You are on page 1of 22

LAMIDEY NOURY MEDICAL

Electrosurgical Generators

SURGILEC MC 2

User Guide
Notice ref : SURGILEC MC2 -NU 0105N001 UK

LAMIDEY NOURY MEDICAL Z.A. Les Godets - 3, rue des Petits Ruisseaux
91370 Verrières-Le-Buisson (France)
Tél : (33) 01.69.20.30.21 - Fax : (33) 01.60.13.97.47
E-mail: info@lamidey-noury.fr
SURGILEC MC2 User guide

IMPORTANT

The instructions herein are the integral part of the high frequency surgical apparatus that you have just
purchased.
They describe the functions of the generator, and must be read carefully before installation and usage.
All the security regulations and warning notices must be scrupulously observed. Make sure that all
members of the surgical team are fully aware of the hazards; understand the principles of the operations and
of the safe handling. No part of this document can be photocopied, reproduced or translated without prior
written consent from LAMIDEY-NOURY MEDICAL. In case of need please contact the Technical assistant.

INDEX

1. INTRODUCTION
1.1. Field of Application.
1.2. Accessories description and compatibility warnings.
1.3. Definitions.
1.4. General description.
1.5. Monopolar Cut.
1.6. Monopolar coagulation.
1.7. Bipolar coagulation.
2. RECEPTION OF THE APPARATUS.
2.1 Reception delivery control.
2.2 Installation of the generator.
2.3 Control before using.
3 PRECAUTIONS AND GENERAL HINTS.
3.1 Instructions for using the High frequency power electro surgery apparatus.
(Meets requirements, standards NF EN 60601-2-2)
3.2 Additional information.
3.3 Advice on the correct plate positioning.
4 CONNECTIONS AND CONTROLS.
4.1 Symbols and pictogram.
4.2 Back panel.
4.3 Front panel.
5 INSTRUCTIONS.
5.1 Starting, alarm patient plate.
5.2 Monopolar cut mode.
5.3 Monopolar coagulation mode.
5.4 Bipolar coagulation mode.
6 ADJUSTMENT AND POWER ACTIVATION.
6.1 H.F.power settings.
6.2 Double foot pedal activation.
6.3 Hand held tactile pencil activation.
6.4 Manilec bipolar forceps activation.
6.5 Audible indicator setting.
6.6 Does and don’ts
6.7 Inert gas activation.
7 TROUBLE SHOOTING
7.1 Unit responds to controls.
7.2 No response from the controls.
7.3 Generator alarm sounds
8 MAINTENANCE.
8.1 Generator.
8.2 Accessories and instruments.
8.3 Service.
9 TECHNICAL SPECIFICATIONS.
9.1 Mechanical and environmental specifications
9.2 Input specifications
9.3 Output specifications
10 Appendix.

LAMIDEY NOURY Medical Page 2/22


SURGILEC MC2 User guide

1 INTRODUCTION
1.1 Field of application.
The Surgilec MC2 generators utilisation is reserved for qualified surgeons. This electrosurgical
unit is to be used intermittently for the surgical operations in the operating theatre.
Used for the following surgical procedures:

Specialties

Out-patients surgery 

Cardiac and thoracic surgery 


Gynaecology and Obstetrics 
Orthopaedics 
Paediatrics 
Plastic and reconstructive procedures 
Vascular surgery 
Visceral and digestive surgery 
Laparoscopic surgery 
Dermatology 
Endoscopy digestive 
Neurosurgery 
Odontology 
Otorhinolaryngology 
Urology 

 Recommended  Usable  Inadvisable

1.2 Accessories description and compatibility warnings.


1.2.1 List of standard and optional compatible accessories.
The following accessories are proven and compatibility tested with the generators SURGILEC. In
the case of using accessories other than listed below, see the paragraph 1.2.2.

References Description
11SM2F Watertight foot pedal bi-circuit for mono and bipolar operations
11K250 Connection cable, disposable length 5m for patient adhesive plate
PLAD2, PLEN2, Adhesive patient split-plate, disposable for adult and child
PLAD2K2, PLEN2K2, Adhesive patient split-plate, disposable for adult and child with cable
11FM40 Cable for foot pedal monopolar forceps operations, sterilizable
11MD25, 11MD40 DERLEC, pencil without cable, foot pedal operation (2-2.4mm or 4mm)
11MC254, 11MC404 DERLEC, Pencil with cable 4m inclu. f.p. operation (2-2.4mm or 4mm)
11MCT9, 11MCT94 TACTILE hand switches pencil, reusable, with cable 4m inclu. (2-2.4mm or 4mm)

LAMIDEY NOURY Medical Page 3/22


SURGILEC MC2 User guide

SUT1, SUTL1 Tactile hand switch pencil, disposable, with electrode needle, 7cm or 15cm
SUC1, SUCL1 Tactile hand switch pencil, disposable with electrode blade, 7cm or 15cm
SUT2, SUC2 Pencil, foot pedal operation, disposable with electrode needle or blade, 7cm

11F242 Cable bipolar, sterilizable, 4m connection 2 pin European


11PM10, 11PM11,… Forceps monopolar all sizes (see catalogue)
11PB11, 11PB13… Forceps bipolar all sizes (see catalogue)
11F341 Cable for forceps MANILEC, Manual control
11PBM1, 11PBM2,… Forceps bipolar MANILEC, manual control

Electrodes See catalogue

1.2.2 Compatibility
Warning: The specification CE mark on the electro surgery accessories don’t guarantee
compatibility with all other generators on the market.
The safety of the patient and the user relies not only on the generator, but also on the
accessories and how they are applied, especially in the case of pedals, plates, electrodes and
pencil holders and their conductors. These have been the object of research, tried and tested to
which the data specifications comply with the CE mark. It is therefore important to utilize the
equipment and accessories, supplied or recommended by the manufacturer, by which the
compatibility with the generator can be proven and certified by the manufacturer or an approved
organization.
In accordance with the norm NF EN 60 601-1(network frequency insulation) regarding the
accessories connected to the patient must be 1500Vrms Minimum.
In accordance with the norm NF EN 60 601-2-2 the electrode cables must resist to a 3000V
network frequency tension, all the pencils, forceps and connector cables, must be insulated HF +
or - 1.5 times the maximum HF output of the generator (on the corresponding outlet)
WARNING: THE USE OF ELECTROSURGERY ACCESSORIES INSUFFICIENTLY ISOLATED RUN
THE RISK OF BURNING, FIRES AND SURGICAL INJURIES.
In accordance to the paragraphs 6.8.2. - (b) of the norm NF EN 60601-1 regarding the security of
the medical electrical equipment, the manufacturer can be held responsible for the effects on the
security, reliability, or the specifications of the equipment or accessories, only if:
- The settings, modifications or repairs have been carried out by the manufacturer or persons
authorized by the manufacturer.
- The electrical installation of the surgical premises, apply within the specifications of the I.E.C.
- The equipment and accessories are used according to the manufacturer’s instructions.
1.3 Definitions
High frequency power: the electrosurgical unit is a HF power generator, according to the norm
NF EN 60 601-2-2, mains voltage output of the electrosurgical generator must be superior or
equal to 300 kHz.
Active electrode: the active electrode is a contact electrode in which the HF power is converted
into thermal energy to obtain the cutting and coagulation effects. The active electrode is often
presented in the form of needles, blade, ball or resection loop.
Neutral electrode: the neutral electrode is an electrode with a large flat surface, placed on the
patient which completes an electrical HF circuit, without producing thermal change. This electrode
is also called an electrode patient plate or dispersive electrode.
Monopolar mode: The application of this mode is when the HF electric current is dispersed
through the patient between the active electrode and the neutral patient plate electrode.

LAMIDEY NOURY Medical Page 4/22


SURGILEC MC2 User guide

Bipolar mode: The application of this mode is when the HF electric current HF is dispersed
through the biological tissues situated between two active electrode tips; the bipolar electrodes
are represented by two jaws of a bipolar forceps
1.4 General descriptions.
The HF power SURGILEC generators are primarily used to obtain a cut or coagulation
(haemostasis), by monopolar or bipolar mode.
The current can be constant during all the operation time from the output circuit, through the
active electrode in monopolar mode or through two active electrodes in bipolar mode.
It is possible to use with the SURGILEC generators, the neutral electrode return plates, a
single or the two sectioned split plate, recommended with monopolar mode.
The Mains power supply is situated at the back of the generator and the power setting touch
controls and visual indicators are found on the front panel.
The automatic control system of the internal parameters, signals all abnormalities.
The last power settings used, are automatically remembered when powered-up. The last
operational settings are also automatically remembered in the case of a power failure which lasts
for less than 20 minutes. If the power failure is more than 20 minutes, it is advised to verify the
last operation selected.
The audible activation indicator is adjustable, so that each user can choose the level desired.
The generator functions with the tactile hand switch pencil or the pencil holder pedal
operation.
1.5 Monopolar cut
The monopolar cut consists of cutting the biological tissue by passing the HF, high-density
current through the active electrode. When the HF current is applied on the tissue by the tip of the
active electrode, this generates intensive heat which bursts the cellule.
The cutting effects are obtained when placing the electrode on the tissue and destroying
therefore the cells one after another. The electrode movement avoids the heat spreading deeper
into the tissue, limiting therefore cell destruction on fine tissue.
The current the best adapted for the cutting, is the pure sinusoidal, without any modulation,
because it makes a very neat cut in producing a reduced amount of thermal and haemostasis
effect. This particular setting allows the user, full security without any necrosis of the surrounding
tissues.
A good coagulation during the cut is one of the principle advantages of the electro surgery;
therefore it is preferable to use a current with a given operating mode, called power haemostatic
cut.
The following indications will help the surgeon to obtain the correct incision:
• Utilise the electrodes: type needle or fine blade ( cutting edge ),
• Keep the tissue humid but not wet,
• Hold the electrode perpendicular to the tissue to be severed,
• Don’t start cutting until the electrode is in contact with the tissue,
• Keep the electrode very clean ( use the electrode scraper pad )
The correct output power allows to not have,
• Any resistance of the movement on the electrode when in contact with the tissue,
• No burning of the cut surfaces,
• No sticking tissue on the electrode.

LAMIDEY NOURY Medical Page 5/22


SURGILEC MC2 User guide

1.6 Monopolar coagulation


The coagulation monopolar provokes capillary vessel haemostasis by the passage of high
frequency current. When the large electrode plate is used to disperse the energy over a large
surface, the high density current is reduced and dries the superficial cells, without penetrating
deeply into the tissue.
The power generally used for coagulation is modulated. This modulation affects the accuracy
of the coagulation, the quality of the haemostasis and the necrosis of the tissue. A higher level of
current modulation produces a thicker coagulation and deeper destruction of tissue, but a better
haemostasis.
A few indications on how to obtain a good coagulation:

• Choose a large electrode, type- ball, blade or large needle,


• After having evacuated the blood, locate the lesion vessel,
• Touch lightly the vessel before activating the HF current,
• Stop as soon as the tissue turns pale, so as not to damage the tissue,
• Keep the electrode very clean (use the electrode scraper pad)

1.7 Bipolar coagulation


The coagulation bipolar or haemostasis is obtained by a reduced-density current which heats
the cellule, ideal coagulation 80°C or slow drying to achieve congealed tissue denatured but not
destroyed. The depth of the coagulation is determined by the power application time. To obtain
the best results for bipolar coagulation:
• Don’t immerse the forceps in the blood, evacuate before coagulation.
• Adjust power and settings to the type of forceps chosen.
• It is better to take more time coagulating with a reduced power, than a quick coagulation
with a stronger power.
• Keep the tissue humid and the jaws of the forceps clean.
• To obtain a coagulation of the tissue between the jaw forceps, these must not touch
themselves.

2 RECEPTION OF THE APPARATUS


2.1 Reception delivery control
On receiving your equipment, inspect carefully for any eventual damage during transport. All
claims will be accepted if they have been notified to the delivery driver or directly to the shipper.
If the equipment is returned, it is important to return it in the original packaging. This is a valid
guaranteed proof.
Remove the apparatus from its packaging and read carefully the enclosed instructions and
information. Ensure that the local supply voltage does not exceed more that 15% of the power
mains rate (230V or 115V). Warning: the timed fuses for 230 and 115V are not calibrated for the
same strength.
2.2 Installation of the generator
The generator’s operating voltage 230V 50Hz (115V 60Hz option) is supplied with a power
cable with a three pin end plug 10 / 16A grounded. With a metallic connector hood, this must be
grounded, for the security of the patient, surgeon and theatre staff. This must be regularly
checked.
IT IS FORBIDDEN TO OPERATE THIS APPARATUS WHICH IS NOT EARTHED.

LAMIDEY NOURY Medical Page 6/22


SURGILEC MC2 User guide

The generator must be placed on a flat, stable surface, at least the same size as the
apparatus, with an area around the apparatus of no less than 25cm. if the unit is placed in an
enclosed area the volume must be no less than 1m3 with protective heat resistant walls.
The stand for the generator must be able to support the weight (see § 9.1) LAMIDEY NOURY
MEDICAL recommend their stainless steel trolley, 2 plateau, without drawers (ref.10GST1) and 2
plateau, with one drawer (ref.10GAT1).
Connection details are described in chapter 4: “Connections and Controls.”
Insert the power cable into the connection at the back of the unit.
Connect the equipotential connection to the area shown at the back bottom left of the unit
with the power supply cable plug (connection optional).
Utilization of the pedal, connect to the appropriate connection at the back of the unit.
Utilization of the hand switch tactile, connection on the front of the unit, the pencil without
controls, connects to the red socket.
Utilization of the bipolar forceps, connect to the socket bipolar.
Before using the unit, it is necessary to place the patient plate on the patient then connect to
the generator. The patient split- plate double and single plate can be used. If the impedance
between the electrode and the skin exceeds certain critical limits, the alarm will sound and stop
the production of the power. (See § 3.3 Advise and 5.1 Activation.)
2.3 Control before using
For the security, a daily general control of the generator and accessories before the first
operation in the day will prevent any possible malfunctions or breakdowns.
2.3.1 Accessory check
Before each operation, assure that all accessories to be used are within easy reach and sterile.
The electro surgery accessories undergo ineluctable wearing, due to mechanical and chemical
factors. Accidental deterioration (knocks, falls, etc…) can happen, therefore a careful visual and
testing check must be conducted before use.
Accessories that show visual signs of wear and tear must not be used, notably – changing of
the accessories shape or connection plug or faulty insulation cables. Excess pulling on the cable
will provoke strain, therefore eventual damage at the mains. Damaged cables must be replaced.
It is possible to test the foot and manual controls, after having adjusted the HF power to “0”.
Press the button on the Yellow cut area (foot control lever) or by pressing the button on the blue
coagulation area (foot control lever) the alarm will sound to show the good working function.
With the accessory TESTELEC (ref. 11VTES), this tests the working function of the controls and
the cable power flow.
WARNING: USING FAULTY ELECTROSURGERY ACCESSORIES, YOU RUN THE RISK OF BURNES,
FIRE RISKS AND MEDICAL INJURY TO YOURSELF, THE PATIENT AND YOUR MEDICAL STAFF.
2.3.2 Generator control
Before the first operation of the day, it is advisable to do a quick visual generator control,
check that the unit is earthed. If the apparatus shows any signs of damage, like knocks or if there
is any deterioration of the connection plug, the unit must not be used, only after verification of the
part or a maintenance control approved by LAMIDEY-NOURY MEDICAL.
• Connect to the patient, a patient plate or a jack plug test ref: 11VJ33.
• Turn the unit on, by precaution do not connect any active electrode with the unit.
• While separating the neutral electrode plate from the cable, check the alarm display
(sounding alarm and visual monitor)
• Check that the digital display lights are on.
• Connect the twin pedal footswitch or the pencil tactile.
• Select the monopolar mode.

LAMIDEY NOURY Medical Page 7/22


SURGILEC MC2 User guide

• Start up successively the operation cut (yellow lever on the double foot pedal or the yellow
button on the hand switch) and the coagulation operation (blue lever or blue button): the
corresponding yellow and blue lights should light up; this produces a two tone warning sound.
• Adjust the HF power to 0, the indicator HF ENERGY during the operation cut and coagulation
should be unlit.
• After, adjust the cut and coagulation power to the maximum level. During the activation of
each function, the indicator HF ENERGY should be lit-up.
• Select the bipolar mode and repeat the same tests as for monopolar mode with the double
pedal operation. The bipolar cut operation will not function.
• Readjust the power levels compatible with the operation to be carried out.

3 PRECAUTIONS AND GENERAL HINTS


3.1 Instructions for using the High Frequency electro surgery apparatus (meets
requirements standards NF EN 60601-2-2)

c1 - The NEUTRAL ELECTRODE PATIENT PLATE must be firmly attached onto the surface of
the patient, and as near as possible to the operating site.
See the §3.4 How to use the neutral electrode plate.

c2 - The PATIENT must not be in contact with the grounded cables or anything in contact
with the ground (for example, operating table, support, etc…). The utilization of antistatic
wrappings is advised.
A common risk is the conduction of the HF current via a fluid. There must not be any fluid
which could make an electric conduct between the patient and the operating table. For that, the
table can be re-covered with a dry surgical sheet or a water-proof under sheet. Check also that
the patient is not excessively painted with antiseptic fluid. The high perspiration areas must be
kept dry.

c3 - The skin to skin contact between the arms and the body of the PATIENT for example,
should be avoided; place dry gauze.

c4 - If using simultaneously the HIGH FREQUENCY ELECTROSURGERY APPARATUS and


other ELECTRONIC THERAPY MONITORING DEVICES on the same patient, the sensors of the latter
should be fitted with HF filters (inductance blocking for example) placed as near as possible to the
sensors. The monitoring electrode needles are forbidden.

c5 - The electro surgery electrode cables, must be placed in a way that there is no contact
with the patient or with other conductors and not to be re-covered around the operating site.
To reduce the risk of interference, avoid positioning the HF cable near to the endoscope camera
cable, for example
.It is recommended to place the ACTIVE ELECTRODES between uses, away from the patient.
In fact, there is a risk of burns to patients, if the electrode is accidentally activated or if the
electrode has not had the time to cool down before being put down. Using a well-insulated sterile
safety holster will protect against these risks.

LAMIDEY NOURY Medical Page 8/22


SURGILEC MC2 User guide

c6 - For the surgical operations on the parts of the body having small confined surfaces, the
use of the bipolar technique is preferred to, in order to avoid undesirable coagulation.
In fact, if the electrosurgical current has to pass an area of confined tissue to the electrode
plate, the high-density in that area risks to coagulate or even burning, for example, in the case of
monopolar coagulation on a testicle, the HF current risks to burn the vas deferens in its path.
c7 - The selected power must be as low as possible for the required purpose.

c8 - A seemingly low power or a fault in the correct operation of the normal functional
settings of the apparatus could indicate an incorrect application of the NEUTRAL ELECTRODE or
the electrode cable is not connected. In this case, before increasing the power settings, the
application of the neutral electrode and connections must be checked.
c9 - The use of flammable anaesthetic, nitrous oxide (N20) and oxygen must be avoided
during the surgical operations on the thorax or the head, unless these agents can be evacuated by
aspiration or if the APPARATUS IS PROTECTED AGAINST THE ANAESTHETICS. It is advisable when
using flammable products (used for cleaning and disinfecting) and flammable solvents for
adhesive products the time to evaporate, before the surgical procedure. The fact that there is a
risk of accumulated flammable solutions under the PATIENT, in the creases or cavities of the body,
such as, the navel or the vagina, all accumulated fluids in this area must be cleaned away before
using the apparatus. The unavoidable electric arc which appears between the active electrode and
the tissue can in the presence of oxygen (contained in the air for example) cause the
inflammation of the endogenous gas. Therefore it is strongly recommended, especially when
operating on the abdomen, the neutralization of the atmosphere around the active electrode by
projecting inert gas (N2, C02,etc…) certain material like cotton wool or sterilized gauze, could
when it is saturated with an antiseptic, be set alight by the sparks produced by the apparatus
when used normally.
The electrosurgical generators must never be used in the presence of flammable liquid
or gas
The SURGILEC MC4 generator, on request, for the protection against the ignition of the gas
endogenous in digestive surgery, can be equipped with the system- projection of nitrogen (N2),
onto the active electrode.
c10 - For the PATIENTS fitted with a pacemaker or a stimulation electrode, there exists a
possible risk of interference to the function or damage to the pacemaker: the BIPOLAR mode is
advisable, if in doubt, refer to the cardiology department.
c11 - The electrosurgical unit’s are high frequency power generators. When the power supply
cables are connected to their terminal, they react like a short wave electromagnetic transmitter.
Other electrical equipment found within the operating theatre, monitors, camera, video etc… If
they have are not designed to function in these surroundings, can be affected by interference.
The ECG monitors can be fitted with special cables including a HF power filter with the inductance
category.
c12 - Although the specific circuit examines permanently, that the output power does not go
beyond 20% of the required setting and the loading value. It must not be forgotten that a serious
failure of the generator can increase the dangerous output power for the patient. In the case
where such an event is perceived, it is important before everything else is to cut the generators
power by operating the on-off switch and/ or disconnect the cable from the mains.

3.2 Additional information


The following warnings must be observed:
• Do not loop the electrosurgical cables while in use.
• The electrodes and instruments must not be place in contact with flammable material.
• The patient must not wear any metal objects, because of metal contact burns.

LAMIDEY NOURY Medical Page 9/22


SURGILEC MC2 User guide

• When using HF power when using the active electrode submerged in a conductive liquid
like saline solution, presents the risk of burning by the overheating of the liquid. A device
capable of evacuating of the liquid must be used, such as the arthropump for the arthroscopy.
The electrode used must have a limited contact area with the fluid with high resistant
temperature sheathing, such as a ceramic sheathing. With the bipolar mode, it is possible to
coagulate tissue which is soaked in saline solution.
• Check carefully the accessories, in particular the cables, before being used, notably
ensuring their good insulation (cf. § 2.3.1).
• So that the correct accessories are compatible with the generator, we recommend to
compare their insulation specifications and those required by the apparatus (see technical
specifications of the generator § 9.3. and standard compatibility requirements § 1.2.2).
• Stimulation of the muscles or nerves of the patient could be caused by low frequency
power sparks between the electrode and the tissue of the patient. If this should happen, stop
immediately the surgical operation and control all the connections of the generator. If no fault is
found, consult a technician approved by LAMIDEY NOURY MEDICAL.
• Any contact with components which bring on a heart condition are to be avoided, if an
electrode catheter current conveys the HF current to the heart this can produce damage to the
cardiac tissue, like atrioventricular bundle or sinusal node.
• Using electrosurgical current on bone tissue risks periosteum damage and jeopardise the
cicatrisation, resulting in necrosis of the tissue.
• The electrical security of the generator is only guaranteed if it is correctly connected to the
power supply and grounded, and meets with the security standards. It is compulsory to check
these essential requirements. If not sure, a thorough control of all the electrical elements should
be undertaken by a qualified person. The manufacturer of the electrical HF generator can not be
held responsible for damage caused by incorrect grounding necessary to the installation. It is
therefore forbidden to proceed in using the apparatus which is not efficiently grounded.
• Before connecting the apparatus, be sure that the voltage (shown on the back panel)
corresponds with that of the mains.
• If there is an incompatibility between the power socket and the power cable, only replace
these with accessories and connectors which meet standard requirements. It is advised against
using extension adapters, multiplugs and extension leads of any type, but if there is no other
choice, make sure that the adapters conform to the standard security requirements.
• Never touch the apparatus with wet or damp hands and feet.
• Do not use the apparatus with bare feet;
• Do not leave the apparatus exposed to excessive light, heat and moisture.
• Do not block the openings; ventilation grill.
• It is prohibited to use the generator around an explosive atmosphere.
• Disconnect the appliance after use.
• When the surgeon has finished with a particular function (monopolar cut or coagulation
bipolar for example) it is recommended to reduce the power level to O.
• The generator must be used exclusively for the purpose in which it has been designed. The
manufacturer cannot be held responsible for damage caused by incorrect use,
• It is dangerous to modify, or to try to modify the specifications of the appliance...
• Before starting any cleaning or upkeep on the appliance, disconnect by removing the
power plug from its socket or switching off at the mains.
• Switch off the generator in case of damage or malfunction. Contact an approved after sales
service technician requesting original spare parts. Any other solution risks jeopardizing the
security of the appliance and particularly the user.

LAMIDEY NOURY Medical Page 10/22


SURGILEC MC2 User guide

• Do not deactivate or reduce the audible activation indicator on the generator. The
activation alarm can minimize or avoid injury to the patient or the personnel in case of accidental
activation.
• Although the appliance has been conceived to function briefly on short circuit, avoid
checking the working order of the appliance by making a short circuit between the active
electrode and the neutral electrode plate or putting the active electrode in contact with metal
objects.
3.3 Advice on correct plate positioning.
The generator cannot be used without previously attaching the neutral electrode plate onto the
patient, to avoid all risks of patient plate contact burns, in monopolar mode, the following security
rules are essential:
3.3.1 Choice of the patient plate:
LAMIDEY-NOURY MEDICAL advice using exclusively the twin conductor split plate. The
disposable adhesive gel plates present the best security advantages against plate discharge and
the adhesive gel improves distinctly the quality of the conduction.
WARNING: USING SINGLE SURFACE PLATES, THE ALARM MONITOR CANNOT DETECT ANY
CONTACT FAULT ON THE PATIENT. ONLY THE USE OF THE TWIN CONDUCTOR SPLIT PLATE
ALLOWS THE GENERATOR TO WARN OF ANY EVENTUAL FAULT.
There exist 3 sizes, adult, child and baby. Whatever the size of the patient, the choice depends
on the surface required to attach the plate. If the size adult is too large for a child, try the child
size plate and if the child plate is too large for a baby, use 2 baby plates (with a double connector
UNIPLAC)
3.3.2 Positioning the plates.
To be sure of the best possible contact with the patient, certain safety measures must be
taken:
- The neutral patient plates contact must be perfectly fitted onto the body of the patient, to be
placed as near as possible to the operating site. The chosen area must be clean, dry, (removal of
all traces of cosmetics) and shaved. Do not use alcohol for cleaning the skin.
- The use of the adhesive neutral electrodes, pre-gelled are advised.
- Choose an area muscular or vascularised (thigh, buttock, abdomen, for example).
- Avoid the areas like body hair, scared tissue, bony areas, joints, metal prosthesis (ECG for
example), areas of the body where there is an accumulation of fluids. Avoiding the lumbar, the
sacrum, the ischium and the scapula areas.
The neutral patient plate’s surface contact must not be reduced by overlapping or by reducing in
size by cutting the plate.
Patient plates must not have any creases or bends. There must be no tissue between the
patient plate and the patient.
There must be no air bubbles trapped between the adhesive plate and the skin (press with the
hand along the length of the plate towards the connection clamp).
Patient plates must be placed as to avoid fluid between the plate and patient, amniotic fluid in
caesarean section, for example and to avoid liquid like iodine under the patient plate.
Do not add extra gel to the patient plate.
Do not remove and reposition the adhesive patient plate if the patient is moved during the
operation, check that the plates contact is correctly positioned.
Do not use the adhesive patient plate after the out of date stamp.
Respect the stocking temperature limit shown on the packaging.
Read with care the utilisation notice for the adhesive patient plates.

LAMIDEY NOURY Medical Page 11/22


SURGILEC MC2 User guide

In the event of the patient plate (SEN) alarm activation, discard and replace patient plate and
connector.
If the output power of the generator seems to be low, do not increase the electrosurgical
current before carrying out a control of the installation of the patient plate.
3.3.3 Advice on how to use
The connector UNIPLAC with the adhesive neutral patient plate
The connector UNIPLAC is designed to be connected to the disposable adhesive patient plates,
single or double surface.
In the event of an alteration or deformation of any kind to the connector UNIPLAC, or bare
cable, carry out a visual control before being used so if need be the article can be replaced or
repaired.
The accessory tester TESTELEC checks the cable power continuity when the UNIPLAC is
connected to the single adhesive plate. The double surface patient plate can be monitored only
when connected to the generator and applied on to the patient.
With the baby plate, use the double connector UNIPLAC which doubles the contact surface.
To connect the patient plate, lift the protected security hand lock, plug-in completely the
patient plates snap connection strip and secure.

Important: the incorrect size of the snap connection strip or incorrectly fitted, can jeopardize
the function of the patient plate alarm system. There is a risk of burning if part of the strip
protrudes from the connection. Using an adhesive bandage to hold together the connections will
avoid disconnection in the event of pulling on the cable. Keep clean the connection clip and plug.
It is advisable to attach the patient plate at least two minutes before starting up the
generator, to give time for the adhesive gel to moisturize the skin.
The UNIPLAC connector does not contain latex. They must not be cleaned by ultrasound or in a
decontamination bath. Wash with soapy water or mild detergent. Rinse well in clear water and
dry, eventually by compressed air. Use only decontamination products for wiping. Do not use the
connector if the connection strip or plug is still wet. UNIPLAC connector is not autoclavable.
SILIPLATES
The SILIPLATES are reusable plates with a conductor which does not contain latex. In the
case, when using on a limb, it must be fixed with an adhesive bandage. The SILIPLATE C (simple
surface) can be used on both sides. The SILIPLATE D (double surface) can be used only on the
side separated by 2 areas. Using the other side is dangerous and forbidden.
Before using, carry out a visual check of the SILIPLATE. In the event of an alteration like
deformation or bare cables, the part must be changed or repaired. Check carefully the condition of
the SILIPLATE area before using. In the event of wear and tear on the surface showing metal, or if
the surface shows any creases or dents, the plate must be immediately discarded and replaced.
The TESTELEC tester allows checking the continuity of the cable of the SILIPLATE C. This
function on the SILIPLATE D is only possible when the generator is connected to the plate after
being applied on to the patient.
The SILIPLATES must be always kept clean and dry. Check that there are no traces of
antiseptic before using. Only use decontaminated wipes to clean thoroughly before using. The
SILIPLATES are not autoclavable.

4 CONNECTIONS AND CONTROLS


4.1 Symbols and pictogram
The apparatus has different symbols at the front and the back which have precise well
recognised meanings, concerning the circuit specifications
Their meanings are explained in the chart below:

LAMIDEY NOURY Medical Page 12/22


SURGILEC MC2 User guide

Read the notice guide before starting.

Output circuits are floating. The parts to which they are connected can be in
direct contact with the heart of the patient. The vertical plates at each side
indicate that it is not necessary to disconnect the electrode patient plate when
applying a defibrillator.

The circuit specifications terminal plate is not grounded, the electrode patient
plate is not grounded due to the high frequency

The metal casing of the apparatus can be connected by the suitable cable with an
equalizing potential rod. (This symbol and connecting terminal is found at the back
of the apparatus.)

Other pictogram showing different functions of the electrosurgical unit SURGILEC:

Patient plate simple conductor Patient double plate conductor


surface

Coagulation soft Coagulation by desiccation

Coagulation by fulguration Coagulation by fulguration


level 1 level 2

Pure cut Haemostatic cut modulation


level 1st

Haemostatic cut modulation ENDO SECTION cut


level 2nd

Foot pedal control Manual control

Coagulation and cut selection


mod

LAMIDEY NOURY Medical Page 13/22


SURGILEC MC2 User guide

4.2 Back panel

1 Switch on/off
2 Mains power input socket
3 Fuses
4 Equipotential socket
5 Description plate
6 Double foot pedal socket

NOTE: The serial number of the generator is shown on the description plate

4.3 Front panel

1 Power setting +/- 7 Monopolar connection socket


2 Bipolar mod selection 8 Bipolar connection socket
3 Monopolar mod selection 9 Patient plate connection socket
4 Haemostatic cut selection and 10 Audio setting +/-
ENDO-SECTION.
11 Visual HF power indicator
5 “Impuls cut” selection
6 Monopolar connection socket
foot pedal control

LAMIDEY NOURY Medical Page 14/22


SURGILEC MC2 User guide

5 INSTRUCTIONS
5.1 Starting, alarm patient plate
The electrical units are conceived to function with neutral electrode
conductors, the single patient plate (as shown) and the double patient
plate, split into two sections. Whichever plate is chosen the touch button or
N°12 can be found on the front panel.

Connect the patient plate to the connection socket N°11, shown on the front panel, pushing in
completely the jack socket finishing with the conductor SILIPLATE or UNIPLAC.
Output circuit are floating in low/high frequency (electrode patient plate not
ground-referenced): All electrical connections between the patient plate and the
operating table are forbidden.

Start-up the generator by pressing the switch on/off, found on the back panel. As soon as the
unit is powered, the units alarm plate will sound in the following cases:
When the single patient plate is connected:
 If there is a fault in the electrical connection between the patient plate and the unit.
 If the double patient plate function is selected.
When the double patient plate is connected:
 If there is a fault in the electrical connection between the double patient plate and the
patient (incorrect contact)
 If the impedance of contact between the patient skin and plate exceeds 60 Ohms, when
the single patient plate is selected.
When the unit alarm sounds, the power is stopped, therefore deactivated.
If there is a fault with the patient plate, only when selecting bipolar mode will the alarm stop.

5.2 Monopolar cut mode


To obtain the monopolar cut operate the double foot pedal (see §6.2) or the hand control
tactile (see §6.3) the power connections are the red power socket N°6 or the three pin socket N°7
found on the front panel.
When activated, the audible alarm and visual yellow light, (on the right (left) of the HF power
indicator) will signal that the cut operation functions.
An array of different cutting effects is available, press the touch selection button N°4: pure
cut, haemostatic cut and Endo-section.
The PURE CUT allows obtaining a cutting effect with minimum haemostasis. If the
used electrode has a sufficiently fine point, this will sever with a low power and the
sides of the incision will be very slightly coagulated. The principle advantage with this
type of cutting power is that there is very little surrounding tissue destruction.
The HAEMOSTATIC CUT releases a modulated power that increases very clearly the
haemostasis effect on the sides of the incision. The higher level of modulation promotes
the priming of the cut. The haemostasis of the incision depends, not only on the power
modulation but also by the size of the active electrode and the speed rate of the cutting : a knife
electrode gives a better haemostasis effect than with a point electrode, the slower the
cutting action, the deeper the haemostasis. The haemostatic cut allows producing a
better result from the self regulation cut for the incision of vascularized tissue.

LAMIDEY NOURY Medical Page 15/22


SURGILEC MC2 User guide

ENDO-SECTION is a high modulated cut designed for a great haemostatic effect. Therefore
when the electrode is penetrating through the tissues, a coagulating phase precedes the cut.
ENDO SECTION is the best way for polypectomy in gastro-enterology.
UNDER FLUID CUTTING (TUR or myomectomy) requires higher power because of the calorific
diffusion in the fluid. In order to achieve priming capable to cut under fluid, it is imperative to
choose the pure cut. Check with the power indication table (appendix).

5.3 Monopolar Coagulation mode


The activation of the monopolar coagulation is achieved by the foot pedal double (§ 6.2) or by
the hand control tactile (§ 6.4).The monopolar coagulation power is released either by the red
socket N°6, or by the three pin socket N°7 found on the front panel.
When activated, the audible alarm and the blue light, shows that the coagulation control
functions.
5.4 Bipolar coagulation mode
The activation of the bipolar coagulation is achieved by the double foot pedal (§ 6.2) or by
MANILEC forceps (§ 6.4). The electrical power bipolar coagulation is released by the socket N°8
found on the front panel.
When activated, the audible alarm and the blue light, shows that the coagulation control
functions.

6 SETTINGS AND POWER ACTIVATION


6.1 H.F. power settings
The H.F. power output is adjustable by the touch button N°1 to
increase or lower the power, as shown for each function. The indication
table (see appendix) at the end of the instruction manual suggests which
selected function and setting is required, regarding the surgical
application.
The values are shown as an indication. It is advised to always look for the minimum power
level.
In the case of power failure which lasts less than 20 minutes, the generator keeps the power
settings and operational selection memorized. If the power failure lasts more than 20 minutes, it
is advisable to verify the parameters before continuing the operation.
During the verification of the different activation controls (§ 6.2 - 6.6), as a precaution, reduce
all the HF powers to 0.
6.2 Double foot pedal activation
The double foot pedal connection is found on the back panel, as indicated (socket N°6).used
for monopolar and bipolar mode. This function commands the cutting mode with the yellow lever
and the coagulation mode with the blue lever.
To change from monopolar mode to bipolar mode or vice versa, press the touch button N°2 or
3 on the front panel.
With the monopolar mode, With the bipolar mode, the
the power is released by the red power is released by the socket
socket N°6 into which the N°8, into which the bi-wire cable
monowire cable ref. 11FM40.is ref. 11F242 is fitted with a 2 pin
connected. plug, type BCP.

LAMIDEY NOURY Medical Page 16/22


SURGILEC MC2 User guide

Cautions:
 Do not pull the foot pedal by its cable.
 The foot pedal can be cleaned and rinsed with running water; without immersion.
 Only use a decontamination surface cleaner.
 Check the foot pedal before using: if there is an adhesive pad missing under the foot
pedal, the controls risk to not function. To check the control system, the tester TESTELEC is
recommended or via the generator (see § 2.3.2)

6.3 Hand held tactile pencil activation


Hand held electrodes are connected on to the front panel by the
connection three pin socket No 7.
For the monopolar mode (touch button No3) the blue switch button on the pencil for
coagulation monopolar and the yellow switch button for cut monopolar.
For the bipolar mode, the blue switch button will activate only coagulation (monopolar) but the
cut switch button will be ineffective.
6.4 MANILEC bipolar forceps activation
The forceps MANILEC are connected on to the front panel, connection plug No 8,
with the cable tri-wire fitted with a three pin plug (BCM) reference No. 11F341.
This bipolar coagulation is obtained by the closing of the forceps, without having
to select the bipolar mode button on the front panel.
The forceps MANILEC are fitted with an automatic releasing contact, with
a square rotation to control the activation. The 4 positions turn, which
alternate between 2 functions: the function for manual bipolar coagulation
activation and the other which restricts this manual activation, this enables the surgeon to dissect
without coagulation.
Using advice: the interior of the connection contact must be kept clean. If the manual activation
does not function, the interior can be cleaned with a mild abrasive.
6.5 Audible indicator setting
The audible sound level is adjustable by the touch button +/ - shown on the front panel
by the pictogram symbolizing a loud speaker. The sound cannot be deactivated: the
minimum sound level continually indicates the activation power levels for cut and
coagulation with monopolar and bipolar mode.

6.6 Does and Don’ts


In Bipolar Mode, monopolar cut is ineffective. Monopolar coagulation is available with a hand
activated pencil. If monopolar mode is not requested, connect a neutral plate will be needless.
In Monopolar Mode, the manual operation bipolar with the forceps MANILEC remains
accessible and takes precedence. - If one of the required functions, cut or coagulation is selected,
the other one is restricted. – The generator can only function if the alarm plate is not primed.
6.7 Inert gas activation
By demand, the generator can be fitted with the inert gas system which is delivered with the
production of the HF current in monopolar mode:
To operate with inert blown gas, use the pedal control electrode DERLEC 25 Nitrogen with
cable, connect to the socket No 10. Connect the plug on the back panel to a source of inert gas
(nitrogen, carbon dioxide etc.), with a maximum pressure of 50 KPa (#0,5 bar).A gas cylinder
fitted with a pressure reducer adjustable from 0, or a distributor socket with a pressure reducer.
The gas flow is delivered by the activation of the generator.

LAMIDEY NOURY Medical Page 17/22


SURGILEC MC2 User guide

7 TROUBLE SHOOTING
7.1 Response from the controls
7.1.1 – When the indicator [HF/ENERG] N°11 is lit up but the electrode stays inactive
when in contact with the patient: - Check that the monopolar/bipolar selection is correct. If the
selection is correct, check the working order of the accessories (connections, cables, electrodes
etc…), if need be, change them.
7.1.2 – When the indicator [HF/ENERG] N° 11 stays inactive, regardless of digital
display settings: The defect appears to originate from the generator, contact the manufacturer or
your authorized representative.
7.2 No Response from the Apparatus
7.2.1 –When the ON indicator does not light up:
Check the sockets, power cables and fuses. If a fuse needs changing, contact the manufacturer or
your authorized representative.
7.2.2 –When the ON indicator lights up: Check the controls (foot pedal, electrodes
pencils or forceps, socket and cables). If need be use the tester TESTELEC.
7.2.3 – If no anomaly is detected, do not attempt to remove the back panel, contact
the manufacturer who will assure the guarantee and is solely responsible for the service and
adjustments.
7.3 Generator Alarm Sounds
7.3.1 – Alarm Plate
If the alarm starts, with the red flashing light (S.E.N): check that the touch selection button
for the patient plate is correct. If the alarm continues, select the bipolar mode. If the alarm stops
this means that there is a fault, check the cables and connections.

7.3.2 - Excess HF Alarm


If the continual alarm and the [EXCES HF] indicator lights up: when the output power level
exceeds the limit programmed on the display panel. Turn the power off for at least 10 seconds
and then turn back on.
If the fault immediately reappears or when power is drawn, the fault must be checked by the
manufacturer or an agreed representative.

8 MAINTENANCE
8.1 Generator
The generator needs no special maintenance apart from regular cleaning of the outer casing
and front panel with a soft clean cloth soaked in alcohol. For decontamination use a sterilized
cleaner.

8.2 Accessories and Instruments


8.2.1 - Foot pedals
The foot pedals are cleaned like the generator, check the power cables for damage or use the
tester TESTELEC. Check the condition of the anti-slip pads under the foot pedals: wear and tear
and missing pads will make the foot pedals unstable and prevent possible activation.
The foot pedals can be decontaminated with a sterilized cleaner, as they are not autoclavable.

LAMIDEY NOURY Medical Page 18/22


SURGILEC MC2 User guide

8.2.2 - Instruments in contact with the patient


The washing of the instruments, the reusable patient plates and their cables are cleaned with
soapy water or a mild detergent cleaner. Do not clean electrosurgical accessories by ultra-wave.
These instruments can be immersed in a decontamination product, except for the UNIPLAC
plates which are treated with a sterilized coating. Respect the manufacturer’s recommendations in
the length of time and temperature of the solution. Rinse thoroughly with filtered running water
and dry completely, if need be use compressed air.
The instruments and cables in contact with the patient are autoclavable at 134°C during
20minutes, except the patient plates and UNIPLAC connections.
A regular accessory check is recommended (see § 2.3.1).

8.3 Service
The generator requires no particular sanction before being commissioned. The casing must not
have been tampered with, unless authorized: as this will nullify the guarantee. To correct any
problems or servicing, contact or send the generator to the technical services department of
LAMIDEY NOURY MEDICAL, with full details of the problem encountered.
The maintenance consists mainly in the cleaning and sterilization of the accessories and the
checking of the good working order of the generator before being used. An annual security control
to check the settings must be done by a specialized technician.
LAMIDEY NOURY MEDICAL recommends an annual preventive maintenance intervention to
control the electrical specifications.
JOURNAL OFFICIEL de la République Française N°66 of 19/03/2003: signed order 3rd March
2003 from the French Department of Health making it an obligation to servicing the medical
devises class IIb.
Changing Fuses
If a power fuse has blown and needs changing, it must be replaced with the same wattage.
Therefore in this situation, we recommend that the generator should be checked before utilization,
as there may be a possibility that the generators power circuits might be at fault.

9 TECHNICAL SPECIFICATIONS
9.1 Mechanical and Environmental
Mobile Unit
- Fixed installation
- Installation on an arm
- Outer protection level: IP 203
- Dimension and weight:
Weight Width Height Depth
9 Kg 345 mm 190 mm 370 mm
- Fixed installation restrictions with the cooling system:
If the unit must be placed in an enclosed area the volume must be no less than 1m3.

LAMIDEY NOURY Medical Page 19/22


SURGILEC MC2 User guide

9.2Input Specifications

Supply : 230 V 50 Hz

Electric shock protection: Class 1 Type CF

Maximum power absorbed : 500 VA

Functioning Mode: Uninterrupted operations with intermittent load


( loaded :1mn ; stopped :3mn)

Maximum current protection : External fuses 2x TFS 5x20 3A 250 V

Air supply (Nitrogen) optional extra : 0,5 Bar


cylinder and pressure reducer 3 to 5 dm3/mn

9.3 Output Specifications


Maximum power Range Monopolar : 200 W / 600 Ω
Bipolar : 70 W /35 Ω
Maximum Pure cut 1200 Vpp
peak to peak MONOPOLAR: Endo-section 2200 Vpp
voltage
Coagulation 2400 Vpp
(Vpp)
BIPOLAR : Coagulation 600 Vpp

(1)
frequency modulation Crest factor
Monopolar pure cut 620 kHz 0 1.4

Endo-section 620 kHz 21 kHz

Monopolar Coagulation 620 kHz 21 kHz

Bipolar Coagulation 620 kHz 21 kHz

(1)
Crest factor: ratio between peak voltage and rms voltage (output unloaded)

10 ANNEXES
- Guidelines to Power Settings.

LAMIDEY NOURY Medical Page 20/22


SURGILEC MC2 User guide

LAMIDEY NOURY Medical Page 21/22


SURGILEC MC2 User guide

- DIAGRAM SHOWING THE CONNECTIONS FOR THE ELECTROSURGICAL INSTRUMENTS:

LAMIDEY NOURY Medical Page 22/22

You might also like