You are on page 1of 5

Coronary

Clinical Impact of Stent Design


R ebec c a L N o a d , 1 Co l m G H a n ra t t y 2 a n d S i m o n J Wa l s h 2

1. Specialist Registrar in Cardiology; 2. Consultant Cardiologist, Belfast City Hospital, Belfast Health and Social Care Trust,
Belfast, Northern Ireland, UK

Abstract
The introduction and widespread adoption of drug-eluting stents into routine clinical practice has seen tremendous changes in the
practice of interventional cardiology. For a prolonged period, manufacturers have focused research on drugs and polymers that are
the key to the prevention of in-stent restenosis. However, stent platform design and its clinical implications have now come back to the
fore. This has occurred for numerous reasons, but has primarily been driven by the need for modern stents to perform well in increasingly
demanding clinical scenarios. This paper reviews the historical evolution of stent platform design. Current manufacturing processes and
materials are also explored. Geometric stent construction and its implications for longitudinal stability and the longer term risks of stent
fracture are reviewed. Finally, the implications of the specific stent chosen for different clinical applications including the treatment of
bifurcations and left main disease are also summarised. This article will familiarise cardiologists with the crucial impact of each of these
factors on modern day practice, as well as acute and long-term outcomes for patients.

Keywords
Stent design, longitudinal compression, stent fracture, stent thrombosis, left main stem stenting

Disclosure: Colm G Hanratty is a consultant to Boston Scientific. Simon J Walsh is a consultant to Abbott Vascular, Biosensors and Boston Scientific. Rebecca L Noad has
no conflicts of interest to declare.
Received: 2 April 2014 Accepted: 28 April 2014 Citation: Interventional Cardiology Review, 2014;9(2):89–93
Correspondence: Simon J Walsh, Consultant Cardiologist, Belfast City Hospital, 51 Lisburn Road, Belfast, BT9 7AB, Northern Ireland, UK.
E: simon.walsh@belfasttrust.hscni.net

As the role of percutaneous coronary intervention (PCI) has evolved, range of stent cell construction (open versus closed), cell geometry,
inevitably, so too has the technology associated with this specialty. strut thickness and different materials were also explored (such as
Initially, plain old balloon angioplasty (POBA) was developed as a gold, carbon and a variety of others).
strategy to ‘stretch’ focal stenoses within the coronary arteries
leading to a relief from ischaemia and angina. Whilst potentially of However, it soon became clear that BMS were beset by the late
great benefit to some patients, POBA was beset by a series of Achilles’ complication of in-stent restenosis (ISR). This late ‘healing phenomenon’
heels. These included early, unpredictable and abrupt vessel closure led to loss of the treated vessel lumen and a recurrence of ischaemia.
due to coronary dissection, subacute recoil of the dilated coronary Repeat procedures (target lesion revascularisation [TLR] and target
lesion with the stenosis recurring relatively early, and later restenosis vessel revascularisation [TVR]) became a frequent occurrence after
due to the healing response invoked by vascular injury from the PCI with BMS. It also became apparent that ISR was associated with
procedure. Ultimately, supplementary technologies were needed to many of the design features of these devices. These included longer
resolve these issues and this led to the concept of deploying bare stent length, smaller stent diameter, a high metal:artery ratio and
metal stents (BMS) in the coronary artery. Early BMS were developed to strut thickness.3–5 Clinical features such as the presence of diabetes
more reliably treat short, focal areas of disease. The aim of using these also contributed to stent failure.3 There were both early and long-term
devices was to cover disruption and stabilise dissection associated adverse consequences of ISR for patients.6,7 These events also led to a
with POBA, thus preventing abrupt vessel closure. In addition, the clinical disadvantage for patients with multivessel disease who were
vessel was scaffolded to prevent early recoil and the initial results treated by PCI compared with those who were offered coronary artery
from widespread adoption of coronary intervention with BMS led to bypass grafting (CABG).8,9
benefits for patients.1 One feature of early BMS was that they were
relatively bulky and stiff devices. This reflected their initial function Ultimately, this brought about the development of drug-eluting
for the treatment of focal and straightforward coronary lesions. As stents (DES). These devices were first introduced in the pivotal
the practice of interventional cardiology evolved to meet the needs of Randomized Study with the Sirolimus-Coated Bx Velocity Balloon-
patients with more complex coronary lesions, so too did the stents. Expandable Stent in the Treatment of Patients with de Novo Native
There was an explosion of different manufacturers, stent designs and Coronary Artery Lesions (RAVEL) trial.10 DES had polymers and a
technologies.2 Therefore, early platforms quickly improved, becoming variety of different drugs coated onto stents to stop late lumen loss.
easier to deliver and thus more user-friendly. Whilst the majority of The therapeutic goal was to reduce the inflammatory and healing
these devices were constructed from stainless steel, there was a response that occurred after stent implantation, with the aim of

© RADCLIFFE CARDIOLOGY 2014 89

walsh_aiza_edited.indd 89 12/05/2014 01:10


Coronary

preventing ISR and its clinical consequences. DES were very effective material’s tendency to be deformed elastically or non-permanently
in reducing levels of ISR relative to BMS and their use quickly became when a force is applied to it), yield strength (the stress at which
widespread. However, their introduction and dissemination meant a material exhibits plastic or permanent deformation) and tensile
that the original design concepts that applied BMS now also needed strength (the maximum stress that a material can withstand whilst
to facilitate timed and consistent drug delivery. By this time, the stretched or pulled before its cross-sectional area significantly
design goals that previously applied to stent manufacture had been contracts).14 In clinical terms, these latter features dictate the overall
expanded and significantly altered. Investigations of different drugs, radial strength of the stent itself, in addition to its susceptibility to
polymers and their combinations became prominent and essentially recoil. These two features are not mutually exclusive. However, these
superseded platform design as manufacturers sought to produce properties are crucial for both the acute and long-term performance
increasingly safe and effective devices. of the stent. The evolution away from stainless steel towards other
alloys has been to allow the stent struts to become thinner whilst
One significant effect of DES was that patients with a previously maintaining the overall radial strength of the device.
difficult disease, such as those with diabetes or multivessel disease,
could be treated.11,12 As a result, more and more challenging lesions Most recently, bioresorbable stents have been introduced to the clinical
became amenable to PCI. In addition, stent design was also pushed arena. The most extensively studied stent is currently the Absorb™
forwards by operator feedback that led manufacturers to encompass (Abbott Vascular, Santa Clara, CA, US). This stent is manufactured
potentially desirable properties in their products. These included from a poly-L-lactic acid (PLLA) polymer. This semi-crystalline polymer
improved radiographic visibility, flexibility, device deliverability and is constructed from a number of linked sinusoidal hoops with stent
conformity to the vessel. In general, the overall trend in design goals struts that are 150  µm thick.15 This particular stent cannot be seen
was to produce thinner stents that are more flexible but where the key radiographically and has two small metallic markers sited at the distal
mechanical property of the stent (the radial strength and hence the and proximal stent edges for intra-procedural identification. A number
scaffold within the vessel) has been preserved. of other bioresorbable materials and platforms are under investigation
at various stages,15 although a review of these materials and devices is
Stent platform design and manufacture has now come back under the beyond the scope of this article. Nevertheless, the same mechanical
clinical spotlight for a variety of reasons. These include the expanding constraints and desirable properties are also directly applicable to
use of these devices in previously ‘taboo’ clinical areas (bifurcations, these devices.
chronic total occlusions, left main disease) with an increasingly
complex demand placed on their mechanical performance. It has A major factor in stent design is the geometry of the stent cell
become apparent that these properties are paramount not only with structure (see Table 1). This is dictated by the number and pattern
regard to the mechanical properties of the stents but also to clinical of connectors between rings or hoops. Reducing the number of fixed
outcomes in these challenging clinical scenarios. connectors is potentially desirable as this improves flexibility, delivery
and decreases the metal:artery ratio. However, this also potentially
Stent Materials, Design, Longitudinal Strength impacts negatively on longitudinal strength. It has recently become
and Stent Fracture apparent that longer, thinner and more flexible stents can be less
Metallic stents are manufactured by different processes. These stable in their longitudinal axis. These stents can be ‘compressed’
include laser cut slotted tubes, multilink hoops and the sinusoidal or distorted along the length of the device creating a ‘concertina’
continuous wire, which is a single unit that is wound, folded and effect or longitudinal stent deformation (LSD).16,17 This phenomenon is
welded into shape. The stent must apply sufficient radial force on now well-understood and usually relates to an interaction between
the wall of the diseased coronary artery so that the vessel lumen is guiding catheters and stents deployed in the aorto-ostial position, or
restored to a near normal diameter whilst subsequently scaffolding stents that are relatively undersized after initial deployment that are
the vessel and preventing collapse of the artery in the longer term. subsequently ‘caught’ and distorted by secondary equipment that is
Desirable performance characteristics include low elastic recoil, passed into the coronary vessel.14,16
conformability, high visibility and ease of deliverability. The latter is a
complex parameter influenced by the flexibility afforded by the stent With regard to longitudinal stability, there are several technical design
itself, the properties of the delivery balloon system and the overall factors that are associated with an increased susceptibility to LSD. It
crossing profile of the entire device. has been shown that at compressive forces of 50 gF (0.5 N) or less, it
is possible to shorten18 or elongate19 modern metallic stents. Stent alloy
Mechanical engineering is a science of compromise. Therefore, and strut thickness appear to be somewhat less important with regard to
altering any single feature of a stent inevitably affects other properties. susceptibility to LSD. The construction of the device, number of connectors
There is a complex interaction between every feature of stent design between rings and their geometrical distribution across the device dictate
and how the device behaves in clinical practice.13 The radio-opacity the longitudinal strength of the platform. In general, more connectors
of a stent is mainly dictated by the material (usually a metallic alloy) between rings correlate with increasing longitudinal strength. Where
from which it is constructed, where the resistance to penetration connectors are present, those that are in longitudinal alignment confer
by X-ray is proportional to the cube of the atomic number of the increasing strength, whilst offset connectors are less strong. However,
elements that make up the alloy.13 There are a range of metallic alloys there is a significant downside to increasing the longitudinal strength of
that are employed in commonly used stents. These include stents that the device. This will increase the ‘stiffness’ of the stent, therefore reducing
are constructed from 316L stainless steel, cobalt chromium alloys its deliverability to and conformability within the vessel.
(MP35N and L605) and platinum chromium alloys (see Figure 1 and
Table 1 for examples). The alloy that the stent is constructed from Whilst unrecognised, LSD has the potential to be clinically disastrous
will not only alter the radio-opacity but also the elastic modulus (a for the patient. However, these events appear to be relatively rare.

90 INTERVENTIONAL CARDIOLOGY REVIEW

walsh_aiza_edited.indd 90 12/05/2014 01:10


Clinical Impact of Stent Design

Figure 1: Examples of Some Drug-eluting and Bare Metal Stents

The drug eluting stent is listed above, with the corresponding bare metal version annotated below. The material that the stents are manufactured from and strut thickness are noted. These
are drawn to scale. Strut thickness refers to the axis measured as if from the lumen to the vessel wall.

Table 1: Examples of Commonly Used Metallic Drug-eluting Stents

Manufacturer Drug-eluting Stents Material Drug Connectors/Ring Geometry


Abbott Vascular (CA, US) Xience Prime™ Cobalt chromium (L605) Everolimus 3


Biosensors (Singapore) Biomatrix™ Stainless steel (316L) Biolimus 2


Boston Scientific (MA, US) Promus Element™ Platinum chromium Everolimus 2


Medtronic (MN, US) Integrity Resolute™ Cobalt chromium (MP35N) Zotarolimus 2


The manufacturer, alloys, drugs eluted, number of connectors between rings and stent geometry are outlined.

Where LSD does occur, a three-pronged approach is required Figure 2). Should this occur, OCT will be necessary to demonstrate
to manage the stent deformation. Ideally, preventative measures the complication.
should be undertaken, such as prudent guide catheter selection in
ostial lesions to allow for sufficient support without excessive guide Nevertheless, there is a clear trade-off with increasing the stiffness
engagement. Choosing a stent of an appropriate size and careful of a stent. Whilst longitudinal stability rises with the number of
proximal stent optimisation (usually with softer semi-compliant stent connectors, the risk of stent fracture increases as the stent
balloons in our practice), particularly in the left main stem, will help becomes stiffer. Therefore, older stent platforms are much more
to ensure that secondary devices are unlikely to catch on struts susceptible to this phenomenon. In one clinical study, the 6-connector
and cause deformation. Secondly, a low threshold is required for Cypher stent was more than four-times more likely to fracture than
suspecting stent deformation, particularly when there is resistance newer platforms.20 The incidence of stent fracture with the Nobori™
in delivering post-dilation balloons. Careful fluoroscopy examination 2-connector Biolimus eluting stent (Terumo Corporation, Tokyo, Japan)
and or additional imaging with intravascular ultrasound (IVUS)/optical is reported at >4 % per lesion at nine-month angiographic assessment
coherence tomography (OCT) may be necessary to diagnose this in a study of >1,000 patients.21 This may relate to the exact shape
problem.16 Finally, if longitudinal stent compression has occurred, of the connector between rings for this particular stent. Fracture
cautious post-dilation should be attempted. Small diameter balloons has also been described in the thinner strut, 3-connector Xience™
may initially be required to cross the damaged stent, and these can stents (Abbott Vascular, Santa Clara, CA, US) where a combination
be gradually upsized as necessary. This method has been used with of LSD and stent fracture were also associated with adverse clinical
success in our experience.16 Further proximal stent deployment may events.22 This was particularly likely to occur at areas of stent overlap.
be needed in the setting of vessel damage.14,16 It is worth noting that Another study has suggested an incidence of stent fracture in Xience
this phenomenon can also occur with bioresorbable platforms (see stents at almost 3 % in a large cohort of >1,000 patients undergoing

INTERVENTIONAL CARDIOLOGY REVIEW 91

walsh_aiza_edited.indd 91 12/05/2014 01:10


Coronary

Figure 2: Optical Coherence Tomography Images of a provided that it is recognised. Our experience is that patients do well
Well-deployed Bioresorbable Vascular Scaffold (A) and the over the long term when this is the case. In contrast, stent fracture
Same Stent After Longitudinal Deformation Caused by a
is much more difficult to predict and it also seems to be a more
Guide Catheter (B)
common phenomenon than was previously considered to be the
case. Furthermore, there is a high chance of an adverse outcome with
stent fracture. Whilst LSD is certainly not desirable, the trade-off of a
more flexible and conformable stent platform with less axial stability
may be worthwhile if these platforms prevent later complications
that are likely to occur in significant numbers. This is becoming
increasingly relevant in the era of treating long segments of disease
with stents that can be as long as 48  mm and the frequent need
for overlapping stents (as is common when treating chronic total
occlusions). Newer approaches have been adopted more recently.
For example, the Promus Premier™ and Synergy™ II stent models
(Boston Scientific, Natick, MA, US) aim to retain flexibility through
the body of the stent by using two offset connectors. However, the
Table 2: Stent Models Within Different Manufacturer’s most proximal three rings are linked by extra connectors to provide
Platforms and Manufacturer Recommended resistance to LSD.
Over-expansion Limits
Stent Models Across a Product Range,
Manufacturer Platform Model’s Over-expansion Bifurcations and Over-expansion
Nominal Limit
Expansion
It is imperative that cardiologists are intimately familiar with the
Abbott Vascular Xience 2.25 mm 3.25 mm stents that they use. A differing range of stents occur within the
(CA, US) Xpedition™ 2.50 mm various manufacturer product ranges (see Table 2). Understanding
2.75 mm 4.00 mm the differences between the stents within these product ranges is key
3.00 mm to obtaining a good clinical outcome for the patient. This is becoming
3.25 mm more important as the lesion complexity increases. For example, using
3.50 mm 4.50 mm a two-stent strategy at a bifurcation could potentially have a vastly
4.00 mm
different mechanical result depending on which product is chosen. If
Biosensors Biomatrix 2.25 mm 3.50 mm
a 2.5 mm diameter product is brought from a side branch back into a
(Singapore) Flex™ 2.50 mm
much larger diameter main vessel, the final expansion in the proximal
3.00 mm
3.50 mm 4.50 mm
main vessel and stent cell diameter (and thus lumen achieved in
4.00 mm the main vessel) will be very different than if a 3  mm or 4  mm
Boston Scientific Promus 2.25 mm 2.75 mm version of the ‘same stent’ is deployed in the side branch instead.
(MA, US) Premier™ 2.50 mm 3.50 mm Understanding the implications of these choices is crucial. This is
2.75 mm particularly pertinent when the left main stem (LMS) is being treated.
3.00 mm 4.25 mm
3.50 mm With LMS intervention, over-expansion of current stent platforms is
4.00 mm 5.75 mm frequently required as the devices are brought back from smaller
Synergy™ 2.25 mm 3.50 mm
diameter daughter vessels (LAD or left circumflex). Despite bench
2.50 mm
testing demonstrating that current drug-eluting stent platforms can
2.75 mm
maintain structural integrity beyond the nominal expansion diameter,24
3.00 mm 4.25 mm
3.50 mm
there are theoretical concerns that over-expansion may affect drug
4.00 mm 5.75 mm delivery and cause mechanical disruption of the stent.
Medtronic Resolute 2.25 mm 3.50 mm
(MN, US) Integrity™ 2.50 mm The first concern regarding over-expansion is that it may damage the
2.75 mm stent polymer, leading to uneven drug elution, with a potential for later
3.00 mm 4.75 mm ISR.25 Bench testing of over-expansion of first generation DES found
3.50 mm that balloon dilation induced only minor polymer coating abnormalities,
4.00 mm which have also been noted on undeployed DES.26,27 It is also possible to
replicate this polymer damage by delivering stents through tortuous and
follow-up angiography between 6 and 9 months.23 Stent fracture is calcified vessels. Another concern with respect to over-expansion is the
likely to lead to ISR, acute or chronic occlusion and is certainly not risk of mechanical disruption of the stent. Stent fracture is possible with
a benign phenomenon.20–23 These events are predisposed by certain overly aggressive stent expansion.
features within the lesion or vessel. These include treating the right
coronary artery, using longer stents, areas of tortuosity, calcification, A further mechanical issue with over-expansion that has been queried
stent malapposition and stenting at hinge points. is of the potential for stent recoil. This is a theoretical concern due to
a reduction in metal:artery ratio and an alteration of the scaffold itself.
Therefore, whilst LSD can occur as a sudden and dramatic event Stent recoil has been described when treating atheromatous vessels.28
that complicates a PCI, this can usually be managed and resolved However, in the LMS the usual reason for over-expansion is apposition

92 INTERVENTIONAL CARDIOLOGY REVIEW

walsh_aiza_edited.indd 92 12/05/2014 01:10


Clinical Impact of Stent Design

to the non-diseased vessel wall rather than scaffolding flow limiting Conclusions and Future Directions
plaque. Furthermore, the radial strength of the stent paradoxically The evolution of the role of stents has resulted in a significant change
increases as it is over-expanded. This is due to straightening of both in stent design, primarily driven by the requirements of the disease
the ring and connectors, thus leading to greater strength, not less. that is being treated. Stent design is crucial for acute and long-term
outcomes for patients and it is critical that cardiologists have a
Our experience of LMS PCI is that this is typically a very large vessel. complete understanding of the design features of the devices that
In a study of 125 patients, the mean cross-sectional area of the distal they are implanting.
LMS was 22.6  mm2 (standard deviation [SD] ± 5.4  mm2) and mean
maximal vessel diameter was 5.7  mm (SD ± 0.7  mm).29 Therefore, It is likely that LMS PCI and the treatment of large vessel bifurcations
the vast majority of patients would consequently require post-dilation will become a mainstream application of PCI over the forthcoming
beyond the nominal diameter of all of the large vessel platforms years. Manufacturers may need to consider producing dedicated
of current generation DES during LMS PCI. We also performed a platforms for the treatment of these vessels. As more patients with
further prospective clinical study where all patients who underwent multivessel disease are treated greater attention will also need to be
IVUS-guided LMS PCI and stent post-dilation beyond suggested expansion placed on longer-term outcomes in more demanding clinical settings.
limits were entered into a registry. In 31 patients, mean maximal stent The risks of latent stent fracture may assume a more prominent role
diameters of >5.0  mm were reliably achieved (using Biomatrix Flex™ in clinical studies in future. Ultimately, as the clinical practice of PCI
9 crown and Promus Element Large vessel platforms) with 5.5  mm continues to evolve, manufacturers and clinicians will have to work
and 6.0  mm post-dilation balloons. No intra-procedural complications closely in partnership to make sure that the devices that are implanted
occurred and in 13.4 months of follow-up only one patient experienced can provide excellent safety and long-term efficacy for patients. The
clinical ISR.29 This indicates excellent short-term efficacy and no increased importance of stent design has been re-emphasised and is likely to
complication rate with over-expansion of current generation DES in the become increasingly relevant in future, where the patient and lesion
LMS. These results compliment the bench data of Foin et al., but in the being treated are likely to mandate very careful selection of the stents
clinical arena.24 It should be noted that under-expansion or incomplete that are deployed in each individual setting. The focus should be
stent apposition of BMS or DES is strongly associated with ISR and stent shifted away from producing ever more deliverable stent platforms
thrombosis.30–32 We would suggest that leaving undersized and unapposed and should be moved back to the fundamental properties of what the
or malapposed stents in the LMS should be avoided at all costs. device has been built to achieve. n

1. Rankin JM, Spinelli JJ, Carere RG, et al., Improved clinical 13. Finet G, Rioufol G, Coronary stent longitudinal deformation implantation, Circ Cardiovasc Interv, 2012;5:663–71.
outcome after widespread use of coronary-artery stenting in by compression: is this a new global stent failure, a 24. Foin N, Sen S, Allegria E, et al., Maximal expansion capacity
Canada, N Engl J Med, 1999;341:1957–65. specific failure of a particular stent design or simply an with current DES platforms: a critical factor for stent
2. Colombo A, Stankovic G, Moses JW, Selection of coronary angiographic detection of an exceptional PCI complication?, selection in the treatment of left main bifurcations?,
stents, J Am Coll Cardiol, 2002;40:1021–33. EuroIntervention, 2012;8:177–81. EuroIntervention, 2013;8:1315–25.
3. Cutlip DE, Chauhan MS, Baim DS, et al., Clinical restenosis 14. Shand J, Ramsewak A, Spence M, et al., The ‘concertina 25. Weimer M, Butz T, Schmidt W, et al., Scanning electron
after coronary stenting: perspectives from multicenter clinical effect’ and the limitations of current drug-eluting stents: is it microscopic analysis of different drug eluting stents after
trials, J Am Coll Cardiol, 2002;40:2082–9. time to revisit and prioritize stent design over efficacy?, Interv failed implantation: from nearly undamaged to major
4. Kastrati A, Mehilli J, Dirschinger J, et al., Intracoronary stenting Cardiol, 2012;4:325–35. damaged polymers, Catheter Cardiovasc Interv, 2010;75:905–11.
and angiographic results: strut thickness effect on restenosis 15. Onuma Y, Serruys PW, Bioresorbable scaffold: the advent 26. Basalus MW, Tandjung K, van Westen T, et al., Scanning
outcome (ISAR-STEREO) trial, Circulation, 2001;103;2816–21. of a new era in percutaneous coronary and peripheral electron microscopic assessment of coating irregularities and
5. Briguori C, Sarais C, Pagnotta P, et al., In-stent restenosis in revascularization?, Circulation, 2011;123:779–97. their precursors in unexpanded durable polymer-based drug-
small coronary arteries: impact of strut thickness, J Am Coll 16. Hanratty CG, Walsh SJ, Longitudinal compression: a “new” eluting stents, Catheter Cardiovasc Interv, 2012;79:644–53.
Cardiol, 2002;40:403–9. complication with modern coronary stent platforms -- time to 27. Basalus MW, Tandjung K, VAN Apeldoorn AA, et al., Effect of
6. Chen MS, John JM, Chew DP, et al., Bare metal stent think beyond deliverability?, EuroIntervention, 2011;7:872–7. oversized partial postdilatation on coatings of contemporary
restenosis is not a benign clinical entity, Am Heart J, 17. Pitney M, Pitney K, Jepson N, et al., Major stent deformation durable polymer-based drug-eluting stents: a scanning
2006;151:1260–4. / pseudofracture of 7 Crown Endeavor/Micro Driver stent electron microscopy study, J Interv Cardiol, 2011;24:149–61.
7. Doyle B, Rihal CS, O’Sullivan CJ, et al., Outcomes of stent platform: incidence and causative factors, EuroIntervention, 28. Aziz S, Morris JL, Perry RA, Stables RH, Stent expansion: a
thrombosis and restenosis during extended follow-up of 2011;7:256–62. combination of delivery balloon underexpansion and acute
patients treated with bare-metal coronary stents, Circulation, 18. Prabhu S, Schikorr T, Mahmoud T, et al., Engineering assessment stent recoil reduces predicted stent diameter irrespective of
2007;116:2391–8. of the longitudinal compression behaviour of contemporary reference vessel size, Heart, 2007;93:1562–6.
8. Serruys PW, Unger F, Sousa JE, et al., Comparison of coronary- coronary stents, EuroIntervention, 2012;8:275–81. 29. Shand JA, Sharma D, Hanratty C, et al., A prospective
artery bypass surgery and stenting for the treatment of 19. Ormiston JA, Webber B, Webster MW, Stent longitudinal intravascular ultrasound investigation of the necessity for and
multivessel disease, N Engl J Med , 2001;344:1117–24. integrity bench insights into a clinical problem, JACC efficacy of postdilation beyond nominal diameter of 3 current
9. The SoS Investigators, Coronary artery bypass surgery versus Cardiovasc Interv, 2011;4:1310–7. generation DES platforms for the percutaneous treatment of
percutaneous coronary intervention with stent implantation 20. Park MW, Chang K, Her SH, et al., Incidence and clinical the left main coronary artery, Catheter Cardiovasc Interv, 2013
in patients with multi-vessel coronary artery disease (the impact of fracture of drug-eluting stents widely used in [Epub ahead of print].
Stent or Surgery trial): a randomised controlled trial, Lancet, current clinical practice: comparison with initial platform of 30. Hassan AK, Bergheanu SC, Stijnen T, et al., Late stent
2002;360:965–70. sirolimus-eluting stent, J Cardiol, 2012;60:215–21. malapposition risk is higher after drug-eluting stent
10. Morice MC, Serruys PW, Sousa JE, et al., A randomized 21. Kuramitsu S, Iwabuchi M, Yokoi H, et al., Incidence compared with bare-metal stent implantation and associates
comparison of a sirolimus-eluting stent with a standard stent and clinical impact of stent fracture after the Nobori with late stent thrombosis, Eur Heart J, 2010;31:1172–80.
for coronary revascularization, N Engl J Med, 2002;346:1773–80. biolimus-eluting stent implantation, J Am Heart Assoc, 31. Fujii K, Carlier SG, Mintz GS, et al., Stent underexpansion and
11. Serruys PW, Morice MC, Kappetein AP, et al., Percutaneous 2014;3(2):e000703. residual reference segment stenosis are related to stent
coronary intervention versus coronary-artery bypass grafting for 22. Inaba S, Mintz GS, Yun KH, et al., Mechanical complications of thrombosis after sirolimus-eluting stent implantation: an
severe coronary artery disease, N Engl J Med, 2009;360:961–72. everolimus-eluting stents associated with adverse events: an intravascular ultrasound study, J Am Coll Cardiol, 2005;45:995–8.
12. Serruys PW, Silber S, Garg S, et al., Comparison of intravascular ultrasound study, EuroIntervention, 2014;9:1301–8. 32. Cook S, Wenaweser P, Togni M, et al., Incomplete Stent
zotarolimus-eluting and everolimus-eluting coronary stents, 23. Kuramitsu S, Iwabuchi M, Haraguchi T, et al., Incidence and Apposition and Very Late Stent Thrombosis After Drug-Eluting
N Engl J Med, 2010;7:363:136–46. clinical impact of stent fracture after everolimus-eluting stent Stent Implantation, Circulation, 2007;115:2426–34.

INTERVENTIONAL CARDIOLOGY REVIEW 93

walsh_aiza_edited.indd 93 12/05/2014 01:10

You might also like