Calibration Guide T21P Indicator

 Make sure appropriate kilogram calibration masses are available before beginning
calibration.
 Make sure the scale base is level and stable during the entire calibration process.
 Allow the Indicator to warm up for five minutes after stabilizing to room temperature.
 To abort calibration, press and hold the SET button for three seconds any time during
the calibration process. Or turn the unit off.

Turn on the Indicator. During startup display sequence press and hold SET until FS__SP
appears. Then follow these steps:

1. Press ENTER to display P___x. (x=position of decimal point: 0, 1, 2, 3, and 4).

2. Press ,  to select the desired setting.

3. Press ENTER to display D___x. (x=divisions: 1, 2,5,10, 20, and 50).


4. Press ,  to select the division.
5. Press ENTER to display Maximum Capacity Setting.
6. Press ,,  to set the desired value.
The LED row below the display refers to the active segment.

7. Press ENTER to accept and noLoad] appears.


Make sure the weighing platform is empty.

8. Press ENTER to accept and Load appears.


9. Press ENTER to display xxxxxx. (xxxxx = Span Weight).
10. Press ,,  to set a different value.
11. Place calibration weight on the weighing platform.
12. Press ENTER to accept and store data.

CALEnd will flash for two seconds; then the Indicator will enter weighing mode.

*80252989* P/N 80252989