ASME BPYCY-2017
ARTICLE2,
ARTICLE 2
RADIOGRAPHIC EXAMINATION
T-210 SCOPE
The radiographic method described in this Article for
examination of materials including castings and welds
shall be used together with Article 1, General Require-
ments, Definitions of terms used in this Article are in
Article 1, Mandatory Appendix f, 1-121.1, RT =
Radiography
Certain product-specific, technique-specific, and
application-specific requirements are also given in other
Mandatory Appendices of this Article aslisted inthe table
of contents, These additional requirements shall also be
complied with when an Appendix is applicable to the
radiographic or radioscopic examination being
conducted.
7-220 GENERAL REQUIREMENTS
7-221 PROCEDURE REQUIREMENTS
T-221.1 Written Procedure. Radiographic examina:
Lion shall be performed in accordance with a writen pro.
cedure Each procedure shal includeat least the following
information, as applicable
{a} material type and thickness range
(6) isotope or maximum X-ray voltage used
{c) source-to-object distance (D in 1-274.1)
{d) distance from source side of abject to film (¢ in
1274.1)
{e) source size (Fin T-274.1)
{f) film brand and designation
{gj screans used
7-221.2 Procedure Demonstration. Demonstration
of the density and image quality indicator (1QN) image re-
quirements of the written procedure on production or
technique radiographs shall be considered satisfactory
evidence of compliance with that procedure.
7-222 SURFACE PREPARATION
7-222.1 Materials Including Castings. Surfaces shall
satisfy the requirements of the applicable materials spec
ification orreferencing Code Section, with additional con
ditioning if necessary, by any suitable process to such a
degree that the images of surface irregularities cannot
mask or be confused withthe image of any discontinuity
on the resulting radiograph
7-222.2 Welds. ‘The weld ripples or weld surface ir-
regularities on both the inside (where accessible) and
outside shall be removed by any suitable process te such
a degree that the images of surface irregularities cannot
mask or be confused with the image of any discontinuity
onthe resulting radiograph
The finished surface of all butt-welded joints may be
flush with the base material or may have reasonably uni-
form crowns, with reinforcement not to exceed that spe.
cified in the referencing Code Section
7-223 BACKSCATTER RADIATION
‘lead symbol "B? with minimum dimensions of Yin
(13mm) in height and %e in. (1.5 mm) in thickness shall
bbe attached tothe back ofeach film holder during each
posure to determine if backscatter radiation is exposing
the film. The lead symbol "B" shall be placed in alocation
so thatit would appear within an area on the radiograph
that meets the requirements of T-282, VIII-288, or
1X-286, as applicable.
7-224 SYSTEM OF IDENTIFICATION
A system shall be used to produce permanent identifi
cation on each radiograph traceable to the contract, com
ponent, weld or weld seam, or part numbers, as
appropriate In addition, the arganization’s symbol or
name and the date of the radiograph shall be plainly
and permanently included on the radiograph. This identi-
fication system does not necessarily require that the in-
formation appear as radiographic images. In any case.
this information shall not abscure the area of interest,
1-225 MONITORING DENSITY LIMITATIONS OF
RADIOGRAPHS
Either a densitometer or step wedge comparison film
shall be used for judging film density
7-226 EXTENT OF EXAMINATION
The extent ofradiographic examination shal be as spe
cified by the referencing Code Section
1-230 EQUIPMENT AND MATERIALS
7-231 FILM
T-231.1 Selection, Radiographs shall be made using
industrial radiographic film.
a7)ARTICLE 2
1-231.2 Processing, Standard Guide for Controlling
the Quality of Industrial Radiographic Film Processing,
SE-999, of Sections 23 through 26 of Standard Guide for
Radiographic Examination, SE-94, may be used as guide
for processing film, except that Section 8.1 of SE-999 is
not required
1-232 INTENSIFYING SCREENS
Intensilying screens may be used when performing
radiographic examination in aceardance with this Article
7-233 IMAGE QUALITY INDICATOR (IQI) DESIGN
7-233.1 Standard iQ! Design. (QIs shall be either the
hole type or the wire type Hole-type IQls shall be manu-
factured and identified in accordance with the require
ments or alternates allowed in SE-1025, Wire-type IQIs,
shall be manutactured and identified in accordance with
the requirements or alternates allowed in SE-747, except
that the largest wire number or the identity number may
be omitted. ASME standard IQIs shall consist of those in
Table T-233.1 for hole type and those in Table T-233.2
for wire type.
7-233.2 Alternative IOI Design. 1Qis designed and
manufactured in accordance with other national or inter
national standards may be used provided the require
ments of sither (a) or ()) below, and the material
requirements of I-276.1 are met
ASME BPVCV-2017
Table 1-233.2
Wire 1Q1 Designation, Wire Diameter, and
Wire Identity
Sea SaB
Wire
yn) identiy
033) 7
oa 8
ost) 9
6s) 10
oor) 1
Sec Sap)
Wire Wire |_ Wire Wire
Diameter in (mm) identity [Diameterin, (mm) identity
(oa) oc 254)
oose (102) 2 "
oot (27) 18 16
aos: (160) 14 19
dose (203) ae 20
aio 1 13) 21
Table T-233.1
Hole-Type 10! Designation, Thickness, and Hole Diameters
1Q1 Thickness
Ql Designation in (mm)
013)
019)
025)
032)
038),
oss)
si)
‘6a
075),
99)
2)
14}
52)
203),
254)
505)
356)
‘.05)
08)
10)
280 7.11
10
b
15
7
20
2
30
a5
40
IT Hole Diameter
aT Hole
Diameter
Toso (1.02)
2.040 (3.02)
2.040 (1.02)
3.050 (1.27)
21060 (1.52)
ao7o(n7
2oao:
00)
20)
27 Hole Diameter
09)
620)
30a7)
ASME BPYCY-2017
(0 Hole-type IQIs. The calculated Equivalent IQ Sensi-
Livity (EPS), per SE-1025, Appendix X1,is equal to or bet-
ter than the required standard hole-type IQ)
(6) Wire-Type 1QIs The alternative wire [QI essential
wire diameter is equal te or less than the required stan-
dard 1Q] essential wire
7-234 FACILITIES FOR VIEWING OF
RADIOGRAPHS
Viewing facilities shall provide subdued background
lighting of an intensity that will not cause reflections sha-
dows, or glare on the radiograph that interfere with the
interpretation process, Equipment used to view radio-
graphs for interpretation shall provide a variable light
source sufficient for the essential IQ! hole ar designated
wire to be visible for the specified density range. The
viewing conditions shall be such that light from around
the outer edge of the radiograph or coming through low.
density portions of the radiograph doesnot interfere with
interpretation,
T-260 CALIBRATION
7-261 SOURCE SIZE
T-2611 Verification of Source Size. ‘The equipment
manufacturer's or supplier's publications, such as techni
cal manuals, decay curves, ar written statements docu-
menting the actual or maximum source size or focal
spot shall be acceptable as source size verification
T-261.2 Determination of Source Size. When manu:
facturer's or supplier's publications are not available
source size may be determined as follows
(0) X-Ray Machines, For X-ray machines operating at
1,000 kV and less, the focal spot size may be determined
in accordance with SE-1165 Standard Test Method for
Measurement of Focal Spots of Industrial X-Ray Tubes
by Pinhole Imaging
(b) lridium-192 Sources, For Iridium-192, the source
size may be determined in accordance with SE-1114
Standard Test Method for Determining the Focal Size of
Iridium-192 Industrial Radiographic Sources
7-262 DENSITOMETER AND STEP WEDGE
COMPARISON FILM
T-262.1Densitometers. Densitometers shall be cali-
brated at least every 90 days during use as follows’
(0) Anational standard step tablet or a step wedge ca-
libration film traceable to a national standard step tablet
and having at least five steps with neutral densities from
atleast 1.0 through 4.0, shall be used. Phe step wedge ca-
libration film shall have been verified within the last year
by comparison with a national standard step tablet un.
less, priar te first use, it was maintained in the original
light-tight and waterproof sealed package as supplied
31
ARTICLE2,
by the manufacturer Step wedge calibration films may
bbe used without verification for one year upon apening
provided it is within the manufacturers stated shelf life
(b) The densitometer manufacturer's step-by-step in
structions far the operation of the densitometer shall be
followed
(6) The density steps closest to 1.0,2.0, 3.0, and 4.0 on
the national standard step tablet or step wedge calibra:
tion film shall be read,
(a) The densitometer is acceptable ifthe density read-
ings de not vary by more than +0.05 density units from
the actual density stated on the national standard step ta-
blet or step wedge calibration film
T-262.2 Step Wedge Comparison Films. Step wedge
comparison films shall he verified prior to first use unless
performed by the manufacturer, as follows
(a) The density of the steps on a ster wedge compari
son film shall be verified by a calibrated densitometer.
() The step wedge comparison filmis acceptable ifthe
density readings do not vary by more than £0.1 density
units from the density stated on the step wedge compar
ison film,
T-2623 Periodic Verification.
(0) Densitometers, Periodic cablibration verification
checks shall be performed as described in T-262.1 at
the beginning of each shift, after 8 hr of continuous use.
or after change of apertures, whichever comes first
(b) Step Wedge Comparisor: Films, Verification checks
shall be performed annually per T-262.2.
T-262.4 Documentation,
(0) Densitometers. Densitometer calibrations required
by T-262.1 shall be documented but the actual readings
for each step de not have te be recorded Periodic densit-
‘ameter verification checks required by -262.3(a) do not
have to be documented.
(6) Step Wedge Calibration Films, Step wedge calibra-
tion film verifications required by -262.1(a) shall be
documented but the actual readings for each step do
not have to be recorded
(c) Step Wedge Comparison Films. Step wedge compat
ison film verifications required by T-252.2 and
1-262.3(b) shall be documented, but the actual readings
for each step do not have to be recorded
1-270 EXAMINATION
T-271 RADIOGRAPHIC TECHNIQUE”
A single-wall exposure technique shall be used for
radiography whenever practical When it is not practical
touse a single-wall technique, a double-wall technique
shall be used An adequate number of exposures shall
bbe made to demonstrate that the required coverage has
been obtained,