You are on page 1of 48

Dental Chair User Manual

Model : FRONTIER

Thank you for purchasing Kuang Yeu Medical Co., Ltd.


dental unit, before using the machine, Please read this
manual so that you properly install and use the machine.
TABLE OF CONTENTS
GENERAL ORDERING

1. Introduction……………………………………………………………………………………………………. 1
2. Security Information………………………………………………………………………………………….. 1
3. Warranty and General……………………………………………………………………………………….. 2
3.1 General……………………………………………………………………………………………………. 2
3.2 Warranty…………………………………………………………………………………………………… 2
3.3 Certifictae Of Compliance……………………………………………………………………………….. 3
4. Caution and Safety Tips……………………………………………………………………………………… 3
4.1 General warnings and safety tips……………………………………………………………………….. 3
4.2 Use of the logo instructions……………………………………………………………………………… 4
4.3 Product Label……………………………………………………………………………………………… 5
5. Considerations………………………………………………………………………………………………… 6
6. Transport and storage………………………………………………………………………………………... 6
6.1 Transport conditions……………………………………………………………………………………… 6
6.2 Transport and storage conditions………………………………………………………………………. 6
6.3 Transport laminate requirements……………………………………………………………………….. 6
6.4 Description labeled with warnings………………………………………………………………………. 7
7. Product Introduction………………………………………………………………………………………….. 8
7.1 Functions and intended use……………………………………………………………………………... 8
8. Conditions……………………………………………………………………………………………………… 9
9. Technical Data………………………………………………………………………………………………… 9
9.1 Patient chair………………………………………………………………………………………………. 9
9.2 Shadowless Lamp………………………………………………………………………………………… 9
9.3 Workspace………………………………………………………………………………………………… 10
10. Product Structure……………………………………………………………………………………………. 12
10.1 Structure Representation………………………………………………………………………………. 12
10.2 Dental Unit description of each unit…………………………………………………………………… 13
11. Product Installation………………………………………………………………………………………….. 18
11.1 Installation space………………………………………………………………………………………... 18
11.2 Installation Handling…………………………………………………………………………………….. 18
11.3 Treatment Basics piping installation instructions…………………………………………………….. 19
12. Instructions for use and precautions………………………………………………………………………. 20
12.1 Dental chair electric power up………………………………………………………………………….. 20
12.2 Instrument tray Instructions…………………………………………………………………………….. 22
12.3 Three-way syringe, Saliva ejectors instructions for use……………………………………………... 25
12.4 Shadowless lamp operating instructions……………………………………………………………… 27
12.5 The passenger's seat function operating instructions……………………………………………….. 30
12.6 Foot controller functions Operating Instructions……………………………………………………… 31
13. Cleaning Care and maintenance instructions…………………………………………………………….. 31
13.1 Dental chair Exterior Cleaning and maintenance instructions……………………………………… 31
13.2 Shadowless lamp, Ceramic dish cleaning and maintenance instructions…………………………. 31
13.3 Filter tank body cleaning and maintenance instructions…………………………………………….. 32
13.4 Three-way syringe, Handpice, Saliva suction cleaning and maintenance………………………… 32
14. Troubleshooting instructions……………………………………………………………………………….. 33
14.1 Shadowless lamp……………………………………………………………………………………….. 33
14.2 Treatment table………………………………………………………………………………………….. 34
14.3 Patient chair……………………………………………………………………………………………… 35
15. Diagram of the air and water pipes………………………………………………………………………… 36
16. Patient chair, Dentist element – Electric Diagram………………………………………………………... 38
17. Water Unit With Swivel Spittoon Bowl – Electric Diagram………………………………………………. 39
18. Foot Control – Electric Diagram……………………………………………………………………………. 40
19. Tubing Identification………………………………………………………………………………………….. 41
1. Introduction
Modern science and technology rapid technological change, technological change
technology successful commercial products, these new products has changed human
living conditions, human beings enjoy the convenience of a comfortable quality of life it
offers. The same treatment station in dental applications, Part multiple functions will be
the trend of the new century, the relationship between the machine and allows the
patient's close, to achieve a comprehensive user-friendly. Its design and the
application complies with the individual's needs. Today's doctors want to provide the
best treatment to the patient, a good environment is not only the need for treatment
and necessary, FRONTIER, it is taking into account the patient comfort and operator
convenience, the entire system operation is very coordinated, he streamlined the style
is very detailed, you can make people happy, and create a good clinic environment.

2. Security Information
Please read the instructions in order to properly use this product. If you have any
special requirements please contact our dealers contact.

Warnings

■ In order to reduce the possibility of fire and electric shock, please do not install the
unit in a humid place, please do not open the box. If you need to repair the machine,
please find professional maintenance staff.
■ If the device has smoke or other abnormal conditions may result in the case of some,
such as the appearance of fire, disconnect the power immediately and contact your
local dealer or service department.
■ If the device is water, it may cause some cases, such as leakage occurs, disconnect
the power immediately and contact your local dealer or service department.
■ If the device fails any unknown situation, disconnect the power immediately and
contact your local dealer.
■ The power cord can not be damaged if the power cord is damaged, turn off the
power immediately, do not directly touch their power. Please contact your local
dealer for replacement of damaged wires.
■ This product is suitable for 220 VAC / 50Hz power at work, if you work at other
voltage may be dangerous.
■ Please pay attention to weather changes, when a lightning storm, disconnect the
power supply.
■ When wet hands, do not touch the power plug, it may cause electric shock.
■ Please do not use pesticides, alcohol, thinner and chemical liquids to clean the
equipment, which will cause the device to surface flaking or rot.

-1-
3. Warranty and General
3.1 General
Qualified personnel of the dental unit product shall be designated in
accordance with the relevant existing industrial safety and read books, use the
functional specification, for use in health and accident prevention regulations and
the job description of this application. According to the above requires the operator
to use only the performance and functionality perfectly trouble-free products. You
must immediately stop using the operation when a failure occurs while the
phenomenon of contact Kuang Yeu Medical Co., Ltd. authorized distribution
agents.
Must ensure that the purpose and design of this product when used, and shall
ensure that its own patient with other people's safety, use of the process should
avoid contamination. The dental unit product, can only be used for dental auxiliary
medical purposes, and forbids the product for other purposes, it may be dangerous.
This product has been in line with current medical electrical safety standard
requirements.
3.2 Warranty
Kuang Yeu Medical Co., Ltd. were awarded operation covering all functional
failure, use of the product or raw material for production of improper defects
warranty. The dental unit will begin from the date of purchase warrant for 12 months
in proving under the regular operating range failure, Kuang Yeu Medical Co., Ltd.
equipment or its authorized sales agents, and so will the warranty obligations of the
Company for the product for a free repair. (In addition to the consumption of goods
and materials), improper damage of any other reason, such as natural disasters,
man-made or force majeure, in particular due to the deliberate destruction or
improper intent interests factors, are no longer covered by the warranty.
Kuang Yeu Medical Co., Ltd. will not be caused by the following factors
responsible for injury or damage to the product:
(1) Improper operation caused by the damage to the product.
(2) Excessive use or exceeding the statutory durable life.
(3) Non-compliance with the instructions for installation of this product,
operation and maintenance.
(4) That is not normal or the battery electrolyte chemical influence of an
unknown liquid.
(5) Water or gas pressure caused by poor power supply terminal is connected
to use it.
If the follow-up impact of the failure of the product which was prompted
not Kuang Yeu Medical Co., Ltd. recognized authorized distribution
agents or personnel, incorrect operation of the product made, repair or
adjust the warranty will be invalidated.

This product is in line with the interests of warranty repair, but is


accompanied by the purchase of the product warranty card or record
product shipments, otherwise will not be accepted.

-2-
3.3 Certifictae Of Compliance

4. Caution and Safety Tips


4.1 General warnings and safety tips

Use safety ˙The products for dentistry for the diagnosis, treatment,
requirements surgery. It can only be used by trained professionals.
˙When using this treatment requires attention
contraindications (if present) in each chapter (for
example: Section "Treatment Devices") will be
described.
˙This device should not be in danger of the use of
explosive areas.
˙Intended use also includes compliance with the
instructions for use and maintenance instructions.
General Safety Tips The device may only trained professionals to operate.
Equipment changes ˙"Installation site" must be in strict accordance with our
and extensions requirements.
˙As the manufacturer of dental self-treatment station, we
only meet the following conditions responsible for the
safety of technical equipment: Maintenance and repair
work performed only by us or expressly authorized by
our units if parts are damaged, you should use original
parts replace.
˙We propose to obtain the type of job and the job scope
and other relevant documents in the implementation of
such work, as necessary or incidental rated data in the

-3-
Equipment changes field changes within, should also include the date,
and extensions company description and signature. Change the device
may affect the operation, patients or third-party security
under the law is not allowed.
˙Change the device may affect the operation, patients or
third-party security under the law is not allowed.
˙Based on consideration of the safety device, this product
is only allowed to use original parts Kuang Yeu Medical
Co., Ltd. or by a third-party vendors expressly approved
by Kuang Yeu Medical Co., Ltd. manufacturer of
accessories. Should the unauthorized use of unlicensed
accessories, the risk borne by the user.
˙If the connection without Kuang Yeu Medical Co.,
Ltd. licensed device, those devices must comply with
existing standards, such as:
Technical data for devices (such as computers) and the
IEC 60950, for medical equipment IEC 60601-1.
˙Treatment table displays must comply with the IEC
60950 standard. Monitor speaker connector can only be
connected in line with IEC 60950 standard (such as a
computer) or IEC 60601-1 standard equipment is
prohibited connect stereo equipment.
˙If you create a new system during the installation
process, the system must meet the IEC 60601-1
Standard Version 3 of the relevant requirements.
Sanitation and ˙Very low immunity in patients with no access to the water
disinfection treatment station. We recommend that you use sterile
solution.
˙Sanitation and disinfection must comply with the relevant
national requirements and recommendations.

4.2 Use of the logo instructions

Mark Description
SEQ ID
Note: This symbol should be indicated together with
the serial number of the instrument.

Waterproof protection strength

Dust protection strength (dental unit)

Products bearing your logo shown at left. In the


European Economic Area, the product complies with
2002/96 / EG Directive and the relevant national
laws. The directive provides that: use of the product /
disposal to comply with environmental requirements.
Not use the product as household waste!

-4-
Please observe the relevant provisions where the
disposal of the country.
Disposal method according to EG Directive 2002/96
provisions on waste electrical and electronic
equipment, we suggest
The product must comply with the provisions of the
directive and within the European Union (EU)
disposal in accordance with the range of specialized
processes.
Please note!
If you can not comply, it may result in damage to the
equipment or the user and / or patient injury.

BF-type electric shock protection grade use


components.

The sign next to the nameplate mounted on the


instrument.
Note: When using the instrument, please follow
instructions for the.
Instructions for use:
Before using this instrument, you need to read the
instructions.

Manufacturer

Grounded Functional earth

4.3 Product Label


75mm

50mm

-5-
5. Considerations
■ This product is only for dentistry diagnosis, treatment and surgery.
■ This product contains electrical mechanical moving parts, make sure not to touch
anything in the range of movement of the work.
■ Patient chair before use, make sure the pillow is locked.
■ When not using this product, please minimize patient chair position, and disconnect
the water, air, power.
■ This product belongs to the patient's chair intermittent operation of the equipment,
continuous exercise time should not exceed 1 minute.
■ Not be in a high temperature, high humidity, vibration, flammable, unbalanced
environment using this product.
■ You must pass before using the water heater, the water heater to prevent dry and
cause a short circuit.
■ Equipment in need of repair or maintenance, please contact qualified service
personnel must disconnect the water, air and power supply before maintenance.
■ When not using the device, be sure to take water, air and power supply is
disconnected.
■ Recommends the use of dental unit for a period of six years, the deadline may cause
the pipe to aging, function failures, and may even cause an accident, please contact a
qualified medical equipment recycling unit contact, promptly eliminated or replaced
with new equipment.
■ Not be used in flammable anesthetics mixed with air or gas mixed with oxygen or
nitrous oxide flammable anesthetic gases situation.

6. Transport and storage


6.1 Transport conditions: up, handled with care, anti-moisture.
6.2 Transport and storage conditions.
■ Transportation environment: Temperature -10℃ ~ +45℃, Relative humidity ≤60%.
■ Storage Environment: Temperature -10℃ ~ +45℃, relative humidity 10%~90%,
non-corrosive gases and ventilated room.
■ Atmospheric Pressure: 500 hpa to 1060 hpa.
6.3 Transport laminate requirements: not more than three wooden laminated
wooden transport or storage time.

-6-
6.4 Description labeled with warnings.

Mark Description

Remarks:
For the user and / or assistant very important
identifying information.

Temperature limits:
Note: The temperature near the upper and lower
limits indicated on the line and below the line at.
-10℃ ~ +45℃

Humidity limits:
NOTE: Humidity limits should be displayed near the
mark on the line and below the line at.
Relative humidity 10%~90%

Atmospheric pressure limits:


NOTE: Atmospheric pressure limits should be
displayed near the mark on the line and below the
line at.
Atmospheric pressure: 500hpa to 1060 hpa

Up, handle with care

Fragile, handle with care

To avoid the rain


NOTE: This symbol means "dry"

Stack limit: no more than 3 layers

-7-
7. Product Introduction
7.1 Functions and intended use:
Products applicable to the medical department dentistry for diagnosis, treatment
and surgery.
The product structure in the form of dental unit and a dental patient chairs integral,
in conjunction with accessories. The dental treatment
Treatment station products are made by physicians operating unit comprises:
instrument tray, shadowless; nursing unit comprising: a water tank, a spittoon; patient
using an electric unit including dental chairs, pedals, three units combined group
together.

7.1.1 Dental Unit LEADER product composition in the table below:


Distinction
Item Device Name With
unit
Cantilevered instrument tray,
Electronic X-ray film viewer,
Instrument tray Slow motor,
Physician
1. Fast-fit with three-way Syringe,
unit
Instrument tray air brake system.
Shadowless Non-contact sensor control LED
Lamp shadowless.
Assistant bit touch panel,
Handpiece back to the air system,
Water heating means,
Air negative pressure strong suction
Nursing Tanks, spittoon, system,
2.
unit Sign tank water negative pressure weak
suction system,
Automatic water supply equipment.
Into the drainage, air, power
installation box.
Patients Biaxially headrest design,
Electric dental
3. make Digital control patient chair,
chair
With unit Multifunction foot control.

7.1.2 Security features


■ Classified according to protection against electric shock type: Class I
■ According to the degree of anti-shock classification: BF type application part
■ According to the degree of protection from liquid classification: IPX0
■ Classified by operation mode: Intermittent loading continuous operation, in line
with the dentist works (fixed connected devices)
■ Devices rated voltage and frequency 220VAC / 50Hz
■ Power input devices: 400VA

-8-
8. Conditions
To ensure the normal use of the product, the working conditions are as follows:
■ Ambient temperature: 10℃ ~ 40℃.
■ Relative Humidity: 10% ~ 90%.
■ Atmospheric pressure: 700hPa ~ 1060hPa
■ Rated voltage: 220±10VAC / 50Hz
■ Water pressure: 200 ~ 400Kpa
■ Air pressure: 5kg/cm2

9. Technical Data
9.1 Patient chair
Rated voltage 220VAC / 50Hz
Rated power 400VA
Master switch fuse 6A (Fast)
The total carrying capacity About 135 kg
Headrest carrying capacity About 30 kg
The maximum size of the seat 1200x550mm
The maximum size of the backrest 580x460mm
Seat surface rises maxmum level 650mm
Seat surface rises minimum level 460mm
The maximum backrest inclination angle 80˚
Backrest tilt angle of the horizontal 15˚

9.2 Shadowless Lamp


Non-contact sensor control LED shadowless (Optional)
Light form LED Chip
Color Temperature 4,300K/5,500K
Spot range (irradiation distance 700mm) 85*150mm
Adjustable brightness range (Manual / Sensor adjustment) 8,000~40,000Lux
Input voltage AC 12V~24V
CRI >85%
Power consumption 10W

-9-
9.3 Workspace

■ Length of not less than 3500 mm


■ Width not less than 2600 mm
■ A height of not less than 2000mm

■ Lowest position the seat surface: 460mm

- 10 -
■ The highest position of the seat surface: 650mm

- 11 -
10. Product Structure
10.1 Structure Representation

2.

5. 4. 8.

1.

3.

6.

7.

Model: FRONTIER
1. Instrument tray
2. Shadowless lamp
3. Water tank
4. Spittoon
5. The passenger's seat
6. Dental electric chair
7. Foot Controller
8. X-ray film viewer

- 12 -
10.2 Dental Unit description of each unit
10.2.1 Physician operation unit (instrument tray, shadowless lamp)
■ Instrument tray
as the picture shows:

5.
1.

4.
2. 3.

1. Instrument tray table


2. Table move handle
3. Molar machine equipment rack
4. Chair Position function operator panel
5. X-ray film viewer

Functional purpose of operation:


1. Instrument tray table
Is designed to provide physicians in clinical operation, placing the mouth
was subject to oral drug auxiliary equipment or small dish of use, and
surgical procedures arising, provide a suitable work table tops.
2. Table move handle
It is an aluminum metal, connected to the main body of the instrument
tray, provides physicians in clinical practice, suitable for the work surface
angle adjustment.
3. Molar machine equipment rack
Molars by two high-speed handpiece, a slow handpiece, an Three-Way
syringe combination, this product will physicians in the clinical course of
the oral.
cavity, and provides the most appropriate power source equipment
(water, air) and give precise control to achieve the repair and cutting of
hard tissues in the oral cavity.
4. Chair Position function operator panel
Integration is the key operating functions of electromechanical dental
chairs, providing physicians due to different clinical treatment chair
position angle position, the operational requirements and settings,
physicians and patients to achieve optimal treatment position adjustment
effect.
5. X-ray film viewer
Is composed of a low current LED bar lights, refracting light through the
mask, and aluminum paste composition, use of light to render the X-ray
image, in order to facilitate analysis of physician medication.

- 13 -
■ LED color temperature of two kinds of non-contact sensor control
shadowless

9.

8.

7.

1. 2. 3. 4. 5. 6.

■ Functional purpose of operation:


1. LED chip driver is fixed pole:
LED light source for fixing, wiring and color temperature of the switching
mechanism, etc., and the reflection from the lens fixed installation, a fixed
distance (700mm) when irradiated to render suitable for clinical and
medical needs of the focal length of the radiation spot (85 * 155mm).
2. LED color temperature switching button (switchable 4,300 / 5,500K color
temperature):
Color temperature designed to meet the medical procedures used in
lighting and colorimetric lighting needs time.
3. The steering angle of the movable arm balance:
A balanced combination of the four-bar linkage arm A and arm B can be
made to move the steering arm, which provides the physician at the time of
the oral cavity lighting, moving lights to the appropriate treatment site.
4. Handle:
Shadowless lamp used to move and adjust the angle.
5. Sensing controller:
You can control the light source power switch and light brightness via a
non-contact sensor, reducing the risk of cross-infection.
6. Handle fixing screws:
With fixed handle, shadowless lamp after use, the handle can loosen fixing
screws, separate the handle to sterilize clear cleaning job.
7. Shadowless lamp manual brightness adjustment knob:
Shadowless lamp used to adjust the brightness manually.
8. Shadowless lamp Power Switch:
Shadowless lamp switch power supply.
9. Shadowless head rotating elastic damping adjustment screw:
Used to adjust the fixed rotary head shadowless lamp tightness.
- 14 -
10.2.2 Nursing unit (Water tank body, Drainage boxes, Spittoon)
■ Water tank body

2.

1.

Objective Function Operation:


1. Water tank body
Attracted by the negative pressure systems, electromechanical integration
control lines. The control signal can be linked to the physician operating unit
Line aggregated main box, which provides water and air supply to the
sub-function to the desired device power source.
2. Passenger's seat
Sucking air from the negative pressure suction fittings, suction flow vacuum
sucking weak joints, three-Way syringe and position control surfaces plate
combination, which gives medical staff physician in the course of treatment
carried out with each other four hands to operate, giving physicians in the
treatment of surgical largest medical assistance.
3. Drainage boxes
It consists of a total intake control valve, the total water control valve, water
filter, air filter, drain assembly, composed of the main power line. It provides
dental unit all power sources, including sewage air purification and water
treatment functions arising from health care to provide a clean clinic
environment.
■ Drainage boxes

■ After opening the drainage box cover, pressure regulator to 4.5~5kg/cm2

- 15 -
■ Spittoon

1.
3.

2.

As illustrated by the mouth dish, water supply and housing, water supply pole with
infrared sensors constitute spittoon group.
Objective Function Operation:
1. Mouth dish
It is made of porcelain body, a funnel shape, flush lever combination, for the
treatment of patients in the process due to the oral wound rinse liquid out of
the vessel after the recovery of the discharge, when patients in the oral fluid
ejection completed by flushing release lever to keep the mouth clean water
dish sanitation.
2. Water supply and housing
Porcelain by the body into a groove shape, guide the course of treatment the
patient needs to clean water cups placed position, and its function is needed
for patients, providing clean water in the course of treatment to flush the
wound in the mouth, to maintain oral health.
3. Water supply pole with infrared sensors
By the infrared sensor and water lever body composition constituted, its
function is to provide patients with the diagnosis and treatment of clean The
water supply.

- 16 -
10.2.3 Patients using the unit (electric dental chair)

5.

4.

3. 1.

2.

As shown by the:
1. Hydraulic pump power pack (Plastic shell inside)
2. The dental chair support base
3. Lift the seat parallel link
4. Tilt the seat recline
5. Headrest
Incorporated in and constitute a dental patient chair (electric dental chair).

Objective Function Operation:


Electric dental chair, which provides dentists at Dental Clinic patients may be due
to the tall, tall, short, fat, thin standing clinics or physicians or clinics seat position,
by lifting the reclining chair adjusted to achieve the desired treatment treatment
site, while the electric chair ergonomic dental treatment process may be moving
chair position, providing the most comfortable patient care environment.

1. The running time: one minute of work, suspended 10 minutes.


2. The maximum load capacity: dental chair maximum weight
capacity of 135kg, can not exceed this value.

1. headrest can carry a maximum weight: 30kg.


2. When the patient's head is in the headrest, do not move.

- 17 -
11. Product Installation
11.1 Installation space
11.1.1 Environmental Requirements
In order to ensure the normal operation of the product, install it in a clean, dry,
well-ventilated place, and ensure that this product is in contact with a solid
floor leveling.
11.1.2 space requirements
To ensure that the movement of products within the scope of work does not
touch anything, it must achieve to ensure the installation space.
The following requirements:
■ length of not less than 3,500 mm
■ width of not less than 2,600 mm
■ height of not less than 2,000 mm

11.2 Installation Handling


■ In the handling process, do not move to lift the boom, arm, assistant rack,
cabinet, plastic case and so easily broken parts, so that its result from excessive
handling force damage.
■ In the handling process, please stable good tool tray, turning left and right and
avoid accidental touch.

Installed Note:
Follow the installation steps must be shown, each part before assembly
installed together. Make sure the power specifications of this machine
with the same specifications are used locally, installed after the ground
should be as tidy, in order to avoid tripping the user or patient. The
equipment to be professionally installed, otherwise malfunction caused
not covered by the warranty.

- 18 -
11.3 Treatment Basics piping installation instructions

Wall

6. 3.
2.
7. 4.
Wall
8. 1.
5.

9.

11.3.1 Power supply


Rigid PVC, 3/8" outlet conduit of the floor stand. Power cord Install the two
sockets, wire and tube length from 300mm long. (Often 5A, 15A at the
same time instant start operation 10A).
11.3.2 Air duct
Rigid PVC, 1/2", an inner diameter of 8mm air pipe (pipe minimum
pressure 8kg/cm2) PVC pipe out of the ground can be.
11.3.3 Inlet
Rigid PVC, 1/2" from the ground and then water, up to but not more than
150mm.
11.3.4 Drain
Rigid PVC, 1 1/2" (inner diameter φ40mm) exports from the ground 50mm.
11.3.5 Reverse osmosis water reserve pipe
Rigid PVC, 1/2", from the ground H = 30mm.
11.3.6 Aspirator vacuum equipment pipe
Rigid PVC, 3/4".
11.3.7 Aspirator vacuum control line pipe
Rigid PVC, 3/8" (can not be shared with AC 220V pipe)
11.3.8 Information pipe
Rigid PVC, 1".
11.3.9 Piping box: Movable or fixed.

Way above the pipe fitted for reference only, as per usage adjustment
fitted.

- 19 -
12. Instructions for use and precautions

When using the product, make sure to open water, air, electricity.

When the power is interrupted and then restored, in addition to


operations carried out by temporarily interrupted, the security issues will
not occur. When To continue working, you can restart.

Please note: In order to take full advantage of all the features of dental
treatment provided, please read the instructions carefully.

12.1 Dental chair electric power up

Main power switch

Dental chair main power switch is located below the dental chair cushion,
turn the power on the main switch LED will light.

It recommended before starting treatment for the first time every time
the handpiece and spittoon cleaning to flush out waste.

For safety considerations, before leaving the chair after use Always
turn off the power off.

Patient chair before use, make sure the headrest is locked.

NOTE: Danger of collision!


When rotating patient chair, careful not to collide with other
components of the treatment station, personnel or equipment
appliance.

- 20 -
12.1.1 Use the headrest

Headrest angle adjustment knob

■ Loosen the rotary knob fixed headrest, the headrest bracket that can be
easily adjusted to the position desired by the user.
■ When the user moves to the desired position, the locking knob fixed
headrest, the headrest can be fixed.

Angle adjustable headrest

During operation, any prohibited items into the headrest moving


range.

Patient chair before use, make sure the headrest is fixed.

- 21 -
12.1.2 Safety Switches Function Description
Safety switches have the following settings:
■ Pinch the seat down to a safe switch.
■ Chair position sleep function stop switch.
■ Chair position actuator intelligent information feedback stop switch.
■ Chair position feature automatically tests the safety function.
■ Traveling between the chair position in case of emergency, press any keys
a chair seat function will immediately stop working.
More security features to ensure that the product design for operator safety in
use to prevent and reduce operational risk on diagnosis and treatment
equipment.

12.2 Instrument tray Instructions


12.2.1 Instrument tray Function Description

2.

1.

5. 3.
4.

1. Instrument tray table


2. X-Ray film viewer
3. Functional operation panel chair position
4. Instrument tray moving handle
5. Handpiece Air / Water adjustment knob

Instrument tray table maximum load of 2 kg, must not exceed


this value

Note: Please attach a weight average countertop basin instruments


assigned to the instrument tray table.

- 22 -
12.2.1.1 Instrument tray function operator panel

6. 5. 4. 1. 10. 8.

7. 3. 2. 9. 11.
1. Key „Programmable Position 1“
2. Key „Programmable Position 2“
3. Key „Programmable Position 3“
4. Key „Patient Chair Reversion“
5. Key „Last Position Memory“
6. Key „Return To Rinse Position“
7. Key „Cup Filler“
8. Key „Patient Chair UP“
9. Key „Backrest DOWN“
10. Key „Backrest UP“
11. Key „Patient Chair DOWN“

Chairs in the road between, as long as the touch panel or any chair
foot controller keys to stop.

12.2.1.2 Chair position memory setting method


1. The position of the chair to adjust the desired position
2. Long press key „Patient Chair Reversion“ (about six seconds time) beep, set
lights flashing, enter Setup state.
3. Press Key „Programmable Position 1“ or „Programmable Position 2“ or
„Programmable Position 3“ position memory to complete this setting.

- 23 -
12.2.1.3 Water and air adjustment

Handpiece water / air pressure adjustment:


With air / water pressure regulating valve for adjusting the water and air supply
handpiece, and has a cooling air, cooling water for use. Its control valve located
below the instrument tray

1.

2.

3. 4. 5. 6.

1. Instrument tray operating handle (when the instrument trays shall need to
move the red circle at the button pressed)
2. On / Off switch
3. Water / air adjustment knob
4. Water / air adjustment knob
5. Water / air adjustment knob
6. Water / air adjustment knob

12.2.1.4 X-RAY film viewer Description

The x-ray film viewer side of the switch to the "On" lights turn on, playing to
"Off" light is turned off.

- 24 -
12.2.1.5 Handpiece installation

No load running and over-pressure start of the handpiece is


strongly forbidden.

• Align four holes of joint core with relevant connection pipe of handpiece
• Align adaptor sleeve with screw mouth of handpiece, gently screw it in,
moderately screw it tight until it won‘t leak.

1. Handpiece
2. Joint Unit
3. Joint Sleeve

12.3 Three-way syringe, Saliva ejectors instructions for use

12.3.1 Three-way syringe 1


2 5/6
Air: press button (5) „air“ 3

Water: press button (6) „water“


4

Mist: press simultaneous buttons + (5) + (6)

1. Nozzle
5.
2. Latch ring 6.
3. Three-way syringe
4. Handle
5. Jet(Air) Button
6. Spray(Water) button

If adjustment is needed, the air and water regulating knob on the bottom
reverse of instrument tray can be adjusted with the same method as above.

Specification:
1. Water flow: pressure 200 kpa, pressure 500 kpa, the water flow is
not less than 120ml / min
2. Spray volume: pressure 200 kpa, pressure 500 kpa, spray amount
not less than 1,000ml / min

- 25 -
12.3.2 Saliva suction

■ Assistant took from the shelf the strong and weak tip, automatic pumping work.

Three-Way Syringe

Strong pipette tips Weak pipette tips

■ How strong saliva suction

Saliva suction switch down Saliva suction switch up


Strong saliva suction:ON Strong saliva suction:OFF

■ How weak saliva suction

Saliva suction switch down Saliva suction switch up


Strong saliva suction:ON Strong saliva suction:OFF

Please replace the original position when not in use saliva suction.

- 26 -
12.4 LED Shadowless lamp operating instructions

6.

5.

1. 2. 3. 4.

Shadowless component parts:


1. LED light chip driver holder
2. Handle
3. The sensor controller
4. Handle fixing screws
5. shadowless lamp manual brightness adjustment knob
6. shadowless lamp power control switch

■ Shadowless lamp illumination distance focal length factory setting is 27.5”


(700mm) need not be adjusted, but in order to maintain the integrity of
shadowless lamp illumination target spot (85*155 mm), there will be a spot
peripheral diffused serious situation or occurrence of color affect groups of
results, we recommend you use to adjust the irradiation distance: 700mm ±
50mm.

Use LED chip specifications:


1. The color temperature of 4,300K/5,500K LED chips
2. Shadowless lamp input voltage 12~24VAC

Never touch the LED light when the light source chip to avoid
scalding.

- 27 -
■ When shadowless lamp balance arm tension is too tight or too loose, You can be
adjusted using the following methods:
1. Open the front cover balance arm, 1/2 "open-end wrench to rotate the tension
adjustment screw, Balance arm too tight, turn the screw clockwise to adjust,
When the balance arm is too loose, turn the screw counterclockwise to adjust.
2. Use M1.5 L-type hex wrench to adjust the balance arm tension release
tightness, Tighten clockwise rotation, counterclockwise rotation relax.
1.

2.

■ When shadowless lamp head rotation too tight or too loose, or the shadowless
lamp balance arm rotates too tight or too loose, You can be adjusted using the
following methods:
1. Use M1.5 L-type hex wrench to adjust the shadowless lamp head screw rotary
damper, Tighten clockwise rotation, counterclockwise rotation relax.
2. Use M1.5 L-type hex wrench to adjust the shadowless lamp balance arm
screw rotary damper, Tighten clockwise rotation, counterclockwise rotation
relax.

1. 2.

- 28 -
■ LED Shadowless lamp operating instructions:

1. Power control switch: left or right to switch


the power switch is turned ON; turn off the
power switch to the middle OFF.
1.
3. 2. shadowless lamp manual brightness
adjustment knob: clockwise rotation
2. brighten, dim the counterclockwise
rotation.
3. Handle fixing screws: grip required
dismantling disinfectant cleaner use,
clockwise rotation of the lock,
counterclockwise rotation loosen.

4. Shadowless induction control:


(1) Before induction control operation, make sure the power control switch
turned ON.
(2) Sensing controller sensing range 10 ~ 150mm, please work within the
sensing distance.
(3) In shadowless lamp off state, the hand slip through the induction range (The
following figure shows the right side), the palm leaves the sensor range,
shadowless lamp will automatically light up. When the shadowless lamp is
on, slide the hand through the sensing range (The following figure shows the
right side), the palm leaves the sensor range, shadowless lamp will
automatically turn off.
(4) In the shadowless lamp lit, the hand moves to the induction direction and
stay within sensing range of about 0.3 to 0.5 seconds (The following figure
shows the right side), keep your hands away, shadowless brightness is
automatically adjusted. Starting reorientation of lightening, as long as the
hand is not removed, shadowless automatically and continuously adjusts the
brightness to the brightest, to emits a short beep sound and up to maximum
brightness, if the hand has not removed automatically and continuously
dimmed shadowless lamp brightness to the darkest, emits a short beep
sound reaches the minimum luminance. Adjust the brightness, have not yet
reached the maximum or minimum brightness, if the hand away,
shadowless lamp brightness will immediately stop to adjust the brightness.

- 29 -
(5) In the shadowless lamp lit, do the sensor to adjust the brightness control
sequence is as follows: If the previous brightness adjustment to brighten
your hand away after stopping to adjust the brightness, again moved his
hand to perform brightness adjustment, adjust the brightness dimmed
direction, brightness adjustment mode so the order cycle.

12.5 The passenger's seat function operating instructions


12.5.1 The passenger's seat Part Description
1. 2. 3. 4.

1. The passenger's seat the operator panel


2. Strong saliva suction
3. Three-way syringe
4. Weak saliva suction

12.5.2 Operation Panel


1. Return to the rinse position
2. Last position memory
1. 2.
3. Patient Chair reset position
4. Patient Chair UP trimming function keys
3.
5. Patient Chair DOWN trimming function keys

4. 6. 6. Backrest DOWN trimming function keys


7. Backrest UP trimming function keys

5. 7.

- 30 -
12.6 Foot controller functions Operating Instructions
12.6.1 Functions Instructions

1. 1. Patient chair Position trimming UP


3. 4. 2. Patient chair Position trimming DOWN
2. 3. Backrest Position trimming DOWN
5. 6. 4. Backrest Position trimming UP
5. Return To Rinse Position
6. Last Position Memory
7. W A 8.
7. Handpiece pedal – Water
8. Handpiece pedal - Air

12.6.2 Handpiece pedal work instructions


■ When only the left pedal (W) is stepped on, the handpiece only water work.
■ When only the right pedal (A) is stepped on, the handpiece only air work.
■ At the same time step on the left and right pedals, handpiece water, air
source simultaneously.
13. Cleaning Care and maintenance instructions
The equipment should be placed in a clean, dry environment,
and use (temperature 0 ℃ -60 ℃; humidity of 10% -90%) within
the following environmental limits, after each use of the machine
after the machine back to its original position. Note that regular
cleaning wipe the surface of the machine. After closing the office
every day, you must supply, water supply, air supply switch is
turned off.

13.1 Dental chair Exterior Cleaning and maintenance instructions

Working day after using a damp cloth, but do not use


solvent-containing cleaning supplies to clean.

13.2 Shadowless lamp, Ceramic dish cleaning and maintenance instructions

Shadowless lamp use dry with a clean cloth to wipe clean.

Use clean water to wash ceramic dish and wipe clean surface.
Note: For health and avoid infection, do not hand in direct
contact with the ceramic dish.

- 31 -
13.3 Filter tank body cleaning and maintenance instructions

Water tank body Filter

■ Filter cleaning method


■ Unscrew the filter cover CLOSE OPEN
■ Remove the filter
■ Will remove the impurities,
and then rinse with water
■ Replace the filter
■ Be sure to completely filter
into the base, To ensure no
leaks
■ Rotate the filter cover tightly

Do not use compressed air source air compressors for industrial use,
and regularly clean or replace the filter.

13.4 Three-way syringe, Handpiece, Saliva suction cleaning and maintenance

Handpiece cleaning and maintenance, follow your handpiece


manufacturer's maintenance manual

■ Three-way syringe, Saliva suction, After the daily work, please disassemble
and use 135˚C / 375˚F disinfection tank disinfection cleaning.
■ Weak saliva suction, Be sure to clean out the filter element.

Working day after using a damp cloth, but do not use


solvent-containing cleaning supplies to clean.

- 32 -
14. Troubleshooting instructions
14.1 LED Shadowless lamp
■ Troubleshooting
1). When the lamp brightness is not enough, you can adjust the brightness
using the brightness manual adjustment knob or the induction control
method..
2). Light does not shine, check Remedy:
a. Check whether the LED bulbs burn and power control wiring is off.
b. Use the correct voltage 12~24VAC input.
c. Switch is bad.
d. Check the power control circuit board is faulty.
3). Sensor control insensitive or impossible to control, control methods:
a. Check the power control circuit board and sensor control circuit board is
faulty.
b. Check the sensor control circuit board is well fixed and the correct
position.
c. Different control modes can be set to match your usage habits (as shown
in Figure 1.).

OFF ON

LED

■The power supply is


normally off

OFF ON

LED

■The power supply is


normally on

OFF ON

DIMER
Figure 1. Control mode setting JUMP
■Only the control power
switch can be sensed

OFF ON

DIMER

■You can simultaneously


control the power switch
and adjust the brightness

- 33 -
14.2 Treatment table

■ Troubleshooting
1). Strong or weak suction aspiration occurs when insufficient attractiveness
excluded methods are:
a. Check attract joint adjustment switch is normally in the ON position.
b. Check the pylons switching valve stuck or if there are leaks.
c. Check attract joint internal filter is clogged with impurities, has attracted
filter strainer cleaning.
2). No water supply problem occurs when the infrared sensor water supply,
excluded methods are:
a. Transmittance glass used to check whether too light, is not conducive to
infrared sensors.
b. Check the glass is placed in the correct position.
c. Check whether the sensor housing water droplets adhered, can be used
to clean three-way syringe favorable induced effects.
d. Use manual pressure water, check electric water inspection is normal.

3). High-speed handpiece, when teeth were cut, such as cutting force is not
enough to exclude methods are:
a. Check the instrument tray on the high-speed handpiece, the rated
working pressure gauge whether 2.5kg/cm2 ~ 3kg/cm2.
b. Check foot control valve is stuck.
c. Use instrument tray beneath the pressure adjusting valve be adjusted.
(Red circle marking the position shown in the following figure).

Water / Air adjustment knob

- 34 -
4). High-speed handpiece, such as water shortage were cutting teeth, excluded
methods are:
a. Use instrument tray beneath the pressure adjusting valve be adjusted.
b. Check foot control valve is stuck.
c. Use water spray three-way syringe tests.
d. Influent water is turned off.

5). Dental Chair, three-way syringe, high-speed handpiece occur no pressure


or no pressure, the problem of exclusion methods are:
a. Check drain tank inside water filter blockages.
b. Check the drainage tank internal pressure valve of the total air pressure
switch is normal to 5kg/cm2.
c. Whether internal curved lines ring true.
d. The air compressor is turned on and running.
14.3 Patient chair
■ Troubleshooting

1). When the patient chairr position adjustment occurs without any response,
troubleshooting methods are:
a. Check the bottom of the chair cushion, the total power of the power
control box is turned on.
b. Check the mains fuse is intact.
c. The power cord is off.
d. Check the power control board inside the box IC Green PU tube,
switching control of the security lodged presented OFF state.
e. The amount of e measured input voltage is normal.

2). Instrument tray chair position button function disorders, excluded methods
are:
a. Check the foot controller to control chair position, the control switch is
sensitive, no stuck (including O key and S key test).
b. Openers reels, checking key IC board control line is off.
c. Open spittoon drainage box, check each power cable connector is
firmly.
d. Anti-drop safety seat chassis switch examine whether abnormal stuck.

- 35 -
15. Diagram of the air and water pipes

- 36 -
- 37 -
16. Patient chair, Dentist element – Wiring Diagram

- 38 -
17. Water Unit With Swivel Spittoon Bowl – Wiring Diagram

- 39 -
18. Foot Control – Wiring Diagram

- 40 -
19. Tubing Identification

- 41 -
- 42 -
MEMO
MEMO
Kuang Yeu Medical Co., Ltd.
Address: No. 10. 7th Road, Industrial Park, Taichung 407, Taiwan.
Tel: +886-4-2350-8866
Fax: +886-4-2350-0099
E-Mail: marketing2@kuang-yeu.com.tw
Website: www.kunag-yeu.com.tw

The Company reserves the right to make any product design improvements of
the product described as corresponding to each function without prior notice, all
interpretation of the Company.