ATE-100 Handheld

Pressure Calibrator
FEATURES The Quick-Select Pressure Module
• Interchangeable pressure and Quick-Select pressure modules are
temperature modules available in full-scale ranges from
• Pressure measurement accuracies of 0.25 inches H2O through 10,000 psi.
±0.025,0.05 and 0.10%, or .06/.07% Gauge, vacuum, absolute, differential or
compound pressure ranges are available.
• Pressure ranges from 0.25 in.H2O to Each base unit can accommodate two
10,000 psi pressure sensor modules simultaneously.
• Supports most standard RTD probes Select from multiple modules which can be
and thermocouples interchanged in seconds without the use of
• Min/max, tare, flow, leak, programmable tools. When installed in the base unit, a
damping, percent function, trip detect, pressure module becomes an integral part
all standard of the measurement system, thereby pro-
• High static DP measurement capability tecting the sensor from damage. Any com-
• Optional: FM Approval for Class I, bination of Quick-Select pressure modules
Div. 1, Groups A, B, C & D hazardous may be used or interchanged as needed
areas, datalogging, hi/low alarm relays in the lab or field.

The AQS-1 Module


The Ashcroft® ATE-100 handheld system The AQS-1 Quick-Select low pressure
LOOK FOR THESE MARKS
introduces a new era in pressure calibration module is based on a proprietary, Ashcroft
ON OUR PRODUCTS standards. This unique instrument meas- engineered, variable-capacitance sensor.
ures pressure, temperature, flow, vacuum, This sensor is available in ranges from
leak, current, voltage and change-of-state 0/0.25 to 0/200 inches H2O. AQS-1 Quick-
for switch testing. A fully equipped ATE- Select pressure modules are available in
100 system provides all calibration capa- accuracies from ±0.1% to ±0.06% of span.
bilities needed in virtually all pressure These Quick-Select pressure modules are
measurement and test applications. available for gauge, differential and com-
Each handheld system consists of a pound measurement requirements.
base unit and one or two Quick-Select™ The variable-capacitance sensing ele-
pressure modules. An array of pressure ment in the AQS-1 incorporates a micro-
modules is available to cover ranges from machined silicon diaphragm. Silicon pro-
0.25 inches H2O through 10,000 psi. vides optimum repeatability with minimal
Modules can be inter-changed quickly and hysteresis. The sensor is extremely stable,
easily without the use of tools. In addition since the sputtered metal capacitance
to pressure measurement, Quick-Select plates are molecularly bonded to the sili-
interface modules are also available to con. This design allows for the elimination
allow the ATE-100 to measure tempera- of epoxies and other types of bonding and
ture with common RTDs or thermocouple adhesive agents from the sensor assembly.
devices. The AQS-1 sensor is also extremely
For the most demanding applications, rugged. It can withstand extreme overpres-
the ATE-100 can be optionally con-figured sure without significant error or calibration
in an intrinsically safe “FM approved” ver- offset. In addition, the sensor is capable of
sion, for use in Class 1, Div. 1 hazardous withstanding significant shock and vibra-
environments. In areas where EMI/RFI tion without damage or loss of measure-
immunity is required, an optional “CE ment output integrity.
Mark” version is also available.
With a broad variety of ranges and fea- The AQS-2 Module
tures, the ATE-100 is one of the most pow- The AQS-2 Quick-Select pressure mod-
erful, versatile and useful test and meas- ule incorporates a piezoresistive, micro-
urement instruments available today. machined silicon sensor. These sensors
are available in ranges from 0/100 inches
H2O through 0/10,000 psi and are also
offered in gauge, absolute, vacuum and
compound variants.

All specifications are subject to change without notice. Ashcroft Inc., 250 East Main Street, Stratford, CT 06614 USA
All sales subject to standard terms and conditions. Tel: 203-378-8281 • Fax: 203-385-0408
BULLETIN ATE-100 © Ashcroft Inc. 2008 07/10 email: info@ashcroft.com • www.ashcroft .com
ATE-100 Handheld
Pressure Calibrator
316 SS sensor isolation is provided on PRODUCT SPECIFICATIONS BASE UNIT PHYSICAL SPECIFICATIONS
all AQS-2 modules. AQS-2 Quick-Select Ashcroft® Dimensions
pressure modules are available in a variety Model No.: ATE-100 7.88 in. (L) x 4.24 in. (W) x 3.25 in. (H)
of accuracies from ±0.1% of span through Dimensions: 7.88 in. (L) x 4.24 in. (W) x 3.25 in. (H) Weight
±0.025% of span. Weight: Max. 2.2 lbs. w/2 pressure modules Max. 2.2 lbs. w/2 pressure modules installed
installed Case Material
Also Measure Temperature Case Material: High impact ABS High impact ABS
The AQS-RT1, AQS-RT2 and AQS-TC1 Sensor Module Sensor Module Capacity
interface modules allow the ATE-100 to Capacity: 2 bays for Ashcroft AQS “Quick 2 bays for Ashcroft AQS “Quick Select®” sensor
measure temperature with common RTD Select®” sensor modules modules
Display
or thermocouple temperature measure- Display: 2 line LCD, 0.37 in. height per line. 2 line LCD, 0.37 in. height per line. Can display
ment devices. Can display simultaneous readings simultaneous readings from 2 modules
from 2 modules Electrical Connection
The Base Units Electrical Miniature recessed banana jacks (one set of test
The base unit provides access to the Connection: Miniature recessed banana jacks (one leads provided with each ATE-100)
complete capabilities of the ATE-100 cali- set of test leads provided with each
bration system. These capabilities extend ATE-100) BASE UNIT OPERATING SPECIFICATIONS
far beyond those of precision pressure Operating Operating Temperature Range
Temp. Range: Standard: 32° to 120°F Standard: 32° to 120°F
measurement. The base unit contains a
Optional: –4 to 120°F Optional: –4 to 120°F
wide variety of application-specific
Storage Storage Temperature
firmware as well as the ability to measure Temperature: –4° to 158°F –4° to 158°F
both current and voltage inputs. In addi- Update Rate
Update Rate: 130 ms (nominal) with one sensor
tion, this unit has the ability to interface installed 130 ms (nominal) with one sensor installed
the system with a computer, data acquisi- Resolution: ±0.002% of span, 60,000 count Resolution
tion system or dumb terminal via the stan- (max) ±0.002% of span, 60,000 count (max)
dard RS-232 interface. Warm-Up: 5 minutes for rated accuracy Warm-Up
Each base unit can accommodate two 5 minutes for rated accuracy
Electrical
Quick-Select pressure modules simultane- Electrical Measurements
Measurements: 0-20mA or 0-30 Vdc
ously. In addition, the modules can be 0-20mA or 0-30 Vdc
Input: (volts) Accuracy Input (volts) Accuracy
interchanged quickly and easily for on- 0/10 Vdc ±0.025% fs 0/10 Vdc ±0.025% fs
the-fly system configuration as application 0/30 Vdc ±0.10% fs 0/30 Vdc ±0.10% fs
requirements change. 0-20mA ±0.03% fs 0-20mA ±0.03% fs
Temperature Temperature Effect Electrical Measurement:
Effect Electrical ±0.001% of span per °F over comensated range
Measurement: ±0.001% of span per °F over comen- Serial Interface
sated range Type: RS-232
Serial Baud Rate: 300, 1200, 2400, 9600
Interface: Type: RS-232 Serial Interface
Baud Rate: 300, 1200, 2400, 9600 Both Quick-Select pressure modules and base unit
Serial electronics can be calibrated in the field.
Interface: Both Quick-Select pressure modues Options
and base unit electronics can be cali- Datalogging with Hi-Lo Relay Feature – Datalogging
brated in the field. manually or automatically stores up to 715 meas-
ured values for upload to PC. Includes upload utility
OPTIONAL FEATURES software. Hi-Lo relay feature allows programming of
• Datalogging with Hi-Lo Relay Feature – setpoints for activation of alarms or control valves.
Datalogging manually or automatically stores up to (Hi-Lo not available with FM approval.)
715 measured values for upload to PC. Includes Enhanced LCD – For –4 to 120°F operating range
upload utility software. Hi-Lo relay feature allows FM Approval – Class 1, Div. 1, Groups A,B,C & D
programming of setpoints for activation of alarms (Not available with CE Mark)
or control valves. (Hi-Lo not available with FM CE Mark – EMI/RFI immunity rating (not available
approval.) with FM approval)
• Enhanced LCD Power Requirements
For –4 to 120°F operating range Standard: (2) 9 Vdc Alkaline Batteries (provides up
• FM Approval – Class 1, Div. 1, Groups A,B,C & D to 30 hours operation per set)
(Not available with CE Mark) Optional: 110 or 220 Vac transformer to power
• CE Mark calibrator from line power
EMI/RFI immunity rating (not available with FM Certification
approval) N.I.S.T. Traceable certification document provided
for base display unit and sensor modules

All specifications are subject to change without notice. Ashcroft Inc., 250 East Main Street, Stratford, CT 06614 USA
All sales subject to standard terms and conditions. Tel: 203-378-8281 • Fax: 203-385-0408
BULLETIN ATE-100 © Ashcroft Inc. 2008 07/10 email: info@ashcroft.com • www.ashcroft .com
ATE-100 Handheld
Pressure Calibrator
PRESSURE SENSOR MODULE Media Compatibility
SPECIFICATIONS Any medium compatible with 316 SS isolation.
Optional: Cleaned for Oxygen Service
AQS-1 MODULE Overpressure Capability
Pressure Types 200% for ranges up to 1000 psi
Gauge, differential & compound 150% for ranges over 1000 psi
Available Ranges Process Connection
(See Chart) Standard: 1⁄8 NPT female
Available Accuracies Optional: 1⁄8 NPT female with flush port
±0.06 (0/1-0/200 in. H2O), ±0.07 (0/0.25-0/0.5 in. Welded VCR fitting with standard finish
H2O) or 0.1% of Span Other Options
Compensated Temperature Range FM Approved (for use with FM approved base unit)
20°F to 120°F CE Mark (for use with CE Mark rated base unit)
Temperature Effect
±.004% of Span per °F over compensated range TEMPERATURE SENSOR MODULE
(from reference temperature range of 70° ±3°) SPECIFICATIONS
Repeatability
±0.01% of span (range 0/1 in. H2O or higher) AQS-RT MODULE (RTD)
±0.02% of span (ranges below 0/1 in. H2O) AQS-RT1 and AQS-RT2 interface modules allow the
Sensitivity ATE-100 to measure temperature with an RTD:
±0.002% of span (typical) AQS-RT1: Accommodates Pt100, Ni120, Cu120 and
Media Compatibility other common 2, 3 or 4 wire probes with resistance
Clean, dry, non-conductive, non-corrosive gas outputs of 400 ohms or less.
Under/Overpressure Capability AQS-RT2: Accommodates Pt1000 and other com-
–15 to 50 psi mon 2, 3 or 4 wire probes with resistance outputs of
Maximum Static (line) Pressure 4000 ohms or less.
100 psi Selectable Units of Measure
Process Connection °C, °F, °K, °R and ohms
Standard: 1⁄8 NPT female Input Receptacle
Other Options Accepts TA4F type RTD connector
FM Approved (for use with FM approved base unit) RTD Probes Available
CE Mark (for use with CE Mark rated base unit) Pt-100 probes, 6˝ or 12˝ length, with or without
handle. DIN Class A accuracy. Includes mating TA4F
AQS-2 MODULE connector. Consult factory for details and availability.
Pressure Types Options
Gauge, absolute, compound & vacuum FM Approved (for use with FM approved base unit)
Available Ranges (CE Mark not available)
(See Chart)
Available Accuracies AQS-TC1 MODULE (Thermocouple)
±0.025, 0.05 or 0.1 % of Span (±0.025 & 0.05% not The AQS-TC1 interface module allows the ATE-100
available on 0/10,000 psi range) to measure temperature with a thermocouple:
Compensated Temperature Range Compatibility
20°F to 120°F Programmed to provide direct temperature readout
Temperature Effect from types J, K, T, E, R, S, B & N thermocouples or
Standard: ±.004% of Span per °F over the compen- direct millivolt readout from any thermocouple.
sated range (from reference temperature range of Reference Junction
70° ±3°) Automatic internal or manual external
Optional: No additional error due to temperature Resolution
over the compensated range Automatic or manually selectable, up to .01°
Repeatability Units of Measure
±0.01% of span Selectable; °C, °F, °K, °R and millivolts
Sensitivity Receptacle
±0.002% of span (typical) Accepts “miniature thermocouple connector”,
Omega® type SMP
Options
FM Approved (for use with FM approved base unit)
(CE Mark not available)

All specifications are subject to change without notice. Ashcroft Inc., 250 East Main Street, Stratford, CT 06614 USA
All sales subject to standard terms and conditions. Tel: 203-378-8281 • Fax: 203-385-0408
© Ashcroft Inc. 2008 07/10 email: info@ashcroft.com • www.ashcroft .com
ATE-100 Handheld
Pressure Calibrator
DIMENSIONS STANDARD RANGES ACCESSORIES
AQS-2 AQS-1 POWER SUPPLIES
.43/.58 4.24 3.25
psi ˝H2O Other Engineering Units** 831X016-01 AC adapter to run system off line
(gauge & (gauge/ power (110 Vac, 60 Hz)
absolute differential 831X016-02 AC adapter to run system off line
pressure) pressure) power (100 Vac, 60 Hz)
ATE-100
831X016-03 AC adapter to run system off line
5 0.25* power (220 Vac, 50 Hz)
10 0.5* psi
831X019-01 AC adapter to run system off line
in. H2O
15 1.0* power (220 Vac, 50 Hz, for CE
7.88 in. Hg
20 2.0* ftSW mark)
25 3.0* Bar ELECTRICAL ACCESSORIES
30 5.0* mBar LPSIIa Loop Power Supply (powers 4-
50 10* kPa 20mA current loops)
60 15* mPa SM-1 Voltage Adapter (Converts voltages
100 25* mmHg of 10/250 Vac and 10/400 Vdc to
150 50* cmH2O levels acceptable to calibrator elec-
3.25 1.73
200 100* mmH2O
tronics; used to test “live” switches)
kg/cm2
250 150* CARRYING, TRANSPORT & PROTECTIVE CASES
User Selectable
300 200* 864X071-01 Contoured Form-Fitting Protective
500 **Note: Engineering units Instrument Case with Shoulder
600 identified above are Strap
1000 accessible through the 864X068-01 Heavy Duty, Watertight, Instrument
1500 unit select feature. How-
ever, readout will default Carrying/Transport Case*
2000 to the primary unit of 864X069-01 Soft Carrying Case*
2500 measure on start-up.
3000 Sensor modules scaled in *Holds base unit, 5 sensors and tools
5000 primary units other than
in. H2O (HQS-1) or psi
6000 (HQS-2) are also available. HOW TO ORDER
7500 Consult factory. Base Display Unit
10,000 1) Specify Model: ATE-100
vacuum 2) Specify Options: (Datalogging, Enhanced LCD,
5 FM Approval, CE Mark)
10 Sensor Modules
15 3) Type (AQS-1 or AQS-2)
compound 4) Pressure Range and Unit of Measure
±5 ±0.125* (see range chart)
±10 ±0.25* 5) Pressure Type (see specifications)
±0.5* 6) Accuracy (see specifications)
±15
7) Specify Options
–15/+30 ±1.0*
a) “zero temperature error over compensated
–15/+60 ±1.5* range” (HQS-2 only)
±2.5* b) Optional fitting (see specifications)
±5.0* c) Clean for Oxygen Service
±7.5* (AQS-2 only)
±12.5* d) FM approval or CE Mark
±25* Temperature Interface Module
±50* 8) Type (AQS-RT1, AQS-RT2 or AQS-TC1)
±75* a) Specify FM approval if required
±100* 9) RTD Probe Type (when required. Consult factory
for probe P/N)
* Non-isolated, for clean dry gas only Accessories
10) Specify required accessories

All specifications are subject to change without notice. Ashcroft Inc., 250 East Main Street, Stratford, CT 06614 USA
All sales subject to standard terms and conditions. Tel: 203-378-8281 • Fax: 203-385-0408
© Ashcroft Inc. 2008 07/10 email: info@ashcroft.com • www.ashcroft.com