Professional Documents
Culture Documents
Prior to use allow all reagents to come to room temperature. SAFETY PRECAUTIONS
Reagents from different kit lot numbers should not be combined or The reagents must be used in a professional setting by trained
interchanged. personnel. The source material for reagents containing bovine serum
was derived in the contiguous 48 United States. It was obtained only
from healthy donor animals maintained under veterinary supervision
and found free of contagious diseases.
3) Reconstitute all assay calibrators level 1 to level 6 (Cat. 4) Remove the aluminum foil and plate sealer. Aspirate the
31061-31066) and controls (Cat. 31067, 31068) by adding contents of each well. Wash each well 5 times by
0.5 mL of demineralized water to each vial. Allow the dispensing 350 µL of working wash solution into each well
standards and controls to sit undisturbed for 5 minutes, and then completely aspirating the contents. Alternatively,
and then mix well by gentle inversions or gentle vortexing. an automated microplate washer can be used.
Make sure that all solid is dissolved completely prior to 5) Immediately add 100 µL of ELISA HRP Substrate (Cat.
use. These reconstituted standards and controls may be 10020) to each well.
stored at 2 – 8 °C for up to 6 hours or at ≤ -20 ºC for long- 6) Cover with foil or other material to protect from light, and
term storage. Do not exceed 3 freeze-thaw cycles. incubate static at room temperature for 20 ± 2 minutes.
4) Test Configuration: 7) Remove the aluminum foil and add 100 µL of ELISA Stop
Solution (Cat. 10030) into each of the wells. Mix gently.
ROW STRIP 1 STRIP 2 STRIP 3 STRIP 4 8) Read the absorbance at 450 nm with reference filter at
A Cal 1 Cal 5 SAMPLE 1 SAMPLE 5 620, 630 or 650 nm immediately.
B Cal 1 Cal 5 SAMPLE 1 SAMPLE 5 PROCEDURAL NOTES
C Cal 2 Cal 6 SAMPLE 2 SAMPLE 6 1. It is recommended that all calibrators, controls and
unknown samples be assayed in duplicate. The average
D Cal 2 Cal 6 SAMPLE 2 SAMPLE 6
absorbance reading of each duplicate should be used for
E Cal 3 C1 SAMPLE 3 Etc. data reduction and the calculation of results.
2. Keep light-sensitive reagents in the original amber bottles.
F Cal 3 C1 SAMPLE 3
3. Store any unused antibody-coated strips in the foil zipper
G Cal 4 C2 SAMPLE 4 bag with desiccant to protect from moisture.
4. Careful technique and use of properly calibrated pipetting
H Cal 4 C2 SAMPLE 4
devices are necessary to ensure reproducibility of the test.
6.
in this insert may affect the results.
Avoid introducing air bubbles into the microwells as this
Human AMH Calibration Curve
could result in lower binding efficiency and higher CV% of
4
duplicate reading.
7. All reagents should be mixed gently and thoroughly prior to 3.5
OD 450/650 nm
use. Avoid foaming. 3
8. If adapting this assay to automated ELISA system such as 2.5
DS-2 or EuChrom DUO a procedural validation is
2
necessary if there is any modification of the assay
procedure. 1.5
9. To ensure the accuracy of samples that test above the 1
dynamic range of the assay (around 20 ng/mL), a special 0.5
diluent is required and can be purchased separately as:
AMH Sample Diluent, Cat. No. 31060
0
0 5 10 15 20 25
INTERPRETATION OF RESULTS
It is recommended to use a cubic plot calibration curve fitting.
Human AMH, ng/mL
1. Calculate the average absorbance for each pair of
duplicate test results.
2. Subtract the average absorbance of the level 1 standard (0
ng/mL) from the average absorbance of all other readings EXPECTED VALUES
to obtain corrected absorbance or use 0 calibrator for
blank in computer program. Each laboratory should establish its own normal range based on
3. The calibration curve is generated by the corrected gender and age.
absorbance of all calibrator levels on the ordinate against
the calibrator concentration on the abscissa using a cubic Eight fresh serum samples were tested side-by-side using Serolisa™
plot. Appropriate computer assisted data reduction human AMH ELISA and a well-known commercial human AMH
programs may also be used for the calculation of results. ELISA. A satisfactory correlation result was obtained as showed in
The AMH concentrations for the controls and the patient samples are the chart below.
read directly from the calibration curve using their respective
corrected absorbance. Serum Age Gender Commercial Serolisa™
0.037
8 38 F 11.95 12.00
Cal-2
0.037 0.023
0.11 ng/mL 0.037
10.0
Cal-5 2.420
2.450 2.436
10 ng/mL 2.481
5.0 y = 0.9034x - 0.4391
R² = 0.9567
Cal-6 3.451 3.398 3.384
20 ng/mL 3.346
0.0
0.0 5.0 10.0 15.0 20.0
Control 1
0.335
0.337 0.323 1.32 ng/ml Commercial AMH Test, ng/mL
0.339
Sample 1 Bias
OBSERVED EXPECTED Test AMH Control AMH Bias
DILUTION RECOVERY (%)
VALUE VALUE Lipid (ng/ml) (ng/ml) (ng/ml)
(dobs)
Neat 19.28 - - 100 mg/dL 1.18 1.05 0.13 -11.9%
80% 15.25 15.42 99% 3.82 4.91 1.10 22.3%
60% 11.63 11.57 101% Bias
40% 8.24 7.71 107% Test AMH Control AMH Bias
(%)
Lipid (ng/ml) (ng/ml) (ng/ml)
20% 4.17 3.86 107% (dobs)
200 mg/dL 1.11 1.05 0.06 -5.5%
Sample 2 OBSERVED EXPECTED 3.64 4.91 1.27
RECOVERY 25.8%
DILUTION VALUE VALUE
Bias
Neat 15.52 - - Test AMH Control AMH Bias
(%)
Lipid (ng/ml) (ng/ml) (ng/ml)
80% 12.07 12.42 97% (dobs)
400 mg/dL 1.23 1.05 0.18 17.2%
60% 9.25 9.31 99%
40% 5.90 6.21 95% 4.31 4.91 0.60 -12.3%
20% 3.08 3.10 99%
REFERENCES
1. Lee, M. et al (1993); Müllerian-inhibiting substance: A gonadal AMH ELISA: Condensed Assay Protocol
hormone with multiple functions. Endocrine Reviews, 142, 152-
164.
2. Hudson, et al (1990); An immunoassay to detect human Müllerian 1. Pipet 50 µl Calibrators, controls and
inhibiting substance in males and females during normal patient samples
development. Journal of Clinical Endocrinology and Metabolism, 2. Pipet 50 µl Tracer Antibody per well
70, 16-22. Incubate @ RT for 4 hours.
3. Lee, M et al (1996); Müllerian Inhibiting Substance in human: on ELISA plate shaker,
normal levels from infancy to adulthood. Journal of Clinical wash 5 x
Endocrinology and Metabolism 81, 571 – 575.
3. 100 µl TMB Substrate
TECHNICAL ASSISTANCE AND CUSTOMER SERVICE
For technical assistance or place an order, please contact Epitope
Incubate @ RT for 20 min
Diagnostics, Inc. at (858) 693-7877 or fax to (858) 693-7678.
static
www.epitopediagnostics.com