Instruction Manual Dear Customer,

LCD DESCRIPTION SPECIFICATIONS


Thank you for choosing a Hanna Instruments Product.
HI 991300 • HI 991301 Please read this instruction manual carefully before using the
HI 991300 HI 991301
Waterproof instrument. If you need additional technical information, do not
hesitate to e-mail us at tech@hannainst.com. Range pH 0.00 to 14.00 pH 0.00 to 14.00 pH
pH/EC/TDS/Temperature Meters EC 0 to 3999 µS/cm 0.00 to 20.00 mS/cm
with Advanced Features TDS 0 to 2000 ppm (mg/l) 0.00 to 10.00 ppt (g/l)
Temperature 0.0 to 60.0°C / 32.0 to 140.0°F
PRELIMINARY EXAMINATION Resolution pH 0.01 pH 0.01pH
EC 1 µS/cm 0.01 mS/cm
Remove the instrument from the packing material and examine TDS 1 ppm (mg/l) 0.01 ppt (g/l)
it carefully to make sure that no damage has occurred during Temperature 0.1°C / 0.1°F
shipment. If noticeable damage is evident, notify your dealer. Accuracy ±0.01 pH
Note: Save all packing material until you are sure that the (@20°C/68°F) ±2 % F.S. (EC / TDS)
instrument functions correctly. All defective items must be ±0.5°C / ±1°F
returned in the original packing together with the pH Calibration Automatic, 1 or 2 point, choose between 2 sets
supplied accessories. of buffers (pH 4.01/7.01/10.01 or
1. Automatic Temperature Compensation indicator Nist pH 4.01/6.86/9.18)
2. Battery percentage (visible at power up) EC/TDS Calibration Automatic, 1 point at
3. Secondary display 1413 µS/cm 12.88 mS/cm
GENERAL DESCRIPTION 4. Low battery indicator or 1382 ppm (CONV=0.5) or 6.44 ppt (CONV=0.5)
5. Stability indicator or 1500 ppm (CONV=0.7) or 9.02 ppt (CONV=0.7)
www.hannainst.com
HI 991300 and HI 991301 have been designed to offer you 6. Calibration indicator Temperature pH: Automatic
the combination of pH, conductivity, total dissolved solids and 7. Selectable temperature unit
WARRANTY Compensation EC/TDS: Automatic with
temperature measurements.
8. Primary display β selectable from 0.0 to 2.4 %/°C with 0.1 increments
To increase precision, two models are available, with different
All Hanna Instruments meters are warranted for two years 9. Measuring unit for primary display TDS Conversion Factor Selectable from 0.45 to 1.00 with
conductivity ranges, for applications from purified to brackish waters.
against defects in workmanship and materials when used for 0.01 increments (default 0.50)
All operations and settings, including calibration buffers and
their intended purpose and maintained according to instructions. Probe (included) HI 1288 pH/EC/TDS/temperature
The probes are warranted for a period of six months.
temperature scale selections, are made through only 2 buttons. ACCESSORIES
The housing is waterproof and rated for IP 67 conditions. with DIN connector and 1 m (3.3’) cable
This warranty is limited to repair or replacement free of charge. The supplied multi-parameter probe, HI 1288, includes pH, EC/TDS Battery Type 3 x 1.5V AAA
HI 1288 pH/conductivity probe with built-in temperature
Damage due to accidents, misuse, tampering or lack of and temperature in one convenient, rugged probe. In addition, to Battery Life Approximately 500 hours of continuous use
sensor, DIN connector and 1 m (3.3’) cable
prescribed maintenance are not covered. ensure against interference from transient electrical noise to pH, a HI 7004M pH 4.01 buffer solution, 230 mL bottle Auto-off After 8 minutes of non-use
If service is required, contact the dealer from whom you solid-state preamplifier is integrated into the probe.
HI 7006M pH 6.86 buffer solution, 230 mL bottle Environment 0 to 50°C (32 to 122°F);
purchased the instrument. If under warranty, report the model Other user selectable features include different TDS factors from RH max. 100%
number, date of purchase, serial number and the nature of the 0.45 to 1.00, and a range of temperature coefficients (β) from 0.0 HI 7007M pH 7.01 buffer solution, 230 mL bottle
Dimensions 152 x 58 x 30 mm (6.0 x 2.3 x 1.2”)
problem. If the repair is not covered by the warranty, you will to 2.4% for greater consistency and reproducibility. HI 7009M pH 9.18 buffer solution, 230 mL bottle
be notified of the charges incurred. If the instrument is to be Weight 205 g (7.2 oz.)
Each meter is supplied with: HI 7010P pH 10.01 buffer solution, 230 mL bottle
returned to Hanna Instruments, first obtain a Returned Goods
• HI 1288 pH/EC/TDS probe with built-in temperature HI 7030M 12.88 mS/cm solution, 230 mL bottle
Authorization Number from the Customer Service department
sensor, DIN connector and 1 m (3.3’) cable
and then send it with shipment costs prepaid. When shipping HI 7031M 1413 µS/cm solution, 230 mL bottle
any instrument, make sure it is properly packaged for complete • Calibration and cleaning solution sachets
HI 7032M 1382 ppm solution, 230 mL bottle Hanna Instruments reserves the right to modify the design,
protection. • 3 x 1.5V AAA batteries
HI 7038M 6.44 ppt solution, 230 mL bottle construction and appearance of its products without advance
• Instruction manual. notice.
All rights are reserved. Reproduction in whole or in part is prohibited HI 70442M 1500 ppm solution, 230 mL bottle
without the written consent of the copyright owner. IST991300Z 01/11
OPERATIONAL GUIDE pH MEASUREMENT & CALIBRATION EC MEASUREMENT & CALIBRATION METER SETUP
First time you use the instrument open the battery compartment • Make sure the meter has been calibrated before use. • Place the probe in the sample to be tested. Use plastic beakers While in measurement mode, press and hold the ON/OFF/MODE
and put the batteries inside observing the polarity. • If the probe is dry, soak it in HI 70300 storage solution for or containers to minimize any electromagnetic interference. button until “TEMP” and the current temperature unit are
To connect the probe 30 minutes to reactivate it. • Tap the probe lightly on the bottom of the container to remove displayed on the lower LCD (eg. “TEMP °C“). Then:
• Submerge the probe in the sample to be tested while stirring it air bubbles that may be trapped inside the tip.
With the meter turned off, connect the HI 1288 probe to the DIN To select the temperature unit (°C / °F)
gently. Wait until the “Not Stable“ tag on the LCD is turned off. • Wait for a few minutes for the temperature sensor to reach
socket on the bottom of the meter by aligning the pins and thermal equilibrium (i.e. until the “Not Stable“ tag disappears). • Press the SET/HOLD button to change the temperature unit,
• The pH value automatically compensated for temperature is
pushing in the plug. Tighten the nut to ensure a good connection. • The LCD displays the EC or TDS value (automatically compen- and then press the ON/OFF/MODE button until the meter
displayed on the primary LCD, while the secondary LCD displays
Remove the protective cap from the probe before taking any the sample temperature. sated for temperature) on the primary LCD, while the secondary returns to normal measurement mode.
measurements. • If measurements are taken in different samples successively, LCD displays the temperature of the solution. To change the type of calibration buffer set,
To turn the meter ON and check the battery status rinse the probe tip thoroughly to eliminate cross-contamination. EC calibration • While in pH measurement mode, after setting the temperature
Press the ON/OFF/MODE button until the display lights up. At After cleaning, rinse the probe tip with some deionized water • From EC or TDS normal mode, press and hold the ON/OFF/ unit press the ON/OFF/MODE button again to show the
start-up, all the LCD segments are displayed for 1 second, then the and some of the sample to be measured. MODE button until “OFF“ on the secondary display is replaced current buffer set: “pH 7.01 BUFF” (for standard buffer set:
percent indication of the remaining battery life is displayed for pH calibration by “CAL“. Release the button.
4.01/7.01/10.01) or “pH 6.86 BUFF” (for NIST buffer set:
another second (E.g. “100 % BATT“). The meter then enters the • While in pH measurement mode, press and hold the • The meter enters the calibration mode. Immerse the probe in
the HI 7031 calibration solution (“µS 1413 USE“) for 4.01/6.86/9.18). Change the set with the SET/HOLD button,
normal measuring mode. ON/OFF/MODE button until “OFF“ on the secondary display then press ON/OFF/MODE to return to normal mode.
HI 991300 or in HI 7030 calibration solution (“mS 12.88
is replaced by “CAL“. Release the button. Place the sensor
Note: If the display needs to be checked, keep the ON button USE“) for HI 991301. To change the EC/TDS conversion factor (CONV) and the
into the first calibration buffer.
pressed while turning the meter on. The meter will display • If the standard value is recognised ''REC'' is displayed until the temperature compensation coefficient (BETA)
• The meter enters the calibration mode, displaying “pH 7.01 reading is stable and calibration is accepted.
all segments as long as the button is pressed. USE” (or “pH 6.86 USE” if the NIST buffer set was selected). • While in EC or TDS measurement mode after setting the
To select the measurement range • The LCD will display “OK” for 1 second and return to normal temperature unit press the ON/OFF/MODE button again to
• For a single-point calibration place the probe in any buffer from measurement mode.
While in normal measurement mode, press the SET/HOLD button the selected buffer set (E.g. pH 4.01, 9.18 or 10.01). The meter show the current conversion factor (eg. “0.50 CONV“), then
• If the standard is not recognised or the slope is out of accepted
quickly to select pH, EC or TDS value on the primary LCD, while will automatically recognize the buffer value. range ''---Wrong'' is display. Change the calibration solution press SET/HOLD to change the value.
temperature will be simultaneously displayed on the secondary • If the buffer is not recognized or the calibration offset is out of or the electrode or press any key to exit calibration. • Press the ON/OFF/MODE button to show the current temperature
LCD (E.g. “pH 5.73“ and “22.5°C“). the accepted range ''---WRONG'' is displayed. • When the calibration procedure is completed, the “Calibrated“ compensation coefficient b (eg. “2.1 BETA“), then press
To freeze the display • If the buffer is recognised ''REC" is displayed until the reading is tag is turned on. SET/HOLD to change the temperature coefficient value.
stable and the calibration is accepted. Notes: • Beta should be set to 1.9 during calibration.
While in measurement mode, press and hold the SET/HOLD • Press the ON/OFF/MODE button to return to normal opera-
• If using pH 7.01 (or pH 6.86 from NIST buffer set), after • Since there is a known relationship between the EC
button until “HOLD“ appears on the secondary display and the tion.
acceptance of the buffer press any key to exit. ''OK1'' message is and TDS reading, it is not necessary to calibrate the
reading will be frozen on the LCD (E.g. “pH 5.73 HOLD“). Press displayed and meter returns to pH measurment mode. meter in TDS. If the conversion factor is either 0.5 or
any button to return to normal mode. • If using 4.01 or 10.01 (9.18) buffer the ''OK1'' message is 0.7, the meter will allow a direct calibration in TDS BATTERY REPLACEMENT
To turn the meter OFF displayed and meter returns to pH measurment mode. by using the Hanna calibration solutions. The meter is supplied with batteries.
While in normal measurement mode, press the ON/OFF/MODE • For a two-point calibration, proceed as for the single point To exit calibration and reset default values The meter displays the remaining battery percentage when turned
button. “OFF“ will appear on the secondary display. Release the calibration selecting 7.01/6.86 pH buffer. on. When the level is below 5%, the “ - + “ symbol on the LCD
• After entering the calibration mode and before the point is
button. • The ''pH 4.01 USE'' message is displayed. accepted, it is possible to quit the procedure and return to the blinks to indicate a low battery condition. If the battery level is low
Note: When the meter detects the absence of probe at its input, • Place the probe in the second calibration buffer (pH 4.01 or last calibration data by pressing the ON/OFF/MODE button. enough to cause erroneous readings, the Battery Error Prevention
the message “Probe not connected“ appears on LCD and ''- 10.01, or, if using NIST, pH 4.01 or 9.18). When the second The LCD displays “---ESC“ for 1 second and the meter returns to System (BEPS) turns the meter off. It is recommended to replace
--'' blinking on LCD lines. When a probe is connected, the buffer is accepted, the LCD will display “OK2” for 1 second and normal mode. the batteries as soon as the display will flash the battery symbol.
“Probe not connected“ tag is turned off, and the readings the meter will return to the normal measurement mode. • To reset the default values and clear a previous calibration, To replace the batteries, follow the next steps:
are displayed on LCD lines. • If the buffer is not recognised or the slope is out of accepted press the SET/HOLD button after entering the calibration
range ''---WRONG'' is displayed. Change the buffer, clean the • Open the battery compartment cap (on the bottom of the
mode, before the first point is accepted.
electrode or press any key to exit calibration. The LCD displays “---CLR“ for 1 second, the meter resets to the instrument).
• When the calibration procedure is completed, the “Calibrated“ default calibration and the “Calibrated” tag on the LCD disap- • Remove old batteries.
tag is turned on. pears. • Replace new batteries observing the polarity on the rear of
• It is always recommended to carry out a two-point calibration for the instrument.
better accuracy. • Close the battery compartment cap.