Professional Documents
Culture Documents
Methods
Chromapedia
Volume I
2
Detection methods in flash and prep HPLC
UV detector
UV detectors are the most frequently used detectors in preparative chromatography. A UV de-
tector is a selective detector, meaning it only measures substances which absorb light of a se-
lected wavelength in the ultraviolet range (200-400nm) or visible range (400-800nm). Substanc-
es which are suitable for UV detection include compounds with a chromophoric group, such as
• aromatic ring
• two conjugated double bonds
• double bond adjacent to an atom with one electron pair
• carbonyl group
• bromine, iodine or sulfur
The UV detector measures the change in intensity of a UV light beam passing through a solution.
The absorption of the light is related to the concentration of the solution which the light beam is
traveling through. This relationship is described by the Lambert-Beer Law:
E = Extinction [dimensionless]
ε = Extinction coefficient [M–1· cm–1]
c = Solution concentration [mol/l]
d = Path length of the light beam through the solution [cm]
Every solvent has a UV absorbance cutoff wavelength. At wavelengths below this value the solvent
itself absorbs all the light. When using a UV detector, it is necessary to select a solvent which has
no significant UV absorption at the wavelength at which measurements are to be taken. Other-
wise, the signal of the substance and the solvent will overlap, resulting in incorrect fractionation.
3
Typical critical solvents used are acetone, toluene and benzene, as indicated in the following
table:
UV limit/cutoff UV limit/cutoff
Solvent Solvent
(nm) (nm)
Tetrahydrofurane 220
If the absorption spectrum of a compound is not known, it is useful to use multiple wavelengths
simultaneously or even a diode array detector (DAD), which can record the whole UV spectrum.
The DAD can also confirm purity and compound identity by showing the absorption spectrum of
each peak, eliminating post fraction thin layer chromatography (TLC).
For example, the DAD analysis of caffeine, vanillin and piperine is shown below:
Figure 1: Typical chromatogram with average UV signal Figure 2: 3D dataset of UV signals (200-400nm) of
(200- 400nm) of the three compounds caffeine, vanillin caffeine, vanille and piperine indicating the absorption
and piperine maxima
4
Scan at 6.95 min
5
Evaporative light scattering (ELS) detector
The general mechanism of ELSD involves measuring the amount of light scattered by particles of
solvent which have been dried through evaporation. The process consists of three steps: neb-
ulization, solvent evaporation and detection (see picture below). During nebulization, a nebulizer
combines a gas flow of air or nitrogen with the column or cartridge effluent to produce an aerosol
of tiny droplets. In the next step, the droplets are introduced into a drift tube where the mobile
phase evaporates and leaves behind a particulate form of the target compound. In the last step,
light strikes the dried particles which exit the drift tube, the light is scattered, and the resulting
photons are detected by a photodiode.
ELSD is described mathematically by equations which are governed by particle size. The peak
area (A) is related to the quantity of analyte in the column (m):
A = amb
Where m is the solute mass, and a and b are constants which depend on variety of factors,
such as the size of the particles, the concentration and type of the target substances, the gas
flowrate, the mobile phase flow rate, and the temperature of the drift tube.
ELS detectors are a preferred choice for purification of compounds without a chromophor-
ic group. Such compounds include carbohydrates, lipids, fats and polymers. The function of
ELS detectors remains undisturbed by mobile phase variations and gradient baseline shifting.
Sensitivity in ELSD is independent of the compound’s physical and chemical properties and is
therefore governed only by the absolute quantity of the compound. ELSD is a mass-dependent
detector, meaning a high signal indicates that a large amount of compound is eluting. As a
semi-quantitative technique, it provides valuable information on the ratio of the compounds in
the sample.
ELSD can detect nearly all compounds, except for highly volatile analytes, such as ethanol in
wine. In general, the compound of interest must be less volatile than the mobile phase. Any
modifier in the mobile phase needs to be also volatile. Since ELSD is a destructive detector,
sample amount transferred to the ELSD should be as minimal as possible.
The lower the boiling point of the mobile phase, the easier it is for the solvent to evaporate. Mobile
phases with high boiling points (e.g. DMS, DMF, toluene or water), either need to be evaporated
at high temperatures, with the risk of destroying the target substances, or nebulized into extreme-
ly tiny droplets, which allow evaporation even at room temperature.
6
Solvent Boiling point °C
N,N-Dimethylformamide 153
Toluene 111
Water 100
Propanol 97
Dipropylether 90
Dichloroethane 84
Acetonitrile 82
Cyclohexane 81
Benzene 80
Ethyl acetate 77
Ethanol 78
Hexane 69
Tetrahydrofuran 66
Methanol 64
Chloroform 62
Acetone 56
Dichloromethane 40
Table 3: Boiling points of typical solvents used in flash and prep HPLC
• Detects any compound less volatile • Only volatile mobile phases can be used
than the mobile phase • The sample must be less volatile than
• Good sensitivity the mobile phase
• Straightforward operation • ELSD is a destructive technique
• Gradient compatible – maintains stable
baselines during gradients
7
Mass spectrometry (MS) detector
The workflow involves converting molecules from the chromatography eluent to a charged or
ionized state. The mass analyzer is the component of the mass spectrometer which takes ionized
masses and separates them based on charge-to-mass ratios. The analyzer then outputs them to
the detector where they are recognized and converted to digital output.
The refractive index detector measures changes in the refraction of light caused by a medium
as it flows through a measuring cell. The detector is non-selective as it registers all substances
which flow through the cell. Refractive index detectors measure according to the following prin-
ciple:
= Difference between the refractive indices
nG = Refractive index of the dissolved sample
nL = Refractive index of the pure solvent
ni = Refractive index of the sample
c = Concentration of the sample
Since the refractive index detector is not specific, it is universally applicable. This type of detector
is however not well suited to applications with gradient elution, since the eluate is always com-
pared with the pure solvent in a reference cell. The system is therefore sensitive to changes in
the composition of the mobile phase. In addition, the detector is highly sensitive to temperature
changes.
8
Advantages of RI detection Limitations of RI detection
Fluorescence detector
When compounds with specific functional groups are excited by shorter wavelength energy, they
emit higher wavelength radiation or fluorescence. Fluorescence intensity is dictated by both the
excitation and emission wavelength, which enables the selective detection of some components
over others. Approximately 15% of all compounds have natural fluorescence. Some of these
compounds include aliphatic and alicyclic compounds with carbonyl groups and compounds
with highly conjugated double bonds. Aromatic components with conjugated pi-electrons give
off the strongest fluorescence activity.
Fluorescence detectors offer one of the highest sensitivity levels among existing detectors. The
sensitivity of fluorescence detectors is 10 to 1000 times higher compared to UV detectors.
9
Combination of UV and ELS detectors
In general, users in preparative chromatography require detectors which are universal and easy
to use, with high sensitivity, with non-destructive technology and without mobile phase effects.
However, none of the currently available detectors fulfill all these criteria as shown in the following
table.
10
Solvent % ELSD UV1 UV2
Time (min)
CH3 O O
CH3
H3C H
CH3
O
H 3C H
OH OCH3
H
H
H3C O H2N HO
Time (min)
O O O
N N H O
HO
N HO
O N
OCH3 O
11
References
Aced G. and Möckel HJ. (1991). Liquid-chromatographie Apparative, theoretische und metho-
dische Grundlagen der HPLC. Weinheim, Germany: VCH.
Hostettmann K. et al. (2010). Handbook of Strategies for the Isolation of Bioactive Natural
Products. Beijing, China: Science Press.
Young C.S. and Dolan J.W. (2003). Success with Evaporative Light-Scattering Detection.
LCGC Europe.
https://pdfs.semanticscholar.org/86fe/6906062968f3e659798669428c1a119f063a.pdf
12