 Clockwise camera rotation IN Optional
X2.7 Anti-clockwise camera rotation Optional
IN
X2.8 Vertical inversion IN Radiological table
X2.9 Horizontal inversion IN Radiological table
X2.10 D.R.F. enable IN A2 board – X4.7 D.R.F. function must be
disabled in pulsed
fluoro
X2.11 Recursive filter x0 IN Radiological table Remotable function
present on Apollo table
only
X2.12 Recursive filter x4 IN A2 board – X4.7 x4 recursive filter must
be selected with pulsed
fluoro because D.R.F. is
disabled
X2.13 Recursive filter x8 IN Radiological table Remotable function
present on Apollo table
only
X2.14 Recursive filter x16 IN Radiological table Remotable function
present on Apollo table
only
X2.15 Video signal area control Radiological table Remotable function
selection for fluoro kV auto present on Apollo table
adjustment (R.O.I.) IN only
IB SYSTEM 36 (Rev. 7)
TECHNICAL MANUAL
Assembly
X6.1 A.D.D. 1
X6.2 A.D.D. 0
X6.3 A.D.D. 2
X6.4 A.D.D. 4 Present only on N33
X6.5 M.V.R. CCU with digital
X6.6 A.D.D. 3 memory
X6.7 A.D.D. 5
X6.8 Not used
X6.9 0V
(Rev. 7) 37 IB SYSTEM
TECHNICAL MANUAL
Connection of accessories
7. CONNECTION OF ACCESSORIES
The IB SYSTEM is compatible with the Genius HF, Genius HF-A, G100RF
and G100RF-A high frequency generator family.
The pulsed fluoro modality, available on TV chain N33, is allowed with
generator Genius HF-A and G100RF-A.
The connection with these generators is shown in the following tables:
A2 board GENIUS
Description
for N23 TV chain generator
X2.1 Analogic driving ABC GTB 402
X2.2 O Vdc GTB 401
X2.3 Fluoro call GTB 429
X2.4 kV/mA automatic or manual selection GTB 454
A2 board GENIUS
Description
for N33 TV chain generator
X2.1 Analogic driving ABC GTB 402
X2.2 O Vdc GTB 401
X2.3 Fluoro call GTB 429
X2.4 X-ray ON in pulsed fluoro GTB 330
IB SYSTEM 38 (Rev. 7)
TECHNICAL MANUAL
Connection of accessories
(Rev. 7) 39 IB SYSTEM
TECHNICAL MANUAL
Connection of accessories
230Vac (refer to
Power supply kV and
X8.4…X8.12 Generator the generator
mA
schematics)
G
NOTE:
For safety reasons, the factory setting of transformer T2 is 380V. This will
avoid possible damages to the components in case of wrong setting.
IB SYSTEM 40 (Rev. 7)
TECHNICAL MANUAL
Connection of accessories
(Rev. 7) 41 IB SYSTEM
TECHNICAL MANUAL
Connection of accessories
(*) Depending by selection of cell 246, X0.54 output can drive R.O.I. or recursive
filter. In the first case X0.55 and X0.56 are disabled, in the second case are
enabled.
IB SYSTEM 42 (Rev. 7)
TECHNICAL MANUAL
Connection of accessories
(Rev. 7) 43 IB SYSTEM
TECHNICAL MANUAL
Connection of accessories
IB SYSTEM 44 (Rev. 7)
TECHNICAL MANUAL
Connection of accessories
(Rev. 7) 45 IB SYSTEM
TECHNICAL MANUAL
Connection of accessories
7.2.5 Assembling and wiring the keypad for over-table serial changers
(SAG 4 - SAG 6) on VIROMATIC
1. Fix the keypad on the head side of the serial changer using the two
M4 screws, as shown in the drawing below:
Over-table
serial changer
Patient
head side
Keypad
2. Run the corrugated tube along the side of the serial changer and
down to the equipment base
3. Fix the male connectors PRT20 and PRT21 to the bar at the foot of
the table
4. The cable departing from connector PRT 20 must be brought to the
component board, making the following connections (see also the
annexed schematic):
• connector X4 (4 pin connector) on circuit board A1 (power supply).
I.I. field select
• connector X7 (2 pin connector) on circuit board A1 (power supply).
24Vdc signal lamps power supply
• terminal block X3 (terminals 1, 2, 4, 5) on table-generator interface
board A2. Inversions and man/auto kV selection
IB SYSTEM 46 (Rev. 7)
TECHNICAL MANUAL
Connection of accessories
For the connection between the trimmer and the generator, refer to
the manual of the generator.
(Rev. 7) 47 IB SYSTEM
TECHNICAL MANUAL
Connection of accessories
Connector on Terminal on
Description
board A1 POD board
X5.1 V ref. for I.I. field 30
X5.2 0 Vdc 31
IB SYSTEM 48 (Rev. 7)
TECHNICAL MANUAL
Functional checks
8. FUNCTIONAL CHECKS
WARNING:
THE FOLLOWING OPERATIONS REQUIRE EMISSION OF AN X-RAY
BEAM. ALTHOUGH THE I.I. ENVELOPE IS SUFFICIENTLY
SCREENED WITH LEAD, THE SERVICE TECHNICIAN MUST
NONETHELESS TAKE EVERY PRECAUTION TO MINIMISE HIS
EXPOSURE TO X RAYS.
The controls for operating the IB SYSTEM are not situated on the IB
SYSTEM itself, but rather in different positions depending on the type of
table. The controls are situated on the control console in the case of
remote controlled tables, and on the serial changer in the case of tilting
tables.
Further details on the positions of the controls can therefore be found in
the documentation supplied with the radiological table.
All the controls are identified by means of standard symbols, in
accordance with the IEC regulations. The following table summarises the
function and position of each control, together with the corresponding
symbol.
(Rev. 7) 49 IB SYSTEM
TECHNICAL MANUAL
Functional checks
IB SYSTEM 50 (Rev. 7)
TECHNICAL MANUAL
Functional checks
G
NOTE:
For TV chain N33 only, is possible to get a test pattern to check the
image quality on the monitor. To do that, set dip-switches S1.4 and S1.7
(A2 board) ON.
(Rev. 7) 51 IB SYSTEM
TECHNICAL MANUAL
Functional checks
TV chains N23 and N33 DRF comes with Dynamic Recursive Filter. This
feature, activated by default, can be disabled acting on board A2 (table-
generator interface).
It is advised to let this feature enabled because it enhances image quality
both statically and during motion.
Before checking image quality make sure that this feature is enabled
and, if necessary increase the filtering action as described at paragraph
6.2 and paragraph 6.3.
G
NOTE:
Filter weight (X0, X4, X8, X16) is more influent when standard recursive
filter is activated; for Dynamic Recursive Filter the default setting is
generally sufficient and acting on X0, X4, X8 or X16 does not affect
significantly the filter action.
The two jumpers X10 and X11 on power supply board A1 connect the
zero volt nodes of the I.I. and CCU power supplies, respectively, to
ground.
IB SYSTEM 52 (Rev. 7)
TECHNICAL MANUAL
Functional checks
(Rev. 7) 53 IB SYSTEM
TECHNICAL MANUAL
Functional checks
1. Position the Leeds Test Object “GS2” phantom at the centre of the
I.I.. The phantom must be aligned with the reference axes
2. Insert a 1.5mm thick copper plate
3. Perform X-ray exposure with automatic regulation of radiological
parameters.
4. For optimal calibration of the system, after adjusting the monitor
brightness and contrast the following conditions must be met:
• white circle visible inside the upper square
• black circle visible inside the lower square
• grey scale completely visible (10 levels of grey).
G
NOTE:
Typically I.I. systems are factory calibrated to deliver a dose in front of
the I.I. entrance surface of 45µR/s.
IB SYSTEM 54 (Rev. 7)
TECHNICAL MANUAL
Functional checks
12
11 1
10 2
18 13
9 17 19 14 3
16 15
8 4
7 5
6
TO.N3
Figure 8-2: Leeds Test Object N3 phantom
1. Position Leeds Test Object “N3” phantom at the centre of the I.I.. The
phantom must be aligned with the orthogonal axes and oriented so
that the reference notch is at the top right
2. Insert a 1.5mm thick copper plate
3. Perform X-ray exposure with automatic regulation of radiological
parameters.
4. Adjust the monitor contrast and brightness to display the greatest
possible level of detail.
In optimal conditions, at least 12 details must be visible.
G
NOTE:
Typically I.I. systems are factory calibrated to deliver a dose in front of
the I.I. entrance surface of 45µR/s.
(Rev. 7) 55 IB SYSTEM
TECHNICAL MANUAL
Functional checks
1. Position the Leeds Test Object “TO.10” phantom at the centre of the
I.I.. The phantom must be aligned with the reference axes and
oriented so that the reference notch is on the upper right.
6 5 4
7 3
8 2
9 1
D
9 8 7 6 5 4 3 2 1 E
9 G F
H
1
J
A
C B
9 8 7 6 5 4 3 2 1
K
8 L 2
M
3
7
6 4
5 TO.10
Figure 8-3: Leeds Test Object TO.10 phantom
Detail number
Group
1 2 3 4 5 6 7 8 9
A, B, C 0.16 0.123 0.086 0.066 0.045 0.032 0.023 0.016 0.012
D, E, F 0.23 0.16 0.123 0.086 0.066 0.045 0.032 0.023 0.016
G, H, J 0.50 0.35 0.23 0.16 0.123 0.086 0.066 0.045 0.032
K, L, M 0.93 0.66 0.50 0.35 0.23 0.16 0.123 0.086 0.066
IB SYSTEM 56 (Rev. 7)
TECHNICAL MANUAL
Functional checks
Group A B C D E F G H J K L M
Minimum number of object 7 7 6 6 5 5 5 5 4 3 2 1
G
NOTE:
Typically I.I. systems are factory calibrated to deliver a dose in front of
the I.I. entrance surface of 45µR/s.
(Rev. 7) 57 IB SYSTEM
TECHNICAL MANUAL
Functional checks
9. ADJUSTMENT
G
NOTE:
The adjustments described below are factory settings made during
the equipment testing phase, and do NOT normally need to be
changed at the time of installation.
The following operations should be performed in the field only when
absolutely necessary, for example in the event of malfunction,
replacement of components, or maintenance.
WARNING:
CERTAIN MAINTENANCE OPERATIONS REQUIRE THE EMISSION OF
AN X-RAY BEAM. ALTHOUGH THE I.I. ENCLOSURE IS
SUFFICIENTLY SCREENED WITH LEAD, THE SERVICE TECHNICIAN
MUST IN ANY CASE TAKE EVERY PRECAUTION TO MINIMISE HIS
EXPOSURE TO RADIATION.
G
NOTE:
The “a” distance is in function of I.I. model used and it is adjusted in
factory during final test.
If the camera must be replaced, recorder the “a” distance and apply on
the new one.
If spacer are present between camera and I.I. re-use that for the new
camera.
WARNING:
Removing the camera exposes the I.I. output window and the camera
lens. Be careful not to touch the output window or the lens with your
fingers or with any type of object.
IB SYSTEM 58 (Rev. 7)
TECHNICAL MANUAL
Functional checks
1. Carefully clean the entire contact area between the camera flange
and the I.I. output surface
2. Carefully clean the camera fixing flange
3. Carefully and very delicately clean the I.I. output window
4. Remove the protective cap from the camera lens and carefully and
very delicately clean it
5. Assemble the camera by inserting the three M4 screws, without
locking them down. Place a metal washer and a nylon washer
between the screw heads and the camera flange, as shown in the
figure:
Camera
body
Metal washer
Screw head
Nylon washer
II body
(Rev. 7) 59 IB SYSTEM
TECHNICAL MANUAL
Functional checks
G
NOTE:
Before making the following adjustments, make sure that the camera
electronic circle is disabled:
• N23 TV chain: Jumper J2 on circuit board N23_ANB1 OFF, pins 1-2
of CCU - see paragraph 9.2
• N33 TV chain: set dip-switches S1.4 and S1.5 ON.
1. Using a felt tip pen, mark the two orthogonal axes with a cross at the
centre of the monitor screen. Also mark the intersection of these axes
with a circle 255mm in diameter
E
D
C
IB SYSTEM 60 (Rev. 7)
TECHNICAL MANUAL
Functional checks
1. If the camera must be replaced, fix the new one in such a way the
cable output will be in the same position of the old one.
2. Position a phantom with centring grid, for example Wellhofer model
ETR 1, at the centre of the radiological table
3. Disable the Vertical Inversion, Horizontal Inversion and zoom
commands
4. Slightly back off the locking ring “C” (Figure 9-2)
5. Perform X-ray exposure. Nominal parameters: 45kV, 1.0mA
6. Adjust the monitor brightness and contrast so that the phantom is
clearly visible
7. Looking at the monitor, rotate the camera so that the image is
aligned with the reference cross marked on the monitor
8. Stop the X-ray exposure
9. Lock down the ring “C” firmly, but without using any tools
10. Lightly back off the three camera fixing screws
11. Perform X-ray exposure. Parameters: 45kV, 1.0mA
12. Looking at the monitor, move the camera so that the image is
centered on the reference cross marked on the monitor.
Pay attention to the parallax errors during these adjustments
13. Stop the X-ray exposure
14. Lock down the three screws.
(Rev. 7) 61 IB SYSTEM
TECHNICAL MANUAL
Functional checks
FUOCO FUOCO
DIAFRAMMA DIAFRAMMA
FOCUS FOCUS
DIAPHRAGM “A” DIAPHRAGM “A”
“B” “B”
IB SYSTEM 62 (Rev. 7)
TECHNICAL MANUAL
Functional checks
5. In order to stabilise the kV for different dose rate values, adjust the
kV, mA and filtration values accordingly.
Note that higher dose rates produce more detailed images, but expose
the patient to higher doses of radiation. It is recommended to not
exceed 60µR/s.
The video signal output by the camera must have a peak value of 0.6V
(measured on the camera output with an oscilloscope). See the trace in
Figure 9-4 below.
0,1 V/div
10 µs/div
Figure 9-4: Waveform at the camera output for adjusting the aperture
G
NOTE:
It is possible to modify the kV automatic regulation window.
The factory calibration is set to cause the automatic correction for values
± 2 kV. To change this setting act on trimmer P4 for N23 camera or
trimmer R106 for N33 camera. Please be aware that changing this setting
moves the balancing regulation, so it will be necessary to adjust also the
setting of trimmer P11 for N23 camera or R79 for N33 camera.
Changing this setting can have adverse effects like:
• image instability due to high gain feedback on kV (if the window is
too narrow)
• different brightness of the same image after automatic regulation (if
the window is too wide).
(Rev. 7) 63 IB SYSTEM
TECHNICAL MANUAL
Functional checks
FUOCO FUOCO
DIAFRAMMA DIAFRAMMA
FOCUS FOCUS
DIAPHRAGM “A” DIAPHRAGM “A”
“B” “B”
1. Place at the centre of the I.I. the “MS 1” phantom of the Leeds
University “SFS Standard Fluoroscopy Set”
2. Select the maximum I.I. zoom
3. Made an X-ray exposure regulating kV and mA in such a way to
optimize spatial resolution (i.e. focus of the system)
4. Adjust the monitor contrast to its maximal value
5. Adjust the monitor brightness so that the image of the monitor is
only just visible
6. Turn the focus adjuster screw “A” on the camera until the image on
the monitor is as bright as possible.
At the end of the procedure, halt the X-ray exposure.
IB SYSTEM 64 (Rev. 7)
TECHNICAL MANUAL
Functional checks
(Rev. 7) 65 IB SYSTEM
TECHNICAL MANUAL
Functional checks
kV- kV+
ROSSO VERDE
RED GREEN
P7 P10 P9 P8 P6 P5 SW3
1 1
R163 R156
J4
P11
SW4
1
1 SW2
J2 J6
1 1 R157
P4
P3 P2 P1
R94
J5
1
N23_ANB1
X5 X5 X1 X5
X6 X2
G
Trimmer R157 Black level NOTE:
Trimmer R163 White level Figure 9-6 also shows the default
Trimmer R156 Automatic black level control
jumper settings.
Trimmer R94 Grey scale correction
IB SYSTEM 66 (Rev. 4)
TECHNICAL MANUAL
Functional checks
kV- kV+
ROSSO VERDE
RED GREEN
J6 J5
1
R61 J7 1
R48 R47 R52
J8 1 1 1
NDR_ADV0
N30_PWR20
V4 X6 (V3)
X4 X5 X2
X1 X3
G
automatic kV control
Trimmer R61 Vertical ramp uniformity NOTE:
Trimmer R52 Vertical parabola uniformity Figure 9-7 also shows the default
Trimmer R47 Horizontal ramp uniformity jumper settings.
Trimmer R48 Horizontal parabola
uniformity
(Rev. 7) 67 IB SYSTEM
TECHNICAL MANUAL
Functional checks
G
NOTE:
Trimmers and jumpers reference indicated as ( ) are relevant to N33 TV
chain.
0,1 V/div
5 µs/div and 2ms/div
IB SYSTEM 68 (Rev. 7)
TECHNICAL MANUAL
Functional checks
0,1 V/div
5 µs/div and 2ms/div
0,1 V/div
5 µs/div and 2ms/div
(Rev. 7) 69 IB SYSTEM
TECHNICAL MANUAL
Functional checks
G
NOTE:
Circle dimension is obtained in digital way for N33 TV chain.
Consequently no adjustment is required.
G
NOTE:
To help make these adjustments, use the reference marks on the
monitor.
Pay attention to the parallax errors during these adjustments.
IB SYSTEM 70 (Rev. 7)
TECHNICAL MANUAL
Functional checks
G
NOTE:
R.O.I. dimensions are obtained in digital way for N33 TV chain.
Dimensions selection is described on paragraph 6.3.
To display the R.O.I. dimensions selected, set dip-switches S1.4 and S1.6
(A2 board) ON.
(Rev. 7) 71 IB SYSTEM
TECHNICAL MANUAL
Functional checks
WARNING:
This is a factory setting. Only change this adjustment when absolutely
necessary.
G
NOTE:
Trimmers and jumpers reference indicated as ( ) are relevant to N33 TV
chain.
10 mV/div
5 µs/div
IB SYSTEM 72 (Rev. 7)
TECHNICAL MANUAL
Functional checks
WARNING:
This is a factory setting. Only change this adjustment when absolutely
necessary.
G
NOTE:
Trimmers and jumpers reference indicated as ( ) are relevant to N33 TV
chain.
0,1 V/div
5 µs/div
(Rev. 7) 73 IB SYSTEM
TECHNICAL MANUAL
Functional checks
7. Adjust R163 (R84) so that the level of the waveform is the same as it
was with the jumper open
8. Visually check that the two waveforms coincide by opening and
closing the jumper
9. Replace jumpers SW2 (J6) and SW3 (J5)
10. Turn R156 (R100) fully counter-clockwise
11. X-ray exposure. Parameters: 80kV, 1.2mA
12. Turn R156 (R100) in a clockwise direction until the image stabilises
without flickering
13. If the waveform is similar to that shown in Figure 9-14, continue
turning R156 (R100) in a clockwise direction until it appears as
shown in Figure 9-13.
0,1 V/div
5 µs/div
IB SYSTEM 74 (Rev. 7)
TECHNICAL MANUAL
Functional checks
1. Place a copper disk and a lead disk of thickness 0.5 mm side by side
at the centre of the I.I.
2. X-ray exposure. Parameters: 70kV, 1.2mA
3. On the monitor, check that the lead disk is shown completely black
and that the copper disk is grey. If this is not the case, adjust the
monitor brightness and contrast controls, and if necessary correct
the kV and/or mA values of the generator
4. Halt the X-ray exposure and remove the lead disk
5. Repeat the exposure with the values set previously
6. Check that the copper disk now appears completely black.
If this is not the case, make the adjustments described in paragraph
9.2.4 and 9.2.5.
(Rev. 7) 75 IB SYSTEM
TECHNICAL MANUAL
Magnetic field correction coil power supply
G
NOTE:
The adjustment must be made with the radiological table in the most
commonly used position, because the ground’s magnetic field causes the
distortion to change considerably as a function of the tilting angle of the
table.
IB SYSTEM 76 (Rev. 7)
TECHNICAL MANUAL
Adjusting the HV power supply
WARNING:
This is a factory setting. Only change this adjustment when replacing the
HV power supply on an existing I.I.
The adjustments must be repeated for the three fields N, M1, M2.
The pulse switch to adjust, the adjustment ranges and the test points are
the following:
Pulse Test
Description Minimum Maximum Ratio
switch point
G1 Voltage G1 (Vdc) 35 350 100 V/V TP4
G2 Voltage G2 (Vdc) 100 2000 1000 V/V TP3
G3 Voltage G3 (kVdc) 2 20 10 kV/V TP2
(Rev. 7) 77 IB SYSTEM
TECHNICAL MANUAL
Adjusting the HV power supply
The storage procedure of the tuned values for any voltage is:
1. Once voltage is correctly tuned, turn ON dip switch 1 of the
configuration switch.
2. After a few seconds, turn OFF dip switch 1 of the configuration
switch.
The pulse switch, test points and configuration switch are illustrated in
Figure 11-1.
IB SYSTEM 78 (Rev. 7)
TECHNICAL MANUAL
Adjusting the HV power supply
(Rev. 7) 79 IB SYSTEM
TECHNICAL MANUAL
Adjusting the HV power supply
G
NOTE:
G4 and G3 voltages may not be read by a common DVM because the
voltage values are very high (more than 1000V). If only a common DVM is
available do not attempt to calibrate these voltages: just leave the preset
values which are usually OK.
The trimmer to adjust, and the adjustment ranges are the following:
IB SYSTEM 80 (Rev. 7)
TECHNICAL MANUAL
Adjusting the HV power supply
The trimmers and the output connectors are illustrated in Figure 11-3.
(Rev. 7) 81 IB SYSTEM
TECHNICAL MANUAL
Maintenance and cleaning
WARNING:
CERTAIN MAINTENANCE OPERATIONS REQUIRE THE
EMISSION OF AN X-RAY BEAM. ALTHOUGH THE I.I. ENVELOPE
IS SUFFICIENTLY SHIELDED WITH LEAD, THE SERVICE
TECHNICIAN MUST IN ANY CASE TAKE EVERY PRECAUTION
TO MINIMISE HIS EXPOSURE TO RADIATION.
(*) Due to the normal drift of the I.I. conversion factor, as time goes by there is a
gradual increase in the kV set-point value. It is possible to compensate (within
limits) for this drift by adjusting the aperture.
IB SYSTEM 82 (Rev. 7)
TECHNICAL MANUAL
Maintenance and cleaning
Power supply
Complete I.I.
Installation
Magn. field
correction
HV Power
Interface
Camera
supply
board
board
board
CCU
Paragraph and description of
the adjustment to be made
Notes:
1) only for 16” I.I. for Digital Acquisition System
2) these are factory settings and do not normally need to be changed. It is
(Rev. 7) 83 IB SYSTEM
TECHNICAL MANUAL
Maintenance and cleaning
12.3 Cleaning
WARNING:
The cleaning operations described below must be performed exclusively
by technical personnel authorised by Villa Sistemi Medicali.
WARNING:
• before proceeding to clean the equipment, it is essential to disconnect
it from the power supply
• take the utmost care to avoid spilling water or other liquids into the
equipment which may cause corrosion or short circuits
• Clean the painted surfaces using only a moist cloth and neutral
detergent, taking care to wipe them afterward with a dry cloth.
Do not use solvents (alcohol, petrol, trichloroethylene) or corrosive or
abrasive products.
The equipment does not have any parts which come into contact with the
patient or the operator, therefore routine cleaning by the operator is not
necessary.
During examinations with contrast medium, the liquid may leak onto the
entrance surface of the I.I.. This condition is readily detected because
very evident spotting appears on the image.
If this occurs, clean the entrance window with a moist cloth and neutral
detergent.
The operation must be carried out with great care because the entrance
window is coated with a very thin (0.5 mm) aluminium sheet.
If the camera or the I.I. are replaced, it is necessary to clean the I.I.
output window and/or the camera optics.
IB SYSTEM 84 (Rev. 7)
TECHNICAL MANUAL
Schematics and drawings
1. Component plate layout for standard I.I. (CCU 625 lines / 50 Hz)
2. Component plate layout for standard I.I. (CCU 625 lines / 100 Hz)
11. Component plate layout for HR I.I. (for Digital Acquisition System)
13. Magnetic field correction board layout (only for 16" I.I.)
14. Magnetic field correction board schematic (only for 16" I.I.).
(Rev. 7) 85 IB SYSTEM
TECHNICAL MANUAL
Schematics and drawings
IB SYSTEM 86 (Rev. 7)
TECHNICAL MANUAL
Schematics and drawings
(Rev. 7) 87 IB SYSTEM
TECHNICAL MANUAL
Schematics and drawings
(Rev. 7) 89 IB SYSTEM
TECHNICAL MANUAL
Schematics and drawings
(Rev. 7) 91 IB SYSTEM
TECHNICAL MANUAL
Schematics and drawings
(Rev. 7) 93 IB SYSTEM
TECHNICAL MANUAL
Schematics and drawings
(Rev. 7) 95 IB SYSTEM
TECHNICAL MANUAL
Schematics and drawings
A1 – Power supply
board layout 6
Code 58161120 - Rev. 1 Page 1 of 1
(Rev. 7) 97 IB SYSTEM
TECHNICAL MANUAL
Schematics and drawings
(Rev. 7) 99 IB SYSTEM
TECHNICAL MANUAL
Schematics and drawings
A2 – Table-generator
interface board layout
for CCU N23
8
Code 58161010 – Rev. 1 Page 1 of 1
A2 – Table-generator
interface board schematic
for CCU N23
9
Code 39169010 – Rev. 1 Page 1 of 1
A2 – Table-generator
interface board layout
for CCU N33
10
Code 58161122 – Rev. 0 Page 1 of 1
A2 – Table-generator
interface board schematic
for CCU N33
11
Code 39169030 – Rev. 1 Page 1 of 2
A2 – Table-generator
interface board schematic
for CCU N33
11
Code 39169030 – Rev. 1 Page 2 of 2
VARIAC module
assy layout 12
Code 72161040 – Rev. 1 Page 1 of 1
VARIAC module
assy schematic 13
Code 39169016 – Rev. 0 Page 1 of 1
4 – Variac module
5 – Options
4 – Variac module
5 – Options
5 2374862900 Lamp
6 4192005400 Potentiometer
G
NOTE:
To order Spare Part for monitors, please provide information about type
of board required, model and serial number of the specific monitor.