Abstract—Objective: Accurate estimation of spatial gait char- costly and restricted to the availability of the physician. With
acteristics is critical to assess motor impairments resulting an ageing world population ([4], [5]) and a resulting increase
from neurological or musculoskeletal disease. Currently, however, in prevalence of chronic age-related disorders such as PD, this
methodological constraints limit clinical applicability of state-
of-the-art double integration approaches to gait patterns with also requires a change in the healthcare system towards a new
a clear zero-velocity phase. Methods: We describe a novel era of diagnostics and individualized treatment [6].
approach to stride length estimation that uses deep convolutional In the past, this has led to the development of several
arXiv:1609.03321v3 [cs.LG] 9 Mar 2017
neural networks to map stride-specific inertial sensor data to electronic measurement systems with the aim to extract spatio-
the resulting stride length. The model is trained on a publicly temporal gait parameters and aid the physician in an objective
available and clinically relevant benchmark dataset consisting of
1220 strides from 101 geriatric patients. Evaluation is done in a gait impairment scoring. These systems include for example
10-fold cross validation and for three different stride definitions. camera based, stationary 3D optical tracking of markers af-
Results: Even though best results are achieved with strides fixed to the body [7] or electronic walkways with embed-
defined from mid-stance to mid-stance with average accuracy and ded pressure sensors [8]. Optical tracking systems require
precision of 0.01±5.37 cm, performance does not strongly depend time-consuming preparation of the subject, but enable three-
on stride definition. The achieved precision outperforms state-
of-the-art methods evaluated on the same benchmark dataset dimensional, sub millimetre accurate tracking. Computerized
by 3.0 cm (36%). Conclusion: Due to the independence of stride walkways with embedded pressure sensors, however, require
definition, the proposed method is not subject to the methodolog- no preparation of the subject, but are limited in accuracy.
ical constrains that limit applicability of state-of-the-art double Furthermore, they can only assess gait parameters related to
integration methods. Furthermore, it was possible to improve ground contact, i.e. foot angles outside the ground plane can
precision on the benchmark dataset. Significance: With more
precise mobile stride length estimation, new insights to the not be tracked. Both of these systems are expensive and need a
progression of neurological disease or early indications might be laboratory environment for gait analysis which prohibits their
gained. Due to the independence of stride definition, previously use in everyday clinical practice or outpatient monitoring.
uncharted diseases in terms of mobile gait analysis can now be Unlike stationary systems, inertial sensors placed at the
investigated by re-training and applying the proposed method.
lower extremity enable mobile and unobtrusive gait analysis
Index Terms—deep learning, convolutional neural networks, that is unconstrained by the laboratory setting [1]. This type of
mobile gait analysis, stride length, regression measurement system is inexpensive and requires little subject-
preparation while at the same time enabling estimation of
I. Introduction three-dimensional spatio-temporal gait parameters ([9]–[15]).
Although mobile systems are increasingly developed to sup-
In a variety of neurological or musculoskeletal diseases,
port clinical diagnostics, monitor medical treatment or provide
motor impairment leads to specific gait characteristics. In
early indicators for gait impairment, they are still an open field
Parkinson’s disease (PD), for example, the resulting symp-
of research [6]. One of the most crucial components of such
toms of gait impairment are often characterised by shuffling
systems is accurate estimation of stride-by-stride parameters.
steps, reduced stride length, or impaired gait initiation [1]. In
Therefore, the purpose of this work is accurate estimation
advanced stages, PD patients can even experience spontaneous
of stride length based on stride-specific inertial sensor data
episodes of gait freezing. These symptoms can lead to a severe
captured at the subject’s feet.
reduction in patient mobility and quality of life [2] and it is
thus important to quantify and treat gait impairments as early
as possible. However, the physician’s rating of gait quality e.g. II. Related Work
in terms of the Movement Disorder Society’s revision of the There is a growing body of literature regarding stride length
Unified Parkinson Disease Rating Scale (MDS-UPDRS) [3] estimation from inertial sensors placed at the lower extremity
is often biased by subjectiveness. Moreover, it requires long- of the human body. The methods covered generally divide
time experience by the movement disorder specialist, is thus into two classes: (Bio)mechanical model based approaches
J. Hannink, T. Kautz, J. Barth and B. M. Eskofier are with the Digital Sports ([11], [12], [16]) and double-integration approaches ([9], [10],
Group, Department of Computer Science, University of Erlangen-Nürnberg [13]–[15], [17], [18]).
(FAU), Germany. C. F. Pasluosta was with the Digital Sports Group (FAU) and
recently joined the Laboratory for Biomedical Microtechnology, Department
of Microsystems Engineering, University of Freiburg, Germany. a) Model based approaches: The model based
S. Schülein, K.-G. Gaßmann are with the Geriatrics Centre Erlangen, approaches employ a double pendulum model for the
Waldkrankenhaus St. Marien, Erlangen, Germany. swing phase and an inverse double pendulum for the stance
J. Klucken is with the Department of Molecular Neurology, University
Hospital Erlangen, University of Erlangen-Nürnberg (FAU), Germany. phase. Three joints (hip and knees, ankle is considered fixed)
Corresponding author: J. Hannink, julius.hannink@fau.de and four segments (shank and thigh) are modelled. Integration
2
of the respective equations of motion is driven by gyroscope Another possibility is to combine all steps above in one
data captured at the shank and thigh of the subjects ([11], Kalman filter estimating orientation, velocity and position
[12], [16]). Salarian et al. [16] additionally try to reduce the from angular rates and acceleration. This is implemented by
number of sensing units either by duplicating data of the Ferrari et al. [15].
right thigh on the left thigh or by simplifying the model to a
single-pendulum with two sensors at the shanks. Apart from c) Clinical applicability: Tab. I summarizes the dataset
the amount of sensors needed on the patient, the models characteristics and results on all related work covered here.
described in [11], [12] and [16] confine the movement-model Regarding the mean accuracies and precisions, the majority of
to the sagittal plane and thus limit their applicability for methods are similar with accuracies within 1 cm and precisions
analysis of impaired gait. Furthermore, specific subject around 6 − 8 cm. There are two exceptions: Rebula et al.
characteristics (segment length etc.) are needed to scale the [13] reach good precision but are outperformed by others
biomechanical models. One of the major benefits, however, in terms of accuracy. The method proposed by Trojaniello
is the intrinsic implementation of biomechanical constraints et al. [14] achieves the best accuracy and precision almost
between the two legs. resolving the reference precision of ±1.27 cm [14]. They are
the only ones employing a magnetometer which might greatly
b) Double-integration approaches: The vast majority of improve their orientation estimation. However, this is only true
stride length estimation algorithms comes from the class of if indoor magnetic field distortions by e.g. ferroconcrete in the
double-integration methods. This family of algorithms makes proximity of the gait acquisition track are small enough. In a
use of the stance phase as a zero-velocity update point general setting without prior knowledge about earth magnetic
(ZUPT) to re-initialize the integration process. Due to intrinsic field disturbances due to construction materials, this out-rules
measurement errors in currently available inertial sensors, the use of magnetometers for indoor gait data acquisitions with
this is necessary in order to minimize the integration drift wearable sensors [21].
[19]. For one gait cycle, the algorithm consists of sensor This also applies to other studies listed in Tab. I. Regarding
orientation estimation from gyroscope measurements in order the number and type of subjects, there are major differences
to convert sensor readings from the local, constantly changing in terms of variability within the datasets. Not listed are
sensor frame to global coordinates. This is followed by gravity other quite important characteristics of the study populations,
cancellation and double-integration of accelerometer readings. i.e. stride length distribution, diversity of gait alterations,
Due to aforementioned limitations in currently available mo- precise age distribution, gender, study protocol or track length
bile sensors, the integration result has to be corrected or de- used during data acquisition. This is greatly threatening the
drifted in a last processing step by enforcing several constrains comparability of methods and their clinical application due to
(e.g. zero-velocity at the start and end of the stride, level non-representative study populations. It shows the urgent need
floor, etc.). This class of methods is mostly evaluated per leg of a unified evaluation of all stride length estimation methods
and biomechanical constrains are hard to implement. However, on the same, clinically relevant benchmark dataset.
three-dimensional sensor trajectories can be reconstructed al- Although there has been much progress on stride length
lowing a deeper analysis of gait characteristics or disease- estimation from inertial sensor data over the last years,
specific alterations. a variety of clinically relevant phenomena still cannot be
The contributions on the double-integration approach resolved due to the achieved measurement precision. These
covered here ([9], [10], [13]–[15], [17], [18]) all employ phenomena include for example the decline in stride length
accelerometer and gyroscope data captured at the subjects’ with age in healthy controls, the reduction in stride length
feet, Trojaniello et al. [14] additionally use magnetometer with disease progression in e.g. PD patients or the effect
information. Sampling rates are comparable and between 100 of medication. Hollman et al. [22] report normative data
and 200 Hz for all studies. The contributions mainly differ on mean stride length in elderly, healthy controls. Between
in the choice of orientation estimation method. Rampp et al. the age groups 70-75 years and 85+ years, stride length
[10] and Mariani et al. [9] apply a quaternion-based time decreases by 20 cm in males and 14 cm in females in the
integration of angular rates to estimate the sensor orientation, mean values over 294 participants [22]. This translates to a
as proposed by Sabatini [18]. Other contributions ([13]–[15]) reduction around 1-2 cm/year. Resolving this difference with
make use of a Kalman filter to obtain sensor orientations. mobile sensors requires the effect to be larger than typically
Integration is mainly done directly, i.e. forward in time. Only twice the standard deviation of the error distribution (i.e.
Trojaniello et al. [14] use a direct and reverse integration precision) obtained in the validation study of a stride length
scheme which spawns two integrations – one from the start estimation method. In order to resolve the yearly decline
and one from the end of the current stride – and computes a with age, the precision of a mobile stride length estimation
weighted mean between the two with weights depending on method would therefore need to be around 5 mm. A second
the distance to the next ZUPT. This is said to further resolve clinical application involves stride length estimation in PD
integration drifts [20]. De-drifting of integrated gravity-free patients. Hass et al. [23] report normative values on mean
acceleration signals is done in the spatial domain in [10], [13] stride length in PD patients over the course of the disease as
and [9] with a linear de-drifting function. Trojaniello et al. measured by the Hoehn&Yahr (H&Y) scale [24]. Between
[14] use a frequency-based de-drifting approach to suppress the groups H&Y < 1.5 (mild) and H&Y = 3-4 (severe), they
the low-frequency drift component in the integrated signals. observed a reduction in mean stride length of 11 cm for
3
TABLE I
Overview of related work: Dataset characteristics vs. mean accuracies and precisions
males and 12 cm for females based on data from 310 PD a convolutional neural network. The network can be trained on
patients [23]. Following the argument above, mobile stride big data sets [32] while at the same time promising excellent
length estimation methods would require precisions around individualisation capabilities to the patient at hand [33]. We
5 cm in order to resolve the effect of stride length reduction evaluate the approach in a cross-validation and obtain mean
with disease progression even on the very coarse H&Y scale accuracy and precision regarding stride length estimation. In
ranging from 0-5. Lastly, Bryant et al. [25] investigated the order to assess the method’s dependency on the data segment
change in stride length between the best possible (on) and provided as input, this is done for three different stride defini-
worst (off) medication state in PD patients. Based on data tions. These are either based on biomechanical events (heel-
from 21 PD patients, they found a stride length difference of strike and mid-stance) or contain the relevant information (i.e.
18 cm between the on- and off-state. These values help to get the swing phase where stride length is produced) but are not
a feeling about the ranges of stride length effects between best directly related to biomechanical events in the datastream.
possible to worst medication levels or mild to severe disease The method we propose here adds a completely novel
progression. Moreover, the clinician is probably interested approach to the problem of stride length estimation from
in more minute changes in clinical scores or medication inertial sensor data. It is not subject to the zero-velocity
levels corresponding to smaller effects on the parameter stride assumption that the double-integration approaches need to re-
length. To this date, mobile stride length estimation is not initialize the integration process. Thereby, the current work
yet precise enough for this kind of tasks. From a clinical enables spatial gait parameter estimation in situations where
perspective, the problem thus stays challenging. this assumption is violated, e.g. in patients with spasticity.
Compared to the long processing chain employed by double-
integration methods that results in amplification of errors
d) Towards a new type of methodology: Until now,
and requires cumbersome, careful and error-prone tuning of
the choice of method for stride length estimation has been
counter measures such as de-drifting, the method proposed
based on biomechanical, physical or geometrical reasoning and
here is purely data-driven. There is no need to carefully
these approaches have proven to work in technical validation
keep track of and account for the accumulating error during
settings. However, one could also aim to learn a regression
processing.
function linking the raw sensor data of a stride directly with the
Our main contributions are: 1) We show feasibility regarding
corresponding stride length. Similar approaches were reported
estimation of spatial gait characteristics with deep convolu-
by Aminian et al. [26] in 1995 where a two-layer perceptron
tional neural networks. 2) With that change in methodology,
was used to estimate speed and incline of human walking.
we drop the zero-velocity phase assumption that constrains
Due to the limits on computing power 20 years ago, the
state-of-the-art double integration methods and limits their
neural network structure presented by Aminian et al. [26] is
applicability.
very shallow with 10 input nodes, one hidden layer of five
units and one output node. A 10 feature parametrization of
the accelerometer signal based on physiological and statistical III. Methods
reasoning is used as input to the network. a) Data Collection and Setup: We use a benchmark
With the recent advances in deep learning ([27]–[29]) and dataset collected by Rampp et al. [10] that is publicly avail-
user-friendly interfaces to such techniques like torch [30] or able at https://www5.cs.fau.de/activitynet/benchmark-datasets/
google’s TensorFlow library [31], the parametrization of the digital-biobank/ and briefly described here. For data collection,
input data has become obsolete. By regressing against raw the Shimmer2R sensor platform [34] was used. It contains
sensor data instead of parametrizations and by employing a 3d-accelerometer (range ±6 g) and a 3d-gyroscope (range
sufficiently deep architectures, the true potential of the neural ±500 ◦ /s) and was attached laterally right below each ankle
network approach can be utilized. Based on a publicly avail- joint (see Fig. 1). The same shoe model (adidas Duramo 3)
able and clinically relevant benchmark dataset, we describe was used by all subjects to avoid gait changes due to different
a method to learn a non-linear relationship between stride- shoe characteristics [35]. Data was captured at 102.4 Hz and
specific sensor data and the corresponding stride length with a resolution of 12 bit. Simultaneously, validation data was
4
Accelerometer [norm]
AccX AccY AccZ
was 609.60 cm × 60.96 cm with a spatial resolution of ±1.27 0.5
cm and a track width of 84 cm. 0.0
−0.5
−1.0
1.0
Gyroscope [norm]
GyrX GyrY GyrZ
0.5
0.0
−0.5
−1.0
0.0 0.5 1.0 1.5 2.0 2.5
Time [s]
Fig. 1. Placement of the inertial sensor and axes definition. Fig. 2. Exemplary input signal for a stride defined from MS→MS after
preprocessing.
In total, 116 geriatric inpatients were assessed with this
setup at the Geriatrics Centre Erlangen (Waldkrankenhaus
St. Marien, Erlangen, Germany). Written informed consent two events within the sensor data is done according to [10].
was obtained prior to the gait assessment in accordance with We call the stride definition provided by the dataset msDTW
the ethical committee of the medical faculty at Friedrich- because of the multi-dimensional subsequence dynamic time
Alexander University Erlangen-Nürnberg (Re.-No. 4208). warping they used for segmentation (for details see [10], [40]).
Patients performed a detailed geriatric assessment, An exemplary, pre-processed input signal for the network
described in detail by Rampp et al. [10]. For the scope of for one stride defined from MS→MS is shown in Fig. 2.
this manuscript, we focus on the free walking test over 10 m
at comfortable walking speed instrumented with the inertial c) Network Architecture: The neural network architec-
sensors and the GAITRite system. Here, the number of strides ture we choose is a two layer convolutional network followed
acquired from each patient ranges from 7 – 32 with a mean by one fully connected layer and a readout-layer (see Fig.
of 12 strides per patient. After excluding datasets from eight 3). The main theoretical motivation for this choice is the
patients due to medical reasons, two due to inertial sensor locality of features gained by this type of architecture [27]. The
malfunction and additional five due to measurement errors network is thus forced to take the true topology of the input
with the GAITRite system, a total of 101 datasets were left data as multi-channelled, synchronized time series into account
for training and evaluation of the stride length estimation and maintain its temporal context. It only considers local
method proposed here. In 54% of the study population, gait connections and does not link information from temporally
disorders or fall proneness were diagnosed. The other top far away regions in the input signal.
three diagnoses were heart rhythm disorder (70%), arterial The preprocessed inertial sensor data xi, j with i = 1 . . . L0
hypertension (69%) and coronary artery disease (41%) all of samples and j = 1 . . . N0 channels serves as input to the
which are associated with gait and balance disorders [38]. network. In the first convolutional layer, feature maps are
Therefore, this dataset constitutes a clinically relevant study constructed from xi, j by convolution with k = 1 . . . N1 kernels
population both in terms of the number of subjects and the ψ(1)
i, j,k of size i = 1 . . . L1 and j ∈ [1, N0 ]. A bias term bk is
(1)
presence of unpredictable gait alterations. added to each feature map before a ReLU activation function
[41] produces the output activations of the first convolutional
b) Preprocessing: Before the inertial sensor data is fed layer
to the convolutional neural network, we perform a series of
preprocessing steps. These include extraction of annotated X
strides (a segmentation of the continuous signal into strides ck = ReLU ψ j,k ∗ x j + bk
(1) (1) (1)
(1)
j
is already supplied by the dataset) and calibration from raw
sensor readings to physical units [39]. Due to different sensor
with k ∈ [1, N1 ] and ReLU(z) = max(0, z).
mounting on the shoe, coordinate system transformations are
needed to align sensor axes on left and right feet. Furthermore, After each convolutional layer, we insert a max-pooling
the signals from accelerometer and gyroscope are normalised layer that downsamples the feature maps by taking the maxi-
w.r.t. the respective sensor ranges and zero-padded to fixed mum in non-overlapping windows of length r. The downsam-
length of 256 samples per stride to ensure equally scaled and pling factor thus is 1/r. The output from the first max-pooling
fixed size input to the network. layer is called o(1) i, j with i ∈ [1, L0 /r] and j ∈ [1, N1 ].
Since we will compare the algorithm on different stride In the second convolutional layer, the same approach is
definitions, we need to detect mid-stance (MS) and heel-strike repeated with different kernels and biases. Given the input
(HS) events in the stride segmentation provided by the dataset signal o(1)
i, j , feature maps are constructed by convolution with
and adjust the stride borders accordingly. Detection of these k = 1 . . . N2 kernels ψ(2) i, j,k of length i = 1 . . . L2 and j ∈ [1, N1 ].
5
(ro)
This is done by multiplication with a weight vector Wi,1 with
i ∈ [1, Nfc ]. A final bias b(ro) is added to arrive at the estimate
of the target variable
X
y= (ro) (fc)
Wi,1 oi + b(ro) (4)
i
Similarly, a target vector (y1 , . . . , yn ) representing a set of stride
features could be estimated by changing the dimensionality of
weight and bias in the readout layer.
For our application of stride length estimation, the input
data is of length L0 = 256 with N0 = 6 channels. On the first
layer of the network, we chose to learn N1 = 32 filters of
length L1 = 30 samples, followed by N2 = 64 and L2 = 15 on
the second. Max-pooling is done in non-overlapping windows
of length r = 2 and the fully connected layer has Nfc = 1024
nodes. Given the downsampling factor and the sampling rate,
the filter lengths both correspond to approximately 0.29 s. The
theoretical motivation for this choice is to keep the resulting
receptive field size constant (L1 = r L2 ) while increasing the
amount of features maps (N2 > N1 ) with network depth.
0.5, so every node on the fully connected layer has a 50/50 can not formally evaluate the performance of individualized
chance of being dropped during training. During testing, models here due to the lack of data (6 strides on average per
however, the full architecture is used and no connections are foot and patient), the results on the training set serve as a
dropped. substitute for the individualized modelling case where prior
The network is implemented and trained using google’s information is available.
TensorFlow library [31].
b) Stride Length Estimation on Unseen Data: Tab. II lists
e) Evaluation Scheme: Evaluation of the proposed average statistics from the 10-fold cross validation for each of
method is based on a 10-fold cross validation scheme. The the three stride definitions. Bland-Altman plots in Fig. 5 assess
strides from 101 patients on the dataset are sorted into training the agreement between the two measurements for all three
and test set depending on the patient identifier to ensure stride definitions. Additionally, Fig. 5 includes regression lines
distinct splits of the dataset. For each of the three stride between the measurement error y − yref and the measurement
definitions msDTW, HS→HS and MS→MS, the model is agreement 12 (y + yref ) in order to assess the dependency of the
evaluated in such a cross-validation scheme and the stride measurement error w.r.t. the stride length.
length is estimated on the test set in each fold. The predictions The model yields best performance when strides are de-
from individual folds are then pooled to arrive at average fined from MS→MS with a mean accuracy and precision of
statistics for the current stride definition. Evaluation statistics 0.01±5.37 cm. The range of precisions achieved for individual
include average accuracy ± precision which correspond to the patients ranges between ±1.54 cm in the best case (patient
mean and standard deviation of the (signed) error distribution. P114) and ±10.43 cm in the worst case (patient P106).
Additionally, we state precision relative to the mean stride Furthermore, the stride length estimation with the proposed
length, average absolute accuracy ± precision as well as Spear- approach is robust w.r.t. the stride definition. When trained
man correlation coefficients between the reference system and and evaluated on the other two stride definitions (HS→HS
our approach. and msDTW), the mean precision differs only marginally from
In order to assess the learning speed and performance of the the one achieved on the MS→MS strides. The mean accuracy,
convolutional neural network, we compute the training error however, decreases to −0.27 cm and −0.42 cm for HS→HS
and the corresponding precision over the training iterations for and msDTW strides respectively.
an exemplary 90/10% (patient-wise) split of the dataset. The regression lines that indicate the dependency of the
measurement error on the stride length regime are comparable
IV. Results for all three stride definitions. The low stride length regime
a) Training: On a mobile workstation equipped with an is on average overestimated while high stride lengths are on
Intel Core i7 processor (8 cores), 16GB of RAM and no GPU average underestimated. This reflects an imbalance found in
acceleration, training takes around 20 min. per fold. the training data. Since a fixed track length was used for data
acquisition, there are more examples from the low stride length
regime on the dataset compared to high stride length strides.
V. Discussion
Overall, our results show that stride length estimation with
deep convolutional neural networks based on stride-specific
inertial sensor data is possible. This is the first major con-
tributions of this work. We show feasibility and contribute a
completely new type of algorithm to the field of stride length
estimation. Although the achieved precision of 5.4 cm is not
sufficient to track the subtle yearly decline in stride length
in healthy individuals, it is sufficient to resolve stride length
differences observed in PD patients [23].
The best result achieved with the method described here out-
performs the approach by Rampp et al. [10] by 3.0 cm (36%)
Fig. 4. Error E() and precision evaluated on the entire training set for an
exemplary 90/10% train/test split of the dataset during training. in precision and also gives a more balanced error distribution.
We can directly compare these two approaches since they are
For an exemplary fold, Fig. 4 shows the error as well as evaluated on an identical dataset. This is not the case for the
the precision on the entire training set over the iterations. other methods found in literature (see Tab. I). Although our
The fixed number of 4000 iterations is sufficient to reach a approach shows similar performance, the discrepancy in eval-
stable regime of the error E() and precision evaluated on uation datasets and their characteristics limits comparability.
the entire training dataset. Moreover, we reach the reference This manifests the urgent need for a standardised and clinically
precision after training which demonstrates the usability of relevant benchmark dataset regarding stride length estimation
the described approach for estimating stride length at high from inertial sensor data. Given the number of subjects and the
precision with a model tuned to the individual. Because we presence of gait alterations in the study population used here, it
7
TABLE II
Results of the deep convolutional neural network approach to stride length estimation for three stride definitions compared to the Gaitrite reference
system. Listed are mean stride lengths for both systems, the error of measurement is reported as mean accuracy and precision, relative precision, mean
absolute accuracy and precision and Spearman correlation (CC).
Stride definition Our approach GAITRite Mean acc. ± prec. Rel. prec. Mean abs. acc. ± prec. CC
MS→MS strides 80.03 ± 22.92 cm 80.02 ± 23.40 cm 0.01 ± 5.37 cm 6.7 % 3.96 ± 3.63 cm 0.97
HS→HS strides 79.75 ± 22.82 cm 80.02 ± 23.40 cm −0.27 ± 5.43 cm 6.8 % 4.11 ± 3.57 cm 0.97
msDTW strides 79.60 ± 22.83 cm 80.02 ± 23.40 cm −0.42 ± 5.42 cm 6.8 % 4.08 ± 3.59 cm 0.97
MS→MS strides
benchmark-datasets/digital-biobank/.
40
Fold 1 Fold 3 Fold 5 Fold 7 Fold 9 The second main contribution of this work is the indepen-
Fold 2 Fold 4 Fold 6 Fold 8 Fold 10
30 dence of stride definition. The mean accuracy and precision of
Measurement error y − yref [cm]
30
Fold 2 Fold 4 Fold 6 Fold 8 Fold 10 thus be detected more accurately. Subsequently, the MS→MS
Measurement error y − yref [cm]
20
Lastly, the high precision on the training dataset shown in
+1.96σ
10 Fig. 4 promises excellent results in case of individualization.
0 If we have a training/calibration dataset to tune the neural
network to the patient at hand, i.e. by collecting one gait
−10 −1.96σ
sequence instrumented with inertial sensors and the GAITRite,
−20 the proposed method could reach the precision of the ref-
−30
Mean acc. ± prec. −0.42 ± 5.42 cm
erence. This would enable highly accurate stride-by-stride
−40
Regression line −0.03 x + 1.61 cm analysis of spatial gait parameters in daily clinical routine or
0 20 40 60 80 100 120 140
outpatient monitoring at a level sufficient to address clinically
Measurement agreement 12 (y + yref ) [cm]
relevant phenomena as for example the decline in stride length
with disease progression in PD patients or medication effects.
Fig. 5. Bland-Altman plots from the evaluation of the proposed algorithm
in a 10-fold cross validation for all three stride definitions. Mean accuracy Certainly, this has to be formally evaluated in the future and
and 95% confidence intervals (±1.96 σ) as well as a regression line between is a possible extension of the presented method.
mean stride length and the measurement error are shown.
Furthermore, our methodology is not limited to stride length
estimation. Targeting the underlying regression task at a dif-
ferent stride-specific gait characteristic e.g. stride width or
would serve well as such a benchmark dataset. Furthermore, timings of individual gait phases does not involve changing
it is publicly available at https://www5.cs.fau.de/activitynet/ the underlying algorithms. The very same technology can be
8
[19] A. Peruzzi, U. Della Croce, and A. Cereatti, “Estimation of stride length Overfitting,” Journal of Machine Learning Research, vol. 15, pp. 1929–
in level walking using an inertial measurement unit attached to the foot: 1958, 2014.
A validation of the zero velocity assumption during stance,” Journal of
Biomechanics, vol. 44, no. 10, pp. 1991–1994, 2011.
[20] M. Zok, C. Mazzà, and U. Della Croce, “Total body centre of mass
displacement estimated using ground reactions during transitory motor
tasks: Application to step ascent,” Medical Engineering and Physics,
vol. 26, no. 9 SPEC.ISS., pp. 791–798, 2004.
[21] W. H. K. de Vries, H. E. J. Veeger, C. T. M. Baten, and F. C. T. van der
Helm, “Magnetic distortion in motion labs, implications for validating
inertial magnetic sensors,” Gait and Posture, vol. 29, no. 4, pp. 535–541,
2009.
[22] J. H. Hollman, E. M. McDade, and R. C. Petersen, “Normative spa-
tiotemporal gait parameters in older adults,” Gait and Posture, vol. 34,
no. 1, pp. 111–118, 2011.
[23] C. J. Hass, P. Malczak, J. Nocera, E. L. Stegemöller, A. Shukala
et al., “Quantitative normative Gait data in a large cohort of ambulatory
persons with parkinson’s disease,” PLoS ONE, vol. 7, no. 8, pp. 4–8,
2012.
[24] M. M. Hoehn and M. D. Yahr, “Parkinsonism: onset, progression, and
mortality,” Neurology, vol. 57, no. 2, p. 318 and 16 pages following,
1967.
[25] M. S. Bryant, D. H. Rintala, J. G. Hou, E. C. Lai, and E. J. Protas,
“Effects of levodopa on forward and backward gait patterns in persons
with Parkinson’s disease,” NeuroRehabilitation, vol. 29, no. 3, pp. 247–
252, 2011.
[26] K. Aminian, P. Robert, E. Jequier, and Y. Schutz, “Estimation of speed
and incline of walking using neural network,” IEEE Transactions on
Instrumentation and Measurement, vol. 44, no. 3, pp. 743–746, 1995.
[27] Y. LeCun, Y. Bengio, and G. Hinton, “Deep learning,” Nature, vol. 521,
no. 7553, pp. 436–444, 2015.
[28] A. Krizhevsky, I. Sutskever, and G. E. Hinton, “ImageNet Classifica-
tion with Deep Convolutional Neural Networks,” Advances In Neural
Information Processing Systems, pp. 1–9, 2012.
[29] G. E. Hinton and R. Salakhutdinov, “Reducing the Dimensionality of
Data with Neural Networks,” Science, vol. 313, pp. 504–507, 2006.
[30] R. Collobert, K. Kavukcuoglu, and C. Farabet, “Torch7: A matlab-
like environment for machine learning,” 2011, software available from
torch.ch.
[31] M. Abadi, A. Agarwal, P. Barham, E. Brevdo, Z. Chen et al., “Tensor-
Flow: Large-scale machine learning on heterogeneous systems,” 2015,
software available from tensorflow.org.
[32] Xue-Wen Chen and Xiaotong Lin, “Big Data Deep Learning: Challenges
and Perspectives,” IEEE Access, vol. 2, pp. 514–525, 2014.
[33] S. Stober, A. Sternin, A. M. Owen, and J. A. Grahn, “Deep Feature
Learning for EEG Recordings,” arXiv:1511.04306, 2015.
[34] A. Burns, B. R. Greene, M. J. McGrath, T. J. O’Shea, B. Kuris et al.,
“SHIMMERTM – A wireless sensor platform for noninvasive biomedical
research,” Sensors, vol. 10, no. 9, pp. 1527–1534, 2010.
[35] J. C. Menant, J. R. Steele, H. B. Menz, B. J. Munro, and S. R.
Lord, “Effects of walking surfaces and footwear on temporo-spatial gait
parameters in young and older people.” Gait & Posture, vol. 29, no. 3,
pp. 392–7, Apr. 2009.
[36] A. L. McDonough, M. Batavia, F. C. Chen, S. Kwon, and J. Ziai, “The
validity and reliability of the GAITRite system’s measurements: A pre-
liminary evaluation.” Archives of Physical Medicine and Rehabilitation,
vol. 82, no. 3, pp. 419–25, Mar. 2001.
[37] K. E. Webster, J. E. Wittwer, and J. a. Feller, “Validity of the GAITRite
R