You are on page 1of 32

S E C T IO N

Foldout pages for:


Theory of operation & Parts list 9 9
Section 2
Figure . . . . . 2-42. Electrical system block diagram (10.4-inch GUI—Trending-enabled CPU PCB
shown) on page 3
Figure . 2-43. Electrical system block diagram (10.4-inch GUI—Non-Trending-enabled CPU PCB
shown) on page 5
Figure . . . . . . 2-44. Electrical system block diagram (9.4-inch GUI—Trending-enabled CPU PCB
shown) on page 7
Figure . . . . . . . . . . . . 2-45. Electrical system block diagram (older 9.4-inch GUI shown) on page 9
Figure . . . 2-86. Puritan Bennett 840 Ventilator System interconnect diagram – Compresor unit on
page 11

Section 9
Figure . . . . . . . . . . . . . . . . . .9-1. 840 Ventilator System patient system and accessories on page 13
Figure . . . . . . . . . . . . . . . . . . . . 9-10. 10.4-inch GUI assembly with TE GUI CPU PCB on page 15
Figure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11. 10.4-inch GUI assembly on page 17
Figure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-15. 9.4-inch GUI assembly on page 19
Figure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-18. Breath delivery unit (BDU) on page 21
Figure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-23. BDU cover assembly on page 23
Figure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-24. BDU chassis assembly on page 25
Figure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-26. 806 Compressor unit on page 27
Figure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-28. Compressor unit enclosure assembly on page 29
Figure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-29. Compressor plenum assembly on page 31

Puritan Bennett 800 Series Ventilator System Service Manual 2-1


2 Theory of operation

This page intentionally left blank.

2-2 Puritan Bennett 800Series Ventilator System Service Manual


Theory of operation 2

INSPIRATORY MODULE
GUI
GUI CPU PCB J3 P1
GUI LED PCB INSPIRATORY
J2 ELECTRONICS
TO UPPER BACKLIGHT P9
PCB SAFETY
LCD PANEL INVERTER
PCBs VALVE
UPPER LCD FROM GUI
TO LOWER J11 PANEL P6
LCD PANEL CPU PCB J2 P2 AIR PSOL
PA (PSOL2)
P1 RS-232 LOWER LCD FROM GUI
PANEL CPU PCB J11 P5 OXYGEN
PI
PSOL
(PSOL1)
BD CPU P2 J7 J20 J19 LCD FLEX
P1 MOTHERBOARD COMPACT
PCB FLASH INSP
PWR CIRCUITS P10 PRES
REMOTE
ALARM XDUCER
P3 J8
A/Z SOL
DIGITAL
TEST J5 (SOL1)
SIGNALS
9 LEDS TOUCH FRAME
KEYBD/KNOB
PCB P3 OXYGEN
DATA FLOW
KEY J12 SENSOR
J1 AI PCB P9 J9 (Q1)
PWR &
ANALOG SIGNALS P1 P4 AIR FLOW
BD +5 V, BD +12 V, GUI +5 V, GUI +12 V, -15 V, +15 V, CTRL/STATUS
J6 P10 J10 SENSOR
2 LEDS DIGITAL (Q2)
SIGNALS J11 P7
P5 BPS
J13 OXYGEN
PTS 2000 BPS PCB J3 PRESS
J2 SWITCH
+36 V
(PS1)
BATTERY J1 J3 POWER SUPPLY
CHARGED I/O CONTROL AIR PRESS
BDU LED SWITCH
J1
PCB F1 (PS2)
PWR
J2 BATTERY P1
SW
PACK OXYGEN
(S1) BATTERY F1 SENSOR
CHARGING S1
EXH MODULE
HUMIDIFIER
OUTLET
EXH PRESS
XDUCER
A/Z SOL
(SOL2) AC PANEL 100-120 V
POWER RELAY
COMPRESSOR UNIT J4 ONLY
COMPRESSOR PCB K1
EXH HEATER J1 CB2
(EXH HTR) J6 BLU
PC RELAY
J1 BRN
UNLOADING
EXH FLOW SOLENOID
SENSOR (SOL3) CB1 E4
(Q3) TH1 AC BLU
J5 FILTER BRN
PCB E3
J2 J3
EXH VALVE
(EV) EQUIPOTENTIALITY
CONNECTOR J4

COMP
EXH PCB
MOTOR
TRANSDUCER
FAN 1 FAN 2
PE STARTING
CAPACITOR

Figure 2-42. Electrical system block diagram (10.4-inch GUI—Trending-enabled CPU PCB shown)

Puritan Bennett 800 Series Ventilator System Service Manual 2-55


2 Theory of operation

This page intentionally blank.

2-56 Puritan Bennett 800Series Ventilator System Service Manual


Theory of operation 2

INSPIRATORY MODULE
GUI
GUI CPU PCB J3 P1
GUI LED PCB INSPIRATORY
J2 ELECTRONICS
P9
TO UPPER PCB SAFETY
BACKLIGHT
LCD PANEL INVERTER VALVE
UPPER LCD FROM GUI
J11 PCBs CPU PCB J2 P6
TO LOWER PANEL
P2 AIR PSOL
LCD PANEL
PA (PSOL2)
P1 RS-232 FLEX CIRCUIT LOWER LCD FROM GUI
PANEL CPU PCB J11 P5
J7 PI OXYGEN
PSOL
(PSOL1)
J19 LCD FLEX
BD CPU P2 J7 MOTHERBOARD J20
P1 CIRCUITS INSP
PCB PWR REMOTE P10 PRES
ALARM XDUCER
P3 J8
DIGITAL A/Z SOL
TEST J5 (SOL1)
SIGNALS
9 LEDS KEYBD/KNOB TOUCH FRAME
PCB P3 OXYGEN
DATA FLOW
KEY J12 SENSOR
J1 AI PCB P9 J9 (Q1)
PWR &
ANALOG SIGNALS P1 P4 AIR FLOW
BD +5 V, BD +12 V, GUI +5 V, GUI +12 V, -15 V, +15 V, CTRL/STATUS
J6 P10 J10 SENSOR
2 LEDS DIGITAL (Q2)
SIGNALS J11 P7
P5 BPS
J13 OXYGEN
PTS 2000 BPS PCB J3 PRESS
J2 SWITCH
+36 V
(PS1)
BATTERY J1 J3 POWER SUPPLY
CHARGED I/O CONTROL AIR PRESS
BDU LED SWITCH
J1
PCB (PS2)
PWR F1
J2 BATTERY P1
SW
PACK OXYGEN
(S1) BATTERY F1 SENSOR
CHARGING S1
EXH MODULE
HUMIDIFIER
OUTLET
EXH PRESS
XDUCER
A/Z SOL
(SOL2) AC PANEL POWER RELAY 100-120 V
COMPRESSOR UNIT J4 COMPRESSOR PCB ONLY
K1
EXH HEATER J1 CB2
(EXH HTR) J6 BLU
PC RELAY
J1 BRN
UNLOADING
EXH FLOW SOLENOID
SENSOR (SOL3) CB1
AC E4 BLU
(Q3) TH1
J5 FILTER BRN
PCB E3
J2 J3
EXH VALVE
(EV) EQUIPOTENTIALITY
CONNECTOR J4

COMP
EXH PCB
MOTOR
TRANSDUCER
FAN 1 FAN 2
PE STARTING
CAPACITOR

Figure 2-43. Electrical system block diagram (10.4-inch GUI—Non-Trending-enabled CPU PCB shown)

Puritan Bennett 800 Series Ventilator System Service Manual 2-57


2 Theory of operation

This page intentionally blank.

2.58 Puritan Bennett 800Series Ventilator System Service Manual


Theory of operation 2

GUI INSPIRATORY MODULE


GUI CPU PCB J3 P1
GUI LED PCB INSPIRATORY
ELECTRONICS
J4 FROM GUI P9
TO UPPER PCB SAFETY
CPU PCB J4, J6
LCD PANEL J6 J2 VALVE
J5 J7
UPPER LCD P6
TO LOWER J12 BACKLIGHT
PANEL P2 AIR PSOL
LCD PANEL INVERTER
J15 J3
PCB PA (PSOL2)
P1
LOWER LCD
PANEL P5 OXYGEN
RS-232 J7 PI
PSOL
FROM GUI (PSOL1)
COMPACT CPU PCB J12, J15
BD CPU P2 J7 MOTHERBOARD J20 J19 COLOR I/O
P1 FLASH INSP
PCB PWR PCBs
REMOTE P10 PRES
ALARM XDUCER
P3 J8
DIGITAL A/Z SOL
TEST J5 (SOL1)
SIGNALS
9 LEDS KEYBD/KNOB TOUCH FRAME
PCB P3 OXYGEN
DATA FLOW
KEY J12 SENSOR
J1 AI PCB P9 J9 (Q1)
PWR &
ANALOG SIGNALS P1 P4 AIR FLOW
BD +5 V, BD +12 V, GUI +5 V, GUI +12 V, -15 V, +15 V, CTRL/STATUS
J6 P10 J10 SENSOR
2 LEDS DIGITAL (Q2)
SIGNALS J11 P7
P5 BPS
J13 OXYGEN
PTS 2000 BPS PCB J3 PRESS
J2 SWITCH
+36 V
(PS1)
BATTERY J1 J3 POWER SUPPLY
CHARGED I/O CONTROL AIR PRESS
BDU LED SWITCH
J1
PCB (PS2)
PWR F1
J2 BATTERY P1
SW
PACK OXYGEN
(S1) BATTERY F1 SENSOR
CHARGING S1
EXH MODULE
HUMIDIFIER
OUTLET
EXH PRESS
XDUCER
A/Z SOL
(SOL2) AC PANEL POWER RELAY 100-120 V
COMPRESSOR UNIT J4 ONLY
COMPRESSOR PCB K1
EXH HEATER J1 CB2
(EXH HTR) J6 BLU
PC RELAY
J1 BRN
UNLOADING
EXH FLOW SOLENOID
SENSOR (SOL3) CB1
AC E4 BLU
(Q3) TH1
J5 FILTER BRN
PCB E3
J2 J3
EXH VALVE
(EV) EQUIPOTENTIALITY
CONNECTOR J4

COMP
EXH PCB
MOTOR
TRANSDUCER
FAN 1 FAN 2
PE STARTING
CAPACITOR

Figure 2-44. Electrical system block diagram (9.4-inch GUI—Trending-enabled CPU PCB shown)

Puritan Bennett 800 Series Ventilator System Service Manual 2-59


2 Theory of operation

This page intentionally blank.

2-60 Puritan Bennett 800Series Ventilator System Service Manual


Theory of operation 2

GUI INSPIRATORY MODULE


GUI CPU PCB J3 P1
GUI LED PCB INSPIRATORY
ELECTRONICS
J6 VGA LCD FROM GUI P9
TO UPPER PCB SAFETY
CONTROLLER J2 CPU PCB J6, J8
LCD PANEL J8 J7 J7 VALVE
UPPER LCD
TO LOWER J16 VGA LCD BACKLIGHT PANEL P6
CONTROLLER P2 AIR PSOL
LCD PANEL J18 INVERTER
PCB J3 PA (PSOL2)
P1 LOWER LCD
REMOTE PANEL P5 OXYGEN
RS-232 ALARM J9 PI
FROM GUI PSOL
CPU PCB J16, J18 (PSOL1)
BD CPU P2 J7 MOTHERBOARD J20 J19 COLOR I/O
P1 INSP
PCB PWR PCBs P10 PRES
XDUCER
P3 J8
DIGITAL A/Z SOL
TEST J5 (SOL1)
SIGNALS
9 LEDS KEYBD/KNOB TOUCH FRAME
PCB P3 OXYGEN
DATA FLOW
KEY J12 SENSOR
J1 AI PCB P9 J9 (Q1)
PWR &
ANALOG SIGNALS P1 P4 AIR FLOW
BD +5 V, BD +12 V, GUI +5 V, GUI +12 V, -15 V, +15 V, CTRL/STATUS
J6 P10 J10 SENSOR
2 LEDS DIGITAL (Q2)
SIGNALS J11 P7
P5 BPS
J13 OXYGEN
PTS 2000 BPS PCB J3 PRESS
J2 SWITCH
+36 V
(PS1)
BATTERY J1 J3 POWER SUPPLY
CHARGED I/O CONTROL AIR PRESS
BDU LED SWITCH
J1
PCB (PS2)
PWR F1
J2 BATTERY P1
SW
PACK OXYGEN
(S1) BATTERY F1 SENSOR
CHARGING S1
EXH MODULE
HUMIDIFIER
OUTLET
EXH PRESS
XDUCER
A/Z SOL
(SOL2)
AC PANEL POWER RELAY 100-120 V
COMPRESSOR UNIT J4 COMPRESSOR PCB K1
ONLY
EXH HEATER
J1 CB2
(EXH HTR) RELAY J6 BLU
PC BRN
J1
UNLOADING
EXH FLOW SOLENOID
SENSOR (SOL3) CB1
(Q3) TH1 AC E4 BLU
J5 FILTER BRN
PCB E3
EXH VALVE J2 J3
(EV) EQUIPOTENTIALITY
CONNECTOR J4

EXH PCB COMP


TRANSDUCER MOTOR
FAN 1 FAN 2
PE
STARTING
CAPACITOR

Figure 2-45. Electrical system block diagram (older 9.4-inch GUI shown)

Puritan Bennett 800 Series Ventilator System Service Manual 2-61


2 Theory of operation

2.4.3 AC distribution components


The AC distribution components include
• a power cord
• AC panel
• power switch
• interfacing cables, wires, and components
Mains (AC) power (or facility power) is applied through the power cord to the AC receptacle
and main and auxiliary circuit breakers. From the main circuit breaker, power is applied
through the AC filter PCB directly to the power supply. Power to the power supply bypasses
the power switch, enabling the power supply to continually charge the BPS. When the power
switch is set to off, the secondaries that power the ventilator electronics are not active. From
the auxiliary circuit breaker, power is applied via the power relay to the compressor
compartment AC receptacle and the humidifier receptacle. The compressor compartment
power cord plugs into receptacle J4, supplying AC power to the compressor unit, while the
humidifier receptacle is intended to power an optional humidifier.

2.4.3.1 Power cord


The 840 Ventilator System includes a detachable power cord. The power cord has an IEC-
standard, three-prong connector. The plug end varies, corresponding to different country
requirements.

2.4.3.2 AC panel
The AC panel ( Figure 2-45 and Figure 2-46), a single field-replaceable unit (FRU), includes
components that apply AC to the ventilator head, compressor unit, and optional humidifier.
The AC panel includes AC receptacle J1, circuit breakers CB1 and CB2, AC filter PCB, power
relay K1, compressor receptacle J4, and potential equalization connector J2. A humidifier
receptacle (J3), which is connected to but is not a part of the AC panel, is on the front of the
ventilator.
• The AC receptacle (J1) receives facility AC power via the power cord.

Warning
Ensure that the power cord retainer bracket is properly installed and secures the power cord
to the AC receptacle (J1).

• The AC panel houses two push-to-reset type circuit breakers. The 5 A main circuit breaker
(CB1) limits current to the power supply. The 10 A (100 – 120 V) or 5 A (220 – 240 V)
auxiliary circuit breaker (CB2) limits current to the compressor and humidifier circuits.
• The AC filter PCB reduces AC line noise to and from the ventilator.
• The power relay (K1) switches power to humidifier receptacle J3 and compressor
receptacle J4 when the power switch (S1) is on. The relay is powered from a dedicated
+12 V power supply output, which is active when the power switch is on.
• The compressor receptacle (J4) receives AC power via K1 and provides the AC power
outlet for the compressor power cord.
• The humidifier receptacle (J3) receives AC power via K1 and provides an AC power outlet
for the humidifier power cord.
• The potential equalization connector (J2) interfaces the equipment and the potential
equalization bus bar.

2-62 Puritan Bennett 800Series Ventilator System Service Manual


Theory of operation 2

To AI PCB BDU compressor Motor mounting


DC cable plate
4-074223-00 Compressor
motor

Compressor
Compressor top view base plate

GRN/YEL

GRN/YEL
GRN/YEL
Capacitor

BLK (4)
Ground wire
from base plate
Grommet GRN/YEL 4-076299-00

Compressor PCB
4-075802-SP

Unloading
solenoid
J5 J6 (SOL 3)
J4

J3 J2 Unloading
J1
Hose to BDU

Compressor AC Fan harness (2)


power cord 4-076236-00
4-076287-00

To AC panel

Figure 2-86. Puritan Bennett 840 Ventilator System interconnect diagram – Compresor unit

Puritan Bennett 800 Series Ventilator System Service Manual 2-106


2 Theory of operation

2.4.12.3 806 Compressor unit operation


The 806 compressor operates in two modes: run and standby. A special start-up sequence
ensures that the motor starts with an unloaded compressor head.

2.4.12.3.1 Run mode


When AC mains power is > 80% of nominal, the compressor is available for use if wall air is
unavailable or the pressure is < 26 psig. In run mode, the compressor acts as the air source for
the ventilator, continuously supplying air to meet the ventilator’s full flow requirements (200
L/min peak flow, 2.5 L breath volume). When accumulator pressure reaches 27 psig, SOL3 is
energized, venting excess flow to assist transporting water vapor out of the air dryer. The
solenoid becomes de-energized when the pressure drops below 22.5 psig (see Figure 2-85).
This cycle repeats as pressure rises and falls in the accumulator.

2.4.12.3.2 Standby mode


The compressor enters stand-by mode when there is sufficient wall air pressure to supply the
ventilator. When the ventilator is powered up, it runs through POST and detects the presence
of the compressor. The compressor turns on and pressurizes the accumulator. During this
cycle, the green compressor ready indicator on the GUI illuminates when the accumulator
pressure reaches at least 13 psig. When the accumulator pressure reaches 27 psig, the
compressor turns off. If pressure in the accumulator drops below 22.5 psig (due to small leaks
in the system or cooling of compressed air), the compressor starts and recharges the
accumulator to 27 psig (see Figure 2-85).

2-107 Puritan Bennett 800Series Ventilator System Service Manual


Parts list 9

24 15
(4 PL)

23

22

19 (4 PL)
Fisher & Paykel humidifier, 2
6
externally-powered (220-240 V)
10
21 (3 PL) 11
16 12
18 7

8
9 15
5
20
17

Fisher & Paykel humidifier,


ventilator-powered (100-120 V)

13
Note: See Sections 9.4.12 and 9.4.13 for accessory bracket part numbers for ventilators 14
mounted on Puritan Bennett 800 Series Ventilator Compressor Mount Cart and Puritan
Bennett 800 Series Ventilator Pole Cart, respectively. Brackets are shipped with separate
installation instructions.

Figure 9-1. 840 Ventilator System patient system and accessories

Puritan Bennett 800 Series Ventilator System Service Manual 9-5


9 Parts list

9.2.1 840 Ventilator System NeoMode patient system and accessories

NeoMode patient system and accessories parts list

Item no.
Part no. Quantity Description
(Figure 9-2)

-- 4-076440-00 1 • Upgrade Kit, NeoMode, English, Non-US

-- 4-076441-00 1 • Upgrade Kit, NeoMode, English, US only

-- 4-076442-00 1 • Upgrade Kit, NeoMode, German

-- 4-076443-00 1 • Upgrade Kit, NeoMode, French

-- 4-076444-00 1 • Upgrade Kit, NeoMode, Italian

-- 4-076445-00 1 • Upgrade Kit, NeoMode, Japanese

-- 4-076446-00 1 • Upgrade Kit, NeoMode, Spanish

-- 4-076447-00 1 • Upgrade Kit, NeoMode, Portuguese


1 307/6922 • Ventilator breathing circuit, neonatal, disposable (DAR)

7441-4S2 • Ventilator breathing circuit, neonatal, disposable, 4 feet


--
(Allegiance Healthcare Corporation )
2 4-076408-00 1 • Filter, expiratory, NeoMode

-- DAR part Box of 12 • Expiratory bacteria filter (DAR), disposable


number
351P19005
Order part
number
4-076408-00
3 4-074601-00 Box of 12 • Inspiratory bacteria filter, 22-mm ISO connectors, disposable,
D/Flex
4 4-076405-00 1 • Adapter plate, NeoMode

-- 4-076589-00 1 • Label, adapter, NeoMode

9-6 Puritan Bennett 800Series Ventilator System Service Manual


Parts list 9

Figure 9-10. 10.4-inch GUI assembly with TE GUI CPU PCB

Puritan Bennett 800 Series Ventilator System Service Manual 9-33


9 Parts list

9.4.3 10.4-inch GUI (graphic user interface) assembly

10.4-inch GUI assembly parts list

Item no. Part no. Quantity Description


(Figure 9-12)
1 4-076672-SP 1 • Housing assembly, GUI rear
2 4-076678-SP 1 • Shield, CPU PCB
3 4-070034-00 47 • Screw, 6-32 X 0.38, sems, SQCN
4 4-075727-SP 1 • PCB, GUI CPU
5 4-076675-00 1 • Bracket, LCD panels and GUI CPU PCB
6 4-076656-00 1 • Cable assembly, GUI alarm
7 4-073500-SP 1 • PCB, GUI LED
8 4-076651-00 1 • Cable assembly, touchframe
9 4-079013-SP 1 • Speaker, GUI alarm, 16 OHM
10 4-076652-00 1 • Cable assembly, keyboard
11 4-076671-SP 1 • Housing assembly, GUI front
12 4-076687-00 4 • Screw, 2-56 X 1/4 PH, nylon
13 4-076855-SP 2 • Inverter, backlight,10.4-inch GUI
14 4-076653-00 2 • Cable assembly, backlight, 10.4-inch GUI
15 4-076658-00 8 • Screw. 6-32 X 1/4 PH sem
16 4-078267-SP 2 • Display, LCD, 10.4-inch color
17 4-076684-00 3 • Gasket, EMC, Chrometrics, 8.5"
18 4-076685-00 2 • Gasket, EMC, Chrometrics, 13"
19 4-076660-00 11 • Rivet, snap, H type
20 4-076676-00 1 • Shield, LCD
21 4-076530-SP 1 • PCB, touchframe
22 4-076686-00 1 • Gasket, EMC, Chrometrics, RS-232, 5"
23 4-076665-00 4 • Screw, lock, fem, 4-40, 0.75L
24 4-076677-00 1 • Shield, RS-232
25 4-071944-SP 1 • Cable, flex circuit assembly, RS-232
26 4-076856-SP 2 • Cable, flex circuit assembly, video
27 4-076688-00 4 • Washer, nylon
28 4-075865-00 1 • Shield, dielectric, GUI

-- 10021120 1 • FRU, IC, real-time clock (not shown)

9-34 Puritan Bennett 800Series Ventilator System Service Manual


Parts list 9

Figure 9-11. 10.4-inch GUI assembly

Puritan Bennett 800 Series Ventilator System Service Manual 9-35


9 Parts list

9.4.4 10.4-inch GUI (graphic user interface) handle

10.4-inch GUI handle parts list

Item no. Part no. Quantity Description


(Figure 9-13)

-- -- 1 Graphic user interface (GUI) (See Section 9.4 for NHA.)

-- 4-078179-00 1 10,000 hour preventative maintenance kit


1 4-070039-00 2 • Screw, PAN, 6-32 x 1.5", sems (Attach top and bottom handles
to rear housing.)
2 4-073008-00 1 • Handle, GUI top
3 4-075864-00 1 • Cable assembly, GUI-to-BDU, 3-ft (0.9-m)

4-071441-00 • Cable assembly, GUI-to-BDU extension, 10-ft (3.0-m) (for shelf


mount)
4 4-075684-00 2 • Jackscrew
5 4-070036-00 1 • Screw, PAN, 6-32 x 0.75, sems
6 4-073007-00 1 • Handle, GUI bottom
7 -- 1 • Kit, Label, Product information (Refer to Section 9.4.1 for part
number.)
8 -- 1 • Label, serial number (Contact your Puritan Bennett
representative for ordering information.)
9 4-076663-00 1 • Window, LED status, GUI

9-36 Puritan Bennett 800Series Ventilator System Service Manual


Parts list 9
9-1. 840 Ventilator System patient system )

1
(2PL)
2
3
4 (2 PL)

5
6 9
7
8
10
33
(13 PL)
11 (5 PL)

12 28 (2 PL)
13
39
32
(13 PL) 37
30 14 38
18
31
(11 PL)
15
40
34

19
16
29
41
35
17
(2 PL)
27 36
(2 PL)
26
25
24 (4 PL)

23

20

22
21

Figure 9-15. 9.4-inch GUI assembly

Puritan Bennett 800 Series Ventilator System Service Manual 9-45


9 Parts list

9.4.7.1 9.4-inch GUI front housing assembly

9.4-inch GUI front housing assembly parts list

Item no.
Part no. Quantity Description
(Figure 9-17)

-- -- 1 Housing assembly, GUI front (See Section 9.4.7 for NHA.)


1 4-073003-00 1 • Bezel, GUI
2 4-073043-00 1 • Gasket, window
3 4-073036-00 1 • Window, GUI
4 4-073033-00 1 • Gasket, housing
5 4-073028-00 1 • Label, alarm graphics, GUI, non-US

4-075429-00 • Label, alarm graphics, GUI, US


6 4-073001-00 1 • Housing, GUI front
7 4-075828-SP 1 • Keyboard assembly, English, non-US (Includes PCB and knob)

4-075834-SP • Keyboard assembly, English, US (Includes PCB and knob)

4-075829-SP • Keyboard assembly, French (Includes PCB and knob)

4-075830-SP • Keyboard assembly, German (Includes PCB and knob)

4-075832-SP • Keyboard assembly, Italian (Includes PCB and knob)

4-075835-SP • Keyboard assembly, Japanese (Includes PCB and knob)

4-075833-SP • Keyboard assembly, Portuguese (Includes PCB and knob)

4-075831-SP • Keyboard assembly, Spanish (Includes PCB and knob)


8 4-071313-00 1 • Cable assembly, keyboard

9-46 Puritan Bennett 800Series Ventilator System Service Manual


Parts list 9

49
2

4
6
(2 PL)
(2 PL)

48 5
12

52

47
13
45

19 10 (2 PL)
(4 PL) 14
16
(2 PL)
46 18

20 17 15 7
21 11
40
23 (4 PL) 11
24 8
25 (2 PL)
(2 PL)
27 9
26 22

29
38
(2 PL)
44 36 33
37
(2 PL) (2 PL) 35
39
51
(2 PL) 30 (2 PL)
28
34 31 (2 PL)
50
32
41
43
42

33 (4 PL)

Figure 9-18. Breath delivery unit (BDU)

Puritan Bennett 800 Series Ventilator System Service Manual 9-53


9 Parts list

9.4.8.1 Exhalation Module

Exhalation module assembly parts list

Item no.
Part no. Quantity Description
(Figure 9-20)

-- 4-072600-SP -- Exhalation module (See Section 9.4.9 for NHA.)


1 4-079022-SP 1 • Heater assembly, exhalation
2 4-070048-00 3 • Screw, PAN, 6-32 x 5/8 (Attach heater assembly to latch
assembly)
3 4-079054-SP 1 • Check valve assembly with seal, exhalation (CV5)
4 -- 1 • • Check valve assembly

5 -- 1 • • Seal, check valve, exhalation

6 -- 1 • Manifold, expiratory filter


7 4-072607-00 1 • Seal, exhalation
8 4-072218-SP 1 • Solenoid valve assembly, 3-way, +6 V (expiratory pressure
transducer autozero solenoid, SOL2)
9 -- 1 • Chassis assembly, exhalation
10 4-072623-00 3 • Screw, FH, 8-32 x 3/8 (Attach exhalation filter manifold and
fascia to exhalation chassis)
11 4-078141-00 1 • Harness assembly, exhalation module
or • Harness assembly, exhalation module, EMI (Phase II) (Use only
4-078220-SP with EMI exhalation cover and EMI exhalation transducer PCB)

12 4-070300-SP 1 • PCB, exhalation transducer


or • PCB, exhalation transducer, EMI (Phase II) (Use only with EMI
4-075226-SP exhalation harness)

13 4-070034-00 3 • Screw, PAN, 6-32 x 0.38, sems (Attach exhalation transducer


PCB to exhalation chassis)
14 4-076461-SP 1 • Valve assembly, exhalation (EV)
15 -- 1 • • Valve, exhalation (EV)

16 -- 1 • • Port, exhalation

17 4-072212-SP 1 • Sensor with O-rings, flow, exhalation (Q3)


18 -- 1 • • Sensor, flow, exhalation (Q3)

19 -- 2 • • O-ring

20 4-070034-00 4 • Screw, PAN, 6-32 x 0.38, sems (Attach exhalation valve to


exhalation chassis)
21 4-070047-00 2 • Screw, PH, 6-32 x 5/16, sems (Attach expiratory pressure
transducer autozero solenoid (SOL2) t o exhalation chassis)
22 4-076392-00 1 • Fascia, exhalation
23 4-072625-00 1 • Latch assembly, exhalation
24 4-072614-00 1 • Ring, retaining (Attaches latch to manifold)
25 4-072623-00 1 • Screw, FH, 8-32 x 3/8 (Attach latch to manifold)

9-54 Puritan Bennett 800Series Ventilator System Service Manual


Parts list 9

2
23 3
(2 PL)
5

22 1

21 (2 PL)

20
26

19 25

6 (4 PL)
18
7 (4 PL)
24
17

27

12

11

14
(2 PL)
8
15 13
(2 PL) 9 (2 PL)
10
(2 PL)

16

Figure 9-23. BDU cover assembly

Puritan Bennett 800 Series Ventilator System Service Manual 9-69


9 Parts list

9.4.8.4 BDU chassis assembly

BDU chassis assembly parts list

Item no.
Part no. Quantity Description
(Figure 9-25)

-- -- 1 Chassis assembly, BDU (See Section 9.4.9 for NHA.)


1 4-079032-SP 1 • Chassis kit, BDU (See Figure 9-24 for parts breakdown.)
2 4-079019-SP 1 • Cable assembly, inspiratory blindmate
3 4-005482-00 3 • Nut, HEX, 8-32, keps (Attach inspiratory blindmate cable to
chassis)
4 4-079018-SP 1 • Cable assembly, DC blindmate
5 4-070055-00 2 • Screw, FH, 8-32 x 1/4 (Attach DC blindmate cable to chassis)
6 4-079017-SP 1 • Cable assembly, AC blindmate
7 4-016142-00 1 • Nut, HEX, 6-32, keps (Attaches AC blindmate assembly ground
wire to chassis)
8 4-072623-00 2 • Screw, FH, 8-32 x 3/8 (Attach AC blindmate assembly to
chassis)
9 4-023138-SP 1 • Audio alarm assembly with barrier, BD
10 4-071452-00 1 • Cable assembly, BDU LED PCB/alarm/power switch
11 4-074901-00 1 • Cable assembly, exhalation I/O
12 4-073041-00 8 • Standoff, #4, M/F (Attach motherboard PCB to enclosure)
13 4-007566-00 8 • Washer, SR, #4 (Attach motherboard PCB to enclosure)
14 4-070900-SP 1 • PCB, motherboard
15 4-070034-00 7 • Screw, PAN, 6-32 x 0.38, sems (Attach motherboard PCB to
enclosure)

9-70 Puritan Bennett 800Series Ventilator System Service Manual


Parts list 9

3
(3 PL) 9
2

10
1

4
5
(2 PL)

15 2
6 (7 PL)
14

11

13
(8 PL)
12
(8 PL)

6
8
(2 PL)

Figure 9-24. BDU chassis assembly

Puritan Bennett 800 Series Ventilator System Service Manual 9-71


9 Parts list

9.4.8.5 BDU chassis kit

BDU chassis kit parts list

Item no.
Part no. Quantity Description
(Figure 9-26)

-- 4-079032-SP 1 Chassis kit, BDU (See Section 9.4.9.4 for NHA.)


1 -- 1 • Chassis, enclosure, BDU
2 4-071549-00 1 • Slide, mount, BDU latch
3 4-070064-00 3 • Screw, FH, 10-32 x 3/8 (Attach slide to chassis)
4 4-079015-SP 1 • Handle kit, BDU release
5 -- 1 • • Release and pin assembly, BDU latch

6 4-072614-00 1 • • Ring, retaining

7 4-071546-00 1 • • Spring, BDU latch

8 4-071558-00 10 • Card guide


9 4-070057-00 9.75 in. • Grommet, serrated strip (Available in multiples of 1-ft lengths.
(24.8 cm), Minimum order is 10 ft.)
cut into 2
4.875 in.
(12.4 cm)
pieces
10 4-070056-00 2 • Grommet
11 4-071583-00 1 • Gasket, chassis, enclosure
12 4-071535-00 4 • Foot
13 4-070041-00 4 • Screw, PAN, 8-32 x 1/2, sems (Attach feet to chassis)

9-72 Puritan Bennett 800Series Ventilator System Service Manual


Parts list 9
9

Figure 9-26. 806 Compressor unit

Puritan Bennett 800 Series Ventilator System Service Manual 9-75


9 Parts list

9.4.9.1 806 compressor base assembly

806 compressor base assembly parts list

Item no.
Part no. Quantity Description
(Figure 9-28)

1 4-076299-00 1 • Cable, ground


2 4-00 1138-00 1 • Lockwasher, #8 (not shown)
3 4-005482-00 1 • Nut, captive, 8-32, zinc (not shown)
4 -- 1 • Assembly, compressor (Refer to appropriate 806 compressor
preventative maintenance kit.)
5 4-075214-00 4 • Lower stopper
6 4-003941-00 4 • Washer, flat
7 4-007654-00 4 • Lockwasher, SR, zinc PL 1/4"
8 4-074339-00 4 • Screw, hex head, 1/4 - 20 x 2
9 4-079058-SP 1 • Assembly, solenoid valve
4-076257-00 1 • • Outlet filter
10 4-076319-00 1 • Clip holder, air dryer assembly
11 4-072623-00 3 • Screw, FH, 8-32 x 3/8
12 4-076220-00 1 • Baffle, sound, base plate
13 4-076241-00 1 • Foam kit, sound, base
14 4-079064-SP 1 • Assembly, base (includes air dryer clamp assembly)
15 4-076289-00 1 • Air dryer filter and muffler assembly (Included in the
preventative maintenance kit)
16 4-079059-SP 1 • Assembly, air dryer
17 4-079067-SP 1 • Assembly, coalescing filter
18 4-076728-00 1 • Coalescing filter element (also included in PM kit)
19 4-076219-00 1 • Bracket, H2O water trap, 806
20 4-076291-00 1 • Elbow, 1/4 NPT x 3/8 TU
21 4-070041-00 18 • Screw, 8-32 x 0.38 (not all screws shown)
22 4-000697-00 2 • Screw, pan, 1/4-20 x 1/2, zinc
23 4-076210-00 1 • Plate, motor mount
24 -- 1 • Capacitor, motor (Refer to appropriate preventative
maintenance kit.)
25 -- 1 • Clamp, capacitor, motor, 1.75" diam (120V & 100V)
-- • Clamp, capacitor, motor, 1.379" (220V)
26 4-075215-00 4 • Upper stopper
27 4-076318-00 4 • Pop rivet (not shown)
28 4-075247-00 4 • Sleeve, shock mount
29 4-076290-00 4 • Vibration mount
30 4-076282-00 1 • Pressure relief valve, 36 PSIG (not shown)

9-76 Puritan Bennett 800Series Ventilator System Service Manual


Parts list 9

Figure 9-28. Compressor unit enclosure assembly

Puritan Bennett 800 Series Ventilator System Service Manual 9-79


9 Parts list

9.4.9.3 806 compressor unit plenum assembly

806 compressor plenum assembly parts list

Item no.
Part no. Quantity Description
(Figure 9-30)

1 4-002568-00 8 • Screw, pan, 6-32 x 2


2 4-001056-00 8 • Washer, #6
3 4-079065-SP 1 • Assembly, outlet air tube
4 -- 1 • Nut, dome, strain relief
5 4-074505-00 1 • Strain relief
6 4-076724-00 1 • Spacer, strain relief
7 -- 1 • Strain relief
8 -- -- • Wire saddle
9 4-079066-SP 1 • Assembly, heat exchanger
10 4-076212-00 1 • Plate, heat exchanger
11 4-072623-00 4 • Screw, FH, 8-32 x 3/8
12 4-070041-00 1 • Screw, 8-32 X 0.5
13 4-0776236-00 2 • Fan cord
14 4-070034-00 4 • Screw, 6-32 x 0.38
15 4-076297-00 1 • Cable, data, DC
16 4-076287-00 1 • Power cord, AC, 806
17 4-019135-00 2 • Cooling fan, 115V
(Included in
4-020303-00 PM kit) • Cooling fan, 100V

4-020305-00 • Cooling fan, 220-240V

9-80 Puritan Bennett 800Series Ventilator System Service Manual


Parts list 9

Figure 9-29. Compressor plenum assembly

Puritan Bennett 800 Series Ventilator System Service Manual 9-81


9 Parts list

9.4.10 Backup power source (BPS)

BPS parts list

Item no.
Part no. Quantity Description
(Figure 9-31)

-- 4-070520-SP -- Backup power source (BPS) (See Section 9.4 for NHA.)
1 4-074678-00 1 • Bezel, BPS, front
2 4-072510-00 2 • Screw, PAN, 6-32 x 1/4, black oxide (Attach bezel)
3 4-074677-00 1 • Panel, BPS front
4 4-071599-00 4 • Screw, FH, 6-32 x 1/4 (Attach front panel)
5 4-074532-00 1 • Fuse, 15 A, fast-acting, cartridge-type
6 4-000003-00 1 • Tie wrap, small
7 4-070523-SP 1 • Battery pack
8 4-074665-00 1 • Floor assembly, BPS enclosure
9 4-079031-SP 1 • Chassis with cable, BPS
10 -- 1 • Standoff, battery
11 4-003443-00 1 • Coupling
12 4-070062-00 2 • Screw, FH, 6-32 x 1/2 (Attach floor to cover)
13 4-016142-00 4 • Nut, HEX, 6-32, keps (Attach PCB to chassis)
14 4-076727-SP 1 • PCB, backup power source (BPS)
15 N-7600028 1 • • Fuse, 15 A, fast-acting

16 4-071599-00 6 • Screw, FH, 6-32 x 1/4 (Attach floor to cover)


17 4-004994-00 2-1/4 in. • Molding, plastic (Available in multiples of 1-ft lengths.
(5.7 cm) Minimum order is 10 ft.)
18 4-072505-00 1 Label, display, BPS, non-US

4-075430-00 Label, display, BPS, US

4-075427-00 1 Label, product information, BPS, English, US (not shown)

4-072504-00 Label, product information, BPS, English/French (not shown)

4-075071-00 Label, product information, BPS, German (not shown)

-- 4-075146-00 Label, product information, BPS, Japanese (not shown)

4-075412-00 Label, product information, BPS, Portuguese (not shown)

4-075085-00 Label, product information, BPS, Spanish/Italian (not shown)

4-075098-00 Label, product information, BPS, Spanish/Portuguese (not shown)

9-82 Puritan Bennett 800Series Ventilator System Service Manual