Slide set of 87 slides based on the chapter authored by
P.B. ZANZONICO
of the IAEA publication (ISBN 78–92–0–143810–2):
Nuclear Medicine Physics:
A Handbook for Teachers and Students
IAEA
International Atomic Energy Agency
CHAPTER 10 TABLE OF CONTENTS
10.1. Introduction
10.2. Operating principles of radiation detectors
10.3. Radiation detector performance
10.4. Detection and counting devices
10.5. Quality control of detection and counting devices
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 2/87
10.1. INTRODUCTION
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 3/87
10.1. INTRODUCTION
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 4/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 5/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 6/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 7/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.1. Ionization detectors
Detector materials
• Most commonly gaseous
• Known as gas filled detectors
• Most widely encountered ones in nuclear medicine are
• Dose calibrators
• Geiger counters
(Difference is the bias voltage magnitude between
anode & cathode)
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 8/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.1. Ionization detectors
At V = 300 V
• All primary electrons are collected at the anode
• Detector signal is maximized
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 9/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.1. Ionization detectors
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 10/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.1. Ionization detectors
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 11/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.1. Ionization detectors
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 12/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.1. Ionization detectors
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 13/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.1. Ionization detectors
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 14/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.1. Ionization detectors
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 15/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.1. Ionization detectors
• Sealed volumes
• Use gases other than air (e.g. argon)
• Gas may be pressurized
• Provides higher stopping power
• Provides higher sensitivity
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 16/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.1. Ionization detectors
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 17/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.1. Ionization detectors
Proportional counters
• Need a stable bias voltage
specialized power supplies
• Restricted to research applications (e.g. in radiobiology)
• Need higher sensitivity
• Need capability of energy discrimination
• Often employ an unsealed, gas flow-through sensitive
volume
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 18/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.1. Ionization detectors
Geiger counters
• High sensitivity and stability with respect to voltage
• Can use a portable power supply such as an ordinary battery
• Widely used as survey meters (sensitivity is critical)
• To measure ambient radiation levels
• To detect radioactive contamination
• Sealed sensitive volumes
• Don’t need temperature–pressure corrections
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 19/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.1. Ionization detectors
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 20/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.2. Scintillation detectors
Scintillation detectors
• Most commonly a crystalline solid such as
Thallium-doped sodium iodide (NaI(Tl))
• Radiation interacts with and deposits energy
in a scintillator
• Deposited radiation energy
is converted to visible light
Emitted isotopically
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 21/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.2. Scintillation detectors
• Photocathode
• Coated on the inner surface of the PMT
• Emits electrons
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 22/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.2. Scintillation detectors
• Light pipe
• Between crystal back and PMT entrance window
• Thin layer of transparent optical gel
• Optically couples crystal to PMT
• Maximizes transmission (>90%) of the light signal from the
crystal into the PMT
• At ground: 0 V
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 23/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.2. Scintillation detectors
• 2nd dynode:
• Even larger +voltage, ~400 V
• Average of 3 e- ’s ejected per incident one
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 24/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.2. Scintillation detectors
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 25/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.2. Scintillation detectors
Energy discriminator
• Known as a pulse height analyser
• Sorts pulses by heights
• Pulses with height (i.e. energy) within the preset energy
window are counted by a timer/scaler
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 26/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.2. Scintillation detectors
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 27/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.2. Scintillation detectors
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 28/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.2. Scintillation detectors
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 29/87
10.2. OPERATING PRINCIPLES OF RADIATION DETECTORS
10.2.2. Scintillation detectors
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 30/87
10.3. RADIATION DETECTOR PERFORMANCE
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 31/87
10.3. RADIATION DETECTOR PERFORMANCE
10.3.1. Sensitivity
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 32/87
10.3. RADIATION DETECTOR PERFORMANCE
10.3.1. Sensitivity
• Geometric sensitivity
• Fraction of emitted radiations which strike the detector
• Directly proportional to the detector area
• Inversely proportional to the square of source–detector
distance (for point source)
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 33/87
10.3. RADIATION DETECTOR PERFORMANCE
10.3.1. Sensitivity
• Intrinsic sensitivity
• Fraction of radiation striking the detector & stopped within the
detector
• Intrinsic sensitivity with:
• Detector thickness
• Zeff
• ρ
• Intrinsic sensitivity with:
• photon energy
(Higher energy photons are more penetrating & are more likely to pass
detector without interacting)
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 34/87
10.3. RADIATION DETECTOR PERFORMANCE
10.3.1. Sensitivity
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 35/87
10.3. RADIATION DETECTOR PERFORMANCE
10.3.1. Sensitivity
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 36/87
10.3. RADIATION DETECTOR PERFORMANCE
10.3.1. Sensitivity
Rg − Rb
S=
A0 e − λ∆t
Rg is the gross (i.e. total) count rate (cpm) of the radionuclide source (RS);
Rb is the background (BG), or blank, count rate (cpm);
A0 is the activity (MBq) of the radionuclide source at calibration;
λ is the physical decay constant
Δt is the time interval (in months or years depending on the half-life) between
the calibration of the radionuclide and the current measurement
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 37/87
10.3. RADIATION DETECTOR PERFORMANCE
10.3.1. Sensitivity
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 38/87
10.3. RADIATION DETECTOR PERFORMANCE
10.3.2. Energy resolution
Energy resolution
• Ability to separate/discriminate different energies
∆E
• Given by the width of photopeak FWHM(%) = ×100%
Eλ
• Related to Poisson ‘noise’
• Statistical uncertainty inherent in the detection process
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 39/87
10.3. RADIATION DETECTOR PERFORMANCE
10.3.2. Energy resolution
Energy resolution
• Scattered and primary radiations
overlap (due to finite resolution) ∆E
FWHM(%) = ×100%
• Better resolution = narrower photopeak Eλ
• Can better separate unscattered &
scattered radiations
• Can eliminate more scattered
radiation counts
• Fewer counts due to unscattered
radiation are discarded
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 40/87
10.3. RADIATION DETECTOR PERFORMANCE
10.3.3. Count rate performance (‘speed’)
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 41/87
10.3. RADIATION DETECTOR PERFORMANCE
10.3.3. Count rate performance (‘speed’)
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 42/87
10.4. DETECTION AND COUNTING DEVICES
10.4.1. Survey meters
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 43/87
10.4. DETECTION AND COUNTING DEVICES
10.4.1. Survey meters
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 44/87
10.4. DETECTION AND COUNTING DEVICES
10.4.1. Survey meters
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 45/87
10.4. DETECTION AND COUNTING DEVICES
10.4.2. Dose calibrator
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 46/87
10.4. DETECTION AND COUNTING DEVICES
10.4.2. Dose calibrator
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 47/87
10.4. DETECTION AND COUNTING DEVICES
10.4.3. Well counter
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 48/87
10.4. DETECTION AND COUNTING DEVICES
10.4.3. Well counter
• Common design
• Cylindrical scintillation crystal
• Circular bore (well) for sample drilled part-way into the crystal backed
by a PMT + electronics
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 49/87
10.4. DETECTION AND COUNTING DEVICES
10.4.3. Well counter
• Alternative design
• ‘Through-hole’
• Hole is drilled through the entire crystal
• Facilitates sample exchange
• Samples are centered lengthwise
• More constant response for different volumes
• Slightly higher sensitivity than well counters
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 50/87
10.4. DETECTION AND COUNTING DEVICES
10.4.3. Well counter
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 51/87
10.4. DETECTION AND COUNTING DEVICES
10.4.3. Well counter
• Integrated computer
• To create & manage counting protocols
• Specify isotope
• Specify energy window
• Specify counting interval
• Manage sample handling
• Apply background, decay, dead time & other corrections
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 52/87
10.4. DETECTION AND COUNTING DEVICES
10.4.4. Intra-operative probes
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 53/87
10.4. DETECTION AND COUNTING DEVICES
10.4.4. Intra-operative probes
• Semiconductor (ionization)
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 54/87
10.4. DETECTION AND COUNTING DEVICES
10.4.4. Intra-operative probes
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 56/87
10.4. DETECTION AND COUNTING DEVICES
10.4.5. Organ uptake probe
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 57/87
10.4. DETECTION AND COUNTING DEVICES
10.4.5. Organ uptake probe
• Generally supplied as
• Integrated computerized system
• With automated
• Data acquisition
• Processing capabilities
• Yield results in terms of % uptake
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 58/87
10.4. DETECTION AND COUNTING DEVICES
10.4.5. Organ uptake probe
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 59/87
10.4. DETECTION AND COUNTING DEVICES
10.4.5. Organ uptake probe
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 60/87
10.4. DETECTION AND COUNTING DEVICES
10.4.5. Organ uptake probe
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 61/87
10.4. DETECTION AND COUNTING DEVICES
10.4.5. Organ uptake probe
• C = total counts
• t = measurement time
• F = fraction of administered activity in the standard
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 62/87
10.4. DETECTION AND COUNTING DEVICES
10.4.5. Organ uptake probe
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 63/87
10.4. DETECTION AND COUNTING DEVICES
10.4.5. Organ uptake probe
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 64/87
10.4. DETECTION AND COUNTING DEVICES
10.4.5. Organ uptake probe
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 66/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 67/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 68/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 69/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.1. Reference sources
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 70/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.1. Reference sources
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 71/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.1. Reference sources
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 72/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.2. Survey meter
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 73/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.2. Survey meter
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 74/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.2. Survey meter
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 75/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.2. Survey meter
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 76/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.3. Dose calibrator
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 77/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.3. Dose calibrator
• Linearity
• ‘decay method’
• At least quarterly
• Daily readings should agree within 10%
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 78/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.3. Dose calibrator
Measurements at installation
• 99mTc geometry dependent response
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 79/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.3. Dose calibrator
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 80/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.3. Dose calibrator
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 81/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.3. Dose calibrator
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 82/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.4. Well counter
• Background
• Check for each energy window used
• Count rate should be checked daily
• Electronic noise & ambient radiation
• May be relatively high and variable in a nuclear medicine facility
• Will produce a non-zero/potentially fluctuating count rate
• Trace contamination will produce inaccurately high count rate values
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 83/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.4. Well counter
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 85/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.5. Intra-operative probe
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 86/87
10.5. QUALITY CONTROL OF DETECTION AND COUNTING DEVICES
10.5.6. Organ uptake probe
IAEA
Nuclear Medicine Physics: A Handbook for Teachers and Students – Chapter 10 – Slide 87/87