You are on page 1of 9

User and

Service Manual

Tube Stirrer

HUMAMIX

!"#$%&'(%)*%%%!"#$%+,-"(%)./.))0

Best.-Nr. / Cat.-No.: 1717001


1234&%5671
+8,9:';-8<,%='>?@A=8@"

BC!DEFEB4DC%GF%BG&FG!3EDH

WE

HUMAN
Gesellschaft für Biochemica und Diagnostica mbH
Max-Planck-Ring 21
D - 65205 Wiesbaden
Germany

declare that Human GmbH has a certified quality system in place according to the harmonised standards DIN
EN ISO 9001 and DIN EN 46001 .

We further declare under our sole responsibility that the product

12343EI%JG!DCI%3EIC!
Cat.-No. 17170/1 & 17170/2

conforms with the standards:

EN 980
EN 50081-1
EN 61010-1
EN 61010-2-020

following the provisions of 73/23/EEC, 89/336/EEC, 92/31/EEC and 93/68/EEC directives.

Wiesbaden, Germany – February 19, 2003 Hans B. Kern


Place and date of issue Marketing Diagnostic Instruments

User/Service Manual HUMAMIX 1/7


K4!!4&DH%BC!DEFEB4DC

1. HUMAN GmbH warrants that the equipment delivered is free from defects in material and
workmanship. This warranty is provided for a period of one year. The warranty period
starts from the delivery date.

2. Warranty does not apply to parts normally consumed in operation, current maintenance
and the adjustments described in the operating instructions provided with the equipment.

3. HUMAN GmbH does not accept any liability in the case where the goods are not used in
accordance with their proper intent.

4. The warranty may not be claimed for damages occurred during the shipment, for damages
resulting from improper handling or use, the defects in maintenance, negligence, bad
functioning of auxiliary equipment, in the case of force majeure or accident and incorrect
power supply.

5. In the event of failure, HUMAN GmbH shall be under no liability for any injury, or any loss
or damage as the result of the failure other than the guarantee conditions.

LCFG!C%GMC!4DE&5%D1C%E&ND!23C&D%D1EN%34&24O%N1G2O+%LC%!C4+
B4!CF2OOH$

E&FG!34DEG&%BG&D4E&C+%E&%D1EN%+GB23C&D%EN%D1C%M!GMC!DH%GF%1234&
5671$%ED%34H%&GD%LC%+2MOEB4DC+%G!%+END!EL2DC+%KED1G2D%1EN%MC!3ENNEG&$

D1C%J4OE+EDH%GF%D1C%524!4&DCC%EN%N2LPCBD%DG%D1C%GLNC!J4DEG&%GF%D1C
E&ND!2BDEG&N%4&+%M!CB42DEG&N%+CNB!ELC+%E&%D1EN%34&24O$

User/Service Manual HUMAMIX 2/7


BG&DC&DN

1 USE AND FUNCTION 4


2 TECHNICAL SPECIFICATIONS 4
2.2 General View 4
3 INSTALLATION PROCEDURE 5
3.1 Lifting And Transport 5
3.2 Unpacking 5
3.3 Environmental Conditions 5
3.4 Mains Supply 5
3.5 Positioning 5
4 OPERATING THE HUMIX 6
4.1 Switching On 6
4.2 Completion of the Operation 6
5 PERIODICAL MAINTENANCE AND CLEANING 6
5.1 Periodical Maintenance 6
5.2 Cleaning 6
6 TROUBLESHOOTING 6
7 ELECTRICAL CIRCUIT DIAGRAM 7
8 SPARE PART LIST 7

User/Service Manual HUMAMIX 3/7


*%2NC%4&+%F2&BDEG&
The HUMIX tube stirrer is used to stirr the liquids in the tubes homogeneously.

The cast iron body and soft rubber cup of the HUMIX secure a stable and vibration free structure.

.%DCB1&EB4O%NMCBEFEB4DEG&N
.$*%D"<Q:8<,?%NR"<8S8<,-8':;%D,7?"

12343EI
Stirring system Vibration
Speed control Single step
Rotation speed (± %5) 2400 rpm
Stirring stroke length 3 mm
Rubber cup dimensions Ø 20 mm
Stirring capacity 50 gr.
Noise and vibration isolation Soft rubber footing
Power supply 230 V, 50 Hz
Power consumption 40 W
External material Epoxy polyester painted cast iron
External dimension ( W x D x H) mm 150 x 200 x 170
Packing dimensions ( W x D x H) mm 350 x 230 x 230
Net weight 6 kg
Gross weight 7 kg

.$.%5":">,?%J8"=

1. On / off switch
2. Stirring rubber cup

User/Service Manual HUMAMIX 4/7


0%E&ND4OO4DEG&%M!GBC+2!C
0$*%O8S-8:9%4:@%D>,:;R'>-
All lifting and transport must be carried out using proper handling equipment. The HUMAMIX must be supported
from underneath and never be turned over.

0$.%2:R,<T8:9
Open the packing box and remove the nylon packing of the HUMAMIX.
Carry the HUMAMIX by holding from the bottom to the place of installation.

Provided with the HUMAMIX are these items, check them.

! Users Manual
! Electrical cable

0$0%C:#8>':6":-,?%B':@8-8':;
Please pay special attention to the followings,

! Indoor use only


! Room temperature from 5°C to 40 ° C
! Humidity level 80% up at 22° C

0$U%3,8:;%NVRR?W
The HUMAMIX requires 230 V ± 10%, 50 Hz, respectively 110 V ± 10%, 50/60 Hz.

Connect the electrical cable to the HUMAMIX and then plug the cable in a 9>'V:@"@%;'<T"-%.

Please make sure that the supplied mains matches the required power ratings. If not, provide an extra line to
support the supply.

4?=,W;%<'::"<-%-Q"%%-V7"%;-8>>">%-'%R>'R">?W%9>'V:@"@%;'<T"-;$

4% ;VRR?W% S8--"@% =8-Q% ,% <8><V8-% 7>",T">% ;Q'V?@% 7"% V;"@% S'>% R>'-"<-8':% ,9,8:;-% ;Q'>-% <8><V8-% 8:% <,;"% 'S% ,
8;'?,-8':%S,V?-$

0$X%M';8-8':8:9
! Lift the HUMAMIX and carry it to its place carefully.
! Check that no damage occurred during transportation.
! Balance it on the four pedestals. If necessary, provide support heights.

Check the followings,

! the proposed site is suitable for users,


! the operator can follow up the tube stirrer even he deals with something else.
! the HUMAMIX does not occupy the utilisation space of other equipment or damage them.

User/Service Manual HUMAMIX 5/7


U%GMC!4DE&5%D1C%12343EI
U$*%N=8-<Q8:9%G:
! Place the tube tightly into the vibrating rubber cup.
! Switch on the tube stirrer.
! See that the led light turns on and the stirring starts.

U$.%B'6R?"-8':%'S%-Q"%GR">,-8':
! Switch off the HUMAMIX and take the tube.
! Remove any undesired effects (spills) of the samples after you have taken out the tube.

X%MC!EG+EB4O%34E&DC&4&BC%4&+%BOC4&E&5
X$*%M">8'@8<,?%3,8:-":,:<"
! There is no requirement for periodical maintenance.

X$.%B?",:8:9
! Do the cleaning after you have disconnected the electrical cable.
! You may wipe the surface against dust and dirt.
! You may use mild liquid soap against difficult dust and dirt.
! Please be aware of the undesirable effects of the cleaning chemicals and be careful while applying them.
! &"#">%V;"%7?",<Q%'>%,<8@8<%;'?#":-;%S'>%<?",:8:9$

Y%D!G2LOCN1GGDE&5
If the tube stirrer does not operate, CHECK WHETHER

! the on/off switch is open ( the stirrer is switched on),


! it is connected well to the power supply,
! the fuse is sound,
! the plug is not defective,
! the installation of the mains supply is not defective,
! there is mains supply.

! the rubber cup is mounted properly,


! the wire connection of the motor is made properly,
! the motor is not defective,
! the stirring system is not defective.

BG&D4BD%DG%HG2!%42D1G!EZC+%1234&%+END!EL2DG!%EF%4&H%C!!G!%GBB2!N$

User/Service Manual HUMAMIX 6/7


[%COCBD!EB4O%BE!B2ED%+E45!43

\%NM4!C%M4!D%OEND
&G +CNB!EMDEG& B,-$A&'$ ]24&DEDH
1 On / off switch 17170/100 1 pc.
2 Rubber cup 17170/101 1 pc.
3 Complete vibration head 17170/102 1 pc.
4 Motor 17170/103 1 pc.
5 Spring in complete vibration head 17170/104 1 pc.

User/Service Manual HUMAMIX 7/7


Human
Gesellschaft für Biochemica
und Diagnostica mbH
Max-Planck-Ring 21 ! D-65205 Wiesbaden
Germany
Telefon: +49 6122 9988 0
Telefax: +49 6122 9988 100

eMail: human@human.de
Internet: http://www.human.de

Ch.-B. / Lot: 030527