The vacuum (400 mmHg or higher) is adjusted to 300 mmHg by the bellows unit in the main unit.
A filter is provided on the bellows unit to prevent the needle in the bellow from accumulating dirt.
Analytical L/M/HFR,
Parameter IRF
Reagent STROMATOLYS STROMATOLYSER-4D STROMAT RET
ER-FB Lysing solution/ Staining Dye (NEW) OLYSER- SERACHII
NR Lysing Diluent
solution/ solution/
Staining Staining
Dye (NEW) Dye (NEW)
*5 IMI Scattergram
Figure 2-1 Scattergrams
NOTE:
Laser Diode is very sensitive to static electricity. Follow the instruction below
when you replace the Laser Diode.
- Remove jumper pin from the terminal on the Laser Diode AFTER connect-
ing to PCB No. 2140. If the jumper pin is removed before connecting to
PCB No.2140, Laser Diode will be damaged.
- Do not disconnect the connector to the PCB No.2140.
In the worst case, the device inside the Laser Diode will be destroyed.
Left side