Professional Documents
Culture Documents
ABSTRACT: The determination of FAME by GC is among the tories and has found broad application to biochemical, biomed-
most commonplace analyses in lipid research. Quantification of ical, microbiological, agricultural, and ecological research.
FAME by GC with FID has been effectively performed for some Presently, a large number of lipid analysts use the FID to
time, whereas detection with MS has been used chiefly for quali- measure FAME separated by GC. Although FID has proven to
tative analysis of FAME. Nonetheless, the sensitivity and selectiv- be a robust tool for FAME determination, the lack of any se-
ity of MS methods advocate a quantitative role for GC–MS in
lectivity can limit the usefulness of this detector when applied
FAME analysis—an approach that would be particularly advanta-
geous for FAME determination in complex biological samples,
to complicated samples, since only instrument response and re-
where spectrometric confirmation of analytes is advisable. To as- tention time information may be gathered. Owing in large part
sess the utility of GC–MS methods for FAME quantification, a to this limitation, misidentification of FAME in the presence of
comparative study of GC–FID and GC–MS methods has been contaminants, artifacts, or coeluting compounds is still of con-
conducted. FAME in prepared solutions as well as a biological cern when using FID (3–5). Thus, much work has been done to
standard reference material were analyzed by GC–FID and maximize the usefulness of retention time for FAME identifi-
GC–MS methods using both ion trap and quadrupole MS systems. cation through methods such as retention time locking, reten-
Quantification by MS, based on total ion counts and processing tion time prediction, and the dependence of retention time on
of selected ions, was investigated for FAME ionized by electron FAME equivalent chain lengths (6–8); however, FAME identi-
impact. Instrument precision, detection limits, calibration behav- ties assigned by such methods are generally considered tenta-
ior, and response factors were investigated for each approach,
tive (9). Hence, FID analysis of complex biological extracts
and quantitative results obtained by each technique were com-
pared. Although there were a number of characteristic differences
may prove inadequate in some situations, particularly for
between the MS methods and FID with respect to FAME analysis, FAME of relatively low abundance (10).
the quantitative performance of GC–MS compared satisfactorily With the coupling of MS methods to GC, much has been ac-
with that of GC–FID. The capacity to combine spectrometric ex- complished in the area of qualitative characterization of FAME
amination and quantitative determination advances GC–MS as a mixtures. Since GC–MS provides spectrometric information
powerful alternative to GC–FID for FAME analysis. on separated compounds, it provides a means of analyte selec-
Paper no. L9654 in Lipids 40, 419–428 (April 2005) tivity; thus, detection with MS also represents a potentially
powerful tool for quantitative analysis of FAME, especially in
the presence of a convoluted biochemical background. Despite
The first of many important advances in the early development the prospective benefits of GC–MS methodologies for quanti-
of GLC for analytical purposes was the separation and deter- tative FAME analysis, the more familiar FID is still favored in
mination of FA reported by James and Martin in 1952 (1). Soon many laboratories, particularly among lipid specialists.
after, the analytical separation of FAME by vapor-phase chro- GC–MS offers a host of tools for qualitative characteriza-
matography was described by Cropper and Heywood (2). Since tion of FA. For example, standard 70 eV EI ionization of pico-
then, the characterization of FA composition by esterification linyl, dimethyloxazoline, pyrrolidide, pentafluorodimethylsilyl,
to FAME and subsequent determination by GC has become one and trimethylsilyl derivatives of FA has been extensively studied
of the most widely performed analyses in lipid research labora- for the purpose of structural determination (11–17). EI ioniza-
tion of these less common FA derivatives yields fragments of di-
Present address of first two authors: Department of Chemistry, University of agnostic value in locating the positions of branching and in some
California Davis, One Shields Ave., Davis, CA 95616.
*To whom correspondence should be addressed at University of Alaska An-
cases the positions of unsaturation in FA (although the geomet-
chorage, Department of Chemistry, 3211 Providence Dr., Anchorage, AK ric configuration of the double bond is not forthcoming).
99508. E-mail address: kennish@uaa.alaska.edu The simplest method of FAME quantification by EI–MS
Abbreviations: AR, area ratio; EI, electron impact; IT, ion trap; LOD, limit may be carried out by monitoring a range of mass-to-charge
of detection; QP, quadrupole; RF, response factor; RSD, relative SD; SIE,
selected ion extraction; SIM, selective ion monitoring; SRM, Standard Ref- (m/z) values that will encompass the fragments expected of the
erence Material; TIC, total ion counts. analytes, then determining amount based on integration of
Copyright © 2005 by AOCS Press 419 Lipids, Vol. 40, no. 4 (2005)
420 E.D. DODDS ET AL.
peaks in the total ion count (TIC) chromatogram. If the identity information to users of ion trap (IT) MS systems (since this
and retention time of a given analyte have been established, the form of MS was still a relatively recent development at that
sensitivity and specificity of EI–MS can be extended through time). Limits of detection (LOD), calibration behavior, and re-
the use of selective ion monitoring (SIM), which involves the producibility of the instrumental methods were not addressed.
exclusive acquisition of a specified ion or group of ions during The present study provides a comparison of GC–FID and
a given time frame. When SIM is used in concert with EI, an GC–MS techniques for quantitative FAME analysis. Using
analyte peak area can be measured reliably regardless of high both IT and QP MS systems, quantification methods based on
background or coeluting peaks, provided that ions unique to the TIC and selected ion techniques were applied to FAME ion-
the analyte are monitored. Ideally, a fragment or fragments of ized by EI. LOD, calibration characteristics, RF, and method
relatively high abundance and characteristic m/z would be precision for determination of a broad range of FAME were as-
monitored. When entire mass spectra are acquired, benefits sessed by analysis of standard mixtures with each detection
similar to those offered by SIM may be gained through quan- method. To determine the applicability of each approach to the
tification based on only a subset of the acquired ions extracted analysis of a biological sample, FAME were prepared from a
from the TIC. This selected ion extraction (SIE) allows com- standard reference fish tissue homogenate with certified values
plete spectra to be recorded and permits the exclusion of unde- for a group of individual FA [NIST Standard Reference Mater-
sirable masses for purposes of quantification. Both approaches ial (SRM) 1946] (19).
generally involve the use of relatively few ions for quantifica-
tion; thus, significant numbers of analyte ions are disregarded
EXPERIMENTAL PROCEDURES
and a corresponding loss of signal is experienced. Nonetheless,
the net result is an enhancement of the signal-to-noise ratio by Calibration standards. A series of standard mixtures, includ-
elimination of most nonanalyte response from the signal. ing all FAME listed in Table 1, was prepared in residue analy-
An investigation of the performance of FID vs. MS in quan- sis-grade hexane (EM Science, Gibbstown, NJ). These FAME
tifying FAME separated by GC has not been conducted since were selected for calibration because of their prevalence in the
the work of Koza et al. in 1989 (18). These authors compared tissues of fish and marine animals. The FAME were obtained
FID response factors (RF) of FAME with those obtained using from Sigma-Aldrich (St. Louis, MO), Supelco (Bellefonte,
EI in quadrupole (QP) as well as sector-type mass spectrome- PA), Matreya (State College, PA), and Nu-Chek-Prep (Elysian,
ters. Unfortunately, only seven FAME were examined, none of MN), and were of the highest purity available. For each level
which was polyunsaturated. In addition, the study provided no of calibration, all FAME were present at equal concentrations,
TABLE 1
Analyte FAME Present in the Calibration Standards and Their Order of Elutiona
Elution FAME
order carbon numer Systematic name Common name
1 14:0 Tetradecanoic acid, methyl ester Methyl myristate
2 16:0 Hexadecanoic acid, methyl ester Methyl palmitate
3 16:1n-7 cis-9-Hexadecenoic acid, methyl ester Methyl palmitoleate
4 17:0 Heptadecanoic acid, methyl ester Methyl margarate
5 18:0 Octadecanoic acid, methyl ester Methyl stearate
6 18:1n-9 cis-9-Octadecenoic acid, methyl ester Methyl oleate
7 18:1n-7 cis-11-Octadecenoic acid, methyl ester Methyl cis-vaccenate
8 18:2n-6 cis,cis-9,12-Octadecadienoic acid, methyl ester Methyl linoleate
10 18:3n-3 cis,cis,cis-9,12,15-Octadecatrienoic acid, methyl ester Methyl linolenate
9 19:0 Nonadecanoic acid, methyl ester None
11 20:0 Eicosanoic acid, methyl ester Methyl arachidate
12 20:1n-9 cis-11-Eicosenoic acid, methyl ester Methyl gondoate
13 20:2n-6 cis,cis-11,14-Eicosadienoic acid, methyl ester None
15 20:4n-6 cis,cis,cis,cis-5,8,11,14-Eicosatetraenoic acid, methyl ester Methyl arachadonate
17 20:5n-3 cis,cis,cis,cis,cis-5,8,11,14,17-Eicosapentaenoic acid,
methyl ester EPA, methyl ester
14 21:0 Heneicosanoic acid, methyl ester None
16 22:0 Docosanoic acid, methyl ester Methyl behenate
18 22:1n-9 cis-13-Docosenoic acid, methyl ester Methyl erucate
19 22:4n-6 cis,cis,cis,cis-7,10,13,16-Docosatetraenoic acid,
methyl ester Methyl adrenate
20 22:5n-3 cis,cis,cis,cis,cis-7,10,13,16,19-Docosapentenoic acid,
methyl ester DPA, methyl ester
21 22:6n-3 cis,cis,cis,cis,cis,cis-4,7,10,13,16,19-Docosahexaenoic acid,
methyl ester DHA, methyl ester
a
The FAME of 21:0 was added pre-analysis as an internal standard.
ranging from 0.02 to 100.0 µg/mL across the series. As an in- Instrumentation, chromatography, and detection methods.
ternal standard, 21:0 FAME was added to each mixture at a FID analyses were performed with an HP 5890 Series II Plus
concentration of 50.0 µg/mL. All standards were analyzed in GC–FID system (Hewlett-Packard, Palo Alto, CA). The detec-
quadruplicate by GC–FID and each of the GC–MS methods. tor temperature was held at 300°C, and the flame was main-
RF standard. A commercially available standard mixture of tained with 30 mL/min H2 and 300 mL/min air. Helium was
37 FAME (Supelco) was used in the determination of RF. The used as the detector auxiliary gas at a flow of 30 mL/min. The
original solution was diluted 10-fold to give a final concentra- same GC system was also used for QP EI-MS analyses by
tion of 1.0 mg/mL total FAME. The RF standard was also equipping the GC instrument described above with an HP 5970
spiked with the FAME of 21:0 to obtain a 50.0 µg/mL internal MS. IT EI-MS analyses were carried out on a Varian CP-3800
standard concentration. This preparation was then analyzed in GC equipped with a Varian Saturn 2200 MS. A 10-min solvent
quadruplicate by GC–FID and all GC–MS methods under delay was applied to all detection methods. On both GC instru-
study. ments, chromatography was carried out using a 60 m × 0.25
NIST SRM. The NIST SRM 1946 (Gaithersburg, MD), Lake mm DB-23 capillary column with a film thickness of 0.25 µm
Superior fish tissue homogenate, was used to evaluate the use- (Agilent Technologies, Wilmington, DE). Helium was used as
fulness of each method in analysis of a realistic sample (19). the carrier gas at a flow rate of 1.0 mL/min with constant flow
The tissue homogenate was extracted using a Dionex ASE 200 compensation. GC inlets were held at a temperature of 300°C,
accelerated solvent extractor (Sunnyvale, CA) (20). Conditions and MS transfer lines were maintained at a temperature of
similar to those previously described for lipid extraction were 250°C. Sample injections of 1 µL were performed without split
used (21–27). for 30 s, followed by a 10:1 split ratio for the remainder of the
To prepare for extraction, the sample was removed from analysis. The oven temperature was programmed from 125 to
−80°C storage and allowed to thaw. A 100-mg portion of the 240°C at a rate of 3°C/min with a final hold of 1.67 min (the
tissue homogenate was weighed and combined with approxi- total analysis time was 40.00 min).
mately 1 g hydromatrix drying agent (Varian, Walnut Creek, IT GC–MS was carried out using 70 eV EI, and these data
CA). The tissue and hydromatrix mixture were transferred to were evaluated using both TIC and SIE. Additionally, QP
an 11-mL stainless steel extraction cell fitted with three cellu- GC–MS with 70 eV EI was carried out in full-scan acquisition
lose filters, and additional hydromatrix was added to fill the (with quantification based on the TIC) as well as in SIM mode.
cell. The sample was then extracted with 60% chloroform/40% All mass spectra were acquired over the m/z range of 40—400
methanol (obtained, respectively, from EM Science and Fisher except during SIM. A dwell time of 100 ms was used for all SIM
Scientific, Fair Lawn, NJ) at 100°C and 13.8 MPa. Two static ions. Ions acquired during QP-SIM or processed in IT-SIE are
extraction cycles were carried out, each 5 min in duration. Both listed in Table 2. In all cases, the three most abundant ions were
extraction solvents were residue-analysis grade and were monitored (SIM) or extracted (SIE) from the complete spectrum.
treated with 100 mg/L BHT as an antioxidant (Sigma). The
crude extract was dried by pouring it through a sintered glass
RESULTS
funnel containing several grams of anhydrous sodium sulfate
(J.T.Baker, Phillipsburg, NJ). This was followed by thorough Reproducibility, calibration, and LOD. Typical chromatograms
rinsing of the funnel with chloroform. The pooled dried extract of a FAME calibration standard obtained by each instrumental
and rinsing were then concentrated to approximately 1 mL approach are shown in Figure 1. Although retention times and
under a stream of nitrogen using a TurboVap apparatus set at a chromatographic resolution are slightly different among the
bath temperature of 50°C (Zymark, Hopkinton, MA). The ex- methods (likely owing to the differing inlets, detector inter-
tract concentrate was transferred to a round-bottomed reaction faces, and column ages involved), all analytes eluted under
vessel, and the remaining solvent was evaporated by imping- comparable conditions regardless of the instrument involved.
ing with nitrogen. The recovered lipids were reconstituted in 1 To compare the precision of each detection technique, the
mL 0.5 M methanolic KOH (VWR, West Chester, PA) and hy- mean area ratio (AR) of each analyte FAME with respect to the
drolyzed at 80°C for 30 min. Once the reaction mixture had internal standard was calculated based on the quadruplicate
cooled to room temperature, 1 mL of freshly opened 10% BF3 analyses of a FAME standard of intermediate concentration (10
in methanol was added (Supelco). Transesterification was per- µg/mL). The variability about each mean was computed and
formed at 100°C for 10 min. After cooling, 1 mL of distilled expressed as the relative SD (RSD). These results are provided
water and 2 mL of hexane were added to the sample with vor- in Table 3 for a representative suite of FAME. The greatest re-
texing. Following phase separation, the organic phase was col- producibility was consistently achieved with detection by FID
lected and transferred to a new vessel. The solvent exchange and QP-SIM, with both techniques giving AR with variabili-
was repeated with a fresh 2-mL portion of hexane. The recov- ties of less than 1% RSD for most FAME. Notable gains in pre-
ered organic phases were pooled, spiked with the methyl ester cision were afforded by QP-SIM as compared with QP-TIC (on
of 21:0 to result in a final concentration of 50.0 µg/mL, and di- average, about 0.9% RSD as compared with about 3% RSD);
luted to a total final volume of 10 mL with hexane. This sam- however, this degree of precision enhancement was not fully
ple was analyzed in quadruplicate by each detection method. realized when IT-SIE was compared with IT-TIC. Although
FIG. 1. Gas chromatograms of a FAME calibration standard with detection by FID (a), quadrupole (QP) MS (b), and
ion trap (IT) MS (c). For (b) and (c), the total ion counts (TIC) are shown for m/z 40–400. All FAME were present at
50.0 µg/mL. The peaks numbered in (a) correspond to the compounds listed in Table 1. The same order of elution
applies to (b) and (c). The internal standard is labeled with an asterisk.
concentrations of the analyte and the internal standard (Ca and FID responses toward all FAME cannot be assumed equivalent
Cs, respectively). RF for a representative sampling of FAME may be argued strictly in terms of the mechanism of detection.
are shown for each instrumental approach in Figure 4. In FID, the production of CHO+ ions, which are collected by
An often-held assumption among lipid analysts is that, in an the cathode and thus generate the detected current, is propor-
FID chromatogram, the peak area of an analyte FAME (AFAME) tional to the number of carbon–hydrogen bonds introduced to
divided by the total peak area of all FAME (ΣAFAME) is equal the flame. Since carbonyl carbons are not FID susceptible, it
to the ratio of the corresponding analyte FA mass (mFA) to the would follow that for an equal mass of two different FAME,
total mass of FA present in the sample (ΣmFA): the FAME of lower carbon number would have the smaller RF
owing to a higher proportion of FID-inactive carbon. Hence,
AFAME m
= FA [4] the FID RF for FAME are expected to depend on both the car-
ΣAFAME ΣmFA
bon number and the number of unsaturations of the analyte
This statement is clearly based on a number of assumptions, FAME in question. The FID RF shown in Figure 4 reinforce
not the least of which is that for the FID, the relative responses this point, clearly illustrating the predicted relationships. In
toward all FAME are equal. The present results, however, are practice, a number of additional factors prevent the FAME area
in agreement with a number of contrary findings that attest to percentage from equating to mass percentage; for example, in-
the questionable legitimacy of Equation 4 (29–31). That the jector bias results in a lesser percentage of more volatile com-
TABLE 3
Reproducibility of FAME Response by the Various Methodsa
pounds being successfully loaded on-column (9,30,32). This appear to be predictably dictated by either carbon number or
phenomenon was likely responsible for the reduced FID RF for degrees of unsaturation. Interestingly, the same ions were ex-
the 20:0 FAME seen in Figure 4(a). tracted to produce the IT-SIE data as were monitored by QP-
Unsurprisingly, the MS detectors also have RF that are sen- SIM, but very different trends in RF were observed, illustrating
sitive to the identity of the analyte, albeit in these cases the re- the impact that continuously sorted ions (QP-SIM) vs. selected
lationships between RF and carbon number or unsaturation extraction of trapped and sequentially released ions (IT-SIE)
number are more varied. The QP RF, for example, showed rel- can exert upon quantification.
atively little variability as a function of carbon number and had NIST SRM. The analyte concentrations determined in the
less scatter than FID; in the case of QP-SIM, the RF were re- quadruplicate analyses of the NIST SRM extract were aver-
produced with exceptional precision. The QP RF, however, ex- aged, and the FA quantities were expressed in terms of the TG
perienced a steep reduction with the addition of even a single mass percentage in tissue such that direct comparison of the
double bond, with additional but smaller reductions following present results with the standard reference values reported by
with additional degrees of unsaturation. The only significant NIST was possible. These results for each detection technique
difference between QP-TIC and QP-SIM appeared to be a mat- are summarized in Figure 5. For the majority of the analytes
ter of precision. For IT-TIC, as in FID, the RF were directly (of which the NIST-certified FA are shown in Fig. 5), most FID
proportional to carbon number and were diminished with in- and MS results were found to be in good agreement with the
creasing unsaturation number, although the range of RF was NIST SRM values (i.e., to within the SE of the measurements).
broader (about 0.6 to 1.1 for IT-SIE as opposed to a range of The procedure used for extraction of the SRM was signifi-
roughly 1.0 to 1.2 for FID). The RF for IT-SIE, though, did not cantly different from that used by NIST to characterize the
TABLE 4
Limits of Detection (LOD) for a Suite of the Standard FAME as Determined by the Various Methods of Analysisa
FID QP-TIC QP-SIM IT-TIC IT-SIE
LOD LOD LOD LOD LOD LOD LOD LOD LOD LOD
FAME (µg/mL) (pmol) (µg/mL) (pmol) (µg/mL) (pmol) (µg/mL) (pmol) (µg/mL) (pmol)
14:0 0.12 0.50 0.61 2.52 0.11 0.45 0.20 0.83 0.09 0.37
16:0 0.05 0.19 0.47 1.74 0.16 0.59 0.20 0.74 0.08 0.30
16:1n-7 0.11 0.41 0.86 3.21 0.18 0.67 0.23 0.86 0.15 0.56
18:0 0.13 0.44 0.64 2.15 0.21 0.70 0.19 0.64 0.09 0.30
18:1n-9 0.16 0.54 0.56 1.89 0.23 0.78 0.32 1.08 0.17 0.57
18:1n-7 0.15 0.51 0.58 1.96 0.23 0.78 0.34 1.15 0.81 2.74
18:2n-6 0.13 0.44 0.92 3.13 0.25 0.85 0.28 0.95 0.20 0.68
18:3n-3 0.14 0.48 1.15 3.94 0.27 0.92 0.24 0.82 0.19 0.65
20:0 0.15 0.46 0.40 1.23 0.23 0.71 0.26 0.80 0.15 0.46
20:1n-9 0.12 0.37 0.52 1.60 0.24 0.74 0.38 1.17 0.18 0.56
20:2n-6 0.11 0.34 0.47 1.46 0.24 0.75 0.40 1.24 0.19 0.59
22:1n-9 0.14 0.40 0.81 2.30 0.20 0.57 0.53 1.51 0.23 0.65
22:4n-6 0.32 0.92 0.65 1.88 0.20 0.58 0.44 1.27 0.42 1.21
22:5n-3 0.24 0.70 0.85 2.47 0.22 0.64 0.50 1.45 0.55 1.60
22:6n-3 0.54 1.58 0.21 0.61 0.25 0.73 0.78 2.28 0.52 1.52
a
For abbreviations see Table 2.
DISCUSSION
The present findings demonstrate that, although FID and vari-
ous MS techniques exhibit distinct behavior where response to
FAME for quantification is concerned, the use of calibration
with internal standardization allows the application of MS to
quantitative FAME analysis with adequate precision, low LOD,
and accurate quantitative results. The hallmarks of FID perfor-
mance in FAME quantification include a broad range of linear
response, excellent precision, and subpicomole LOD. Of
course, the main limitation of this detector is that the response
is nonspecific and provides no information on the identity of
the analyte.
Overall, the best MS performance was obtained using the
QP instrument operated in SIM mode. Although some selec-
tivity was afforded by virtue of the monitoring of selected ions,
the ability to acquire complete mass spectra was precluded.
Whereas this may be satisfactory for quantitative analysis of
well-characterized samples, samples of unknown FAME com-
position should be screened with full-scan MS prior to SIM
analysis for quantification. QP-SIM allowed better linearity of
response and better precision when compared with acquisition
of TIC on the same instrument and when compared with the IT
instrument. Subpicomole LOD were also obtainable in SIM
mode, a marginal improvement in the TIC LOD (typically, a
few picomoles). Even so, it should be noted that the majority
of these results occur within a rather small range regardless of
detection method, with each method being capable of a level
of performance that is acceptable for many applications.
The response of the IT instrument was less linear and, as has
been reported, gave slightly different spectra from those pro-
duced using QP instruments (33). The main advantage of the
FIG. 2. Calibration curves for the 18:0 FAME as determined by FID (a), IT instrument was that acquisition of full-scan mass spectra
QP-MS (b), and IT-MS (c). Error bars, where large enough to be visible, (IT-TIC) gave better quantitative performance than the QP-
represent the SD. SIM, selective ion monitoring; SIE, selected ion ex- TIC; thus, although QP-SIM offered the best MS performance
traction; for other abbreviations see Figure 1. overall, IT-TIC was superior to QP-TIC for quantification with
full-scan acquisition. The IT-MS analysis also afforded the
sample. The present method made use of high-pressure, high- ability to collect full spectra, yet still offered the capability to
temperature accelerated solvent extraction with chloroform/ extract ions for a given analyte later if enhanced signal-to-noise
methanol, employing less than 20 mL of solvent to extract 100 was needed. Indeed, SIE of FAME chromatograms did allow
mg of tissue homogenate in roughly 20 min. By contrast, the some improvement in LOD and method precision.
SRM values were determined by 18–24 h Soxhlet extractions These results also support the need for rigorous calibration,
of 2.5 g tissue homogenate samples with hexane/acetone. Al- irrespective of the detection method chosen. Whereas appro-
though the performance of accelerated solvent extraction for priate calibration is particularly critical for MS work, the as-
lipid extraction has been addressed elsewhere and deemed sat- sumption of uniform FID RF is also inadequate for precise
FIG. 3. Expanded chromatograms at the retention time of 18:0 FAME with detection by QP-MS showing the TIC (a)
and SIM of m/z 43 + 74 + 87 (b). Approximately 20 pg of the 18:0 FAME was injected in both cases. For abbrevia-
tions see Figures 1 and 2.
FIG. 4. Response factors for a series of saturated FAME (a) and a series of unsaturated FAME (b)
as determined by each detection method. Error bars represent the SD. For abbreviations see
Figure 2.
M.M., Eulitz, K., and Ku, Y. (2000) Gas Chromatography–High try Methods for Structural Analysis of Fatty Acids, Lipids 33,
Resolution Selected-Ion Mass Spectrometric Identification of 343–353.
Trace 21:0 and 20:2 Fatty Acids Eluting with Conjugated 16. Choi, M.H., and Chung, B.C. (2000) Diagnostic Fragmentation
Linoleic Acid Isomers, Lipids 35, 797–802. of Saturated and Unsaturated Fatty Acids by Gas Chromatogra-
11. Janssen, G., and Parmentier, G. (1978) Determination of Dou- phy–Mass Spectrometry with Pentafluorophenyldimethylsilyl
ble Bond Positions in Fatty Acids with Conjugated Double Derivatization, Anal. Biochem. 277, 271–273.
Bonds, Biomed. Mass Spectrom. 5, 439–443. 17. Hamilton, J.T., and Christie, W.W. (2000) Mechanisms for Ion
12. Christie, W.W., Brechany, E.Y., Johnson, S.B., and Holman, R.T. Formation During the Electron Impact–Mass Spectrometry of
(1986) A Comparison of Pyrrolidide and Picolinyl Ester Deriva- Picolinyl Ester and 4,4-Dimethyloxazoline Derivatives of Fatty
tives for the Identification of Fatty Acids in Natural Samples by Acids, Chem. Phys. Lipids 105, 93–104.
Gas Chromatography–Mass Spectrometry, Lipids 21, 657–661. 18. Koza, T., Rezanka, T., and Wurst, M. (1989) Quantitative
13. Christie, W.W., Brechany, E.Y., and Holman, R.T. (1987) Mass Analysis of Fatty Acid Methyl Esters by Capillary Gas Chroma-
Spectra of the Picolinyl Esters of Isomeric Mono- and Dienoic tography with Flame-Ionization Detection: Quadrupole and Sec-
Fatty Acids, Lipids 22, 224–228. tor Mass Spectrometer, Folia Microbiol. 34, 165–169.
14. Harvey, D.J. (1998) Picolinyl Esters for the Structural Determi- 19. May, W.E., and Rumble, J. (2002) Certificate of Analysis, Stan-
nation of Fatty Acids by GC/MS, Mol. Biotechnol. 10, 251–260. dard Reference Material 1946, Lake Superior Fish Tissue, Na-
15. Christie, W.W. (1998) Gas Chromatography–Mass Spectrome- tional Institute of Standards and Technology, Gaithersburg, MD.
FIG. 5. Mean mass percentage (as TG) of representative certified FA in the NIST Standard Ref-
erence Material (SRM) 1946 as determined by each method. Error bars represent the SD, ex-
cept for those associated with the NIST SRM values; for these values, the error bars represent
the uncertainties reported by NIST. For other abbreviations see Figure 2.
20. Richter, B.E., Jones, B.A., Ezzell, J.L., Porter, N.L., Avdalovic, 28. Miller, J.N., and Miller, J.C. (2000) Statistics and Chemomet-
N., and Pohl, C. (1996) Accelerated Solvent Extraction: A Tech- rics for Analytical Chemistry, 4th edn., pp. 120–123, Pearson
nique for Sample Preparation, Anal. Chem. 68, 1033–1039. Education, Essex.
21. Schafer, K. (1998) Accelerated Solvent Extraction of Lipids for 29. Ackman, R.G. (2002) The Gas Chromatograph in Practical
Determining the Fatty Acid Composition of Biological Mater- Analyses of Common and Uncommon Fatty Acids for the 21st
ial, Anal. Chim. Acta 358, 69–77. Century, Anal. Chim. Acta 465, 175–192.
22. Boselli, E., Velazco, V., Caboni, M.F., and Lercker, G. (2001) 30. Ulberth, F., Gabernig, R.G., and Schrammel, F. (1999) Flame-
Pressurized Liquid Extraction of Lipids for the Determination Ionization Detector Response to Methyl, Ethyl, Propyl, and
of Oxysterols in Egg-Containing Food, J. Chromatogr. A 917, Butyl Esters of Fatty Acids, J. Am. Oil Chem. Soc. 76, 263–266.
239–244. 31. Ackman, R.G., and Sipos, J.C. (1964) Application of Specific
23. Richardson, R.K. (2001) Determination of Fat in Dairy Products Response Factors in the Gas Chromatographic Analysis of
Using Pressurized Solvent Extraction, J. Assoc. Off. Anal. Methyl Esters of Fatty Acids with Flame Ionization Detectors,
Chem. Int. 84, 1522–1533. J. Am. Oil Chem. Soc. 41, 377–378.
24. Moreau, R.A., Powell, M.J., and Singh, V. (2003) Pressurized 32. Craske, J.D., and Bannon, C.D. (1987) Gas Liquid Chromatog-
Liquid Extraction of Polar and Nonpolar Lipids in Corn and raphy Analysis of the Fatty Acid Composition of Fats and Oils:
Oats with Hexane, Methylene Chloride, Isopropanol, and Etha- A Total System for High Accuracy, J. Am. Oil Chem. Soc. 64,
nol, J. Am. Oil Chem. Soc. 80, 1063–1067. 1413–1417.
25. Toschi, T.G., Bendini, A., Ricci, A., and Lercker, G. (2003) 33. Horman, I., and Traitler, H. (1989) Routine Gas Chromato-
Pressurized Solvent Extraction of Total Lipids in Poultry Meat, graphic/Mass Spectrometric Analysis of Fatty Acid Methyl Es-
Food Chem. 83, 551–555. ters Using the Ion Trap Detector, Biomed. Environ. Mass Spec-
26. Zhuang, W., McKague, B., Reeve, D., and Carrey, J. (2004) A trom. 18, 1016–1022.
Comparative Evaluation of Accelerated Solvent Extraction and 34. Seppanen-Laakso, T., Laakso, I., and Hiltunen, R. (2002)
Polytron Extraction for Quantification of Lipids and Extractable Analysis of Fatty Acids by Gas Chromatography, and Its Rele-
Organochlorine in Fish, Chemosphere 54, 467–480. vance to Research on Health and Nutrition, Anal. Chim. Acta
27. Dodds, E.D., McCoy, M.R., Geldenhuys, A., Rea, L.D., and 465, 39–62.
Kennish, J.M. (2004) Microscale Recovery of Total Lipids from
Fish Tissue by Accelerated Solvent Extraction, J. Am. Oil Chem.
Soc. 81, 835–840. [Received November 18, 2004; accepted March 28, 2005]