Professional Documents
Culture Documents
Stokes1988 PDF
Stokes1988 PDF
Stokes1988 PDF
or Not?
INTRODUCTION
228
lytically stable at neutral and basic pH. Thus, they have been exten-
sively evaluated for use in chronically implantable devices such as arti-
ficial hearts, heart valves, shunts, indwelling catheters and many
more-including cardiac and neurologic pacing leads.
The stability of several polyether polyurethanes has been established
in long-term animal tests [2,3]. The human experience in pacing with
Pellethanes 2363-55D and 80A exceeds 12 years with implants begin-
ning in 1975 (Neuro) and 1977 (cardiovascular). In general, the human
experience has been good. In spite of this, controversy developed over
the biostability of these polymers in the mid 1980s when it was dis-
covered that the insulation failure rates on certain lead models was
higher than others. Two mechanisms have been discovered that can af-
fect implanted polyether polyurethanes, stress cracking and autooxida-
tion [4,5]. Thus, we find an apparent paradox where some materials
have been shown to be biostable, yet may be affected by degradation
mechanisms. Are polyether polyurethanes biostable or are they not?
We do know that the answer to this question is not a simple yes or no.
Certainly no material is entirely stable in an absolute sense. Thus, the
existence of degradation mechanisms in implanted devices must be ac-
ceptable if the rate of change is slow enough to assure that clinical fail-
ure does not occur within the expected lifetime of the device. This must,
in fact, be the definition of biostability.
Because our experience has dealt primarily with cardiac pacing, our
testing naturally concentrates on this technology. Nonetheless, it is
possible that our experience will apply to other fields of biomedical
device technology
11IATERIALS
turing human use devices, other companies have developed their own
versions. Surethane (Cardiac Control Systems) is believed to be equiva-
lent to Biomer and is used to insulate one company’s cardiac pacing
leads [8]. An extrusion grade of Biomer using water as the chain ex-
tender (thereby reducing the urea linkages by a factor of 2) has also
been evaluated in our studies.
Cardiothane-51 (Kontron, formerly Avcothane-51 from Avco) is the
product of MDI and PTMO with a butane diol (BD) chain extender and
acetoxy terminated polydimethyl siloxane. Theoretically, this is a mois-
ture cured (cross-linked) coating material.
Tecoflex (Thermedics) is formed through the reaction of dicyclohexyl
methane diisocyanate and PTMO using BD as a chain extender. Sev-
eral hardness grades, including Shore 80A (Tecoflex EG 80A) and Shore
60D (T~coflex EG 60D) are available. The differences in hardness are
achieved by using a higher molecular weight PTMO in Tecoflex EG
80A compared to %coflex EG 60D. A series of polymers apparently sim-
ilar in composition to Ibcoflex was offered by K. J. Quinn (Q-thane).
The Pellethane resins, 2363-80A and 55D have been the most exten-
sively tested polymers in our studies since they are used on all but a
few models of cardiac pacing leads. These polymers are the reaction
products of MDI, PTMO and BD, with a slight excess of isocyanate. The
isocyanate end groups can react with urethane groups to form ther-
mally labile allophonate cross-links (allowing thermoplastic extrusion
and injection molding). Several Shore A hardnesses, 80A, 90A and 55D,
are believed to be achieved by varying the ratios of PTMO to MDI and
BD. Pellethane 2363-80AE is similar to Pellethane 2363-80A, but with
hydroxyl end groups.
Several additional polymers will be briefly mentioned. These include
peroxide cured (Vesta) and enhanced tear resistant (Dow-Corning) sili-
cone rubbers, Rimplast PTSE BS-1204 (Petrarch, silicone-styrene/eth-
* 16 to 18 months implant.
233
months, then stabilized as about 1.6 to 1.8% weight was gained. When
specimens were vacuum dried, their weight returned to original values
(see Table 1). The mechanical properties of dried specimens increased
upon drying, but not quite to the original values. This was not entirely
unexpected since a slight nonreversible change could occur as a result
of extraction or absorption. At high magnification, evidence of &dquo;cold
flows&dquo; (a molding defect) was found at one location on the specimen
which may also have resulted in some slight nonreversibility (Figure 1)
[10]. Other surfaces of the specimens had a wrinkled appearance, due to
mold shrinkage (Figure 2). There was no microscopic evidence of crack-
ing or any other untoward results. Density remained virtually un-
changed at 1.120 (80A) and 1.162 (55D). DSC showed only that mois-
ture was absorbed. There was no evidence of morphologic or chemical
change at the end of the 2 year study. There was no evidence of degra-
dation and no trends showing continuously declining properties. Thus,
the aged polymer properties remained acceptable for the intended pur-
pose (pacing lead insulation).
Subsequently, other polymers have been similarly tested. These in-
clude Tecoflex EG 80A and EG 60D, Pellethane 2363-80AE, Q-thane
PE97 and Cardiothane-51 coatings on Pellethane 2363-80A. In these
cases, the tensile specimens were injection molded on more modern
screw machines, using conditions recommended by the polymer manu-
facturers. This produced more uniform specimens (without cold flows).
Pellethane 2363-80AE gained 2.2% weight after 24 months implant,
which was completely reversed by vacuum drying (Table 1). Density in-
creased from 1.116 t .003 to 1.127 di .002. There were no changes in
the (bulk) infrared spectrum. Inherent viscosity decreased about 15%.
tensile strength decreased 32% while elongation did not change appre-
ciably. Vacuum drying completely reversed the change in tensile
strength. At high magnifications, however, shallow cracks were found
in the surfaces (Figure 3). Thus, the bulk polymer appears to be reason-
ably biostable, but some question exists about the material-tissue in-
terface.
Tecoflex EG 80A gained 2.6% weight after 16 months implant (table
1). This was not completely reversed by vacuum drying, suggesting that
some non-volatile material had been absorbed. The tensile strength of
the wet specimen increased 22% after implant. Drying the specimen
caused an even greater increase of 69% total. Elongation increased
only 13% (dry). No changes in the (bulk) infrared spectrum were ob-
served. The inherent viscosity of bulk polymer increased slightly
(16%), indicating that an increase in molecular weight occurred. By the
time the Thcoflex resins were evaluated, questions had been posed
Figure 3. Cracks in Pellethane 2363-80AE tensile specimen surface after 24 months im-
plant in a rat. 500 x.
about the biostability of surfaces. Thus, slices were taken from the
specimens’ surfaces for molecular weight determination (Table 2). Both
the weight and number average molecular weights of the surface mate-
rial decreased by 40% after 18 months implant. Shallow cracks in the
surfaces were seen on the scanning electron microscope. Thus, this
polymer did exhibit evidence of degradation both on surfaces and in its
bulk.
Tecoflex EG 60D gained 2% weight in vivo which was reversible (Ta-
ble 1). No changes in (bulk) infrared spectra were observed. The in-
herent viscosity of bulk polymer increased 50% in 18 months, then de-
creased to within 10% of the original value within the next 8 months.
,
STRESS CRACKING
237
Figure 6. A stress crack in explanted Pellethane 2363-80A which still retains substan-
tial craze structure. 700 x.
239
Figure 9. A cross section showing very shallow cracks in the tissue contacting surface of
an explanted pacing lead near a tight ligature. Note the very uniform depth of crack-
ing. 14 x.
Figure 10. An explanted specimen similar to that shown in Figures 5, 8, and 9 has been
mounted under fixed strain. This strain pulled the cracks apart revealing what looks like
an undamaged substrate (no craze tips are visible at up to 5000 x). No crack propagation
occurred. Thus, these are stable, nonpropagating cracks. 510 x.
frared spectra of frosted material does not show any evidence of chemi-
cal degradation in our studies [12].
Initial Hypotheses .
Tight ligatures were placed at each end of the mandrels to fix the
strain. Up to 4 such specimens were tied end to end in strings of up to
4 pieces each. Then up to 4 such ethylene oxide sterilized strings of
specimens were implanted subcutaneously in rabbits.
In early studies, specimens included strains ranging from 0 to 500%
elongation in increments of 100%. Explants were made weekly for 12
weeks. All specimens were examined by optical and scanning electron
microscopy. Pellethane 2363-80A tubing extruded into a very cold tank
developed cracks at 100% elongation in 1 week. Simply switching to a
hot take-off tank (slowing the heat transfer rate) increased the induc-
tion period to 4 to 5 weeks and the critical strain to 200-300% elonga-
tion [12]. Some variation in these findings occurred in later tests as a
function of raw material lot, but process sensitivity remained a consis-
tent finding. Annealing the specimen to produce 100% stress relaxa-
tion completely eliminated cracking in this test [4,12]..
A factorial experiment was done to evaluate and verify the effects
and interactions of various processing parameters on the stress crack
resistance of Pellethane 2363-80A. Since the existence of induction pe-
riods, critical strains, etc. had been demonstrated in previous experi-
ments, the new test could be simplified. We implanted all specimens for
12 weeks only, on a go/no go basis. If cracking occurred, the timing was
not important since the ultimate goal was no cracks. The variables
tested were hot and cold heat transfer in the extrusion take-off tank,
baking or not baking tubing in air prior to strain, no strain and 400%
elongation, annealing to 100% stress relaxation or not annealing speci-
mens after strain and the effects of extraction of tubing prior to speci-
men assembly. Each variable combination was represented once in each
of 36 rabbits for a total of 576 specimens. The optical microscopic re-
sults were analyzed by BMDP, SAS and StatGraphics Statistical pro-
grams.
There were several statistically significant effects and interactions in
the 32 processing combinations evaluated. The outstanding primary ef-
fect (cause of ESC) was clearly shown to be excessive stress (strain) in
the absence of annealing. The single most important factor preventing
ESC was annealing to 100% stress relaxation. Slow heat transfer on ex-
trusion also improved stress crack resistance, but to a lesser degree
than annealing. Thus, this study confirmed with high statistical con-
fidence that the mechanism is a form of stress cracking, that it criti-
cally depends on applied and morphologic strain and that it can be pre-
vented by .annealing processes. A very interesting observation in this
study was the fact that most process combinations exhibited significant
animal to animal variation, e.g., &dquo;biologic variance&dquo;’ In fact, cracking
only occurred in 40% of the rabbits. Thus, unless the processing was
just right (or just wrong), the animals’ chemistry did or did not interact
with residual stress to cause cracking. We can, therefore, redefine the
mechanism of in vivo stress cracking as the formation of crazes or
cracks through the interaction of residual stress with the biologic
(mammalian) environment of certain individuals. -
In fact, stress cracking has not been duplicated in vitro in spite of ex-
tensive attempts. A partial list of the media we have tried so far is pre-
sented in Table 3. It is particularly interesting to note that in vitro in
blood, plasma, or serum, stress cracking does not occur, whereas it does
occur in vivo in blood, subcutaneous, intramuscular and other living
tissues. Thus, we still have not identified the chemical portion of the
stresscracking mechanism. A new hypothesis that is currently under
investigation is that a living cellular interaction is required.
stress can be built into the polymer. Even more residual stress can be
built in during device manufacture, during implant and by normal
body motions. Relatively large localized stress can be generated at sur-
faces by imposed deformations such as bending or fixation ligatures.
Cracks will propagate until the driving stress is relieved, sometimes at
neutral axes. The depth of cracks will vary with the intensity of resid-
ual stress. Cracks, therefore, must be deepest adjacent to stress risers,
decreasing in depth and density as a function of distance from the
source of strain.
The above theory may help explain why Pellethanes 2363-80A and
55D tensile specimens did not crack in vivo, while other polyether poly-
urethanes did in early studies. The Pellethane 2363-80A and 55D spec-
imens were molded on a ram press. The resin must be very hot (in fact,
overheated) because heat is transferred to the pellets from the wall of
the ram barrel by convection only. The mold was not temperature con-
trolled, so it also heated up. Thus, these specimens were made with a
very hot melt and a warm mold, which appears to be optimum condi-
tions to minimize stress cracking. The later specimens were made on a
screw machine. This allowed molding at much lower temperatures, us-
ing both shear and heat to melt the polymer, shooting it into a rela-
tively cool mold. These are conditions that can maximize residual
stresses. Extruders are also screw machines which push molten poly-
mer into a cold take off tank to form tubing. Thus, the apparent para-
dox of in vivo cracking on devices made from polymers shown to be
biostable in early tests can be explained on the basis of different pro-
cessing conditions, maximizing or minimizing residual molecular
strains, causing or preventing stress cracking.
f~
8.
to
::3
0
s
a
0)
c
0
U
(I)
M
.~
to
a
.3:
.!E;
q
S2
-i:l
!~
248
ø
3
N
.2
&dquo;U
JD
’E
<ï:I
a.
E
d
~~ p
cEEC:(Q~/)§)’~ E CO &
E
&$dquo;0cO:C
.-
’v
Q gcho
fl
~
if CUD
c
O £
n
a !! «
s ~
~ . 00
<D
0
-2
z ~ ~-¡¡¡-ae
0 u m
dk_C
&dquo; oT5~5
.0 ~
W ~B~
1 $t55
m (/) 2 CD
-i:~
g g g~£
~
C ~ clJ
f ’~-0)S
g~-6~:g
~0~x=
gf Q) ~ 0 :ð OJ.!Q ae=~ (JO ~ QC
..c:=ø .
M-c~§~~
co ~..c -= ID >
i~~i .
Lð~;ae~2~
u~ t ~ c9 .~ O Q
L~s~~s~
p O 0) p ~ p
’SQ).!)~~ü.9
~ -6 &dquo;* ~ ~ ’* .ç
~ TI . - 0>&dquo; OJ
~~-g.850.8
-Q2C©-QJm
~,lflo~£~§
~t ~ a~ E ’4 .~c ° v
~§§~~~@8~
Q) ~Q)(J_O
Q pE U?c~LQ ~au~
C
t
~~§~°@-§
~ ’g t5 ; .S ~
~ 8 g.,ê ~ õ !-
~ ~ ~’y E ~.~
-5 .ç ~ ~ °b fô ~
.çW~Or
~ ~ Q) Q)&dquo;O
ocn0
tItQJOO#fi
O>&dquo;&dquo;’*--£. Q).2
~~ q p ~ N
’*~9ii~~
249
Figure 12. Cracks the surface of an oxidized piece of Pellethane 2363-80A. Notice the
random morphology of cracks and lack of crazing. 480 x.
250
AUTOOXIDATION
Figure 14. Tracings of Pellethane 2363-80A bulk infrared spectra. Top: spectrum of
unimplanted polymer. Center: spectrum from an explanted, severely cracked lead or lead
insulation degraded in 3% hydrogen peroxide at 37°C. Bottom: spectrum of nearly disin-
tegrated polymer in the presence of cobalt after 6 months in 3%, 37 °C hydrogen peroxide
or 1 year subcutaneous implant in a rabbit.
252
tion and propagation. The venous system and capillary bed where car-
diac pacing leads are placed is a low oxygen environment compared to
air. One possible source of oxygen in vivo is the inflammatory process,
especially as related to exocytosis [21,22]. Various phagocytes, such as
macrophages, are known to release hydrogen peroxide, hydroxyl radi-
cal, hydroperoxyl radical and superoxide anion. These chemicals and
many others must be deposited on the outer surfaces of an implanted
device as a consequence of the normal inflammatory &dquo;foreign body&dquo; pro-
cess. Pacing leads were immersed in 37°, 3fob HzO2 for periods of up to
six months [23]. The peroxide was replaced with fresh solution every
other day to assure maximum activity. A good qualitative correlation
with the actuarial performance of implanted leads was found. Those
lead models with the lowest in aivo actuarial survival also performed
poorly in the peroxide test [23,24]. Those that had not experienced oxi-
dative degradation in humans were not damaged in the peroxide test.
The same analytical results were obtained with leads immersed in per-
oxide and explanted from humans and animals. The same corrosion of
conductor coils was observed, the same optical and electron microscopic
appearance and texture of cracked material was seen, the same infra-
red spectra were obtained, etc. Thus, this test may have some predictive
value to evaluate the oxidative stability of an implantable device.
Figure 15. Nearly disintegrated Pellethane 2363-80A tubing over a cobalt mandrel after
6 months immersion in 3%, 37°C hydrogen peroxide. A comparable appearance is found
on some similar specimens after 2:1 1 year implant in a rabbit. 46 x.
(Pt and Ti). Controls were empty tubing and glassballs. The tubing was
dilated in acetone, then shrunk fit over the coil/mandrel/powder to
assure about .13 mm interference fit over the 1.9 mm o.d. coils or man-
drels. The tubing was extruded Pellethane 2363-80A (P80A) or 55D
(P55D) with a wall thickness of .13 mm. The ends of the 6&dquo; long speci-
mens were sealed to assure that no direct fluid leakage could occur. The
tubing to swell over Co, Ag, Pt and possibly Cr. By the end of the experi-
ment, substantial residues were found inside P80A samples containing
Fe (rust red) and Ti (cream-colored) > MP35N (cream-colored) > DBS
(white flakes) and silver (purple) > 304SS demonstrating that corro-
sion had occurred. Only traces of residues were found inside P55D spec-
imens in the presence of chromium, DBS, Elgiloy, glass (all whitish)
and cobalt (blue flakes). The P80A tubing disintegrated over cobalt
(Figure 15). Fine shallow cracks were observed in the presence of Ti
and DBS. Pellethane 2363-55D lumen surfaces cracked in the pressure
of Co and Ma The other P80A and P55D samples appeared to be un-
damaged optically, but some changes in tensile strength and elonga-
tion occurred (see Table 5).
Infrared spectra of P80A over Co showed extensive loss of PTMO
ether (1110 cm-i) relative to urethane ether (1080 crri 1) and amide III
(1220 cm-l) with the generation of a relatively large ester carbonyl at
1175 cm-1, the loss of CH, symmetrical stretch at 2860 cm-’ vs. asym-
metrical stretch at 2940 cm-1 and substantial changes in the complex
between 760 and 860 (Figure 14). Pellethane 2363-55D has less PTMO
ether compared to P80A to begin with. Nonetheless, there were clear
changes in the P55D infrared spectra in the presence of Co, which were
essentially the same as those found in Pellethane 2363-80A in the pres-
ence of Co. The other tubing specimens exhibited little to no change in
Figure 16. Pellethane 2363-80A lumen surface after 1 year subcutaneous implant in rab-
bits. Stripes of cracks are located at places where the metallic coil wires touch the poly-
mer. 100 x .
infrared spectra except for the lumens of P80A tubing over DBS and
Elgiloy. These spectra were similar to those described above. The bulk
polymer spectra were unchanged.
Thus, this in vitro study shows that significant bulk degradation of
P80A and P55D can occur in the presence of hydrogen peroxide and Co
and Mo, whereas surface damage was done where polymer contacted Ti
and DBS. Some bulk changes in the presence of Co bearing alloys,
MP35N and Elgiloy are presumed to be due to release of Co ions.
2Range 0-100%.
CONCLUSIONS
REFERENCES