Professional Documents
Culture Documents
auto-B-D
auto liquicolor
System Reagent for HumaStar 600 (HS600SR)
C
1 Introduction .................................................................................................................................................................. 2
2 Imprecision .................................................................................................................................................................... 2
3 Linearity and Detection Limit ...................................................................................................................................... 2
3.1 Linearity ....................................................................................................................................................................................... 2
3.2 Detection Limit ......................................................................................................................................................................... 3
4 Comparison of methods ............................................................................................................................................... 3
5 Stability, Accelerated Stress ........................................................................................................................................ 5
5.1 Linearity ....................................................................................................................................................................................... 5
5.2 Recovery of control sera ........................................................................................................................................................ 6
6 Real Time Stability ........................................................................................................................................................ 6
7 On Board stability / Calibration stability .................................................................................................................... 6
7.1 Recovery of control sera ........................................................................................................................................................ 6
7.2 Calibration stability ................................................................................................................................................................. 7
8 Interferences ................................................................................................................................................................. 7
9 Traceability .................................................................................................................................................................... 7
Form: 4.2-01.06-04
Rev. 007 | valid of 26.11.2013 1/7
1 Introduction
The performance characteristics of the auto-BILIRUBIN-D liquicolor HS600SR have been tested and documented in
order to verify the clinical usefulness and compliance with the essential requirements of directive 98/79/EC.
2 Imprecision
The imprecision (within-run and day-to-day) of the auto-BILIRUBIN-D liquicolor HS600SR was calculated from six
determinations on five consecutive days using reagent lot # H022 (expiry 2006-12). Low, medium and high level
control sera (lot # CE 127 A, CE 127 B and CE 127 C) were employed as sample material. Acceptance criteria were a
max. CV of 4% (intra-assay) and 5% (inter-assay).
3.1 Linearity
The linearity of the auto-BILIRUBIN-D liquicolor HS600SR, lot 0042/0001A was controlled by employing a high
concentrated pool serum successively diluted with phys. saline. The analysed concentrations were compared with
the theoretical concentrations obtained from a linear regression. Deviations of measured against theoretical
concentrations are expressed in absolute and relative values. Linearity was assumed within an accepted deviation
of ± 10% starting at high pool content of 10%.
High Pool Analytical Data Regressed Data Deviation from Regression Line
Content (%) mg/dl mg/dl mg/dl (%)
0 0.1 0.5 -0.40
10 1.3 1.5 -0.14 -9.6
20 2.5 2.4 0.07 2.9
30 3.5 3.4 0.14 4.1
40 4.5 4.3 0.15 3.5
50 5.6 5.3 0.29 5.5
60 6.4 6.3 0.09 1.4
70 7.3 7.2 0.12 1.7
80 8.2 8.2 -0.02 -0.2
90 9.0 9.1 -0.10 -1.1
100 9.8 10.1 -0.29 -2.9
4 Comparison of methods
The auto-BILIRUBIN-D liquicolor HS600SR lot 0013 run on HumaStar 600 has been compared against the BILIRUBIN-
D reagent (Beckman) lot 9967 method measured on Olympus AU 400. Control sera have been employed in the
comparison (N=74). The results have been evaluated by a non-parametric regression model according to Bablok &
Passing, Acceptance criteria were r > 0.95 and a slope between 0.95 and 1.05 and could be described as follows:
Sample-
Sample- Mean Reference Mean Test
no. Beckman Olympus AU 400 HumaStar 600 SR
1 0.59 0.64
2 0.66 0.61
3 0.39 0.30
4 0.08 0.19
5 0.72 0.71
6 0.13 0.18
7 0.94 0.82
8 0.23 0.30
9 0.10 0.16
10 0.67 0.88
11 1.16 1.19
12 0.06 0.11
13 0.14 0.16
14 0.60 0.50
15 0.08 0.15
16 0.63 0.73
17 0.08 0.11
18 0.11 0.16
19 0.92 0.84
20 0.15 0.18
21 1.02 1.07
22 0.09 0.12
23 0.72 0.79
24 0.20 0.25
25 2.23 2.08
26 0.45 0.51
27 0.06 0.26
28 0.07 0.22
29 0.46 0.43
30 0.10 0.14
31 0.17 0.22
32 1.06 1.14
33 0.34 0.30
34 1.89 2.00
35 1.79 1.82
36 0.55 0.67
37 0.60 0.71
38 0.67 0.70
39 1.08 1.21
40 0.52 0.50
41 0.55 0.51
42 1.11 1.18
43 0.51 0.56
44 0.66 0.70
45 0.52 0.54
46 3.53 3.46
47 3.76 3.89
48 3.60 3.55
49 4.85 4.79
50 3.16 2.22
51 3.50 3.66
52 2.05 2.06
53 3.58 3.76
54 3.51 3.53
55 3.61 3.43
56 5.64 5.57
57 5.17 4.65
58 3.08 3.15
59 1.39 1.43
60 1.65 1.76
61 1.59 1.47
62 3.56 3.47
Results from a non-parametric regression analysis acc. to Passing & Bablok are shown in the graph below.
Conclusion: Both methods showed a good agreement and no significant deviation could be observed with
any specific sample.
5.1 Linearity
The linearity of the auto-BILIRUBIN-D liquicolor HS600SR was controlled according to the procedure already
described in 3.1. The reagents were stressed in original HumaStar 600 plastic bottles 14 days at 37°C, lot
0042/0001, expiry date 2013-05.
Conclusion: The test is specified to cover a linear range up to 10 mg/dl as indicated in section 3.1. The above
results clearly confirm that the test achieves its specification after accelerated stress.
Conclusion: Accelerated stress data are indicative for a real-time stability of 24 months as determined for the
corresponding multipurpose reagent, supporting the claimed shelf life of 18 months.
Conclusion: The calibration stability of the auto-BILIRUBIN-D liquicolor HS600SR is confirmed for a minimum
of 7 days.
8 Interferences
Interference by hemoglobin, lipemia (intralipid) and ascorbic acid have been studied by adding known amounts of
the potentially interfering substance to a known sample.
Recoveries have been analysed according to the method of Glick et al. (Clin.Chem. 1986, 32 470-5).
Criterion Recovery within ± 10 % of initial value.
The results are summarised in the following table.
Hemoglobin Analyt. Result Result Intralipid Analyt. Result Result % Ascorbate Analyt. Result Result %
mg/dl mg/dl % mg/dl mg/dl mg/dl mg/dl
0 0.81 100 0 0.94 100 0 0.91 100.0
50 0.24 29.0 100 1.04 110.6 2 0.91 99.5
100 0.09 11.1 200 0.61 64.4 4 0.92 100.5
150 0.07 8.6 300 0.07 7.4 6 0.89 97.3
200 0.08 9.3 400 8 0.91 100.0
250 0.08 9.9 500 10 0.90 98.4
300 0.09 11.1 600 12 0.90 98.9
350 0.11 13.0 700 14 0.90 98.9
400 0.12 14.8 800 16 0.89 97.8
450 0.14 17.3 900 18 0.90 98.4
500 0.15 18.5 1000 20 0.91 99.5
Glick 4 n.a 1
Conclusion: This test is not significantly influenced by ascorbate < 20 mg/dl. Lipemic samples strongly
interfere and should not be used, there is poor correlation between intralipid and triglycerides. A
Glick value of 4 for hemoglobin indicates a strong interfering effect. Hemolytic samples should
be avoided.
9 Traceability
The auto-BILIRUBIN-D liquicolor HS600SR calibrated with AUTOCAL revealed recovery well to standardised
competitor methods within the reproducibility limits of the method. Reference material for direct bilirubin is not
available.