Professional Documents
Culture Documents
Picture credits:
https://encrypted-tbn0.gstatic.com/images?
q=tbn:ANd9GcRxiXuiTxtx3pld2nAR7KG76e_X84yKHKPHjhedQ_BiVkPJTfFqcZKXHYsdhxm3TBiZ_Ww&
usqp=CAU
10
8
mass(g)
0
0 2 4 6 8 10 12
run no
Figure 1 above graph of repeated measured mass of certified calibrated mass vs number of
runs
f(x) = NaN x + NaN
R² = 0 CMM 10g
12
10
8
mass(g)
0
0 2 4 6 8 10 12
run no
Figure 2 above graph of repeated measured mass of certified calibrated mass vs number of
runs
10
8
mass(g)
0
0 2 4 6 8 10 12
run no
Figure 3 above graph of repeated measured mass of certified calibrated mass vs number of
runs
f(x) = NaN x + NaN
R² = 0 stone
12
10
8
mass(g)
0
0 2 4 6 8 10 12
run no
10
8
mass(g)
0
0 2 4 6 8 10 12
run no
Figure 5 above graph of repeated measured mass of 125 conical flask vs number of runs
f(x) = NaN x + NaN
R² = 0 25ml conical flask
12
10
8
mass(g)
0
0 2 4 6 8 10 12
run no
Figure 6 above graph of repeated measured mass of 25 ml conical flask vs number of runs
10
8
mass(g)
0
0 2 4 6 8 10 12
run no
Figure 7 above graph of repeated measured mass of water sample vs number of runs
4. DISCUSSION
The experiment was successfully done even though there were some challenges such as
uncertainties due to some results of the measurement. The means of different samples were
obtained successfully as they were more accurate and precise. The means of certified
calibration masses of 50g was obtained to be 50.1601, 10g being 10.02085714 and 2g being
2.010029. These values are so close to the literature review results which estimates high
accuracy and precision since they were so close to each other. Similarly, the means of
materials was also found so close to the literature review as the mean of stone was10.47606,
125ml conical flask 93.99839, 25ml conical flask 29.05303 and water 5.239029. In addition
to that standard deviation also supported these results as it showed some smaller Lerona,
2014) that the standard deviation measures how closely the data are clustered around the
mean. The smaller the standard deviation the more closely the data are clustered about the
mean as it was shown on table 2 and 3 that the standard deviations of all samples were so
close to the mean hence an indication that the results were more precise and accurate.
95% confidence interval sets the confidence interval limits of the mean in which there is 95%
probability that true mean lies in that interval miller (2010). The confidence interval for the
mean of CCM 50g at 95% confidence interval the true value lies between 50.1601±0.000316
g, CCM 10g indicates that its true value lies between 10.02085714±0.000158912 and CCM
2g lies between 2.010021±0.000138, on the other hand stone its true value lies between
10.47606±0.001714, 125ml conical flask lies between 93,988839±0.000546,25ml conical
flask lies between 29.053303 ±0.000212 and finally water mass has its true value lying on the
range of 5.239029±0.005411.
However, the experiment was not as successful to its best because of some errors that have
been encountered during the process. Zero error was one them as the instrument had a
problem to be zeroed when taking measurements, Wells (1992). This may have resulted on
occurrence of some outliers which was visibly shown on the graph. Also, another thing which
could have affected the results was contamination of the instrument by dust and some salts
that are normally measured on it. Also, some degree of human reaction error might occur
which contributed to bad results Populaire (2004). In order to reduce human reaction error,
the experiment must be repeated for several times in order to try to minimize it.
5.CONCLUSION
Since the analytical balances provide accurate measurements to four decimal places, it is
concluded that they have a standard uncertainty of approximately 0.00010g. The mass
uncertainty of CCM 50 was 50.1601±0.000129, CCM 10g 10.02085714 ±6.49437E-05, CCM
2g is 2.010029±5.65E-05, stone is 10.47606±0.0007, 125ml conical flask
93.99839±0.000223, 25ml conical flask 29.05303±8.65E-05 and finally water mass was
5.239029±0.002211.
REFERENCE
Amilbhar, S. V., & Lerona, P. E. (2004, September 19). Using Analytical Balance and
Piso Statistics. Retrieved:January 7, 2019, from
http://www.academia.edu/901227/Using_Analytical_Balanceand_Piso_Statistics
Miller, J. N., & Miller, J. C. (2010). Statistics and chemometrics for analytical chemistry.
pearson .
Populaire, S., & Giménez, E. C. (2006). A simplified approach to the estimation of analytical
measurement uncertainty. Accreditation and quality assurance, 10(9), 485-493.
Wells, D. E., Maier, E. A., & Griepink, B. (1992). Calibrants and calibration for analytical
balances. International journal of environmental analytical chemistry, 46(4), 255-264.