3. METROLOGICAL VERIFICATION OF BREATH
ALCOHOL DETECTOR
-0,1
Error
2628
Table 2: Table of EMA according to OIML R126 The error is the difference between the average indication of
equipment and reference value.
EMA Analysing the chart we can see that the breath analyzer
BAC–Blood alcohol Type approval/ Periodic meets the established by Decree 1556/2007 of 10 December.
rate (g/l) initial verification verification/ The analysis of this equipment in particular has shown also
Extraordinary that meet the metrological and technical requirements
verification
defined by Recommendation OIML R 126. So the
BAC< 0,920 ±0,046 g/l ±0,074 g/l
0,920≤BAC≤4,600 ±5 % ±8 % instrument is sealed and a verification certificate is issued. If
BAC > 4,600 ±20 % ±30 % not meet the requirements is issued a rejection report and the
instrument cannot be used in the following year, until be
repaired.
4. CONCLUSIONS
Table 3: Results of a breath alcohol detector verification (g/l).
In recent decades has been remarkable increase in the use of
breath analyzers either by private entities or by entities with
Reference Reference Average Error Standard legal competence. This reflects an increased surveillance in
value uncertainly indication deviation the blood alcohol content of individuals, which carries a
0,335 0,0046 0,336 0,00 0,01 greater security for society, especially in the context of
0,539 0,0083 0,548 0,01 0,01
0,883 0,0124 0,91 0,03 0,00
driving under the influence of alcohol.
1,310 0,0166 1,35 0,04 0,00 The improvement of methods for quantification of alcohol
2,052 0,0249 2,13 0,08 0,02 concentration in blood and the easy way they can be
implemented has contributed to a greater use of these
resources and a more frequent monitoring.
As the tests are carried out with gas mixtures prepared and
certified by IPQ laboratory, the national traceability is thus
0,5
ensured.
REFERENCES
0,25
[1] Vocabulário Internacional de Metrologia, VIM, IPQ 2008.
[2] “Guia para a expressão da incerteza de medição nos
laboratórios de calibração”, IPQ 2005.
Error
0
0 1 2 3 4 5 6 7
[3] Furtado, C.M. Pires, M.C. Ferreira, O. Pellegrino ”Controlo
Metrológico dos alcoolímetros, Cinemómetros no IPQ”,
Actas da 2ª Conferência Nacional da Sociedade Portuguesa
-0,25 de Metrologia, 2007, Funchal.
[4] M.C. Ferreira, A. Cruz “Controlo Metrológico no Instituto
Português da Qualidade”, Actas do 2º Encontro Nacional da
-0,5
Sociedade Portuguesa de Metrologia, 2006, Lisboa
Calibration points [5] OIML R 126, “Evidential breath analyzers”, 1998,
2629