YDT-IDT-2 Operation Manual PDF
YDT-IDT-2 Operation Manual PDF
YDT-IDT-2
Operation Manual
Hunan Taiyanglong Medical Technology Co., Ltd.
1. Introduction
1.1 Use of this operation manual
This operation manual provides important information on how to use this product to ensure
safe operation. Besides observe all safety instructions and operating instructions given in this
manual, user should also comply with location accident prevention regulations and general safety
regulations in relevant to this product.
User should assure all the staffs who operate the pendant read and fully understand this
manual. The manual is an integral part of the product, it should be placed next to the pendant
during the work for operator review it any time. If this equipment is transferred to third parties, in
the time of transfer, be sure to keep this manual be transferred to a third party.
In order to facilitate your understanding of this equipment, this manual provides some
photos and diagrams, but they are not in very accurate proportions, these may cause slight
deviation from the actual product. This manual applies to YDT-IDT-2 products. There may be
differences in some images and functions, specific please in kind prevail.
2. Safety
In order to prove best protection for the operator and make sure this device can be used
safely and smoothly, this section gives safety instructions which should be specially paid attention
during operation. Failure to follow the safety measures and safety instructions which given in this
manual will result in serious harm.
WARNING!
z Explosion hazard˖The equipment is not suitable for operation in areas with explosion hazard.
z Non-observance may cause personal injury or devices damage. The rack installation must be
carried out by persons with the appropriate qualifications, the expansion bolt must be
tightened and inspect its fixed.
z Electric shock ˖ Electric part repair or maintain must be carried out by skilled technicians.
Non-professional or before full understand the technical performance of this equipment, any
repair or maintenance is prohibited.
z Personal injury˖The installation, connection or adjustment of gas pipe and outlets must only
be operated by professional person. Any accident because by wrong connection of gas hoses
are forbidden.
z Personal injury˖Any cleaning for surface of power sockets are forbidden before power off.
z Personal injury ˖ The arm and service head adjustment must be carried out by skilled
technicians
z Personal injury˖The pendant remove work must be carried out by professional technicians or
persons with appropriate qualifications.
1
Hunan Taiyanglong Medical Technology Co., Ltd.
z Personal injury or devices damage˖The pendant is used for supporting or positioning various
devicesǃequipmentsǃIV poles etc, so the maximal load must not be exceeded. Otherwise it
may cause falling off of the pendant which cause personal injury or equipment damage
z The pendant is a heavy weight suspension device. Although it is easy to move, but the mobile
arm and the terminal box will produce a very strong potential energy, this may bring a
collision cause damage to the equipment or injury to personnel.
z Inadvertently larger force may make the terminal box, infusion pumps move, so need to make
sure provide long enough infusion tubes and power cables to fit for unintentional movement.
Otherwise it may cause patient injury or equipment damage.
z ATTENTIONS !
z This equipment is only applicable for single-phase AC voltage 220V+/-10% ǃ 50Hz. Any
replaces in input voltage of the device will cause serious damage. ˄ Except another design
which the user required˅
z Leak oxygen cause fire. When working with oxygen ,please do not use open flame. Do not
smoking! And operations with burning objects or open flame are not permitted.
z Do not pull and plug the pin from the power sockets directly. Please make sure you have held
on the pendant or press the power sockets, then pull and plug the pin.
z Be care for devices damage˖when the extension arm or service head rotating or move, please
be careful and don’t collide with the other equipment.
z When rotating the extension arm and service head, please keep it away from other staffs.
Please do not use too much force to avoid personnel injure, limit device damage or hitting to
the wall .
z To avoid damage of surface, do not use any scouring agents and strong alkaline, strong acidic
and alcoholic-containing to clean pendants. Please use neutral scouring agent and soft cloth.
And make sure there is no fluid drop into the pendant.
z No authorization from Hunan Taiyanglong Medical Technology Co., Ltd. Any modification
to the pendant is not permitted. Modifications without authorization will not be counted into
product warranty.
Current
Danger!
Current creates the risk of death!
Once exposed to the live conductors will cause an
immediate life-threatening danger. Damaged
electrical insulators or individual parts may pose a
threat to life. Therefore˖
2
Hunan Taiyanglong Medical Technology Co., Ltd.
Hot surface(on the lighting system). Hot surface can’t see by eye. Please
wait it for cool down before touching these surfaces.
3 Product introduction
ICU motorized pendant is a modern and practical medical subsidiary equipment in modern
operation room. It provides loading platform for medical equipment and interface of medical gas,
vacuum suction and power supply for operation room of hospital.
3
Hunan Taiyanglong Medical Technology Co., Ltd.
with the electric safety requirement: GB9706.1-2007. The gas outletsǃnetwork interfaceǃ
information socketǃvideophoneǃbackground musicǃteleconsultation and other data port are
optional as customer’s request.
3.2 Application
ICU motorized pendant is used for hospital ICU. It provides interface for medical gas,
vacuum suction, anesthetic gas scavenging, electricity, network and platforms for medical
equipment like monitors and high frequency electrotomes.
The pendant can work continuously, but can not load over the maximum weight.
3.4 Structure
6 Control handle
5 Terminal Box
4 IV pole
3 Tube
2 Single arm
1 Ceiling cover
4
Hunan Taiyanglong Medical Technology Co., Ltd.
5
Hunan Taiyanglong Medical Technology Co., Ltd.
ATTENTIONSφ
Ƹ
! This equipment only for single-phase AC voltage 220V+10% ǃ 50Hz, any change in
input voltage of the device will cause serious damage.˄Except another design which the user
required˅
1˅The pendant power supply˖power grid; rated voltage: 220V+10%, rated frequency: 50Hzf1Hz
or rated voltage: 380V+10%, rated frequency: 50Hzf1Hz to customer’s requirement.
2˅Users must provide a separate phase of the loop lineǃcenter line and RCDs to match the input
power.
3˅Must provide the special grounding terminal line. Water hose or heating pipe are not permitted
to use as protection of grounding devices
WARNINGφ
!
Ƹ Personal injury or devices damage: The rack installation must only be carried out by
persons with appropriate qualification, the expansion bolt must be tightened and inspect its
fixed.
6
Hunan Taiyanglong Medical Technology Co., Ltd.
floor slab and ceiling is qualified for installation and Concrete strength requirements C30.
4.1.2 installation
1˅According the layout design, find the center points for the pendant installation on the false
ceiling. Cut a round hole as 400mm around the center point.
2˅Fix the pre-built brackets center to the ceiling hole center. Use the expansion screw to fix the
installation rack on the concrete floor slab.
3˅Make sure the bracket bottom is parallel with the ceiling and with a distance of 200mm to the
ceiling. Then tighten the expansion screw, strengthen the bracket with steel wire rope. (refer to
Figure 1 Bracket Installation Diagram˅
º
7
Hunan Taiyanglong Medical Technology Co., Ltd.
3 Connection board
2 Screw nut
1 connection screw
No. Item name
8
Hunan Taiyanglong Medical Technology Co., Ltd.
6 Upper tube
5 Pneumatic brake
2 Connection board
6 Screw
5 Spring washer
4 Flat washer
3 Single arm
1 Connection board
9
Hunan Taiyanglong Medical Technology Co., Ltd.
7 Terminal box
6 Dropping Tube
3 Single arm
1 Connection board
1 terminal box
10
Hunan Taiyanglong Medical Technology Co., Ltd.
2 Ceiling cover
1 Connection board
5 Single arm
11
Hunan Taiyanglong Medical Technology Co., Ltd.
4 Decoration strip
1 Ceiling cover
WARNING!
! Personal injury˖Please note that the installations, electrical connections or the gas outlets
Ƹ
adjustments must only be carried out by persons with skilled installation knowledge.
Caution ˖Medical gas hoses are differentiated by international standard color: Oxygen
by white, VAC by yellow, Air by black, Nitrous Oxide by green, Nitrogen by blue, Carbon
dioxide by grey, AGSS by red or with transparent grid tube.
1˅Before installation, check the marks on the gas hoses and gas pipelines and then match and
connect them together. Clamp them to prevent leakage.
2˅ Connect the electric lines to the line bank by fixing the terminals to the criterion
Caution: Disconnecting the electric mains before install the power system.
Ref.: circuit diagram
3˅Connect the grounding lines which connected to earth bus bar to the equipment and make sure
no looseness.
4˅Connect the interface which connect the power lines, control lines of the pendant to the terminal
rows and drives.
12
Hunan Taiyanglong Medical Technology Co., Ltd.
ATTENTIONS˖
! Leak oxygen cause fire. When working with oxygen, please avoid any flames. No
Ƹ
smoking! Touch to burning objects or fire are forbidden.
Taiyanglong pendant medical gas outlets use special design of push-pull type protection shield
cover.When use push the shield cover up, plug the fast connection or other matched adaptor(like
humidifier) into the gas outlets. Double check the tightness by pull and check if it is well fixed or
if any leakage. Then regulate the gas flow to requirement. When take off the connection, hold the
the connection by one hand first, and clip the retaining key by other hand before take it off, and
then pull down the shield cover to protect the outlets.(refer to Figure 20)
13
Hunan Taiyanglong Medical Technology Co., Ltd.
Figure 20
ATTENTIONS˖
! Do not pull off the plug from the electric sockets directly, use one hand to hold the
Ƹ
electric sockets and disconnect the plug by the other hand.
All the power sockets are in build-in concealed design and with protective door. Please open
protective door when use and then plug the related instruments or devices in related power sockets.
Tidy up the cables by cable clips and close the protective door. Switch on the grid power switch, if
the instruction light is on, it means the power sockets are ready for work.
Caution: Input power of the instruments or devices must be consistent with input power of
this product. Current for every single outlet must not exceed 10A. Otherwise, the pendant
system will be damaged.
14
Hunan Taiyanglong Medical Technology Co., Ltd.
ATTENTIONφ
Ƹ
! Be ware of damage devices˖ The moving, rotating range is limited. To avoid
The clinical management Section, monitor and ventilator section and infusion pumps can be
levelly rotated in a limited range. The rotation and moving is operated by handle control.
ATTENTIONφ
Ƹ
! Personal injury˖Cleaning pendant main socket are forbidding without power off.
Ƹ
! Personal injury or damage devices˖The pendant is used for loading various devicesǃ
equipmentsǃIV poles etc, so the maximal load must not be exceeded otherwise will cause fall down
problems.
CAUTION:
Ƹ
! Translation or rotation of terminal box of wet side and suspension equipment shelf of
dry side should avoid other staff. Do not push too hard to prevent injuries and limit device
rupture and cause equipment damage.
Ƹ
! To avoid damage to pendant painting surface, do not use alcohol cleaning agents,
solvents and do not use strong acid or alkali cleaners, apply a neutral soap or detergent solution
and a soft cloth to clean the drawbridge, and ensure there is no cleaning solution into the pendant.
1) Make sure the electric lines are powered off and gases are shut off after every day’s work.
Check the power indicator light is off and adaptors of gas outlet are pulled out.
2) The loading capacity of platform is limited, supporting various devices in accordance with their
specifications. Please refer to the equipment technical data. The maximal load must not be
exceeded !
3) When operate translation or rotation of terminal box of wet side and suspension equipment
shelf of dry side, do not push too hard to prevent the positioning pin broken cause limit device
15
Hunan Taiyanglong Medical Technology Co., Ltd.
rupture .
4) The electric sockets should always keep dry. When cleaning or disinfection are needed, please
use the soft dry cloth and make sure that no cleaning fluid enters the sockets.
Please adjust the pressure according to requirement before use the medical gas˄pressure
regulating valve in the hospital supply system). Over pressure or leak gas adaptors are forbidden.
Check gas outlets from time to time to see retaining key is flexible and reliable. Please promptly
cleaned when it is not in use. When cleaning and disinfection, please avoid any water to avoid
impact airtight performance.
6.1 Maintenance
The pendant designed service life is 10 years, and life for gas hoses is 5years. The normal
maintenance period of pendant system is twelve months. When time is due please ask professional
maintenance person open the ceiling cover, terminal box cover, joint parts blanking caps for
inspecting as below:
1) Inspect all gas outlets with gas hoses, whether any joint loose, leakage, the airtight performance
of gas outlet meet to the requirement. If any loose or leakage please use the locking clamp to repair
the leakage, or replace connector and gas outlets. Check if the gas hose are aging with cracks. If
any hose is aging with cracks, kindly replace with new hoses.
2) Inspect any loose for all power sockets or connection terminals and if the spring pieces loses its
elasticity or spark erosion caused by poor contact. If this is the case then tighten the screws or
replace the socket.
3) Inspect the grounding line is solid and reliable: If any loose, fasten it right away.
4) Inspect the medical gas pipe and power wire in rotating place, if there is any abrading, please
replace the gas pipe and power wire and take additional precautionary measures to deal with the
abrading, such as a case or sleeve.
5) Inspect the expansion bolts are tightly fixed into the concrete floor slab. If any loose, tighten it
right away to avoid falling down. And pay attention to check the tightness of the rope and seed if it
is appropriate to play a role of protection.
6) Inspect the positioning pin of each joints and positioning block of bridge pulley and see if any
abrading or loose. Please replace them if any limit function problem found.
Special Note: The bracket part inspection should be focused. If there is any rust, welding
collapse, faulty soldering and other hazards should be immediately removed to prevent
serious accidents and personal injury!
WARNINGφ
16
Hunan Taiyanglong Medical Technology Co., Ltd.
Ƹ
! Personal injury˖The remove of pendant must only be carried out by skilled
Any problems cause during use the pendant. please contact your supplier or my company promptly.
Only skilled technician is permitted to repair the pendant system (except the replace of fuse)ˈ
otherwise, our company do not take any responsibility to the devices damage or personal injury
caused by non-professional persons.
We offer one year of free maintenance for the pendant from date of shipment and provide lifelong
paid maintenance and accessories.
17
Hunan Taiyanglong Medical Technology Co., Ltd.
7.1 Mark
7.1.1 Label
This pendant uses stick-on label, its production specifications comply to GB / T13306-91.
The Logo contents˖
1˅The manufacturer nameǃaddress and contacts˗
2˅Product name, model No, series No. and production date˗
3˅Rated voltage, power and frequency˗
18
Hunan Taiyanglong Medical Technology Co., Ltd.
Medical gas pipeline marks: This pendant medical gas pipeline color differentiate by different
color in accordance with international standards, and use word mark "clear and recognizable".
7.2 Packingφ
The pendant packed with high-quality bubble film first then put into cartons, and then put
into the wooden cases. Fix with bolts. For empty spaces, fill and press with foam and square wood
strip to prevent any loose or worn .
Operating manual, quality certification and packing list are attached in the packing case.
19
Hunan Taiyanglong Medical Technology Co., Ltd.
8.1.3 The product standard the pendant performs: YZB /XIANG 0067-2010
Thank you choose Taiyanglongÿs products. If you have any suggestions and comments when
you use the pendant, including the update of the product ˄ increase the number of platform ǃ
networkǃphone or video interfaceǃVGA interface˅or other technical requirement. Please be free
to contact us:
E-mail address˖hnlytyl@163.com
20