Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.

0
Interferential Stimulator

Portable Interferential stimulator

SI-500P
Portable ICT

User manual
(Version 1.0)

GOODPL INC.
#1-114, 115 Wonju Medical Instrument Industry Park, 1720-26 Taejang-dong,
Wonju-si, Gangwon-do 220-962, Korea
Tel. +82-1644-5866, Fax. +82-33-746-1545
Webpage: http://www.goodpl.kr
Email address: kckang@goodpl.kr

1 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

Product and components

1 2 3 4 5

No Name Function

1 Control panel Possible to check control and status of machine

2 Connector for silicon pad Part that connect with silicon pad cable

3 Connector for suction cup Part that connect with suction cup cable

4 Power switch ON/OFF switch

5 Power cable connector Power connector(AC220V, 50/60Hz)

6 Accessories Suction cup, silicon pad, cable

2 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

SI-500P features

Converting digital to analogue current signals makes the stimulation softer and enhances the product
durability further. 9 auto modes and 3 manual modes are supported and cup and pad can be selectively used.

1. Computerized electromagnetic interferential stimulator


- Unlike already existing electric stimulators used to relieve pain and facilitate regenerate of injured
ligaments, muscles, tendons, and other soft-tissue, SI-500P uses computer controlled electric
pulses to stimulate cell correspondence and injured soft tissue nerves to remove acute and
chronic pain. SI-500P has especially demonstrated excellent results for chronic disk pains, low
mobility induced joint pains, and post surgery muscle regeneration

2. Automated computer controlled programs


- Nine types of fully computer controlled standard mild stimulation frequencies were programmed
into the device. Newly developed programs can be uploaded and used on the device in the future.
Simply, through the use of the treatment functions of fixed currents, pure 2,500/ 4,000/ 5,000 Hz
signals and the interferential current therapy functions of certain frequencies were programmed
into the device to allow for greater autonomy of treatment options. It is also possible to periodically
change the fixed current using sweep frequencies and sweep time

3. F/A (Fine Adjustment) function


- Unlike previous analog devices, the digital signal generating SI-500P increases intensity in units of
1mA. To accommodate this characteristic fine adjustment of signal adjustment was made possible.
At threshold limits, it is possible to fine adjust intensity in units of 0.2mA by simply pushing the
encoder button. This feature minimizes the intensity felt by the patient

4. Internal voice function


- The status of the device can easily be checked due to a melody that signals the start and finish of
treatment

5. Other functions
- SI-500P features safety functions such as the zero start function, which initially sets the treatment
intensity at zero. Excess current prevention circuits were used to prevent higher than desirable
currents from reaching the area being treated. Also a moisture prevention system was applied to
protect the suction area to enable safe treatment and protection of the device

3 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

Operational panel configuration and functions

Operational panel configuration and functions

1-INTENSITY
By turning the Encoder to the right the output intensity can be adjusted
- After making adjustments to increase the output current and no further increases occur for the
next 10 seconds, an automatic lock function kicks into prevent any further increases and only
decreases are possible. In this case, in order to increase the output current the user must first
decrease the current and then increase it
- Under normal conditions the intensity can be increased in units of 1mA. By pushing the
Encoder button fine adjustments in units of 0.2mA is possible
- Changing basic frequency (2,500Hz, 4,000Hz, 5,000Hz)
(Push the encoder button to change the basic frequency [Can only be configured before
treatment begins]

2-STOP
Stop treatment mode

3-AUTO: Auto programs (9 Types) – [A-1 ~ A-9]

4 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

4-MANUAL: User configured programs (3 Types) – [P-1 ~ P-3]

5-SET
(a) Time configuration: 1 min ~ 30 min (Units of 1 min)
(b) When pressing manual mode for 2 seconds the device returns to user configure mode
- START: Configures the start frequency (1Hz~200Hz)
- END: Configures the end frequency (1Hz ~ 200Hz)
- ro: Round time (sec) from the start frequency to end frequency (20, 30, 40, 50, 60 sec)
- ti: Repeat time from the start frequency to end frequency (20, 30, 40, 50, 60 sec)
- ue(VECTOR): Interferential low current application time (CON, V1, V2, V3)
- nn(MODULATION): Amplitude between the fixed current and interferential current (0%, 25%,
50%, 75%, 100%)
- Press the ‘SET’ button when moving on to the next level through Encoder configuration

6-WHITE
WHITE electrodes can be separately configured.

7-RED
RED electrodes can be separately configured.

8-TIME
Time configuration: 1 min ~ 30 min (Units of 1 min)

9-MODE

10-PULS: Adjusts the suction cycle (Total four levels)


(Level 1: CON, Level 2: 15/min, Level 3: 20/min, Level 4: 30/min)

11-VAC: Adjusts the pressure intensity (Total four levels)


(Level 1: Off, Level 2: 100mmHg, Level 3: 150mmHg, Level 4: 200mmHg)
Without further changes to the VAC, treatment will begin using the most recently configured VAC
value

12-F/A ,MODE : when use intensity function will displayed on the panel

Purpose
The purpose of using the GP-500P can be largely classified as nerve suppression, muscle
treatment, promotion of blood circulation, massaging, and fat cell reduction. A more in-depth
classification is shown as the following;

1. Physical and chemical recuperation of muscles and muscle related functions


2. Acceleration of muscle physiological activities and recovery (muscle development)
3. Maintenance and promotion of blood and lymph circulation
4. Mitigation of abnormal muscle spasms

5 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

5. Prevention of loss of muscle proteins


6. Promotion of blood circulation
7. Stabilization of nerves
8. Curtailment of obesity

Items of caution
The following users should consult a physician before using this device

-Persons undergoing menstruation or are pregnant


-Persons with malignant tumors
-Persons with high temperatures (over 38℃)
-Persons with sensitive skin or skin irritations
-Persons with high blood pressure (over 180mmHg)
-Do not use this device on the heart, head, face, mouth, genitals, or skin irritations
-Persons with change in temperature induced heart diseases
-Persons with epilepsy
-Persons experiencing abnormalities on the treated area following treatment
-Persons with acute diseases
-Persons with hearing disabilities
-Persons with infectious diseases

※ Please do not put a heavy thing to the top of product.


※ Accessories of product are composed in suction cup, suction catheter, gel pad, and
electrode catheter.
※ Please check the catheter all times before using the device and replace it with a new one
if necessary

6 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

How to install and items of caution

※ Please adhere to the following instructions before installing this product.

1. How to install
① Check the power supply (AC230V/50Hz)
② Connect the catheter to the device by color
(The color of the catheter cup may differ according to the product)
③ Soak the sponge, which is an accessory, and squeeze the sponge for adequate dampness

2. Items of caution
- Please designate a manager responsible for the device and do not have anyone other than
this person operate the device. Also, please thoroughly read through the following items of
caution to ensure safe operation of this device
① Items of caution during installation
(Please take note of the appropriate location to install this device)
- Install in areas that are not wet
- Install in areas unexposed to dust and are well ventilated
- Install in areas that can supply the appropriate voltage, frequency, and consumed power
- Install in areas that are not affected by mechanical vibrations or electric abnormalities
- Install in areas that are not exposed to gases produced by chemical substances
- Check to see if the power cord is appropriately connected to the socket
- Install in areas that are not affected by extreme temperatures, humidity, salinity, or ions
- Areas that are exposed to electromagnetic pulses may cause the devise to function
abnormally
- The device may generate unstable output if the device is used in close proximity (1m) to
shortwave or microwave devices

② Items of caution before using the device


- Check the supplied voltage and see if the power plug is appropriately connected
- Check to see if each catheter is appropriately connected with the device
- Check to see if the timer and output configurations are accurately set
- Please use the suction catheter supplied for this product by Jungwoo TM and refrain from
using catheter manufactured by other companies
- Please read the user manual regarding the functions and programs of this device and
operate the device in accordance with the conditions of the patient

③ Items of caution while using this device


- Check to see if the device is operating properly
- If any abnormalities are detected, follow the appropriate troubleshooting procedures and
continue to operate only after confirming the device is back to normal

7 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

④ Items of caution after using the device


- Turn the power switch off and separate the power plug from the socket
- Keep the low frequency catheter cord clean
- Use a clean towel or tissue to wipe the low frequency catheter cup and maintain it in clean
conditions for storage

⑤ Storage conditions and storage method


- Store in areas that are not humid
- Do not store in areas affected by high saline/ sulfuric air
- Do not store in areas affected by vibration and impact or areas that are sloped
- Do not store in areas affected by gases produced from chemical substances

⑥ Operational taboos(Please take extra caution before using this device for the
following patients)
- Do not use this device for patients with pacemakers or other implanted electronic devices
without the consent of a physician
- The use of a high frequency stimulator in conjunction with this device may result in burns
near the stimulator electrode and damages to the electrode
- Do not use this device for patients deemed by a physician as being too physically weak

⑦ Other items
- Do not perform unauthorized alterations or repairs to this device
- Please contact our headquarters or a nearby sales representative for any problems with the
device
- Perform regular inspections of the device to maintain the device in optimal conditions

8 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

How to use the SI-500P


① Using the ON/OFF switch located on the back of the device, turn the device ON
② Select AUTO or MANUAL mode
Auto mode : AUTO1 (Peripheral diseases), AUTO2 (Pain alleviation), AUTO3 (Multi-function
treatment), AUTO4 (External wounds), AUTO5 (Muscle atrophy), AUTO6 (Chronic), AUTO7
(Promotion of circulation), AUTO8 (Synthetic treatment – edema), AUTO9 (Circulation diseases)
Manual mode: P-1, P-2, P-3

③ Select Time(1~30minutes)

④ Select the fixed current (2.5KHz, 4.0KHz, 5.0KHz)


⑤ Adjusts the pressure intensity (Total four levels)
(Level 1: Off, Level 2: 100mmHg, Level 3: 150mmHg, Level 4: 200mmHg)
Without further changes to the VAC, treatment will begin using the most recently configured
VAC value
⑥ Patch suction cup as the same color positioned in the direction of the cross.

<Cup> <Pad>

9 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

⑦ Configure the suction cycle of the suction catheter by pressing the PULS button (OFF:CON,
P1:15/Min, P2:20/Min, P3:30/Min)
- The suction cycle will begin 10 seconds after treatment begins
⑧Turn the intensity volume and adjust the output intensity according to the patient’s condition
⑨ If the Red/White balance is wrong, turn the appropriate button and adjust accordingly

10 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

What is the ICT AUTO function?


1) AUTO1 (Peripheral diseases)
- By mitigating pain and inflammation transmitters by accelerating the process of tissue expansion
and promoting metabolic circulation and maintaining reversible and partial depolarization of vascular
smooth muscles(anti-bleeding effect), the contraction of vascular smooth muscles and afferent
contraction of the vein and lymph are facilitated to produce anti-edema effects(cellular membrane
permeation is promoted through fine vibration of cellular ions) and promote the re-absorption of
tissue fluid through capillary vessel walls. Also the depolarization of pain transmitting nerve fibers
cut off pain transmission to produce peripheral pain alleviation.

2) AUTO2 (Pain alleviation)


- Central pain alleviation is achieved through the anti-stimulation effect and discharge of endorphins
through stimulation of sensory fibers. Vasodilation is induced following initial increases in reversible
sympathetic nerve activities
- Used for pain alleviation by removing pain alleviation metabolites and transferring exudates.

3) AUTO3 (Multi function treatment)


- Used on dispersed treatment areas. This treatment option induces vasodilation by applying various
forms of stimulation to the motor nerves and tissue and promotes pump and stimulation effects on
various tissue exudates.

4) AUTO4 (External wounds)


- Peripheral sympathetic nerves are stimulated to activate sympathetic nerve receptors, which in
turn results in vasoconstriction and reduction of blood flow to damaged tissue and the reduction of
capillary vessel pressure to increase re-absorption of tissue fluid through capillary vessel walls.
- This induces contraction of the vein and increases afferent lymph transfer, which reduces external
fluids from the blood vessel. In other words tissue fluids resulting from damaged areas are
disintegrated and transferred. Appropriate fixture and pressure is required considering that shortly
following this process, vasoconstriction is reduced and vasodilatation occurs.

5) AUTO5 (Muscle atrophy)


- Excessive muscle spindle activities are prevented to help muscles recover adequate tension by
providing soft and intense constriction to muscles experiencing spasms and contracture failures.

6) AUTO6 (Chronic)
- Ion permeation to cell membranes is increased by repeating a process of exiting and relaxing
tissues to increase congestion within blood vessels that activate ion activities within cells (various
alterations).

7) AUTO7 (Promotion of circulation)


- Oxygen and antibody supply activities are promoted through vasodilation and improved circulation
by inducing fine vibrations of cellular ions. During this process, enzyme activities and cellular mitosis
increase resulting in the facilitation of healing

8) AUTO8 (Synthetic treatment – edema)

11 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

- This treatment is an option for patients suffering from intractable edema and patients that respond
relatively less well to electric therapy.

9) AUTO9 (Circulation diseases)


- This treatment is an option for patients suffering from long term fixation following injuries such as a
bone fracture or patients experiencing geriatric diseases.

12 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

13 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

14 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

Product after service (A/S)


1. Guidelines to device malfunctions

This device requires the utmost safety precautions considering it is a device that discharges electric
currents to an injured area of the human body. If any of the following malfunctions are detected
please visit the manufacturer or the sales representative to receive assistance. For any cases in
which the device is deemed to be operating abnormally, please make sure the device is fully
repaired before further use

[Malfunctions]
 Cases in which output discharge does not change even after applying the catheter and
turning the intensity knob
 Cases in which the device occasionally does not operate normally
 Cases in which the display window does not operate normally even though the power
switch is on
 Cases in which the suction cup does not function properly on the area of treatment

2. After service contact information

Please receive A/S immediately for any cases in which the device is operating abnormally
 Tel. no: +82-1644-5866 (A/S information)
 Fax. no: +82-33-746-1545
 Webpage: http://www.goodpl.kr
 Email address: kckang@goodpl.kr

Product Specifications

No Item Specification

1 Power AC

2 Voltage 220V

3 Frequency 50/60 ㎐

4 Power consumption rate 50 VA

Manual mode: 60 mA(500Ω)


5 Max output current
Auto mode: 50 mA(500Ω)

6 Max output voltage 90Vp-p(500Ω)


7 Pre-programmed frequency 2,500Hz, 4,000Hz, 5,000Hz
8 Working time 1~30minutes(per 1 minute)

15 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

9 Output wave

Level 1 OFF

Level 2 V1 150mmHg
10 Suction pressure
Level 3 V2 180mmHg
Level 4 V3 250mmHg

Level 1 CON

Level 2 P1 13times/Min
11 Suction cycle
Level 3 P2 15times/Min
Level 4 P3 20times/Min

13 Dimension 420× 380× 180mm

14 Weight 8 kg

16 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator

Warranty

Name of model SI-500P


Type of product Portable Interferential Stimulator
Duration of warranty 1 year after purchase
Purchase address (signature)
Purchase date yyyy/mm/dd
Manufacture no.
Name of hospital (signature)
Hospital contact info

- This product has undergone stringent quality management and has passed testing
procedures in accordance with CE (MDD 93/42/EEC), EN ISO 13485:2003 and KGMP
standards. This warranty will cover for damages or abnormalities that occur during normal
conditions of operation

- Please submit your warranty to a GOODPL INC service center or sales representative for
after service and receive assistance regarding repairs

- If the sales representatives or place of purchase does not offer repairs please contact the
service center or hotline as shown on the bottom of the warranty

- If this warranty is sent via fax to our headquarters, we will make a phone call to confirm the
device number and register the device in our database. For all other cases, the warranty
begins on the day the device is sent out for delivery from our headquarters

GOODPL INC.
#1-114, 115 Wonju Medical Instrument Industry Park,
1720-26 Taejang 2-dong, Wonju-si, Gangwon-do 220-962,
Korea
Tel. +82-1644-5866, Fax. +82-33-746-1545
Webpage: http://www.goodpl.kr
Email address: kckang@goodpl.kr

17 of 15.