Professional Documents
Culture Documents
ISSN 1533-0346
Volume 1, Number 2, April (2002)
©Adenine Press (2002)
Introduction
119
120 Haus
Table I
Technical Advances in Mammography
Year Development
Prior to 1969 Conventional tungsten target x-ray tubes with direct exposure industrial type films were used
1969 Dedicated mammographic unit with molybdenum target tube and compression cone introduced (CGR Senographe)
1971 Xeroradiography system introduced for mammography (Xerox)
1972 Screen-film system introduced for mammography (DuPont Lo-dose system)
1976 Rare earth screen-film system and special cassette introduced for mammography (Kodak Min-R system)
1977 Mammography x-ray unit for magnification with microfocal spot introduced (Radiological Sciences Inc.)
1978 Mammography unit with grid introduced (Philips)
1987 American College of Radiology Mammography Accreditation Program (ACR MAP) begins
1992 American College of Radiology Mammography Quality Control Manual for Radiologists, Radiologic
Technologists, and Medical Physicists, introduced
1994 The Food and Drug Administration (FDA) implements the Mammography Quality Standards Act (MQSA)
2000 Digital mammography system approved by the FDA for clinical use (GE Senographe 2000D)
as those produced by the 17.9 and 19.5 keV characteristic object-to-image receptor distances affect geometric blurring.
lines from the molybdenum target, provide high subject con- To minimize geometric blurring, the focal spot size and
trast for breasts of average thickness. When a 0.03mm object-to-image receptor distance should be minimized,
molybdenum filter is used, the spectrum is strongly sup- whereas focal spot-to-object distance should be maximized.
pressed at photon energies greater than 20 keV because of the Focal spot sizes and shapes for the dedicated and conven-
k-shell absorption edge of molybdenum at that energy (9-11). tional x-ray units used for mammography vary considerably
This allows more radiation from the characteristic lines to be (14, 15). Likewise, the focal spot-to-breast surface distances
used in image formation. Other tube target and filter combi- for different mammographic units vary.
nations used in dedicated mammographic units include a (1)
tungsten target with approximately 0.06-mm molybdenum Several years ago it was common to compress the breast
filter, (2) tungsten target with approximately 0.05-mm rhodi- directly on top of the mammographic cassette. The distance
um filter, and (3) rhodium target with approximately 0.025- between the chest wall edge of the breast and the screen-film
mm rhodium filter (Figure 3). It has been shown that molyb- combination was very small. Today, many mammographic
denum target with molybdenum filter combinations are procedures re performed with moving Bucky-type-grid.
appropriate for breasts less than 5 cm thick, and molybdenum With the grid in place, a gap of 1 to 2 cm in distance between
target with rhodium filters or rhodium target with rhodium the edge of the breast and the screen-film combination may
filters may be appropriate for breasts greater than 5 cm thick occur. The size of the focal spot, therefore, needs to be
(12). For mammography, kVp settings between 22 and 32 smaller in terms of limiting geometric resolution for a given
are used depending on breast tissue thickness and composi- focal spot-to-breast surface distance. The typical nominal
tion and exposure time. It has been shown that increasing focal spot sizes for todays mammography units are approxi-
kVp settings between 24 and 32 kVp, at comparable optical mately 0.3mm (non magnification) and 0.10mm (magnifica-
densities causes a slight decrease in lesion detections (8). tion technique) (4, 7).
Image Receptors
The xeroradiography system and the Xonics electronic radi- (2) phosphor particle size, (3) light-absorbing dyes and pig-
ography system are no longer commercially available. ments in the screen, and (4) screen-film contact. In recent
years, screen-film combinations for mammography have
Screen-Film Combinations used a single high-definition screen in contact with a single-
emulsion film.
Most single-screen, single-emulsion film combinations com-
monly used today have higher film contrast and require sig- The single screen is used as a back screen for mammography
nificantly lower radiation exposure than those used a few because x-ray absorption (and emission of screen light) is
years ago (Figure 7). It is interesting to note that today’s highest on the side of the screen where the x-rays enter. If the
screen-film combinations require approximately 50 times less screen were used s a front screen, x-ray absorption would be
radiation than direct-exposure films. Films currently used in higher in the plane of the screen that is the farthest distance
mammography usually have a single emulsion and are used in from the screen-emulsion contact surface. This causes
combination with a single back screen. Films are sensitized greater light spread (blur) than when x-ray absorption is high-
to match the spectral emission of the intensifying screen. est near the screen-emulsion contact surface, as is the case
when it is used as a back screen. Both parallax and crossover
The majority of mammographic images are produced with are eliminated in a single-back-screen configuration.
single intensifying screen used as a back screen in combina-
tion with single-emulsion film (Figure 8). Many mammo- Figure 9 shows modulation transfer function (MTF) curves
graphic screens incorporate phosphors containing metals for a direct-exposure film and five mammographic screen-
from the lanthanide series of elements such as terbium-acti- film combinations. Also shown is an MTF curve for a
vated gadolinium oxysulfide (Gd2O2S:Tb). Screens may screen-film combination used for conventional diagnostic
incorporate light absorbers in the phosphor that are used to radiology procedures. These curves show that screen-film
increase sharpness. Intensifying screens have a protective combinations used for mammography have much higher
overcoat to resist surface abrasion, are edge-sealed to mini- spatial resolution than do those used for conventional diag-
mize edge wear, and the base includes a backing layer to elim- nostic procedures (11)
inate screen curl. Mammographic screens consisting of ter-
bium-activated gadolinium oxysulfide material emit light in
the visible spectral region from 382 to 622 nm, although the
primary emission peak is in the green spectral region, 545 nm.
Film Processing
Viewing Conditions
low-density areas in the mammogram being examined. On October 1, 1994, the Mammography Quality Standards
Subdued lighting is preferred in the viewing room. It is also Act (MQSA) went into effect. To operate lawfully under
important to have a variable high output light source (with MQSA, mammography facility must be certified by the
appropriate masks) to view high-density areas on mammo- Food and Drug Administration (FDA). In order to be certi-
grams, and to make sure mammograms are properly fied, a facility must be accredited by a federally approved,
exposed. private non-profit or state accreditation body. The FDA has
approved the ACR, the states of Iowa, California, Arkansas
Quality Control and Texas as accreditation bodies.
In addition to the technical development advancements in With the implementation of MQSA, all mammography facil-
mammography, quality control programs have also had ities must apply to a federally-approved accreditation body,
major impact on the improvement in image quality of mam- undergo periodic review of its clinical images, have an annu-
mograms. The American College of Radiology (ACR) al survey by a medical physicist, and meet federally-devel-
Mammography Accreditation Program which began in late oped quality standards for personnel qualifications, quality
1987 has had a significant impact on improved and consis- assurance programs, and record keeping and reporting. The
tent image quality in mammography (7). The program facility must also undergo an annual inspection conducted by
requires (1.) questionnaire to be completed which includes federal or state personnel.
information such as credentials of staff, type of x-ray equip- Table IV
ment, image receptors, processing and quality control proce- Medical Physicist Tests (Annually) (7)
dures used, (2.) mammograms of two patients (one with fatty
breasts, the other with dense breasts), for evaluation by panel 1. Mammographic unit assembly evaluation
2. Collimation assessment
of radiologists, (3.) a breast phantom image for evaluation of 3. Evaluation of system resolution
medical physicists, (4.) radiation dose evaluation, and film 4. AEC system performance
processor control data. 5. Uniformity of screen speed
6. Artifact evaluation
7. Image quality evaluation
The ACR also developed quality control manuals with rec- 8. kVp accuracy and reproducibility
ommendations on quality control procedures and tests to be 9. Beam quality assessment
10. Breast exposure and AEC reproducibility
done on a routine basis (7). The manuals define the respon- 11. Average glandular dose
sibilities of the radiologist, medical physicist, and technolo- 12. Radiation output rate
gist for mammography quality control. There are separate 13. Measurement of viewbox luminance and
recommendations on quality control tests and frequency of room illuminance
tests to be performed by the medical physicist and technolo-
gist (Tables 3 &4) (7). Conclusion
Weekly
Acknowledgements
1. Screen cleanliness
2. View boxes and viewing conditions The author thanks Wende Logan-Young, MD, Breast Clinic
3. Phantom images of Rochester, New York, for providing the clinical mammo-
Monthly gram comparisons; Joel E. Gray, PhD, Trex Medical
1. Visual check list Corporation, Lorad Division, for the figure on the HTC grid.
Quarterly References &Footnotes
1. Repeat analysis
2. Analysis of fixer retention in film 1. Haus, A.G.: Physical principles and radiation dose in mammogra-
phy. Breast Carcinoma Current Diagnosis and Treatment, Masson
Semi-Annually Publishing pp 99-114. Eds. S. A. Feig and R. McLelland. NY (1983).
1. Darkroom fog 2. NCRP Report 85 Mammography-A users’s guide. National Council
2. Screen-film contact of Radiation Protection and Measurements. Bethesda, MD (1986).
3. Compression 3. Rothenberg, L. N., Haus, A. G.: Physicists in mammography-A his-
torical perspective. Medical Physics 22 (11) pp 1923-1934,
November(1995).
4. Recommended Specifications for New Mammography Equipment: 18. Sickles, E. A., Weber,. N.: High-contrast mammography with mov-
Screen-Film X-ray Systems, Image Receptors, and Film Processors. ing grid: Assessment of clinical utility Am. J. Roentgenol 146 pp
American College of Radiology, Reston, VA (1995). 1137-1139, (1986).
5. Barnes, G. T.: Mammography imaging physics: x-ray equipment 19. LaFrance, R. L., Gelskey, D. E., Barnes, G. T: A circuit modification
considerations. Syllabus, Categorical Course in Diagnostic that improves mammographic phototimer performance. Radiology
Radiology Physics: Physicial Aspects of Breast Imaging-Current 166 pp 773-776, (1988).
and Future Considerations, Radiological Society of North America 20. Thourson, T. L., Xeroradiography: SPIE 56 pp225-235 (1975).
pp 41-57, Eds. A. G. Haus and M. J. Yaffe, Oak Brook, IL (1999). 21. Wayrynen, R. E.: Fundamental aspects of mammographic receptors
6. Haus, A. G.: Mammography imaging physics: screen-film process- film process. Reduced Dose Mammography, Masson Publishing pp
ing and viewing condition considerations. In ref 5 pp 59-77. 521-528, Eds. W. W. Logan, E. P. Muntz, New York, NY, (1979).
7. American College of Radiology (ACR) Mammography Quality 22. Muntz, E. P, Welkowsky, M., Kaegi, E., Morsell, L, Wilkinson, E.,
Control Manual for Radiologists, Medical Physicists, and Techno- Jacobson, G.: Optimization of electrostatic imaging systems for min-
logists, American College of Radiology, Reston VA (1999). imum patient dose or minimum exposure in mammography.
8. Gabbay, E.: Mammography x-ray source, Syllabus, A Categorical in Radiology 127 pp 517 (1978).
Physics: Technical Aspects of Breast Imaging, Radiological Society 23. Kimme-Smith, C., Rothchild, P. A., Bassett, L. W., Gold, R. H.,
of North America pp 47-62, Eds. A. G. Haus and M. J. Yaffe, Oak Moler, C.: Mammographic film processor temperature, development
Brook, IL (1994). time, and chemistry: Effect on dose, contrast and noise. Am. J.
9. Haus, A. G., Metz, C. E., Chiles, J. T., Rossman, K.: The effects of Roentgenol 152 pp 34-40 (1989).
x-ray spectra from molybdenum and tungsten target tubes on image 24. Tabar, L., Haus, A. G.: Processing mammographic films: Technical
quality in mammography. Radiology pp 705-709 (1976). and clinical considerations. Radiology 173 pp 65-69 (1989).
10. Fewell T. R., Shupping, R. E.: Handbook of Mammographic X-ray 25. Haus, A. G., Jaskulski, S, M.: The Basics of Film Processing
Spectra. HEW Publication (FDA) (1979). Medical Imaging. Medical Physics Publishing, Madison, WI (1997).
11 Haus, A. G.: Dedicated mammographic x-ray equipment, screen- 26. Kimme-Smith, C., Haus, A. G., DeBruhl, N., Bassett, L. W.,: Effects
film processing systems and viewing conditions in mammography. of ambient light and viewbox luminance on the detection of calcifi-
Seminars in Breast Disease Vol 2 No 1, pp 30-54 March (1999). cations in mammography. Am. J. Roentgenol 168 pp 775-778
12. Hendrick, R. E., Burns, E. A.: Optimizing mammographic tech- (1997).
niques. In ref 5 pp79-89 27. Haus, A. G., Gray, J. E., Daly, T. R.: Evaluation of mammographic
13. Haus, A. G., Doi, K.,Chiles, J. T., Rossmann, K., Mintzer, R. A: The viewbox luminance, illuminance, and color. Medical Physics 20 pp
Effect of geometric and recording system unshapness in mammog- 819-821 (1993)
raphy, Invest Radiology 10 No 1 , pp 43-52 (1975). 28. Karellas, A, Harris, L. J., D’Orsi, C. J.: Small field digital
14. Haus, A. G., Cowart R. W., Dodd, G. D., Bencomo, J.: A method of Mammography with a 2048 x 2048 pixel charge coupled device.
evaluation and minimizing geometric unsharpness for mammo- Radiology 177 pp 288 (1990)
graphic x-ray units. Radiology 128 pp 775-778 (1978). 29. Yaffe, M. J.: Digital mammography. In Ref 5 pp 229-238
15. Kimme-Smith, C, Bassett, L. W., Gold, R. H.: Focal spot measure- 30. Hendrick, R. E., Parker, S. H.: Stereotactic imaging. In Ref 8 pp 263-
ments with pin hold and slit for microfocus mammography units. 274
Medical Physics 15 pp 293-298 (1988). 31. Haus, A. G., Yaffe, M. J.: Screen-film and digital mammography-
16. Barnes, G. T., Brezovich, I. A.: The intensify of scattered radiation image quality and radiation dose considerations.: Radiologic Clinics
in mammography. Radiology 126 pp 243-247 (1978). of North America Vol 38 No 4 pp 871-898 (2000)
17. Chan, H. P., Frank, P. H., Doi, K.et al: Development of ultra-high 32. Pisano, E. P.: Current clinical status of full-field digital
strip density (UHSD) grids: A new anti-scatter technique for mam- Mammography. In Ref 5 pp 239-247
mography. Radiology 154 pp 807-815 (1985).