Professional Documents
Culture Documents
10 - Chapter 3
10 - Chapter 3
Abstract
discussed.
CHAPTER III
INSTRUMETATION TECHNIQUES OF INFRARED AND
RAMAN SPECTROSCOPY
3.1 INTRODUCTION
and thus yielding new information about the atomic or molecular species, or
50
can be viewed as a communication device between the system under study
and high energy sources like lasers, which permit spectrum storage and
simulation, automatic control of slit, etc. Often these are under the control of a
during vibration. The intensity of the infrared and Raman spectra mainly
51
is currently more popular as compared to other physical techniques (X-ray
unknown compounds.
spectrum which lies between the visible and microwave regions. The infrared
from about 12,800 to 10 crn' or wavelengths from 0.78 to 1000 urn. From
instrumentation and application point of view, the infrared region has been
rotation region) and far IR region (rotation region). The techniques and the
applications of methods based upon the three infrared spectral regions differ
infrared spectrum, apart form the prominent modes of vibrations, there are
other vibrations due to coupling and overtones that may lead to additional
species can be rationalized by assuming that they all arise from various
radiation is thus confined largely to molecular species that have small energy
52
these circumstances can an alternating electrical field of the radiation interact
with the molecule and cause changes in the amplitude of one of its motions
[101,102].
3.3 INSTRUMENTATION
3.3.1 Introduction
3.1. A source provides radiation over the entire range of the infrared
spectrum. The monochromator disperses the light and then selects a narrow
wave number range. The detector measures the energy and transforms it into
recorder. The FTIR spectra of most of the samples were recorded in mid and
far IR region and are used for the present investigation [7, 61,103-106].
Radiation
source
. H Sampling area 1-1 Monochromator ~~
condensing optics
infrared spectrophotometer
3.3.2 Source
from a binder and oxides of thorium, cerium, zirconium and yttrium. The
diameter. The maximum radiation for the Globar occurs in the 5500-5000
ern" region. Nichrome wire, carbon arc, rhodium wire and tungsten filament
53
In a commercial infrared spectrometer either a nichrome wire or a
3.3.3 Monochromator
prisms or diffraction grating. Materials for prism construction, which are found
most suitable in the infrared region, are listed in the following Table 3.1.
Table 3.1
Lithium fluoride
Calcium fluoride
from different optical materials, so that each could be used in sequence in its
54
Si02, NaCI and KBr prisms. Low-cost instruments use a NaCI prism over the
full range. They give highest resolution in the vital finger - print region.
radiation according to its wave number components (ii) it restricts the radiation
falling on the detector into a narrow wave number range, and (iii) it maintains
[61]. Some instruments use a double monochromator. That is, the exist slit of
the first monochromator serves as the entrance slit for the second
3.3.4 Detector
plotted on a chart recorder. Two Types of the detectors are commonly used:
a. Thermal Detectors
to the intensity of the incident radiation over the whole spectral range of the
instrument. The most desirable features of the detectors are the closeness
with which they approach the behavior of a black body, high sensitivity, high
55
The infrared detectors may be selective or non-selective. The selective
fabricated from crystals are known as pyroelectrics and take less time than
other thermal detector. Hence with these crystals, radiation can be chopped
at a higher rate. However, these are more expensive and not widely
employed.
b. Photon Detector
photon that strikes an electron in the detector can raise that electron from a
photoconductive cell.
or Hg-Cd- Te (MCT), which have high sensitivity and very quick response
temperature which reduces noise. The radiant energy falling on the detector
56
Photovoltaic Cells are essentially electrical resistors, which decrease
in resistance in relation to the intensity of light striking their surface and are
constructed from a thin layer (0.1 u) of semiconductor, like lead sulphide and
lead tin telluride. These detectors extend spectral sensitivity up to 3.5 and 6 Jl.
respectively [104].
number of ways. Polyethylene pellets are used for recording the far IR
spectra.
57
Solidsare sampled in a wide variety of ways. If the sample is soluble, it
pressed
pellet of alkali halide, usually potassium bromide. Sample is mixed
transparent
plate or disc which can be inserted into the spectrophotometer.
mullingagent. KBr does not absorb infrared light in the region 2.5-15 urn and
Solid samples have also been examined in the form of a thin layer
powderedsample is mixed to form a paste with little heavy paraffin oil. The
mullis sandwiched between salt plates for measurement. Mulls are formed by
b. Liquids
pure samples can be measured as very thin films squeezed betwee~ \'(1Q
alkali halide windows of a demountable cell. This technique can produce a
film of thickness 0.01mm or less. This method is most useful for qualitative
work only because the sample thickness cannot be controlled. Liquid cells
58
consist of two alkali halide windows usually NaCI or KBr, separated by a
spacer of suitable thickness made of Teflon or lead which limits the volume of
C. Gases
cells ranging from a few centimeters to several meters that can be directly
placed in the path of the infrared beam. The end walls of the cell are usually
vibrational changes in the gaseous phase often split the high frequency
D. Solvents
range are available and the spectra of the sample dissolved in carbon
two parts: (a) an optical system, which uses an interferometer, and (b) a
59
plots the spectra. A schematic diagram of the essential components of a FT
mirror at a constant velocity. Between these two mirrors is beam splitter set at
45° from the initial position of the movable mirror. A parallel beam of radiation
from an infrared source is passed to the mirrors through the beam splitter.
The beam splitter reflects about half of the beam to the fixed mirror, which
reflects it back to the beam splitter and transmits the other half to the movable
mirror, which reflects it back to the beam splitter. The returning beams are
again split and mixed about half going back to source and half passing
transmits most of the desirable radiation. In the far infrared region, germanium
terephthalete) (650-10 ern") are used as beam splitters. A thin film of the
The return beams from both the mirrors along the same path length as
their incident path are recombined into a single beam at the beam splitter. The
path length of one of the return beams is changed in order to create phase
directed through the sample and focused on to the detector. The detector
60
The movable mirror can be moved in a range of say ± 5 cm. The mirror
detectors with response times short enough to detect and transmit rapid
changes to the recorder. The detector used in conjunction with rapid scanning
energy passes through the instrument and the resolving power is constant
over the entire spectrum. The signal to noise ratio is also improved [109].
61
SrirtitlnalY
lU1n~' 'ufmred 'S(ttl rce
'~~
IBtfSfi11:f>,lll1r11f
0\
N
F1ft
Diti;~tQr
SormtlTe
c;lj).m"",l~nS
r;;;w:·~'~~~-.,
t. t
~.·I::
t' *.--=-"
f
I
q •
L __,__"",~j
C:olllm~ttDg
mirror
lasers and computers. The advantages of lasers are their high intensity, high
the time of invention of Raman Effect, both infrared and Raman spectra of
molecular structure and also for the quick identification of the presence of the
that water does not cause interference; indeed, Raman spectra can be
3.7.1 Instrumentation
light and the scattered light is observed at right angles to the incident radiation
63
and a detector. Fig. 3.3 represents schematic diagram of a Raman
spectrometer.
Monochromator H'---_D_e_te_ct_o_r __
Filter
(a) Source
to-noise ratio. Five of the most common lasers along with their
wavelength (nm) used for Raman spectroscopy are; Argon ion (488 or
514.5 nm), Krypton ion (530.9 or 647.1 nrn), Helium / Neon (632.8 nm),
Diode laser (782 or 830 nm) and Nd;YAG (1064nm). Because the
argon and krypton ion sources that emit in the blue and green region of the
spectrum have an advantage over the other sources. Diode and Nd:YAG
wavelength lasers. The first is that they can be operated at much higher
The second is that they are not energetic enough to populate a significant
64
molecules. Consequently, fluorescence is generally much less intense or
effective in eliminating fluorescence. The two line of the diode array laser
simpler than for infrared spectroscopy because glass can be used for
windows, lenses and other optical components instead of the more fragile
source is easily focused on a small sample area and the emitted radiation
~ Liquid Samples
measured directly from the reaction vessel. Water is a good solvent for
recording the Raman spectra. Water absorbs strongly in the infrared but
~ Solid Samples
needed. A few milligrams of the solid samples are required. Solid can be
65
packed into a capillary tubes as a powder. The crystal can be mounted in
different orientation of the crystal. For single crystals, the Raman spectrum
varies depending on the direction of the crystal axis, when polarized light
~ Gas Samples
liquids or solids and hence may require cells of larger path length. The
for a wide variety of sizes and forms of the sample. Samples in gas, liquid
and solid states can be examined easily. In this study the spectral data
were recorded on BRUCKER IFS 66V, FTIR with FRA 106 FT-Raman
Fig 3.4.
(c) Filter
Liquid filters are placed between the source and the samples tube.
Different filters are used for different excitation radiation. Filters made of
66
0/ To isolate a single exciting line
or fluorescence.
Raman lines.
(d) Monochromators
In the monochromator, both lenses and mirrors have been used. Most
(e) Detectors
Basically there are two different ways to detect and record Raman lines.
The easiest way is to gather the scattered light emerging through a glass
then developed and both the line frequencies and intensities can be
Spectrum.
67
I NO. YAG LASER
i INTERFEROMETER
r ':"
J
I
I
I
I
..--. APERTURE
I
I
I
1
t tR
r SOtJIitCE
___________ ...J
I r--'- -----~----1
I I
/-,
\..Y.~ I ~ I
~.
,_'-
~
I -~ ft'· ~
,\r :ri'"'-
I
...,,'"
..
~~
I
I
I
,I
FILfER
MODULE
0\
,,-,_L___ t ly: (
6
00 I __ -4-
I ~.-r--- ----T-
I
L': -.:~ --'-L---.--
I
-------
- I
I I
, I ~_,-----
t¥1 : 1ft SAMPLE I p~I _ RAMAN
.!::::::::!r 1 POSITION I 1>------ OETECTOR
IR DETECTORS: . : ! r I
I I I t ~~~ j
SAMPLE
COMPARTMENT
69