You are on page 1of 4

Fresenius J Anal Chem (1999) 364 : 705–708 © Springer-Verlag 1999

O R I G I N A L PA P E R

Vít Šprta · Bohumil Knob · Pavel Janoš

X-ray fluorescence determination of total sulfur in fly ash

Received: 27 October 1998 / Revised: 5 March 1999 / Accepted: 11 March 1999

Abstract Standard addition, double dilution and stan- production, concrete mixing, ceramics and other products.
dard calibration were used for x-ray fluorescence (XRF) For those applications, it is important to know the content
determinations of sulfur in fly ashes. Samples were analysed of the major elements, both metals (Al, Fe, Ca) as well as
as pellets prepared by mixing with acrylate copolymer or non-metals (Si, S). Sulfur may cause a degradation of the
with microcrystalline cellulose (in the case of the double concrete and therefore its amount has to be controlled [3].
dilution method). Lithium sulfate was used for the stan- Various analytical methods have been used for an ele-
dard addition method and also as standard with known mental analysis of the fly ashes. Most of them (gravime-
sulfur content for the double dilution method. Fly ashes try, titrimetry, photometry, AAS, ICP-OES, ICP-MS),
analysed by optical emission spectrometry with an induc- however, require a laborious and time consuming sample
tively coupled plasma (ICP-OES) were used as standards dissolution. From the non-destructive techniques, x-ray
for the standard calibration XRF method. Sulfur was de- emission methods, such as particle-induced x-ray emis-
termined in the range of ca. 10–1–100 % S. For the fly ashes sion (PIXE) and x-ray fluorescence spectrometry (XRF)
from the North-Bohemian brown coals, the differences have been employed most frequently for the bulk ele-
between the XRF determinations and the ICP-OES deter- mental analyses of the fly ashes. Sixteen metals, both ma-
minations ranged from ca. 1.4 to 10% rel. and precision jor and microelements, were quantified in fly ash samples
(repeatability) was better than 10% (RSD). The standard from the coal-fired power plants in India using PIXE and
calibration method is suitable for routine analyses of real energy-dispersive XRF [4]. XRF together with instru-
samples of similar nature. The methods of standard addi- mental neutron activation analysis (INAA) were used for
tion and double dilution are rather laborious in sample the determination of 23 elements in Hungarian brown
preparation compared with the standard calibration. coals, fly ash, bottom ash and dust particulates [5]. Parti-
cle size/density separated fractions of the coal fly ash
were characterised by energy-dispersive XRF and other
Introduction instrumental techniques [6]. Metalic elements were in-
vestigated in [4–6] from the point of view of their envi-
Fly ash is a solid waste produced in great quantities par- ronmental impacts. Major as well as minor elements in-
ticularly during combustion of fossil fuels in power cluding Si, but not sulfur, were determined by automated
plants, but also in many other industrial processes. Its pro- XRF of the pressed fly ash pellets [7]. Sulfur together
duction as well as its handling and dumping causes poten- with a number of major and trace elements were deter-
tial hazards to plants, animals and human health [1]. On mined in a study of the elemental distribution in Greek
the other hand, the utilisation of the fly ash in waste man- lignites and their ashes [8]. Direct XRF measurements
agement, e.g. for wastewater sludge conditioning, is ex- were recommended for the determinations of 22 elements
tensively studied [2]. Therefore, the contents of toxic ele- including some non-metals (Si, S) in airborne particles
ments such as heavy metals are usually determined to as- [9]. Relatively little experimental details are reported on
sess environmental impacts. the determination of sulfur in the above mentioned stud-
Recently, fly ash has gained much attention as a raw ies. Depth selective x-ray absorption fine structure spec-
material in building industry, where it is used for cement trometry (XAFS), in which both the x-ray fluorescence
yield and the total electron yield are recorded concomi-
tantly, seems to be very promising for element-speciation
and element-distribution studies [10]. XAFS allows to
V. Šprta, B. Knob, P. Janoš (쾷)
Research Institute of Inorganic Chemistry, Revoluční 84, obtain information about the chemical state and distribu-
CZ-400 01 Ústí nad Labem, Czech Republic tion of sulfur in particles of the fractionated coal fly ash
706

[11]; this technique, however, is not commonly available


in most laboratories.
In the present work, XRF was used to determine the to-
tal sulfur content in fly ash. Various sample preparation
methods were applied to several kinds of fly ashes and
various methods of calibration and data evaluation were
compared.

Experimental
Fly ashes, chemicals

Samples of the fly ashes were collected from electrostatic precipi-


tators in various power plants fuelled with North-Bohemian brown Fig. 1 Determination of sulfur by multiple addition – relative in-
coals. The samples were dried at 60 °C for 5 h and then ho- tensity vs. sulfur addition
mogenised in a ball mill with corundum balls for 5 h. The CW3
sample was prepared in the Joint Research Centre, Ispra, Italy, as a
candidate reference material from incinerated chemical waste col-
lected in Germany and represents the ash produced in a specialised the Kα spectral lines were measured at the maximum
incineration plant. (Imax) and at the baseline (Ibas). Simultaneously with the
For the standard addition measurements, 10 g of the sample sample, the instrumental standard was measured (Ist). The
was mixed with 5 mL of a solution of lithium sulfate of known relative intensity was calculated according to the equa-
concentration, the mixture was homogenised by stirring and dried tion:
at 60 °C. The dry mixture was mixed with acrylate copolymer,
dried at 60 °C again, ground and pressed in an Al dish (160 MPa). Irel = (Imax – Ibas)/Ist (1)
For standard calibration measurements, the samples were treated
in a similar way without addition of lithium sulfate. This procedure was applied to fly ashes from the North-
Microcrystalline cellulose was used as an inert diluent for dou- Bohemian brown coal (P4, P6, P9) and to the CW3 ash, af-
ble dilution measurements. Two pellets containing 60 and 30% of
fly ash were prepared as follows: the fly ash was mixed with cel- ter additions ranging from 0.053 to 1.53% S. The relative
lulose at ratios 6 : 4 and 3 : 7, respectively. The mixtures were intensities were plotted against the added amount of sulfur
ground, homogenised and pressed in an Al dish (160 MPa). and the concentration of sulfur in the original sample was
Lithium sulfate LiSO4 · H2O served as a standard with a known obtained from the intersection of an extension of the
content of sulfur – it was treated in the same way as the fly ash.
Conditions of the x-ray measurements are given in Table 1. A spe- straight line (relative intensity vs. addition of sulfur) with
cial glass U4 (2THETA, Český Těšín, Czech Republic) containing the x-axis (Fig. 1). As can be seen from this figure, a lin-
1.44% S was used as an internal standard to determine the relative ear dependence is obtained only up to ca. 1% S. The
intensities of the spectral lines. ICP-OES was used as reference points besides the linear range were not taken into account
method for the sulfur determination. The measurements were per- during the evaluation of the measurements. In addition to
formed with the Optima 3000, Perkin-Elmer, spectrometer at the
emission line 180.669 nm. the graphical evaluation presented in Fig. 1 also a numer-
ical approach was applied utilising the least-square
method. Correlation coefficients were typically 0.994–
Results and discussion 0.996, except for the P4 sample (0.978). The results of
the determinations are listed in Table 2 together with esti-
mated standard deviations and together with the results
Standard addition method
obtained by other methods.
Several samples with various amounts of lithium sulfate
(as suggested by Chinchón et. al. [3]) were prepared as
Double dilution method
described in the experimental section. The intensities of
In dilution methods, a sample is mixed with an inert elu-
Table 1 Apparatus and conditions of measurements
ent containing no analyte. Two intensity measurements at
two different concentrations of the sample permit correc-
X-ray spectrometer: PHILIPS PW 1404, Rh lamp, 3 kV tions of the measured intensities for absorption-enhance-
Spectral line: Kα ment effects. The standard with a known concentration of
Voltage: 40 kV the analyte is diluted in the same way as the sample (for a
Current: 75 mA more detailed discussion see [12, 13]).
Collimator: Fine For the determination of sulfur in fly ashes, microcrys-
Crystal: GE talline cellulose and lithium sulfate were used as a diluent
Angle: 110.665° and standard, respectively (see experimental section). The
Detector: FL relative intensities measured in a similar way as stated
Time of measurement: 30 or 60 s above were used for the calculations of the sulfur con-
Software: X40, X44 tents; results are listed in Table 2. Double dilution was not
707

Table 2 Determination of sul-


fur in fly ashes by various Sample XRF – standard XRF – double XRF – standard ICP-OES
XRF techniques, comparison addition dilution calibration
with ICP-OES. Content of S
in % P4 0.21 ± 0.01 – 0.10 ± 0.03 0.11 ± 0.02
P6 0.75 ± 0.02 0.92 ± 0.04 0.65 ± 0.05 0.67 ± 0.04
Confidence interval (α = 0.05)
estimated from at least 3 deter- P9 1.69 ± 0.07 1.85 ± 0.11 1.46 ± 0.09 1.48 ± 0.08
minations according to Dean – CW3 2.36 ± 0.08 2.02 ± 0.07 2.24 ± 0.18 1.82 ± 0.14
Dixon: x ± Kn R

Table 3 Precision of the measurement and repeatability of the ple. The repeatability of the whole procedure including
XRF determination sample preparation was tested for the calibration stan-
Precision of Repeatability – Repeatability – dards with the lowest (level 1) and the highest (level 2)
measurement level 1 level 2 concentration of sulfur (Table 3).
(n = 10) (n = 10) (n = 10) The results of the sulfur determination in the P4, P6,
Average (%S) 0.652 0.116 1.410 P9 and CW3 samples are listed in Table 2.
Standard deviation 0.002 0.009 0.035
(%S)
Conclusions

applied to the P4 fly ash, due to the low concentration of The results of the sulfur determinations with various XRF
sulfur in the original sample, so that a sufficient precision methods are summarised in Table 2 and compared with
under the given conditions cannot be obtained. the results of the ICP-OES determinations. The best
agreement between the XRF and ICP-OES analyses was
found for the method of standard calibration. This is quite
Standard calibration method understandable, because the XRF standard calibration
method was calibrated using fly ashes of a similar compo-
The standard calibration method is most suitable for rou- sition previously analysed by ICP-OES. An exception is
tine analyses provided that the analytical function can be the CW3 sample, in which the sulfur content is above the
determined by means of materials with a matrix identical upper limit of the calibration range (an extrapolation of
to that of the real samples. Unfortunately, there is a lack of the calibration line was necessary). Moreover, the origin
reference materials (fly ashes) with certified values of the and to a certain degree also the matrix of the CW3 sample
sulfur content in a sufficient range of concentrations. Of- was different from those of other samples and calibration
ten, the sulfur content in the fly ash reference materials is standards, which caused a lower accuracy of the determi-
indicated as an “information value” only. Sets of samples nation.
previously analysed by other analytical methods, such as The methods of standard calibration and standard addi-
AAS or ICP-OES, are frequently used for the calibration tion can be applied to fly ashes with sulfur contents from
of the XRF method, especially in environmental analysis ca. 0.1% to ca 1.5–2%. The double dilution method can-
[14]. In this work, a series of the fly ashes was analysed not be applied to samples with a low sulfur content, be-
by ICP-OES after microwave digestion and these results cause the intensities of the spectral line are poorly de-
were used for the calibration of the XRF method. The re- tectable for the diluted samples. The critical step for the
liability of the ICP-OES determinations was confirmed by standard addition as well as for the double dilution meth-
comparison with other methods and a small-scale inter- ods is the sample preparation: mixing/homogenisation
laboratory comparison [15]. with additive or diluent. The precision obtained for the
The calibration standards as well as samples for the determination of sulfur in fly ashes is comparable to that
XRF measurements were prepared as described in the ex- achieved by XRF determination of sulfur in sulfide-con-
perimental section and the respective intensities of the Kα taining rocks [17].
spectral line were measured. Parameters of the calibration During all the XRF measurements, no decrease of the
line signal intensity with time (testifying the presence of ele-
mental sulfur [18]) was observed. It is supposed that sul-
y=Ex+D (2) fur is present as sulfate (mainly in the surface layer), or
were: E = 0.2765, D = –0.00723. The limit of detection minor (residual) as sulfide in the bulk [11].
(LOD) was calculated from the following equation [16]
Acknowledgements This work was supported by the INCO-
COPERNICUS Program, project ERB-IC15-CT96-0811. Finan-
–D/ E
LOD = 3000 E (3) cial support from the Czech Ministry of Education (grant No. OK
10 t 296) is also gratefully acknowledged.
where t is the measuring time. LOD was 8 mg/kg for t =
30 s. The precision of the XRF measurements was deter-
mined from 10 repetitive measurements on the same sam-
708
10. Kawai J, Adachi H, Hayakawa S, Zhen S-Z, Kobayashi K,
References Gohshi Z, Maeda K, Katajima Z (1994) Spectrochim Acta
49B: 739
1. Adriano DC, Page PL, Elseewi AA, Straughan I (1980) J Env- 11. Kawai J, Hayakawa S, Esaka F, Zheng S, Kitajima Y, Maeda
iron Qual 9: 330 K, Adachi H, Gohshi Y, Furuya K (1995) Anal Chem 67: 1526
2. Wang S, Viraraghavan T (1997) Waste Management 17: 443 12. Bertin EP (1970) Principles and Practice of X-Ray Spectromet-
3. Chinchón JS, López-Soler A, Travería A, Vaguer R (1991) ric Analysis, Plenum Press, New York, pp 581–583
Materials and Structures 24: 13 13. Tertian R, Claisse F (1982) Principles of Qantitative X-Ray
4. Vijayan V, Behera SN, Ramamurthy VS, Puri S, Shahi JS, Fluorescence Analysis, Heyden, London
Singh N (1997) X-Ray Spectrom 26: 65 14. Driscoll JN (1998) Int Environ Technol 8 (No.5): 10
5. Rausch H, Sziklai IL, Sándor S, Szabó T (1995) J Radioanal 15. Ariese F (ed) (1998) Analeach, Annual report W-98/14, IVM,
Nucl Chem 1996: 25 Amsterdam
6. Furuya K, Miyajima Y, Chiba T, Kikuchi T (1987) Environ Sci 16. X44 Software for XRF, Operational Manual, Philips Analyti-
Technol 21: 898 cal, 4th edn, p 7–78
7. Bettineli M, Taina P (1990) X-Ray Spectrom 19: 227 17. Patsitschniak A (1986) X-Ray Spectrom 15: 197
8. Boganovic I, Fazinic S, Itskos S, Jakšic M, Karydas E, Katselis 18. Chinchón JS, López-Soler A, Travería A, Vaquer R (1988) X-
V, Paradellis T, Tatic T, Valkovic O, Valkovic V (1995) Nucl Ray Spectrom 17: 217
Instrum Methods in Phys Res B 99: 402
9. Wang CF, Chang EE, Chiang PC, Aras NK (1995) Analyst
120: 2521

You might also like