( Technical Specification

Analogue Radiography System


EVA-HF525
INTRODUCTION

G EMS S EVA - HF 52 5 d ia g no stic x -r ay sy stem i s


designed with simply configuration for convenience
and user- friendly in a general radiographic procedure.
It features a high frequency X-ray generator and high
quality X-ray output, giving you a high definition images.

FEATURES

Great Expandability

It can be set up, such as CR, film, portable, wireless and even fixed, depending on users’ demand, further
leading to easy entry into digital radiography.

Economic Feasibility and Convenience

By using a wired flat-panel detector, it provides better outcome while establishing the environment as the
existing conventional X-ray and CR environments at the same time. This compact-type system helps
equipment move easily, and the table designed to fit for this system makes it possible to minimize the
range of patients’ movements further improving users’ workflow and minimizing their operation time.

Compact Design Maximizing the Spatial Availability

Simply consists of floor-mounted and wall stands, this product can even support special shooting positions
in association with a portable detector, while increasing the spatial availability of an X-ray room.
EVA-HF525
SYSTEM CONFIGURATION

Standard Configuration
Unit Model Name Remarks
High Frequency Generator – 40kW
X-ray Tube – E7239X
Analogue Radiography
EVA-HF525 4-way floating table
System
Operating Console system
X-ray Grid
Option
Unit
PAU (Power Assistant Unit)
PSU (Power Supply Unit)
AEC (Auto Exposure Control)

Specification

High frequency inverted Generator Collimator


Type Manual
40kW Generator Power output DC 24V
Output power 40kW Automatic illumination
Line voltage 220VAC+-10%,Single phase 50/60Hz Lamp timer With timer for lamp
(30sec)/configurable timer
Current 55A
APR Memory No
Power consumption 55KVA Laser guide No
kVp step 40 - 125kvp (1kvp step)(±5%) Max. Window 47 x 47 cm (SID 100cm)
10/12.5/16/20/25/32/40/50/63/ 80/100/ Min. Window 5 x 5 cm (SID 100cm)
mA step
125/160/ 200/250/320/400/500 (±10%)
mAs 0.4mAs - 630mAs (±10%+0.5mAs) Grid
Time 4msec - 10sec(±2%+0.1msec) Materials Aluminum interspaced x-ray grid
Dimension 43.8 x 46.3 cm
Maximum condition for
320mA@125kVp at 0.1sec Densities 85L/inch
voltage
Ratios 8:1
Maximum condition for
80kVp@500mA at 0.1sec Focus Range Table: 33”~44” / Bucky Stand: 40”~72”
current
Communication RS-232C Full Digital
Movement

E7239X Tube Stand


Item E7239x Longitudinal Travel 194cm
Focal Spot value 1.0 / 2.0mm 65-172cm
Vertical Travel
Operating Voltage 40 ~ 125 kV
X-ray Tube rotation ±90°
Target Angle 16°
Transverse Tube Travel ±30cm
Permanent Filtration 0.9mmAl/75kV
Anode Heat Content 140KHU(100kJ) Movement arrested by Electromagnetic brakes
Maximum Anode 667HU/s (475W) 4-way floating table
Heat Dissipation rate Forward-Backward movement ±200mm
Housing Heat Storage 1250 kHU(900 kJ ) Left-Right movement ±600mm
Capacity Table Dimension 2000mm x 740mm
Anode Rotation speed Min 2700 rpm / 50Hz
Table Top to Film distance 78mm
Min 3200 rpm / 60Hz
Input Energy Small focus:
21kW(50Hz) 22.5kW (60Hz) * Technical specification can be changed w ithout notification
Large focus:
42.5kW(50Hz) 47kW (60Hz):
Cooling Method Natural / Forced Air
Operating Temp. 10 ~ 40℃
Operating Hum. 30 ~ 85%
EVA-HF525
OUTLINE DRAWINGS

Manufacturer
GEMSS MEDICAL SYSTEMS CO.,LTD.

Head Office: 61, Dunchon-daero 541 beon-gil, Jungwon-


gu, Seongnam-si, Gyeonggi-do, Korea
Tel: +82-31-764-7321
Fax: 82-31-764-7324
E-mail: sales@gemss-medical.com