Professional Documents
Culture Documents
12 t h October 2007
Place des Maïeurs 2 B-1150 Brussels +32 2 772 2225 tel +32 2 772 2329 fax edma@edma-ivd.be www.edma-ivd.be
Page 1 of 4
the laboratory (validation of the the presence of many factors intrinsic to
measurement procedure including the individual laboratory (which need to
preanalytical steps, and any other be taken into account when establishing a
factors relevant for the use in the SOP’s), manufacturers of devices cannot
laboratory, not validation of the provide generic SOP’s suitable for use
device) within multiple laboratories.
2) The laboratory is using a device
which has not been CE marked, or Laboratories should in any case ensure
is using a CE marked device for a that the final established SOP’s are in
use which was not intended by its agreement with the manufacturer’s
manufacturer – In this case, the recommendations, for instance as regards
specific use of the device has not to the use of reagents, operator training
been previously validated, thus it is or respecting environmental conditions for
the responsibility of the laboratory to operation (temperature, vibrations etc.)
provide a validation not only of the
procedures but also, insofar as This does not preclude manufacturers
possible, of the device itself, helping laboratories to establish their
according to the relevant requirements SOP’s, either as part of their usual after
of the IVD directive. sales service or as additional contractual
arrangements with the laboratories. Such
Depending on the nature of the device, an arrangement would be a collaborative
the nature of the measurement procedure effort between the manufacturer and the
and the relationship between the laboratory as it would be impossible for a
manufacturer and the laboratory, it may manufacturer to establish SOP’s for a
occur in some cases that a manufacturer laboratory on its own. However,
decides to assist the laboratory in the regardless of any assistance provided by
validation of a new use for his device. the manufacturer, the responsibility for
However the manufacturer is in no way the establishment of laboratory SOP’s lies
obliged to do so. with the laboratory.
Place des Maïeurs 2 B-1150 Brussels +32 2 772 2225 tel +32 2 772 2329 fax edma@edma-ivd.be www.edma-ivd.be
Page 2 of 4
maintenance should be made available reference population used to establish
to the user. those values.
• Any responsibilities of the user (e.g.
decontamination of instruments prior However it is the responsibility of the
to servicing) laboratory to determine whether the
reference population is comparable to the
It should be noted that the report, while population from which the samples
clear, need not be exhaustive. However, measured in the laboratory originate.
all information which may affect the Should the two populations be
performance of the device should be significantly different, the laboratory may
made available to the user. decide to establish new reference values.
This can be done either experimentally or
Following these maintenance operations if available via appropriate bibliographical
the manufacturer or his authorized references.
subcontractor should perform a functional
validation of the instrument in order to Metrological Traceability of
ensure it will meet stated performance Measurement Results 2
criteria as specified on the report is prior
to use by the laboratory. Adequate Where appropriate, the manufacturer will
documentation will be presented to the provide the laboratory with information, on
laboratory to demonstrate that this is the the traceability of the values assigned to
case. trueness controls and calibrators provided
to the user. This information includes the
It should be noted that information on the reference material of higher order to
internal calibrations of the device (e.g. which values are traceable.
dispensing of volumes by automatic
pipettes, temperature control, flow rates The final steps in the metrological
of pumps) need not always be provided, traceability chain are performed in the
so long as the overall performance of the laboratory. This usually includes the
device is unaffected. calibration of the measurement system
and measurement procedure on a patient
It should also be noted that the sample. This represents the laboratories
manufacturers responsibility in terms of important contribution to the
maintenance is to ensure and establishment of a full chain of
demonstrate that instruments meet the metrological traceability.
stated performance criteria as specified in
the maintenance report after For details on the establishment of a
maintenance. It is the responsibility of the chain of metrological traceability, see EN
laboratory to demonstrate to accreditation ISO 17511: “In vitro diagnostic medical
bodies that they are meeting the devices – Measurement of quantities in
maintenance requirements necessary for biological samples – Metrological
accreditation. traceability of values assigned to
calibrators and (trueness) control
Following maintenance the laboratory may materials.”
decide to re-validate the procedures
performed with that instrument, if he
deems it necessary.
Place des Maïeurs 2 B-1150 Brussels +32 2 772 2225 tel +32 2 772 2329 fax edma@edma-ivd.be www.edma-ivd.be
Page 3 of 4
Uncertainty of Measurement Results It should be noted, that because of the
nature of the biological samples, and in
Upon request, manufacturers should particular because often the analytes are
provide the laboratories with the poorly defined, the uncertainty of results
uncertainty data associated with the may be large when compared with the
calibrators and trueness control materials reproducibility of results. Similarly the
provided to the user. uncertainty associated with biological
samples is much greater than that
It is the responsibility of the laboratory to associated with direct measurements of
do the following: physical properties, such as weight,
temperature or distance.
• Calculate the final measurement
uncertainty of the result For further guidance on this issue, please
• Decide whether and how to present see the EDMA position paper on the
that measurement uncertainty to the estimation of uncertainty of measurement
clinician. in medical laboratories.
References:
Place des Maïeurs 2 B-1150 Brussels +32 2 772 2225 tel +32 2 772 2329 fax edma@edma-ivd.be www.edma-ivd.be
Page 4 of 4