You are on page 1of 18

GE Medicalsystems

Technical
Publications

Direction 46-019530
Revision 0

<@
L - Advantx@Console
Central Listing

Copyrights 1991 By General Electric Co

Operating Documentahon

.
.

Direction 46-0 19530


Revision 0

Advantx@Console
Central Listing

IMPORTANT! rn rn rn X-RAY PROTECTION


X-ray equipment if not properly erator to take adequate precau- tion and take adequate steps to
used may cause injury Accord- tions to prevent the possibility of insure protection against injury
ingly, the instructions herein any persons carelessly unwisely
contained should be thoroughly or unknowingly exposing them- All persons authorized to use the
read and understood before you selves or others to radiation equipment must be cognizant of
attempt to place this equipment the danger of excessive exposure
in operation The General Elec- to x-radiation and the equipment is
tric Company, Medical Systems It is important that everyone hav- sold with the understanding that
Group, will be glad to assist and ing anything to do with x-radiation the General Electric Company,
cooperate in placing this equip- be properly trained and fully ac- Medical Systems Group its
ment in use quainted with the recommenda- agents and representatives have
tions of the National Council on no responsibility for injury or dam-

-*
Although this apparatus incorpo- Radiation Protection and Measure- age which may result from expo-
rates a high degree of protection ments as published in NCRP Re- sure to x-radiation
against x-radiation other than the ports available from NCRP Publica-
useful beam no practical design tions 7910 Woodmont Avenue Various protective material and
of equipment can provide com- Room 1016, Bethesda Maryland devices are available It is urged
plete protection Nor can any 20814 and of the International that such materials or devices be
practical design compel the op- Commission on Radiation Protec- used
. ADVANTX CONSOLE
GE MEDICAL SYSTEMS CENTRAL LISTING
REV 0 DIRECTION 46-019530

TABLE OF CONTENTS

CHAPTER TITLE PAGE

1 ADVANTX JUMPER AND SWITCH POSITIONS 1-1


SECTION 1 - ABBREVIATED GUIDE 1-1
SECTION 2 - DETAILED DESCRIPTION 1-2

FUSES AND CIRCUIT BREAKERS 2- 1

LED S 3- 1

TEST POINTS 4-1

VARIABLE COMPONENTS 5-1

1/11 (Blank)
b

ADVANTX CONSOLE
GE MEDICAL SYSTEMS CENTRAL LISTING
I REV 0 DIRECTION 46-01 9530

REVISION HISTORY
~~ ~ ~~~ ~~

REV DATE REASON FOR CHANGE


0 Nov 6, 1991 Initial release

LIST OF EFFECTIVE PAGES

PAGE REVISION PAGE REVISION PAGE REVISION


NUMBER NUMBER NUMBER NUMBER NUMBER NUMBER

Title Page 0

X-ray Protection 0

1 0
.
11 Blank
111 0

1-1 and 1-2 0


0

3- 1 0

4- 1 0
5- 1 0
ADVANTX CONSOLE
GE MEDICAL SYSTEMS CENTRAL LISTING
REV 0 DIRECTION 46-019530

CHAPTER 1 - ADVANTX JUMPER AND SWITCH POSITIONS


SECTION 1
ABBREVIATED GUIDE
1

The followng table serves as a brief verification guide for ‘normal” Advantx
swtch and jumper positions A detailed description of the positions follows this
table If your system does not match the unique configurations listed for a panic
ular circuit board you must consult the detailed description section of this docu
ment to determine the proper setting on that circuit board
Note Some of the circuit boards listed are optional and may not be present on all
systems

Damage to equipment may occur I f the jumpers and switches are not
set properly

Circult Board Normal Setting I sLco~~2~


I Unique Configuration I
scc1 A2 JP1-C&D Room #1 Only
JP2-A Room #1 Only

I JP3-OFF I I Room#l Only I


JP4-IN Room #1 Only
JP4-G2 Console Only
S94-(all positions) ON Room#l ’
I Wsitions except 1
~273-(a11 2) ON I I Room#1 Only I

1-1
ADVANTX CONSOLE
GE MEDICAL SYSTEMS CENTRAL LISTING
REV 0 DIRECTION 46-019530

SECTION 2
DETAILED DESCRIPTION

SCCl A2 CPUBoard 46264036

Posrtions Description
s94 Other settings used for Diagnostic Tests
1 2 3 4 5 6 7 8 consult GE Propnetary Documentation
ALL ON
S273
1 2 3 4 5 6 7 8 Node ID S
Room1 OFF OFF ON ON ON ON ON ON
Room2 ON ON OFFON ON ON ON ON
Room3 OFF ON OFFON ON ON ON ON
Room4 ON OFF OFFON ON ON ON ON
JP1 Firmware A --
B --
C Normal
D Normal
E Not Used
JP2 DuaRReset A Normal
B --
JP3 Audio Tone ON --
OFF Normal
JP4 Buzzer Enable IN -- G2 Version Console
OUT -- G1 Version Console

1-2
e
ADVANTX CONSOLE
GE MEDICAL SYSTEMS CENTRAL LISTING
t REV 0 DIRECTION 46-019530

CHAPTER 2 - FUSES AND CIRCUIT BREAKERS

Ckt Brkr / Schematic


Fuse Location Rating Voltage Description

CB1 2 5 A 250V 120 VAC 120 VAC Line Supply

2-1
ADVANTX CONSOLE
GE MEDICAL SYSTEMS CENTRAL LISTING
8
REV 0 DIRECTION 46-019530

CHAPTER 3 - LED’S

SCCl A2 CPU 46-264036


Schematic
Number Location Color Description
D S 41 Red CPU Halt
DS 44 Red Report Diagnostic Error Codes
DS 45 Red Report Diagnostic Error Codes
DS 46 Red Report Diagnostic Error Codes
I DS48
~~~
I I Red I Remrt Diaanostic Error Codes
DS 49 Red Report Diagnostic Error Codes
DS 51 Red Report Diagnostic Error Codes
D S 52 Red Report Diagnostic Error Codes
D S 53 Red Report Diagnostic Error Codes

3- 1
\
ADVANTX CONSOLE
GE MEDICAL SYSTEMS CENTRAL LISTING
a
REV 0 DIRECTION 46-019530

CHAP'IER 4 - TEST POINTS

SCCl A2 CPU 46-264036

Description I vpical Range I


I TP1 I 14-A10 I +12v I +12VDC I
TP2 14-D10 -12 v -12 VDC

TP3 14-810 +5v +5 VDC

TP4 14-C10 Logic Ground ov

4- 1
3

ADVANTX CONSOLE
GE MEDICAL SYSTEMS CENTRAL LISTING
a REV 0 DIRECTION 46-01 9530

CHAPTER 5 - VARIABLE COMPONENTS

SCCl A2 CPU 46-264036

Schematic
Component Location Description Function
R386 200KO 0 5 W , 1 0 % 25turn Gain Adjust for Audio Amplifier