Professional Documents
Culture Documents
Iodimetria PDF
Iodimetria PDF
pubs.acs.org/ac
ppm I (Figure S9). Squares 3−5 quantify iodate in the range of color measurement is particularly challenging if there are
0−4 ppm I; squares 6−8 quantify iodate in the range of 3−7 ppm variations of images captured in sun, shade, incandescent or
I; square 9 quantifies iodate in the range of 7−11 ppm I; and fluorescent lighting, and so on. Images were therefore acquired in
square 10 quantifies iodate in the range of 11−15 ppm I (see a light box (Supporting Information p S-3) that provided
Figure 3 for calibration curves). Square 11 is a limit test that consistent illumination. The blue color in the central circle of
detects iodate in concentrations greater than 30 ppm I (Figure each reaction area was measured by computer image analysis, see
S9). This level of iodine corresponds to overiodized salt that Figure 3. Each reaction zone displays a sigmoidal increase in
could present a health risk to some consumers. Square 12 is a
positive control containing potassium iodate, potassium iodide,
tosic acid, and starch to produce a blue response no matter what
test solution is applied. Additional features on the test card
include a QR code, fiducial marks, and color standards to
facilitate computer image analysis.
Calibration of device. Iodate solutions in a 3.7 M sodium
chloride matrix were run on the test card in order to quantify
iodine levels in fortified table salt (Figure 2). Quantification by
The average absolute interoperator precision based on eye ppm I (about 3SD of the method); these samples were retained
analysis was 0.5 ppm I, as evaluated by eq 1 below: in the determination of the analytical metrics. The average
n absolute accuracy for the computer image analysis of the PADs
∑i = 1 |X1̅ − X̅ 2 |
was 0.9 ppm I as calculated by eq 2 where Xreal is the known
n (1) concentration, Xmeasured is the saltPAD response as measured by
where X̅ 1 is the average response of 5 PADs from expert 1; X̅ 2 is ImageJ, and n is the number of saltPADs used (n = 106). The
the average response of 5 PADs from expert 2; and n, the total main source of inaccuracy is systematic underestimation of the
number of solutions containing a different concentration, is 11. iodine concentration at high iodine concentrations (Figure 4b;
The accuracy was calculated by eq 2, see Figure S10 for a Bland-Altman plot). In the high
concentration ranges, which are less critical for analysis of
n
∑i = 1 |X real − X measured| minimum iodine content in market samples of salt, there is only a
n (2) single point measurement, not a triplicate measurement as for the
lower concentrations. Another cause of this systematic error is
where Xreal is the known concentration, Xmeasured is the saltPAD the camera software. Camera software often adjusts the exposure
response as measured by eye, and n is the number of saltPADs time of images to make them pleasing to the eye. When many
used (n = 110). dark blue circles appear, the software adjusts the coloration of the
When 2 newly trained users analyzed the same 110 images, image (compare the 50 ppm image with the 10 ppm image in
they achieved an average accuracy of 1.4 ppm I and an average Figure S7), which lowers the apparent iodine concentration.
precision of 0.9 ppm I (Figure 4a). The precision was determined Using a camera with manual exposure could help avoid this issue.
by using eq 3, The average precision of the test card (method precision) for
n computer image analysis was determined using eq 3. For samples
∑i = 1 SD
in the full range of 0−15 ppm I, the average precision is 0.9 ppm I,
n (3)
and for samples in the range of 0−7 ppm I (where the test card
where SD is the standard deviation of the 10 replicates per includes triplicate reaction zones), the average precision is 0.3
sample and n is the number of unknown concentrations (n = 11). ppm I. These measurements show that the computer image
analysis is more accurate and precise than visual estimates of
concentrations by newly trained users but also that expert readers
can surpass the accuracy and precision of the computer image
analysis program.
There are advantages and disadvantages to visual interpreta-
tion and computer image analysis of the cards. Visual analysis
increases the usability in low resource settings but introduces
human error in reading or recording results. Capturing the data
with a cell phone camera and interpreting the data electronically
preserves the primary data and facilitates centralized record
keeping through the mobile phone network. This would
empower monitoring agencies to track the quality of salt in
different geographical areas over time.
Robustness. The robustness of the PAD was tested to see
how storage and different water sources affected the response.
Figure 4. (A) Visual analysis of the saltPAD: Each data point is the SaltPADs were wrapped in aluminum foil to exclude light,
average of the visual reads of 10 images by one of two newly trained vacuum packed in Ziploc baggies, and stored in a 40 °C
analysts. Error bars show one SD. The data set is the same as the one convection oven to accelerate the aging of the test cards.19 The
used for computer image analysis. The true values for both analysts were saltPAD response was tracked over time by analyzing low,
identical, but the average and error bars for Analyst 2 are offset slightly to moderate, and high concentration standards as seen in Figure 5.
ease the reading of the data. Analyst 1 had an accuracy of 1.7 ± 1 ppm I, Of 23 samples, 18 assayed within 2 ppm. To test how alternate
and Analyst 2 had an accuracy of 1.2 ± 0.7 ppm I. (B) Computer analysis water supplies may affect the analysis, an 8.0 ppm I (from iodate)
of the saltPAD: each data point is the average of 8−10 replicates per standard was made up in a matrix of 3.7 M salt using tap water
concentration. Error bars show one SD. Four PADs gave discrepant
with high mineral content (170 ppm calcium) and also in a
results and are not included; 4 samples gave an absolute error >3 ppm
but are included in the analysis. matrix of water from St. Mary’s Lake on Notre Dame’s campus
(high content of natural organic matter). More information
about the composition of these water samples is listed in Table
In the computerized image analysis, the intensities of the S3. The solutions were then run on the PADs. The standards
colors in the indicator dots were measured using ImageJ and made in hard water had an error of 8% (n = 2) while the standards
concentrations were calculated using the calibration curves in the lake water produced an error of 17% (n = 2).
shown in Table 1. Four of the 110 saltPADs gave discrepant Comparison with other field assays for iodized salt.
results (color intensities that did not fall within any of the Current methods for iodine measurement in salt include
calibration ranges) and could not be quantified. Based on the glassware titration, rapid test kits (RTKs), and spectroscopic
appearances of these cards, we believe these discrepant PADs test kits. Each method requires trade-offs in terms of accuracy,
contained fabrication errors (e.g., wax lines that failed to seal cost, and ease of use. When conducted by a skilled operator in a
properly or reagent spotting errors). In a field analysis situation, well-controlled lab setting, titration with a buret gives excellent
these samples would have had to be reassayed on new cards. Four accuracy and precision. However, under field conditions in
other samples assayed at greater than an absolute error of 3.0 developing world settings, the accuracy and precision are
D DOI: 10.1021/ac504269q
Anal. Chem. XXXX, XXX, XXX−XXX
Analytical Chemistry Article
compromised by practices such as use of old or impure reagents, cards could be used to detect antibiotics that contain less than
lack of calibration or service for analytical balances, and use of 83% of the labeled API content.
overconcentrated titrant.21 Rapid test kits, in which test reagents Suitability for use in low-resource settings. The World
are applied directly from dropper bottles to solid salt to detect the Health Organization uses ASSURED (affordable, sensitive,
presence of iodate, are very inexpensive, but an external specific, user-friendly, rapid and robust, equipment free, and
validation study conducted by the WHO showed that the tests deliverable to the end user) criteria to assess devices meant for
do not give reliable quantification.20 In the past decade, use in low resource settings. The ASSURED criteria were
spectrophotometric assays using the characteristic purple color considered at every stage of design of the test cards. For our
of the starch-triiodide complex have been used to monitor salt current in-house fabrication process, materials costs are $0.09
quality. These assays show good analytical metrics, but they have USD (see Supporting Information Table S5 for breakdown of
a limited quantification range and require purchase of both a costs) and labor about $0.50. This cost is already less than the
specialized reader and kits of the necessary reagent chemicals and per-assay cost of kit-based spectrophotometric iodate assays, and
vials.21,22 Table 2 summarizes the reported analytical metrics, we think it could be greatly reduced if manufacturing were
costs, and technical expertise required to use each of these automated using a roll-to-roll process. The paper test card
methods and the saltPAD test card. performs iodine analysis with accuracy and precision that rivals
Utility of test card for other analyses: iodometric back- other technology currently used in salt fortification plants in the
titration of amoxicillin. Iodometric titration is a workhorse developing world, and it can perform an iodometric back-
technique for quantitative analysis.23 The United States titration with sufficient accuracy to detect substandard anti-
Pharmacopeia (USP) section ⟨425⟩ contains a titrimetric biotics. The salt assay is easy to perform; a one page set of
procedure for quantification of β-lactam antibiotics, which can instructions and a set of comparison images is sufficient guidance
be used to ensure the quality of a finished pharmaceutical even for first time users. For salt analysis, the user does not have
product. We studied whether the paper test card analysis could to handle any chemicals other than salt and water. The test can be
be useful for quality testing in low resource settings where run under conditions (high heat, humidity, small amount of
analytical chemistry laboratories are rare and substandard space) likely to be found in a small salt fortification plant. The
antibiotics are unfortunately all too common.24 sample can be prepped and run on the test card in about 5 min.
Solutions with amoxicillin concentrations between 0 and 0.9 The test cards can be easily transported or mailed to the end user
mg/mL underwent the chemical transformations shown in and withstand even the conditions of a warm shipping van.
Reactions 4 and 5, and then were diluted 2:25 to get into the Finally, the test cards can be analyzed using computer image
working range of the saltPAD. The triiodide remaining in each processing of mobile phone photographs, which enables use of
test solution was titrated in triplicate on the saltPAD. The test the mobile phone network for data collection and archiving.
card response (Figure 6) showed good differentiation of levels of While a full assessment of how well the paper analytical device
amoxicillin that differed by 0.15 mg/mL. This shows that test meets the ASSURED criteria will require field testing of
a
See Table S6 for calculations of imprecision. bThe study was conducted as a categorical response and it is unclear if replicates of the same solution
were used, so precision data is not definable.
E DOI: 10.1021/ac504269q
Anal. Chem. XXXX, XXX, XXX−XXX
Analytical Chemistry Article
manufactured products, the results from the prototype device that they could not be quantified; four other cards gave results
suggest that adequate analytical performance for analysis of some with absolute errors greater than 3 ppm. The card images were
fortified foods and pharmaceuticals can be obtained with this also evaluated blind by two newly trained readers. Each reader
robust and easy to use paper platform. watched a Powerpoint presentation (Figure S11) that explained
■ EXPERIMENTAL SECTION
Materials. 2% Starch indicator (BDH-VWR International),
how to evaluate the different reaction areas including the
controls, the triplicate reaction areas, and the single range
reaction areas. Before evaluating the blinded images, the readers
p-toluenesulfonic acid (tosic acid, Sigma-Aldrich), potassium then took a quiz (Figure S11) in which they classified ten
iodide (Alfa Aesar), cadmium chloride (Alfa Aesar), sodium unknowns. Each reader scored at least 8/10 correct on the quiz.
nitrite (Mallinckrodt), anhydrous sodium thiosulfate (J.T. Stability testing. Accelerated aging of the test cards was
Baker), secondary standard of potassium iodate, 100.2% (J.T. simulated by storage at 40 °C in a convection oven.19 The cards
Baker), sodium chloride (Macron), amoxicillin (Sigma-Aldrich), were run with standard iodate solutions.
0.01 N iodine solution (Alfa Aesar), hydrochloric acid (Fisher), Back-titration to quantify amoxicillin. Solutions of
sodium hydroxide (Fisher). The paper source was Ahlstrom 319 amoxicillin were created in deionized water at concentrations
(Midland Scientific). The wax printing was done with a Xerox ranging from 0 to 1.0 mg/mL. To 2.0 mL of each solution, 2.0 mL
ColorQube 8570N. A Biomek FX (Beckman Coulter) liquid of 1.0 M NaOH was added and allowed to sit for 15 min. After
transfer robot was used to create the saltPADs. A convection acidifying with 2.0 mL of 1.2 M HCl, 10.00 mL of 0.0050 M
oven (GCA Corporation) was used to perform stability studies. triiodide was pipetted into the solution and allowed to sit for an
Fabrication of saltPAD. SaltPADs were created by printing additional 15 min. Diluting the solutions 2:25 with DI water
onto Ahlstrom 319 paper using solid wax ink.7,25,16,17 The allowed the response on the test card to appear in the most
designs are created in and printed from Adobe Illustrator. Labels accurate and precisely quantified ranges. New calibration curves
and alignment marks were printed with a laser printer; then the for the iodine response were created due to the matrix change
thick and thin lines for the reaction zones and the back side from salt solutions to water solutions. The slopes of the two sets
hydrophobic barrier were printed on the wax printer. The layers of calibration curves were statistically equivalent (95%
of wax were melted to form hydrophobic barriers by baking at confidence level) but the intercepts were not equivalent (see
100 °C for 14 min. A solution deposition robot (Biomek FX) Table S4).
deposited 60 aliquots of reagent solutions into the loading zones
on each test card at a rate of 30 test cards per hour. The spotting
error by the robot is approximately 1%. Any missed spots (visible
■ CONCLUSIONS
A paper card that stores multiple reagents and recombines them
as whiter areas against the gray background of damp loading through surface-tension-enabled mixing can carry out an
zones) were pipetted by hand. iodometric titration. This paper analytical device performs
Running and analyzing the test cards. To create a test accurate and precise measurement of iodate in iodized salt. The
solution, the user dilutes 3.25 g of salt into 15 g of water, which is test outcomes can be read directly by eye with an accuracy and
a 1:5 dilution. Test cards were dosed with 125 μL of solution per precision of under 1.7 ppm, or by electronic image analysis with
reaction zone using an Eppendorf pipet, and then they were an accuracy and precision of under 1 ppm I (expressed as mg
gently shaken by hand for 3 min. This requires some practice iodine atoms/L) over a range of 0.8−15 ppm I. Four out of 110
because if the paper is shaken too vigorously, solution will spill cards gave discrepant readings, probably due to fabrication
out of the reaction zones. Using an orbital shaker with a 22 mm errors. In addition to performing the iodometric titration, the
diameter orbit, 4 revolutions per second provides great mixing card can detect analytes such as iodide and contains positive and
but at 5 revolutions per second the solution spills. Images were negative controls. The iodometric titration test card can be used
taken in a light box equipped with two strings of plug-in 3 W LED to assay other redox-active analytes, including analytes in
lights that provided a light output of 162 Lumens and were rated solutions of lower ionic strength than the iodized salt solutions.
82 on the Color Rendering Index. Images were taken with an When used to perform the iodometric back-titration called for in
iPhone4 and imported into ImageJ for analysis.26 The images a pharmacopeia analysis of β-lactam antibiotics, the test card
were inverted, and the average unweighted gray value over the provided good differentiation among amoxicillin solutions that
entire reaction zone circle was measured. varied by 0.15 mg/mL, which could enable field detection of
SaltPAD calibration curves. Working iodate standards substandard β-lactam pharmaceuticals. Test cards able to
ranging from 0 to 15 ppm I in 3.7 M NaCl were created by perform quantitative chemical analysis in settings outside the
dilution from a stock potassium iodate solution. The working laboratory could help to solve quality and regulatory compliance
problems in the developing world.
■
standards were pipetted on the saltPADs, and the blue color of
the PAD response was measured with ImageJ to create
calibration curves. ASSOCIATED CONTENT
Internal lab validation. After construction of the calibration *
S Supporting Information
curves in Figure 3, an in-lab study was conducted using the image Fabrication details, full set of standard images for known iodine
analysis as a tool to assess the accuracy and precision of the concentrations, additional analysis of data from expert readers
PADs. 3.7 M sodium chloride brine was spiked with 11 levels of and computer image analysis, calibration for amoxicillin analysis,
iodate, and the coded samples were analyzed blind by two Adobe Illustrator print files, and Powerpoint presentation used
operators, who each independently applied test samples, ran test for training card readers. This material is available free of charge
cards, visually interpreted the test card by comparison to printed via the Internet at http://pubs.acs.org.
standard images, and then used ImageJ to evaluate test card
results for 5-fold replicates. Four of the resulting 110 images gave
discrepant results (e.g., one or more of the low range dots was
■ AUTHOR INFORMATION
Corresponding Author
uncolored, and higher range dots were colored) which meant *Phone: (574) 631-4665. E-mail: mlieberm@nd.edu.
F DOI: 10.1021/ac504269q
Anal. Chem. XXXX, XXX, XXX−XXX
Analytical Chemistry Article
Notes
The authors declare the following competing financial
interest(s): M.L. and N.M. are inventors on a pending U.S.
patent application describing this technology.
■ ACKNOWLEDGMENTS
N.M. was the recipient of a Global Fellowship from US
Pharmacopeia. We acknowledge the Notre Dame Faculty
Research Support Program and the Global Alliance for Improved
Nutrition (GAIN) for their financial support. A special thank you
is extended to our main contacts at GAIN, Rebecca Spohrer and
Rizwan Yusufali. Undergraduate researchers Evan Graham,
Kellie Raddell, and Emily Mediate developed and tested earlier
versions of the test cards. Doa’a Aldulaimy assisted in fabricating
the test cards used in this study. Michelle Pillers and Abigail
Weaver served as card readers.
■ REFERENCES
(1) Martinez, A. W.; Phillips, S. T.; Whitesides, G. M. Anal. Chem.
2010, 82, 3−10.
(2) Martinez, A. W.; Phillips, S. T.; Butte, M. J.; Whitesides, G. M.
Angew. Chem., Int. Ed. Engl. 2007, 46, 1318−1320.
(3) Noh, H.; Phillips, S. T. Anal. Chem. 2010, 82, 8071−8078.
(4) Li, X.; Ballerini, D. R.; Shen, W. Biomicrofluidics 2012, 6, 011301.
(5) Vella, S. J.; Beattie, P.; Cademartiri, R.; Laromaine, A.; Martinez, A.
W.; Phillips, S. T.; Mirica, K. a; Whitesides, G. M. Anal. Chem. 2012, 84,
2883−2891.
(6) Muller, R. H.; Clegg, D. L. Anal. Chem. 1949, 21, 1123−1125.
(7) Yagoda, H. Ind. Eng. Chem. 1937, 9, 79−82.
(8) Weaver, A.; Reiser, H.; Barstis, T. L. O.; Benvenuti, M.; Ghosh, D.;
Hunckler, M.; Joy, B.; Koenig, L.; Raddell, K.; Lieberman, M. Anal.
Chem. 2013, 85, 6453−6460.
(9) Fu, E.; Lutz, B.; Kauffman, P.; Yager, P. Lab Chip 2010, 10, 918−
920.
(10) Karita, S.; Kaneta, T. Anal. Chem. 2014, 86, 12108−12114.
(11) Lisowski, P.; Zarzycki, P. K. Chromatographia 2013, 76, 1201−
1214.
(12) Sonntag, O. In Laboratory techniques in biochemistry and molecular
biology; van der Vliet, P. C., Ed.; Elsevier:Amsterdam, 1993.
(13) Scida, K.; Li, B.; Ellington, A. D.; Crooks, R. M. Anal. Chem. 2013,
85, 9713−9720.
(14) Fujita, H. J. Phys. Chem. 1952, 56, 625−629.
(15) Colin, H. H.; de Claubry, G. Ann. Physik 1814, 48, 297.
(16) Carrilho, E.; Martinez, A. W.; Whitesides, G. M. Anal. Chem.
2009, 81, 7091−7095.
(17) Lu, Y.; Shi, W.; Jiang, L.; Qin, J.; Lin, B. Electrophoresis 2009, 30,
1497−1500.
(18) Arthur, P.; Moore, T. E.; Lambert, J. J. Am. Chem. Soc. 1949, 71,
3260.
(19) Anderson, G.; Scott, M. Clin. Chem. 1991, 37, 398−402.
(20) Pandav, C. S.; Arora, N. K.; Krishnan, A.; Sankar, R.; Pandav, S.;
Karmarkar, M. G. Bull. World Health Organ. 2000, 78, 975−980.
(21) Rohner, F.; Garrett, G. S.; Laillou, A.; Frey, S. K.; Mothes, R.;
Florian, J.; Locatelli-Rossi, L. Food Nutr. Bull. 2013, 33, S330−S335.
(22) Dearth-Wesley, T.; Makhmudov, A.; Pfeiffer, C. M.; Caldwell, K.
Food Nutr. Bull. 2004, 25, 130−136.
(23) Harris, D. C. Quantitative Chemical Analysis, 8th ed.; W.H.
Freeman and Company: New York, NY, 2010; pp 351−356.
(24) Newton, P. N.; Green, M. D.; Fernández, F. M. Trends Pharmacol.
Sci. 2010, 31, 99−101.
(25) Yagoda, H. U.S. Patent 2,129,754, 1938.
(26) Ferreira, T.; Rasband, W. ImageJ User GuideIJ 1.46; imagej.nih/
gov/ij/docs/guide.
G DOI: 10.1021/ac504269q
Anal. Chem. XXXX, XXX, XXX−XXX