Professional Documents
Culture Documents
MET ALB 2 06533638B EN
MET ALB 2 06533638B EN
2008-06
Microalbumin_2 (µALB_2)
System New Information
ADVIA® 1200 Updates to Reagents, Reagent Preparation and Use, Calibration
Frequency, Quality Control, Analytical Range, System Correlation
ADVIA 1650/1800 Updates to Reagents, Reagent Preparation and Use, Calibration
Frequency, Quality Control, Analytical Range, System Correlation
ADVIA 2400 Updates to Reagents, Reagent Preparation and Use, Calibration
Frequency, Quality Control, Analytical Range, System Correlation
NOTE: For an explanation of the symbols in this document and associated products,
refer to Understanding the Symbols located on the introductory page of the Methods
Directory.
Method Summary
Item Description
Method Principle PEG-enhanced immunoturbidimetric
Specimen Type Human urine
On-board Stability Without
Reagent With Reagent
ADVIA Chemistry Container Container
Systems Insert Insert
ADVIA 1200 60 days 60 days
ADVIA 1650/1800 60 days 60 days
ADVIA 2400 60 days 60 days
Reagent Storage Temperature 2–8°C
Calibration Frequency Without
Reagent With Reagent
ADVIA Chemistry Container Container
Systems Insert Insert
ADVIA 1200 20 days 30 days
ADVIA 1650/1800 60 days 60 days
ADVIA 2400 20 days 60 days
Reagent Blank (RBL) Frequency At time of method calibration
Reaction Type Endpoint (EPA)
Measurement Wavelength 340 nm
Standardization Internal standard
Item Description
Analytical Range Urine Specimen*
0.3–(38.0–42.0) mg/dL
(3–[380–420]) mg/L
*The albumin (ALB) concentration in the
Microalbumin_2 Calibrators, Level 5, varies from
38.0–42.0 mg/dL
(380–420 mg/L).
Expected Values < 30 mg/day
Intended Use
For in vitro diagnostic use in the quantitative determination of albumin in human urine
on the ADVIA Chemistry systems. Such measurements are used in the diagnosis and
treatment of microalbuminuria and are helpful for the detection and treatment of
patients at risk from renal disease.
Reagents
The reagents are packaged as listed below. Components of the package are available
as a kit only.
REF
Container Size Symbol Contents Amount No. of Tests
03051194 Microalbumin 4 x 105
Reagents (ADVIA
1650/1800)
20-mL Reagent 1 4 x 20 mL 4 x 130
(ADVIA
1200/2400)
20-mL Reagent 2 4 x 4.3 mL
CAUTION! This device contains material of animal origin and should be handled as a
potential carrier and transmitter of disease.
NOTE: Sodium azide can react with copper and lead plumbing to form explosive metal
azides. If disposal into a drain is in compliance with federal, state, and local
requirements, flush reagents with a large amount of water to prevent the buildup of
azides.
For all systems, unopened reagents are stable until the expiration date printed on the
product label when stored at 2–8°C. Do not freeze reagents.
For additional details, refer to the Methods Introduction section of the system-specific
Operator’s Guide.
Sample Handling
We recommend using urine for this method. Collect the urine as a 24-hour, timed, or
random midstream sample (spot collection) in a clean, unused glass or plastic
collection container. Preservatives are not recommended for this method.
Specimen Storage 3
For timed or 24-hour samples, if possible, use fresh specimens. You can also use
samples that have been stored up to 2 weeks at 4°C or up to 5 months at -70°C. Avoid
repeated freeze-thaw cycles.
For additional details, refer to Sample Collection and Preparation in the Methods
Introduction section of the system-specific Operator’s Guide.
For instructions on how to load reagents and run samples, refer to the Daily Operations
section of the system-specific Operator’s Guide.
NOTE: Mix stored specimens adequately and centrifuge prior to testing.
Calibration
Calibrate this method using ADVIA Chemistry Microalbumin_2 Calibrators
(REF 06487733) as follows:
1. Refer to the package insert supplied with the calibrator for handling instructions
and values.
2. Place 1 sample container filled with distilled water as the blank sample onto the
sample tray (SST) cup position assigned to the blank for multipoint calibration.
3. At the Setup menu, select the Con/Cal Sample Setup window.
4. Enter the expected sample container type and assigned SST cup position.
5. If no sample type and cup position are defined, you must do so before running the
calibration.
6. Place each of the 5 levels of the Microalbumin Calibrator onto the SST cup position
assigned for each level.
For setup and use instructions, refer to the Calibration Overview section of the
system-specific Operator’s Guide.
Calibration Frequency
Perform a calibration when this method is implemented on the system. You must
recalibrate after the following events:
• when the reagent lot number changes
• after replacing critical optical or hydraulic components
• when indicated by quality control procedures
Siemens has validated the calibration stability for this method as shown in the following
table:
Siemens recommends calibrating new reagent packs if the previous reagent pack was
calibrated any time during its on-board stability, other than as a fresh pack.
Individual laboratory quality control programs and procedures may require more
frequent calibration.
Quality Control
Siemens recommends the use of quality control material from Bio-Rad Laboratories
with at least 2 levels (low and high). A satisfactory level of performance is achieved
when the analyte values obtained are within the Acceptable Control Range for the
system or within your range, as determined by an appropriate internal laboratory
quality control scheme.
For more information, refer to the Quality Control Overview section of the
system-specific Operator’s Guide.
As with any chemical reaction, you must be alert to the possible effect on results of
unknown interferences from medications or endogenous substances. The laboratory
and physician must evaluate all patient results in light of the total clinical status of the
patient.
Interference
Siemens tested the following potential interferents with a urine pool of approximately
3 mg/dL albumin concentration and found the results shown below:
µALB_2 Sample
Interferent Interferent Level Concentration Interference*
µALB_2 Sample
Interferent Interferent Level Concentration Interference*
Performance Characteristics
Precision 5
Each sample was assayed 2 times per run, 2 runs per day, for at least 10 days.
Precision estimates were computed according to CLSI document EP05-A2, Evaluation
of Precision Performance of Quantitative Measurement Methods; Approved Guideline.5
ADVIA 1650/1800
Within-Run Total
Specimen Type Level SD CV (%) SD CV (%)
Common Units (mg/dL)
Pool 1 2.9 0.06 2.1 0.10 3.3
Pool 2 29.3 0.34 1.2 0.49 1.7
Control 1 1.9 0.08 4.1 0.12 6.1
Control 2 9.1 0.16 1.8 0.23 2.5
SI Units (mg/L)
Pool 1 29 0.6 2.1 1.0 3.3
Pool 2 293 3.4 1.2 4.9 1.7
Control 1 19 0.8 4.1 1.2 6.1
Control 2 91 1.6 1.8 2.3 2.5
ADVIA 2400
Within-Run Total
Specimen Type Level SD CV (%) SD CV (%)
Common Units (mg/dL)
Pool 1 2.8 0.05 1.6 0.08 2.9
Pool 2 28.8 0.24 0.8 0.28 1.0
Control 1 1.7 0.04 2.1 0.07 3.8
Control 2 8.9 0.08 0.9 0.10 1.1
SI Units (mg/L)
Pool 1 28 0.5 1.6 0.8 2.9
Pool 2 288 2.4 0.8 2.8 1.0
Control 1 17 0.4 2.1 0.7 3.8
Control 2 89 0.8 0.9 1.0 1.1
Analytical Range
This method measures the ALB concentration in urine ranging from the minimum
detectable concentration (MDC) to the ALB concentration in the highest level of the
calibrator according to the table shown below. MDC is an estimation based on two
times the within-run standard deviation of the zero calibrator.
Siemens has validated an automatic rerun condition for this method that extends the
reportable ranges for urine up to 380 mg/dL (3800 mg/L) for all ADVIA Chemistry
systems.
Prozone Effect
A prozone effect was not demonstrated for analyte concentrations up to at least
20,000 mg/dL (200,000 mg/L) for all ADVIA Chemistry systems.
Expected Values 6
The expected value for this method is < 30 mg/day for urine.
We provide this information for reference. Each laboratory should establish its own
normal range. You can enter normal range values and abnormal range values at the
Analytical Parameters (Chemistry) window.
System Correlation
The performance of the applicable method (y) was compared with the performance of
the same method on the comparison system (x).
ADVIA 1200
Specimen Type Comparison System (x) N Regression Equation Sy.x r Sample Range
Urine ADVIA 1650/1800 95 y = 0.97x - 0.1 0.4 0.996 0.6–16.7 mg/dL
µALB_2 y = 0.97x - 0.7 4.4 0.996 6–167 mg/L
Urine ADVIA 1200 µALB 96 y = 0.99x - 0.1 0.2 0.999 0.8–16.9 mg/dL
y = 0.99x - 1.0 2.2 0.999 8–169 mg/L
ADVIA 1650/1800
Specimen Type Comparison System (x) N Regression Equation Sy.x r Sample Range
Urine ADVIA 1650/1800 98 y = 0.99x - 0.1 0.1 0.999 0.6–16.8 mg/dL
µALB y = 0.99x - 0.7 1.2 0.999 6–168 mg/L
ADVIA 2400
Specimen Type Comparison System (x) N Regression Equation Sy.x r Sample Range
Urine ADVIA 1650/1800 95 y = 0.97x - 0.0 0.4 0.997 0.6–16.7 mg/dL
µALB_2 y = 0.97x - 0.0 3.5 0.997 6–167 mg/L
Urine ADVIA 2400 µALB 97 y = 1.00x - 0.2 0.6 0.991 0.5–17.1 mg/dL
y = 1.00x - 2.4 6.4 0.991 5–171 mg/L
Standardization
The ADVIA µALB_2 method is traceable to an internal standard manufactured using
highly purified material. Currently, no reference standard for this method exists.
Assigned values of ADVIA Chemistry Microalbumin_2 Calibrators are traceable to this
standardization.
References
1. Fielding BA, Price DA, Houlton CA. Enzyme immunoassay for urinary albumin. Clin
Chem.1983;29:355-357.
2. Hellsing K. Influenced polymers on the antigen-antibody reaction in a continuous
flow system. In: Automated Immunoprecipitin Reactions. Colloquium on AIP.
Tarrytown, NY: Technicon Inst. Corp.; 1972:17, 798-9.
3. Burtis CA, Ashwood ER, Bruns DE. Tietz Textbook of Clinical Chemistry and
Molecular Diagnostics, 4th edition. St. Louis, MO.: Elsevier Saunders;
2006:886-888.
4. Young DS. Effects of Drugs on Clinical Laboratory Tests. 5th ed. Washington:
AACC Press (2000).
5. Clinical and Laboratory Standards Institute (formerly NCCLS). Evaluation of
Precision Performance of Quantitative Measurement Methods; Approved Guideline
- Second Edition. Wayne, PA: Clinical and Laboratory Standards Institute; 2004.
NCCLS Document EP05-A2.
6. Wu AHB. Tietz Clinical Guide to Laboratory Tests, 4th edition, Saunders Elsevier,
St. Louis, MO: 2006:316.
Technical Assistance
For customer support, please contact your local technical support provider or
distributor.
Trademarks
ADVIA is a trademark of Siemens Medical Solutions Diagnostics.