Professional Documents
Culture Documents
йŽĨƚŽƚĂůĐŽƵŶƚ
A cleanroom can be classified according
to its PDR by the method given in the ƌĞƐƵůƚƐ
VCCN Guidelines 9 (2014). In these ϭ /^KϭϰϲϰϰͲϭ
guidelines, the Particle Deposition Class ;ĞdžƚƌĂƉŽůĂƚĞĚͿ
(PDC) of a cleanroom is determined by
the following equation. Ϭ͘ϭ
Equation 6
Ϭ͘Ϭϭ
Ϭ ϮϬ ϰϬ ϲϬ ϴϬ ϭϬϬ
Where, PDRD is the maximum permitted Figure 1: Airborne particle counts as a percentage of the total particle counts *5µm.
ĐƵŵƵůĂƚŝǀĞƉĂƌƚŝĐůĞƐŝnjĞ;ђŵͿ
ISO 14644-1 (extrapolated) is calculated using Equation 7.
PDR (number/m2/h) of particles that are
equal to, or larger than, the considered
be calculated and the ISO class obtained particle size are measured, but the
particle size, D (µm).
by reference to the classification Table 1 deposition velocity of cumulative counts
Classification according to airborne given in ISO 14644-1. However, particles have not been previously available.
particle concentration with cumulative counts greater than These were determined in our second
The airborne cleanliness of a cleanroom 5µm are not given in this table and article, and given in Table 1 in cm/s.
is classified by ISO 14644-1 in terms of therefore it was necessary to calculate The cumulative deposition velocities
the concentration of particle sizes in the the ISO class by use of Equation 8. given in Table 1 can be used in Equation
range between *0.1µm and *5µm, and However, it is not clear how accurate 3 to calculate PDRs from knowledge
use of the following equation: this calculation would be, and this was of the cumulative particle concentration
investigated. in the cleanroom air. However,
Equation 7
Shown in Figure 1 are the investigations reported in our second
concentrations of cumulative counts of article showed that to calculate the most
different sizes of airborne particles accurate PDRs, the following
obtained in the cleanroom experiments restrictions should apply.
Where, Cn is the maximum permitted
reported in the second article of this 1. The calculations should only be
concentration/m3 of airborne particles
series (Whyte, Agricola and Derks, applied to ‘operational’ conditions in
that are equal to, and greater than,
2016) and given as a percentage of the a cleanroom i.e. during manufacturing,
the considered particle size, N is the
total count of all particles *5µm. Also and not in ‘at rest’ conditions.
ISO class number, and D (µm) is the
shown is an extrapolation of the particle
considered particle size. 2. The calculation of PDRs for particles
size concentrations of ISO 14644-1
Rewriting Equation 7 in terms of the above about *30 µm should be
calculated by Equation 7. It can be seen
cleanroom class (N) gives the following avoided as the cumulative size
that these diverge from the actual
equation, which allows the ISO distribution, and therefore the
concentrations at about *20µm, and by
Class (N) to be calculated from deposition velocity, is affected by
about *40µm the concentration is about
a concentration of particles of a variations in surface cleanliness and
10 times greater. Therefore, for more
considered cumulative size. redispersion of particles by activity.
accurate calculations it is better not to
Equation 8 use sizes above *20µm 3. With the exception of particles
*0.3µm, the deposition velocities
Deposition velocity of particles
given in Table 1 were obtained from
The deposition velocity of particles is
observations in an ISO Class 8 room.
their velocity through air towards a
However, it was found that the PDR
Later in this article, a method is given to cleanroom surface. The deposition
increased as the particle concentration
calculate the maximum ISO class required velocities of ‘discrete’ sizes of particles
decreases. This was considered to be
to ensure that airborne contamination have been obtained both theoretically
caused by lower particle concentrations
of a product is not greater than a specified and experimentally by various researchers,
being associated with higher air supply
and acceptable amount. If the critical and this information is discussed in our
rates, where smaller particles would
particle size that causes the contamination second article. However, the normal
be quickly swept from the cleanroom
is within the normal range of particles method of measuring particles in a
with little time to deposit, but larger
used in ISO 14644-1 i.e. *0.1µm to *5µm, cleanroom is by cumulative counts,
particles would still deposited by
the maximum particle concentration can where all particles above a considered
gravity. This effect was expected
ϭ
Table 1: Deposition velocities of cumulative counts of particle sizes
Cumulative particle diameter *0.3µm *0.5µm *5µm *10µm *25µm *40µm *50µm *100µm
Deposition velocity (cm/s) 0.003 0.006 0.3 0.9 4.2 9.1 13 41
WĂƌƚŝĐůĞĐŽŶĐ͘;ŶŽͬŵϯͿ
time of the air reduced. An increase
in the turbulent intensity of the air
was also thought to be a contributing
ϭϬϬϬϬ
factor. For a range of particles between
about *5µm and *30µm, the ϱϬϬϬ
deposition velocities would be
expected to increase by about
1.7-fold if applied to an ISO Class 7 Ϭ
cleanroom, about 3-fold if applied to Ϭ ϮϬ ϰϬ ϲϬ ϴϬ ϭϬϬ
an ISO Class 6 cleanroom, and about
dŝŵĞ;ŵŝŶͿ
5-fold if applied to an ISO Class 5.
Figure 2: Concentration of particles *10µm at a sampling location over time
4. Particles *0.3µm or *0.5µm were
expected to be less influenced by
a. Establish the contamination rate product, or other critical surface,
gravity, and the same deposition
of products manufactured in so as to reflect the actual airborne
velocity applied over the range of
a cleanroom, when the PDR or contamination adjacent to the surface.
cleanroom cleanliness classes.
airborne particle concentration Figure 2 shows the airborne
is known; concentration of particles *10µm
The main, and usually only, source
measured during an experiment reported
of MCPs in cleanroom air is personnel. b. Establish the maximum airborne
in the second article in this series. It can
Micro-organisms grow on the skin of cleanliness class, in terms of PDR or
be seen that the airborne concentration
personnel and, during activity, microbes airborne particle concentration, for
varied by more than 10 times, and if
are dispersed into the air on skin cells, an acceptable and specified amount
particle deposition is to be accurately
or fragments of skin cells. MCPs have of product contamination;
predicted, sampling must be carried out
various shapes, and it is normal to consider
c. Repeat (a) and (b) for MCPs. at the same time as products are exposed
particle movement in air and deposition
to deposition, and include periods of
onto surfaces, in terms of equivalent
Before discussing the methods used low and high activity. Sampling should
aerodynamic diameter, which is the
to carry out these three objectives, also be carried out over a suitably
diameter of a sphere of unit density that
it is necessary to consider how the long period of time or, if only short
settles in air at the same rate as the
PDR or MDR, and airborne concentrations manufacturing times occur, by multiple
particle being considered. The equivalent
of particle and microbial contamination, sampling.
aerodynamic particle diameter is not the
should be sampled to ensure accurate
same as the equivalent optical particle
results. Calculation of the number
diameter, the latter being used for
of airborne particles
the sizing of particles by optical particle
Measurement of airborne deposited onto product
counters. The average equivalent
contamination in cleanrooms The number of airborne particles that
aerodynamic particle diameter of
It has been demonstrated that particle may deposit onto a product, or another
airborne MCPs has been reported
concentrations will vary about a critical surface, can be calculated from
by Noble et al (1963) and Whyte and
cleanroom. In non-unidirectional airflow knowledge of either the PDR or the
Hejab (2007) to be about 12µm
cleanrooms, Whyte et al (2010) airborne particle concentration. Both
The deposition velocity of the average
have shown that the airborne particle these methods are now considered
size of airborne MCPs in different
concentration can vary about a cleanroom, by use of the following example.
cleanliness conditions in cleanrooms
depending on the performance of the air A product has a horizontal upper-
has been recently investigated by Whyte
supply diffusers, temperature difference surface area of 2 cm2 (0.0002m2) that
and Eaton (2016), and reported in an
between supply and room temperature, is exposed for 10 min (600s) to airborne
article to be published soon. It was found
position of the air extracts, and position contamination in a non-unidirectional
that the deposition velocity increased
of sources of contamination. Carr (1994) ISO Class 8 cleanroom. The reliability
as the cleanliness of the cleanroom
also reported that the particle of the product is known to be affected
increased, in a similar way to particles,
concentrations in a unidirectional by contamination with particles greater
and the results of the study are given in
cleanroom will vary by up to 10-fold than 10µm, and the likely number
Table 2.
between locations. It is therefore of these particles that will be deposited
The purpose of this article is to
important to ensure that sampling is onto a product is required.
explain calculation methods that can
carried out as close as possible to the
be used for the following purposes:
Calculation of number of times and the number of MCPs deposited of MCPs of 1.61cm/s (0.0161m/s) found.
MCPs deposited was found to average 3 per plate. The average number of MCPs that might
To calculate the surface contamination The microbial deposition rate (MDR) deposit onto a single product is calculated
of products, or other critical surfaces, was calculated by means of Equation 2 as follows:
by microbe-carrying particles (MCPs), as follows:
similar methods are used to those
described for particles in the previous
sections of this article. These methods
are based on information obtained from This is equivalent to a contamination
either a) settle plates that determine the rate of 1 in 518 products.
This MDR is then used to calculate
microbial deposition rate (MDR) or b)
the product contamination. As the
an airborne microbial sampler that Calculation of the maximum MCPs
product has an exposed horizontal area
ascertains the concentration of MCPs deposition rate, or maximum
of 2cm2 (0.0002m2) and MCPs have 10
in the cleanroom air. The methods airborne microbial concentration,
minutes (600s) to deposit, the product
are illustrated by means of the example for a specified product
contamination rate is as follows:
used in the previous two sections, contamination rate
in which a product with a horizontal Methods have been outlined above for
surface area of 2 cm2 (0.002m2) is calculating the maximum airborne
exposed to airborne deposition of MCPs cleanliness class of cleanroom, according
This is equivalent to 1 in 463 products
for 10 minute (600s). to ISO 14644-1, for an acceptable product
being contaminated by a MCP.
contamination rate of particles. A similar
Calculation of microbial
Calculation of MCP deposition method can be used with MCPs, and
deposition onto product by
onto product by means of microbial is demonstrated by use of the same
use of settle plate counts
air sampler concentrations manufactured product whose horizontal
The method of calculating the product
The method of calculating product area is 2 cm2 (0.0002m2) and exposure
contamination from the deposition rate
contamination from the airborne time to airborne contamination is
of MCPs onto settle plates has been
concentration of MCPs found by an 10 minutes (600s). The acceptable
previously discussed by Whyte (1986).
airborne sampler has been described microbial product contamination
This method is analogous to measuring
by Whyte and Eaton (2015), but that during manufacturing is set at 1 product
the PDR by witness plates, and is likely
method used a single deposition velocity in a 1000.
to be a more accurate method than that
to cover all ventilation conditions in
involving the air concentration of MCPs Calculation of maximum
a cleanroom. In view of the recently
obtained by microbial air samplers. microbial deposition rate
obtained deposition velocities given in
Settle plates are Petri dishes that contain The maximum MDR for a specified
Table 2, the method should be modified
nutrient agar, and when exposed in a contamination rate of 1 in 1000 products
to use different deposition velocities
cleanroom, MCPs will deposit from the can be calculated from the rewritten
in different airborne cleanliness.
air onto their surface. Settle plates with Equation 3.
The measurement of the airborne
a diameter of 90mm are commonly
concentration of MCPs by a microbial
used, but 140mm diameter plates are
air sampler is analogous to using an
more accurate for use in the low airborne
airborne particle counter to measure
concentrations found in cleanrooms;
the concentration of particles. However,
multiple settle plates are also more
microbial samplers measure all of the
accurate. After sampling, the settle
MCPs in the cleanroom air, and not the This MDR can be used to obtain the
plates are incubated at a suitable
concen tration above a threshold size as required cleanliness of the cleanroom
temperature and time, so that MCPs
measured by a particle counter. As all in terms of the maximum number of
grow into microbial colonies, which can
of the MCPs are counted, the required microbes deposited on a settle plate.
be counted and used to find the number
deposition velocity is the average If the settle plate used to sample the air
of MCPs that have deposited onto the
deposition velocity of all airborne is 14 cm diameter (surface area 0.0154/
settle plate in a given time. Settle plates
MCPs sampled. The deposition m2) and exposed for 3 hours (3600s),
should be laid out adjacent to product
velocity is given in Table 2 over the maximum microbial count on the
to ensure that the same deposition rate
a range of ventilation conditions. settle plate is calculated as follows:
of MCPs is likely to occur on the settle
The example used to illustrate this
plates as on product, and should be
method is again a product with an
exposed for several hours. The calculation
exposed surface area of 2cm2 (0.0002m2)
of microbial deposition is illustrated by
and time of exposure of 10 min (600s).
an example.
The product was manufactured in This result of 1.4 per settle plate is
A settle plate of 14cm diameter (area
unidirectional airflow, and the average the number of MCPs that should not
= 0.0154m2) was laid out adjacent to
airborne concentration of MCPs found be exceeded if the specified product
product for 3 hours during manufacturing.
adjacent to the exposed product during contamination rate of 1 in 1000 products
To obtain an accurate result, the
manufacturing was 1/m3. Table 2 was is not to be exceeded.
measurement was repeated several
consulted and a deposition velocity
Calculation of maximum airborne concentration of either type of particle. cumulative particle size should not
concentration of MCPs Also given in the Introduction are exceed 30µm. Also, the deposition
The maximum concentration of airborne equations that can be used to calculate velocity should be modified for different
MCP in a cleanroom can be calculated the maximum airborne cleanliness class ventilation conditions. The deposition
in a similar way to that previously for either particle deposition (according velocities given in Table 1, with the
described for airborne particles, and to VCCN Guidelines 9) or airborne exception of 0.3 and 0.5 µm particles,
illustrated by an example. To start particle concentration (according to were obtained in ISO Class 8 conditions
the calculation, it is necessary have an ISO 14644-1). These same equations can and should be increased by 1.7, 3 and 5
estimate of the deposition velocity. As a also be used to calculate the maximum times for ISO Class 7, Class 6, and
first estimate, the cleanroom is assumed number of MCPs that will deposit onto Class 5 conditions, respectively. The
to be a non-UDAF type with an airborne settle plates, or maximum concentration deposition velocity of particles *0.3µm
concentration of 50/m3, and by consulting of MCPs measured by an airborne and *0.5µm should be kept constant
Table 2 it can be seen that the deposition microbial sampler. The most accurate for different ISO Classes. Our previous
velocity of the average size of MCPs at method of calculating the amount article recommended that additional
that airborne concentration is 0.42cm/s of deposition, or maximum airborne research is needed to obtain more
(0.0042m/s). The maximum MDR concentration, is by the PDR or MDR as accurate deposition velocities in different
has been calculated in the previous this deposition closely simulates the ventilation conditions, but irrespective
section and, knowing it is 0.0083m2/s, actual deposition, and does not require of the outcome of these investigations,
the maximum MCP concentration the use of the deposition velocity. the present method will give useful
is calculated as follows: To obtain the most accurate results results, especially as no method is
from the equations given in the presently available to calculate the
Introduction, the deposition rate and expected rate of product contamination.
airborne concentration of contamination A method is also given in this article
should be measured adjacent to the for calculating the maximum airborne
product, and over the time the product class of cleanroom required for a specified
The calculated concentration of 1.98/m3 is open to airborne contamination, level of product contamination. At present,
is lower than the concentration first so that sampling accurately reflects the when a cleanroom is designed, the
estimated at the start of the calculation airborne concentration and deposition required cleanliness class of a cleanroom
(50/m3) and the calculation should be rate at the product. is typically obtained by an informed
repeated using a more accurate estimate The number of particles that deposit guess. This often leads to cleanrooms
of the deposition velocity. By consulting onto a product from cleanroom air can that are much cleaner than needed, with
Table 2 it can be seen that the deposition be calculated from the measurement unnecessarily-high capital and running
velocity for an airborne concentration of of the PDR and use of Equation 4. costs. Occasionally, and more seriously,
1.98/m3 should be closer to 1.5cm/s i.e. Equation 4 also requires the minimum cleanrooms can be built that are not clean
0.015m/s. Using this deposition velocity, size of particle that causes contamination enough to avoid excessive airborne
the maximum MCP concentration is problems (the critical size), the surface contamination. To avoid these problem,
re-calculated and found to be 0.55/m3. area of product exposed to deposition, a method is required that matches an
and time of exposure. Instruments have acceptable and specified contamination
been available for some time to measure rate of product to a maximum cleanliness
the PDR onto cleanroom surfaces such class of cleanroom. An example is given
as silicon wafers, but it is only recently to show how this can be calculated for
that relatively inexpensive and portable the PDC (according to VCCN 9), and
instruments have become available airborne particle class (according to
Discussion (Agricola, 2015 and 2016). However, ISO 14644-1).
This is the third and last article of a series where an instrument to measure PDR is It is accepted that the calculation of
that discusses the deposition of airborne not available, the airborne concentration the maximum ISO class for a cleanroom
particles onto critical surfaces in a of particles above the critical size can be using the methods suggested in this
cleanroom. The Introduction to this measured by an airborne particle counter, article is unlikely to be exact, but should
present article discusses the inter- and the PDR calculated by knowledge of give a good indication of the type
relationship between the concentration its corresponding deposition velocity for of cleanroom required e.g. non-
of airborne contamination, the particle cumulative counts. unidirectional with a high or low air
or microbial deposition rate (PDR or Table 1 gives the deposition supply rate, unidirectional airflow, or
MDR), and the deposition velocity of velocities of cumulative counts of a the requirement of separative devices,
particles and MCPs moving through air range of particle sizes. However, the and should be a considerable advance
under the influence of deposition forces, deposition velocities may vary from over the present method of using an
which are mainly gravity. Equations those given in the table. It has been informed guess and act as a useful
are given that relate these variables and demonstrated in our second article additional aid.
allow the number of airborne particles (Whyte, Agricola and Derks, 2016) that Included in this article are methods
and MCPs that deposit onto critical to minimise variation, the calculations for calculating the amount of product
surfaces to be calculated from knowledge should be confined to operational contamination by MCPs. The most
of either the PDR or MDR, or the airborne conditions in the cleanroom, and the accurate method of estimating product
contamination will be obtained from organisms. Journal of Hygiene, 61, CFD airflow studies of a cleanroom
the use of counts obtained from settle pp.385-391. with special respect to air supply
plates, as they directly measure the inlets. International Journal
8. VCCN Guideline 9 (2014). Particle
number of MCPs that will deposited of Ventilation, 9(3), pp.197-210.
deposition in cleanrooms and
onto a given surface area in a given time.
associated controlled environments. 14. Whyte W and Eaton T (2015).
Alternatively, the microbial airborne
Vereniging Contamination Control Assessment of degree of risk from
concentration can be measured and,
Nederland, 3831 NV Leusden, The sources of microbial contamination
by use of the appropriate deposition
Netherlands. in cleanrooms; 1: Airborne.
velocity given in Table 2, the amount
European Journal of Parenteral
of deposition calculated. This second 9. Whyte W (1986). Sterility assurance
and Pharmaceutical Science, 20(2),
method is unlikely to be as accurate, and models for assessing airborne
pp.52-62.
as the collection efficiency can vary bacterial contamination. Journal of
between air samplers and can be Parenteral Science and Technology, 15. Whyte W, Agricola K and Derks
low (Whyte et al, 2007) and an extra 40, pp.188-197. M (2015). Airborne particle
calculation is also required that is deposition in cleanrooms:
10. Whyte W (1996). In support of settle
dependent on the accuracy of the Deposition mechanisms. Clean
plates. PDA Journal of Pharmaceutical
deposition velocity. A method is also Air and Containment Review,
Science and Technology, 50(4),
described that calculates the maximum Issue 24, pp.4-9.
pp.201-204.
deposition of MCPs on settle plates,
16. Whyte W, Agricola K and Derks M
or the airborne concentration, for 11. Whyte W and Hejab M (2007).
(2016). Airborne particle deposition
a specified and acceptable amount Particle and microbial airborne
in cleanrooms: Relationship
of product contamination. These dispersion from people. European
between deposition rate and
calculations should be a useful tool in Journal of Parenteral and
airborne concentration. Clean Air
contamination control in cleanrooms. Pharmaceutical Science, 12(2),
and Containment Review, Issue 25,
pp.39-46.
pp.4-10.
References
12. Whyte W, Green G and Albisu
1. Agricola K (2015). Practical 17. Whyte W and Eaton T (2016).
A (2007). Collection efficiency
experiments in practical deposition Deposition velocity of airborne
and design of microbial air sampler.
monitoring. Clean Air and microbe-carrying particles.
Journal of Aerosol Science, 38,
Containment Review. Issue 21, pp.4-8. European Journal of Parenteral and
pp.101-114.
Pharmaceutical Science, (in press)
2. Agricola K (2016). Real-time particle
13. Whyte W, Hejab M, Whyte WM and
deposition monitoring of operational
Green G (2010). Experimental and
cleanroom quality. Journal of the
IEST, 59(1), pp.1-13.
W (Bill) Whyte, B.Sc. (microbiology), D.Sc. (mechanical engineering) and Honorary
3. Carr PE, Rapa AC, Fosnight WJ, Research Fellow at Glasgow University, has been involved with cleanrooms for
Baseman and Cooper, RJ (1994). over 50 years. He has published over 140 reports and papers and written two major
Measured effects of reduced flow books on the subject. He is a founder and former chairman of the Scottish
velocity in a laminar flow cleanroom Society for Contamination Control and the Cleanroom Testing and Certification
(1994). Journal of the IES, May/June, Board - International. He is a member of the BSI committee involved in the
pp. 41-46. writing of cleanroom standards. He has extensive experience as an industrial
consultant and running cleanroom courses.
4. Cheng Y-S, Yeh H_C and Allen MD
(1988). Dynamic shape factor of Koos Agricola is an Applied Physicist and has worked in R & D at Océ Technologies,
plate-like particles. Aerosol Science a Canon Company, since 1986. His responsibilities include contamination control
and Technology, 8, pp.109-123. in cleanrooms for the manufacture of critical parts. In his spare time, Koos
assists Technology of Sense b.v. as a Contamination Control Specialist. Koos
5. ISO 14644-1: 2015. Cleanrooms and is secretary of the VCCN (Dutch Contamination Control Society), ICCCS
Associated Controlled environments. (International Confederation of Contamination Control Societies) and ICEB
Part 1: Classification of air cleanliness (International Cleanroom Education Board) and a technical expert on ISO/TC
by particle concentration. International 209 Working Groups 1, 3, 4, 11, 12 and 13 and CEN/TC243 Working Group 5.
Organization for Standardization, Koos is treasurer of the CTCB-I (Cleanroom Testing and Certification Board –
Geneva, Switzerland. International) and regularly teaches various cleanroom technology subjects.
6. Mackintosh CA, Lidwell OM, Martin Derks is a skilled electronics engineer and has worked as calibration
Towers AG and Marples RR (1978). engineer and contamination control equipment expert for Philips Semiconductor
The dimensions of skin fragments where his role included responsibilities for cleanrooms, contamination control
dispersed during activity. Journal equipment for cleanrooms, maintenance and service. After 10 years as sales
of Hygiene, 81, pp.471- 479. engineer for a manufacturer of particle counters, Martin became managing
director for the European Headquarters of Lighthouse Worldwide Solutions.
7. Noble WC, Lidwell OM and Kingston This company supplies contamination monitors for various cleanroom
D (1963). The size distribution of industries, like pharma, medical and semiconductor.
airborne particles carrying micro-