MP DIAGNOSTICS INTERNAL PROCEDURE GUIDELINES

Area: IPG number: MP/CC/002


LABORATORY Date of Submission Date of Revision
January, 2017
Prepared by: Approved by:
Marco B. Espina, R.M.T. Demetrio L. Valle Jr., M.D
Laboratory Manager Clinical Pathologist
Subject: Distribution:
Turning ON Prestige 24i machine Clinical Laboratory

1. Purpose:
1.1 To know how to initialize the Prestige 24i machine for chemistry.

2. Responsibility:
2.1 Performance: Medical Technologist
2.2 Supervision: Laboratory Manager

3. Procedure:
3.1 Turn ON the main power switch.
3.2 Turn ON system power switch. When main power switch is ON, and this
switch is turned ON, analyzer starts up automatically and waits in stand-by
state. If you turn OFF this switch, all the functions stops, except reagent
cooling function.

[Program Downloading] is shown in the center of the screen and the


percentage is also shown. When loading is completed, this message
disappears and changes into the Main control screen.

3.3 The temperature monitor and the status monitor at the upper right of the
screen change into [WARM UP] and [IDLE].

The temperature monitor at the upper right turns from [WARM UP] to
[TEMP OK].

Note: The stand by state means the temperature monitor shows [TEMP
OK], and it will take about 25 minutes after turning on the system
switch.

3.4 Click [OK] button to clear the dialogue box.

Approval Process Name Position Signature Date

Prepared by: Marco B. Espina, R.M.T. Laboratory Manager

Approved by: Demetrio L. Valle Jr., M.D. Clinical Pathologist

Page 1 of 1