Interferential Stimulator
SI-500P
Portable ICT
User manual
(Version 1.0)
GOODPL INC.
#1-114, 115 Wonju Medical Instrument Industry Park, 1720-26 Taejang-dong,
Wonju-si, Gangwon-do 220-962, Korea
Tel. +82-1644-5866, Fax. +82-33-746-1545
Webpage: http://www.goodpl.kr
Email address: kckang@goodpl.kr
1 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
1 2 3 4 5
No Name Function
2 Connector for silicon pad Part that connect with silicon pad cable
3 Connector for suction cup Part that connect with suction cup cable
2 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
SI-500P features
Converting digital to analogue current signals makes the stimulation softer and enhances the product
durability further. 9 auto modes and 3 manual modes are supported and cup and pad can be selectively used.
5. Other functions
- SI-500P features safety functions such as the zero start function, which initially sets the treatment
intensity at zero. Excess current prevention circuits were used to prevent higher than desirable
currents from reaching the area being treated. Also a moisture prevention system was applied to
protect the suction area to enable safe treatment and protection of the device
3 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
1-INTENSITY
By turning the Encoder to the right the output intensity can be adjusted
- After making adjustments to increase the output current and no further increases occur for the
next 10 seconds, an automatic lock function kicks into prevent any further increases and only
decreases are possible. In this case, in order to increase the output current the user must first
decrease the current and then increase it
- Under normal conditions the intensity can be increased in units of 1mA. By pushing the
Encoder button fine adjustments in units of 0.2mA is possible
- Changing basic frequency (2,500Hz, 4,000Hz, 5,000Hz)
(Push the encoder button to change the basic frequency [Can only be configured before
treatment begins]
2-STOP
Stop treatment mode
4 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
5-SET
(a) Time configuration: 1 min ~ 30 min (Units of 1 min)
(b) When pressing manual mode for 2 seconds the device returns to user configure mode
- START: Configures the start frequency (1Hz~200Hz)
- END: Configures the end frequency (1Hz ~ 200Hz)
- ro: Round time (sec) from the start frequency to end frequency (20, 30, 40, 50, 60 sec)
- ti: Repeat time from the start frequency to end frequency (20, 30, 40, 50, 60 sec)
- ue(VECTOR): Interferential low current application time (CON, V1, V2, V3)
- nn(MODULATION): Amplitude between the fixed current and interferential current (0%, 25%,
50%, 75%, 100%)
- Press the ‘SET’ button when moving on to the next level through Encoder configuration
6-WHITE
WHITE electrodes can be separately configured.
7-RED
RED electrodes can be separately configured.
8-TIME
Time configuration: 1 min ~ 30 min (Units of 1 min)
9-MODE
12-F/A ,MODE : when use intensity function will displayed on the panel
Purpose
The purpose of using the GP-500P can be largely classified as nerve suppression, muscle
treatment, promotion of blood circulation, massaging, and fat cell reduction. A more in-depth
classification is shown as the following;
5 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
Items of caution
The following users should consult a physician before using this device
6 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
1. How to install
① Check the power supply (AC230V/50Hz)
② Connect the catheter to the device by color
(The color of the catheter cup may differ according to the product)
③ Soak the sponge, which is an accessory, and squeeze the sponge for adequate dampness
2. Items of caution
- Please designate a manager responsible for the device and do not have anyone other than
this person operate the device. Also, please thoroughly read through the following items of
caution to ensure safe operation of this device
① Items of caution during installation
(Please take note of the appropriate location to install this device)
- Install in areas that are not wet
- Install in areas unexposed to dust and are well ventilated
- Install in areas that can supply the appropriate voltage, frequency, and consumed power
- Install in areas that are not affected by mechanical vibrations or electric abnormalities
- Install in areas that are not exposed to gases produced by chemical substances
- Check to see if the power cord is appropriately connected to the socket
- Install in areas that are not affected by extreme temperatures, humidity, salinity, or ions
- Areas that are exposed to electromagnetic pulses may cause the devise to function
abnormally
- The device may generate unstable output if the device is used in close proximity (1m) to
shortwave or microwave devices
7 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
⑥ Operational taboos(Please take extra caution before using this device for the
following patients)
- Do not use this device for patients with pacemakers or other implanted electronic devices
without the consent of a physician
- The use of a high frequency stimulator in conjunction with this device may result in burns
near the stimulator electrode and damages to the electrode
- Do not use this device for patients deemed by a physician as being too physically weak
⑦ Other items
- Do not perform unauthorized alterations or repairs to this device
- Please contact our headquarters or a nearby sales representative for any problems with the
device
- Perform regular inspections of the device to maintain the device in optimal conditions
8 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
③ Select Time(1~30minutes)
<Cup> <Pad>
9 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
⑦ Configure the suction cycle of the suction catheter by pressing the PULS button (OFF:CON,
P1:15/Min, P2:20/Min, P3:30/Min)
- The suction cycle will begin 10 seconds after treatment begins
⑧Turn the intensity volume and adjust the output intensity according to the patient’s condition
⑨ If the Red/White balance is wrong, turn the appropriate button and adjust accordingly
10 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
6) AUTO6 (Chronic)
- Ion permeation to cell membranes is increased by repeating a process of exiting and relaxing
tissues to increase congestion within blood vessels that activate ion activities within cells (various
alterations).
11 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
- This treatment is an option for patients suffering from intractable edema and patients that respond
relatively less well to electric therapy.
12 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
13 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
14 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
This device requires the utmost safety precautions considering it is a device that discharges electric
currents to an injured area of the human body. If any of the following malfunctions are detected
please visit the manufacturer or the sales representative to receive assistance. For any cases in
which the device is deemed to be operating abnormally, please make sure the device is fully
repaired before further use
[Malfunctions]
Cases in which output discharge does not change even after applying the catheter and
turning the intensity knob
Cases in which the device occasionally does not operate normally
Cases in which the display window does not operate normally even though the power
switch is on
Cases in which the suction cup does not function properly on the area of treatment
Please receive A/S immediately for any cases in which the device is operating abnormally
Tel. no: +82-1644-5866 (A/S information)
Fax. no: +82-33-746-1545
Webpage: http://www.goodpl.kr
Email address: kckang@goodpl.kr
Product Specifications
No Item Specification
1 Power AC
2 Voltage 220V
3 Frequency 50/60 ㎐
15 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
9 Output wave
Level 1 OFF
Level 2 V1 150mmHg
10 Suction pressure
Level 3 V2 180mmHg
Level 4 V3 250mmHg
Level 1 CON
Level 2 P1 13times/Min
11 Suction cycle
Level 3 P2 15times/Min
Level 4 P3 20times/Min
14 Weight 8 kg
16 of 15.
Copyright 2011, GOODPL INC. All rights reserved USER MANUAL of GP-500P, Rev 1.0
Interferential Stimulator
Warranty
- This product has undergone stringent quality management and has passed testing
procedures in accordance with CE (MDD 93/42/EEC), EN ISO 13485:2003 and KGMP
standards. This warranty will cover for damages or abnormalities that occur during normal
conditions of operation
- Please submit your warranty to a GOODPL INC service center or sales representative for
after service and receive assistance regarding repairs
- If the sales representatives or place of purchase does not offer repairs please contact the
service center or hotline as shown on the bottom of the warranty
- If this warranty is sent via fax to our headquarters, we will make a phone call to confirm the
device number and register the device in our database. For all other cases, the warranty
begins on the day the device is sent out for delivery from our headquarters
GOODPL INC.
#1-114, 115 Wonju Medical Instrument Industry Park,
1720-26 Taejang 2-dong, Wonju-si, Gangwon-do 220-962,
Korea
Tel. +82-1644-5866, Fax. +82-33-746-1545
Webpage: http://www.goodpl.kr
Email address: kckang@goodpl.kr