USP Requirements for

Volumetric Instruments
How to Work with Different Standards in the Lab

volumetric instruments: the class A


error limits for the types and sizes
described by USP/ASTM.
▪▪ For volumetric instruments
with nominal capacities not stat-
ed in a standard, the USP/ASTM
standards typically refer to speci-
fications of the next smaller ca-
pacity. In contrast ISO standards
refer to specifications of the next
nearest capacity (or prescribe to
keep fundamental specifications
like shape and some important di-
mensions).
▪▪ According to ISO standards for
‘Ex’ adjusted volumetric instru-
ments (pipets and burets) class AS
– with S for swift delivery – is de-
fined as highest quality grade.
USP/ASTM standards do not con-
tain such a specification. They just
define class A for highest quality
grade of volumetric instruments.

© pilipphoto - Fotolia.com
Transfer Pipets and
Graduated Pipets

Companies producing pharmaceutical products for the error limits required by USP as well For transfer pipets the USP refers to
as the described nominal capacities the ASTM standard E 969. Contrari-
US market are monitored by US authorities, e.g. the FDA are different from the correspon- ly to DIN EN ISO 648 transfer pipets
dent ISO data. This diversity ham- with two marks are not mentioned
(Food and Drug Administration) and have to meet the re- pers an easy and appropriate work in the ASTM E 969. Hence, this kind
with glass volumetric instruments of volumetric instruments should
quirements of the USP (United States Pharmacopeia). Be- in European laboratories, if they not be used for applications super-
have to follow the demands of the vised by US authorities (FDA). Spec-
sides many instructions for analytical procedures the USP USP. Thus, the subsequent sections ifications of transfer pipets with
will explain the important require- one mark in accordance with ISO
defines the equipment to be used for these processes. ments, namely nominal capacity are only included for the nominal
(Figure 1) and error limits (Figure 2) capacities 0.5, 1, 2, 5, 10, 20, 25, 50
However, the differences between USP and DIN EN ISO for all types of USP glass volumetric and 100 ml, while ASTM standard E
instruments and compare these 969 defines the capacities 3, 4, 6, 7,
can pose a difficult task for work with glass volumetric data with the relevant DIN EN ISO 8, 9, 15, 30 and 40 ml in addition to
standards. the aforementioned values. The ma-
instruments in European labs. First of all, some general re- jority of nominal capacities for
marks that should be considered: transfer pipets have to meet the
▪▪ Minimum and maximum capac- same error limits whether in accor-
ities stated in the respective USP/ dance with ASTM or ISO. Just class
Section 31 outlines the requirements On the one hand, the require- ASTM and ISO standards should be A ASTM error limits of transfer pi-
for volumetric instruments to be uti- ments for volumetric flasks, transfer interpreted as sharp limits. There- pets with nominal capacities of 1,
lized. A most important requirement pipets, micro-burets, and graduated fore additional nominal capacities 2, 4, 5, 6 and 7 ml are tighter than
of the USP is the mandatory use of burets in the USP refer to ASTM below or above these limits cannot class AS ISO error limits. The 0.5 ml
glass volumetric instruments with standards. On the other hand, for refer to any of these standards. transfer pipets represent a particu-
accuracy according to class A. Due graduated pipets and graduated ▪▪ Since the construction specifi- lar case as the class AS ISO error
to some clauses this seemingly easy cylinders the USP specifies explicitly cations of USP/ASTM and ISO dif- limit of this pipet is the only one to
demand eventually turns out to be their requirements in the aforemen- fer completely, manufacturers usu- be tighter than the class A ASTM
difficult to understand. tioned section 31. Moreover, class A ally just focus on the usability of error limits.

G.I.T. Laboratory Journal 9–10/2013, pp 33–35, WILEY-VCH Verlag GmbH & Co. KGaA, GIT VERLAG, Weinheim www.gitverlag.com www.laboratory-journal.com
The USP defines the specifica- the respective DIN EN ISO 4788 4
tions for graduated pipets with sizes, namely the capacities 5, 10,
nominal capacities 1, 2, 5, and 10 100, and 250 ml, show the same
ml as well as the respective class A error limits. Class A error limits of
error limits directly in the section the other 5 nominal capacities (25,
31. In contrast to the DIN EN ISO 50, 500, 1000, and 2000 ml) de-
835 just type 2 graduated pipets fined by USP are tighter than the
are mentioned. As a consequence corresponding limits of the ISO
the commonly known types 1 and standard. Please note that class A
3 pipets should not be used for ap- error limits for mixing cylinders are
plications supervised by US author- the same as for graduated cylin-
ities. Class A USP error limits of the ders regardless whether they are
5 and 10 ml graduated pipets are with reference to USP or ISO.
tighter than class AS error limits of For burets the USP refers to
the correlated ISO standard, while ASTM standards E 287 for graduat-
the 2 ml graduated pipet exhibits ed burets (nominal capacities from
the same error limit with reference 10 to 100 ml) and E 1189 for micro-
to both standards. The 1 ml gradu- Fig. 1: Comparison of minimal (left) and maximal nominal capacities (right) burets (nominal capacities from 1 to
ated pipet represents another par- defined by USP/ASTM and ISO 10 ml) while the respective DIN EN
ticular case as the class AS ISO er- ISO standard 385 describes specifi-
ror limit of this pipet is tighter than not state nominal capacities below standard E 288, which describes cations for nominal capacities from
the class A USP error limit. 5 ml and above 2000 ml. Hence, the tighter class A error limits than the 1 to 100 ml. Class A error limits of
commonly known 1 and 2 ml trape- comparable DIN EN ISO 1042 for all USP/ASTM are the same as the
Flasks, Cylinders, Burets zoidal flasks and the 5000 ml bulb- sizes of volumetric flasks. class AS error limits of ISO. However,
shaped flasks should not be used for Specifications for graduated they may differ with regard to the
For volumetric flasks the USP refers applications supervised by US au- cylinders are mentioned in section commonly known 25 ml buret,
to ASTM standard E 288, which does thorities. The USP refers to the ASTM 31 of the USP. In comparison with which is not described by USP/

Fig. 2: Error limits of highest quality grade volumetric instruments in accordance with USP/ASTM (class A) or ISO (class A or AS): a) transfer pipets: Capacities
3, 4, 6, 7, 8, 9, 15, 30 and 40 ml are not stated in ISO; b) graduated pipets; c) volumetric flasks: Capacity 20 ml is not stated in USP/ASTM; d) graduated cylin-
ders. Please note general remark 3 for capacities not stated in a standard.
ASTM with a subdivision of 0.05 ml USP and highest quality grade
and, thus, should not be used for (class A or AS) following ISO, man-
applications supervised by US au- ufacturers print USP on the volu-
thorities. All other kinds of class AS metric instrument (Fig. 3) and de-
ISO burets can be used if they are liver either batch or individual
marked with the inscription USP. certificates with USP and/or ASTM
references. For applications super-
Summary vised by US authorities like the
FDA these USP volumetric instru-
As mentioned at the start of this ments should be chosen instead of
report, the diversity of commonali- the ISO volumetric instruments.
ties and differences hampers an The calibration could be done in
easy and appropriate work with accordance with the established
glass volumetric instruments for working plans as the DIN EN ISO
end-users in European laborato- 4787 shall be used for volumetric
ries, if they have to follow the re- instruments whether according to
quirements of the USP/ASTM. USP/ASTM or ISO standards.
Some manufacturers have modi- Fig. 3: Examples for inscriptions of 5 ml transfer pipet in accord with ISO
fied their production lines to be (left) or USP/ASTM (right)
able to offer volumetric instru-
ments that fulfill both the class A over, the DIN EN ISO standards are cases of 1 ml graduated pipet and
error limits required by USP/ASTM, also printed on these certificates as the 0.5 ml transfer pipet, whose
and the construction specifications the construction specifications and class AS ISO error limits are smaller
of the respective DIN EN ISO stan- maximal permissible errors are met than the class A USP/ASTM error Contact
dards. Usually changes are made as well. It is possible to confirm limits, the smaller one can be cho- Dr. Christian Schurz
to the certificates which refer to smaller values for the maximal per- sen. Brand GmbH + Co KG
the USP/ASTM standards and, missible errors, because standards To differentiate between volu- Wertheim, Germany
therefore, to the class A error limits prescribe only the maximal permis- metric instruments in accordance christian.schurz@brand.de
demanded by USP/ASTM. More- sible error. Hence, in the particular with class A error limits required by www.brand.de