You are on page 1of 8

Technology in Cancer Research & Treatment

ISSN 1533-0346
Volume 1, Number 2, April (2002)
©Adenine Press (2002)

Historical Technical Developments in Mammography Arthur G.Haus

www.tcrt.org Medical Physicist


In mammography it is most important to consistently produce high-contrast, high resolution 112 Tar Heel Drive
images at the lowest radiation dose consistently with high image quality. Today it is possible Delaware,OH 43015
to obtain mammograms with higher image quality that require significantly lower radiation
dose compared to mammograms dating back to the 1970’s and early 1980’s. Some of the
technical developments due to x-ray equipment and image receptors and quality control pro-
grams that have let to todays high quality mammographic images are discussed in this article.

Introduction

In recent years, there have been many significant technological improvements in


mammographic x-ray equipment, image recording systems, and viewing condi-
tions. Some of the major technical development and quality control milestones
in mammographic imaging are shown in Table 1 (1-7). Until the mid 1980’s,
many x-ray units were used that were not dedicated to mammography. These x-
ray units had tungsten target tubes that were designed originally for medical
imaging procedures, such as chest radiography. Some of these units had com-
pression devices that were home made;therefore, breast compression was less
than optimal by today ’s standards. Many of these units had very large focal
spots or short focal spot-to-breast surface distances that could result in signifi-
cant geometric blur (unsharpness). Direct exposure (industrial type) x-ray films
were being used, which often required long exposure times (causing blur by
motion) and which resulted in high radiation exposure. In addition, viewing con-
ditions were inadequate.

Today, mammography is performed with dedicated mammographic x-ray equip-


ment. These units have specially designed tube targets, smaller focal spots, and
significantly improved breast compression devices, among other features.
Cassettes and screen-film combinations are designed specifically for mammog-
raphy. Film processing and viewing conditions also have improved significant-
ly over the years. In 2000, the first digital mammography system was approved
by the Food and Drug Administration (FDA) for clinical use.

The American College of Radiology (ACR) Mammography Accreditation


Program introduced in 1987, the ACR Quality Control Manuals introduced in
1992, and the Mammography Quality Standards Act which was implemented in
1994, have also had a significant impact on the improvement of the technical
quality of mammographic images in the United States.

Today it is possible to obtain mammograms with higher image quality that


Corresponding Author:
require significantly lower radiation doses compared with mammograms dating Arthur G. Haus
back to the 1970’s and early 1980’s, (Figure 1). E-mail: hausag@aol.com

119
120 Haus

Table I
Technical Advances in Mammography
Year Development
Prior to 1969 Conventional tungsten target x-ray tubes with direct exposure industrial type films were used
1969 Dedicated mammographic unit with molybdenum target tube and compression cone introduced (CGR Senographe)
1971 Xeroradiography system introduced for mammography (Xerox)
1972 Screen-film system introduced for mammography (DuPont Lo-dose system)
1976 Rare earth screen-film system and special cassette introduced for mammography (Kodak Min-R system)
1977 Mammography x-ray unit for magnification with microfocal spot introduced (Radiological Sciences Inc.)
1978 Mammography unit with grid introduced (Philips)
1987 American College of Radiology Mammography Accreditation Program (ACR MAP) begins
1992 American College of Radiology Mammography Quality Control Manual for Radiologists, Radiologic
Technologists, and Medical Physicists, introduced
1994 The Food and Drug Administration (FDA) implements the Mammography Quality Standards Act (MQSA)
2000 Digital mammography system approved by the FDA for clinical use (GE Senographe 2000D)

X-ray Equipment aluminum (xeroradiography). The unit also had a compres-


sion cone, a most important feature that identified it as a ded-
Prior to the late 1960’s, there were no commercially available, icated mammographic unit (Figure 2 left). Between 1969
dedicated mammographic x-ray units. X-ray units with tung- and 1973, more than 1000 units were installed. In the early
sten targets which were originally designed for medical imag- 1970’s, dedicated mammographic units became commercial-
ing procedures such as chest radiography were being used. ly available from several companies, including General
Some of these units had compression devices that were home- Electric, Philips, Picker, and Siemens (1, 5, 8).
made;therefore, breast compression was less than optimal by
today ’s standards. Many of these units had very large focal
spots and/or short focal spot-to-breast surface distances which
could result in significant geometric blur (unsharpness).
Direct exposure (industrial type) x-ray films were used which
often required long exposure times (causing blur due to
motion) and which resulted in very high radiation exposures.

Figure 2: The first commercially available mammographic x-ray unit, the


CGR Senographe, was produced in 1967. The unit featured a molybde-
num target x-ray tube nd a compression cone (left). The CGR Senographe
unit with long cone technique (right) (Haus et al 13))

In 1977, dedicated unit with microfocal spot for magnifica-


tion mammography became commercially available from
Figure 1: Mammograms of the same patient, same breast, taken in (A) Radio-logic Sciences Inc., a subsidiary of Pfizer (1). This
1977, (B) 1984, (C) 1990 and (D) 1998. (Images courtesy of Wende Logan- unit had a tungsten target with beryllium window and two
Young, MD) focal spot sizes – microfocal spot (0.09mm nominal) for
magnification and a larger (0.45mm) focal spot for conven-
Tube Target and Filtration tional mammography. In 1984 (CGR) introduced a higher
current microfocus tube which allowed a longer SID with
Charles-Marie Gros, a physicist and physician at the less scatter for the same magnification factor (8).
University of Strasbourg, using a phantom, experimented
with crystallographic x-ray tubes having different anode In 1992, double track x-ray source with a molybdenum tar-
materials and filters. He showed that molybdenum anode get, molybdenum filter, a rhodium target, and a rhodium fil-
and molybdenum filter produced improved images (higher ter was introduced by General Electric.
contrast) of the breast. In 1969, the first of series of dedi-
cated mammographic x-ray units – the CGR Senographe unit Dedicated molybdenum target x-ray units are widely used,
contained a molybdenum target tube which filtered the beam and settings of less than 28 kVp are generally recommended
using either 0.03 mm of molybdenum (film) or 0.5 mm of with these units (2, 4). The use of low-energy photons, such

Technology in Cancer Research & Treatment, Volume 1, Number 2, April 2002


Historical Developments in Mammography 121

as those produced by the 17.9 and 19.5 keV characteristic object-to-image receptor distances affect geometric blurring.
lines from the molybdenum target, provide high subject con- To minimize geometric blurring, the focal spot size and
trast for breasts of average thickness. When a 0.03mm object-to-image receptor distance should be minimized,
molybdenum filter is used, the spectrum is strongly sup- whereas focal spot-to-object distance should be maximized.
pressed at photon energies greater than 20 keV because of the Focal spot sizes and shapes for the dedicated and conven-
k-shell absorption edge of molybdenum at that energy (9-11). tional x-ray units used for mammography vary considerably
This allows more radiation from the characteristic lines to be (14, 15). Likewise, the focal spot-to-breast surface distances
used in image formation. Other tube target and filter combi- for different mammographic units vary.
nations used in dedicated mammographic units include a (1)
tungsten target with approximately 0.06-mm molybdenum Several years ago it was common to compress the breast
filter, (2) tungsten target with approximately 0.05-mm rhodi- directly on top of the mammographic cassette. The distance
um filter, and (3) rhodium target with approximately 0.025- between the chest wall edge of the breast and the screen-film
mm rhodium filter (Figure 3). It has been shown that molyb- combination was very small. Today, many mammographic
denum target with molybdenum filter combinations are procedures re performed with moving Bucky-type-grid.
appropriate for breasts less than 5 cm thick, and molybdenum With the grid in place, a gap of 1 to 2 cm in distance between
target with rhodium filters or rhodium target with rhodium the edge of the breast and the screen-film combination may
filters may be appropriate for breasts greater than 5 cm thick occur. The size of the focal spot, therefore, needs to be
(12). For mammography, kVp settings between 22 and 32 smaller in terms of limiting geometric resolution for a given
are used depending on breast tissue thickness and composi- focal spot-to-breast surface distance. The typical nominal
tion and exposure time. It has been shown that increasing focal spot sizes for todays mammography units are approxi-
kVp settings between 24 and 32 kVp, at comparable optical mately 0.3mm (non magnification) and 0.10mm (magnifica-
densities causes a slight decrease in lesion detections (8). tion technique) (4, 7).

Grids for Mammography

Scattered radiation can significantly reduce image contrast in


mammography (Figure 5) (16). In 1978 Philips introduced
moving grid for mammography.

The use of specifically designed grids for mammography


reduce scattered radiation and improves subject contrast,
which is especially significant when imaging thick, dense
breasts (5, 9, 16, 17, 18). Grids are now included with ded-
icated mammographic x-ray units. The majority of grids
used for mammography consist of lead strips separated by
spacers of radiolucent material such as carbon fiber. Most of
these grids are the moving type, which blur the grid lines.
Moving-type grids are preferred for mammography.

Figure 3: Typical x-ray spectra comparing molybdenum target and molyb-


denum filter with rhodium target and rhodium filter (Haus 11).

Geometry and Focal Spots

In 1975, a study showed that, although direct exposure films


have higher spatial resolution than mammographic screen-
film combinations, total resolution can be optimized with
mammographic screen film combinations by using longer
cones (increased focal spot-to-film distances) and/or smaller
focal spots on dedicated mammographic units (Figure 2
Figure 4: Modulation transfer functions (MTF’s) of geometric unsharp-
right, Figure 4) (13). ness when the conventional cone and the long cone with improved geome-
try are used at an object-to-recording system distance of 5 cm. MTF’s are
The size, shape and intensity distribution of the x-ray tube of Kodak RP/M non-screen film and the DuPont Lo-dose screen-film sys-
focal spot in combination with focal spot-to-object and tem for mammography are also shown (Haus et l. 13).

Technology in Cancer Research & Treatment, Volume 1, Number 2, April 2002


122 Haus

Automatic Exposure Control

Automatic exposure control systems, also referred to as pho-


totimers, are designed to automatically provide the radiation
exposure needed to produce a mammogram with an accept-
able (and consistent) optical density. Most dedicated mam-
mographic x-ray units have automatic exposure control
devices. New circuit modifications and developments in
detector arrangements, which take into account changes in
breast tissue thickness and composition, can provide more
consistent film optical density (19).

Image Receptors

In the 1950’s and 1960’s, direct exposure with medical film or


Industrial type films was used for mammography (Figure 7).
In 1971, a xeromammography unit became available (20).
Until then, very little attention was primarily directed to the
use of an image receptor that could provide a mammogram of
adequate contrast, high resolution and low noise.

Figure 5: Dependence of the ratio of scattered-to-primary radiation on the


diameter of the radiation field for 3 and 6 cm thickness of a Lucite phantom:
“A at 32 kVp and “B ”at 42 kVp (Barnes and Brezovich 16)

Recently, rhombic cellular structure air grids have been


introduced. These grids offer the potential of improved
image contrast and transmission efficiency compared with
conventional grids (Figure 6).

Figure 7: Characteristic curves for (1) a direct-exposure film, (2) a single-


screen, single–emulsion film used in the 1970’s and early 1980’s and (3) a
single-screen, single–emulsion film combination used for mammography
today. (Haus 11)

In 1972, the first screen-film combination designed for


mammography was introduced by DuPont (21). With this
single-screen, single-emulsion combination, radiation dose
was reduced 10 to 20 times compared to direct exposure
films. DuPont also introduced a vacuum bag system to pro-
Figure 6: Contrast improvement factor versus compressed breast thickness vide good screen-film contact. In 1975, Kodak brought to
for HTC and linear grid. This is a relative curve. It changes with different market mammography screen-film combination which uti-
breast composition and film gradient. (Courtesy of Trex Medical lized rare earth phosphor screens. This combination pro-
Corporation, Lorad Division, Danbury, CT)
duced sharper images and required one half the radiation
exposure of earlier screen-film combinations. A special low-
Grids designed for mammography generally require expo- absorption cassette was also introduced for mammography.
sure increases of approximately 2 to 2.5 times the exposure In 1977, Xonics introduced an electron radiography system
required for nongrid techniques. for mammography (22).

Technology in Cancer Research & Treatment, Volume 1, Number 2, April 2002


Historical Developments in Mammography 123

The xeroradiography system and the Xonics electronic radi- (2) phosphor particle size, (3) light-absorbing dyes and pig-
ography system are no longer commercially available. ments in the screen, and (4) screen-film contact. In recent
years, screen-film combinations for mammography have
Screen-Film Combinations used a single high-definition screen in contact with a single-
emulsion film.
Most single-screen, single-emulsion film combinations com-
monly used today have higher film contrast and require sig- The single screen is used as a back screen for mammography
nificantly lower radiation exposure than those used a few because x-ray absorption (and emission of screen light) is
years ago (Figure 7). It is interesting to note that today’s highest on the side of the screen where the x-rays enter. If the
screen-film combinations require approximately 50 times less screen were used s a front screen, x-ray absorption would be
radiation than direct-exposure films. Films currently used in higher in the plane of the screen that is the farthest distance
mammography usually have a single emulsion and are used in from the screen-emulsion contact surface. This causes
combination with a single back screen. Films are sensitized greater light spread (blur) than when x-ray absorption is high-
to match the spectral emission of the intensifying screen. est near the screen-emulsion contact surface, as is the case
when it is used as a back screen. Both parallax and crossover
The majority of mammographic images are produced with are eliminated in a single-back-screen configuration.
single intensifying screen used as a back screen in combina-
tion with single-emulsion film (Figure 8). Many mammo- Figure 9 shows modulation transfer function (MTF) curves
graphic screens incorporate phosphors containing metals for a direct-exposure film and five mammographic screen-
from the lanthanide series of elements such as terbium-acti- film combinations. Also shown is an MTF curve for a
vated gadolinium oxysulfide (Gd2O2S:Tb). Screens may screen-film combination used for conventional diagnostic
incorporate light absorbers in the phosphor that are used to radiology procedures. These curves show that screen-film
increase sharpness. Intensifying screens have a protective combinations used for mammography have much higher
overcoat to resist surface abrasion, are edge-sealed to mini- spatial resolution than do those used for conventional diag-
mize edge wear, and the base includes a backing layer to elim- nostic procedures (11)
inate screen curl. Mammographic screens consisting of ter-
bium-activated gadolinium oxysulfide material emit light in
the visible spectral region from 382 to 622 nm, although the
primary emission peak is in the green spectral region, 545 nm.

For screen-film radiography, light diffusion (spreading of the


light emitted by the intensifying screen before it is recorded
by the film) causes blurring (Figure 8) (11). Factors
involved include (1) phosphor layer thickness in the screen,

Figure 9: Modulation transfer function curves are for a direct-exposure


film; three mamographic screen-film combinations; and a two-screen, dou-
ble-emulsion film combination commonly used for conventional radiogra-
phy. (Haus 11)

Film Processing

In the late 1980’s and early 1990’s, the importance of film


processing in mammography became recognized (Figure 10)
(23, 24)

Film processing must be considered as part of system that


Figure 8: Diagrams compare physical configurations for single-emulsion includes the automatic film processor, film type, and chemi-
film in contact with single (back) intensifying screen (as used for mam- cals (25). Today it is recognized that these components must
mography) and a double-emulsion film sandwiched between two intensify- be considered together as a system and must be properly
ing screens (used for other radiological procedures. (Haus 11) optimized, maintained, and sustained to obtain appropriate

Technology in Cancer Research & Treatment, Volume 1, Number 2, April 2002


124 Haus

Several companies are providing viewbox systems designed


for reading mammograms with appropriate luminance levels
(Table 2) (27). The ACR Mammography Quality Control
Manual indicates that the mammographic viewboxes should
be capable of producing a luminance of at least 3000 nit
(candelas per meter square) (7, 27).
Table II
Mammography View Box Measurements from Five Institutions (27)
Number
Average Minimum Maximum
of Panels
2
Luminance (cd/m ) 23 2,920 1,620 3,630
Illuminance (lux) 23 40 6 97
Color temperature (0K) 23 8,400 4,950 10,900

The illuminator surface should provide diffused light of uni-


form brightness, and variations in surface luminance should
Figure 10: Gradient versus optical density curves for Kodak ortho M film
in standard nd extended processing. (Tabar and Haus 24)
be gradual. The luminance level must be sufficient to illu-
minate areas of interest in the mammogram. Ideally, all
image quality in terms of film contrast of the processed radi- viewboxes should have the same color.
ograph (7, 25). The resulting film speed affects radiation
dose to the patient. Automatic film processor variables The contrast sensitivity of the eye (the ability to distinguish
include (1) processing cycle time, (2) temperature, (3) chem- small luminance differences) is greatest when surroundings
icals, (4) replenishment, (5) agitation, and (6) drying. are of about the same brightness as the area of interest.
Therefore, to see detail in a mammogram, it is important to
Digital Mammography reduce glare to a minimum, to avoid surface reflections, and
to reduce ambient light level to approximately that reaching
Digital image receptors for mammography have been under- the eye through the mammogram. Glare and reflections can
going development in recent years (28, 29). Small area dig- be reduced by locating illuminators away from bright sur-
ital detectors have been available for a few years on two dif- roundings such as windows, by turning off surrounding
ferent commercially available stereotactic breast localization viewboxes when not in use, and by using masks to cover
units (30). Most recent efforts for obtaining digital mammo-
grams include units which “stitch together ”several small
area images, scanning slot digital detectors, and large area
multi-detector arrays. In 2000 the first digital mammogra-
phy unit was approved for clinical use by the FDA and
became commercially available. Today several companies
have received FDA approval for digital mammography sys-
tems which are commercially available.

A main advantage of digital mammography compared to


screen-film mammography is that the image acquisition,
image display, and storage are decoupled (Figure 11) (31).
Clinical studies are underway to evaluate the clinical per-
formance of digital mammography systems (32).

Viewing Conditions

Much of the information contained in the mammographic


image cannot be visualized if proper consideration is not
given to viewing conditions. Mammograms should be inter-
preted under conditions that provide good visibility, comfort,
Figure 11: Characteristic curves for a digital mammography system. The
and minimal fatigue (26).
acquisition and display have two separate curves. The acquisition system
has a linear response to x-ray intensity whereas the display curve can be
adjusted by the viewer. (Haus and Yaffe 31)

Technology in Cancer Research & Treatment, Volume 1, Number 2, April 2002


Historical Developments in Mammography 125

low-density areas in the mammogram being examined. On October 1, 1994, the Mammography Quality Standards
Subdued lighting is preferred in the viewing room. It is also Act (MQSA) went into effect. To operate lawfully under
important to have a variable high output light source (with MQSA, mammography facility must be certified by the
appropriate masks) to view high-density areas on mammo- Food and Drug Administration (FDA). In order to be certi-
grams, and to make sure mammograms are properly fied, a facility must be accredited by a federally approved,
exposed. private non-profit or state accreditation body. The FDA has
approved the ACR, the states of Iowa, California, Arkansas
Quality Control and Texas as accreditation bodies.

In addition to the technical development advancements in With the implementation of MQSA, all mammography facil-
mammography, quality control programs have also had ities must apply to a federally-approved accreditation body,
major impact on the improvement in image quality of mam- undergo periodic review of its clinical images, have an annu-
mograms. The American College of Radiology (ACR) al survey by a medical physicist, and meet federally-devel-
Mammography Accreditation Program which began in late oped quality standards for personnel qualifications, quality
1987 has had a significant impact on improved and consis- assurance programs, and record keeping and reporting. The
tent image quality in mammography (7). The program facility must also undergo an annual inspection conducted by
requires (1.) questionnaire to be completed which includes federal or state personnel.
information such as credentials of staff, type of x-ray equip- Table IV
ment, image receptors, processing and quality control proce- Medical Physicist Tests (Annually) (7)
dures used, (2.) mammograms of two patients (one with fatty
breasts, the other with dense breasts), for evaluation by panel 1. Mammographic unit assembly evaluation
2. Collimation assessment
of radiologists, (3.) a breast phantom image for evaluation of 3. Evaluation of system resolution
medical physicists, (4.) radiation dose evaluation, and film 4. AEC system performance
processor control data. 5. Uniformity of screen speed
6. Artifact evaluation
7. Image quality evaluation
The ACR also developed quality control manuals with rec- 8. kVp accuracy and reproducibility
ommendations on quality control procedures and tests to be 9. Beam quality assessment
10. Breast exposure and AEC reproducibility
done on a routine basis (7). The manuals define the respon- 11. Average glandular dose
sibilities of the radiologist, medical physicist, and technolo- 12. Radiation output rate
gist for mammography quality control. There are separate 13. Measurement of viewbox luminance and
recommendations on quality control tests and frequency of room illuminance
tests to be performed by the medical physicist and technolo-
gist (Tables 3 &4) (7). Conclusion

Table III Today, due to technical advancements in x-ray equipment,


Technologists Tests (7)
image receptors (screen-film and digital), film processing
and viewing conditions together with implementation of
Daily
1. Darkroom cleanliness required quality control, it is possible to consistently produce
2. Processor quality control mammograms of high image quality.
3. Mobile unit quality control

Weekly
Acknowledgements
1. Screen cleanliness
2. View boxes and viewing conditions The author thanks Wende Logan-Young, MD, Breast Clinic
3. Phantom images of Rochester, New York, for providing the clinical mammo-
Monthly gram comparisons; Joel E. Gray, PhD, Trex Medical
1. Visual check list Corporation, Lorad Division, for the figure on the HTC grid.
Quarterly References &Footnotes
1. Repeat analysis
2. Analysis of fixer retention in film 1. Haus, A.G.: Physical principles and radiation dose in mammogra-
phy. Breast Carcinoma Current Diagnosis and Treatment, Masson
Semi-Annually Publishing pp 99-114. Eds. S. A. Feig and R. McLelland. NY (1983).
1. Darkroom fog 2. NCRP Report 85 Mammography-A users’s guide. National Council
2. Screen-film contact of Radiation Protection and Measurements. Bethesda, MD (1986).
3. Compression 3. Rothenberg, L. N., Haus, A. G.: Physicists in mammography-A his-
torical perspective. Medical Physics 22 (11) pp 1923-1934,
November(1995).

Technology in Cancer Research & Treatment, Volume 1, Number 2, April 2002


126 Haus

4. Recommended Specifications for New Mammography Equipment: 18. Sickles, E. A., Weber,. N.: High-contrast mammography with mov-
Screen-Film X-ray Systems, Image Receptors, and Film Processors. ing grid: Assessment of clinical utility Am. J. Roentgenol 146 pp
American College of Radiology, Reston, VA (1995). 1137-1139, (1986).
5. Barnes, G. T.: Mammography imaging physics: x-ray equipment 19. LaFrance, R. L., Gelskey, D. E., Barnes, G. T: A circuit modification
considerations. Syllabus, Categorical Course in Diagnostic that improves mammographic phototimer performance. Radiology
Radiology Physics: Physicial Aspects of Breast Imaging-Current 166 pp 773-776, (1988).
and Future Considerations, Radiological Society of North America 20. Thourson, T. L., Xeroradiography: SPIE 56 pp225-235 (1975).
pp 41-57, Eds. A. G. Haus and M. J. Yaffe, Oak Brook, IL (1999). 21. Wayrynen, R. E.: Fundamental aspects of mammographic receptors
6. Haus, A. G.: Mammography imaging physics: screen-film process- film process. Reduced Dose Mammography, Masson Publishing pp
ing and viewing condition considerations. In ref 5 pp 59-77. 521-528, Eds. W. W. Logan, E. P. Muntz, New York, NY, (1979).
7. American College of Radiology (ACR) Mammography Quality 22. Muntz, E. P, Welkowsky, M., Kaegi, E., Morsell, L, Wilkinson, E.,
Control Manual for Radiologists, Medical Physicists, and Techno- Jacobson, G.: Optimization of electrostatic imaging systems for min-
logists, American College of Radiology, Reston VA (1999). imum patient dose or minimum exposure in mammography.
8. Gabbay, E.: Mammography x-ray source, Syllabus, A Categorical in Radiology 127 pp 517 (1978).
Physics: Technical Aspects of Breast Imaging, Radiological Society 23. Kimme-Smith, C., Rothchild, P. A., Bassett, L. W., Gold, R. H.,
of North America pp 47-62, Eds. A. G. Haus and M. J. Yaffe, Oak Moler, C.: Mammographic film processor temperature, development
Brook, IL (1994). time, and chemistry: Effect on dose, contrast and noise. Am. J.
9. Haus, A. G., Metz, C. E., Chiles, J. T., Rossman, K.: The effects of Roentgenol 152 pp 34-40 (1989).
x-ray spectra from molybdenum and tungsten target tubes on image 24. Tabar, L., Haus, A. G.: Processing mammographic films: Technical
quality in mammography. Radiology pp 705-709 (1976). and clinical considerations. Radiology 173 pp 65-69 (1989).
10. Fewell T. R., Shupping, R. E.: Handbook of Mammographic X-ray 25. Haus, A. G., Jaskulski, S, M.: The Basics of Film Processing
Spectra. HEW Publication (FDA) (1979). Medical Imaging. Medical Physics Publishing, Madison, WI (1997).
11 Haus, A. G.: Dedicated mammographic x-ray equipment, screen- 26. Kimme-Smith, C., Haus, A. G., DeBruhl, N., Bassett, L. W.,: Effects
film processing systems and viewing conditions in mammography. of ambient light and viewbox luminance on the detection of calcifi-
Seminars in Breast Disease Vol 2 No 1, pp 30-54 March (1999). cations in mammography. Am. J. Roentgenol 168 pp 775-778
12. Hendrick, R. E., Burns, E. A.: Optimizing mammographic tech- (1997).
niques. In ref 5 pp79-89 27. Haus, A. G., Gray, J. E., Daly, T. R.: Evaluation of mammographic
13. Haus, A. G., Doi, K.,Chiles, J. T., Rossmann, K., Mintzer, R. A: The viewbox luminance, illuminance, and color. Medical Physics 20 pp
Effect of geometric and recording system unshapness in mammog- 819-821 (1993)
raphy, Invest Radiology 10 No 1 , pp 43-52 (1975). 28. Karellas, A, Harris, L. J., D’Orsi, C. J.: Small field digital
14. Haus, A. G., Cowart R. W., Dodd, G. D., Bencomo, J.: A method of Mammography with a 2048 x 2048 pixel charge coupled device.
evaluation and minimizing geometric unsharpness for mammo- Radiology 177 pp 288 (1990)
graphic x-ray units. Radiology 128 pp 775-778 (1978). 29. Yaffe, M. J.: Digital mammography. In Ref 5 pp 229-238
15. Kimme-Smith, C, Bassett, L. W., Gold, R. H.: Focal spot measure- 30. Hendrick, R. E., Parker, S. H.: Stereotactic imaging. In Ref 8 pp 263-
ments with pin hold and slit for microfocus mammography units. 274
Medical Physics 15 pp 293-298 (1988). 31. Haus, A. G., Yaffe, M. J.: Screen-film and digital mammography-
16. Barnes, G. T., Brezovich, I. A.: The intensify of scattered radiation image quality and radiation dose considerations.: Radiologic Clinics
in mammography. Radiology 126 pp 243-247 (1978). of North America Vol 38 No 4 pp 871-898 (2000)
17. Chan, H. P., Frank, P. H., Doi, K.et al: Development of ultra-high 32. Pisano, E. P.: Current clinical status of full-field digital
strip density (UHSD) grids: A new anti-scatter technique for mam- Mammography. In Ref 5 pp 239-247
mography. Radiology 154 pp 807-815 (1985).

Date Received: January 15, 2002

Technology in Cancer Research & Treatment, Volume 1, Number 2, April 2002

You might also like