BRONCHOSCOPE DISINFECTION GUIDE

for OES BF-40 Series Bronchofiberscopes


®

Refer to the companion BRONCHOSCOPE CLEANING GUIDE for detailed cleaning information

WARNING: This guide is only a summary of the steps necessary to properly disinfect your bronchoscope. Be sure to follow the
detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that was included with your bronchoscope when purchased.

HIGH-LEVEL DISINFECTION
1 2 3 4
Suction Cleaning Adapter

Fill basin with disinfectant. Attach a Attach a syringe to the adapter. While immersed, detach the adapter Soak all equipment according
suction cleaning adapter (MAJ-222) to Pull the plunger of the syringe to from the bronchoscope. Use a clean, to the time and temperature
the previously cleaned bronchoscope. fill the instrument/suction channels lint-free cloth to wipe away any bub- recommended by the disinfectant
Immerse completely in disinfectant and channel cleaning adapter bles clinging to external surfaces. manufacturer. Use a timer to mea-
solution. with disinfectant solution. sure the exact contact time.

RINSING
5 6 7 To Suction Pump 8 To Suction Pump

Alcohol

Connect the suction cleaning adapter Disconnect the suction cleaning Connect suction cleaning adapter
and syringe. Inject air to remove adapter and immerse the broncho- to bronchoscope and to a tube Immerse the bronchoscope's distal
disinfectant from the instrument/suction scope and adapter in sterile water. from a suction pump. Aspirate end in 70% alcohol and aspirate
channels, then remove the broncho- Using a sterile, lint-free cloth, thor- water through channels for 30 alcohol for 5 seconds. Remove the
scope from the disinfectant. oughly rinse and wipe all surfaces. seconds. Remove from water and distal end from the alcohol and
aspirate air through channels for aspirate air for 20 seconds.
60 seconds.

DRYING TRANSPORTATION
9 10

Turn the suction pump off. Hold the control Dry the bronchoscope and adapter with an
section with the openings pointing down and alcohol-moistened cloth. Make sure that the Hold all parts of the endoscope when carrying. Do
disconnect the suction cleaning adapter. equipment is completely dry prior to storage. not let the ends of the bronchoscope swing freely.

NOMENCLATURE IMPORTANT
© 2004 Olympus America Inc. / Printed in USA 01/04 — PWC-100020

• Wear appropriate personal protective


equipment during reprocessing.
Single-Use Suction Valve (MAJ-209)
Universal Cord Light Guide • Autoclave all autoclavable accessories.
Suction Cylinder
• Store the bronchoscope hanging vertically
with all valves and caps removed.
Single-Use Biopsy Valve (MAJ-210) • Use a compatible disinfectant and test it
for proper concentration as recommended
Instrument/Suction Channel Port Venting Connector
by the manufacturer.
• Do not use the carrying case to store
"patient-ready" endoscopes. Any
bronchoscope placed in the carrying case
Control Section Insertion Tube
Boot must be reprocessed prior to patient use.
Bending Section
UP/DOWN Angulation Control Lever Distal Tip
Clean the bronchoscope
meticulously prior to disinfection.